CLINICAL RESEARCH: HEART FAILURE
Extended Mechanical Circulatory Support With a Continuous-Flow Rotary Left Ventricular Assist Device
Francis D. Pagani, MD, PhD*,*,
Leslie W. Miller, MD ,
Stuart D. Russell, MD ,
Keith D. Aaronson, MD*,
Ranjit John, MD ,
Andrew J. Boyle, MD ,
John V. Conte, MD ,
Roberta C. Bogaev, MD||,
Thomas E. MacGillivray, MD¶,
Yoshifumi Naka, MD#,
Donna Mancini, MD#,
H. Todd Massey, MD**,
Leway Chen, MD**,
Charles T. Klodell, MD ,
Juan M. Aranda, MD ,
Nader Moazami, MD ,
Gregory A. Ewald, MD ,
David J. Farrar, PhD ,
O. Howard Frazier, MD|| for the HeartMate II Investigators
* University of Michigan, Ann Arbor, Michigan
Washington Hospital Center, Washington, DC
Johns Hopkins Hospital, Baltimore, Maryland
University of Minnesota, Minneapolis, Minnesota
|| Texas Heart Institute, Houston, Texas
¶ Massachusetts General Hospital, Boston, Massachusetts
# Columbia University, New York, New York
** University of Rochester, Rochester, New York
 University of Florida, Gainesville, Florida
 Barnes-Jewish Hospital, St. Louis, Missouri
 Thoratec Corporation, Pleasanton, California
Manuscript received December 1, 2008;
revised manuscript received March 3, 2009,
accepted March 10, 2009.
* Reprint requests and correspondence to: Dr. Francis D. Pagani, Section of Cardiac Surgery, University of Michigan Health System, Cardiovascular Center, Room 5161, 1500 East Medical Center Drive, SPC#5864, Ann Arbor, Michigan 48109 (Email: fpagani{at}umich.edu).
 |
Abstract
|
|---|
Objectives: This study sought to evaluate the use of a continuous-flow rotary left ventricular assist device (LVAD) as a bridge to heart transplantation.
Background: LVAD therapy is an established treatment modality for patients with advanced heart failure. Pulsatile LVADs have limitations in design precluding their use for extended support. Continuous-flow rotary LVADs represent an innovative design with potential for small size and greater reliability by simplification of the pumping mechanism.
Methods: In a prospective, multicenter study, 281 patients urgently listed (United Network of Organ Sharing status 1A or 1B) for heart transplantation underwent implantation of a continuous-flow LVAD. Survival and transplantation rates were assessed at 18 months. Patients were assessed for adverse events throughout the study and for quality of life, functional status, and organ function for 6 months.
Results: Of 281 patients, 222 (79%) underwent transplantation, LVAD removal for cardiac recovery, or had ongoing LVAD support at 18-month follow-up. Actuarial survival on support was 72% (95% confidence interval: 65% to 79%) at 18 months. At 6 months, there were significant improvements in functional status and 6-min walk test (from 0% to 83% of patients in New York Heart Association functional class I or II and from 13% to 89% of patients completing a 6-min walk test) and in quality of life (mean values improved 41% with Minnesota Living With Heart Failure and 75% with Kansas City Cardiomyopathy questionnaires). Major adverse events included bleeding, stroke, right heart failure, and percutaneous lead infection. Pump thrombosis occurred in 4 patients.
Conclusions: A continuous-flow LVAD provides effective hemodynamic support for at least 18 months in patients awaiting transplantation, with improved functional status and quality of life. (Thoratec HeartMate II Left Ventricular Assist System [LVAS] for Bridge to Cardiac Transplantation; NCT00121472)
Key Words: heart failure left ventricular assist device heart transplantation
|
Abbreviations and Acronyms
| | BTT = bridge-to-transplantation | | FDA = Food and Drug Administration | | LVAD = left ventricular assist device | | MCS = mechanical circulatory support | | NYHA = New York Heart Association |
|
Heart transplantation remains the most successful treatment option for patients with advanced heart failure refractory to medical therapy (1). As a consequence of limited donor availability (1), left ventricular assist device (LVAD) therapy has become an established treatment for patients with advanced heart failure as either a bridge to transplantation (BTT) (2) or as permanent support as an alternative to transplantation (3,4). Historically, patients have been supported by devices engineered with pulsatile design (i.e., HeartMate IP1000, VE, or XVE, Thoratec PVAD or IVAD, Thoratec Corporation, Pleasanton, California; or Novacor, World Heart Corporation, Oakland, California) (2,5–8). These devices are designed with an internal pumping chamber and inflow and outflow valves permitting cyclic filling and emptying with pump actuation elicited by either pneumatic or electrical systems (9). Previous studies have demonstrated the efficacy of these devices with regard to improvement in survival to transplantation (2) and improvement in survival compared with optimal medical management for patients with advanced heart failure, not candidates for transplantation (3).
Pulsatile devices have limitations in their design that preclude their practical use for extended mechanical circulatory support (MCS). These limitations include a large pump size, requirement for extensive surgical dissection for implant, a large body habitus of the recipient, the presence of a large-diameter percutaneous lead for venting air, and audible pump operation (2,7). A critical limitation of the majority of these devices has been the high incidence of reoperation for device exchange for device infection or malfunction (3,10–12).
The REMATCH (Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure) trial demonstrated a survival advantage for LVAD therapy over optimal medical management for patients with advanced heart failure who were not eligible for transplantation. Although demonstrating the potential of MCS in providing improved survival, this trial demonstrated the risk of mechanical failure and device-related complications inherent in the pulsatile HeartMate VE LVAD (3,12). In patients surviving up to 2 years on device support, nearly 65% underwent replacement for infection or malfunction (12).
The development of continuous-flow rotary pump technology represents an innovative design for LVADs (13–18). These devices have the advantage of a smaller pump size and potential for greater mechanical reliability by simplification of the pumping mechanism (13,14). Reports from clinical trials of these newer pump designs have demonstrated efficacy in providing hemodynamic support and favorable risk-to-benefit ratio (15–18).
The HeartMate II LVAD is a continuous-flow rotary pump that has completed a U.S. Food and Drug Administration (FDA)-approved pivotal trial in 133 BTT patients (18). Since this report, 336 additional patients have undergone implantation of the HeartMate II LVAD as of April 2008 through a continued-access protocol approved by the FDA. We report on the first 281 patients entered into this clinical evaluation who have completed study end points or at least 18-month follow-up after LVAD implantation.
 |
Methods
|
|---|
Study design.
Patients were enrolled in the study conducted at 33 centers in the U.S. between March 2005 and April 2008 (18). The study was supervised by the Thoratec Corporation. The clinical affairs and biostatistics departments at Thoratec designed the trial in consultation with the FDA and clinical investigators. Coordinators at each site collected study data, which was forwarded to the data analysis center of the sponsor. The academic authors vouch for the completeness and accuracy of the data and the analyses. A data and safety monitoring board, consisting of 4 independent physicians and 1 biostatistician who were not investigators in the study, met routinely to review study compliance, adverse events, quality of life, and outcomes of patients. These 5 committee members were compensated for their time, but none have any financial interest in the Thoratec Corporation or stand to gain financially from the outcome of the trial. A clinical events committee of 4 independent physicians who were not involved in the conduct of the trial reviewed, classified, and adjudicated the causes of death and all adverse events. The study was conducted in compliance with FDA regulations for Good Clinical Practices. The protocol was approved by the FDA and the institutional review board at each participating center.
Study subjects.
Patients with heart failure who were on a waiting list for heart transplantation at each center were eligible for study enrollment. Patients were required to have symptoms of New York Heart Association (NYHA) functional class IV heart failure and to be ill enough to have high priority for transplantation (United Network for Organ Sharing status 1A or 1B). A complete list of study inclusion and exclusion criteria have been reported (18). All participating patients provided written informed consent before enrolling in the study.
Data collection baseline assessment.
Baseline data were obtained upon patient consent and enrollment into the study. Information collected for baseline data have been previously reported (18).
Continuous-flow pump.
The continuous-flow LVAD used in this study was the HeartMate II LVAS (Thoratec Corporation), which is a rotary pump with axial flow design (Fig. 1) (18). The system design and operating performance of the device have been previously described (18).

View larger version (54K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 1 Components of the Continuous-Flow LVAD
(A) The inflow cannula is inserted into the apex of the left ventricle, and the outflow cannula is anastomosed to the ascending aorta. Blood exits through the left ventricular apex and into the left ventricular assist device (LVAD), which pumps throughout cardiac diastole and systole into the ascending aorta. (B) The LVAD pump is placed within the abdominal wall or peritoneal cavity. A percutaneous lead carries the electrical cable to an electronic controller and battery packs, which are worn on a belt and shoulder holster, respectively.
|
|
Surgical implantation.
Surgical implantation of the HeartMate II LVAS was conducted according to the HeartMate II LVAS "Instructions for Use." Post-operative treatment included initiation of an anticoagulation regimen (18).
Follow-up after device implantation.
Post-operative medical care was managed according to each investigator's usual practice. Details of the protocol for patient follow-up after device implantation have been previously reported (18). Readmissions to the hospital and adverse events were recorded throughout the study as they occurred with the use of standardized definitions (18). All deaths of patients and causes of death were determined at autopsy when possible, examination of medical records, or by interviews with family members. Final adjudication was determined by the clinical events committee. Major device malfunctions were defined as those resulting in death or requiring device replacement.
Outcomes.
The principal outcomes assessed through 18 months after enrollment were the proportions of patients who had undergone transplantation, had undergone explantation of the device because of recovery of ventricular function, or continued with ongoing MCS. The proportion of patients who died on support and those who were withdrawn from the study were also followed for 18 months. Patients who had the original continuous-flow pump replaced with another identical device remained in the study, whereas patients who had the original pump replaced with a different type of device were withdrawn from the study.
Statistical analysis.
Differences between measures of continuous variables before and after implantation were analyzed by a paired t test. McNemar's test was used for comparisons between paired categorical variables. The level of statistical significance was set at p < 0.05. All statistical comparisons were 2 sided. Biochemical and hemodynamic variables are presented as mean (±SD), and median and range were used where appropriate. Discrete variables are presented as percentages. Adverse events are presented both as the percentage of patients who had the event and as event rates per patient-year. Survival analysis for patients continuing on mechanical support was performed with the Kaplan-Meier method. Patients were censored for transplantation, recovery of the natural heart, and withdrawal from the study. The Kaplan-Meier method also was used for analysis of freedom from death due to major device malfunction or need for device replacement for all causes.
 |
Results
|
|---|
Study patients.
A total of 469 patients meeting study-entry criteria were enrolled into the study as of April 2008 and underwent implantation of the continuous-flow LVAD as a BTT. The initial 281 of the 469 patients that completed study end points or have at least 18 months of follow-up with ongoing device support comprise the patients in this report. These 281 patients include additional follow-up of the first 133 patients implanted as part of a primary cohort and previously reported (18). Most of the subjects were men with a median age of 54 years (Table 1). The most frequent heart failure etiology was nonischemic cardiomyopathy. All patients had symptoms of advanced heart failure despite optimal medical management with oral medications. The majority of patients in the study were receiving an intravenous inotrope, with approximately one-third requiring 2 inotropes. Patients not administered inotrope therapy were intolerant to it as the result of ventricular arrhythmias. Forty-five percent of patients were on concomitant support with an intra-aortic balloon pump. Previous biventricular pacing therapy failed in a significant proportion of these patients. Pre-operative hemodynamic and laboratory assessment were consistent with a group of patients with advanced heart failure requiring inotrope and/or intra-aortic balloon pump support.
Outcomes.
Of the 281 patients, 222 (79%) either received a transplant, recovered cardiac function and underwent device explantation, or remained alive with ongoing LVAD support at 18-month follow-up (Fig. 2). At 18 months, 157 (55.8%) patients had received a heart transplant, 58 (20.6%) remained alive with ongoing LVAD support, 56 (19.9%) patients died, 7 (2.5%) patients recovered cardiac function and underwent device explantation, and 3 (1%) patients were withdrawn from the study after device explantation and exchange for another type of LVAD. The median time to transplantation was 118 days (range 10 to 545 days), median time to death was 64 days (range 0 to 797 days), and the median time to pump removal after cardiac recovery was 302 days (range 161 to 558 days). The median duration of support for all patients was 155 days (range 0 to 1,026 days), with a cumulative follow-up of 181 patient-years.

View larger version (28K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 2 Outcomes for 281 Patients After Implantation of the Continuous-Flow Left Ventricular Assist Device
Competing outcomes analysis of patients undergoing implantation of the continuous-flow left ventricular assist device for the first 18 months after device implantation.
|
|
Overall survival for the patients who continued on LVAD support was 82% (95% confidence interval [CI]: 77% to 87%) at 6 months, 73% (95% CI: 66% to 80%) at 1 year, and 72% (95% CI: 65% to 79%) at 18 months (Fig. 3). Of the 157 patients that received a transplant, post-transplant survival was 96% at 30 days and 86% at 1 year. As of June 2008, 42 patients were alive with device support with a median duration of 1.6 years (longest duration 3.1 years). Of these patients, 71% remained active on the transplant list, and 29% were not listed (Table 2). Only 3 patients were not listed for irreversible medical conditions or degree of organ dysfunction.

View larger version (15K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 3 Kaplan-Meier Survival Analysis
Survival analysis for patients who continued to receive support with the continuous-flow left ventricular assist device censored at the time of heart transplantation and device explantation for cardiac recovery.
|
|
Average LVAD estimated blood flow at 6 months of support was 5.6 ± 0.9 l/min (flow index 2.83 ± 0.45 l/min/m2) at a pump speed of 9,467 ± 499 rpm. Left ventricular end-diastolic dimension determined by echocardiography reduced from 69.7 ± 12.3 mm at baseline to 59.2 ± 15.1 mm at 1 week and 56.7 ± 14.5 mm at 6 months. Anticoagulation with warfarin resulted in an average international normalized ratio throughout support of 2.1 ± 0.8 (median 2.0). Twenty-five (8.9%) patients underwent transplantation during their initial hospital stay, and 33 (11.7%) patients died on LVAD support before discharge.
Two hundred twenty (78%) patients were discharged from the hospital with the LVAD, with a median hospital stay after surgery of 25 days (range 8 to 180 days). The median number of days out of hospital before transplantation, readmission, or death was 55.5 (range 1 to 892 days). One hundred forty nine (68%) patients required rehospitalization after discharge, with a median duration of rehospitalization of 5 days (range 0 to 209 days).
Causes of death.
The primary causes of death were sepsis in 11 patients (4%), stroke in 10 (4%; ischemic: n = 5, 2%; hemorrhagic: n = 5, 2%), and right heart failure in 7 (3%). Other causes included device-related deaths in 7 patients (3%), multiorgan failure in 5 (2%), anoxic brain injury in 3 (1%), bleeding in 3 (1%), and other causes in 10 (4%) (cancer, respiratory failure, hyperthermia, air embolism, and unknown). There were a total of 7 device-related deaths, 4 that were attributed to malfunction of implanted components (pump thrombosis in 2, inflow graft that was twisted during implantation in 1, and outflow elbow disconnect in 1) and 3 that were attributed to external components (severed percutaneous lead in 1, power loss in 2). Only 13 of the deaths in the 281 (4.6%) patients (or 23% of the 56 deaths) occurred after 6 months of device support and included 5 from sepsis, 2 from power loss of the LVAD, 2 from withdrawal of support, and 1 each from cancer, respiratory failure, hemorrhage, and unknown.
Adverse events.
Bleeding requiring transfusion and surgery were the most common adverse events (see Miller et al. [18] for definition) in the study and were primarily observed within the first 30 days of device implantation (Table 3). Stroke was observed in 25 (8.9%) patients, with the event rate greatest in the first 30 days. Ischemic strokes were more common than hemorrhagic strokes. Five of the 15 (33%) ischemic strokes occurred within 48 h. Almost one-half of the strokes were fatal (10 of 25; 40%). After the first 30 days, the event rates for ischemic and hemorrhagic stroke were 0.05 and 0.03 events per patient year, respectively. Six (2%) additional patients had transient ischemic attacks that completely reversed. Other nonstroke neurologic events, including seizures and confusion, were observed in 15 (5%) patients. Psychological symptoms developed in 16 (6%) patients. Localized infection not related to the device occurred in 84 (30%) patients. Infection associated with the percutaneous lead was observed in 41 (14%) patients, and there were 5 (2%) pre-peritoneal pump pocket infections. Respiratory and renal failure occurred in 72 (26%) and 30 (11%) patients, respectively. Thirty-six (13%) patients had right heart failure requiring inotrope support for more than 14 days. Of the 53 (19%) patients that developed post-operative right heart failure, survival with continued LVAD support at 18 months was 62 ± 8%. Twenty-six of these 53 (49%) patients with right heart failure received a transplant. Seventeen (6%) patients received temporary support with right ventricular assist devices (median time of support 11.5 days; range 0 to 148 days). The median duration of post-operative inotrope support was 9 days.
A total of 11 (4%) patients underwent 12 HeartMate II LVAD replacements for either device thrombosis (n = 4; 1.4%) (2 of whom subsequently died), surgical complications at the time of LVAD implantation (n = 3; 1.1%), percutaneous lead wire damage (n = 4; 1.4%), or for development of device infection (n = 1; 0.3%). One patient had 2 of the pump replacements (thrombosis and infection). Of the 12 pump replacements, 9 (75%) were replaced with another HeartMate II LVAD, and the patients continued in the study, and 3 (25%) were replaced with another type of LVAD device, and the patients were withdrawn from the study. There were no failures of the mechanical pumping mechanism. The median time to pump replacement was 106 days (range 0 to 672 days). The freedom from major device malfunction resulting in death (n = 4) or device replacement for all causes (malfunction, thrombosis, or infection; 9 without deaths) was 96% (95% CI: 95% to 99%) at 6 months, 93% (95% CI: 90% to 98%) at 1 year, and 92% (95% CI: 88% to 97%) at 18 months (Fig. 4).

View larger version (16K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 4 Kaplan-Meier Analysis of the Freedom From Major Device Malfunction
Analysis of freedom from major device malfunction was defined as the freedom from death as the result of major device malfunction or need for device replacement for all causes including device malfunction, thrombosis, or infection.
|
|
Fifty-nine percent of patients (165 of 281) required 368 operations or procedures after device implantation. The majority of these occurred within 30 days of device implantation (245 of 368; 67%), and most were required for re-explorations or sternal closures for bleeding complications (177 of 245; 72%) followed by temporary RVAD insertion or removal (n = 18; 7%), tracheostomy (n = 12; 5%), implantable cardioverter defibrillator insertion or replacement (n = 9; 4%), infection (n = 7; 3%); pump replacement (n = 4; 2%), and various other cardiac (n = 15; 6%) and noncardiac (n = 3; 1%) procedures. After 30 days, the most frequent indication for reoperation was for infection complications (49 of 123; 40%), bleeding (n = 19; 9%), pump replacement (n = 8; 7%), RVAD insertion or removal (n = 3; 2%), implantable cardioverter defibrillators (n = 5; 4%), and various other noncardiac (n = 24; 20%) and cardiac procedures (n = 14; 11%).
Functional assessment, quality of life, and end-organ function.
Functional assessment with 6-min walk and NYHA functional classification were performed for patients remaining on device support up to 6 months (Table 4). Of 109 patients with paired values at baseline and at 6 months, only 14 of 109 (13%) were able to perform a 6-min walk test at baseline, compared with 97 (89%) patients after 6 months of support, and there was a significant improvement in distance walked between baseline and 6 months.
View this table:
[in this window]
[in a new window]
|
Table 4 Functional Status, Quality of Life, and End-Organ Function for Patients With Paired Data at Baseline and at 6 Months
|
|
Additionally, NYHA functional classification improved from 3.9 ± 0.3 at baseline (with 0% of patients in NYHA functional class I or II) to 1.8 ± 0.7 at 6 months (with 83% in functional class I or II). Quality of life assessed by both Minnesota Living With Heart Failure and the Kansas City Cardiomyopathy Questionnaire was significantly improved at 6 months compared with baseline (Table 4), with mean scores improving over 25 U, or 41% and 75%, respectively. Hepatic (total bilirubin, serum aspartate aminotransaminase, and serum alanine aminotransaminase) and renal (blood urea nitrogen) function significantly improved from baseline to 6 months, but changes in serum creatinine were not statistically significant.
 |
Discussion
|
|---|
Continuous-flow rotary pumps represent a novel design concept for LVADs. Findings from this study have validated the efficacy and safety profile of this design for patients awaiting transplantation. At 18 months after LVAD implantation, a majority of patients (79%) underwent transplantation, had cardiac recovery, or remained alive with ongoing LVAD support. Of the patients remaining alive with ongoing LVAD support, the majority were actively listed for transplant, obtained satisfactory renal and hepatic function and functional status, and were free from significant neurological injury. For patients that received a transplant, 1-year survival after transplant survival was similar (86% at 1 year) to that reported for transplant recipients by the Scientific Registry of Transplant Recipients (1). These data confirm the feasibility of extended LVAD support with a continuous-flow LVAD and importantly provide support for evaluating this technology for permanent MCS as an alternative to transplantation.
Major adverse events and deaths were associated with LVAD implantation (Table 3). Most deaths occurred within the first 3 months and were attributable to stroke, infection, or multiorgan failure. The skewed distribution of mortality and morbidity events early after implantation suggests that the acuity of patient illness contributes significantly to early deaths and complications. Improvements in patient selection, timing of LVAD implantation, and improvements in device technology should lead to improvements in outcomes. The refinement of risk assessment scores for patients being evaluated for LVAD therapy will assist in defining optimal patient selection and timing of LVAD intervention (19,20). Multivariate analyses have shown the importance of pre-existing liver and renal dysfunction, coagulation abnormalities, and lower serum albumin as risk factors that can identify patients who may be too ill for LVAD surgery (19).
There have been no randomized comparisons of implantable LVAD therapy with optimal medical management in patients awaiting transplantation because of the perceived benefits of LVAD therapy in patients at imminent risk of death and the potential ethical issues of withholding life-saving therapy. Further, there have been no randomized comparisons of different types of implantable LVAD designs in this population. Data pertaining to the efficacy and safety profile of a continuous-flow, rotary pump from this study can be placed in perspective when compared historically with previous FDA-approved clinical studies investigating the use of devices with pulsatile design for patients undergoing transplantation.
In these studies, which evaluated >500 patients during a 10-year period, overall survival to transplantation or survival on LVAD support at 6 months was approximately 70% with an average support duration typically of 1 to 3 months and was a consistent finding in each of the clinical studies (2,6,8,21). In a comparison with devices with pulsatile design, the continuous-flow, rotary pump demonstrated at least equal efficacy with regard to hemodynamic support, ability to improve renal and hepatic function, rates of transplantation, and overall patient survival but typically over longer durations of support compared with earlier studies with pulsatile devices. Actuarial survival on LVAD support for patients in this study was 73% at 1 year and 72% at 18 months and appeared superior to survival observed during pump support with a pulsatile device (53% at 1 year) (2).
Importantly, significantly fewer deaths were observed during late follow-up (6 to 18 months) in the present study, suggesting that the incidence of major events contributing to deaths such as stroke, infection, and device malfunction were lower with the new design. This late reduction in deaths has not been previously observed with pulsatile pumps for either BTT (2) or destination therapy (3) and suggests that refinements in device technology have improved late outcomes. Importantly, excellent late survival on LVAD support (12 to 18 months) was maintained in the absence of continuing high rates of attrition to transplantation (Fig. 2), suggesting that significant device complications requiring urgent transplantation were not prevalent.
The rates of major adverse events appear less for continuous-flow rotary pumps when compared with devices with pulsatile design when comparing rates of adverse events with similar definitions to historical trials (2). The adverse event rates (in events per patient-year) of bleeding requiring surgery (0.45 vs. 1.47), percutaneous lead infection (0.26 vs. 3.49), stroke (0.14 vs. 0.44), nonstroke neurological events (0.09 vs. 0.67), and right heart failure requiring a right ventricular assist device (0.09 vs. 0.30) were significantly less in this current study compared with those observed during clinical evaluation of a pulsatile pump, respectively (2). This comparison is historical in nature, and differences in rates of adverse events may have been influenced by differences in acuity of patient illness or improvements in patient management over time. However, these improvements are consistent, at least in part, with improvements in LVAD technology.
Durability and reliability of LVAD design are among the most significant features for extended use of MCS devices. Device replacements in this study were relatively infrequent despite a cumulative support duration of approximately 180 years. When device replacement was required, it was largely for device thrombosis and infection. No mechanical failures of the pumping mechanism were observed, although there was damage to wires in the percutaneous leads and 2 deaths from problems with grafts and connectors.
Recently, a safety advisory regarding the potential for wear and fatigue of the percutaneous lead to occur over time was released by Thoratec Corporation. The advisory provided an estimate of the probability of the need for pump replacement due to percutaneous lead damage to be 1.3% at 12 months, 6.5% at 24 months, and 11.4% at 36 months and also noted percutaneous lead care instructions for patients and caregivers as well as methods to diagnose clinically-relevant wear and fatigue (22). As patients with LVADs become more active and are supported for longer durations, design improvements are being made to improve the durability of the percutaneous lead, and modular designs are under consideration that will permit external exchange of the percutaneous lead in the event of a lead fracture without the need for pump exchange.
The absence of mechanical failures of the pumping mechanism is very encouraging in the prospect of long-term therapy for patients with advanced heart failure. Device exchange is a significant cause of morbidity and mortality (11,12). Data from the present study with a continuous-flow rotary pump demonstrate an important trend toward a low rate of device replacement required from all causes at 18 months. These data suggest that small durable devices can lead to a reduction in infectious and mechanical pump complications and support the concept and feasibility of extended pump support.
The use of continuous-flow rotary pumps has raised a number of concerns. Continuous-flow rotary pumps significantly reduce but do not eliminate the normal pressure amplitude associated with the native circulation that may influence long-term organ function. This initial concern has now been refuted by longer-term follow-up from this present study that has demonstrated preservation of organ function (23). Continuous-flow rotary pumps are currently designed to operate at fixed rotor speeds that do not automatically adjust to changes in left ventricular pre-load. Pre-load changes could limit the capacity of the pump output and patient exercise capacity or alternatively might result in excessive flow in low filling conditions associated with left ventricular collapse and ventricular arrhythmias. However, recent reports have suggested that the hydrodynamic characteristics of the continuous-flow rotary pump used in this clinical study allow flow to increase during exercise and appear to have no or insignificant limitations in patient exercise performance compared with pulsatile pumps (24).
Additionally, although the risk of pump thrombosis and thromboembolism (particularly stroke) is low, it has not been eliminated with the new pump design. Current management strategies for continuous-flow devices require long-term antithrombotic therapy with warfarin, thereby increasing the patient's risk of hemorrhagic events. The optimum level of anticoagulation to minimize both thromobembolic and hemorrhagic risks has not been determined from this present study because there was variability in the anticoagulation management among the participating centers. Although the smaller percutaneous lead in the continuous-flow device has appeared to reduce the rates of device-associated infection, infection remains a potential concern. Additional follow-up is needed to establish the durability and reliability of the pump mechanism over the longer term (>18 months).
Study limitations.
The study was a nonrandomized comparison. Thus, direct comparisons of the current study of a continuous-flow rotary pump to medical therapy or to other studies of pulsatile- or continuous-flow pump designs cannot be performed. Second, the study was performed in a BTT population. There was a significant attrition of patients to transplantation (50%) by 1 year. Although this was the intent of the study, attrition of patients impacts the interpretation of data obtained from the patients remaining on device support as a result of the potential bias from giving priority to transplantation of patients experiencing the least benefit from the device.
 |
Conclusions
|
|---|
A continuous-flow rotary pump LVAD with axial design provides safe, reliable, and effective hemodynamic support in patients awaiting transplantation with improved quality of life and functional capacity. Furthermore, LVAD therapy with continuous-flow rotary pumps with extended support is associated with a very low rate of device malfunction or infection requiring device exchange. Continuous-flow rotary pumps provide a superior alternative to pumps with a pulsatile design in patients awaiting transplantation.
 |
Footnotes
|
|---|
A complete list of study investigators has been previously published (N Engl J Med 2007;357:885–96). Dr. Pagani has received research grant support from Thoratec Corporation, was site PI for the HeartMate LVAD trial, has received research grant support from Terumo Heart Corporation, and was national Co-PI for the DuraHeart LVAD trial. Dr. Miller has received research grant support and honoraria for talks at academic medical centers from Thoratec Corporation. Dr. Russell has been a consultant and received research grant support from Thoratec Corporation. Dr. Aaronson has been a consultant for Thoratec Corporation (reimbursed for travel expenses only). Dr. John has received research support from Thoratec Corporation and has served on the Scientific Advisory Board of Ventracor. Dr. Boyle has been a consultant for Thoratec Corporation. Dr. Conte has been a PI for Thoratec Corporation, ABiomed, and Heartware, and has been a trainer for Thoratec/Abiomed. Dr. Bogaev has been a consultant to Thoratec Corporation. Dr. Naka has been a speaker for Thoratec Corporation and a consultant for Cardiomems. Dr. Farrar is an employee and stock holder of Thoratec Corporation. Drs. Miller and Pagani contributed equally to this work.
 |
References
|
|---|
1. Mulligan MS, Shearon TH, Weill D, Pagani FD, Moore J, Murray S. Heart and lung transplantation in the United States, 1997–2006 Am J Transplant 2008;8:977-987.[CrossRef][Web of Science][Medline]2. Frazier OH, Rose EA, Oz MC, et al. Multicenter clinical evaluation of the HeartMate vented electric left ventricular assist system in patients awaiting heart transplantation J Thorac Cardiovasc Surg 2001;122:1186-1195.[Abstract/Free Full Text] 3. Rose EA, Gelijns AC, Moskowitz AJ, et al. Long-term use of a left ventricular assist device for end-stage heart failure N Eng J Med 2001;345:1435-1443.[CrossRef][Web of Science][Medline] 4. Park SJ, Tector A, Piccioni W, et al. Left ventricular assist devices as destination therapy: a new look at survival J Thorac Cardiovasc Surg 2005;129:9-17.[Abstract/Free Full Text] 5. Joyce LD, Noon GP, Joyce DL, et al. Mechanical circulatory support—a historical review ASAIO J 2004;50:x-xii.[CrossRef][Web of Science][Medline] 6. Slaughter MS, Tsui SS, El-Banayosy A, et al. Results of a multicenter clinical trial with the Thoratec Implantable Ventricular Assist Device J Thorac Cardiovasc Surg 2007;133:1573-1580.[Abstract/Free Full Text] 7. El-Banayosy A, Arusoglu L, Kizner L, et al. Novacor left ventricular assist system versus HeartMate vented electric left ventricular assist system as a long-term mechanical circulatory support device in bridging patients: a prospective study J Thorac Cardiovasc Surg 2000;119:581-587.[Abstract/Free Full Text] 8. Farrar DJ, Hill JD, Pennington DG, et al. Preoperative and postoperative comparison of patients with univentricular and biventricular support with the Thoratec ventricular assist device as a bridge to heart transplantation J Thorac Cardiovasc Surg 1997;113:202-209.[Abstract/Free Full Text] 9. Baughman KL, Jarcho JA. Bridge to life—cardiac mechanical support N Engl J Med 2007;357:846-849.[CrossRef][Web of Science][Medline] 10. Dowling RD, Park SJ, Pagani FD, et al. HeartMate VE LVAS design enhancements and its impact on device reliability Eur J Cardiothorac Surg 2004;25:958-963.[Abstract/Free Full Text] 11. Pagani FD, Long JW, Dembitsky WP, et al. Improved mechanical reliability of the HeartMate XVE Left Ventricular Assist System Ann Thorac Surg 2006;82:1413-1419.[Abstract/Free Full Text] 12. Dembitsky WP, Tector AJ, Park S, et al. Left ventricular assist device performance with long term circulatory support: lessons from the REMATCH trial Ann Thorac Surg 2004;78:2123-2130.[Abstract/Free Full Text] 13. Takatani S. Progress of rotary blood pumps: Presidential Address, International Society for Rotary Blood Pumps 2006, Leuven, Belgium Artif Organs 2007;31:329-344.[CrossRef][Web of Science][Medline] 14. Hoshi H, Shinshi T, Takatani S. Third-generation blood pumps with mechanical noncontact magnetic bearings Artif Organs 2006;30:324-338.[CrossRef][Web of Science][Medline] 15. Goldstein DJ, Zucker M, Arroyo L, et al. Safety and feasibility trial of the MicroMed DeBakey ventricular assist device as a bridge to transplantation J Am Coll Cardiol 2005;45:962-963.[Free Full Text] 16. Esmore D, Spratt P, Larbalestier R, et al. VentrAssist left ventricular assist device: clinical trial results and clinical development plan update Eur J Cardiothorac Surg 2007;32:735-744.[Abstract/Free Full Text] 17. Göttel P, Hetzer R, Schmid C, et al. Clinical results of the first 99 patients with the axial flow pump INCOR J Heart Lung Transplant 2005;24(Suppl 1):S57. 18. Miller LW, Pagani FD, Russell SD, et al. Use of a continuous-flow device in patients awaiting heart transplantation N Engl J Med 2007;357:885-896.[CrossRef][Web of Science][Medline] 19. Lietz K, Long JW, Kfoury AG, et al. Outcomes of left ventricular assist device implantation as destination therapy in the post-REMATCH era: implications for patient selection Circulation 2007;116:497-505.[Abstract/Free Full Text] 20. Matthews JC, Koelling TM, Pagani FD, Aaronson KD. The right ventricular failure risk score a pre-operative tool for assessing the risk of right ventricular failure in left ventricular assist device candidates J Am Coll Cardiol 2008;51:2163-2172.[Abstract/Free Full Text] 21. Frazier OH, Rose EA, McCarthy P, et al. Improved mortality and rehabilitation of transplant candidates treated with a long-term implantable left ventricular assist system Ann Surg 1995;222:327-336.[Web of Science][Medline] 22. United States Food and Drug Administration, MedWatch, Food and Drug Administration Safety Information and Adverse Event Reporting Program, 2008 Safety Alerts for Human Medical Products (Drugs, Biologics, Medical Devices, Special Nutrionals, and Cosmetics)Reported October 2008 http://www.fda.gov/medwAtch/safety/2008/safety08.htm#HeartMate 1995Accessed January 10, 2009. 23. Radovancevic B, Vrtovec B, de Kort E, et al. End-organ function in patients on long-term circulatory support with continuous- or pulsatile-flow assist devices J Heart Lung Transplant 2007;26:815-818.[CrossRef][Web of Science][Medline] 24. Haft J, Armstrong W, Dyke DB, et al. Hemodynamic and exercise performance with pulsatile and continuous flow left ventricular assist devices Circulation 2007;116(Suppl I):I8-I15.[CrossRef][Web of Science][Medline]
Related Articles
-
Continuous Flow Rotary Left Ventricular Assist Device: Mechanical Circulatory Support 2.0
- Howard J. Eisen and Shelley R. Hankins
J. Am. Coll. Cardiol. 2009 54: 322-324.
[Full Text]
[PDF]
-
Inside This Issue
J. Am. Coll. Cardiol. 2009 54: A28.
[Full Text]
[PDF]
This article has been cited by other articles:

|
 |

|
 |
 
S. E. A. Felix, J. R. Martina, J. H. Kirkels, C. Klopping, H. Nathoe, E. Sukkel, N. Hulstein, F. Z. Ramjankhan, P. A. F. M. Doevendans, J. R. Lahpor, et al.
Continuous-flow left ventricular assist device support in patients with advanced heart failure: points of interest for the daily management
Eur J Heart Fail,
February 3, 2012;
(2012)
hfs012v1.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Hasin, Y. Topilsky, J. A. Schirger, Z. Li, Y. Zhao, B. A. Boilson, A. L. Clavell, R. J. Rodeheffer, R. P. Frantz, B. S. Edwards, et al.
Changes in Renal Function After Implantation of Continuous-Flow Left Ventricular Assist Devices
J. Am. Coll. Cardiol.,
January 3, 2012;
59(1):
26 - 36.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. J. Spurlock, K. Koch, D. E. Mazur, E. M. Fracz, R. H. Bartlett, and J. W. Haft
Preliminary In Vivo Testing of a Novel Pump for Short-Term Extracorporeal Life Support
Ann. Thorac. Surg.,
January 1, 2012;
93(1):
141 - 146.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Suarez, C. B. Patel, G. M. Felker, R. Becker, A. F. Hernandez, and J. G. Rogers
Mechanisms of Bleeding and Approach to Patients With Axial-Flow Left Ventricular Assist Devices
Circ Heart Fail,
November 1, 2011;
4(6):
779 - 784.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. John, F. Kamdar, P. Eckman, M. Colvin-Adams, A. Boyle, S. Shumway, L. Joyce, and K. Liao
Lessons Learned From Experience With Over 100 Consecutive HeartMate II Left Ventricular Assist Devices
Ann. Thorac. Surg.,
November 1, 2011;
92(5):
1593 - 1600.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Brouwers, J. Denollet, N. de Jonge, K. Caliskan, J. Kealy, and S. S. Pedersen
Patient-Reported Outcomes in Left Ventricular Assist Device Therapy: A Systematic Review and Recommendations for Clinical Research and Practice
Circ Heart Fail,
November 1, 2011;
4(6):
714 - 723.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. John, Y. Naka, N. G. Smedira, R. Starling, U. Jorde, P. Eckman, D. J. Farrar, and F. D. Pagani
Continuous Flow Left Ventricular Assist Device Outcomes in Commercial Use Compared With the Prior Clinical Trial
Ann. Thorac. Surg.,
October 1, 2011;
92(4):
1406 - 1413.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Zahr, E. Genovese, M. Mathier, M. Shullo, K. Lockard, R. Zomak, D. McNamara, Y. Toyoda, R. L. Kormos, and J. J. Teuteberg
Obese Patients and Mechanical Circulatory Support: Weight Loss, Adverse Events, and Outcomes
Ann. Thorac. Surg.,
October 1, 2011;
92(4):
1420 - 1426.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Krabatsch, E. Potapov, A. Stepanenko, M. Schweiger, M. Kukucka, M. Huebler, E. Hennig, and R. Hetzer
Biventricular Circulatory Support With Two Miniaturized Implantable Assist Devices
Circulation,
September 13, 2011;
124(11_suppl_1):
S179 - S186.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. S. Kiernan, D. T. Pham, D. DeNofrio, and N. K. Kapur
Management of HeartWare left ventricular assist device thrombosis using intracavitary thrombolytics
J. Thorac. Cardiovasc. Surg.,
September 1, 2011;
142(3):
712 - 714.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. Brassard, A. S. Jensen, N. Nordsborg, F. Gustafsson, J. E. Moller, C. Hassager, S. Boesgaard, P. B. Hansen, P. S. Olsen, K. Sander, et al.
Central and Peripheral Blood Flow During Exercise With a Continuous-Flow Left Ventricular Assist Device: Constant Versus Increasing Pump Speed: A Pilot Study
Circ Heart Fail,
September 1, 2011;
4(5):
554 - 560.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. M. Adamson, M. Stahovich, S. Chillcott, S. Baradarian, J. Chammas, B. Jaski, P. Hoagland, and W. Dembitsky
Clinical Strategies and Outcomes in Advanced Heart Failure Patients Older Than 70 Years of Age Receiving the HeartMate II Left Ventricular Assist Device: A Community Hospital Experience
J. Am. Coll. Cardiol.,
June 21, 2011;
57(25):
2487 - 2495.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. N. Hong, A. Iribarne, J. Yang, B. Ramlawi, H. Takayama, Y. Naka, and M. J. Russo
Do Posttransplant Outcomes Differ in Heart Transplant Recipients Bridged With Continuous and Pulsatile Flow Left Ventricular Assist Devices?
Ann. Thorac. Surg.,
June 1, 2011;
91(6):
1899 - 1906.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Kennon and A. Jain
Circulatory support in severe aortic stenosis
Heart,
May 15, 2011;
97(10):
783 - 784.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. C. Starling, Y. Naka, A. J. Boyle, G. Gonzalez-Stawinski, R. John, U. Jorde, S. D. Russell, J. V. Conte, K. D. Aaronson, E. C. McGee Jr, et al.
Results of the Post-U.S. Food and Drug Administration-Approval Study With a Continuous Flow Left Ventricular Assist Device as a Bridge to Heart Transplantation: A Prospective Study Using the INTERMACS (Interagency Registry for Mechanically Assisted Circulatory Support)
J. Am. Coll. Cardiol.,
May 10, 2011;
57(19):
1890 - 1898.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. L. Williams, J. R. Trivedi, K. C. McCants, S. D. Prabhu, E. J. Birks, L. Oliver, and M. S. Slaughter
Heart Transplant vs Left Ventricular Assist Device in Heart Transplant-Eligible Patients
Ann. Thorac. Surg.,
May 1, 2011;
91(5):
1330 - 1334.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Singh, M. Shullo, R. L. Kormos, K. Lockard, R. Zomak, M. A. Simon, C. Bermudez, J. Bhama, D. McNamara, Y. Toyoda, et al.
Impact of Renal Function Before Mechanical Circulatory Support on Posttransplant Renal Outcomes
Ann. Thorac. Surg.,
May 1, 2011;
91(5):
1348 - 1354.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. W. Miller
Left Ventricular Assist Devices Are Underutilized
Circulation,
April 12, 2011;
123(14):
1552 - 1558.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. C. Stewart and L. W. Stevenson
Keeping Left Ventricular Assist Device Acceleration on Track
Circulation,
April 12, 2011;
123(14):
1559 - 1568.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. H. Adams, J. Chikwe, F. Filsoufi, and A. C. Anyanwu
The Year in Cardiovascular Surgery
J. Am. Coll. Cardiol.,
March 29, 2011;
57(13):
1425 - 1444.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. V. Pamboukian
Mechanical Circulatory Support: We Are Halfway There
J. Am. Coll. Cardiol.,
March 22, 2011;
57(12):
1383 - 1385.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Goda, H. Takayama, S.-W. Pak, N. Uriel, D. Mancini, Y. Naka, and U. P. Jorde
Aortic Valve Procedures at the Time of Ventricular Assist Device Placement
Ann. Thorac. Surg.,
March 1, 2011;
91(3):
750 - 754.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. R. Ensor, C. A. Paciullo, W. D. Cahoon Jr, and P. E. Nolan Jr
Pharmacotherapy for Mechanical Circulatory Support: A Comprehensive Review
Ann. Pharmacother.,
January 1, 2011;
45(1):
60 - 77.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E. J. Birks and M. S. Slaughter
51 Ventricular assist devices, including intra-aortic balloon pumps
Oxford Textbook of Heart Failure,
January 1, 2011;
1(1):
med-9780199577729-chapter - med-9780199577729-chapter.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
In the Literature
Clinical Infectious Diseases,
December 15, 2010;
51(12):
iii - iii.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. E. Achneck, B. Sileshi, A. Parikh, C. A. Milano, I. J. Welsby, and J. H. Lawson
Pathophysiology of Bleeding and Clotting in the Cardiac Surgery Patient: From Vascular Endothelium to Circulatory Assist Device Surface
Circulation,
November 16, 2010;
122(20):
2068 - 2077.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. V. Letsou, T. D. Pate, J. R. Gohean, M. Kurusz, R. G. Longoria, L. Kaiser, and R. W. Smalling
Improved left ventricular unloading and circulatory support with synchronized pulsatile left ventricular assistance compared with continuous-flow left ventricular assistance in an acute porcine left ventricular failure model
J. Thorac. Cardiovasc. Surg.,
November 1, 2010;
140(5):
1181 - 1188.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E. Z. Gorodeski, E. C. Chu, C. H. Chow, W. C. Levy, E. Hsich, and R. C. Starling
Application of the Seattle Heart Failure Model in Ambulatory Patients Presented to an Advanced Heart Failure Therapeutics Committee
Circ Heart Fail,
November 1, 2010;
3(6):
706 - 714.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Authors/Task Force Members, K. Dickstein, P. E. Vardas, A. Auricchio, J.-C. Daubert, C. Linde, J. McMurray, P. Ponikowski, S. G. Priori, R. Sutton, et al.
2010 Focused Update of ESC Guidelines on device therapy in heart failure: An update of the 2008 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure and the 2007 ESC guidelines for cardiac and resynchronization therapy Developed with the special contribution of the Heart Failure Association and the European Heart Rhythm Association
Eur. Heart J.,
November 1, 2010;
31(21):
2677 - 2687.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Authors/Task Force Members, K. Dickstein, P. E. Vardas, A. Auricchio, J.-C. Daubert, C. Linde, J. McMurray, P. Ponikowski, S. G. Priori, R. Sutton, et al.
2010 Focused Update of ESC Guidelines on device therapy in heart failure: An update of the 2008 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure and the 2007 ESC Guidelines for cardiac and resynchronization therapy * Developed with the special contribution of the Heart Failure Association and the European Heart Rhythm Association
Eur J Heart Fail,
November 1, 2010;
12(11):
1143 - 1153.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Authors/Task Force Members, K. Dickstein, P. E. Vardas, A. Auricchio, J.-C. Daubert, C. Linde, J. McMurray, P. Ponikowski, S. G. Priori, R. Sutton, et al.
2010 Focused Update of ESC Guidelines on device therapy in heart failure: An update of the 2008 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure and the 2007 ESC Guidelines for cardiac and resynchronization therapy * Developed with the special contribution of the Heart Failure Association and the European Heart Rhythm Association
Europace,
November 1, 2010;
12(11):
1526 - 1536.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N. Uriel, S.-W. Pak, U. P. Jorde, B. Jude, S. Susen, A. Vincentelli, P.-V. Ennezat, S. Cappleman, Y. Naka, and D. Mancini
Acquired von Willebrand Syndrome After Continuous-Flow Mechanical Device Support Contributes to a High Prevalence of Bleeding During Long-Term Support and at the Time of Transplantation
J. Am. Coll. Cardiol.,
October 5, 2010;
56(15):
1207 - 1213.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. W. Miller
The Development of the von Willebrand Syndrome With the Use of Continuous Flow Left Ventricular Assist Devices: A Cause-and-Effect Relationship
J. Am. Coll. Cardiol.,
October 5, 2010;
56(15):
1214 - 1215.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
V. K. Topkara, S. Kondareddy, F. Malik, I.- W. Wang, D. L. Mann, G. A. Ewald, and N. Moazami
Infectious Complications in Patients With Left Ventricular Assist Device: Etiology and Outcomes in the Continuous-Flow Era
Ann. Thorac. Surg.,
October 1, 2010;
90(4):
1270 - 1277.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E. I. Charitos and H.-H. Sievers
Antithrombotic therapy in patients with left ventricular assist devices: a critical view of the data and lessons from the past
Interact CardioVasc Thorac Surg,
October 1, 2010;
11(4):
505 - 505.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. A. Thompson, K. J. Philip, S. Simsir, and E. R. Schwarz
Review: The New Concept of ''Interventional Heart Failure Therapy'': Part 2--Inotropes, Valvular Disease, Pumps, and Transplantation
Journal of Cardiovascular Pharmacology and Therapeutics,
September 1, 2010;
15(3):
231 - 243.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
D. G. Jakovljevic, R. S. George, D. Nunan, G. Donovan, R. S. Bougard, M. H. Yacoub, E. J. Birks, and D. A. Brodie
The impact of acute reduction of continuous-flow left ventricular assist device support on cardiac and exercise performance
Heart,
September 1, 2010;
96(17):
1390 - 1395.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. A. Miller, K. Ulisney, and J. T. Baldwin
INTERMACS (Interagency Registry for Mechanically Assisted Circulatory Support): A New Paradigm for Translating Registry Data Into Clinical Practice
J. Am. Coll. Cardiol.,
August 24, 2010;
56(9):
738 - 740.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. Mancini and K. Lietz
Selection of Cardiac Transplantation Candidates in 2010
Circulation,
July 13, 2010;
122(2):
173 - 183.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Ohata, H. Ueda, and Y. Yamada
Cibenzoline intoxication necessitating implantation of a biventricular assist system in a patient with severe cardiomyopathy
Interact CardioVasc Thorac Surg,
July 1, 2010;
11(1):
95 - 97.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N. G. Smedira, K. J. Hoercher, D. Y. Yoon, J. Rajeswaran, L. Klingman, R. C. Starling, and E. H. Blackstone
Bridge to transplant experience: Factors influencing survival to and after cardiac transplant
J. Thorac. Cardiovasc. Surg.,
May 1, 2010;
139(5):
1295 - 1305.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. L. Kormos, J. J. Teuteberg, F. D. Pagani, S. D. Russell, R. John, L. W. Miller, T. Massey, C. A. Milano, N. Moazami, K. S. Sundareswaran, et al.
Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: Incidence, risk factors, and effect on outcomes
J. Thorac. Cardiovasc. Surg.,
May 1, 2010;
139(5):
1316 - 1324.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. H. Lund, J. Matthews, and K. Aaronson
Patient selection for left ventricular assist devices
Eur J Heart Fail,
May 1, 2010;
12(5):
434 - 443.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. G. Rogers, K. D. Aaronson, A. J. Boyle, S. D. Russell, C. A. Milano, F. D. Pagani, B. S. Edwards, S. Park, R. John, J. V. Conte, et al.
Continuous Flow Left Ventricular Assist Device Improves Functional Capacity and Quality of Life of Advanced Heart Failure Patients
J. Am. Coll. Cardiol.,
April 27, 2010;
55(17):
1826 - 1834.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
W.H. W. Tang and G. S. Francis
The Year in Heart Failure
J. Am. Coll. Cardiol.,
February 16, 2010;
55(7):
688 - 696.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. N. DeMaria, J. J. Bax, O. Ben-Yehuda, G. K. Feld, B. H. Greenberg, J. Hall, M. Hlatky, W. Y.W. Lew, J. A.C. Lima, A. S. Maisel, et al.
Highlights of the Year in JACC 2009
J. Am. Coll. Cardiol.,
January 26, 2010;
55(4):
380 - 407.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. W. Miller
Is Pulsatile Blood Flow No Longer Essential?
Circulation,
December 8, 2009;
120(23):
2313 - 2314.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. D. Pal, C. T. Klodell, R. John, F. D. Pagani, J. G. Rogers, D. J. Farrar, C. A. Milano, and for the HeartMate II Clinical Investigators
Low Operative Mortality With Implantation of a Continuous-Flow Left Ventricular Assist Device and Impact of Concurrent Cardiac Procedures
Circulation,
September 15, 2009;
120(11_suppl_1):
S215 - S219.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. J. Eisen and S. R. Hankins
Continuous Flow Rotary Left Ventricular Assist Device: Mechanical Circulatory Support 2.0
J. Am. Coll. Cardiol.,
July 21, 2009;
54(4):
322 - 324.
[Full Text]
[PDF]
|
 |
|
|