J Am Coll Cardiol, 2007; 49:1474-1481, doi:10.1016/j.jacc.2006.11.040
(Published online 20 March 2007). © 2007 by the American College of Cardiology Foundation |
,*






,
* Department of Dermatology, Harvard Medical School and Massachusetts General Hospital, Boston, Massachusetts
Department of Radiology, Harvard Medical School and Massachusetts General Hospital, Boston, Massachusetts
Department of Pathology, Harvard Medical School and Massachusetts General Hospital, Boston, Massachusetts
Wellman Center for Photomedicine, Harvard Medical School and Massachusetts General Hospital, Boston, Massachusetts.
|| Department of Emergency Medicine and Department of Anesthesiology, University of Massachusetts Medical School, Worcester, Massachusetts.
Manuscript received August 3, 2006; revised manuscript received October 31, 2006, accepted November 27, 2006.
* Reprint requests and correspondence: Dr. Seemantini K. Nadkarni, Harvard Medical School, Massachusetts General Hospital and Wellman Center for Photomedicine, 40 Blossom Street, BAR 718, Boston, Massachusetts 02114. (Email: snadkarni{at}hms.harvard.edu).
| Abstract |
|---|
|
|
|---|
Background: A method capable of evaluating plaque collagen content and SMC density can provide a measure of the mechanical fidelity of the fibrous cap and can enable the identification of high-risk lesions. Optical coherence tomography has been demonstrated to provide cross-sectional images of tissue microstructure with a resolution of 10 µm. A recently developed technique, PSOCT measures birefringence, a material property that is elevated in tissues such as collagen and SMCs.
Methods: We acquired PSOCT images of 87 aortic plaques obtained from 20 human cadavers. Spatially averaged PSOCT birefringence,
, was measured and compared with plaque collagen and SMC content, quantified morphometrically by picrosirius red and smooth muscle actin staining at the corresponding locations.
Results: There was a high positive correlation between PSOCT measurements of
and total collagen content in all plaques (r = 0.67, p < 0.001) and in fibrous caps of necrotic core fibroatheromas (r = 0.68, p < 0.001). Polarization-sensitive optical coherence tomography measurements of
demonstrated a strong positive correlation with thick collagen fiber content (r = 0.76, p < 0.001) and SMC density (r = 0.74, p < 0.01).
Conclusions: Our results demonstrate that PSOCT enables the measurement of birefringence in plaques and in fibrous caps of necrotic core fibroatheromas. Given its potential to evaluate collagen content, collagen fiber thickness, and SMC density, we anticipate that PSOCT will significantly improve our ability to evaluate plaque stability in patients.
| ||||||||||
A new imaging technique, termed polarization-sensitive optical coherence tomography (PSOCT), has been developed that may enable the quantification of collagen and SMC content in atherosclerotic plaques. Optical coherence tomography (OCT) is a high-resolution (
10 µm) imaging method that has demonstrated high accuracy for the characterization of plaque microstructure, identification of thin-cap fibroatheromas, and quantification of macrophage content in vivo (9,10). Polarization-sensitive OCT enhances conventional OCT by measuring tissue birefringence (1113), a material property that is elevated in tissues containing proteins with an ordered structure, such as organized collagen and SMC actin-myosin. A stable plaque is associated with high collagen content, thicker collagen fibers, and large numbers of SMCs. In contrast, unstable plaques likely have lower collagen content, thinner collagen fibers, and fewer SMCs (5,6,8). In this study, we investigated the relationship between PSOCT measurements of birefringence and collagen content, fiber thickness, and SMC density in atherosclerotic plaques ex vivo. Results from this work will aid in determining the suitability of this technology for providing information on these critical plaque constituents in patients.
| Materials and Methods |
|---|
|
|
|---|
A total of 87 aortic plaques were randomly obtained from 20 human cadavers and imaged using PSOCT. The time between death and imaging ranged from 12 to 24 h. The specimens were cut open and placed in a phosphate-buffered saline bath maintained at 37°C during imaging. To ensure accurate registration with histopathology, the imaging site was marked with 2 India ink spots at the edges of the OCT scan. Polarization-sensitive OCT and conventional OCT images were simultaneously obtained within 1 s, and all images were 5 x 1.2 mm (2,000 tranverse pixels x 256 depth pixels).
Birefringence measurement using PSOCT.
The imaging system in the current study provides both conventional OCT and PSOCT images with a single lateral scan. When light traverses birefringent tissue, light polarized along directions parallel and perpendicular to the fiber orientation of the tissue travels at different velocities, incurring a relative phase retardation,
, which results in a change in its polarization state. The phase retardation,
, accumulates as light travels deeper through tissue at a rate proportional to the magnitude of birefringence. To obtain a PSOCT image, mathematical techniques described in previous work (15) are used to measure Stokes parameters at each depth and depth-resolved
values are determined. The accumulated phase retardation,
(L), at each depth, L, is then displayed with respect to the tissue surface as a grayscale image with black corresponding to 0° (at the tissue surface) and white to 180° (or 360°) (Figs. 1B and 1F).
|
(L), was calculated by averaging across the width of the ROI at each depth, L. Tissue birefringence was then measured over each ROI by calculating the slope,
L) of the ROI (Fig. 2).
|
Collagen content was measured morphometrically using digitized circularly polarized light microscopy images of PSR-stained sections within a software-selected 500 x 200 µm ROI centered between the ink marks (8,18). Because mechanical stability of the plaque may also depend on collagen fiber thickness, morphometric differentiation between thicker (orange-red) and thinner (yellow-green) collagen fibers was conducted (19,20). A hue saturation value color coordinate transformation was performed, and color separation of thick and thin collagen fibers was achieved by dividing the number of hue values within a range corresponding to orange-red (thick fibers) and yellow-green (thin fibers) by the total pixel area of the ROI (20,21). Total collagen content was computed by summing the percentage of orange-red and yellow-green fibers within the ROI. Smooth muscle cell content was measured using digitized immunohistochemistry sections stained for alpha-smooth muscle actin, and the percent staining within the ROI was calculated using automated bimodal histogram thresholding and image segmentation (22).
Statistical analysis.
Multiple regression analysis was performed to test the relationship between PSOCT birefringence,
, with histologic measurements of thick collagen fiber, thin collagen fiber, and SMC content using stepwise regression. The correlation of PSOCT birefringence,
, with each independent variable was also computed using linear regression analysis. For all analyses, a p value <0.05 was considered statistically significant. All statistical analyses were performed using MicroCal Origin (MicroCal Software Inc., Northampton, Massachusetts).
| Results |
|---|
|
|
|---|
Collagen birefringence.
Figure 1 shows illustrative PSOCT images with the corresponding histopathology of 2 atherosclerotic plaques. The fibrous plaque in Figures 1A to 1D shows abundant thicker (orange-red in Fig. 1C) collagen fibers, constituting 88% of the ROI. In the corresponding PSOCT image (Fig. 1B), the presence of these highly birefringent collagen fibers causes a sharp transition from black (
= 0°) at the tissue surface to white (
= 180°). The fibroatheroma in Figures 1E to 1H shows evidence of depleted collagen constituting 20% of the ROI, with small amounts of thin collagen fibers (yellow-green in Fig. 1G). Low birefringence in PSOCT presents as a black region corresponding to low
(Fig. 1F). In Figure 2,
values averaged over the width of the central ROI in the two PSOCT images shown in Figures 1B and 1F are plotted as a function of depth. The
value remains low at the surface and increases at different rates with depth. Least-squares fits over a depth of 200 µm for each plot indicate that for the highly birefringent fibrous plaque constituting 88% collagen (Figs. 1A to 1D),
(L) increases at a rate of
= 0.60°/µm, compared with a significantly lower rate of
= 0.04°/µm in the plaque constituting 20% collagen (Figs. 1E to 1H). Figure 3
shows illustrative PSOCT images of a NCFA with the corresponding histology sections. The OCT image in Figure 3A shows a signal-poor lipid pool with poorly delineated borders beneath a signal-rich band corresponding to the fibrous cap. Birefringence of the fibrous cap is seen in the PSOCT image (Fig. 3B) as a transition from black at the surface to gray. Beneath the fibrous cap, the necrotic core appears noisy because low signal from the lipid pool causes PSOCT measurements of
to be unreliable in this region.
|
for the 73 atherosclerotic plaques. Linear regression analysis showed good correlation between total collagen content and PSOCT measurements (r = 0.67, p < 0.0001). Twenty-two atherosclerotic plaques that were histologically classified as NCFAs had average cap thicknesses ranging from 75 to 490 µm, and the minimum cap thickness ranged from 26 to 215 µm. Linear regression analysis showed good correlation between total collagen content and
in fibrous caps of NCFAs (r = 0.68, p < 0.001) (Fig. 4B).
|
(thick collagen fibers, p < 0.0001, and thin collagen fibers, p < 0.05). In Figures 4C and 4D,
is plotted versus thick and thin collagen fiber content, respectively. Linear regression analysis demonstrated a high positive correlation between thick collagen fiber content and
(r = 0.76, p < 0.0004) and an inverse relationship between thin collagen fiber content and
(r = 0.48, p < 0.0001).
SMC birefringence.
Figure 5
illustrates a fibrous plaque with low collagen content (
7%) (Fig. 5C); however, the corresponding PSOCT image (Fig. 5B) shows a rapid transition from black to white, indicating high birefringence. The accompanying immunohistology section shows extensive alpha-smooth muscle actin staining, confirming the presence of SMCs constituting 55% of the ROI (Fig. 5D), with
= 0.46°/µm. The results of the multiple regression analysis showed that SMC content, when incorporated in the model jointly with collagen content, did not demonstrate a statistically significant relationship with
(p = 0.25). When SMC content was considered individually using linear regression analysis, we found a statistically significant correlation between SMC content and
(r = 0.34, p < 0.01). To evaluate the relationship between SMC content and
, independent of collagen content, we analyzed plaques with low collagen content (<50% of the ROI: n = 30 plaques). For these plaques, SMC content showed high correlation with
(r = 0.74, p < 0.01) (Fig. 6).
|
|
| Discussion |
|---|
|
|
|---|
In this study, we measured tissue birefringence from PSOCT images of human atherosclerotic plaques. The PSOCT birefringence,
, was highly related to total collagen content in all atherosclerotic plaques, as well as in fibrous caps of NCFAs.
also demonstrated a strong positive correlation with thick collagen fiber content and a negative correlation with thin collagen fiber content. The inverse relationship between
and thin collagen fibers may be explained by the replacement of highly birefringent, thick organized fibers by thin, more randomly oriented fibers. In support of this hypothesis, we found an inverse relationship between thick and thin collagen fibers. Studies have shown that PSOCT measurements of birefringence are influenced by collagen fiber organization;
is low when collagen fibers are less organized and high when fibers are well organized (27). Recent work in animals suggests that thicker collagen fibers in intimal lesions are more aligned and circumferentially oriented than thinner collagen fibers, which show increased structural disorganization (8). This difference in fiber organization of thick and thin collagen fibers is, therefore, a plausible explanation for our findings that PSOCT birefringence was related positively with thick collagen content and inversely with thin collagen fiber content. We also found that fibrous cap thickness in NCFAs did not bear a statistically significant relationship with PSOCT birefringence (p = 0.81), suggesting that the birefringence measured from PSOCT images may be independent of fibrous cap thickness.
Birefringence exhibited by intimal SMCs may be attributed to their structural construct, comprising a myosin containing thick filament co-assembled with an actin-containing thin filament (28). The results of our multiple regression analysis showed that when considered jointly with collagen content, SMC content did not demonstrate a significant relationship with
. Because plaques in our dataset contained both collagen and SMCs, and because of the strong positive correlation of thick collagen fibers with
, it is possible that the variable SMC content dropped out of the multiple regression model. However, when plaques with low collagen content were analyzed separately,
showed high positive correlation with SMC content, demonstrating that PSOCT measures birefringence exhibited by SMCs. Previous work has shown that the normalized standard deviation calculated within OCT images is highly correlated with macrophage content (29). In a separate analysis, we found a statistically significant inverse correlation (r = 0.6, p < 0.005) between the normalized standard deviation computed from OCT images within fibrous caps of NCFAs and
in this study. This inverse relationship may be explained by the possible depletion of collagen by metalloproteinases associated with increased numbers of macrophages. In support of this hypothesis, we found a similar inverse relationship between cap collagen content and OCT-normalized standard deviation (r = 0.56, p < 0.01). Taken together, our results indicate that PSOCT birefringence may provide a powerful new index related to plaque stability. Because increased birefringence was correlated to abundant thick collagen fibers and/or SMCs, the detection of high birefringence in PSOCT images may imply increased plaque stability. Conversely, low PSOCT birefringence may indicate compromised plaque stability owing to low collagen content, fewer thick collagen fibers, and/or reduced numbers of SMCs. In all images, phase retardation angles were averaged over the ROI to reduce measurement uncertainty at each depth and facilitate more accurate linear fits for measuring
. We observed a significant reduction in statistical uncertainty of depth-resolved phase retardation angles by averaging over 500 µm.
Previous studies have shown that OCT measures multiple factors contributing to plaque instability, including detecting NCFAs, measuring microstructural details such as thin fibrous caps, and identifying cholesterol crystals and quantifying macrophage content (10,17,29). Polarization-sensitive OCT images are always obtained simultaneously with conventional OCT images, providing additional measurements of birefringence related to collagen and SMC content. Thus, OCT enhanced with the capability for PSOCT imaging can identify multiple factors associated with plaque rupture to provide a more comprehensive understanding of plaque stability.
Study limitations.
The effects of arterial pulsation on PSOCT measurements have not been analyzed here and warrant further investigation. However, on the basis of prior clinical studies in patients, we have observed that plaque birefringence is preserved during coronary pulsation. New technology has been developed to enable high-speed PSOCT imaging for clinical applications in the near future (30). Intimal plaque composition of aortas is quite similar to that of coronaries; however, there are significant differences in the media for the 2 arterial types. For this reason, we selected our ROIs to ensure that they were contained only within the intima, and we anticipate that our results can be generalized to coronary lesions. To minimize the influence of tissue shrinkage attributable to formaldehyde fixation and histologic processing (31) on the data, collagen and SMC content in histologic sections were measured as a percentage of the total area of the ROI. Registration of PSOCT images with corresponding histology sections was performed using fiducial ink marks at the lesion site applied on the plaque using a 26-gauge needle (450 µm); hence, we anticipate that the registration precision between PSOCT and histology was
500 µm.
Cholesterol crystals, which appear as linear signal-rich regions, may offer another source of birefringence in PSOCT images (Fig. 3). In OCT images, cholesterol crystals can be easily distinguished from other plaque components by their linear and highly reflecting appearance. Three FC plaques in our analysis contained calcific nodules beneath the 200 µm deep ROI, and these regions of calcification were not included in the analysis. Calcific nodules did not contain enough signal for PSOCT measurements, and, therefore, the birefringence of calcifications could not be established. However, calcific nodules can be easily detected with high sensitivity in OCT images as sharply delineated regions within the plaque having signal-poor interiors (17). Likewise, birefringence in the necrotic cores of NCFAs could not be measured using PSOCT. When light enters the core, because of large variance in scattering structures in the necrotic debris the polarization state of light becomes randomized after multiple scattering events, resulting in unreliable phase retardation measurements (32) (Fig. 3). Deep within the lipid pool, the signal is greatly attenuated, and PSOCT measurements of birefringence cannot be obtained.
A recent study (13) has demonstrated a qualitative assessment of plaque collagen using OCT by displaying resulting changes in backreflected light achieved by manually altering the incident polarization state. The PSOCT system used in our current study provides a quantitative evaluation of plaque birefringence,
, by measuring the accumulated phase retardation as light travels through birefringent tissue. Our current system measures
independent of the incident polarization state of light and sample orientation, thus facilitating intracoronary imaging in patients using catheters similar to those used in conventional intracoronary OCT (9,10). A previous feasibility study has demonstrated intracoronary PSOCT in ex vivo coronary arteries using rotary scanning fiber-optic catheters identical to those used in previous clinical trials (33). The strength of this technique may be further emphasized by noting that PSOCT birefringence is measured in a cross-sectional image, allowing evaluation of discrete microanatomic structures such as fibrous caps. Natural history studies are underway to address questions regarding the role of systemic and local predictors of plaque rupture risk. If the results of these studies show that focal plaque stabilization provides clinical benefits, therapeutic intervention could be guided by information such as that provided by PSOCT, potentially improving patient outcome.
Polarization-sensitive OCT is unique in that it provides images of birefringence, which are co-registered with high-resolution cross-sectional images of plaque morphology obtained by conventional OCT. Beyond the measurement of cap thickness, PSOCT provides additional information about the composition of plaques and NCFA fibrous caps, where low birefringence likely indicates increased instability. Given the potential significance of the additional information provided by PSOCT, and its promise for intracoronary application, we anticipate that this technology will be useful for improving our understanding of the mechanisms of plaque progression and rupture and for the detection of high-risk plaques before the occurrence of an acute coronary event.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
O C Raffel, T Akasaka, and I-K Jang Cardiac optical coherence tomography Heart, September 1, 2008; 94(9): 1200 - 1210. [Full Text] [PDF] |
||||
![]() |
C. Hu, A. Dandapat, L. Sun, J. Chen, M. R. Marwali, F. Romeo, T. Sawamura, and J. L. Mehta LOX-1 deletion decreases collagen accumulation in atherosclerotic plaque in low-density lipoprotein receptor knockout mice fed a high-cholesterol diet Cardiovasc Res, July 15, 2008; 79(2): 287 - 293. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Honda and P. J. Fitzgerald Frontiers in Intravascular Imaging Technologies Circulation, April 15, 2008; 117(15): 2024 - 2037. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |