Background: Percutaneous aortic valve replacement represents an emerging alternative therapy for high-risk and inoperable patients with severe symptomatic aortic valve stenosis.
Methods: Patients with: 1) symptomatic, severe aortic valve stenosis (area <1 cm2); 2) age
80 years with a logistic EuroSCORE
20% (21-F group) or age
75 years with a logistic EuroSCORE
15% (18-F group); or 3) age
65 years plus additional prespecified risk factors were included. Introduction of the 18-F device enabled the transition from a multidisciplinary approach involving general anesthesia, surgical cut-down, and cardiopulmonary bypass to a truly percutaneous approach under local anesthesia without hemodynamic support.
Results: A total of 86 patients (21-F, n = 50; 18-F, n = 36) with a mean valve area of 0.66 ± 0.19 cm2 (21-F) and 0.54 ± 0.15 cm2 (18-F), a mean age of 81.3 ± 5.2 years (21-F) and 83.4 ± 6.7 years (18-F), and a mean logistic EuroSCORE of 23.4 ± 13.5% (21-F) and 19.1 ± 11.1% (18-F) were recruited. Acute device success was 88%. Successful device implantation resulted in a marked reduction of aortic transvalvular gradients (mean pre 43.7 mm Hg vs. post 9.0 mm Hg, p < 0.001) with aortic regurgitation grade remaining unchanged. Acute procedural success rate was 74% (21-F: 78%; 18-F: 69%). Procedural mortality was 6%. Overall 30-day mortality rate was 12%; the combined rate of death, stroke, and myocardial infarction was 22%.
Conclusions: Treatment of severe aortic valve stenosis in high-risk patients with percutaneous implantation of the CoreValve prosthesis is feasible and associated with a lower mortality rate than predicted by risk algorithms.
| ||||||
75 years of age, aortic stenosis was present in as many as 4.6%. Of note, not only does symptomatic valvular heart disease become more prevalent with age but also comorbidities that increase the risk for an operative valve replacement. The latter is the current standard therapy for aortic stenosis, with an operative mortality of <5% for first-time isolated aortic valve replacements (2). However, several factors have been identified as independently predictive for an increased risk of periprocedural or postprocedural mortality (3). For example, in patients with reduced systolic left ventricular ejection fraction, the rate of mortality increases to 10% (4). Furthermore, advanced age and renal disease increase the operative mortality risk markedly (odds ratio 4.2) (4,5). The need for alternative treatment options for patients with severe aortic stenosis (AS), particularly in combination with comorbidities, is justified by the fact that as many as one-third of elderly comorbid patients with symptomatic AS were denied surgery in the EuroHeart survey (6). Especially for high-risk patients, the "gold standard" treatment with a conventional surgery may not be the best option. A less-invasive procedure to minimize cardiovascular complications associated with general anesthesia, thoracotomy, and heart-lung machine is required. After the evaluation of a percutaneous valve replacement strategy in animal models (79), Cribier et al. (10) performed the first human implantation of a balloon-expandable aortic valve prosthesis with encouraging results. Nevertheless, the primarily used antegrade approach for valve implantation has shown to be very challenging (11). The innovation of a more flexible delivery catheter for the retrograde approach recently improved the procedural outcome (12). To avoid access site problems and to enhance stabilization during valve implantation, a transapical approach has been developed and is currently being tested (13,14).
An alternative technique with retrograde implantation of a self-expanding valve prosthesis (CoreValve prosthesis, CoreValve Inc., Irvine, California), which uses a porcine bioprosthesis within a nitinol frame, was first described in 2005 by our group (15). Subsequently, we have shown that the implantation procedure of the first- (24-French [F]) and second-generation (21-F) CoreValve device in 10 and 15 patients, respectively, is feasible and when successful results in marked hemodynamic and clinical improvements (16). Further device modifications have been realized since then and have reduced the sheath size from the initial 24-F to the present 18-F device (third-generation) (Figs. 1 and 2).
In this report, we describe the procedural success and clinical outcome up to 30 days after implantation of the second- (21-F) and current third-generation (18-F) of the CoreValve revalving system.
|
|
| Methods |
|---|
|
|
|---|
Patient population. A total of 86 consecutive patients were included in the present analysis. The study started in August 2005 with the 21-F device and enrolled a total of 50 patients (n = 25 in Siegburg, n = 14 in Leipzig, and n = 11 in Montreal). At the end of September 2006, the 18-F device became available and was later used exclusively for 36 patients until February 2007 (n = 18 in Siegburg, n = 10 in Leipzig, and n = 8 in Montreal).
Inclusion criteria were the following: 1) severe native aortic valve stenosis with an area <1 cm2 or <0.6 cm2/m2, with or without aortic valve regurgitation and age
80 years or a logistic EuroSCORE of
20% for the 21-F group and age
75 years or logistic EuroSCORE
15% for the 18-F group, respectively, or age
65 years and at least one of the following complications: cirrhosis of liver, pulmonary insufficiency (forced expiratory volume in one second <1 l), previous cardiac surgery, pulmonary hypertension >60 mm Hg, porcelain aorta, recurrent pulmonary embolus, right ventricular insufficiency, thoracic burning sequelae with contraindication for open chest surgery, history of mediastinum radiotherapy, severe connective tissue disease with contraindication for surgery, or cachexia (body mass index
18 kg/m2). 2) Echocardiographic aortic valve annulus diameter
20 and
27 mm. 3) Diameter of the ascending aorta
45 mm at the sinotubular junction.
Exclusion criteria included hypersensitivity or contraindication to any study medication; sepsis or active endocarditis; excessive femoral, iliac or aortic atherosclerosis, calcification, or tortuosity; aortic aneurysm; bleeding diathesis or coagulopathy; recent myocardial infarction or cerebrovascular accident; mitral or tricuspid valvular insufficiency (> grade II); left ventricular or atrial thrombus; uncontrolled atrial fibrillation; previous aortic valve replacement; polyarterial patients with either severe iliac or aortic vascular condition that make an insertion impossible or symptomatic carotid or vertebral arteries narrowing (>70%) disease or abdominal/thoracic aortic aneurysm; progressive disease with life expectancy <1 year; pregnancy; or creatinine clearance <20 ml/min.
Preinterventional morphological patient screening included transthoracic as well as transesophageal echocardiography, carotid and arteriovenous duplex ultrasonography, computed tomographic angiography, optional cardiac magnetic resonance imaging, and invasive cardiac evaluation with coronary angiogram and left ventriculography. The baseline operative risk of the patients was estimated by the logistic EuroSCORE (18). The patient was considered high risk if there was a consensus among an independent cardiologist and cardiac surgeon that conventional surgery would be associated with excessive morbidity and mortality.
Device description and procedure. The CoreValve aortic valve prosthesis consists of a trileaflet bioprosthetic porcine pericardial tissue valve, which is mounted and sutured in a self-expanding nitinol stent (Fig. 1). The prosthetic frame (stent) is manufactured by laser cutting and has an overall length of 50 mm. Further details of the device have been described previously (17). The second-generation (21-F) and third-generation (18-F) (Fig. 1) devices were used in the present study.
Vascular access was obtained either with or without standard surgical cutdown of the common iliac artery, the common femoral artery, or the subclavian artery. The procedure was performed with the patient under general anesthesia or with just local anesthesia in combination with a mild systemic sedative/analgesic treatment. The use of transesophageal echocardiographic guidance and the type of hemodynamic support (extracorporeal percutaneous femorofemoral bypass, tandem heart, extracorporeal membrane oxygenation, or none) was left to the discretion of the operator. Balloon valvuloplasty with a 20- to 23-mm balloon under rapid pacing was performed before device placement, after which over a stiff guidewire, placed in the left ventricle, the device was deployed retrogradely under fluoroscopic guidance. If used, extracorporal circulatory support was activated just before device placement across the native valve position and terminated immediately after withdrawal of the delivery catheter and confirmation of adequate valve function. Hemodynamic and echocardiographic outcomes were assessed serially during the procedure. Evaluation of postprocedural regurgitations was performed using a supra-aortic angiogram and echo. After the procedure, the patients were transferred to the intensive care unit.
Clinical follow-up and transthoracic echocardiography were performed postprocedure, at hospital discharge, and 30 days after device implantation. Additional clinical and echocardiographic follow-up is planned on a yearly basis for at least 4 years after the implantation.
Antiplatelet and antithrombotic medication. Acetylsalicylic acid (100 mg/day) was administered before the procedure and continued indefinitely. In addition, all patients received clopidogrel (300-mg loading dose), followed by 75 mg daily for at least 6 to 12 months. During the intervention, the patient received weight-adjusted intravenous heparin to achieve an activated clotting time of 300 to 350 s for the duration of the procedure.
Definitions and statistical analysis. Clinical adverse events were adjudicated by an independent clinical events committee. Device success was defined as stable device placement and adequate function as assessed by angiography and echocardiography. Acute procedural success was defined as device success with absence of periprocedural major adverse cardiovascular and cerebral events (MACCEs) in the first 48 h after device implantation. Major adverse cardiovascular and cerebral events consisted of death from any cause, myocardial infarction (creatine kinase-myocardial band >2 x the upper limit of normal), cardiac tamponade, stroke (as assessed by routine neurological assessment before and after procedure and before hospital discharge), urgent or emergent conversion to surgery or balloon valvuloplasty, emergent percutaneous coronary intervention, cardiogenic shock, endocarditis or aortic dissection. Major bleeding was defined as hemorrhage requiring surgery and/or 3 or more units of blood transfusion.
Categorical variables are presented as frequencies and were compared by chi-square test. Continuous variables are presented as mean ± standard deviation. A 2-tailed unpaired Student t test for comparison between groups and a paired Student t test for intragroup comparison was used. A p value of <0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
|
Acute procedural and clinical results. Acute device success was achieved in 76 (88%) of 86 enrolled patients (Table 2) with no difference between the 2 groups (88% vs. 89%, p = NS) (Table 2). In 6 patients, misplacement of the valve led to urgent conversion to operative valve replacement (Table 2). In 2 patients, the device did not cross the heavily calcified native valve despite a balloon predilatation and, therefore, only a balloon valvuloplasty was performed. In 2 patients, a suboptimal placement of the prosthesis with remaining aortic regurgitation had to be corrected by implantation of a second CoreValve prosthesis (prosthesis-in-prosthesis).
|
|
Follow-up clinical results. Overall mortality at 30 days was 12% in the intent-to-treat population with a combined rate of death, stroke and myocardial infarction of 22%. In patients with device and procedural success, the mortality was 9% and 5%, respectively (Table 4). With respect to the functional class, a remarkable relief of symptoms was observed with a decline from a mean New York Heart Association functional class of 2.85 ± 0.73 before to 1.85 ± 0.60 after valve implantation (p < 0.001).
|
|
|
| Discussion |
|---|
|
|
|---|
However, the CoreValve design provides several advantages that facilitate device deployment and reliable implantation: 1) a certain deployment error margin; 2) self-axing properties; 3) beneficial anchoring characteristics in native valve area as well as the ascending aorta; and 4) the ability for device retrieval after partial implantation of the first two-thirds of the prosthesis. This modified deployment technique, as opposed to the rapid complete deployment as favored in the beginning, has been introduced during the study course. Having deployed the distal two-thirds of the prosthesis (Fig. 2), the valve is already sufficiently functioning, whereas the device position can still be adjusted or the device can be pulled back completely.
The encouraging device success rate we have observed in this study goes along with a low procedural and 30-day mortality, which is lower than the predicted operative risk of these patients using the EuroScore risk algorithm (17). The accuracy and value of these risk assessment scores is sometimes controversially discussed. However, these tools are widely used for outcome stratification in high-risk surgical candidates in which large randomized clinical trails are missing. The population enrolled in our series presented with an average logistic EuroScore of 21.7%. The overall 30-day mortality of 12% as well as the 30-day mortality of 9% in patients with acute device success in our study is therefore encouraging, proving the safety of the procedure with regard to this important clinical end point. In addition, if procedural success is achieved, the 30-day event rate is just 6%.
However, there was a relatively high rate of adverse events at 30 days in the overall intent-to-treat population, largely driven by cerebrovascular events and cardiac tamponades. The latter were induced in 6 cases, most likely by wire perforations during valve insertion and placement. Interestingly, most of these events occurred in the middle of the transition phase from the 21-F to 18-F period, when the less-invasive approach without hemodynamic support was introduced and a new device deployment technique was established. In this phase, it was realized that meticulous distal wire tip control is of particular importance for this procedure, given the characteristics of very stiff wires with a suboptimal tip shape for this procedural setting and their known potential for perforations.
Consequently, to address this issue, mandatory manual wire shaping is now performed before insertion, forming a less traumatic pigtail-like wire tip. In combination with continuous fluoroscopic tip position control, this measure reduced the incidence of wire perforations effectively in the later study phase. A specifically manufactured wire addressing these specific requirements would be certainly helpful to avoid the risk of perforations, a potentially life-threatening event which can occur in surgically treated patients in up to 17% (cardiac effusion) and 4% (tamponade) (18).
The problem of cerebrovascular events when treating patients with severe aortic stenosis is also already known from both the era of balloon valvuloplasty as well as surgical series. The risk of a perioperative cerebrovascular event for patients age
80 years of age undergoing coronary bypass surgery or combined bypass surgery and aortic valve replacement has been reported to be as high as 10% and 15%, respectively (3). Therefore, an overall stroke rate of 10%, including minor and major events in our study, is comparable with surgical data in this clinical setting. However, further progress is certainly needed to sufficiently prevent these embolic events during this kind of percutaneous approach. Whether these events are caused by thrombi or liberated plaque particles from the native valve, the ascending aorta or aortic arch, or air emboli is currently unclear. Careful device preparation to avoid air emboli, optimal device positioning without extensive placement maneuvers within the native valve, as well as adequate antiplatelet medication is certainly mandatory to reduce the incidence of embolic events.
The hemodynamic results of this study clearly demonstrate the efficacy of this new technique. As soon as the valve is adequately implanted, there is a striking reduction of the transvalvular gradient, usually without significant regurgitations. However, dilation afterward to fully expand the prosthesis, mainly in heavily calcific degenerated valves, is sometimes needed (24% in this study) to achieve a good hemodynamic outcome. However, this step can be performed safely and reliably, even with slightly oversized balloons, without risk for structural device damage. If regurgitation is still detected, it is usually located in the paravalvular area. We found that the acute postprocedural grade of regurgitation can still change in the following days, perhaps as the result of factors such as the self-expanding properties of the device (improvement) or a kind of recoil of the valve segment due to heavy calcifications (worsening). However, we have not observed any significant clinical worsening of regurgitation in the entire population. More long-term data are certainly needed to assess the durability of the acute hemodynamic results. The degeneration pattern of this pericardial valve is expected to be comparable with common aortic valve bioprostheses.
Although the CoreValve technique is still in its infancy, device modifications and procedural advances are proceeding. The device profile reduction to the 18-F catheter resulted in remarkable procedural improvements without different safety outcomes. The procedural duration is significantly reduced as the result of a less-invasive technique without need for ventricular assist devices, general anesthesia, and surgical access-site preparations. This third-generation device allows now a truly percutaneous approach to aortic valve replacement which has the potential to change the standards of care particularly for high-risk surgical candidates with severe AS in the near future.
Study limitations. The current multicenter study describes only the short-term results after CoreValve implantation; assessment of the long-term durability of this prosthesis will require at least a 5-year follow-up. There was no control group in this study, which limited our ability to assess the device efficacy. The results of this study apply only to the patient population enrolled (high-risk patients with aortic stenosis and multiple comorbid conditions). Additional studies are required to determine the suitability of this device for patients who are otherwise good candidates for surgical aortic valve replacement and those with predominant aortic regurgitation.
| Conclusions |
|---|
|
|
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
80 years: results from the National Cardiovascular Network J Am Coll Cardiol 2000;35:731-738.This article has been cited by other articles:
![]() |
A. Vahanian, O. R. Alfieri, N. Al-Attar, M. J. Antunes, J. Bax, B. Cormier, A. Cribier, P. De Jaegere, G. Fournial, A. P. Kappetein, et al. Transcatheter valve implantation for patients with aortic stenosis: a position statement from the European Association of Cardio-Thoracic Surgery (EACTS) and the European Society of Cardiology (ESC), in collaboration with the European Association of Percutaneous Cardiovascular Interventions (EAPCI). Eur. J. Cardiothorac. Surg., July 1, 2008; 34(1): 1 - 8. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Kempfert, J. M. Blumenstein, M. A. Borger, A. Linke, S. Lehmann, P. Pritzwald-Stegmann, M. W.A. Chu, G. Schuler, V. Falk, F. W. Mohr, et al. Minimally invasive off-pump valve-in-a-valve implantation: the atrial transcatheter approach for re-operative mitral valve replacement Eur. Heart J., June 20, 2008; (2008) ehn285v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Walther, V. Falk, J. Kempfert, M. A. Borger, J. Fassl, M. W.A. Chu, G. Schuler, and F. W. Mohr Transapical minimally invasive aortic valve implantation; the initial 50 patients Eur. J. Cardiothorac. Surg., June 1, 2008; 33(6): 983 - 988. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Zegdi, P. Achouh, and J.-N. Fabiani Percutaneous aortic valve replacement with the CoreValve bioprosthesis. J. Thorac. Cardiovasc. Surg., June 1, 2008; 135(6): 1407 - 1408. [Full Text] [PDF] |
||||
![]() |
F. Descoutures, D. Himbert, L. Lepage, B. Iung, D. Detaint, D. Tchetche, E. Brochet, Y. Castier, J.-P. Depoix, P. Nataf, et al. Contemporary surgical or percutaneous management of severe aortic stenosis in the elderly Eur. Heart J., June 1, 2008; 29(11): 1410 - 1417. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Antunes Percutaneous aortic valve implantation. The demise of classical aortic valve replacement? Eur. Heart J., June 1, 2008; 29(11): 1339 - 1341. [Full Text] [PDF] |
||||
![]() |
A. Vahanian, O. Alfieri, N. Al-Attar, M. Antunes, J. Bax, B. Cormier, A. Cribier, P. De Jaegere, G. Fournial, A. P. Kappetein, et al. Transcatheter valve implantation for patients with aortic stenosis: a position statement from the European Association of Cardio-Thoracic Surgery (EACTS) and the European Society of Cardiology (ESC), in collaboration with the European Association of Percutaneous Cardiovascular Interventions (EAPCI) Eur. Heart J., June 1, 2008; 29(11): 1463 - 1470. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Piazza, Y. Onuma, E. Jesserun, P. P. Kint, A.-M. Maugenest, R. H. Anderson, P. P. T. de Jaegere, and P. W. Serruys Early and Persistent Intraventricular Conduction Abnormalities and Requirements for Pacemaking After Percutaneous Replacement of the Aortic Valve J. Am. Coll. Cardiol. Intv., June 1, 2008; 1(3): 310 - 316. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Wendt, M. Thielmann, T. Buck, R.-A. Janosi, T. Bossert, N. Pizanis, M. Kamler, and H. Jakob First clinical experience and 1-year follow-up with the sutureless 3F-Enable aortic valve prosthesis Eur. J. Cardiothorac. Surg., April 1, 2008; 33(4): 542 - 547. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Grunenfelder, A. Plass, H. Alkadhi, and M. Genoni Evaluation of biological aortic valve prostheses by dual source computer tomography and anatomic measurements for potential transapical valve-in-valve procedure Interactive CardioVascular and Thoracic Surgery, April 1, 2008; 7(2): 195 - 200. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Webb Percutaneous Aortic Valve Replacement Will Become a Common Treatment for Aortic Valve Disease J. Am. Coll. Cardiol. Intv., April 1, 2008; 1(2): 122 - 126. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Walther, J. Kempfert, M. A. Borger, J. Fassl, V. Falk, J. Blumenstein, M. Dehdashtian, G. Schuler, and F. W. Mohr Human Minimally Invasive Off-Pump Valve-in-a-Valve Implantation Ann. Thorac. Surg., March 1, 2008; 85(3): 1072 - 1073. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Murtuza, J. R. Pepper, R. DeL Stanbridge, C. Jones, C. Rao, A. Darzi, and T. Athanasiou Minimal Access Aortic Valve Replacement: Is It Worth It? Ann. Thorac. Surg., March 1, 2008; 85(3): 1121 - 1131. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. N. DeMaria, J. J. Bax, O. Ben-Yehuda, P. Clopton, G. K. Feld, G. S. Ginsburg, B. H. Greenberg, J. D. Knoke, W. Y.W. Lew, J. A.C. Lima, et al. Highlights of the year in JACC 2007. J. Am. Coll. Cardiol., January 29, 2008; 51(4): 490 - 512. [Full Text] [PDF] |
||||
![]() |
R. Zegdi, G. Sleilaty, A. Lafont, and J.-N. Fabiani Percutaneous Aortic Valve Replacement With the CoreValve Prosthesis J. Am. Coll. Cardiol., January 15, 2008; 51(2): 170 - 170. [Full Text] [PDF] |
||||
![]() |
E. Grube, U. Gerckens, P. Wenaweser, and L. Buellesfeld Reply J. Am. Coll. Cardiol., January 15, 2008; 51(2): 170 - 171. [Full Text] [PDF] |
||||
![]() |
W. J. Stewart Imaging the Future of Transcatheter Aortic Valve Replacement J. Am. Coll. Cardiol. Img., January 1, 2008; 1(1): 25 - 28. [Full Text] [PDF] |
||||
![]() |
Promising Initial Results with CoreValve Percutaneous Aortic Valve Placement Journal Watch Cardiology, July 25, 2007; 2007(725): 3 - 3. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |