J Am Coll Cardiol, 2007; 49:1965-1973, doi:10.1016/j.jacc.2006.10.080
(Published online 30 April 2007). © 2007 by the American College of Cardiology Foundation |
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
University of California and Veterans Administration Medical Centers, San Diego, California.
Manuscript received March 8, 2006; revised manuscript received October 12, 2006, accepted October 17, 2006.
* Reprint requests and correspondence: Dr. Sanjiv M. Narayan, Veterans Affairs and University of California Medical Centers-San Diego, Cardiology/111A, 3350 La Jolla Village Drive, San Diego, California 92161. (Email: snarayan{at}ucsd.edu).
| Abstract |
|---|
|
|
|---|
Background: Focal AT often cannot be distinguished from macrore-entrant AT until the time of electrophysiology study (EPS). We hypothesized that quantitative ECG metrics should separate focal AT, using its short activation relative to tachycardia cycle length (CL), from macrore-entrant AT, whose activation should span the CL. We developed tools to accurately quantify CL and P- or F-wave duration even when overlying T waves, then prospectively applied them to patients during focal or macrore-entrant AT ablation and compared them to the gold standard EPS diagnosis.
Methods: We studied 41 patients (27 men, 14 women) age 57 ± 17 years. In the training group (n = 20), tachycardia P or F waves overlying T waves were identified from transitions in slope (dV/dt) relative to "expected" T waves generated from scaling of the sinus-rate T-wave. Electrocardiographic P-wave duration agreed with the duration of intra-atrial activation. Autocorrelation was used to estimate ECG atrial CL (p < 0.001).
Results: Compared to macrore-entry (n = 13), focal AT (n = 7) had shorter P waves (115 ± 31 ms vs. 227 ± 67 ms; p < 0.001) that were smaller ratios of CL (28 ± 7% vs. 85 ± 21%; p < 0.001). Receiver-operating characteristic curve areas for AT were 0.92 for P(F)-wave duration and 0.99 for P(F)/CL ratio. On blinded prospective analysis (n = 21), P(F)-wave duration <160 ms identified focal (n = 7) from macrore-entrant AT (n = 14) with 90% sensitivity and 90% specificity, and a P(F)/CL ratio <45% gave 86% sensitivity and 98% specificity.
Conclusions: Quantitative ECG indexes of shorter atrial activation and longer diastolic interval separate focal from macrore-entrant AT without diagnostic maneuvers.
| ||||||||
In principle, differences between focal and macrore-entrant AT should be detectable at the bedside using the ECG. An established mechanistic feature of macrore-entry is that activation spans the tachycardia cycle length (CL), whereas activation in focal AT is shorter (1,2). However, difficulties in visually measuring P or F waves across leads or when overlying T waves hinder accurate ECG estimation of the P(F)-wave duration or CL.
Therefore, we developed ECG tools to precisely measure atrial CL and P- or F-wave duration even if overlying T waves. We hypothesized that these quantitative ECG indexes in focal AT should reveal shorter P waves because of activation of a smaller ratio to CL that reflects longer diastolic intervals compared to F waves in macrore-entrant AT. We used receiver-operating characteristic curves to find optimal cutpoints in a training study, then applied these cutpoints prospectively to patients with focal and macrore-entrant AT referred to the diagnostic gold standard of EPS.
| Methods |
|---|
|
|
|---|
All patients underwent clinical EPS in the post-absorptive state after discontinuing anti-arrhythmic medications except amiodarone (Table 1). Routine diagnostic catheters were advanced transvenously, including a 6-F quadrapolar catheter to the His bundle position, a 6-F decapolar catheter into the coronary sinus, an 8-F ablation catheter (EP Technologies, Sunnyvale, California) for mapping, and other catheters as clinically required.
|
Acquisition of data. We recorded 12-lead surface ECGs (0.05 to 100 Hz bandpass-filtered) and corresponding bipolar intracardiac electrograms (30 to 250 Hz) that were digitized at 1 kHz and exported with 16-bit resolution from our recorder (Bard, Billerica, Massachusetts). Electrocardiographic analysis was performed on a personal computer using software developed by the authors in Labview (National Instruments, Austin, Texas) (6). Electrocardiographic analysis used leads V5, aVF, and V1 to represent orthogonal leads X, Y, and Z, respectively. Intracardiac electrograms were analyzed at 200 mm/s scale.
ECG identification of tachycardia P or F waves. We attempted to identify P or F waves in each orthogonal ECG lead (V5, aVF, or V1) regardless of the conducted atrial:ventricular ratio. When tachycardia P or F waves overlay T waves, they were revealed by comparison against "expected" T waves, computed for each patient using rate-adjusted linear scaling from the sinus-rate T-wave (Fig. 1).
|
(sinus RR interval [s])} then scaled this to the tachycardia rate {QTe = QTc ·
(tachycardia RR interval (s)} (7) (Fig. 1). Second, we linearly scaled the sinus rhythm T-wave to yield an expected T-wave shape at the tachycardia rate (Fig. 1). Assuming that QRS duration (QRSd) does not alter significantly between sinus rhythm and tachycardia, the "expected" JT duration (JTe) = QTe QRSd. Thus, the expected T-wave was generated by sampling the sinus T-wave (duration JTs = QTs QRSd) every JTs/JTeth point. For example, if tachycardia JTe is half JTs, every second sinus T-wave point was sampled to yield the expected tachycardia T-wave (Fig. 1).
Finally, we compared observed with expected tachycardia T waves using subtraction to identify the obscured P- or F-wave. Figure 1 shows this for a case of focal AT.
Determining P- or F-wave onset and offset. Slope transitions from the isoelectric baseline or T-wave (if superimposed) were used to reveal subtle changes in contour indicating P(F) onset and offset. Figure 1 shows dV/dt of tachycardia T waves in a case of focal AT, smoothed by applying a 3-point median filter. The upstroke, turning point, and downstroke in X-axis dV/dt (lead V5) correspond with visual P-wave onset, peak, and offset. We applied dV/dt separately for each lead, then expressed measurements in each axis and as the 3-axis mean. Leads in which P- or F-wave onset or offset remained obscure were not included. The 3-lead mean of P(F)-wave measurements was reported for each patient.
ECG estimates of tachycardia cycle length. Because manual CL measurement may introduce errors, given difficulties in precisely identifying P- or F-wave onsets, we used novel autocorrelation and spectral methods to quantify CL.
First, atrial activity was emphasized using our described sliding correlation method (6,8). Briefly, a 120-ms ECG template was selected overlying a P- or F-wave (Figs. 2A.I and 2B.I). Sequential Pearson cross-correlations of this template across the ECG, indicated by the dashed templates in Figure 2A.I (left), yield a correlation-time series ranging from 1 to + 1 in which values near 1.0 indicate recurrent P or F waves. This correlation-time series reflects atrial activity across the ECG.
|
Spectral CL estimates were obtained from an 8,192-point fast Fourier transform (0.12 Hz resolution) (Figs. 2A.III and 2B.III). The dominant frequency was the largest magnitude between 2 and 6.8 Hz (reflecting 500 to 150 ms), and estimated CL was given by its reciprocal. Cycle length estimates were validated against measured atrial CL for 10 cycles at 200 mm/s scale.
Diagnostic criteria for focal AT and receiver-operating characteristics. In the training population (n = 20), we derived ECG P- or F-wave durations and P(F) duration-to-CL ratios to diagnose focal AT. Using Excel (Microsoft, Inc., Redmond, Washington), we created receiver operating characteristic curves for the diagnosis of focal AT to determine the optimal cutpoint for each index. This was applied in a blinded fashion in our pilot validation study (n = 21 patients).
Statistical analysis and sample size considerations for the validation study. Continuous variables were presented as mean ± SD. The 2-tailed t test was used to compare continuous variables between groups. The Fisher exact test was applied to contingency tables. Because of the small sample sizes, exact confidence limits for sensitivity and specificity were computed using binomial probabilities. Probabilities below 5% (p < 0.05) were considered significant.
For the diagnosis of focal AT in the training population, P-wave duration <160 ms had an accuracy of 80%, and P-wave duration to tachycardia CL <45% had an accuracy of 95%. To estimate sample sizes required for a prospective study of these tests, assuming an accuracy of 80%, a validation study would require 20 subjects to separate focal AT from macrore-entrant AT with 80% power at the 5% level of significance.
| Results |
|---|
|
|
|---|
Differences in sinus rhythm ECG. P-wave durations (3-axis means) in sinus rhythm were shorter in focal AT (120 ± 33 ms) than macrore-entrant AT (163 ± 39 ms; p < 0.001), despite no difference in sinus CL between groups (Table 2). For focal AT patients, P-wave durations did not differ between sinus rhythm and tachycardia, even though the P-axis typically changed (p = 0.86). Conversely, for macrore-entrant AT patients, F-wave durations in tachycardia were significantly longer than P-wave durations in sinus rhythm (p < 0.001).
|
Training group: ECG estimated P(F)-wave duration and ratio to CL (Table 2). In the training group (focal AT, n = 7; macrore-entrant AT, n = 13), P-wave duration was substantially shorter in patients with focal AT compared to F-wave duration in macrore-entrant AT (p < 0.001 in each orthogonal axis) (Table 2). There was no significant difference in ventricular rate between groups (134 ± 51 beats/min vs. 114 ± 56 beats/min; p = 0.48).
Figure 3A shows an activation map of focal AT in the lateral right atrium in a 28-year-old man. From the ECG, the measured P-wave duration (3-axis mean) was 127 ms, representing 43% of autocorrelation (estimated CL 284 ms). Detailed intracardiac mapping (NavX, Endocardial Solutions, Minneapolis, Minnesota) corroborated these values, showing that the extent of atrial activation was 125 ms, with measured CL 274 ms.
|
For the training group (Table 2), compared to macrore-entrant AT, focal AT had a shorter P(F)-wave duration in all axes (p < 0.001), which represented a smaller ratio of CL (p < 0.001). For P-wave duration <160 ms (3-axis mean), sensitivity for AT was 90% (3-axis mean 95% confidence interval [CI] 0.70 to 0.99); specificity was 77% (95% CI 0.61 to 0.89). For P-wave duration to CL <45%, sensitivity for AT was 100% (95% CI 0.82 to 1.0); specificity for AT was 90% (95% CI 0.76 to 0.97).
Receiver-operating characteristics for focal AT. Receiver-operating characteristic curves for the diagnosis of focal AT were generated from the training population. The area under the curve for P(F)-wave duration was 0.92 (Fig. 4), and P-wave duration <160 ms provided sensitivity 91%, specificity 89%, positive predictive value 83%, and negative predictive value 95% for focal AT (Table 3). The P(F) wave-to-CL ratio provided area under curve = 0.99 (Fig. 5), and ratio <45% provided sensitivity 86%, specificity 98%, positive predictive value 95%, and negative predictive value 91% for focal AT.
|
|
|
|
| Discussion |
|---|
|
|
|---|
Electrophysiologic separation of focal AT from macrore-entrant AT. Shorter atrial activation and longer diastolic intervals are a robust identifier of focal AT from macrore-entrant AT (1). At EPS, however, high-spatial-resolution mapping may be required to define the duration of intra-atrial activation and determine whether it is focal or continuous (Fig. 3). Because the ECG represents bi-atrial activation, P or F waves ideally represent the duration of atrial activation. These ECG results confirm that focal AT can be separated from macrore-entrant AT by short atrial activation and a long diastolic interval. Accordingly, activation in focal AT was not prolonged compared to sinus rhythm, reflecting the long diastolic time of each, but was prolonged relative to sinus rhythm in macrore-entrant AT. We applied our indexes to 3 orthogonal leads because P or F waves may not be represented equally in all leads (5,9,10), although inter-lead variations were minimal (Table 2).
These indexes could be influenced by many factors. First, left atrial sizes were similarly and mildly dilated in each group (Table 1), although structural remodeling may be more likely in patients with macrore-entrant AT (11). Second, none of our focal AT patients, but one-third of our macrore-entrant AT patients, had a history of AF (Table 1). Remodeling could explain the prolonged sinus P waves, and possibly tachycardia F-waves, seen in macrore-entrant AT patients (12,13). Third, differences in ventricular rates could theoretically influence our results. However, our methods were effective whether ventricular rates differed (validation study) or did not differ (training study) between groups. Fourth, the relatively well-preserved left ventricular ejection fractions in both groups largely exclude heart failure-related atrial remodeling (14). Nevertheless, these methods should be applied to patients with a significant history of prior AF or heart failure.
Prior ECG estimates of P(F)-wave duration and tachycardia CL. Few methods have been described to separate focal from macrore-entrant AT from the ECG, possibly because of difficulties in precisely measuring P- or F-wave onset, particularly if superimposed on T waves. Prior studies mostly compared P- or F-wave shape in AT (9,10) or macrore-entrant AT (4,5), although the waves are known to overlap. One study of selective pulmonary vein pacing to simulate focal AT (10) reported P-wave durations of 110 to 130 ms, in close agreement with our results. However, that study examined only P waves unobscured by T waves. Those data, and comparisons of our P(F)-wave durations against intra-cardiac atrial activation times (Fig. 3), validate and strengthen the clinical utility of our approach.
Spectra have been used to estimate CL from the ECG (15,16), yet they were inaccurate in this study, hindered by harmonics and noise (Fig. 2B.III). For this reason, our novel use of autocorrelation more accurately estimated CL.
Existing methods to identify and separate focal from macrore-entrant AT. Visible ECG criteria for separating focal from macrore-entrant AT have largely been discarded in favor of mapping at EPS (1) because many P- or F-wave shapes, including even sawtooth patterns (1), may occur in focal AT. Although F waves may have low amplitude in macrore-entrant AT, particularly if left atrial (5), this can also occur in focal AT (1,2,10). Electrocardiogram shape criteria may thus confuse focal with macrore-entrant AT (1) and multifocal AT with AF (17), and even atrial rates are similar between arrhythmias, as supported by this study (1) (Table 1).
Clinical significance. Accurate noninvasive identification of focal AT may enable better prognostication because AF often follows macrore-entrant AT (18) and requires anticoagulant as well as anti-arrhythmic drugs. Because ablation approaches for focal and macrore-entrant AT ablation may differ, noninvasive diagnosis may also guide ablation from the bedside 12-lead ECG. Identifying focal AT from the ECG enables the use of P-wave polarity to predict the site of tachycardia origin (9). Notably, polarity may not effectively localize macrore-entry because the exit direction of activation is not predictable for many circuit locationswitness the positive F-wave in lead V1 in both typical right atrial flutter and left atrial macrore-entry.
Study limitations. This is a small study and, in particular, these results require validation in patients after ablation and with coexisting AF and heart failure, in whom conduction slowing may further prolong P waves at baseline (14). Second, macrore-entrant AT patients had higher amiodarone usage than those with focal AT, which may have lengthened tachycardia F waves or sinus rhythm P waves. However, 75% of macrore-entrant AT patients did not use amiodarone, and analysis of P(F)-wave durations excluding patients taking amiodarone still showed a highly significant difference between focal and macrore-entrant AT (p < 0.001 for each lead, data not shown). Third, P(F)-wave durations overlying T waves were intended as estimates, although, in fact, they approximated atrial activation times from detailed mapping (Fig. 3). Generating an "expected" T-wave by linear scaling from sinus rhythm assumes that the T-wave does not change in shape as rates vary. Although these simplifications were empirically successful, T-wave dynamics are far more complex (19). Future work could thus use the T-wave at the closest rate from a series of native or paced atrial rates. Alternatively, atrial separation could be achieved via differences between sinus P-wave (20) and T-wave (19) spectra, although spectral differences between P waves in focal AT and F waves in macrore-entrant AT (8) are less well described.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|