J Am Coll Cardiol, 2009; 53:1855-1858, doi:10.1016/j.jacc.2008.07.075
© 2009 by the American College of Cardiology Foundation
This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow View Related Genuine Article on CVN
Right arrow View Related Cardiosource Journal Scan
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (7)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Webb, J. G.
Right arrow Articles by Boone, R. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Webb, J. G.
Right arrow Articles by Boone, R. A.
Related Collections
Right arrowRelated Article

FOCUS ISSUE: VALVULAR HEART DISEASE: CLINICAL RESEARCH

A New Transcatheter Aortic Valve and Percutaneous Valve Delivery System

John G. Webb, MD*, Lukas Altwegg, MD, Jean-Bernard Masson, MD, Saad Al Bugami, MD, Abdullah Al Ali, MD and Robert A. Boone, MD

Division of Cardiology, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada

Manuscript received April 29, 2008; revised manuscript received July 15, 2008, accepted July 21, 2008.

* Reprint requests and correspondence: Dr. John G. Webb, McLeod Professor of Heart Valve Intervention, St. Paul's Hospital, 1081 Burrard Street, Vancouver, British Columbia, Canada V6Z 1Y6 (Email: webb{at}providencehealth.bc.ca).


    Abstract
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusion
 References
 
Objectives: We describe procedural and clinical outcomes in a high-risk cohort undergoing transcatheter aortic valve replacement with early next-generation transcatheter valve and delivery systems.

Background: Percutaneous aortic valve replacement is gaining acceptance as a viable option in patients at high surgical risk. Broader application will require further advances in valve and delivery system technology.

Methods: Transarterial aortic valve replacement was attempted in 25 patients (mean age 85 years) determined to be at high surgical risk due to comorbidities. A new delivery catheter system (RetroFlex 2, Edwards Lifesciences, Irving, California) was utilized in combination with either a balloon-expandable SAPIEN (Edwards Lifesciences) valve or a next-generation low-profile cobalt-chromium bovine pericardial SAPIEN XT (Edwards Lifesciences) valve.

Results: Percutaneous valve replacement was successful in all 25 high-risk patients. Aortic valve area increased from 0.59 ± 0.15 cm2 to 1.60 ± 0.27 cm2. In this high-risk cohort (Society of Thoracic Surgeons and logistic EuroSCORE estimates of surgical mortality were 8.9% and 21.0%, respectively), 30-day mortality was 0%.

Conclusions: Technical and procedural advances in catheter systems and prosthetic valves designed for percutaneous aortic valve delivery may contribute to increased procedural success and improved clinical outcomes.

Key Words: aortic stenosis • percutaneous • valve • valvuloplasty

Abbreviations and Acronyms
  AVR = aortic valve replacement
  IQR = interquartile range


Initial attempts at percutaneous aortic valve replacement (AVR) were compromised due to the difficulties associated with vascular passage of a relatively bulky prosthetic valve compressed onto a standard valvuloplasty catheter (1,2). Subsequent technical and procedural improvements allowed the development of a more predictable therapy (3). Although outcomes continue to improve, limitations (aortic injury, atheroembolism, difficulty crossing the native valve) remain apparent (4). We describe a new delivery system and next-generation balloon-expandable valve in a case series of 25 high-risk patients undergoing transarterial AVR.


    Methods
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusion
 References
 
Percutaneous AVR was approved for clinical use in patients with symptomatic aortic stenosis in whom the risk associated with open heart surgery was considered prohibitive by a team of cardiologists and cardiac surgeons. Written informed consent was obtained. Patients were excluded if the diameter of the aortic annulus was <18 or >26 mm as assessed by transesophageal echocardiography, if there was severe iliofemoral arterial disease, or if a reasonable quality or duration of life was considered unlikely.

Procedures were performed in a catheterization laboratory under general anesthesia. Percutaneous femoral artery puncture was utilized followed by percutaneous suture "pre-closure" (Prostar, Abbott, Boston, Massachusetts) or surgical cutdown. The basic procedure of transarterial AVR has been previously described (3,4).

Delivery system.   The RetroFlex 2 delivery system (Edwards Lifesciences, Irvine, California) incorporates a coaxial nose cone catheter, a balloon catheter, and a deflectable pusher catheter (Fig. 1). The catheter is advanced through the sheath and into the aorta. An articulation control knob allows steering around the aortic arch and through the central orifice of the stenotic native valve. The deflectable pusher catheter is retracted, and the nose cone is advanced so as to expose the deployment balloon (Fig. 2). The prosthesis is positioned within the native valve. The deployment balloon is inflated and deflated while right ventricular burst pacing is used to temporarily reduce transvalvular flow. The balloon is utilized to guide the withdrawal of the nose cone through the newly implanted valve. The nose cone and pusher catheter are approximated and removed through the delivery sheath.


Figure 1
View larger version (87K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 1 Diagrammatic Representation Showing the RetroFlex 2 Catheter System

The catheter system is shown during passage through the aorta (left) and with the nosecone advanced and flexion catheter withdrawn to allow expansion of the deployment balloon (right).

 

Figure 2
View larger version (16K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 2 Photograph of RetroFlex 2 Catheter

The catheter is shown during introduction (top) and with the balloon exposed to allow deployment of the prosthesis (bottom).

 
Valves.   Initial reports of balloon-expandable valve replacement utilized a valve constructed from a stainless steel tubular frame with equine pericardial leaflets and a fabric sealing cuff (Cribier-Edwards, Edwards Lifesciences). This basic valve design was subsequently modified utilizing bovine pericardial leaflets and lengthening the sealing cuff without modification of the frame (SAPIEN, Edwards Lifesciences).

A next-generation transcatheter heart valve (SAPIEN XT, Edwards Lifesciences) was utilized in 3 patients. A cobalt-chromium frame permits thinner struts without loss of structural integrity. Thinner struts and a more open design allow for a lower crimped profile (Fig. 3), while maintaining the valve's radial stiffness. Bovine pericardial leaflets are matched for thickness and elasticity and incorporate Thermafix anticalcification treatment. The scalloped geometry and attachment method of the leaflets have been modified to achieve a naturally closed design and enhance valve durability (Fig. 4).


Figure 3
View larger version (80K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 3 Low-Profile Next-Generation SAPIEN XT Valve

The thin cobalt-chromium struts and relatively open frame allow a low delivery profile without loss of radial strength (A). Unlike the SAPIEN valve, the scalloped leaflets are closed at rest. The scalloped geometry of the bovine pericardial leaflets results in superior durability as assessed by in vitro testing (B).

 

Figure 4
View larger version (105K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 4 Low-Profile Next-Generation Cobalt-Chromium SAPIEN XT Valve Explanted at 20 Weeks From a Sheep

Tissue growth over the inflow portion of the frame can be seen.

 
The femoral arterial sheath required is determined by the crimped diameter of the prosthetic valve and its delivery system. A 23-mm SAPIEN valve crimped on the RetroFlex 2 delivery catheter requires a 22-F internal diameter sheath, whereas a 26-mm SAPIEN valve requires a 24-F sheath. When crimped on the current RetroFlex 2 catheter, the 26-mm next-generation low-profile valve allows use of a 22-F sheath.

Statistics.   Continuous variables are presented as mean ± SD, medians, and interquartile ranges (IQRs), as appropriate. Categorical variables are stated as frequencies and percentages. Normality was tested with the Kolmogorov-Smirnov goodness-of-fit test. For comparison of continuous variables before and after transarterial AVR, the Student t test was used, and for comparison of groups, a 1-way analysis of variance F test was utilized. Categorical variables were compared by the Fisher exact test. A 2-sided value of p < 0.05 was considered statistically significant.


    Results
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusion
 References
 
The RetroFlex 2 system was utilized in 25 patients. Mean age was 85 years (IQR: 79 to 88 years). Patients had multiple comorbidities, including prior thoracotomy (28%), renal insufficiency (40%), porcelain aorta (12%), grade 3 or 4 mitral regurgitation (12%), severe lung disease (24%), and severe frailty (40%). Baseline characteristics are presented in Table 1.


View this table:
[in this window]
[in a new window]

 
Table 1 Baseline Characteristics of 25 High-Risk Patients
 
Procedural success, defined as implantation of a prosthesis at the intended site, was achieved in all patients. Tortuosity of the thoracic aorta, horizontal aortic roots, and severely stenotic, heavily calcified valves, although problematic with earlier delivery systems, were relatively easily managed with the new delivery system. Percutaneous needle puncture was utilized for femoral arterial access in all patients and percutaneous suture closure in the majority. Outcomes are presented in Table 2. Society of Thoracic Surgeons and logistic EuroSCORE estimates of operative 30-day mortality were 8.9% and 21.0%, respectively. All patients remained alive at 30 days.


View this table:
[in this window]
[in a new window]

 
Table 2 Outcome at 30 Days
 
The SAPIEN valve was utilized in 22 patients and the next generation low profile cobalt-chromium SAPIEN XT valve in 3. At discharge, echocardiographic aortic mean gradient had fallen from 49.3 ± 17.9 mm Hg to 10.6 ± 2.9 mm Hg and effective orifice area had risen from 0.59 ± 0.15 cm2 to 1.6 ± 0.27 cm2 (Table 3). No patient had more than mild valvular regurgitation, and only 1 patient had more than mild paravalvular regurgitation (graded moderate). All patients had normal prosthetic valve function at 1-month echocardiographic follow-up.


View this table:
[in this window]
[in a new window]

 
Table 3 Echocardiographic Valve Function
 

    Discussion
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusion
 References
 
Transcatheter valve replacement continues to evolve rapidly. In the initial experience with transarterial AVR, procedural success was achieved in 78% of patients with a 30-day mortality of 11% (3). The current success rate of 100% and 30-day mortality of 0% in 25 high-risk patients is encouraging. Technological and procedural improvements as well as increasing experience likely share responsibility for these improving outcomes.

Although mortality was not observed, morbidity was. The major source of morbidity was related to vascular access. Pre-procedural screening and improved vascular management appear to be reducing vascular morbidity and mortality risk (4). Consequently, a major thrust of current development is to reduce the diameter of the valve and delivery system. The development of a new, thinner cobalt-chromium SAPIEN XT valve frame with a more open design allows for tighter crimping of the valve and an improved delivery profile. Further modification of the delivery system to accommodate this next-generation valve is expected to reduce the delivery profile to 18- or 19-F (Fig. 5). Such a reduction in profile will likely increase the population of patients amenable to this therapy and improve outcomes further. The thinner struts of the next-generation cobalt-chromium low-profile device did not appear to result in a reduction in orifice area (Fig. 6). We have not observed structural valve failure in our larger experience with almost 200 valve implants. Comparable or improved durability with the next-generation valve is anticipated by in vitro testing but remains to be clinically demonstrated.


Figure 5
View larger version (26K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 5 Photograph of RetroFlex 3 Catheter

More recently the RetroFlex 3 catheter has seen early clinical use. This system incorporates a shorter fixed nose cone for ease of use and safety. Future modifications in RetroFlex 4 will reduce profile further.

 

Figure 6
View larger version (141K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 6 Fluoroscopic Appearance of the Low-Profile Cobalt-Chromium Next-Generation SAPIEN XT Valve

Aortic angiography shows optimal positioning with no evidence of paravalvular regurgitation (left). The valve can be seen to be fully and symmetrically expanded (right).

 

    Conclusion
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusion
 References
 
Advances in transcatheter valves, delivery systems, and procedural technique may result in improved outcomes and suggest the potential for a broader role of this therapeutic modality.


    Footnotes
 
Dr. Webb is a consultant to Edwards Lifesciences Inc.


    References
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusion
 References
 
1. Cribier A, Eltchaninoff H, Tron C, et al. Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosis J Am Coll Cardiol 2004;43:698-703.[Abstract/Free Full Text]

2. Cribier A, Eltchaninoff H, Bash A, et al. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description Circulation 2002;106:3006-3008.[Abstract/Free Full Text]

3. Webb JG, Chandavimol M, Thompson C, et al. Percutaneous aortic valve implantation retrograde from the femoral artery Circulation 2006;113:842-850.[Abstract/Free Full Text]

4. Webb JG, Pasupati SJ, Humphries K, et al. Percutaneous transarterial aortic valve replacement in selected high risk patients with aortic stenosis Circulation 2007;116:755-763.[Abstract/Free Full Text]

5. Nashef SA, Roques F, Hammill BG, et al. Validation of European System for Cardiac Operative Risk Evaluation (EuroSCORE) in North American cardiac surgery Eur J Cardiothorac Surg 2002;22:101-105.[Abstract/Free Full Text]


Related Article

Inside This Issue
J. Am. Coll. Cardiol. 2009 53: A28. [Full Text] [PDF]



This article has been cited by other articles:


Home page
J Am Coll CardiolHome page
R. Gurvitch, A. Cheung, J. Ye, D. A. Wood, A. B. Willson, S. Toggweiler, R. Binder, and J. G. Webb
Transcatheter Valve-in-Valve Implantation for Failed Surgical Bioprosthetic Valves.
J. Am. Coll. Cardiol., November 15, 2011; 58(21): 2196 - 2209.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
A. C. T. Ng, K.-H. Yiu, S. H. Ewe, F. van der Kley, M. Bertini, A. de Weger, A. de Roos, D. Y. Leung, J. D. Schuijf, M. J. Schalij, et al.
Influence of left ventricular geometry and function on aortic annular dimensions as assessed with multi-detector row computed tomography: implications for transcatheter aortic valve implantation
Eur. Heart J., November 2, 2011; 32(22): 2806 - 2813.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
R. Gurvitch, J. G. Webb, R. Yuan, M. Johnson, C. Hague, A. B. Willson, S. Toggweiler, D. A. Wood, J. Ye, R. Moss, et al.
Aortic Annulus Diameter Determination by Multidetector Computed Tomography: Reproducibility, Applicability, and Implications for Transcatheter Aortic Valve Implantation
J. Am. Coll. Cardiol. Intv., November 1, 2011; 4(11): 1235 - 1245.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
J. L. Zamorano, L. P. Badano, C. Bruce, K.-L. Chan, A. Goncalves, R. T. Hahn, M. G. Keane, G. La Canna, M. J. Monaghan, P. Nihoyannopoulos, et al.
EAE/ASE recommendations for the use of echocardiography in new transcatheter interventions for valvular heart disease
Eur. Heart J., September 1, 2011; 32(17): 2189 - 2214.
[Abstract] [Full Text] [PDF]


Home page
Journal of Medical Marketing: Device, Diagnostic and Pharmacetical MarketingHome page
R. Tarricone and M. Drummond
Challenges in the clinical and economic evaluation of medical devices: The case of transcatheter aortic valve implantation
Journal of Medical Marketing: Device, Diagnostic and Pharmacetical Marketing, August 1, 2011; 11(3): 221 - 229.
[Abstract] [PDF]


Home page
Eur Heart J Cardiovasc ImagingHome page
J. L. Zamorano, L. P. Badano, C. Bruce, K.-L. Chan, A. Goncalves, R. T. Hahn, M. G. Keane, G. La Canna, M. J. Monaghan, P. Nihoyannopoulos, et al.
EAE/ASE recommendations for the use of echocardiography in new transcatheter interventions for valvular heart disease
Eur Heart J Cardiovasc Imaging, August 1, 2011; 12(8): 557 - 584.
[Abstract] [Full Text] [PDF]


Home page
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in MedicineHome page
E. E. Roselli
Engineering Transcatheter Aortic Valve Devices: A Surgeon's Perspective
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, March 1, 2011; 225(3): 207 - 219.
[Abstract] [PDF]


Home page
J Am Coll Cardiol IntvHome page
C. Godino, F. Maisano, M. Montorfano, A. Latib, A. Chieffo, I. Michev, R. Al-Lamee, M. Bande, M. Mussardo, F. Arioli, et al.
Outcomes After Transcatheter Aortic Valve Implantation With Both Edwards-SAPIEN and CoreValve Devices in a Single Center: The Milan Experience
J. Am. Coll. Cardiol. Intv., November 1, 2010; 3(11): 1110 - 1121.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
R. Gurvitch, D. A. Wood, E. L. Tay, J. Leipsic, J. Ye, S. V. Lichtenstein, C. R. Thompson, R. G. Carere, N. Wijesinghe, F. Nietlispach, et al.
Transcatheter Aortic Valve Implantation: Durability of Clinical and Hemodynamic Outcomes Beyond 3 Years in a Large Patient Cohort
Circulation, September 28, 2010; 122(13): 1319 - 1327.
[Abstract] [Full Text] [PDF]


Home page
ANN INTERN MEDHome page
R. R. Coeytaux, J. W. Williams Jr., R. N. Gray, and A. Wang
Percutaneous Heart Valve Replacement for Aortic Stenosis: State of the Evidence
Ann Intern Med, September 7, 2010; 153(5): 314 - 324.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
H. Jilaihawi, R. Bonan, A. Asgar, R. Ibrahim, T. Spyt, D. Chin, and J. Kovac
Anatomic Suitability for Present and Next Generation Transcatheter Aortic Valve Prostheses: Evidence for a Complementary Multidevice Approach to Treatment
J. Am. Coll. Cardiol. Intv., August 1, 2010; 3(8): 859 - 866.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
S. R. Dixon, C. L. Grines, and W. W. O'Neill
The Year in Interventional Cardiology
J. Am. Coll. Cardiol., May 18, 2010; 55(20): 2272 - 2286.
[Full Text] [PDF]


Home page
Oxford Textbook of Interventional CardiologyHome page
J. G. Webb and F. Nietlispach
Chapter 39 Transcatheter aortic valve replacement
Oxford Textbook of Interventional Cardiology, January 1, 2010; 1(1): med-9780199569083-chapter - med-9780199569083-chapter.
[Abstract] [Full Text]


Home page
CirculationHome page
J. G. Webb, L. Altwegg, R. H. Boone, A. Cheung, J. Ye, S. Lichtenstein, M. Lee, J. B. Masson, C. Thompson, R. Moss, et al.
Transcatheter Aortic Valve Implantation: Impact on Clinical and Valve-Related Outcomes
Circulation, June 16, 2009; 119(23): 3009 - 3016.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow View Related Genuine Article on CVN
Right arrow View Related Cardiosource Journal Scan
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (7)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Webb, J. G.
Right arrow Articles by Boone, R. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Webb, J. G.
Right arrow Articles by Boone, R. A.
Related Collections
Right arrowRelated Article

 
  CME Topic Collections Past Issues Search Current Issue Home

Advertisement