J Am Coll Cardiol, 2009; 54:822-828, doi:10.1016/j.jacc.2009.06.004
© 2009 by the American College of Cardiology Foundation
This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (1)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Tereshchenko, L. G.
Right arrow Articles by Berger, R. D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tereshchenko, L. G.
Right arrow Articles by Berger, R. D.
Related Collections
Right arrowRelated Articles

QUARTERLY FOCUS ISSUE: HEART RHYTHM DISORDERS: CLINICAL RESEARCH

Transient Local Injury Current in Right Ventricular Electrogram After Implantable Cardioverter-Defibrillator Shock Predicts Heart Failure Progression

Larisa G. Tereshchenko, MD, PhD*,{dagger},*, Mitchell N. Faddis, MD, PhD{dagger}, Barry J. Fetics, MS*, Karl E. Zelik, MS{ddagger},§, Igor R. Efimov, PhD{ddagger} and Ronald D. Berger, MD, PhD*

* Division of Cardiology, Department of Medicine, Johns Hopkins Hospital, Baltimore, Maryland
{dagger} Cardiovascular Division, Washington University School of Medicine, St. Louis, Missouri
{ddagger} Biomedical Engineering Department, Washington University in St. Louis, St. Louis, Missouri
§ University of Michigan, Ann Arbor, Michigan

Manuscript received February 4, 2009; revised manuscript received June 1, 2009, accepted June 11, 2009.

* Reprint requests and correspondence: Dr. Larisa G. Tereshchenko, Carnegie 592, 600 North Wolfe Street, Baltimore, Maryland 21287 (Email: lteresh1{at}jhmi.edu).


    Abstract
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
Objectives: This study aimed to identify an early marker of functional impairment after an implantable cardioverter-defibrillator (ICD) shock as a predictor of heart failure progression.

Background: The ICD population has substantial risk of death due to progressive pump failure.

Methods: Near-field (NF) bipolar right ventricular (RV) electrograms (EGMs) during induced ventricular fibrillation (VF) and 10 s after rescue ICD shock were analyzed in 310 patients (mean age 59 ± 14.5 years, 219 men [71%]) with structural heart disease, New York Heart Association functional class I to III, and implanted with a single- or dual-chamber Medtronic (Minneapolis, Minnesota) ICD for primary (245 patients, 79%) or secondary prevention of sudden cardiac arrest. A local injury current (LIC) on NF RV EGM was defined as a deviation of EGM potential ≥1 mV or ≥15% of the preceding R-wave peak-to-peak amplitude.

Results: During mean follow-up of 29.3 ± 15.0 months, the combined end point of death or hospitalization due to congestive heart failure (CHF) exacerbation was documented in 40 patients (12.9%, or 5.3% per person-year of follow-up). LIC was observed in 106 patients. In multivariate risk analysis, after adjustment for baseline prognostic factors (ejection fraction, history of atrial fibrillation, diabetes mellitus) and appropriate ICD shocks during follow-up, patients with observed LIC after induced VF rescue ICD shock at ICD implantation were more likely to die or to be hospitalized (hazard ratio: 2.69; 95% confidence interval: 1.41 to 5.14; p = 0.003).

Conclusions: Transient LIC on bipolar NF RV EGM after induced VF rescue ICD shock is associated with increased risk of CHF progression, future hospitalizations due to CHF exacerbation, and subsequent heart failure death.

Key Words: congestive heart failure • implantable cardioverter-defibrillator • ventricular tachyarrhythmia

Abbreviations and Acronyms
  CHF = congestive heart failure
  CI = confidence interval
  DFT = defibrillation threshold testing
  EGM = electrogram
  HR = hazard ratio
  ICD = implantable cardioverter-defibrillator
  LIC = local injury current
  LVEF = left ventricular ejection fraction
  NF = near field
  NYHA = New York Heart Association
  RV = right ventricle/ventricular
  SCA = sudden cardiac arrest
  VF = ventricular fibrillation
  VT = ventricular tachycardia


Implantable cardioverter-defibrillators (ICDs) improve survival of patients who are at risk for sudden cardiac arrest (SCA) (1–4). However, long-term follow-up of ICD patients with congestive heart failure (CHF) has shown that both appropriate (5–7) and inappropriate (5,8) ICD shocks are associated with increased risk of death, predominantly from progressive heart failure. Thoughtful medical management of heart failure and programming of ICD therapies in this patient cohort might improve the prognosis, but no early markers of heart failure progression available at the time of ICD implantation are known.

Extensive data indicate that defibrillation shocks are accompanied by transient adverse effects. These adverse effects include: 1) transient ectopy, tachycardia, or induction of ventricular fibrillation (VF) (9,10); 2) complete heart block and increased pacing thresholds (10,11); 3) atrial and ventricular mechanical dysfunction (stunning) (12–15); 4) significant elevation of troponin I serum level (16); and 5) decrease of the myocardial lactate extraction rate by mitochondria (17). Whether transient signs of myocardial injury after an ICD shock could predict future progression of CHF remains unclear.

Changes of electrocardiogram (ECG) and intracardiac electrograms (EGMs) during ICD implantation procedures were observed previously. Transient ST-segment elevation on surface ECG after induced VF rescue ICD shock was described in 19% of patients (18), but its prognostic significance was not studied. Other investigators have shown that a current of injury on intracardiac EGM within 10 min of lead fixation serves as a marker of adequate active lead fixation during an ICD or pacemaker implantation procedure (19,20). The prognostic significance of transient local injury current (LIC) on near-field (NF) right ventricular (RV) EGM after induced VF rescue ICD shock is unknown. We hypothesized that LIC on bipolar NF RV EGM after induced VF rescue ICD shock predicts future CHF progression in patients with New York Heart Association (NYHA) functional class I to III CHF.


    Methods
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
The study protocol was approved by the Johns Hopkins University and the Washington University Human Studies Committees, and all patients gave written informed consent before entering the study.

Study population.   This is prospective observational study. Male and female patients older than 18 years with structural heart disease and NYHA functional class I to III CHF were eligible for the study if they had a Medtronic (Minneapolis, Minnesota) transvenous single- or dual-chamber ICD device with dedicated bipolar ICD lead implanted for primary or secondary prevention of SCA within 1 week before enrollment. Exclusion criteria were indications for cardiac resynchronization therapy defibrillator and NYHA functional class IV, contraindications for defibrillation threshold testing (DFT), pregnancy, inherited channelopathies, and concomitant conditions other than cardiac diseases that were associated with a high likelihood of death during 1 year after enrollment.

Ventricular tachycardia (VT)/VF was induced with a shock-on T-wave protocol. Stored intracardiac EGMs recorded during DFT (induced tachyarrhythmia and 10 s post-ICD shock) were extracted from the ICD memory 7 days after procedure, converted into digital format using proprietary Medtronic software, and further analyzed using custom Matlab (The MathWorks, Inc., Natick, Massachusetts) software application. Control recordings of NF RV EGM at rest simultaneously with 1-lead (lead II) surface ECG were obtained via Medtronic programmer 2090 using the NI USB-9215A portable data acquisition system (National Instruments, Austin, Texas) 7 days after the procedure.

Programming of the ICD device was based on the attending electrophysiologist's clinical evaluation. Patients were followed-up in the Washington University Arrhythmia Clinic and via the Internet-based CareLink (Medtronic, Inc., Minneapolis, Minnesota) remote monitoring system. All ICD interrogation data were adjudicated by an ICD end point committee (attending electrophysiologist and 2 of the investigators [L.G.T. and R.D.B.]). ICD shocks occurring for VT or VF were classified as appropriate.

Measurement of LIC on the bipolar NF RV EGM after induced VF rescue ICD shock.   Endocardial NF RV EGM was recorded as the difference of potentials between the tip and the ring of the dedicated bipolar ICD lead implanted in the RV apex. The LIC was characterized as the magnitude of elevated or depressed potential immediately after the major fast EGM deflection (Fig. 1), measured from the baseline (the isoelectric portion before the major EGM deflection) to its highest point in millivolts. Peak-to-peak amplitude of major fast EGM deflection (R-wave) was measured to assess relative LIC on average representative beat. Significant LIC was defined as a deviation of EGM potential ≥1 mV or ≥15% of preceding R-wave peak-to-peak amplitude. Digital EGM (bandpass filter 2 to 100 Hz) was magnified and measured after separate calibration of each recording (1 mV equal to 30 to 40 pixels, Screen Calipers 4.0, Iconico, Inc., New York, New York). The first 2 s after shock were excluded. LIC was measured on every sinus beat and averaged. Ventricular-paced beats, distorted beats of undetermined origin, and ectopic beats were excluded.


Figure 1
View larger version (14K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 1 Measurement of LIC After Induced VF Rescue ICD Shock

ICD = implantable cardioverter-defibrillator; LIC = local injury current; R = peak-to-peak R-wave amplitude; VF = ventricular fibrillation.

 
End points.   Either of 2 major CHF events—death or hospitalization due to CHF exacerbation, whichever came first—served as the primary end point. We use the term "CHF event" to refer to this combined end point throughout the report. Cases of death with clear confirmed noncardiovascular cause were censored at the time of the last office visit. Time to event was measured from the day of ICD implantation.

Statistical analysis.   Results are presented as mean ± SD for normally distributed variables, and as median and interquartile range for skewed distributions. Continuous variables were compared using the independent samples t test if normally distributed and the Wilcoxon rank sum test if skewed. The Pearson chi-square test was used to compare categorical variables. A p value of <0.05 was considered significant. Kaplan-Meier survival analysis was used to compute mean and median survival time. The log-rank (Mantel-Cox) statistic was computed to test the equality of survival distributions. Cox multivariate regression model was used for adjustment by known predictors of CHF progression. Appropriate ICD shock for VT/VF at follow-up was treated as a time-dependent covariate. SPSS version 17.0.0 (SPSS Inc., Chicago, Illinois) and STATA version 10 (StataCorp LP, College Station, Texas) software packages were used for calculations.


    Results
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
Study population.   The study population consisted of 310 patients (mean age 59.0 ± 14.5 years, 219 [71%] men) who underwent ICD implantation for primary (n = 245, 79%) or secondary (n = 65, 21%) prevention of SCA. Ischemic cardiomyopathy with myocardial infarction history was diagnosed in 187 (60.3%) patients and nonischemic cardiomyopathy in 123 (39.7%) patients. A single-chamber ICD was implanted in 175 (56.5%) patients, and dual-chamber ICD in 135 (43.5%) patients. A new dedicated bipolar transvenous ICD lead was implanted in 264 (85.1%) patients, and an ICD generator change procedure was performed in 46 (14.9%) patients who had had an ICD lead implanted more than 1 year ago. Only the first induced VT/VF and EGM after the first rescue ICD shock was analyzed.

Bipolar NF RV EGM changes after induced VF rescue ICD shock.   Significant LIC after induced rescue ICD shock was found in 106 (34.2%) patients. The baseline characteristics of the patients are summarized in Table 1, and ICD shock characteristics are summarized in Table 2. Figure 2 shows examples of EGM changes after induced VF rescue ICD shock. Control EGMs obtained 7 days after the procedure confirmed that observed changes were temporary and demonstrated isoelectric potential.


View this table:
[in this window]
[in a new window]

 
Table 1 Clinical Characteristics of Patients With and Without LIC After ICD Shock
 

View this table:
[in this window]
[in a new window]

 
Table 2 Characteristics of Induced VF Events and ICD Rescue Shocks
 

Figure 2
View larger version (23K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 2 RV EGMs: Control and After ICD Shock

(A) Typical post-shock implantable cardioverter-defibrillator (ICD) recording of right ventricular (RV) near-field electrogram (EGM) in control recording, and after ICD shock in patients with local injury current (A), and in patients with lack of injury (B).

 
Death, CHF hospitalizations, and appropriate ICD shocks.   During a mean follow-up of 29.3 ± 15.0 months, the combined end point of death or hospitalization due to CHF exacerbation was documented in 40 patients (12.9%, or 5.3% per person-year of follow-up). Appropriate ICD shocks were observed in 78 patients (25.2%, or 10.3% per person-year of follow-up); of these patients, 3 patients died median 113 days after appropriate ICD shocks, and 3 patients underwent successful heart transplantation. CHF events were twice as frequent among patients with appropriate ICD shocks (n = 16, 20.5%) compared with patients without sustained arrhythmia (n = 24 of 232, 10.3%; p = 0.020). ICD shock preceded CHF event by median 132 days (interquartile range 6 to 627 days).

Risk of CHF progression associated with LIC at ICD implantation.   LIC (–) patients had a higher CHF event-free survival rate during follow-up (88.1% vs. 71.1%, p = 0.015) (Fig. 3A). Cox proportional hazard ratio (HR) for the newly implanted ICD lead subgroup was higher (HR: 3.29, 95% confidence interval [CI]: 1.54 to 7.06, p = 0.002) than for all patients (HR: 2.61, 95% CI: 1.37 to 4.99, p = 0.004). Figure 3B shows Kaplan-Meier curves when the analysis was confined to patients with newly implanted leads. This effect was not significant for the chronic ICD lead subgroup (HR: 0.78, 95% CI: 0.09 to 6.67, p = 0.820). Multivariate Cox model that included LIC, time-dependent appropriate ICD shocks at follow-up, new/chronic lead factor, and interaction between LIC and the lead factor confirmed effect modification (p < 0.0001).


Figure 3
View larger version (19K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 3 LIC Predicts Risk of CHF

Kaplan-Meier curves for freedom from congestive heart failure (CHF) events in patients with significant local injury current (LIC) (+) and those with the LIC (–) for all leads (A) and for newly implanted leads only (B).

 
In patients with LIC, subsequent sustained VT/VF events with appropriate ICD shocks predicted CHF progression (Fig. 4A) (event-free survival 40% vs. 80%, p < 0.006), whereas in patients without LIC, subsequent VT/VF was not predictive (Fig. 4B) (event-free survival 87% vs. 88%, p = 0.683). Multivariate Cox regression model that included LIC, time-dependent appropriate ICD shocks, and interaction between LIC and ICD shocks confirmed significant effect modification (p = 0.001).


Figure 4
View larger version (21K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 4 LIC Modifies CHF Risk After ICD Shock

Kaplan-Meier curves for freedom from CHF events in patients with and without appropriate ICD shocks at follow-up among LIC (+) patients (A), and LIC (–) patients (B). Abbreviations as in Figures 1 and 2.

 
After adjustment for baseline factors (age, race, left ventricular ejection fraction [LVEF], NYHA functional class, history of diabetes mellitus, atrial fibrillation or flutter, renal failure, hypertension, use of digoxin and aldosterone antagonists) and time-dependent appropriate ICD shocks during follow-up, LIC signified a highly increased risk of subsequent CHF events. Each Cox model included LIC, time-dependent appropriate ICD shocks during follow-up, NYHA functional class, and other covariates tested one by one as listed in Table 3. LIC was a significant predictor in all tested Cox models with HR from 2.2 to 2.6 (p < 0.01). HRs of time-dependent appropriate ICD shock at follow-up ranged from 2.5 to 7.1. Time-dependent appropriate ICD shock at follow-up was not a significant predictor in the models that included cycle length of VF and renal failure. NYHA functional class HRs ranged from 2.7 to 3.1 and were significant in all models (p < 0.001).


View this table:
[in this window]
[in a new window]

 
Table 3 Univariate and Multivariate HRs of Tested Predictors
 

    Discussion
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
To our knowledge, this is the first description of LIC on bipolar NF RV EGM after ICD shock. Our results demonstrate for the first time that transient myocardial injury after induced VF rescue ICD shock manifesting as LIC on bipolar NF RV EGM is associated with increased risk of CHF progression, future hospitalizations due to CHF exacerbation, and heart pump failure death. LIC after induced VF rescue ICD shock was a predictor of adverse CHF outcomes after adjustment by traditional risk factors, including appropriate ICD shocks and LVEF or NYHA functional class, and provided additional prognostic information.

We propose a "triple-hit" hypothesis to explain the genesis of LIC on the NF RV EGM: 1) cardiac myocytes are fragile due to an underlying condition that leads to progressive CHF; 2) mechanical injury occurs due to lead placement; and 3) a rescue ICD shock elicits LIC, especially if the first 2 "hits" are present.

CHF progression in ICD patients.   High risk of death due to pump failure in ICD patient populations without or after appropriate ICD therapies remains an important health care problem. Several clinical factors elucidated to be prognostic for CHF progression in ICD patients in previous studies are appropriate and inappropriate ICD shocks (5), renal failure (21), NYHA functional class, and LVEF (22). Our study is the first to show that the LIC phenomenon after induced VF rescue ICD shock carries an independent high risk, if observed in newly implanted ICD leads.

It is known that neurohumoral and cytokine activations contribute to the inflammatory and oxidative characteristics of CHF patients (23). We speculate that these pathways that have been activated for the long term in at-risk CHF patients result in a dramatic response to induced VF rescue ICD shock. Since patients without subsequent CHF were less likely to exhibit LIC, susceptibility for heart failure progression appears to be the "first hit" prerequisite for the LIC phenomenon we observed, and allows the appearance of LIC to serve as a marker of CHF risk.

Local mechanical myocardial injury and injury current on bipolar intracardiac EGM.   Transient LIC presenting on NF RV EGM during the acute placement of an ICD or pacemaker lead is well known. Several groups of investigators linked characteristics of LIC at the time of an active-fixation lead placement with subsequent adequate lead fixation (19,20) and with lead perforation (24). Transvenous insertion of endocardial leads for permanent pacing (25) or use with an ICD (26) is accompanied by acute injury, followed by a sequence of cardiac histopathological changes starting with acute inflammation and leading eventually to the formation of a fibrous connective tissue scar (27,28). Maximum ventricular lead diameter, number of implanted leads (25), and CRT device left ventricular lead placement (29) were independent predictors of peak cardiac troponin I levels in patients undergoing conventional pacemaker/ICD implantation. Less frequently observed LIC in patients with chronic leads is a finding that suggests recent local mechanical injury is usually required as the "second hit" for LIC to occur. Future study is needed to determine the optimal prognostic time window from lead fixation to VT/VF induction.

Transient myocardial injury after ICD shock.   According to the excitation theory of defibrillation, electrical shocks depolarize the membranes of most cardiac cells, resulting in resynchronization of electrical activity of the heart. If shock-induced changes in transmembrane potential are excessively large, they can cause transient cell membrane damage due to electroporation (30–32). Other potential causes of myocardial injury after an ICD shock include free radical formation (33,34) and conformation changes of the membrane ion channels (35). We speculate that enhanced LIC on bipolar NF RV EGM after an ICD shock in patients prone to subsequent CHF progression is produced by local voltage gradients, resulting from potential differences between electroporated myocardial cells and normal cells. In our study, LIC was observed after ICD shock, but not in subsequent control EGMs, thus supporting the importance of the shock and possibly the induced arrhythmia) for the "third hit."

Our results show that appropriate ICD shock predicts future CHF exacerbation and death only in patients with significant LIC after rescue ICD shock. Conversely, patients without LIC and subsequent appropriate ICD shocks during follow-up had the same favorable course as patients without ICD shocks. This important clinical finding suggests that ICD shock does not cause, but rather unveils, risk of progressive CHF.

Study limitations.   Our observations were limited by the 10-s post-shock EGM recording storage. We were unable to determine a final recovery time point and duration of EGM changes. EGM after ICD lead fixation but before induced VT/VF was not available for analysis.

Specific filter settings on bipolar NF RV EGM may preclude analysis of other manufacturers' EGMs. The small number of end point events limited multivariate Cox regression analysis.

The small number of chronic leads in this study prompts further investigation of LIC after ICD shock in chronic leads to determine its predictive value for subsequent CHF exacerbation.

In this study, we did not test the effect of ICD shock alone, without preceding induced arrhythmia, on the genesis of LIC. Theoretically, this could have been assessed at device implantation through the use of a protocol for upper limit of vulnerability testing (36,37), instead of that used to determine DFT.


    Conclusions
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
The observed LIC phenomenon predicts progression of CHF in ICD patients with appropriate ICD shocks, and with otherwise stable NYHA functional class I to III CHF. Early awareness of the high risk of CHF exacerbation and thoughtful medical management may improve CHF prognosis in ICD patients.


    Acknowledgments
 
The authors thank Jane Chen, Timothy Smith, Marye Gleva, and Bruce Lindsay for providing medical care for study participants, and Judy Osborn for help with collection of follow-up data.


    Footnotes
 
This study was supported by Medtronic, Inc., as an Investigator-Initiated Research Project (awarded to Drs. Berger and Tereshchenko). Dr. Faddis has served as a consultant for Boston Scientific Corp., Stereofaxis, Inc., and St. Jude Medical, Inc. Dr. Efimov has been a stockholder and Scientific Advisory Board Chair with CARDIALEN, and has received National Institutes of Health grants HL67322, HL074283, and HL082729.


    References
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
1. Anderson JL, Hallstrom AP, Epstein AE, et al. Design and results of the Antiarrhythmics Vs Implantable Defibrillators (AVID) registry. The AVID Investigators. Circulation 1999;99:1692-1699.[Abstract/Free Full Text]

2. Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G. A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators. N Engl J Med 1999;341:1882-1890.[CrossRef][Web of Science][Medline]

3. Moss AJ, Hall WJ, Cannom DS, et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators. N Engl J Med 1996;335:1933-1940.[CrossRef][Web of Science][Medline]

4. Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction N Engl J Med 2002;346:877-883.[CrossRef][Web of Science][Medline]

5. Poole JE, Johnson GW, Hellkamp AS, et al. Prognostic importance of defibrillator shocks in patients with heart failure N Engl J Med 2008;359:1009-1017.[CrossRef][Medline]

6. Moss AJ, Greenberg H, Case RB, et al. Long-term clinical course of patients after termination of ventricular tachyarrhythmia by an implanted defibrillator Circulation 2004;110:3760-3765.[Abstract/Free Full Text]

7. Goldenberg I, Moss AJ, Hall WJ, et al. Causes and consequences of heart failure after prophylactic implantation of a defibrillator in the multicenter automatic defibrillator implantation trial II Circulation 2006;113:2810-2817.[Abstract/Free Full Text]

8. Daubert JP, Zareba W, Cannom DS, et al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact J Am Coll Cardiol 2008;51:1357-1365.[Abstract/Free Full Text]

9. Fast VG, Cheek ER. Optical mapping of arrhythmias induced by strong electrical shocks in myocyte cultures Circ Res 2002;90:664-670.[Abstract/Free Full Text]

10. Waldecker B, Brugada P, Zehender M, Stevenson W, Welens HJ. Ventricular arrhythmias after precordial electric shock Am J Cardiol 1986;57:120-123.[CrossRef][Web of Science][Medline]

11. Eysmann SB, Marchlinski FE, Buxton AE, Josephson ME. Electrocardiographic changes after cardioversion of ventricular arrhythmias Circulation 1986;73:73-81.[Abstract/Free Full Text]

12. Sparks PB, Kulkarni R, Vohra JK, et al. Effect of direct current shocks on left atrial mechanical function in patients with structural heart disease J Am Coll Cardiol 1998;31:1395-1399.[Abstract/Free Full Text]

13. Sparks PB, Jayaprakash S, Mond HG, Vohra JK, Grigg LE, Kalman JM. Left atrial mechanical function after brief duration atrial fibrillation J Am Coll Cardiol 1999;33:342-349.[Abstract/Free Full Text]

14. Grimm RA, Stewart WJ, Arheart K, Thomas JD, Klein AL. Left atrial appendage "stunning" after electrical cardioversion of atrial flutter: an attenuated response compared with atrial fibrillation as the mechanism for lower susceptibility to thromboembolic events J Am Coll Cardiol 1997;29:582-589.[Abstract]

15. Kam RM, Garan H, McGovern BA, Ruskin JN, Harthorne JW. Transient right bundle branch block causing R wave attenuation postdefibrillation Pacing Clin Electrophysiol 1997;20:130-131.[CrossRef][Medline]

16. Hasdemir C, Shah N, Rao AP, et al. Analysis of troponin I levels after spontaneous implantable cardioverter defibrillator shocks J Cardiovasc Electrophysiol 2002;13:144-150.[CrossRef][Web of Science][Medline]

17. Osswald S, Trouton TG, O'Nunain SS, Holden HB, Ruskin JN, Garan H. Relation between shock-related myocardial injury and defibrillation efficacy of monophasic and biphasic shocks in a canine model Circulation 1994;90:2501-2509.[Abstract/Free Full Text]

18. Gurevitz O, Lipchenca I, Yaacoby E, et al. ST-segment deviation following implantable cardioverter defibrillator shocks: incidence, timing, and clinical significance Pacing Clin Electrophysiol 2002;25:1429-1432.[CrossRef][Medline]

19. Saxonhouse SJ, Conti JB, Curtis AB. Current of injury predicts adequate active lead fixation in permanent pacemaker/defibrillation leads J Am Coll Cardiol 2005;45:412-417.[Abstract/Free Full Text]

20. Redfearn DP, Gula LJ, Krahn AD, Skanes AC, Klein GJ, Yee R. Current of injury predicts acute performance of catheter-delivered active fixation pacing leads Pacing Clin Electrophysiol 2007;30:1438-1444.[Medline]

21. Schefer T, Wolber T, Binggeli C, Holzmeister J, Brunckhorst C, Duru F. Long-term predictors of mortality in ICD patients with non-ischaemic cardiac disease: impact of renal function Europace 2008;10:1052-1059.[Abstract/Free Full Text]

22. Lee DS, Tu JV, Austin PC, et al. Effect of cardiac and noncardiac conditions on survival after defibrillator implantation J Am Coll Cardiol 2007;49:2408-2415.[Abstract/Free Full Text]

23. Tang WH, Francis GS. The year in heart failure J Am Coll Cardiol 2008;52:1671-1678.[Free Full Text]

24. Van Gelder BM, Bracke FA. Current of injury (COI) pattern recorded from catheter delivered active fixation pacing leads Pacing Clin Electrophysiol 2008;31:786-787.[CrossRef][Medline]

25. Nikolaou NI, Spanodimos SG, Tsaglis EP, et al. Biochemical evidence of cardiac damage following transvenous implantation of a permanent antibradycardia pacemaker lead Pacing Clin Electrophysiol 2005;28:1174-1181.[CrossRef][Medline]

26. Hurst TM, Hinrichs M, Breidenbach C, Katz N, Waldecker B. Detection of myocardial injury during transvenous implantation of automatic cardioverter-defibrillators J Am Coll Cardiol 1999;34:402-408.[Abstract/Free Full Text]

27. Ford SE, Manley PN. Indwelling cardiac catheters. An autopsy study of associated endocardial lesions. Arch Pathol Lab Med 1982;106:314-317.[Web of Science][Medline]

28. Beyersdorf F, Schneider M, Kreuzer J, Falk S, Zegelman M, Satter P. Studies of the tissue reaction induced by transvenous pacemaker electrodes. I. Microscopic examination of the extent of connective tissue around the electrode tip in the human right ventricle. Pacing Clin Electrophysiol 1988;11:1753-1759.[CrossRef][Medline]

29. Altin T, Akyurek O, Vurgun K, et al. Effect of transvenous cardiac resynchronization therapy device implantation on cardiac troponin I release Pacing Clin Electrophysiol 2007;30:1356-1362.[CrossRef][Medline]

30. Al Khadra A, Nikolski V, Efimov IR. The role of electroporation in defibrillation Circ Res 2000;87:797-804.[Abstract/Free Full Text]

31. Nikolski VP, Sambelashvili AT, Krinsky VI, Efimov IR. Effects of electroporation on optically recorded transmembrane potential responses to high-intensity electrical shocks Am J Physiol Heart Circ Physiol 2004;286:H412-H418.[Abstract/Free Full Text]

32. Nikolski VP, Efimov IR. Electroporation of the heart Europace 2005;7(Suppl 2):146-154.[Free Full Text]

33. Caterine MR, Spencer KT, Pagan-Carlo LA, Smith RS, Buettner GR, Kerber RE. Direct current shocks to the heart generate free radicals: an electron paramagnetic resonance study J Am Coll Cardiol 1996;28:1598-1609.[Abstract]

34. Ravingerova T, Slezak J, Tribulova N, Dzurba A, Uhrik B, Ziegelhoffer A. Free oxygen radicals contribute to high incidence of reperfusion-induced arrhythmias in isolated rat heart Life Sci 1999;65:1927-1930.[CrossRef][Web of Science][Medline]

35. Chen W, Lee RC. Altered ion channel conductance and ionic selectivity induced by large imposed membrane potential pulse Biophys J 1994;67:603-612.[Web of Science][Medline]

36. Day JD, Doshi RN, Belott P, et al. Inductionless or limited shock testing is possible in most patients with implantable cardioverter-defibrillators/cardiac resynchronization therapy defibrillators: results of the multicenter ASSURE study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations) Circulation 2007;115:2382-2389.[Abstract/Free Full Text]

37. Swerdlow CD, Shehata M, Chen PS. Using the upper limit of vulnerability to assess defibrillation efficacy at implantation of ICDs Pacing Clin Electrophysiol 2007;30:258-270.[CrossRef][Medline]


Related Articles

Predicting Outcome After Implantable Cardioverter-Defibrillator Therapy: A New Piece to the Puzzle?
Richard K. Shepard and Kenneth A. Ellenbogen
J. Am. Coll. Cardiol. 2009 54: 829-831. [Full Text] [PDF]

Inside This Issue
J. Am. Coll. Cardiol. 2009 54: A28. [Full Text] [PDF]



This article has been cited by other articles:


Home page
Circ Arrhythm ElectrophysiolHome page
E. H. Weiss, F. M. Merchant, A. d'Avila, L. Foley, V. Y. Reddy, J. P. Singh, T. Mela, J. N. Ruskin, and A. A. Armoundas
A Novel Lead Configuration for Optimal Spatio-Temporal Detection of Intracardiac Repolarization Alternans
Circ Arrhythm Electrophysiol, June 1, 2011; 4(3): 407 - 417.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
A. N. DeMaria, J. J. Bax, O. Ben-Yehuda, G. K. Feld, B. H. Greenberg, J. Hall, M. Hlatky, W. Y.W. Lew, J. A.C. Lima, A. S. Maisel, et al.
Highlights of the Year in JACC 2009
J. Am. Coll. Cardiol., January 26, 2010; 55(4): 380 - 407.
[Full Text] [PDF]


Home page
J Am Coll CardiolHome page
R. K. Shepard and K. A. Ellenbogen
Predicting Outcome After Implantable Cardioverter-Defibrillator Therapy: A New Piece to the Puzzle?
J. Am. Coll. Cardiol., August 25, 2009; 54(9): 829 - 831.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (1)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Tereshchenko, L. G.
Right arrow Articles by Berger, R. D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tereshchenko, L. G.
Right arrow Articles by Berger, R. D.
Related Collections
Right arrowRelated Articles

 
  CME Topic Collections Past Issues Search Current Issue Home

Advertisement