J Am Coll Cardiol, 2006; 48:1365-1368, doi:10.1016/j.jacc.2006.07.017 (Published online 11 September 2006).
© 2006 by the American College of Cardiology Foundation
This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
j.jacc.2006.07.017v1
48/7/1365    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via ISI Web of Science (6)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Zegdi, R.
Right arrow Articles by Fabiani, J.-N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Zegdi, R.
Right arrow Articles by Fabiani, J.-N.

FOCUS ISSUE: CARDIAC INTERVENTION: PERCUTANEOUS VALVE PROSTHESIS

A Repositionable Valved Stent for Endovascular Treatment of Deteriorated Bioprostheses

Rachid Zegdi, MD, PhD*,{dagger},*, Ziad Khabbaz, MD{dagger}, Nicolas Borenstein, DVM{ddagger} and Jean-Noël Fabiani, MD*,{dagger}

* Université René Descartes, Paris V, France
{dagger} AP-HP, Service de Chirurgie Cardio-Vasculaire, Hôpital Européen Georges Pompidou, Paris, France
{ddagger} IMM Recherche, Institut Mutualiste Montsouris, Paris, France.

Manuscript received February 27, 2006; revised manuscript received July 6, 2006, accepted July 10, 2006.

* Reprint requests and correspondence: Dr. Rachid Zegdi, Hôpital Européen Georges Pompidou–Service de Chirurgie Cardiovasculaire, 20 Rue Leblanc, 75908 Paris, France. (Email: rzegdi{at}hotmail.com).


    Abstract
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
OBJECTIVES: We report our animal experience of endovascular valve replacement (VR) of failed bioprosthesis (BP) using an original delivery catheter allowing repositioning of the valved stent (VS).

BACKGROUND: Among the different devices designed for percutaneous VR, none has the potential for repositioning of a fully deployed VS.

METHODS: Five sheep underwent, on beating heart, tricuspid VR with a stented BP. Prolapse of 1 leaflet was induced by tearing. For the endovascular tricuspid VR, we used a VS constructed with a nitinol self-expandable stent and a porcine stentless aortic valve. We also used an original delivery catheter, allowing repositioning of the VS through a compression or relaxation mechanism of the stent.

RESULTS: Epicardial echocardiography and right ventriculography showed severe tricuspid regurgitation, with a regurgitant jet extending to the inferior vena cava. After surgical exposure to the infrarenal inferior vena cava, the VS was successfully implanted inside the failed BP in all cases. Repositioning of the fully deployed VS was always possible. Echocardiographic and macroscopic studies revealed adequate VS positioning, excellent leaflet opening, and absence of any intraprosthetic or periprosthetic leak.

CONCLUSIONS: Endovascular VR was easily performed in sheep with failed BP in the tricuspid position. The novel delivery catheter allowed adequate repositioning of our fully deployed VS before its definitive release. One may anticipate that the safety improvement conferred by this new technology will certainly favor the development of percutaneous VR in clinical practice.

Abbreviations and Acronyms
  BP = bioprosthesis
  SVD = structural valve deterioration
  VR = valve replacement
  VS = valved stent


Structural valve deterioration (SVD) is the most frequent long-term complication of implanted bioprosthesis (BP). When severe enough, reoperation may be necessary with an increased operative risk in comparison with the first operation (1).

Since the recent publications of percutaneous pulmonary or aortic valve replacement (VR), the enthusiasm of the medical community toward these potential alternatives to surgery has been dramatically increasing (2,3). However, encountered problems are numerous, and there is still a long way before a safe and effective procedure becomes available.

A major problem encountered with percutaneous VR is the impossibility of readjusting the position of a valved stent (VS) once fully deployed. Bad positioning of the VS could lead to a fatal per-procedural issue (3). None of the available VS have the potential to be repositioned after complete deployment (4). Thus, being able to reposition a fully deployed VS could prove of crucial importance and would certainly favor the development of percutaneous VR by improving the safety of the procedure.

The present animal study had 2 main objectives: first, to test the feasibility of the endovascular treatment of a failed BP with a VS, and second, to test a novel delivery device permitting the repositioning of a fully deployed VS.


    Methods
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
Valved stent and delivery device.   A 25-mm porcine aortic stentless valve (Toronto SPV, model SPA-101-25, St. Jude Medical, Minneapolis, Minnesota) was mounted onto a 30-mm nitinol self-expandable stent. After compression, the VS was introduced inside a 13-mm delivery catheter.

The originality of the delivery system consisted of the possibility of controlling precisely and reversibly the deployment of the VS after the delivery catheter has been retrieved. This was because of the presence of 2 sutures encircling the stent, whose traction through a proximal handle could reversibly affect the state (deployed or compressed) of the stent (Fig. 1). Therefore, if the location of the deployed VS was judged unsatisfactory, the operator had the possibility of repositioning the VS as often as necessary before final delivery. Once deployed in an adequate position, the VS was definitely liberated from the delivery catheter by pulling on a wire that controlled the attachment of the 2 sutures. The delivery catheter could then be withdrawn, leaving the VS in place.


Figure 1
View larger version (69K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 1 Representation of the principle of compression-relaxation of the valved stent. (Top, left) The stent has been compressed by exerting traction on the 2 encircling sutures. (Top, right) The stent has been reversibly deployed. (Bottom) The 2 sutures are proximally connected to a handle.

 
Experimental protocol.   Five sheep (>50 kg) were used for our experiments. All of the animals were treated according to the European regulations for animal experimentation (5). Through a right thoracotomy, tricuspid valve replacement was first performed on a beating heart with a Carpentier-Edwards Perimount number 27 aortic BP (model 2,800, Edwards Lifesciences, Irvine, California). The anterior leaflet of the BP was cut along its 2 commissural attachments to create prosthesis failure.

The infrarenal inferior vena cava was then accessed through a right retroperitoneal approach. A guidewire was inserted into the vein and passed under fluoroscopic guidance into the pulmonary artery, after crossing the tricuspid BP. The VS was positioned inside the BP. The external sheath was then pulled back and the VS deployed by releasing the traction on the 2 encircling sutures. When correctly positioned, the VS was definitively released from the delivery catheter.

Evaluation of the VS.   After insertion of the failed BP, tricuspid regurgitation was assessed semi-quantitatively with epicardial echocardiography and right ventriculography. After implantation of the VS and using standard echo-Doppler techniques, the transvalvular gradient was determined and any intraprosthetic or periprosthetic leak was recorded. Sheep were killed 1 to 2 h after the procedure, and hearts were then harvested. Positioning, stent deformation, and leaflet movement of the VS were carefully analyzed.


    Results
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
The experiment was completed in all 5 sheep. All animals were weaned from bypass without inotropic support. Duration (median time) of extracorporeal circulation was 48 mn (range 38 to 52 mn).

Intraprosthetic tricuspid insufficiency was severe in all sheep. Bidimensional echocardiography showed prolapse of the anterior leaflet of the BP into the right atrium. The vena contracta width was 8 mm (range 6 to 10 mm). The color flow regurgitant jet was large, eccentric, and extending into the inferior vena cava (Fig. 2). Pulsed Doppler study showed frank systolic blood flow reversal inside the inferior vena cava. Right ventriculography confirmed these findings by showing opacification of the inferior vena cava by the regurgitation jet (data not shown).


Figure 2
View larger version (47K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 2 Echocardiographic color Doppler study before and after endovascular valve replacement of a failed bioprosthesis. (Left) Tricuspid regurgitant jet extending into the inferior vena cava (IVC). (Right) Absence of any tricuspid regurgitation after valved stent delivery. RA = right atrium.

 
The metallic frame of the BP being radio-opaque, the target site of the VS implantation was therefore easily identified. Under fluoroscopic guidance, the VS was properly positioned inside the BP in all sheep. To evaluate the effectiveness of the delivery system, the VS was deliberately malpositioned inside the BP (Fig. 3). It was then very easy to correct its position using the compression-relaxation technique mentioned above (Fig. 3). Deployment of the VS in its correct and final position took no more than 2 min. During these maneuvers, a transient and mild decrease in systolic blood pressure (<20 mm Hg in all sheep) was observed.


Figure 3
View larger version (117K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 3 Representation of valved stent repositioning. (Top, left) The valved stent has been voluntarily misplaced too proximally inside the failed bioprosthesis. (Top, right) Recompression of the valved stent. (Bottom, left) The valved stent has been ultimately delivered in an adequate position. (Bottom, right) Right lateral view of the heart (the right atrium and the anterior part of the right ventricle have been resected) showing excellent positioning of the valved stented inside the failed bioprosthesis.

 
Unrestricted movement of the VS leaflets was easily observed with epicardial bi-dimensional echocardiography. There was neither residual tricuspid regurgitation nor periprosthetic leak in any case (Fig. 2). Median transvalvular gradient was 7 mm Hg (range 6 to 8 mm Hg).

At final analysis of the explanted hearts, all of the VS were in good position (Fig. 3). Leaflets of the VS were perfectly mobile without any evidence of macroscopic injury. Anchorage of the VS inside the BP appeared to be firm.


    Discussion
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
With the extension of the indication of bioprosthetic VR and the ageing of the population, cardiologists and surgeons will certainly face more and more cases of SVD. Besides being far less invasive, endovascular replacement of deteriorated BP should theoretically reduce the operative mortality by preventing complications of repeat heart dissection and the deleterious effects of cardiopulmonary bypass and myocardial ischemia caused by aortic cross-clamping. This should also imply a dramatic reduction of hospital costs.

Regurgitating forms of SVD of BP are easily accessible to an endovascular treatment. The actual self-expandable stents develop a satisfactory radial force permitting a firm anchorage of the VS inside a BP. However, in case of severely calcified BP, a balloon predilation of the BP might be necessary.

Several clinical or experimental observations have reported a number of fatal accidents caused by poor positioning of a VS (3,6). To the best of our knowledge, none of the VS reported in the literature offers the possibility to be readjusted after full deployment. The delivery device we tested permitted the reversibility of the VS positioning as many times as needed before the final release. This relied on a simple maneuver (compression-relaxation of the VS) that could be performed by 1 operator only.

The present delivery system offers other advantages. Mounting of the VS onto its delivery catheter is greatly facilitated by the traction on the encircling sutures, which would compress the valve. Thus, there is no need for any crimping device, as is the case for some other VS models. To achieve adequate positioning of the VS, right ventricular pacing or use of extracorporeal circulation during VS delivery has been recommended by others. With the present device, these potentially deleterious strategies are likely to be unnecessary, as was the case in our experimental study.

In conclusion, endovascular VR was easily performed in sheep with a failed tricuspid BP. The delivery catheter allowed adequate repositioning of our fully deployed VS before its definitive release. One may anticipate that the safety improvement conferred by this new technology will certainly favor the development of percutaneous VR in clinical practice.


    Footnotes
 
This work was financially supported by the Laboratoire Perouse company.


    References
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 

  1. Akins CW, Buckley MJ, Dagget WM, et al. Risk of reoperative valve replacement for failed mitral and aortic bioprostheses Ann Thorac Surg 1998;65:1545-1552.[Abstract/Free Full Text]
  2. Bonhoeffer P, Boudjemline Y, Qureshi SA, et al. Percutaneous insertion of the pulmonary valve J Am Coll Cardiol 2002;39:1664-1669.[Abstract/Free Full Text]
  3. Cribier A, Eltchaninoff H, Tron C, et al. Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosis J Am Coll Cardiol 2004;43:698-703.[Abstract/Free Full Text]
  4. Lutter G, Ardehali R, Cremer J, et al. Percutaneous valve replacement: current state and future prospects Ann Thorac Surg 2004;78:2199-2206.[Abstract/Free Full Text]
  5. Convention européenne sur la protection des animaux vertebras utilises a des fins expérimentales ou a d’autres fins scientifiques Journal Officiel des Communautés Européennes 1999;L222:29-37.
  6. Boudjemline Y, Bonhoeffer P. Steps toward percutaneous aortic valve replacement Circulation 2002;105:775-778.[Abstract/Free Full Text]




This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
j.jacc.2006.07.017v1
48/7/1365    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via ISI Web of Science (6)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Zegdi, R.
Right arrow Articles by Fabiani, J.-N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Zegdi, R.
Right arrow Articles by Fabiani, J.-N.

 
  cardiology careers collections past issues search home