STATE-OF-THE-ART PAPER
Transcatheter Aortic Valve Implantation for Stenosed and Regurgitant Aortic Valve BioprosthesesCoreValve for Failed Bioprosthetic Aortic Valve Replacements
Muhammed Z. Khawaja, MBBS*,*,
Peter Haworth, MBBS*,
Azad Ghuran, MBChB, MD*,
Lorraine Lee, BSE*,
Adam de Belder, MD*,
Neville Hutchinson, MD*,
Uday Trivedi, MBBS*,
Jean-Claude Laborde, MD and
David Hildick-Smith, MD*
* Sussex Cardiac Centre, Brighton and Sussex University Hospitals NHS Trust, Brighton, United Kingdom
Clinique Pasteur, Toulouse, France
Manuscript received May 13, 2009;
revised manuscript received June 22, 2009,
accepted June 29, 2009.
* Reprint requests and correspondence: Dr. Muhammed Z. Khawaja, Specialist Registrar in Cardiology, Sussex Cardiac Centre, Brighton & Sussex University Hospitals NHS Trust, Eastern Road, Brighton, East Sussex BN2 5BE, United Kingdom (Email: drzeeshan{at}doctors.org.uk).
 |
Abstract
|
|---|
Transcatheter aortic valve implantation is increasingly being used to treat severe aortic stenosis in patients with high operative risk. In an aging population the incidence of aortic stenosis is rising, and increasing numbers of elderly patients are undergoing aortic valve replacement with bioprosthetic valves. Therefore, there is a corresponding increase in prosthetic degeneration. This presents cardiologists with a cohort of patients for whom the risk of re-do aortic valve surgery is prohibitive. We present the first series of such patients with degenerative bioprosthetic stenosis or regurgitation successfully treated with CoreValve (Medtronic, Luxembourg) implantation.
Key Words: aortic bioprosthesis TAVI
|
Abbreviations and Acronyms
| | AR = aortic regurgitation | | AS = aortic stenosis | | AVR = aortic valve replacement | | LV = left ventricular/ventricle | | TAVI = transcatheter aortic valve implantation |
|
Transcatheter aortic valve implantation (TAVI) has entered the mainstream as a viable treatment option for patients with symptomatic, severe aortic stenosis (AS) who are at prohibitively high surgical risk. The initial success in 2002 (1) has been followed by a surge of interest, strengthened by encouraging medium-term results with both the self-expanding CoreValve (Medtronic, Luxembourg) (2,3) and balloon-expandable Edwards Sapien (Edwards Lifesciences, Irvine, California) systems (4,5). Although not initially designed for the purpose, the CoreValve (Medtronic) implant is also an option for patients with a degenerative aortic valve bioprosthesis who would be at high surgical risk from repeat thoracotomy. We present a case series of 4 patients in whom this treatment has been used for failing bioprostheses.
 |
Case Histories
|
|---|
Patient #1.
A 76-year-old man underwent isolated aortic valve replacement (AVR) in 1986 for nonrheumatic AS with a 23-mm Carpentier-Edwards valve. During 2007 he became increasingly dyspneic and was found to have severe aortic regurgitation (AR) with moderate AS. His renal function was impaired (creatinine 201 mmol/l, estimated glomerular filtration rate 28 ml/min), and given his resultant high operative risk score (logistic EuroSCORE 19%), he was referred for TAVI. Transthoracic echocardiography confirmed severe eccentric aortic regurgitation, and the "aortic annulus" measured 21 mm. The procedure was performed under conscious sedation as previously described (6). The right radial artery was cannulated, and a 5-F pigtail catheter was advanced to the aortic root. A 4-F balloon flotation temporary pacing wire was placed via the right internal jugular vein. With the left common femoral approach, the arterial access was prepared for closure with the Prostar (Abbot Vascular Devices, Redwood City, California). The valve was crossed with an Amplatz L1 (Cordis Corp., Miami, Florida) catheter, a hand-curved J-tip Amplatz SuperStiff guidewire (Boston Scientific Corp., Miami, Florida) was coiled in the left ventricle, and an 18-F Cook sheath was introduced over the wire. Without valvular pre-dilation, a 26-mm CoreValve (Medtronic) was introduced and positioned under fluoroscopic guidance, with concurrent pacing at 150 beats/min to limit the effects of aortic regurgitation on valve advancement during deployment. Aortography and hemodynamic monitoring confirmed resolution of the aortic regurgitation (Fig. 1). There was no gradient across the new bioprosthesis on invasive assessment. The femoral access site was closed with the pre-laid ProStar suture pair. A post-procedure echocardiogram the next day showed normal functioning of the supra-annular valve, with a peak flow velocity of 2.7 m/s across it, and trivial aortic regurgitation.

View larger version (65K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 1 Successful Treatment of AR in a Bioprosthetic Aortic Valve Replacement
(A) Aortography demonstrating aortic regurgitation (AR) of the bioprosthesis. (B) With the bioprosthesis as a target for placement of the CoreValve (Medtronic). (C) Aortography showing the final position of the CoreValve (Medtronic) within the scaffold of the original device without AR.
|
|
Patient #2.
A 66-year-old man underwent AVR in 2001 for symptomatic AS with a 21-mm Mitroflow (Sorin Group Canada, Inc., Burnaby, British Columbia, Canada) bioprosthesis. Echocardiography revealed recurrence of aortic stenosis with regurgitation and significant left ventricular (LV) dilation. Dobutamine stress echo revealed an ejection fraction of 30% with evidence of contractile reserve. The patient was considered by the multi-disciplinary team to be at prohibitively high surgical risk (logistic EuroSCORE 25%) and was accepted for TAVI. Via the common femoral artery, aortography revealed grade IV AR with a gradient of 100 mm Hg across the valve. A 26-mm CoreValve (Medtronic) device was deployed under rapid ventricular pacing just inferior to the sewing ring of the original bioprosthesis, with resolution of the AR and a transvalvar gradient of 5 mm Hg.
The patient had a diuresis of 4 l over the first 6 h after the procedure and immediately felt much better. On the first post-operative day, he asked to go home but was persuaded to stay for an echocardiogram, which showed some early improvement in LV function (LV velocity time integral rising from 17.4 to 18.8 cm).
At 2-month follow-up he was asymptomatic and walking 4 miles/day. Repeat echocardiography revealed a well-positioned implant and improved hemodynamic parameters with a peak gradient of 50 mm Hg.
Patient #3.
An 85-year-old woman was diagnosed with degeneration of her 18-year-old AVR—a flail leaflet causing grade IV AR. Due to her high logistic EuroSCORE of 38%, she was referred for TAVI. Via the femoral approach, a 26-mm CoreValve (Medtronic) implant was deployed into the existing xenograft (Fig. 2) as previously described, with no procedural complications. At 5-month follow-up the patient reported a vast improvement in her quality of life with minimal dyspnea on exertion (New York Heart Association [NYHA] functional class I). A transthoracic echocardiogram showed that the CoreValve (Medtronic) was well-seated with only a mild paraprosthetic leak and a peak velocity of 2.4 m/s.

View larger version (60K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 2 CoreValve Deployment Within a Bioprosthesis
(A) Pre-procedure severe aortic regurgitation (AR). (B) Fully deployed CoreValve (Medtronic) prosthesis with (C) almost complete resolution of the AR.
|
|
Patient #4.
This patient underwent implantation of a 25-mm Cryolife-O'Brien (Cryolife International, Atlanta, Georgia) bioprosthesis in 2002 for AS. At 84 years of age, she developed degenerative stenosis of her AVR with a gradient of 83 mm Hg and mild AR. Iliac angiography revealed vessels of insufficient caliber to allow a transfemoral approach (Table 1). The multidisciplinary team agreed that TAVI via the left subclavian approach would be her best option.
The procedure was performed under general anesthesia. The left subclavian artery was accessed via a subclavicular surgical cut-down. Introduction of the 18-F St. Jude (St. Jude Medical, St. Paul, Minnesota) sheath required gradual rotation during advancement, because the fit in the subclavian artery was tight. The marker-tip of the sheath dehisced during introduction and was retrieved with an ENSnare (HATCH Medical, L.L.C., Duluth, Georgia) device (Fig. 3). The 26-mm CoreValve (Medtronic) bioprosthesis was advanced through the tipless sheath and deployed. The skirt remained constrained, but given the decrease in gradient to 50 mm Hg and complicated procedure, it was accepted pending further spontaneous valvular expansion.

View larger version (187K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 3 Retrieval of Fractured Sheath-Tip and Post-Implant Valvuloplasty
(A) The detached sheath tip within the ascending aorta, subsequently secured with an ENSnare (HATCH Medical, L.L.C., Duluth, Georgia) device (B). (C) Post-procedural valvuloplasty with a BALT (Montmorency, France) balloon improving the proximal deployment (note the temporary pacing wire within the right ventricle providing pace control). (D) Final position of the CoreValve (Medtronic) device.
|
|
Later that evening the patient developed a pale, cold, and pulseless left arm. Emergency subclavian arteriography revealed occlusive subclavian artery dissection. The vessel was approached retrogradely via percutaneous brachial artery access. A 7 x 80 mm LifeStent FlexStar (Edwards Lifesciences) was implanted, and flow restored. After a stormy recovery involving inotropic support and hemofiltration, repeat echocardiography revealed an increase in transvalvular gradient to 70 mm Hg. Therefore, we proceeded to aortic valvuloplasty with a 20-mm BALT (Montmorency, France) balloon. This decreased the trans-aortic gradient from 70 to 30 mm Hg without demonstrable AR (Figs. 3C and 3D). Review at 6 months showed the patient to be well, NYHA functional class II.
Echocardiography demonstrated good LV function with a peak gradient across the valve of 29 mm Hg and no AR.
 |
Discussion
|
|---|
Our case series demonstrates that the CoreValve (Medtronic) prosthesis can be safely and effectively deployed in stenotic and regurgitant degenerative surgical aortic valve bioprostheses. Implantation is facilitated by the existing valve structure, and immediate results show good hemodynamic status with a low transvalvular gradient and no AR. Medium- and long-term results are awaited, but in the meantime, CoreValve (Medtronic) TAVI seems to be an acceptable alternative to re-do surgery for patients with degenerative aortic bioprostheses who are at high perioperative surgical risk.
Aortic stenosis remains the most common valvular complaint in Europe (7,8), and age is a significant contributing factor in its natural history (8). The gold standard for treatment remains surgical AVR (9). The combined effects of an aging population and the prevalence of AS in this cohort have led to increasing numbers of elderly patients being referred for valvular surgery. The valve of choice in these patients is usually a bioprosthetic device, to avoid the need for anticoagulation with its associated complications (10). As a result, there is an increasing prevalence of failing bioprosthetic AVRs. Concurrently, the spectrum of comorbidities in these patients grows more complex, and therefore the risks of re-do surgery increase. Absolute surgical risk for re-do aortic valve operations might be raised by up to 15% compared with initial implants (11,12).
The feasibility of implanting a transcatheter valve into a surgical bioprosthesis was first demonstrated in a pig model in 2007. The investigators used the Cribier-Edwards system to transapically implant 23-mm devices into 5 Carpentier Edwards porcine aortic valve prostheses to good effect (13). A case report of successful treatment of a severely regurgitant aortic valve bioprosthesis with the CoreValve (Medtronic) system has also been described—with retention of good function and no complications at 1-year follow-up (14).
The risks inherent in the TAVI procedure still exist in these patients, as illustrated by the subclavian artery dissection in Patient #4. However, there are some features that make this option attractive. The original bioprosthesis gives a highly visible target for placement of the new implant and offers a scaffold that immediately grips the CoreValve (Medtronic) as it is deployed, facilitating accurate placement of the device. Additionally, the uniform circular sewing ring allows for equal and symmetrical expansion of the skirt of the valve such that paraprosthetic AR is unlikely.
Several other aspects of this technique warrant discussion in further detail. Firstly, in native cases, measurement of the aortic annulus is critical to the success of the procedure. In the cases presented, however, it is simply necessary to know the valvular diameter of the previously implanted valve. A 26-mm CoreValve (Medtronic) will be the correct option in the vast majority of cases. It is important to note that internal diameters of <21 mm in the aortic valve implant would be insufficient for CoreValve (Medtronic) implantation, exceeding its design limitations, with a resultant gradient and possible malfunction of the leaflets resulting in a suboptimal outcome.
Second, although native cases require balloon pre-dilation under rapid pacing control, this is not necessary with re-stenotic valves and indeed is likely to be counterproductive for bioprosthetic valves, which might disintegrate with risk of stroke or torrential regurgitation. Patients with severe aortic regurgitation, however, do benefit from pacing at implant both to reduce systolic pressure and to reduce reflux of blood into the LV, which will augment any tendency for the valve to "move forward" into the ventricle on deployment. This pacing rate, however, should aim to achieve a systolic blood pressure of 100 mm Hg rather than the 40 mm Hg used during balloon dilation and can therefore persist throughout the valve deployment. Currently, tamponade from temporary pacing wires has ironically proved one of the most risk-prone aspects of CoreValve (Medtronic) TAVI cases (15), so the lack of need for a temporary pacing wire in cases without severe AR adds to the safety of the procedure.
The need for permanent pacing after the procedure is a limitation of the CoreValve (Medtronic) system. The registry data suggest a rate of 20% to 25%. In this case series, none of the patients required a permanent pacemaker, presumably because the skirt of the valve is primarily restricted by the sewing ring of the original surgically implanted valve, such that compression of the conduction system is much less likely. Waiting to see whether a pacemaker might be required is a drain on bed-days, and it is advantageous to the patient to be able to mobilize freely on day 1 without significant risk of heart block.
The technique described in this report makes the use of a surgical bioprosthesis rather than a mechanical valve potentially more attractive in patients undergoing conventional valve replacement (16), and already this trend is being observed. However, the resulting internal diameter of the valve decreases with every deployment. This "Russian doll" effect might become more evident in light of successful procedures within both bioprosthetic and TAVI replacements and might serve as an additional driver for the development of valves with smaller diameters. Indeed the longest surviving recipient of the CoreValve (Medtronic) system was also the first to have a "valve-in-valve" procedure, and 3-year follow-up has shown maintenance of clinical status and a well-functioning valve (17). It is notable in our series that the gradients obtained on initial follow-up are not negligible. Although this is unlikely to be important in patients for whom there was no other acceptable surgical alternative, this would not be the case for a patient with a longer life-expectancy. Similarly, if the valvular gradient remains stable over time, this is less of a concern than if the gradient begins to rise with ingrowth of tissue.
The evidence for the use of TAVI is limited, as with any new technology, and consists of mainly of registry data. This can often include the steeper initial learning curve—although impressive procedural success rates of 97% and mortality of 1.5% have recently been reported by a multicenter group (18).
 |
Conclusions
|
|---|
We have confirmed the viability of TAVI in degenerative bioprostheses, both stenotic and regurgitant. Long-term follow up of these patients is required to demonstrate ongoing benefits and limitations of this technique.
 |
Footnotes
|
|---|
Dr. Laborde is a consultant and employee for CoreValve. Dr. Hildick-Smith is a proctor for CoreValve (Medtronic, Minneapolis, Minnesota).
 |
References
|
|---|
1. Cribier A, Eltchaninoff H, Bash A, et al. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description Circulation 2002;106:3006-3008.[Abstract/Free Full Text]2. Grube E, Laborde JC, Gerckens U, et al. Percutaneous implantation of the CoreValve self-expanding valve prosthesis in high-risk patients with aortic valve disease: the Siegburg first-in-man study Circulation 2006;114:1616-1624.[Abstract/Free Full Text] 3. Grube E, Schuler G, Buellesfeld L, et al. Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation self-expanding CoreValve prosthesis: device success and 30-day clinical outcome J Am Coll Cardiol 2007;50:69-76.[Abstract/Free Full Text] 4. Cribier A, Eltchaninoff H, Tron C, et al. Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosis J Am Coll Cardiol 2004;43:698-703.[Abstract/Free Full Text] 5. Cribier A, Eltchaninoff H, Tron C, et al. Treatment of calcific aortic stenosis with the percutaneous heart valve: mid-term follow-up from the initial feasibility studies: the French experience J Am Coll Cardiol 2006;47:1214-1223.[Abstract/Free Full Text] 6. Behan M, Haworth P, Hutchinson N, Trivedi U, Laborde JC, Hildick-Smith D. Percutaneous aortic valve implants under sedation: our initial experience Catheter Cardiovasc Interv 2008;72:1012-1015.[CrossRef][Web of Science][Medline] 7. Lindroos M, Kupari M, Heikkila J, Tilvis R. Prevalence of aortic valve abnormalities in the elderly: an echocardiographic study of a random population sample J Am Coll Cardiol 1993;21:1220-1225.[Abstract] 8. Iung B, Baron G, Butchart EG, et al. A prospective survey of patients with valvular heart disease in Europe: The Euro Heart Survey on Valvular Heart Disease Eur Heart J 2003;24:1231-1243.[Abstract/Free Full Text] 9. Kojodjojo P, Gohil N, Barker D, et al. Outcomes of elderly patients aged 80 and over with symptomatic, severe aortic stenosis: impact of patient's choice of refusing aortic valve replacement on survival QJM 2008;101:567-573.[Abstract/Free Full Text] 10. Accola KD, Scott ML, Palmer GJ, et al. Surgical management of aortic valve disease in the elderly: a retrospective comparative study of valve choice using propensity score analysis J Heart Valve Dis 2008;17:355-364discussion 365.[Web of Science][Medline] 11. Eitz T, Fritzsche D, Kleikamp G, Zittermann A, Horstkotte D, Korfer R. Reoperation of the aortic valve in octogenarians Ann Thorac Surg 2006;82:1385-1390.[Abstract/Free Full Text] 12. Jamieson WR, Burr LH, Miyagishima RT, et al. Re-operation for bioprosthetic aortic structural failure—risk assessment Eur J Cardiothorac Surg 2003;24:873-878.[Abstract/Free Full Text] 13. Walther T, Falk V, Dewey T, et al. Valve-in-a-valve concept for transcatheter minimally invasive repeat xenograft implantation J Am Coll Cardiol 2007;50:56-60.[Abstract/Free Full Text] 14. Wenaweser P, Buellesfeld L, Gerckens U, Grube E. Percutaneous aortic valve replacement for severe aortic regurgitation in degenerated bioprosthesis: the first valve in valve procedure using the Corevalve Revalving system Catheter Cardiovasc Interv 2007;70:760-764.[CrossRef][Web of Science][Medline] 15. Piazza N. CoreValve Implantation Complications. Paper presented at: Medtronic CoreValve 1st International Proctors Meeting; July 2009; Frankfurt, Germany. 16. Thomas M. Trans-catheter aortic valve implantation in the United Kingdom: NICE guidance Heart 2009;95:674-675.[Free Full Text] 17. Ruiz CE, Laborde JC, Condado JF, Chiam PT, Condado JA. First percutaneous transcatheter aortic valve-in-valve implant with three year follow-up Catheter Cardiovasc Interv 2008;72:143-148.[CrossRef][Web of Science][Medline] 18. Piazza N, Grube E, Gerckens U, et al. Procedural and 30-day outcomes following transcatheter aortic valve implantation using the third generation (18 Fr) corevalve revalving system: results from the multicentre, expanded evaluation registry 1-year following CE mark approval EuroIntervention 2008;4:242-249.[Medline]
Related Article
-
Inside This Issue
J. Am. Coll. Cardiol. 2009 55: A28.
[Full Text]
[PDF]
This article has been cited by other articles:

|
 |

|
 |
 
A. N. Azadani and E. E. Tseng
Transcatheter Heart Valves for Failing Bioprostheses: State-of-the-Art Review of Valve-in-Valve Implantation
Circ Cardiovasc Interv,
December 1, 2011;
4(6):
621 - 628.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. Gurvitch, A. Cheung, J. Ye, D. A. Wood, A. B. Willson, S. Toggweiler, R. Binder, and J. G. Webb
Transcatheter Valve-in-Valve Implantation for Failed Surgical Bioprosthetic Valves.
J. Am. Coll. Cardiol.,
November 15, 2011;
58(21):
2196 - 2209.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Eggebrecht, U. Schafer, H. Treede, P. Boekstegers, J. Babin-Ebell, M. Ferrari, H. Mollmann, H. Baumgartner, T. Carrel, P. Kahlert, et al.
Valve-in-Valve Transcatheter Aortic Valve Implantation for Degenerated Bioprosthetic Heart Valves
J. Am. Coll. Cardiol. Intv.,
November 1, 2011;
4(11):
1218 - 1227.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Bedogni, M. L. Laudisa, S. Pizzocri, C. Tamburino, G. P. Ussia, A. S. Petronio, M. Napodano, A. Ramondo, P. Presbitero, F. Ettori, et al.
Transcatheter Valve-in-Valve Implantation Using CoreValve Revalving System for Failed Surgical Aortic Bioprostheses
J. Am. Coll. Cardiol. Intv.,
November 1, 2011;
4(11):
1228 - 1234.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Descoutures, D. Himbert, C. Radu, B. Iung, C. Cueff, D. Messika-Zeitoun, G. Ducrocq, E. Brochet, P. Nataf, and A. Vahanian
Transarterial Medtronic CoreValve System Implantation for Degenerated Surgically Implanted Aortic Prostheses
Circ Cardiovasc Interv,
October 1, 2011;
4(5):
488 - 494.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. N. Azadani, N. Jaussaud, L. Ge, S. Chitsaz, T. A. M. Chuter, and E. E. Tseng
Valve-in-Valve Hemodynamics of 20-mm Transcatheter Aortic Valves in Small Bioprostheses
Ann. Thorac. Surg.,
August 1, 2011;
92(2):
548 - 555.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Pasic, A. Unbehaun, S. Dreysse, S. Buz, T. Drews, M. Kukucka, and R. Hetzer
Transapical aortic valve implantation after previous aortic valve replacement: Clinical proof of the "valve-in-valve" concept
J. Thorac. Cardiovasc. Surg.,
August 1, 2011;
142(2):
270 - 277.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N. Piazza, S. Bleiziffer, G. Brockmann, R. Hendrick, M.-A. Deutsch, A. Opitz, D. Mazzitelli, P. Tassani-Prell, C. Schreiber, and R. Lange
Transcatheter Aortic Valve Implantation for Failing Surgical Aortic Bioprosthetic Valve: From Concept to Clinical Application and Evaluation (Part 1)
J. Am. Coll. Cardiol. Intv.,
July 1, 2011;
4(7):
721 - 732.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N. Piazza, S. Bleiziffer, G. Brockmann, R. Hendrick, M.-A. Deutsch, A. Opitz, D. Mazzitelli, P. Tassani-Prell, C. Schreiber, and R. Lange
Transcatheter Aortic Valve Implantation for Failing Surgical Aortic Bioprosthetic Valve: From Concept to Clinical Application and Evaluation (Part 2)
J. Am. Coll. Cardiol. Intv.,
July 1, 2011;
4(7):
733 - 742.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Drews, M. Pasic, S. Buz, A. Unbehaun, S. Dreysse, M. Kukucka, A. Mladenow, and R. Hetzer
Transapical aortic valve implantation after previous heart surgery
Eur J Cardiothorac Surg,
May 1, 2011;
39(5):
625 - 630.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
I. J. Nunez-Gil, A. Goncalves, E. Rodriguez, J. Cobiella, P. Marcos-Alberca, L. Maroto, C. Fernandez-Golfin, M. Carnero, C. Macaya, and J. L. Zamorano
Transapical mitral valve-in-valve implantation: a novel approach guided by three-dimensional transoesophageal echocardiography
Eur Heart J Cardiovasc Imaging,
April 1, 2011;
12(4):
335 - 337.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. N. DeMaria, J. J. Bax, O. Ben-Yehuda, G. K. Feld, B. H. Greenberg, J. Hall, M. Hlatky, W. Y. W. Lew, J. A. C. Lima, A. S. Maisel, et al.
Highlights of the Year in JACC 2010
J. Am. Coll. Cardiol.,
January 25, 2011;
57(4):
480 - 514.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Webb and A. Cribier
Percutaneous transarterial aortic valve implantation: what do we know?
Eur. Heart J.,
January 2, 2011;
32(2):
140 - 147.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. R. Coeytaux, J. W. Williams Jr., R. N. Gray, and A. Wang
Percutaneous Heart Valve Replacement for Aortic Stenosis: State of the Evidence
Ann Intern Med,
September 7, 2010;
153(5):
314 - 324.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|