Figure 1


Figure 1 Technique of EECP

Three pairs of pneumatic cuffs are applied to the calves, lower thighs, and upper thighs. The cuffs are inflated sequentially during diastole, distal to proximal. The compression of the lower-extremity vascular bed increases diastolic pressure and flow and increases venous return. The pressure is then released at the onset of systole. Inflation and deflation are timed according to the R-wave on the patient’s cardiac monitor. The pressures applied and the inflation–deflation timing can be altered by using the pressure waveforms and electrocardiogram on the enhanced external counterpulsation (EECP) therapy monitor.





Right arrow Return to article

 