CLINICAL RESEARCH: CARDIAC IMAGING
Characterization of Peri-Infarct Zone Heterogeneity by Contrast-Enhanced Multidetector Computed TomographyA Comparison With Magnetic Resonance Imaging
Karl H. Schuleri, MD*,
Marco Centola, MD*,
Richard T. George, MD*,
Luciano C. Amado, MD*,
Kristine S. Evers, AA*,
Kakuya Kitagawa, MD, PhD*,
Andrea L. Vavere, MS*,
Robert Evers, BSRT*,
Joshua M. Hare, MD ,
Christopher Cox, PhD ,
Elliot R. McVeigh, PhD ,
João A.C. Lima, MD* and
Albert C. Lardo, PhD*, , ,*
* Department of Medicine, Division of Cardiology, Johns Hopkins University, Baltimore, Maryland
Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland
Department of Medicine, Division of Cardiology, University of Miami, Miami, Florida
Johns Hopkins Bloomberg School of Public Health, Baltimore, Maryland
Manuscript received September 8, 2008;
revised manuscript received January 9, 2009,
accepted January 16, 2009.
* Reprint requests and correspondence: Dr. Albert C. Lardo, Image Guided Cardiotherapy Laboratory, Johns Hopkins School of Medicine, Division of Cardiology, 1042 Ross Building, Baltimore, Maryland 21205 (Email: al{at}jhmi.edu).
 |
Abstract
|
|---|
Objectives: This study examined whether multidetector computed tomography (MDCT) improves the ability to define peri-infarct zone (PIZ) heterogeneity relative to magnetic resonance imaging (MRI).
Background: The PIZ as characterized by delayed contrast-enhancement (DE)-MRI identifies patients susceptible to ventricular arrhythmias and predicts outcome after myocardial infarction (MI).
Methods: Fifteen mini-pigs underwent coronary artery occlusion followed by reperfusion. Both MDCT and MRI were performed on the same day approximately 6 months after MI induction, followed by animal euthanization and ex vivo MRI (n = 5). Signal density threshold algorithms were applied to MRI and MDCT datasets reconstructed at various slice thicknesses (1 to 8 mm) to define the PIZ and to quantify partial volume effects.
Results: The DE-MDCT reconstructed at 8-mm slice thickness showed excellent correlation of infarct size with post-mortem pathology (r2 = 0.97; p < 0.0001) and MRI (r2 = 0.92; p < 0.0001). The DE-MDCT and -MRI were able to detect a PIZ in all animals, which correlates to a mixture of viable and nonviable myocytes at the PIZ by histology. The ex vivo DE-MRI PIZ volume decreased with slice thickness from 0.9 ± 0.2 ml at 8 mm to 0.2 ± 0.1 ml at 1 mm (p = 0.01). The PIZ volume/mass by DE-MDCT increased with decreasing slice thickness because of declining partial volume averaging in the PIZ, but was susceptible to increased image noise.
Conclusions: A DE-MDCT provides a more detailed assessment of the PIZ in chronic MI and is less susceptible to partial volume effects than MRI. This increased resolution best reflects the extent of tissue mixture by histopathology and has the potential to further enhance the ability to define the substrate of malignant arrhythmia in ischemic heart disease noninvasively.
Key Words: MDCT delayed contrast-enhancement peri-infarct zone MRI
|
Abbreviations and Acronyms
| | CNR = contrast-to-noise ratio | | DE = delayed contrast-enhancement | | LAD = left anterior descending coronary artery | | LV = left ventricle/ventricular | | MDCT = multidetector computed tomography | | MI = myocardial infarction | | MRI = magnetic resonance imaging | | PIZ = peri-infarct zone | | SEM = scanning electron microscopy | | SNR = signal-to-noise ratio | | TEM = transmission electron microscopy |
|
The success of myocardial protection after myocardial infarction (MI) has led to an ever-increasing number of patients with advanced heart disease and chronic heart failure (1) in which questions of viability and risk stratification are of crucial importance. Clinical assessment of myocardial viability is performed by single-photon emission computed tomography, positron emission tomography, stress echocardiography, or magnetic resonance imaging (MRI) (2–5). Evaluation of infarct size and morphology with delayed contrast-enhancement (DE) MRI has been well validated and is currently considered the clinical gold standard for viability assessment (6–8). Recently, DE-MRI was used to assess heterogeneity of the peri-infarct zone (PIZ), which was shown to identify patients susceptible to ventricular arrhythmias (9) and predict post-MI mortality (10).
The use of cardiac multidetector computed tomography (MDCT) has expanded beyond assessment of coronary atherosclerotic burden (11–13) to include myocardial viability by DE in a similar fashion as MRI (14–16) as well as myocardial perfusion assessment (17,18). Limited spatial resolution of in vivo MRI in the z or axial direction (slice thickness) and the resulting partial volume effect at the infarct border is regarded as a limiting factor for the assessment of nonviable myocardial tissue by MRI and is likely partly responsible for the presence and appearance of the PIZ documented by MRI (7). With its high isotropic resolution, MDCT has the potential to greatly decrease partial volume effects and accurately characterize the PIZ. However, the high z-axis spatial resolution of MDCT comes at the expense of higher image noise, which may preclude accurate assessment of the PIZ. The effect of slice thickness and partial volume effects on infarct size and infarct heterogeneity relative to MRI has not been systematically studied for DE-MDCT imaging. Accordingly, the purpose of this study was 2-fold: 1) to compare DE-MRI and -MDCT assessment of the PIZ in a chronic 6-month occlusion/reperfusion model of healed MI; and 2) to characterize the effect of slice thickness (partial volume averaging) on DE-MDCT and -MRI infarct and PIZ measurements.
 |
Methods
|
|---|
Animal model.
All animal studies were approved by the Johns Hopkins University Institutional Animal Care and Use Committee and comply with the Guide for the Care and Use of Laboratory Animals (National Institutes of Health Publication no. 80-23, revised 1985). Female Göttingen mini-pigs were purchased from Marshall BioResources (North Rose, New York).
Animals with chronic infarcts were created as previously described in detail (19). Briefly, MI was induced by engaging the left anterior descending coronary artery (LAD) with an 8-F hockey stick catheter under fluoroscopic guidance. Then, a 0.014-inch angioplasty guidewire was inserted into the LAD and a 2.5 x 12-mm Maverick balloon (Boston Scientific, Natick, Massachusetts) was inflated to 4 atm just distal to the second diagonal branch of the LAD. After 120 min, occlusion of the vessel was terminated by deflating the balloon, and restoration of flow in the LAD was confirmed by angiography.
Animals progressed to heart failure (Table 1), and all MDCT and MRI studies were performed in random order on the same day 191 ± 4 days after MI induction (Table 1). A total of 15 animals were studied. One animal died immediately after the MDCT scan before MRI data could be acquired. Pathology at 2-mm slice thickness was achieved in 12 animals. In a subset of 5 animals, ex vivo MRI images were obtained after the last in vivo experiment.
MDCT.
Each animal was scanned, with electrocardiographic monitoring, using a 0.5-mm x 64-detector scanner (Aquilion 64, Toshiba Medical Systems Corporation, Otawara, Japan). Animals received intravenous metoprolol (2 to 5 mg) and/or amiodarone (50 to 150 mg) to achieve a heart rate <100 beats/min for DE studies. After scout acquisition and slice prescription, a 150-ml bolus of iodixanol (Visipaque, 320 mg iodine/ml, Amersham Health, Amersham, United Kingdom) was injected intravenously (0.91 ± 0.04 x 10–3 mg iodine/kg body weight). During MDCT acquisition, respiration was suspended and imaging was performed using a retrospectively gated MDCT protocol with the following parameters: gantry rotation time 400 ms, temporal resolution 212 ± 5 ms, detector collimation 0.5 mm x 64 (isotropic voxels 0.5 x 0.5 x 0.5 mm3, 13 line pairs/cm), helical pitch variable depending on heart rate (range 6.4 to 6.8), tube voltage 120 kV, tube current 400 mA. The DE images were acquired during the first pass and at 5, 10, and 15 min after contrast delivery.
Raw data were reconstructed as contiguous 0.5-mm slice thickness by an adaptive multisegment reconstruction algorithm (20). Images were reconstructed at 80% of the R-R interval using a standard kernel (FC43). Electrocardiographic editing to account for arrhythmias was performed when necessary. Multiplanar reformation in the short axis of the heart was implemented to achieve 8-, 4-, 2-, and 1-mm slice thickness to isolate effects of partial volume and signal-to-noise ratio (SNR).
MRI.
In vivo MRI images were acquired using a 1.5-T MR scanner (CV/i, GE Medical Systems, Waukesha, Wisconsin) 24 weeks after MI. Global left ventricular (LV) function was assessed using a steady-state free-precession pulse sequence (21). A total of 8 to 10 contiguous short-axis slices were prescribed to cover the entire LV, from base to apex. Image parameters were the following: echo time/repetition time 1.9/4.2 ms, flip angle 45°; 256 x 160 matrix; 8-mm slice thickness/no gap; 125 kHz; 28-cm field of view; 1 number of signal averages (NSA).
The DE images were acquired 15 min after intravenous injection of gadolinium diethylenetriamine penta-acetic acid (0.2 mmol/kg body weight, Magnevist, Berlex, Wayne, New Jersey), using an electrocardiogram-gated, breath-hold, interleaved, inversion recovery, fast gradient repetition echo pulse sequence. The DE-MRI images were acquired using the same slice prescription used for short-axis cine images. Imaging parameters were: echo time/repetition time 3.3/7.3 ms; flip angle 25°; bandwidth 31.2 kHz; field of view 280 mm; 256 x 196 matrix; 1.3 x 1.3 x 8.0 mm3; 8-mm slice thickness/no gap. Inversion recovery time ( 200 ms) was manually adjusted as needed to null the normal myocardium (22).
Ex vivo MRI was conducted in a subset of 5 animals to visualize infarct morphology at submillimeter resolution. After intravenous administration of heparin 5,000 IU and 0.20 mmol/kg gadolinium diethylenetriamine penta-acetic acid, the animals were euthanized and the hearts were removed and filled with vinyl polysiloxane to ensure diastolic arrest. The heart was scanned in a 1.5-T Sonata MR scanner (Siemens Medical Solutions, Erlangen, Germany) with a 3-dimensional gradient echo sequence. The imaging parameters were as follows: echo time/repetition time 4.02/9.7 ms; flip angle 20°; bandwidth ±130 Hz/pixel; field of view 100 mm; image matrix 256 x 256; spatial resolution 0.39 x 0.39 x 0.39 mm3. Images were then averaged to achieve multiple slice thicknesses (1, 2, 4, and 8 mm) for assessment of partial volume effects on the PIZ and infarct core.
Imaging analysis.
Cine MRI images, DE-MDCT, and DE-MR images were analyzed using a custom research software package (Cine Tool, GE Medical Systems). To evaluate resting LV function, endocardial and epicardial borders of the LV were defined each in the end-diastolic and -systolic frame in contiguous slices and end-diastolic LV volume, end-systolic LV volume, LV stroke volume, and LV ejection fraction were calculated (Table 1). Infarct size in DE-MDCT and -MR images was defined by SDs of the signal intensity to the remote mean (noninfarct) myocardium for the core infarct and 2 SD of the signal intensity from the remote mean to the PIZ as described previously (23).
The SNRs were calculated for MDCT and MRI by dividing the mean signal intensity in the infarct area with the SD of signal intensity outside of the thorax. The SD of the air measurements reflects the degree of image noise. Mean signal intensity and SD were obtained by measuring air/noise outside the body or in the lung field. Image contrast-to-noise ratios (CNR) were calculated by using the following equation: (mean signal intensity of the infarct – mean signal intensity of the remote region)/SD of noise.
Pathology and histology.
Animals were immediately killed after the imaging study. Hearts were arrested in diastole by slow retrograde infusion of an ice-cold solution of 4 mmol/l potassium chloride (KCl) and 4% paraformaldehyde dissolved in phosphate-buffered saline. The porcine hearts were excised, filled with dental rubber to avoid uneven tissue shrinkage, and post-fixed in 4% paraformaldehyde overnight at 4°C. For post-mortem infarct size assessment, the fixed porcine hearts were sectioned from apex to base at 2-mm slice thickness with a commercial available meat slicer. Myocardial slices were weighed, and the apical and basal aspect of each slice was digitally photographed. The infarct scar and areas of viable myocardium were manually traced using imaging analysis software (Sigma Scan Pro5, Systat Software Inc., San Jose, California). The infarct mass was calculated according to slice weight, and infarct size was expressed as percentage of LV mass excluding the papillary muscle.
Fixed tissue samples from the peri-infarct area were embedded in paraffin, and 5-µm sections were cut and stained with a Masson trichrome protocol. Light microscopy images were obtained on a Zeiss Axiophot microscope (Carl Zeiss NTS GmbH, Oberkochen, Germany).
Electron microscopy.
Small tissue samples including both chronic infarct scar and viable myocardium were dissected from the porcine hearts after perfusion with ice-cold phosphate-buffered saline solution. For transmission electron microscopy (TEM) and scanning electron microscopy (SEM), the cardiac tissue was fixed in 2% glutaraldehyde in 0.1 M cacodylate pH 7.4 with 3% sucrose and 3 mM CaCl2 overnight at 4°C. The TEM samples were post-fixed with 1% osmium tetroxide reduced in potassium ferrocyanide for 1 h at 4°C. Tissue was incubated in 0.15% tannic acid for 1 min to enhance the visualization of collagen fibers, and placed in 2% uranyl acetate for 1 h at room temperature.
After fixation, the tissue was stained en bloc with a 2% aqueous solution of uranyl acetate and dehydrated in graded ethanol and embedded. Ultra-thin sections (70 to 90 nm) were cut and stained with uranyl acetate and lead citrate. Samples were viewed and photographed on a Hitachi 7600 TEM (Hitachi High Technologies America, Inc., Gaithersburg, Maryland).
The SEM samples were fixed and dried with liquid CO2, mounted onto SEM stubs, and sputter coated with 20-nm gold palladium. The SEM images were obtained on a LEO1530 field emission SEM (Carl Zeiss NTS GmbH).
Statistical analysis.
All data are presented as mean ± standard error of the mean unless otherwise stated. For infarct size and infarct volume/mass comparison, Pearson correlation and linear regression analysis were used to compare MDCT versus post-mortem pathology and MRI. Results were confirmed by Bland-Altman analysis and agreement expressed as mean ± SD difference between methods at 95% confidence intervals. The MDCT and MRI data at 8-mm slice thickness (Table 2) were evaluated with a paired Student t test. These analyses were performed in MedCalc (MedCalc Software, Mariakerke, Belgium).
Because the study involved repeated measurements of individual animals, an alinear mixed model (24) was used for analysis of the data depending on slice thickness. The model included fixed effects for imaging modality and/or slice thickness, as well as their interaction when both were included in the model, and a random animal effect. Post-hoc comparisons were made among adjusted means using Wald tests. The linear mixed model analyses were performed in SAS (SAS/STAT Software, Cary, North Carolina). In all analyses, values of p < 0.05 were considered significant.
 |
Results
|
|---|
MDCT viability imaging.
First-pass DE-MDCT imaging showed homogenous myocardial enhancement of the remote myocardium and hypodense regions in the LAD territory. Post-infarct MDCT imaging at 5, 10, and 15 min after contrast injection identified collagenous myocardial scar that was characterized by well-delineated hyperdense regions. Density analysis of the LV blood pool, infarct region, and remote myocardium at all time points confirmed that optimal image quality (highest CNR) occurred at 10-min post-contrast, and thus all DE-MDCT images were analyzed at this time point.
DE-MDCT versus -MRI and pathology.
The morphology and distribution of chronic infarct scar assessed by DE-MDCT correlated well with gross pathology and with ex vivo and in vivo MRI (Fig. 1). Infarct mass was 4.3 ± 0.6 g evaluated by gross pathology. Total infarct size as percent of LV mass by MDCT at 8 mm correlated well with in vivo MRI (r2 = 0.92; p < 0.0001) and post-mortem pathology (r2 = 0.97; p < 0.0001) with a 8.2% mean underestimation by MDCT versus MRI and a qualitative trend toward overestimation by MDCT versus pathology with mean difference –8.2%. Matching slice thickness of 2 mm for DE-MDCT images and post-mortem pathology, the overestimation of DE-MDCT decreased to a mean difference of –7.3%.

View larger version (19K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 1 Comparison of Infarct Morphology by MDCT With Post-Mortem Pathology and MRI
Example images in short-axis orientation for (A) delayed contrast-enhancement multidetector computed tomography (DE-MDCT) at 2 mm, (B) post-mortem pathology at 2 mm, and (C) ex vivo and (D) in vivo magnetic resonance imaging (MRI) at 4-mm slice thickness show a similar location and distribution pattern of the chronic collagenous scar in the anterior septal left ventricular wall.
|
|
PIZ imaging.
A PIZ could be measured for both DE-MDCT and -MRI in all animals at all slice thicknesses (Table 2, Fig. 2). We used standard histology TEM and SEM as a reference to investigate the tissue composition of the PIZ and to clarify the imaged substrate. Gross pathology and Masson trichrome staining showed transmural infarct scars with densely packed collagen fibers (Figs. 2C and 2F) in all animals. Infarct scar and viable myocardium were clearly delineated in all chronic MI (Figs. 2G to 2I), and islands of viable myocytes could be detected by Masson trichrome staining at the border zone of the infarcts (Figs. 2C and 2F). All myocardial infarcts showed this intermingling of viable and nonviable tissue at the border zone of the infarct scar.

View larger version (116K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 2 Characterization of Tissue Heterogeneity in the PIZ by MDCT
Example of matching 8-mm slice section in chronic infarcts for DE-MDCT (A and B) and -MRI (D and E). (A and B) The viable myocardium (*) shows lower attenuation values than enhanced scar tissue (white arrows). The peri-infarct zone (PIZ) is visualized between 2 black arrows by intermediate signal intensity (white circle) in DE-MDCT images. (D and E) Viable myocardium (*), DE infarct scar (white arrows), and the PIZ (white circle) show different signal intensity in DE-MRI. The viable myocardium is darker in the MRI acquisition, and the PIZ appears larger in DE-MDCT images compared with DE-MRI. (C and F) Masson trichrome stain depicts viable myocardium in red (*) from nonviable tissue in blue. At higher magnification the densely packed collagenous extracellular matrix (white arrows) in the chronic infarct scar can be appreciated. Islands of viable myocytes (red) within the scar tissue are visualized, showing the heterogeneity of the PIZ (area between black arrows). Scanning electron microscope images of the densely packed collagen fibers 6 months after MI (G) and viable myocytes (H) characterize the ultra-structure of the chronic infarct and the PIZ. (I) Transmission electron microscopy of the PIZ shows the clear delineating of the collagenous scar (white arrows) and viable tissue (*). Abbreviations as in Figure 1.
|
|
Effect of slice thickness on the PIZ.
Slice thickness had a marked effect on the calculated mass of the PIZ for both in vivo and ex vivo DE-MRI experiments (Fig. 3), suggesting a strong partial volume effect. For ex vivo MRI, the PIZ mass decreased from 0.9 ± 0.2 mg to 0.2 ± 0.1 mg comparing 8- and 1-mm thickness, respectively (p = 0.0001). The DE-MDCT PIZ volume showed a trend toward increasing values with decreasing slice thickness (Fig. 4). Additionally, the relative proportion of the PIZ expressed as a percentage of the total infarct core changed significantly with slice thickness (Table 3, Fig. 5). These results were consistent with increased CNR and decreased SNR in the PIZ with thinner slices by visual and quantitative assessment (Figs. 6B and 6C).

View larger version (11K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 4 Comparison of PIZ Assessment With Ex Vivo DE-MRI and -MDCT
The volume of the PIZ (A) and PIZ expressed as percentage of the total infarct size (B) decreases in a linear fashion with reduced slice thickness evaluated by ex vivo MRI, suggesting a pronounced partial volume effect, whereas the amount of PIZ volume is less affected by slice thickness in MDCT acquisitions until a slice thickness of 1 mm is reached. This implies that MDCT assessment of the PIZ is less susceptible to partial volume effects, but affected by image noise at 1 mm. There are marked differences in the PIZ volume and percentage assessment between MDCT and ex vivo MRI assessment at 1 and 8 mm, *p < 0.01 and p < 0.001, respectively. Differences between slice thicknesses: 1 mm versus 4 and 8 mm, p < 0.01 and p < 0.001, respectively; 2 mm versus 8 mm, ¶p = 0.01 for ex vivo MRI, and 8 mm versus 2 and 4 mm, #p < 0.05 and **p < 0.05, respectively. Abbreviations as in Figure 2.
|
|

View larger version (60K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 5 Effect of Slice Thickness on DE-MDCT PIZ
(A) Long-axis orientation of DE-MDCT showing the location of the short axis images (white box). (B and C) DE-MDCT images in the short axis reconstructed at 8 and 1 mm. Note the smooth appearance of the tissue versus the more heterogeneous appearance at 1 mm. (D and E) The same myocardial slices with a computer-generated mask depicting the core infarct (red) and the PIZ (yellow). The increase of the PIZ (yellow) can be appreciated with decreased slice thickness. Abbreviations as in Figure 1.
|
|

View larger version (15K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 6 Effect of Slice Thickness on DE-MDCT Density and SNR and CNR
(A) Signal density in Hounsfield units for the LVBP, IN, PIZ, and RE; the LVBP and RE do not show different attenuation values at different slice thickness, whereas IN values (*p < 0.05, 1 mm vs. 4 mm and 2 mm vs. 4 mm; p = 0.01, 1 mm vs. 8 mm, 2 mm vs. 8 mm) and PIZ values ( p < 0.01, 1 mm vs. 4 and 8 mm, and 2 mm vs. 8 mm) change with the reconstructed slice thickness parameters ( p = 0.007 and ¶p = 0.002, respectively). (B) Mean MDCT SNR (8 mm vs. 2 and 1 mm ( p = 0.01 and p < 0.01, respectively; and 4 mm vs. 1 mm, *p < 0.05; p = 0.003), and (C) mean MDCT CNR (8 mm vs. 2 and 1 mm, p < 0.01; **p = 0.008) change with the reconstructed slice thickness, which reflects reduced imaging quality of thinner reconstructed slices. CNR = contrast-to-noise ratio; IN = infarcted myocardium; LVBP = left ventricular blood pool; RE = remote myocardium; SNR = signal-to-noise ratio; other abbreviations as in Figures 1 and 2.
|
|
 |
Discussion
|
|---|
To the best of our knowledge, this is the first report of DE-MDCT and -MRI PIZ imaging with post-mortem pathology in dense remodeled scar 6 months' post-MI. The major findings of this study are that DE-MDCT is superior to MRI in its ability to detect and quantify the PIZ after contrast delivery in chronic MI. Furthermore, we characterize the influence of partial volume effects on the measurement of the PIZ by imaging methods and provide detailed comparisons against histopathology.
Scar enhancement by MDCT: comparison with MRI.
Delayed hyperenhancement in chronic collagenous myocardial scar results from an accumulation of contrast media in the interstitial space between collagen fibers, resulting in an increased volume of contrast distribution in the scar tissue compared with that of tightly packed myocytes (7). The accumulation of contrast material in the collagenous scar matrix is a passive process, and the timing between administration of contrast agents and imaging is crucial to differentiate blood pool from scar and the PIZ. The DE-MDCT density values are somewhat unique and determined by the physical properties of individual constituents of the heart including blood and viable and nonviable myocardium that result from direct X-ray attenuation by iodine. Although not a perfect binary system, such imaging is theoretically well suited to characterize infarct heterogeneity.
Because of its high CNR and lack of ionizing radiation, DE-MRI is a highly attractive modality for myocardial viability imaging. This approach is based on gadolinium-induced alterations of water relaxivity and thus represents a surrogate measure of the amount and distribution of contrast in the extracellular space. Recent studies have shown that DE-MDCT and -MRI infarct size compare well in patients (14,25), despite the lower image quality of DE-MDCT examinations. The lower image quality of DE-MDCT is likely related to technical factors such as temporal resolution as well as differences in the mechanisms of hyperenhancement between DE-MDCT and -MRI (e.g., the lack of iodine–water interactions that modulate signals). The quality of acute DE-MDCT infarct imaging is improved over that of scar imaging because iodine is able to access the intracellular space rather than the limited extracellular space generated by a collagen matrix (15). Other features unique to DE-MRI data acquisition, such as inversion recovery based myocardial nulling, further improve scar visualization over DE-MDCT but may also partly explain the overestimation of infarct size by DE-MRI in our chronic infarcts and in the acute setting (26).
Partial volume effects and the appearance of the PIZ.
The appearance of the PIZ by DE-MDCT and -MRI can be explained by 2 possible orientations of the normal myocardium with respect to scar: 1) an area of uniformly dense scar adjacent to preserved myocardium with a sharp binary transition; or 2) the intermingling of bundles of fibrotic scar with viable myocytes. Partial volume effects are related to the 3-dimensional spatial resolution of a tomographic image. If a given voxel at the infarct periphery contains both infarct and noninfarct tissue, the 2 different signal intensities will be averaged, and this particular voxel will be represented by an intermediate attenuation value. For standard clinical DE-MRI, voxel volumes are typically >13 mm3 (1.3 x 1.3 x 8.0 mm3) and thus approximately 200,000 cells at the interface between infarcted and viable tissue in the PIZ can occur in a single voxel. To quantify the effect of partial volume averaging for DE-MRI, we acquired high-resolution ex vivo MR images at 0.4-mm slice thickness using a 3-dimensional sequence with an isotropic voxel size 0.39 x 0.39 x 0.39 mm3, resulting in a pixel volume 59 µm3, which allowed for reproduction of various slice thicknesses. The measured PIZ decreased 350% when comparing a clinically used slice thickness (8 mm) with 1-mm slices. We also observed a similar decrease in the PIZ when comparing in vivo DE-MRI images. These data suggest a strong partial volume effect that greatly decreases local CNR in the PIZ and support the hypothesis that the PIZ measured by DE-MRI may overestimate the actual PIZ.
In MDCT imaging with 64-detector-row scanners, raw data can be acquired and displayed at a high isotropic resolution with a voxel volume of 43 µm3 (0.35 x 0.35 x 0.35 mm3), which is 2 orders of magnitude smaller than that of in vivo MRI. During the MDCT reconstruction process it is possible to reconstruct and analyze multiple slice thicknesses from the original data set, which allows partial volume effects to be quantified. At thinner slices, partial volume averaging decreases and CNR increases at the expense of decreased SNR. Unlike the DE-MRI derived PIZ, the DE-MDCT PIZ volume increased with increasing slice thickness. This is likely explained by an increase in CNR in the PIZ that allows better differentiation of viable cells and scar because partial volume averaging is greatly decreased. These advantages, however, are partly offset by increased image noise that may limit PIZ assessment.
Overall, the PIZ in these experiments in chronic infarct scars was small. This is partly because of the lack of collateral coronary circulation in swine species that generates well-delineated infarcts with sharp borders. Further, the iodine washout in the chronic porcine model is more rapid compared with our previous experiments in a canine model of acute MI (15). The rich collateral supply in dog hearts facilitates the passive accumulation and distribution of contrast material in the infarct area and delays the clearance of iodine, creating a more intense infarct signal and greater PIZ. However, our PIZ size correlates well with human data reported for patients with small PIZ volumes (10).
Clinical implications.
The potential value of imaging the PIZ has been recently reported (9,10). This very pathologically complex region contains a mixture of both viable and nonviable electrically inactive scar that is thought to provide substrate for ventricular tachycardia via macrore-entry mechanisms. Thus, accurate characterization and phenotyping of the PIZ may allow selection of optimal candidates for implantable defibrillators and identify targets for ablation therapy during interventional electrophysiology procedures. We are currently using DE-MDCT of the PIZ for this purpose at our institution in a porcine model of ventricular tachycardia.
Standardized low-radiation DE-MDCT protocols are required for ultimate translation of PIZ assessment into the clinical setting. We have recently described a prospectively gated protocol for high-resolution DE-MDCT imaging that lowers radiation dose by an order of magnitude (27). The focus of such protocols is achieving a reasonable balance of SNR and CNR for visual and quantitative infarct measurements while minimizing partial volume effects. Based on the results of these studies, a slice thickness of 1 to 2 mm seems to provide this balance. Further studies are required to determine optimal threshold cutoffs and determine whether the evaluation of the PIZ with DE-MDCT has prognostic value as suggested for MRI (9,10). The ability to image healed MI in heart failure with DE-MDCT and to quantify the PIZ may enhance identification of patients susceptible to life-threatening arrhythmias and sudden cardiac death.
Study limitations.
We applied previously published signal density threshold algorithms to define the infarct core and peri-infarct region. However, specific optimal cutoff values for normal, PIZ, and core zones of myocardium are unknown and are likely dependent on study quality. Additionally, because of the passive contrast kinetics of iodinated contrast and the limited collateral circulation in pigs, DE-MDCT imaging of collagenous scar in this model requires a relatively large dose of iodine to cause a sufficient change in the volume of distribution in that myocardial bed. In this study we used 1.5x a human equivalent dose, although successful viability imaging in humans has been reported in several studies with standard contrast volumes used for coronary CT angiography (25). A further limitation of the study is the lack of quantitative histological gold standards to index the degree of heterogeneity of the PIZ.
 |
Conclusions
|
|---|
This study found that DE-MDCT provides an accurate measure of the spatial extent of chronic collagenous infarct scars relative to MRI and post-mortem pathology. The improved spatial resolution of DE-MDCT minimizes partial volume averaging and allows for quantification of complex tissue substrates in the PIZ.
 |
Acknowledgments
|
|---|
The authors thank Michael Delannoy, MS, and Carol Cooke, MS, at the JHMI Microscope Facility for excellent technical assistance performing the electron microscopy studies. The authors also thank Norman J. Barker, MS, RBP, for his expertise and helpful suggestions obtaining pathology and histology images.
 |
Footnotes
|
|---|
Supported by National Institutes of Health grant U54 HL081028 (Specialized Center for Cell Therapeutics) and The Donald W. Reynolds Foundation. Christopher M. Kramer, MD, served as Guest Editor for this article.
 |
References
|
|---|
1. American Heart Association. Heart and Stroke Statistics—2008 Update. http://circ.ahajournals.org/cgi/content/full/117/4/e25Accessed June 10, 2008.2. Burt RW, Perkins OW, Oppenheim BE, et al. Direct comparison of fluorine-18-FDG SPECT, fluorine-18-FDG PET and rest thallium-201 SPECT for detection of myocardial viability J Nucl Med 1995;36:176-179.[Abstract/Free Full Text] 3. Mahrholdt H, Wagner A, Judd RM, Sechtem U. Assessment of myocardial viability by cardiovascular magnetic resonance imaging Eur Heart J 2002;23:602-619.[Free Full Text] 4. Sicari R, Pasanisi E, Venneri L, Landi P, Cortigiani L, Picano E. Stress echo results predict mortality: a large-scale multicenter prospective international study J Am Coll Cardiol 2003;41:589-595.[Abstract/Free Full Text] 5. Iskandrian AS, Heo J, Schelbert HR. Myocardial viability: methods of assessment and clinical relevance Am Heart J 1996;132:1226-1235.[CrossRef][Web of Science][Medline] 6. Kim RJ, Wu E, Rafael A, et al. The use of contrast-enhanced magnetic resonance imaging to identify reversible myocardial dysfunction N Engl J Med 2000;343:1445-1453.[CrossRef][Web of Science][Medline] 7. Wu KC, Lima JA. Noninvasive imaging of myocardial viability: current techniques and future developments Circ Res 2003;93:1146-1158.[Abstract/Free Full Text] 8. Kim RJ, Albert TS, Wible JH, et al. Performance of delayed-enhancement magnetic resonance imaging with gadoversetamide contrast for the detection and assessment of myocardial infarction: an international, multicenter, double-blinded, randomized trial Circulation 2008;117:629-637.[Abstract/Free Full Text] 9. Schmidt A, Azevedo CF, Cheng A, et al. Infarct tissue heterogeneity by magnetic resonance imaging identifies enhanced cardiac arrhythmia susceptibility in patients with left ventricular dysfunction Circulation 2007;115:2006-2014.[Abstract/Free Full Text] 10. Yan AT, Shayne AJ, Brown KA, et al. Characterization of the peri-infarct zone by contrast-enhanced cardiac magnetic resonance imaging is a powerful predictor of post-myocardial infarction mortality Circulation 2006;114:32-39.[Abstract/Free Full Text] 11. Poon M, Rubin GD, Achenbach S, et al. Consensus update on the appropriate usage of cardiac computed tomographic angiography J Invasive Cardiol 2007;19:484-490.[Medline] 12. Schroeder S, Achenbach S, Bengel F, et al. Cardiac computed tomography: indications, applications, limitations, and training requirements: report of a writing group deployed by the Working Group Nuclear Cardiology and Cardiac CT of the European Society of Cardiology and the European Council of Nuclear Cardiology Eur Heart J 2008;29:531-556.[Abstract/Free Full Text] 13. Budoff MJ, Achenbach S, Blumenthal RS, et al. Assessment of coronary artery disease by cardiac computed tomography: a scientific statement from the American Heart Association Committee on Cardiovascular Imaging and Intervention, Council on Cardiovascular Radiology and Intervention, and Committee on Cardiac Imaging, Council on Clinical Cardiology Circulation 2006;114:1761-1791.[Free Full Text] 14. Gerber BL, Belge B, Legros GJ, et al. Characterization of acute and chronic myocardial infarcts by multidetector computed tomography: comparison with contrast-enhanced magnetic resonance Circulation 2006;113:823-833.[Abstract/Free Full Text] 15. Lardo AC, Cordeiro MA, Silva C, et al. Contrast-enhanced multidetector computed tomography viability imaging after myocardial infarction: characterization of myocyte death, microvascular obstruction, and chronic scar Circulation 2006;113:394-404.[Abstract/Free Full Text] 16. Mahnken AH, Koos R, Katoh M, et al. Assessment of myocardial viability in reperfused acute myocardial infarction using 16-slice computed tomography in comparison to magnetic resonance imaging J Am Coll Cardiol 2005;45:2042-2047.[Abstract/Free Full Text] 17. George RT, Silva C, Cordeiro MA, et al. Multidetector computed tomography myocardial perfusion imaging during adenosine stress J Am Coll Cardiol 2006;48:153-160.[Abstract/Free Full Text] 18. George RT, Jerosch-Herold M, Silva C, et al. Quantification of myocardial perfusion using dynamic 64-detector computed tomography Invest Radiol 2007;42:815-822.[CrossRef][Web of Science][Medline] 19. Schuleri KH, Boyle AJ, Centola M, et al. Novel use of the adult Göttingen minipig as a chronic heart failure model post myocardial infarction: focus on cardiovascular imaging and regenerative therapies Comp Med 2008;58:568-579.[Web of Science][Medline] 20. Dewey M, Laule M, Krug L, et al. Multisegment and halfscan reconstruction of 16-slice computed tomography for detection of coronary artery stenoses Invest Radiol 2004;39:223-229.[CrossRef][Web of Science][Medline] 21. Slavin GS, Saranathan M. FIESTA-ET: high-resolution cardiac imaging using echo-planar steady-state free precession Magn Reson Med 2002;48:934-941.[CrossRef][Web of Science][Medline] 22. Simonetti OP, Kim RJ, Fieno DS, et al. An improved MR imaging technique for the visualization of myocardial infarction Radiology 2001;218:215-223.[Abstract/Free Full Text] 23. Amado LC, Gerber BL, Gupta SN, et al. Accurate and objective infarct sizing by contrast-enhanced magnetic resonance imaging in a canine myocardial infarction model J Am Coll Cardiol 2004;44:2383-2389.[Abstract/Free Full Text] 24. In: Molenberghs G, Verbeke G, editors. Linear Mixed Models for Longitudinal Data. New York, NY: Springer; 2000. 25. Nieman K, Shapiro, MD, Ferencik M, et al. Reperfused myocardial infarction: contrast-enhanced 64-section CT in comparison to MR imaging Radiology 2008;247:49-56.[Abstract/Free Full Text] 26. Saeed M, Wendland MF, Bremerich GL, Weinmann HJ, Higgins CB. Assessment of myocardial viability using standard extracellular and necrosis specific MR contrast media Acad Radiol 2002;9(Suppl 1):S84-S87.[CrossRef][Web of Science][Medline] 27. Chang HJ, George RT, Schuleri KH, et al. Prospectively ECG-gated delayed enhanced multi-detector computed tomography accurately quantifies infarct size and reduces radiation exposure by an order of magnitude J Am Coll Cardiol Img 2009;2:412-420.[Abstract/Free Full Text]
Related Article
-
Inside This Issue
J. Am. Coll. Cardiol. 2009 53: A24.
[Full Text]
[PDF]
This article has been cited by other articles:

|
 |

|
 |
 
K. H. Schuleri, M. Centola, S. H. Choi, K. S. Evers, F. Dawoud, R. T. George, J. A. C. Lima, and A. C. Lardo
CT for Evaluation of Myocardial Cell Therapy in Heart Failure: A Comparison With CMR Imaging
J. Am. Coll. Cardiol. Img.,
December 1, 2011;
4(12):
1284 - 1293.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Bamberg, A. Becker, F. Schwarz, R. P. Marcus, M. Greif, F. von Ziegler, R. Blankstein, U. Hoffmann, W. H. Sommer, V. S. Hoffmann, et al.
Detection of Hemodynamically Significant Coronary Artery Stenosis: Incremental Diagnostic Value of Dynamic CT-based Myocardial Perfusion Imaging
Radiology,
September 1, 2011;
260(3):
689 - 698.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Carlsson, D. Saloner, A. J. Martin, P. C. Ursell, and M. Saeed
Heterogeneous Microinfarcts Caused by Coronary Microemboli: Evaluation with Multidetector CT and MR Imaging in a Swine Model
Radiology,
March 1, 2010;
254(3):
718 - 728.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. J. Gibbons, P. A. Araoz, and E. E. Williamson
The Year in Cardiac Imaging
J. Am. Coll. Cardiol.,
February 2, 2010;
55(5):
483 - 495.