Advertisement

Click here for more guidelines.

 
 




CME Topic Collections Past Issues Search Current Issue Home
     

J Am Coll Cardiol, 2010; 55:543-554, doi:10.1016/j.jacc.2009.08.067
© 2010 by the American College of Cardiology Foundation
This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Online Appendix
Right arrow View Related Cardiosource Journal Scan
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (2)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Leon, M. B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Leon, M. B.
Related Collections
Right arrowRelated Articles

CLINICAL RESEARCH: INTERVENTIONAL CARDIOLOGY

A Randomized Comparison of the Endeavor Zotarolimus-Eluting Stent Versus the TAXUS Paclitaxel-Eluting Stent in De Novo Native Coronary Lesions

12-Month Outcomes From the ENDEAVOR IV Trial

Martin B. Leon, MD*,*, Laura Mauri, MD, MSc{ddagger}, Jeffrey J. Popma, MD§, Donald E. Cutlip, MD||, Eugenia Nikolsky, MD, PhD*, Charles O'Shaughnessy, MD, Paul A. Overlie, MD#, Brent T. McLaurin, MD{ddagger}{ddagger}, Stuart L. Solomon, MD**, John S. Douglas, Jr, MD§§, Michael W. Ball, MD||||, Ronald P. Caputo, MD{dagger}, Ash Jain, MD¶¶, Thaddeus R. Tolleson, MD{dagger}{dagger}, Bernard M. Reen, III, MD{ddagger}{ddagger}{ddagger}, Ajay J. Kirtane, MD*, Peter J. Fitzgerald, MD, PhD##, Kweli Thompson, MD***, David E. Kandzari, MD{dagger}{dagger}{dagger} for the ENDEAVOR IV Investigators

* Columbia University Medical Center and the Cardiovascular Research Foundation, New York, New York
{dagger} St. Joseph Hospital, Syracuse, New York
{ddagger} Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts
§ Caritas Cardiovascular Center, St. Elizabeth's Medical Center, Boston, Massachusetts
|| Beth Israel Deaconess Medical Center, Boston, Massachusetts
Elyria Memorial Hospital, Elyria, Ohio
# Lubbock Heart Hospital, Lubbock, Texas
** Methodist Hospital, Houston, Texas
{dagger}{dagger} Mother Frances Health System, Tyler, Texas
{ddagger}{ddagger} Anderson Medical Center, Anderson, South Carolina
§§ Emory University, Atlanta, Georgia
|||| Heart Center of Indiana, Indianapolis, Indiana
¶¶ Washington Hospital, Fremont, California
## Stanford University, Stanford, California
*** Medtronic CardioVascular, Santa Rosa, California
{dagger}{dagger}{dagger} Scripps Clinic, La Jolla, California
{ddagger}{ddagger}{ddagger} Presbyterian Hospital, Charlotte, North Carolina

Manuscript received May 19, 2009; revised manuscript received August 26, 2009, accepted August 30, 2009.

* Reprint requests and correspondence: Dr. Martin B. Leon, Columbia University College of Physicians and Surgeons, Center for Interventional Vascular Therapy, New York–Presbyterian Hospital/Columbia University Medical Center, New York, New York 10032 (Email: mleon{at}crf.org).


    Abstract
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 Appendix
 References
 
Objectives: The ENDEAVOR IV (Randomized Comparison of Zotarolimus-Eluting and Paclitaxel-Eluting Stents in Patients with Coronary Artery Disease) trial evaluated the safety and efficacy of the zotarolimus-eluting stent (ZES) compared with the paclitaxel-eluting stent (PES).

Background: First-generation drug-eluting stents have reduced angiographic and clinical restenosis, but long-term safety remains controversial. A second-generation drug-eluting stent, which delivers zotarolimus, a potent antiproliferative agent, via a biocompatible phosphorylcholine polymer on a cobalt alloy thin-strut stent has shown promising experimental and early clinical results.

Methods: This is a prospective, randomized (1:1), single-blind, controlled trial comparing outcomes of patients with single de novo coronary lesions treated with ZES or PES. The primary end point was noninferiority of 9-month target vessel failure defined as cardiac death, myocardial infarction, or target vessel revascularization.

Results: Among a total of 1,548 patients assigned to ZES (n = 773) or PES (n = 775), at 9 months, ZES was noninferior to PES with rates of target vessel failure 6.6% versus 7.1%, respectively (pnoninferiority ≤ 0.001). There were fewer periprocedural myocardial infarctions with ZES (0.5% vs. 2.2%; p = 0.007), whereas at 12 months, there were no significant differences between groups in rates of cardiac death, myocardial infarction, target vessel revascularization, or stent thrombosis. Although incidence of 8-month binary angiographic in-segment restenosis was higher in patients treated with ZES versus PES (15.3% vs. 10.4%; p = 0.284), rates of 12-month target lesion revascularization were similar (4.5% vs. 3.2%; p = 0.228), especially in patients without planned angiographic follow-up (3.6% vs. 3.2%; p = 0.756).

Conclusions: These findings demonstrate that ZES has similar clinical safety and efficacy compared with PES in simple and medium complexity single de novo coronary lesions. (ENDEAVOR IV Clinical Trial; NCT00217269)

Key Words: drug-eluting stents • target lesion revascularization • zotarolimus-eluting stent

Abbreviations and Acronyms
  ARC = Academic Research Consortium
  DAPT = dual antiplatelet therapy
  DES = drug-eluting stent(s)
  DS = diameter stenosis
  IVUS = intravascular ultrasound
  MI = myocardial infarction
  PES = paclitaxel-eluting stent(s)
  SES = sirolimus-eluting stent(s)
  TLR = target lesion revascularization
  TVF = target vessel failure
  TVR = target vessel revascularization
  ZES = zotarolimus-eluting stent(s)


Since the introduction of drug-eluting stents (DES) in the U.S. more than 5 years ago, angiographic restenosis and repeat revascularization procedures after percutaneous coronary interventions have been significantly reduced for both simple ("on-label") and complex ("off-label") lesions (1–5). Safety concerns, however, have been raised for both sirolimus-eluting stents (SES) (Cordis Corporation, Warren, New Jersey) and paclitaxel-eluting stents (PES) (Boston Scientific, Natick, Massachusetts), focusing on a small increase in stent thrombosis occurring longer than 1 year after the index procedure (6–8). Consequently, treatment practices have shifted, resulting in a more selective use of DES, extended dual antiplatelet therapy (DAPT) regimens (to 1 year or longer), and concerns about premature cessation of DAPT due to either patient noncompliance or intervening medical problems (9,10).

The design features of an "optimally safe" DES remain elusive, although most observers agree that impaired early and late healing responses, perhaps associated with certain biostable polymers, may be partially causative (11,12). The zotarolimus-eluting stent (ZES) (Medtronic CardioVascular, Santa Rosa, California) combines a more rapid elution profile of the antiproliferative drug zotarolimus with a thinner, more biocompatible phosphorylcholine polymer placed on a cobalt alloy thin-strut stent. ZES have been clinically tested in a first-human-use study with extended follow-up (13,14), a double-blind randomized trial versus bare-metal stents (15), and a small randomized trial versus SES (16). These studies have indicated that ZES, when compared with bare-metal stents (15), had an advantageous safety profile and reduced target lesion revascularization (TLR) despite a somewhat higher angiographic late loss than observed with SES and PES (1–3,16). The purpose of the current trial was to compare ZES with commonly used PES in a large randomized trial, emphasizing clinical end points, to limit the impact of systematic angiographic follow-up on repeat revascularization decisions.


    Methods
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 Appendix
 References
 
Study design and patient population.   The ENDEAVOR IV (Randomized Comparison of Zotarolimus-Eluting and Paclitaxel-Eluting Stents in Patients with Coronary Artery Disease) trial was a prospective, multicenter, single-blinded, randomized, controlled clinical trial that compared clinical and angiographic outcomes between patients treated with ZES and patients treated with PES. Consecutive adult patients with clinical evidence of ischemic coronary disease or a positive functional study were enrolled at 80 centers in the U.S.

Key clinical exclusion criteria included recent acute myocardial infarction (MI), another planned percutaneous coronary intervention within the next 30 days or previous percutaneous coronary intervention in the target vessel within the previous 9 months, recent stroke or transient ischemic attack, left ventricular ejection fraction less than 30%, and contraindication to DAPT (aspirin and a thienopyridine).

Angiographic requirements were the presence of a single de novo native coronary lesion with a diameter stenosis (DS) of at least 50% but <100% by visual estimate, reference vessel diameter ≥2.5 and ≤3.5 mm, and lesion length ≤27 mm. A target vessel with evidence of thrombus or excessive tortuosity or a target lesion that was in a left main or an ostial location, or with severe calcification, or at a bifurcation involving a side branch >2.0 mm in diameter were excluded.

The institutional review board at each site approved the protocol, and each eligible patient provided written, informed consent before the index procedure.

Study device.   The Endeavor ZES (Medtronic) consists of a cobalt-based alloy stent with a phosphorylcholine coating that releases the drug zotarolimus. The synthetic phosphorylcholine drug carrier is composed of an outer phospholipid portion that mimics the outer membrane of red blood cells to provide biocompatibility and an inner hydrophobic region to confer stability and adhesion to the stent surface (17). A confluent coating of phosphorylcholine on bare-metal stents reduces thrombogenicity in in vitro models (18). Zotarolimus (ABT-578) is a tetrazole ring containing macrocyclic lactone analogue of sirolimus and is the first cytostatic agent developed exclusively for delivery from a DES to prevent restenosis (19). The phosphorylcholine coating thickness is approximately 4 µm and the concentration of zotarolimus is 10 µg/mm stent length. The ZES formulation results in rapid drug elution from the stent surface (~80% in 1 week and ~95% in 2 weeks). In animal studies, ZES were associated with less platelet adhesion, improved early healing responses compared with other first-generation DES, and reduced neointimal area (20,21).

The TAXUS Express PES (Boston Scientific) consists of a 316L surgical grade stainless steel Express stent, a Translute (Boston Scientific) polymer carrier, and paclitaxel loaded in a concentration of 1 µg/mm2 in a slow release formulation.

Randomization, implantation procedure, adjunct pharmacology, and follow-up.   Patients were stratified by diabetic status and clinical site and were randomized (using an Interactive Voice Response System) in a 1:1 manner to receive either ZES or PES. Diabetes was defined as treatment for diabetes mellitus with insulin, oral antidiabetic agents, or a modified diet.

Target lesions were pre-treated using standard balloon angioplasty, followed by stent implantation and post-dilation (as needed) to achieve a final lumen DS of <10%. Patients were treated with 75 mg of aspirin within 24 h before the procedure, which was continued indefinitely. All patients received a loading dose of clopidogrel of at least 300 mg followed by 75 mg/day for at least 6 months. Continuation of clopidogrel beyond 6 months was at the operator's discretion. During the procedure, bivalirudin or unfractionated heparin (to maintain an activated clotting time >250 s) was administered. Platelet glycoprotein IIb/IIIa receptor inhibitors were used at the discretion of the operator.

Following the interventional procedure, patients were assessed at 30 days; 6, 9, and 12 months; and yearly thereafter up to 5 years after the procedure. Patients in the angiographic and intravascular ultrasound (IVUS) subgroups were evaluated at 8 months after the procedure.

Data management and analysis.   An independent data management organization (Harvard Clinical Research Institute, Boston, Massachusetts) collected and managed all data analyses. All authors had full access to the database and analysis upon which this manuscript is based. An independent clinical events committee, blinded to the study stent identity, adjudicated all primary and secondary end point events. An independent data safety monitoring board, composed of 4 independent physicians and 1 biostatistician blinded to study treatment, was responsible for regular review of the clinical safety data and could recommend study discontinuation or modification.

End points and definitions.   The primary clinical end point was target vessel failure (TVF), defined as the composite of cardiac death, MI, or clinically driven target vessel revascularization (TVR) at 9 months after the procedure. Secondary clinical end points were acute device, lesion, and procedure success and major adverse cardiac event, defined as death, MI, or clinically driven TLR. Secondary angiographic end points were in-segment and in-stent late lumen loss and binary restenosis. The IVUS end points at follow-up were neointima hyperplasia volume, percent volume obstruction, stent strut apposition to the vessel wall, and positive vessel remodeling.

Stent thrombosis as defined by the clinical protocol required 1 of the following: 1) coronary symptoms and target vessel angiographic confirmation of thrombus or occlusion; 2) pathologic confirmation of acute thrombosis in the target vessel; 3) unexplained death within 30 days; or 4) target vessel MI without angiographic confirmation of thrombosis or other identified culprit lesions within 30 days. Patients with intervening TLR events were excluded and protocol-defined stent thrombosis was assessed within 24 h (acute), at 1 to 30 days (subacute), and after 30 days following stent placement (late). Stent thrombosis (both early and late) was also adjudicated by a subcommittee of the clinical events committee blinded to study stent identity and according to the definitions proposed by the Academic Research Consortium (ARC) (22).

Angiography and IVUS core laboratories.   Coronary angiograms were obtained at baseline in all patients, and at 8 months after the procedure in the first 328 consecutive patients enrolled and assigned to routine angiographic follow-up and in any patient who received multiple stents. All angiograms were evaluated by reviewers blinded to the study stent, by an independent core laboratory (Brigham and Women's Angiographic Core Laboratory, Boston, Massachusetts).

All patients in the angiographic subgroup underwent IVUS imaging at 8 months, which was evaluated by reviewers blinded to the study stent at an independent IVUS core laboratory (Cardiovascular Core Analysis Laboratory, Stanford University, Palo Alto, California).

Statistical methods.   This randomized study was designed to determine if the primary end point (TVF at 9 months) of ZES was noninferior to PES. The sample size estimation was based on a noninferiority test using the Farrington-Manning approach at alpha = 0.05 (1-sided) significance level with 84% statistical power to reject the null hypothesis of inferiority (23). Assuming an expected TVF rate for ZES and PES to be 7.6%, a noninferiority margin of 3.8% with at least 90% clinical follow-up and 1:1 randomization, a total sample size of 1,548 enrolled patients was required.

In-segment late lumen loss at 8 months comparing ZES and PES was a secondary powered noninferiority angiographic end point. The estimated sample size for the angiographic subset was based on a 2-sample t test at alpha = 0.05 (1-sided) significance level with 90% statistical power to reject the null hypothesis of inferiority. Assuming a true equivalence of the means between the 2 groups, a noninferiority margin of 0.2 mm, with a common SD of 0.55 mm, 80% angiographic follow-up rate, and 1:1 randomization, a total sample size of 328 patients was required for the angiographic substudy.

For other secondary clinical and angiographic outcomes, the 2-sample t test (2-sided) was used to compare continuous variables, and Fisher exact test was used to compare binary variables. Kaplan-Meier cumulative incidence estimates were used to analyze outcome events, which were compared between treatment groups using the log-rank test.

To identify the predictors of TLR and TVF at 12 months, a stepwise multivariable logistic regression analysis was performed for overall patients and for ZES and PES patient groups separately. The level of significance to enter and stay in the model was set at 0.05. Appropriate candidate variables were entered into the model, including assignment to angiographic follow-up and ZES versus PES for the overall patients.

To explore whether the treatment effects of TVF and TLR were consistent across important subgroups, logistic regression analysis was performed to test the interaction among the treatment assignment and the subsets.

All analyses were performed with SAS software (version 8.2 or higher, SAS Institute, Cary, North Carolina).


    Results
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 Appendix
 References
 
Patient enrollment and baseline characteristics.   Between April 11, 2005, and June 27, 2006, a total of 1,548 patients were enrolled at 80 clinical sites in the U.S. (ZES: n = 773, PES: n = 775) (Fig. 1). Clinical follow-up at 30 days was obtained in 770 ZES and 773 PES patients (99.6% and 99.7%, respectively). Clinical follow-up for ZES patients was 98.3% at 9 months and 97.5% at 12 months, and for PES patients was 97.5% at 9 months and 96.9% at 12 months. Angiographic follow-up at 8 months was obtained in 144 ZES patients (87.8% of eligible) and in 135 PES patients (82.3% of eligible).


Figure 1
View larger version (14K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 1 Patient Disposition and Follow-Up

Randomization, angiographic follow-up, and clinical follow-up. mo = month; QCA = quantitative coronary angiography.

 
Baseline clinical characteristics for ZES and PES cohorts were well-balanced (Table 1). For the overall study population, mean age was 63.5 years, 67.7% were male, and 30.8% were diabetic. Approximately one-half of the patients had stable angina, unstable angina, or Canadian Cardiovascular Society class III/IV angina.


View this table:
[in this window]
[in a new window]

 
Table 1 Baseline Clinical and Angiographic Characteristics of the Study Population
 
Target vessel location was in the left anterior descending coronary in 41.8% of patients and 56.1% had single-vessel disease (Table 1). Mean reference vessel diameter for all patients was 2.72 mm, mean lesion length was 13.60 mm, and pre-procedure percent DS was 65.3%.

Procedural results and angiographic outcomes.   Although stent length and stent diameter were similar between ZES and PES patients, the numbers of stents per lesion and maximum inflation pressures were lower in ZES patients (Table 2). Post-procedural minimal lumen diameter, percent DS, and acute gain were similar among ZES and PES treatment arms (Table 2). Procedure success was significantly higher for ZES when compared with PES (98.7% vs. 96.8%, p = 0.015).


View this table:
[in this window]
[in a new window]

 
Table 2 Procedural Characteristics and Angiographic Results
 
In those patients assigned to angiographic follow up at 8 months, there were statistically significant differences in minimal lumen diameter, percent DS, and late lumen loss (both in-stent and in-segment) favoring PES (Table 3). The powered secondary end point, in-segment late lumen loss, was 0.36 mm for ZES and 0.23 mm for PES (pnoninferiority = 0.089). Binary angiographic in-stent restenosis was 13.3% for ZES and 6.7% for PES (p = 0.075), and in-segment restenosis was 15.3% for ZES and 10.4% for PES (p = 0.284) (Table 3).


View this table:
[in this window]
[in a new window]

 
Table 3 Angiographic Outcomes at 8 Months
 
IVUS results.   Quantitative IVUS imaging was performed at 8 months follow-up in 164 ZES patients and 164 PES patients. Neointimal volume was 24.14 mm3 for ZES and 14.88 mm3 for PES (p = 0.002) and percent volume obstruction was 15.72 for ZES and 9.88 for PES (p < 0.001). Incomplete stent strut apposition immediately after the procedure was similar in both groups and late acquired incomplete apposition was present in 1 ZES patient (0.9%) and 3 PES patients (3.2%; p = 0.346) (Table 4).


View this table:
[in this window]
[in a new window]

 
Table 4 IVUS Findings at 8 Months
 
30-day clinical outcomes.   At 30 days, there were fewer non–Q-wave MIs in patients receiving ZES versus patients receiving PES (0.5% vs. 2.2%, p = 0.007) (Table 5). Similarly, ZES patients had reduced rates of cardiac death or MI, TVF, and major adverse cardiac events (p = 0.042, p = 0.010, and p = 0.019, respectively). There were no significant differences between treatment arms for death, cardiac death, and stent thrombosis (protocol definition and ARC definition) (Table 5).


View this table:
[in this window]
[in a new window]

 
Table 5 Clinical Outcomes at 30 Days
 
9- and 12-month clinical outcomes.   The primary end point, TVF at 9 months, was similar when comparing ZES and PES patients (6.6% for ZES and 7.1% for PES, pnoninferiority < 0.001) (Table 6, Fig. 2A). At 9 and 12 months, there were trends showing fewer non–Q-wave MIs with ZES compared with PES (p = 0.118 and p = 0.095, respectively) (Table 6). There were no significant differences at 9 or 12 months between treatment arms for death, cardiac death, any MI, and major adverse cardiac events (Table 6, Fig. 2B).


View this table:
[in this window]
[in a new window]

 
Table 6 Clinical Outcomes at 9 and 12 Months
 

Figure 2
View larger version (15K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 2 Cumulative Incidence of Target Vessel Failure, Cardiac Death and Myocardial Infarction, and TVR at 360 Days (ZES vs. PES)

Time-to-event curves through 1 year for target vessel failure (A), cardiac death or myocardial infarction (MI) (B), and target vessel revascularization (TVR) (C) in patients treated with zotarolimus-eluting stent (ZES) versus paclitaxel-eluting stent (PES). Event rates represent Kaplan-Meier estimates. The p values are based on the log-rank test. CI = confidence interval; HR = hazard ratio.

 
Although there were no significant differences in stent thrombosis among the treatment cohorts, there was a trend toward more frequent protocol definition and ARC definition (definite or probable) stent thrombosis with ZES compared with PES at 9- and 12-month follow-up (Table 6). Of the 7 ZES patients with ARC stent thrombosis (definite or probable), 3 occurred before 30 days (2 definite and 1 probable) and 4 occurred between 30 days and 6 months (3 definite and 1 probable). In the 3 ZES patients with ARC definite or probable stent thrombosis before 30 days, 2 were associated with edge dissection and incomplete stent expansion and 1 had an unplanned surgical procedure. In the 4 ZES patients with ARC definite or probable stent thrombosis between 30 days and 6 months, 3 patients were no longer taking DAPT (time of DAPT cessation to stent thrombosis was 2 days, 20 days, and 2 months), 1 of which also had an unplanned surgical procedure.

There were no differences in overall clinically driven TLR or TVR comparing ZES with PES at either 9- or 12-month follow-up (Table 6, Fig. 2C). In those patients with angiographic follow-up (18% of the total study population), TLR at 12 months was 8.5% for ZES and 3.0% for PES (p = 0.071) (Fig. 3). In those patients without angiographic follow-up (82% of the study population), TLR was 3.6% for ZES and 3.2% for PES (p = 0.756) (Fig. 3).


Figure 3
View larger version (15K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 3 TLR at 12 Months for Angiographic and Nonangiographic Follow-Up Subgroups

Rates of target lesion revascularization (TLR) at 12 months in patients with and without angiographic follow-up treated with ZES versus PES. Abbreviations as in Figure 2.

 
Subgroup analysis and predictors of TLR.   Analyses of clinical subgroups emphasizing high restenosis-risk patients revealed no significant differences (nonsignificant interaction p values) between ZES and PES at 1-year clinical follow-up for either TVF or TVR (Figs. 4A and 4B).


Figure 4
View larger version (21K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 4 Subgroup Analyses of the 1-Year Rates of TVF and TVR (ZES vs. PES)

Subgroup analyses of the 1-year target vessel failure (TVF) (A) and TVR (B) among the patients randomized to receive a ZES versus PES. Probability for interaction represents the likelihood for interaction between the variable and the relative treatment effect. LAD = left anterior descending artery; other abbreviations as in Figure 2.

 
Multivariate predictors of TLR at 12 months for all patients were history of diabetes, left anterior descending target vessel and the presence of multiple stents (Table 7). A history of diabetes, multiple stents, and age were significant predictors of TLR for the PES group. The only significant predictor of TLR for patients in the ZES group was assignment to angiographic follow-up.


View this table:
[in this window]
[in a new window]

 
Table 7 Multivariate Predictors of TLR and TVF at 12 Months
 

    Discussion
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 Appendix
 References
 
The main findings from the randomized ENDEAVOR IV study comparing ZES versus PES in patients with simple and medium complexity single de novo coronary lesions are:
1 The primary clinical end point, 9-month TVF, was similar in ZES and PES patients.
2 The powered secondary end point was not met in this study with in-segment late lumen loss being significantly higher in ZES versus PES patients.
3 Despite the less robust reduction in intimal hyperplasia seen on angiography with ZES versus PES, the clinical expression of restenosis (TLR and TVR) were similar among the 2 study cohorts after 9- and 12-month follow-up.

The primary clinical end point, 9-month TVF, which includes both safety and efficacy measures, was very close in ZES and PES patients (6.6% and 7.1%, respectively, pnoninferiority < 0.001). Individual safety end points such as cardiac death and MI were also similar in the 2 groups, though non–Q-wave MIs were still more frequent in PES patients (19 vs. 10 events; p = 0.095), undoubtedly a reflection of the earlier increase in periprocedural non–Q-wave MIs after PES treatment. Similar reduction in periprocedural non–Q-wave MIs was observed when ZES was compared with SES in a smaller randomized trial (from 3.5% with SES to 0.6% with ZES, p = 0.042) (16). Importantly, in ENDEAVOR IV, of the 17 periprocedural non–Q-wave MIs after PES, 8 (47%) were large MIs with >10x upper limit of normal creatine kinase-myocardial band rises. Other studies have indicated a higher than expected frequency of periprocedural MIs associated with PES implantation with a reported incidence of 3.5% to 5.4% (2,3) and post hoc angiographic analyses have suggested that disproportionate changes in PES covered side branches (both occlusion and/or reduced flow) may be partially responsible for these early clinical events (3,24).

During the 12-month follow-up, there was an increase in both protocol-defined and ARC-defined stent thrombosis after ZES therapy, which although not significant, requires further comment. Early stent thrombosis (within 30 days of the procedure) for both bare-metal stents and DES is usually influenced by anatomic, procedural, and clinical factors (25). Of the 7 ARC definite or probable ZES thrombosis events, 3 occurred within 30 days and all were associated procedural mishaps (edge dissection or incomplete stent expansion) or clinical comorbid situations (unplanned surgery). The remaining 4 ZES thrombosis events occurred between 1 and 6 months, and in 3 patients, clopidogrel therapy was discontinued before the event, and 1 of these patients also had unplanned surgery immediately before the event. Study-to-study variations in stent thrombosis within the first year of follow-up are commonly observed among DES trials and the frequency of stent thrombosis for ZES in this trial was greater than reported in previous studies (13,15,16,26,27), whereas the stent thrombosis rates for PES in this trial were lower than previously reported (2,3,28). Such variations in the incidence of stent thrombosis during the first year of follow-up are exemplified in the reports from 2 recent randomized Danish trials with similar study designs comparing the safety and efficacy of SES versus PES (SORT OUT II [Comparison of Paclitaxel- and Sirolimus-Eluting Stents in Everyday Clinical Practice] trial) (29) or ZES (SORT OUT III [Comparison of Zotarolimus-Eluting Stents and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease] trial) (30) in a real-world study population. The rates of definite stent thrombosis at 9 months in patients randomized to SES were unusually low in SORT OUT III (0.2%) and considerably higher in SORT OUT II (1.7%), despite similar methodology and patient populations. Importantly, DES safety has usually been associated with increased stent thrombosis events occurring longer than 1 year after the index procedure. Therefore, long-term patient follow-up from the ENDEAVOR IV trial is required to determine if the frequency of very late stent thrombosis is impacted by ZES therapy. Moreover, the ongoing PROTECT (Patient Related OuTcomes with Endeavor versus Cypher stenting Trial), the largest randomized trial focusing on the comparative safety of DES (n = 8,800) will assess rates of stent thrombosis in patients assigned to SES versus ZES after 3 years of clinical follow-up (31).

Angiographic follow-up, performed at 8 months after the index procedure in 18% of the patients indicated significantly lower in-segment minimum lumen diameter, a higher in-segment percent diameter stenosis, and a higher in-segment late loss with ZES compared with PES. The angiographic follow-up findings after ZES in this study were comparable to multiple other ZES clinical trials, which have shown a consistent in-stent late loss of approximately 0.60 mm (13,15,16,26). Although in-segment binary restenosis was higher with ZES than with PES (15.3% vs. 10.4%), the difference was not statistically significant (p = 0.284). The rates of in-segment restenosis in the PES arm in this study are somewhat higher than in the PES arm in the pivotal TAXUS IV trial (8.6%) and lower than in PES arm in the TAXUS V trial (18.9%), which likely corresponds to the different complexity of treated lesions (2,3).

Despite the less robust reduction in intimal hyperplasia seen on angiography with ZES versus PES, the clinical expression of restenosis (TLR and TVR) were similar among the 2 study cohorts after 9- and 12-month follow-up both in the entire trial population and in high restenosis subgroups (including diabetics, longer lesions, smaller vessels, and multiple stent cases). The most likely explanation for the disparity between ZES versus PES differences in rates of angiographic and clinical restenosis is the remarkable influence of routine angiographic follow-up examinations on the frequency repeat revascularization events known as "oculo-stenotic reflex" (32). In this study, there was more than a 2-fold higher frequency of TLR after ZES in those patients with versus those without systematic angiographic follow-up. The higher late loss after ZES, eliciting many situations in which the visual estimate of follow-up angiographic diameter stenosis approaches or exceeds 50%, likely heightens this tendency. Further confirmation is derived from the multivariable analysis, wherein unlike PES, the only multivariable predictor of TLR after ZES was assignment to routine angiographic follow-up.

Study limitations.   As with most pivotal early stage randomized clinical trials used for Food and Drug Administration approval of a DES in the U.S. (1,2,4,17), the lesion and patient inclusion criteria for ENDEAVOR IV are somewhat restrictive. Thus, the reported findings can only be applied to the simple-to-medium complexity lesions and stable patients that were treated. This study was designed to test the noninferiority of ZES versus PES for TVF and in-segment late loss; the p values for all the other end points should be viewed with caution. Definitive conclusions regarding ZES safety, especially very late stent thrombosis, will necessitate prolonged clinical follow-up for several additional years. In addition, the trial was not specifically powered to evaluate the differences in rates of individual clinical end points including MI and TLR, and therefore these results should be viewed as hypothesis-generating rather than definitive.


    Conclusions
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 Appendix
 References
 
The ENDEAVOR IV trial should be viewed as a component of the larger comprehensive assessment of the new Endeavor ZES. Compared with the well-characterized PES, findings from this randomized trial indicate that in single de novo coronary lesions, the Endeavor ZES has improved periprocedural safety, similar 12-month clinical safety and efficacy outcomes, and despite more frequent angiographic restenosis, similar clinical repeat revascularization events.


    Appendix
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 Appendix
 References
 
For a list of principal investigators and participating institutions for ENDEAVOR IV, please see the online version of this article.


    Acknowledgments
 
The authors thank Naureen Sheikh, PhD, and Denise Jones, BSN, of Medtronic CardioVascular, for management of study clinical operations. They would also like to acknowledge the important contributions of Manuela Negoita, MD, Minglei Liu, PhD, Jane Moore, MS, of Medtronic CardioVascular, and Colleen Gilbert, PharmD, of CommGeniX, with manuscript preparation. Similarly, Leroy LeNarz, MD, was an invaluable collaborator in the design, execution, and analysis of results from the ENDEAVOR IV trial. For a full list of principal investigators and participating institutions for ENDEAVOR IV, please see the Online Appendix.


    Footnotes
 
The ENDEAVOR IV clinical trial was funded by Medtronic CardioVascular, Santa Rosa, California. Dr. Leon has served as a consultant to Volcano Corporation, Abbott Vascular, Boston Scientific, Cordis, and Medtronic, Inc. Dr. Mauri has served as a consultant to Abbott, Boston Scientific, Cordis, and Medtronic, Inc. Dr. Popma has received research grants from Cordis, Boston Scientific, Medtronic, Abbott Vascular, Biosensors, and ev3; has served as a consultant for Medtronic, Boston Scientific, Cordis, Abbott Vascular, and Lilly; and has served as a speaker for Pfizer, Bristol-Myers Squibb, Sanofi, Lilly, Boston Scientific, Medtronic, Cordis, and The Medicines Co. Dr. O'Shaughnessy has received research grants from Medtronic, Inc. and Boston Scientific, and is on the advisory board and Speakers' Bureau of Boston Scientific. Dr. Overlie has served as a clinical investigator for Abbott Vascular, Boston Scientific, Cordis, and Medtronic, Inc. Dr. Solomon is a registered speaker for the Medtronic Corporation. Dr. Douglas is a research investigator for Medtronic, Cordis, Boston Scientific, St. Jude, and Abbott. Dr. Caputo is a consultant to Boston Scientific and Cordis, and is on the Speakers' Bureau of Medtronic and Abbott. Dr. Kirtane has served as a consultant and speaker for Medtronic CardioVascular, Abbott Vascular, and Boston Scientific. Dr. Fitzgerald has served as a consultant for Abbott, Boston Scientific, Cordis, EndoTex, St. Jude Medical, Biosensor, ev3, Medtronic, Inc., GlaxoSmithKline, XTENT, ATI, Volcano Corporation, Novadaq, AorTx, CardioMind, Cytograft Tissue Engineering, FlowCardia, Cardio-Optics, Optics, CardioMind, RTI Medical, SurModics, Hospira, and CatherosMed. Dr. Thompson is a full-time employee and stockholder of Medtronic CardioVascular. Dr. Kandzari receives research/grant support and consulting honoraria from Medtronic CardioVascular, Inc. and Cordis Corporation.


    References
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 Appendix
 References
 
1. Moses JW, Leon MB, Popma JJ, et al. SIRIUS Investigators Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery N Engl J Med 2003;349:1315-1323.[CrossRef][Web of Science][Medline]

2. Stone GW, Ellis SG, Cox DA, et al. TAXUS-IV Investigators A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease N Engl J Med 2004;350:221-231.[CrossRef][Web of Science][Medline]

3. Stone GW, Ellis SG, Cannon L, et al. TAXUS V Investigators Comparison of a polymer-based paclitaxel-eluting stent with a bare metal stent in patients with complex coronary artery disease: a randomized controlled trial JAMA 2005;294:1215-1223.[Abstract/Free Full Text]

4. Stone GW, Midei M, Newman W, et al. SPIRIT III Investigators Comparison of an everolimus-eluting stent and a paclitaxel-eluting stent in patients with coronary artery disease. A randomized trial. JAMA 2008;299:1903-1913.[Abstract/Free Full Text]

5. Marroquin OC, Selzer F, Mulukutla SR, et al. A comparison of bare-metal and drug-eluting stents for off-label indications N Engl J Med 2008;358:342-352.[CrossRef][Medline]

6. Camenzind E, Steg PG, Wijns W. Stent thrombosis late after implantation of first-generation drug-eluting stents: a cause for concern Circulation 2007;115:1440-1445.[Free Full Text]

7. Daemen J, Wenaweser P, Tsuchida K, et al. Early and late coronary stent thrombosis of sirolimus-eluting and paclitaxel-eluting stents in routine clinical practice: data from a large two-institutional cohort study Lancet 2007;369:667-668.[CrossRef][Medline]

8. Lagerqvist B, James SK, Stenestrand U, Lindback J, Nilsson T, Wallentin L. Long-term outcomes with drug-eluting stents versus bare-metal stents in Sweden N Engl J Med 2007;356:1009-1019.[CrossRef][Medline]

9. Grines CL, Bonow RO, Casey Jr. DE, et al. Prevention of premature discontinuation of dual antiplatelet therapy in patients with coronary artery stents: a science advisory from the American Heart Association, American College of Cardiology, Society for Cardiovascular Angiography and Interventions, American College of Surgeons, and American Dental J Am Coll Cardiol 2007;49:734-739.[Abstract/Free Full Text]

10. Hodgson JM, Stone GW, Lincoff AM, et al. Society for Cardiovascular Angiography and Interventions Late stent thrombosis: considerations and practical advice for the use of drug-eluting stents: a report from the Society for Cardiovascular Angiography and Interventions Drug-eluting Stent Task Force Catheter Cardiovasc Interv 2007;69:327-333.[CrossRef][Web of Science][Medline]

11. Joner M, Finn AV, Farb A, et al. Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk J Am Coll Cardiol 2006;48:193-202.[Abstract/Free Full Text]

12. Joner M, Nakazawa G, Finn AV, et al. Endothelial cell recovery between comparator polymer-based drug-eluting stents J Am Coll Cardiol 2008;52:333-342.[Abstract/Free Full Text]

13. Meredith IT, Ormiston J, Whitbourn R, et al. First-in-human study of the Endeavor ABT-578-eluting phosphorylcholine-encapsulated stent system in de novo native coronary artery lesions: Endeavor I trial EuroIntervention 2005;1:157-164.[Medline]

14. Meredith IT, Ormiston J, Whitbourn R, et al. ENDEAVOR I Investigators Four-year clinical follow-up after implantation of the endeavor zotarolimus-eluting stent: ENDEAVOR I, the first-in-human study Am J Cardiol 2007;100:56M-61M.[Web of Science][Medline]

15. Fajadet J, Wijns W, Laarman GJ, et al. ENDEAVOR II Investigators Randomized, double-blind, multicenter study of the Endeavor zotarolimus-eluting phosphorylcholine-encapsulated stent for treatment of native coronary artery lesions: clinical and angiographic results of the ENDEAVOR II trial Circulation 2006;114:798-806.[Abstract/Free Full Text]

16. Kandzari DE, Leon MB, Popma JJ, et al. ENDEAVOR III Investigators Comparison of zotarolimus-eluting and sirolimus-eluting stents in patients with native coronary artery disease: a randomized controlled trial J Am Coll Cardiol 2006;48:2440-2447.[Abstract/Free Full Text]

17. Burke SE, Kuntz RE, Schwartz LB. Zotarolimus (ABT-578) eluting stents Adv Drug Deliv Rev 2006;58:437-446.[CrossRef][Web of Science][Medline]

18. Hayward JA, Chapman D. Biomembrane surfaces as models for polymer design: the potential for haemocompatibility Biomaterials 1984;5:135-142.[CrossRef][Web of Science][Medline]

19. Baldo P, Cecco S, Giacomin E, et al. mTOR pathway and mTOR inhibitors as agents for cancer therapy Curr Cancer Drug Targets 2008;8:647-665.[CrossRef][Web of Science][Medline]

20. Whelan DM, van der Giessen WJ, Krabbendam SC, et al. Biocompatibility of phosphorylcholine coated stents in normal porcine coronary arteries Heart 2000;83:338-345.[Abstract/Free Full Text]

21. Garcia-Touchard A, Burke SE, Toner JL, Cromack K, Schwartz RS. Zotarolimus-eluting stents reduce experimental coronary artery neointimal hyperplasia after 4 weeks Eur Heart J 2006;27:988-993.[Abstract/Free Full Text]

22. Cutlip DE, Windecker S, Mehran R, et al. Academic Research Consortium Clinical endpoints in coronary stent trials: a case for standardized definitions Circulation 2007;115:2344-2351.[Abstract/Free Full Text]

23. Farrington CP, Manning G. Test statistics and sample size formulae for comparative binomial trials with null hypothesis of non-zero risk difference or non-unity relative risk Stat Med 1990;9:1447-1454.[Web of Science][Medline]

24. Popma JJ, Mauri L, Kirtane A, Cutlip D, Almonacid A, Leon MB. Frequency and clinical consequences associated with sidebranch occlusion during stent implantation using zotarolimus-eluting and paclitaxel-eluting coronary stents Circ Cardiovasc Intervent 2009;2:133-139.[CrossRef]

25. Cutlip DE, Baim DS, Ho KK, et al. Stent thrombosis in the modern era: a pooled analysis of multicenter coronary stent clinical trials Circulation 2001;103:1967-1971.[Abstract/Free Full Text]

26. Gershlick A, Kandzari DE, Leon MB, et al. ENDEAVOR Investigators Zotarolimus-eluting stents in patients with native coronary artery disease: clinical and angiographic outcomes in 1,317 patients Am J Cardiol 2007;100:45M-55M.[Web of Science][Medline]

27. Mauri L, Cutlip DE, Leon MB, et al. Long term safety results from the Endeavor program: four-year follow-up (abstr) Am J Cardiol 2008;102(Suppl 8):124i.

28. Laskey WK, Yancy CW, Maisel WH. Thrombosis in coronary drug-eluting stents: report from the meeting of the Circulatory System Medical Devices Advisory Panel of the Food and Drug Administration Center for Devices and Radiologic Health, December 7–8, 2006 Circulation 2007;115:2352-2357.[Free Full Text]

29. Galløe AM, Thuesen L, Kelbaek H, et al. SORT OUT II Investigators Comparison of paclitaxel- and sirolimus-eluting stents in everyday clinical practice: the SORT OUT II randomized trial JAMA 2008;299:409-416.[Abstract/Free Full Text]

30. Lassen JF, Rasmussen K, Galløe A, et al. SORT-OUT III: A Prospective, Randomized Comparison of Zotarolimus-eluting and Sirolimus-eluting Stents in Patients with Coronary Artery Disease Paper presented at: Transcatheter Therapeutics. Washington, DC. October 16, 2008.

31. Brugaletta S, Burzotta F, Sabaté M. Zotarolimus for the treatment of coronary artery disease: pathophysiology, DES design, clinical evaluation and future perspective Expert Opin Pharmacother 2009;10:1047-1058.[CrossRef][Web of Science][Medline]

32. Topol EJ, Nissen SE. Our preoccupation with coronary luminology: the dissociation between clinical and angiographic findings in ischemic heart disease Circulation 1995;92:2333-2342.[Abstract/Free Full Text]


Related Articles

When Can Noninferior Be Superior?: The Multidimensional Nature of Clinical Decision-Making Calls for Innovative Approaches to Clinical Trials
E. Magnus Ohman and Robert M. Califf
J. Am. Coll. Cardiol. 2010 55: 555-557. [Full Text] [PDF]

Inside This Issue
J. Am. Coll. Cardiol. 2010 55: A32. [Full Text] [PDF]



This article has been cited by other articles:


Home page
Eur Heart JHome page
G. Sarno, B. Lagerqvist, O. Frobert, J. Nilsson, G. Olivecrona, E. Omerovic, N. Saleh, D. Venetzanos, and S. James
Lower risk of stent thrombosis and restenosis with unrestricted use of 'new-generation' drug-eluting stents: a report from the nationwide Swedish Coronary Angiography and Angioplasty Registry (SCAAR)
Eur. Heart J., January 9, 2012; (2012) ehr479v1.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
D. E. Kandzari, C. S. Barker, M. B. Leon, L. Mauri, W. Wijns, J. Fajadet, and R. Mehran
Dual Antiplatelet Therapy Duration and Clinical Outcomes Following Treatment With Zotarolimus-Eluting Stents
J. Am. Coll. Cardiol. Intv., October 1, 2011; 4(10): 1119 - 1128.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
V. Klauss, P. W. Serruys, T. Pilgrim, P. Buszman, A. Linke, T. Ischinger, F. Eberli, R. Corti, W. Wijns, M.-C. Morice, et al.
2-Year Clinical Follow-Up From the Randomized Comparison of Biolimus-Eluting Stents With Biodegradable Polymer and Sirolimus-Eluting Stents With Durable Polymer in Routine Clinical Practice
J. Am. Coll. Cardiol. Intv., August 1, 2011; 4(8): 887 - 895.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
B. C. Brott
Linking Drug-Eluting Stent Kinetics and Clinical Outcomes: Insights From Optical Coherence Tomography
J. Am. Coll. Cardiol. Intv., July 1, 2011; 4(7): 786 - 788.
[Full Text] [PDF]


Home page
CirculationHome page
D. E. Cutlip
Five-Year Outcomes of Sirolimus-Eluting and Paclitaxel-Eluting Stents in Routine Clinical Practice: Are They Equally Good (or Bad)?
Circulation, June 21, 2011; 123(24): 2779 - 2781.
[Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
T. Pilgrim, L. Raber, A. Limacher, L. Loffel, P. Wenaweser, S. Cook, J.-C. Stauffer, M. Togni, R. Vogel, A. Garachemani, et al.
Comparison of Titanium-Nitride-Oxide-Coated Stents With Zotarolimus-Eluting Stents for Coronary Revascularization: A Randomized Controlled Trial
J. Am. Coll. Cardiol. Intv., June 1, 2011; 4(6): 672 - 682.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
S. R. Dixon and C. L. Grines
The Year in Interventional Cardiology
J. Am. Coll. Cardiol., May 31, 2011; 57(22): 2207 - 2220.
[Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
D. E. Kandzari, L. Mauri, J. J. Popma, M. A. Turco, P. A. Gurbel, P. J. Fitzgerald, and M. B. Leon
Late-Term Clinical Outcomes With Zotarolimus- and Sirolimus-Eluting Stents: 5-Year Follow-Up of the ENDEAVOR III (A Randomized Controlled Trial of the Medtronic Endeavor Drug [ABT-578] Eluting Coronary Stent System Versus the Cypher Sirolimus-Eluting Coronary Stent System in De Novo Native Coronary Artery Lesions)
J. Am. Coll. Cardiol. Intv., May 1, 2011; 4(5): 543 - 550.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
A. C. Yeung, M. B. Leon, A. Jain, T. R. Tolleson, D. J. Spriggs, B. T. Mc Laurin, J. J. Popma, P. J. Fitzgerald, D. E. Cutlip, J. M. Massaro, et al.
Clinical Evaluation of the Resolute Zotarolimus-Eluting Coronary Stent System in the Treatment of De Novo Lesions in Native Coronary Arteries: The RESOLUTE US Clinical Trial
J. Am. Coll. Cardiol., April 26, 2011; 57(17): 1778 - 1783.
[Abstract] [Full Text] [PDF]


Home page
Circ Cardiovasc IntervHome page
C. Briguori, F. Airoldi, G. Visconti, A. Focaccio, G. Caiazzo, B. Golia, G. Biondi-Zoccai, B. Ricciardelli, and G. Condorelli
Novel Approaches for Preventing or Limiting Events in Diabetic Patients (Naples-Diabetes) Trial: A Randomized Comparison of 3 Drug-Eluting Stents in Diabetic Patients
Circ Cardiovasc Interv, April 1, 2011; 4(2): 121 - 129.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
A. N. DeMaria, J. J. Bax, O. Ben-Yehuda, G. K. Feld, B. H. Greenberg, J. Hall, M. Hlatky, W. Y. W. Lew, J. A. C. Lima, A. S. Maisel, et al.
Highlights of the Year in JACC 2010
J. Am. Coll. Cardiol., January 25, 2011; 57(4): 480 - 514.
[Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
L. Mauri, J. M. Massaro, S. Jiang, I. Meredith, W. Wijns, J. Fajadet, D. E. Kandzari, M. B. Leon, D. E. Cutlip, and K. P. Thompson
Long-Term Clinical Outcomes With Zotarolimus-Eluting Versus Bare-Metal Coronary Stents
J. Am. Coll. Cardiol. Intv., December 1, 2010; 3(12): 1240 - 1249.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
M. B. Leon, E. Nikolsky, D. E. Cutlip, L. Mauri, H. Liberman, H. Wilson, J. Patterson, J. Moses, D. E. Kandzari, and for the ENDEAVOR IV Investigators
Improved Late Clinical Safety With Zotarolimus-Eluting Stents Compared With Paclitaxel-Eluting Stents in Patients With De Novo Coronary Lesions: 3-Year Follow-Up From the ENDEAVOR IV (Randomized Comparison of Zotarolimus- and Paclitaxel-Eluting Stents in Patients With Coronary Artery Disease) Trial
J. Am. Coll. Cardiol. Intv., October 1, 2010; 3(10): 1043 - 1050.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
J. M. de la Torre Hernandez, F. Alfonso, F. Gimeno, J. A. Diarte, R. Lopez-Palop, A. Perez de Prado, F. Rivero, J. Sanchis, M. Larman, J. F. Diaz, et al.
Thrombosis of Second-Generation Drug-Eluting Stents in Real Practice: Results From the Multicenter Spanish Registry ESTROFA-2 (Estudio Espanol Sobre Trombosis de Stents Farmacoactivos de Segunda Generacion-2)
J. Am. Coll. Cardiol. Intv., September 1, 2010; 3(9): 911 - 919.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
S. Garg and P. W. Serruys
Coronary Stents: Current Status
J. Am. Coll. Cardiol., August 31, 2010; 56(10_Suppl_S): S1 - S42.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
D. J. Kereiakes, L. A. Cannon, R. L. Feldman, J. J. Popma, R. Magorien, R. Whitbourn, I. M. Dauber, A. C. Rabinowitz, M. W. Ball, B. Bertolet, et al.
Clinical and Angiographic Outcomes After Treatment of De Novo Coronary Stenoses With a Novel Platinum Chromium Thin-Strut Stent: Primary Results of the PERSEUS (Prospective Evaluation in a Randomized Trial of the Safety and Efficacy of the Use of the TAXUS Element Paclitaxel-Eluting Coronary Stent System) Trial
J. Am. Coll. Cardiol., July 20, 2010; 56(4): 264 - 271.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
R. A. Byrne, A. Kastrati, K. Tiroch, S. Schulz, J. Pache, S. Pinieck, S. Massberg, M. Seyfarth, K.-L. Laugwitz, K. A. Birkmeier, et al.
2-Year Clinical and Angiographic Outcomes From a Randomized Trial of Polymer-Free Dual Drug-Eluting Stents Versus Polymer-Based Cypher and Endeavor, Drug-Eluting Stents
J. Am. Coll. Cardiol., June 8, 2010; 55(23): 2536 - 2543.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
A. K. Jain, C. Lotan, I. T. Meredith, F. Feres, R. Zambahari, N. Sinha, M. T. Rothman, and for the E-Five Registry Investigators
Twelve-month outcomes in patients with diabetes implanted with a zotarolimus-eluting stent: results from the E-Five Registry
Heart, June 1, 2010; 96(11): 848 - 853.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
E. M. Ohman and R. M. Califf
When Can Noninferior Be Superior?: The Multidimensional Nature of Clinical Decision-Making Calls for Innovative Approaches to Clinical Trials
J. Am. Coll. Cardiol., February 9, 2010; 55(6): 555 - 557.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Online Appendix
Right arrow View Related Cardiosource Journal Scan
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (2)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Leon, M. B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Leon, M. B.
Related Collections
Right arrowRelated Articles

 
  CME Topic Collections Past Issues Search Current Issue Home

Advertisement