ARTICLE
Direct-Access Valve Replacement
A Novel Approach for Off-Pump Valve Implantation Using Valved Stents
Christoph H. Huber, MD*,*,
Lawrence H. Cohn, MD* and
Ludwig K. von Segesser, MD
* Division of Cardiac Surgery, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts
Service de Chirurgie Cardiovasculaire, Centre Hospitalier Universitaire Vaudois, CHUV, Lausanne, Switzerland
Manuscript received November 14, 2004;
revised manuscript received March 25, 2005,
accepted April 5, 2005.
* Reprint requests and correspondence: Dr. Christoph H. Huber, Division of Cardiac Surgery, Brigham and Women's Hospital, 75 Francis Street, Boston, Massachusetts 02115
(Email: huberch{at}dr.com).
OBJECTIVES: This study validates the off-pump antegrade transventricular route for ultrasound-guided direct-access aortic valved stent implantation.
BACKGROUND: Direct-access aortic valved stent implantation offers numerous advantages over the remote-access percutaneous approach and may one day provide an alternative to surgical aortic valve replacement.
METHODS: Valved stents were implanted off-pump in 12 pigs (68.5.0 ± 7.3 kg) via direct-access transapical approach using a left-sided mini-thoracotomy and continuous ultrasonic and fluoroscopic guidance. Acute valved stent function was studied with intravascular and intracardiac ultrasound. All valved stents were tested in vitro before insertion. Macroscopic analysis was performed at necropsy.
RESULTS: In 8 of 12 pigs, valved stents were delivered to the target site over the native aortic valve leaflets without interference of coronary blood flow and with good acute valve function. Two valved stents were deployed and supra-annularly occluded the coronary orifice, leading to fatal outcome. Two valved stents dislodged into the left ventricle, one because of size mismatch and one that failed to unfold correctly.
CONCLUSIONS: Twelve pigs underwent deployment of a valved stent in the aortic position. Six valves observed for an average 4.5-h period showed satisfactory postimplantation valve function.
|
Abbreviations and Acronyms
| | AcuNav = intracardiac ultrasound | | AVR = aortic valve replacement | | CBF = coronary blood flow | | IVUS = intravascular ultrasound | | LCA = left coronary artery | | LV = left ventricle/ventricular | | RCA = right coronary artery | | RCSSI = residual coronary sinus stent index |
|
This article has been cited by other articles:

|
 |

|
 |
 
R. Quaden, T. Attmann, M. Schunke, D. Theisen-Kunde, J. Cremer, and G. Lutter
Percutaneous aortic valve replacement: Endovascular resection of human aortic valves in situ.
J. Thorac. Cardiovasc. Surg.,
May 1, 2008;
135(5):
1081 - 1086.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. Wendt, M. Thielmann, T. Buck, R.-A. Janosi, T. Bossert, N. Pizanis, M. Kamler, and H. Jakob
First clinical experience and 1-year follow-up with the sutureless 3F-Enable aortic valve prosthesis
Eur. J. Cardiothorac. Surg.,
April 1, 2008;
33(4):
542 - 547.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. K. Rosengart, T. Feldman, M. A. Borger, T. A. Vassiliades Jr, A. M. Gillinov, K. J. Hoercher, A. Vahanian, R. O. Bonow, and W. O'Neill
Percutaneous and Minimally Invasive Valve Procedures: A Scientific Statement From the American Heart Association Council on Cardiovascular Surgery and Anesthesia, Council on Clinical Cardiology, Functional Genomics and Translational Biology Interdisciplinary Working Group, and Quality of Care and Outcomes Research Interdisciplinary Working Group
Circulation,
April 1, 2008;
117(13):
1750 - 1767.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. J. Davidson and D. S. Baim
Percutaneous Aortic Valve Interventions
Card. Surg. Adult,
January 1, 2008;
3(2008):
963 - 971.
[Full Text]
|
 |
|

|
 |

|
 |
 
E. M. Flecher, J. W. Curry, T. M. Joudinaud, C. L. Kegel, P. A. Weber, and C. M.G. Duran
Coronary flow obstruction in percutaneous aortic valve replacement. An in vitro study
Eur. J. Cardiothorac. Surg.,
August 1, 2007;
32(2):
291 - 294.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. H. Huber, B. Marty, and L. K. von Segesser
Acceptance and introduction of disruptive technologies simple steps to build a fully functional pulmonary valved stent
Interactive CardioVascular and Thoracic Surgery,
August 1, 2007;
6(4):
430 - 432.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Walther, V. Falk, T. Dewey, J. Kempfert, F. Emrich, B. Pfannmuller, P. Broske, M. A. Borger, G. Schuler, M. Mack, et al.
Valve-in-a-Valve Concept for Transcatheter Minimally Invasive Repeat Xenograft Implantation
J. Am. Coll. Cardiol.,
July 3, 2007;
50(1):
56 - 60.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Walther and F. W. Mohr
Aortic valve surgery: time to be open-minded and to rethink
Eur. J. Cardiothorac. Surg.,
January 1, 2007;
31(1):
4 - 6.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Walther, V. Falk, M. A. Borger, T. Dewey, G. Wimmer-Greinecker, G. Schuler, M. Mack, and F. W. Mohr
Minimally invasive transapical beating heart aortic valve implantation -- proof of concept
Eur. J. Cardiothorac. Surg.,
January 1, 2007;
31(1):
9 - 15.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. M. Dewey, T. Walther, M. Doss, D. Brown, W. H. Ryan, L. Svensson, T. Mihaljevic, R. Hambrecht, G. Schuler, G. Wimmer-Greinecker, et al.
Transapical aortic valve implantation: an animal feasibility study.
Ann. Thorac. Surg.,
July 1, 2006;
82(1):
110 - 116.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Walther, T. Dewey, G. Wimmer-Greinecker, M. Doss, R. Hambrecht, G. Schuler, F. W. Mohr, and M. Mack
Transapical approach for sutureless stent-fixed aortic valve implantation: experimental results.
Eur. J. Cardiothorac. Surg.,
May 1, 2006;
29(5):
703 - 708.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. H. Huber and L. K. von Segesser
Direct Access Valve Replacement (DAVR) -- are we entering a new era in cardiac surgery?
Eur. J. Cardiothorac. Surg.,
March 1, 2006;
29(3):
380 - 385.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|