J Am Coll Cardiol, 2006; 47:1840-1845, doi:10.1016/j.jacc.2005.11.078
(Published online 17 April 2006). © 2006 by the American College of Cardiology Foundation |



* Department of Cardiology, Bristol Heart Institute, Bristol Royal Infirmary, Bristol, United Kingdom
Department of Medical Physics and Bioengineering, Bristol Heart Institute, Bristol Royal Infirmary, Bristol, United Kingdom
Department of Clinical Radiology, Bristol Heart Institute, Bristol Royal Infirmary, Bristol, United Kingdom.
Manuscript received April 16, 2005; revised manuscript received August 14, 2005, accepted November 16, 2005.
* Reprint requests and correspondence: Prof. Karl R. Karsch, Bristol Heart Institute, University of Bristol, Level 7, Bristol Royal Infirmary, Upper Maudlin Street, Bristol BS2 8HW, United Kingdom. (Email: K.R.Karsch{at}bristol.ac.uk).
| Abstract |
|---|
|
|
|---|
BACKGROUND: Multislice computed tomography is now a viable modality for cardiac imaging. However, for any diagnostic use of ionizing radiation, the risk to the patient must be considered and justified.
METHODS: Multislice computed tomography angiography and conventional angiography were used to assess 180 patients with suspected coronary artery disease. Estimates of effective dose were derived from exposure data recorded for each patient examination. For each modality, a comparable calculation technique was used, based on Monte Carlo modeling of the standard Cristy phantom.
RESULTS: In a subset of 91 directly comparable patients the mean effective dose for MSCT coronary angiography was 14.7 mSv (SD 2.2) and that for conventional angiography was 5.6 mSv (SD 3.6). A significant difference in effective dose was seen between the two protocols.
CONCLUSIONS: The mean effective dose for MSCT coronary angiography was significantly higher than that for conventional angiography. As MSCT cardiac scanners become increasingly available, operators must be aware of the radiation dose and the factors that affect it.
| |||||||||
Radiation dose levels in diagnostic conventional coronary angiography (CCA) are well known (10,11). Less is known about radiation doses for cardiac MSCT in the clinical setting. Coronary imaging with four-slice MSCT may give larger radiation doses than either conventional angiography or electron-beam computed tomography (CT) (12). Electrocardiogram-controlled tube current modulation has the potential to reduce patient dose in cardiac MSCT protocols (13,14). However, the advent of 16-slice scanners, and more recently 64-slice scanners, has brought increased use of MSCT in the diagnosis of coronary artery disease with little data on the accompanying radiation doses.
We compare the effective doses for a large series of patients investigated for coronary artery disease using both 16-slice CT coronary angiography and CCA. Radiation dosimetry is complex and can only estimate dose based on information about the radiographic procedure and an anatomic model used to represent the patient. We use the same anatomic model for both conventional and MSCT coronary angiography (CA), allowing a direct comparison between these techniques.
| Methods |
|---|
|
|
|---|
The study was fully approved by the local hospital ethics committee. All patients recruited during the study period from April 2003 to August 2004 were included in the evaluation.
MSCT coronary angiography protocol. Informed consent was obtained from each patient. Patients heart rates were assessed 1 h before scanning, and if their heart rate was above 65 beats/min, 50 to 100 mg of oral metoprolol tartrate (Berk Pharmaceuticals, Eastbourne, United Kingdom) was administered.
Patients were scanned on a Sensation 16 (Siemens, Forchheim, Germany) with retrospective electrocardiogram-gated reconstruction. The scan sequence was a topogram, calcium-scoring scan, test bolus scan, and coronary angiogram.
The MSCT coronary angiography initially used 12 detectors (software Version 60A) (group 1), then 16 detectors once this became available (software Version 70A) (group 2). Scan parameters are listed in Table 1.
|
The angiogram covered from the pulmonary trunk to just below the base of the heart. Patients with known coronary artery bypass graft (CABG) were scanned from the sternoclavicular joint with their arms cradled above their head. All scans were performed in a cranial to caudal direction. The timing of the main injection of contrast medium (105 ml Iomeron 400 at 4 ml/s) was determined by the test bolus data.
MSCT radiation dose. Many different dose indicators are used to describe medical exposures. The computed tomography dose index (CTDI) represents the radiation dose within a single slice under controlled conditions.
The CTDI may be measured free in air or within a cylindrical acrylic phantom to represent dose within the body. The CTDIw is a weighted average of CTDI values within a phantom and may be multiplied by the scan length and corrected for spiral pitch to give the dose-length product (DLP). The CTDIw and DLP are broadly indicative of patient dose and are often displayed on the scanner console (15).
To maximize accuracy, we measured CTDI in air using a Radcal 2025 electrometer and 3 cm3 pencil ionization chamber (Radcal Corp., Monrovia, California) with a traceable calibration (Physikalisch-Technische Bundesanstalt [PTB], Berlin, Germany).
MSCT effective dose. The calculation of effective dose in MSCT is an approximation based on a number of factors. The CTDI is modified to take into account the best estimate of dose absorbed by the patient; achieved using a number of models or conversion factors. These are designed to reproduce the absorption characteristics of a "typical" patient as calculated using the established phantoms of recognized radiation dosimetry organizations. We used the basic conversion established by the National Radiological Protection Board (NRPB) in the United Kingdom from measured values of CTDI in air (16).
The NRPB used Monte Carlo modeling to estimate organ doses within an anthropomorphic geometrical phantom based on that of Cristy (17). The Imaging Performance Assessment of CT Scanners (ImPACT) group (London, United Kingdom) has matched a large range of contemporary CT scanners to the most appropriate NRPB data sets. Effective doses for MSCT were estimated using the ImPACT dosimetry calculator.
CCA technique. Experienced operators performed all conventional angiograms during routine sessions. Femoral access was used in the majority of cases. A standard sequence of projections was used, with variations according to need. Most patients (84%) also had a ventriculogram. Where a diagnostic procedure progressed to intervention, only the diagnostic part of the procedure was included in the radiation dose estimate.
Three Siemens cardiac catheterization units were used: an Axiom Artis FC single-plane, an Axiom Artis BC biplane, and an older HiCor single-plane unit. Both Axiom units used automatic selection of additional copper filtration (from 0 to 0.9 mm). The HiCor unit was used with a fixed 0.2-mm copper filter.
CCA radiation dose. The Axiom study report for each procedure gave the projection, field size, kVp, additional filtration, and dose-area product (DAP) for each acquisition, plus the DAP for each plane for fluoroscopy. Only DAP was available from the HiCor unit, so patients were preferentially assigned to the Axiom units, depending on availability and clinical urgency.
The DAP takes into account the intensity, duration, and area of the X-ray beam and is measured by an integral device each time an exposure is made. We checked the calibration of each DAP meter using a Radcal 9010 electrometer and 90X5-6 ionization chamber with a traceable calibration (PTB). The DAP was corrected for attenuation by the couch and mattress where appropriate.
CCA effective dose. A PC-based X-ray Monte Carlo program (18) was used to estimate effective doses from conventional angiography (Radiation and Nuclear Safety Agency, Helsinki, Finland). This software is more flexible than the NRPB conversion factors used in previous studies. Like that used to estimate MSCT doses, it is based on the Cristy model. Although it can be adjusted to reflect patients of other weights and heights, this was not done, because the CT dose estimate used a standard-sized patient.
The geometry for each projection was described by field size, angulation (right-left and cranial-caudal), and focus-skin distance. The study report indicated which field was used for each run. Actual field dimensions were measured using conventional X-ray film for each nominal field size, assuming no further collimation. We assumed that the focus-image intensifier distance was 105 cm and that the patient was 5 cm from the image intensifier face.
We assumed that the fluoroscopy geometry matched that of the acquisitions. For cases that proceeded to intervention, the contribution from fluoroscopy was included in proportion to the duration of the diagnostic part of the procedure.
Statistics. Differences between parameters for the two patient groups were tested by unpaired t tests at the 95% confidence level (two-tailed). We assumed equal variances.
| Results |
|---|
|
|
|---|
A further 62 patients had incomplete data for CCA, mainly where the HiCor unit was used owing to clinical urgency, and insufficient information was available to calculate effective dose accurately. Consequently 91 patients had effective dose measurements from both MSCT and conventional coronary angiography for direct comparison. The characteristics of this population are detailed in Table 2.
|
|
Effective dose from CCA. For the directly comparable patients the mean DAP for CCA was 27.6 Gy·cm2, and the mean effective dose was 5.6 mSv. There was no significant difference in DAP or fluoroscopy time between the directly comparable patients and those excluded due to incomplete data. Dose data are shown in Table 4. Figure 1 compares the effective doses for coronary angiograms using both techniques.
|
|
| Discussion |
|---|
|
|
|---|
Values such as CTDI and DAP do not allow comparison between different imaging modalities. Effective dose takes into account the amount of radiation absorbed in different organs and tissues, the relative biologic effectiveness of the radiation used, and the differing radiosensitivities of each organ and tissue. It can therefore be used to compare different imaging modalities. We directly compared the effective dose for coronary angiography using two different techniques, which had not been performed in a patient population before.
In comparing dose estimates, differences in scanner type and scan protocol are key factors to consider. Initial dose estimates for MSCT coronary angiography using four-slice scanners were lower than our data. Those estimates used phantom measurements (12) and a DLP method (19) on a small series of patients, unlike our larger patient cohort in a real clinical setting. We found a higher effective dose than studies (14,20) assessing 16-slice CT angiography, because our average clinical scan length was longer than the scan lengths used in those phantom studies.
When we moved from 12 to 16 detectors, the radiation dose increased in proportion to the manufacturers recommended effective mAs. This does not predict what will happen for greater numbers of detectors, because various factors influence radiation exposure, including the geometric efficiency of the beam and the resultant overscan. Bolus tracking may reduce radiation exposure, but as the test bolus scan contributed 5% to overall dose, this reduction is likely to be small.
Clearly our results refer to the protocols in place at the time of the study. Fixed exposure settings were used for all patients. Ideally, operators should optimize protocols to achieve adequate image quality for different-sized patients and should be aware of the impact this will have on patient dose.
Patients also had a calcium-scoring scan using 12 detectors, which covered from the carina to just below the base of the heart (mean scan length 182 mm). For the directly comparable patients, the mean effective dose was 4.1 mSv for group 1, but only 2.6 mSv for group 2 when electrocardiogram-controlled tube current modulation was used. The effective dose from the topogram is small, typically in the range of 0.1 to 0.2 mSv.
Our population, studied in a modern cardiac catheterization laboratory, gave similar DAP values to those in the literature for diagnostic angiography (21). Our estimates of effective doses were also similar to those from other studies. However, our use of the Monte Carlo program allowed us to better model the geometries and beam characteristics used for individual patients.
The International Commission on Radiological Protection has estimated the additional lifetime risk of fatal cancer as approximately 1 in 20,000 per mSv for the whole population (22). Although this risk is lower for the geriatric population generally investigated for coronary artery disease, significant reductions in risk will only be seen in patients in their 70s and 80s. Our subjects had an average age of 63 years.
A coronary CT angiogram with an effective dose of 14.7 mSv has a risk of inducing a fatal cancer of 1 in 1,400. Conventional coronary angiography (5.6 mSv) has a risk of 1 in 3,600, and a calcium-scoring scan (2.6 mSv) a risk of 1 in 7,700. To put this in context, a typical effective dose for a rest-stress myocardial single photon emission CT scan using technetium-99m is 8 mSv and that for a myocardial scan using thallium-201 is 18 mSv (23).
Coronary CT gives a higher risk of cancer than conventional angiography, although it is without the invasive mortality and morbidity risks. Radiation-induced skin injuries have been reported for interventional procedures carried out under fluoroscopic guidance (24,25); the risk of skin injury is greater when long or multiple procedures are carried out. Diagnostic procedures are unlikely to result in skin injury, and a comparison of skin doses from conventional angiography and MSCT is beyond the scope of this paper. Risks from contrast administration are inherent to both imaging techniques.
Study limitations. The anatomic model used to calculate effective dose will influence accuracy. Patient-specific dosimetry is not generally available for MSCT. Therefore, calculations do not allow for individual variation in weight, height, or gender but are based on a "standard" hermaphrodite adult with a weight of 71 kg and height of 174 cm. Our population was shorter and heavier than this model, reflecting the overweight population investigated for coronary artery disease. Fluoroscopy systems automatically adjust the amount of radiation delivered for body size, whereas current cardiac CT protocols do not offer this facility.
Conclusions. Coronary CT delivers a relatively high radiation dose for a purely diagnostic procedure. It also delivers a significantly higher effective dose than conventional angiography to patients being investigated for coronary artery disease. With the increasing availability of MSCT cardiac scanners, operators must be aware of the radiation dose, and the factors that affect it, for both clinical and research protocols.
As the clinical role of noninvasive CT coronary angiography develops, clinicians should remember that the use of ionizing radiation in medical exposures should be both justified and optimized. Further work is needed to determine optimal scan protocols. Future studies, using the next generation of CT scanners with 64 or more detectors, must provide information on their radiation exposure when evaluating the clinical use of cardiac CT.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
W T Roberts, J J Bax, and L C Davies Cardiac CT and CT coronary angiography: technology and application Heart, June 1, 2008; 94(6): 781 - 792. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Kramer All High-Risk Patients Should Not Be Screened With Computed Tomographic Angiography Circulation, March 11, 2008; 117(10): 1333 - 1339. [Full Text] [PDF] |
||||
![]() |
J. P. Earls, E. L. Berman, B. A. Urban, C. A. Curry, J. L. Lane, R. S. Jennings, C. C. McCulloch, J. Hsieh, and J. H. Londt Prospectively Gated Transverse Coronary CT Angiography versus Retrospectively Gated Helical Technique: Improved Image Quality and Reduced Radiation Dose Radiology, March 1, 2008; 246(3): 742 - 753. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Schroeder, S. Achenbach, F. Bengel, C. Burgstahler, F. Cademartiri, P. de Feyter, R. George, P. Kaufmann, A. F. Kopp, J. Knuuti, et al. Cardiac computed tomography: indications, applications, limitations, and training requirements: Report of a Writing Group deployed by the Working Group Nuclear Cardiology and Cardiac CT of the European Society of Cardiology and the European Council of Nuclear Cardiology Eur. Heart J., February 2, 2008; 29(4): 531 - 556. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Horiguchi, M. Kiguchi, C. Fujioka, Y. Shen, R. Arie, K. Sunasaka, and K. Ito Radiation Dose, Image Quality, Stenosis Measurement, and CT Densitometry Using ECG-Triggered Coronary 64-MDCT Angiography: A Phantom Study Am. J. Roentgenol., February 1, 2008; 190(2): 315 - 320. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Abdulla, S. Z. Abildstrom, O. Gotzsche, E. Christensen, L. Kober, and C. Torp-Pedersen 64-multislice detector computed tomography coronary angiography as potential alternative to conventional coronary angiography: a systematic review and meta-analysis Eur. Heart J., December 2, 2007; 28(24): 3042 - 3050. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. M. Hurwitz, R. E. Reiman, T. T. Yoshizumi, P. C. Goodman, G. Toncheva, G. Nguyen, and C. Lowry Radiation Dose from Contemporary Cardiothoracic Multidetector CT Protocols with an Anthropomorphic Female Phantom: Implications for Cancer Induction Radiology, December 1, 2007; 245(3): 742 - 750. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. B Meijboom, N. R Mollet, C. A Van Mieghem, A. C Weustink, F. Pugliese, N. van Pelt, F. Cademartiri, E. Vourvouri, P. de Jaegere, G. P Krestin, et al. 64-Slice CT coronary angiography in patients with non-ST elevation acute coronary syndrome Heart, November 1, 2007; 93(11): 1386 - 1392. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Einstein, K. W. Moser, R. C. Thompson, M. D. Cerqueira, and M. J. Henzlova Radiation Dose to Patients From Cardiac Diagnostic Imaging Circulation, September 11, 2007; 116(11): 1290 - 1305. [Full Text] [PDF] |
||||
![]() |
J. Ector, O. Dragusin, B. Adriaenssens, W. Huybrechts, R. Willems, H. Ector, and H. Heidbuchel Obesity Is a Major Determinant of Radiation Dose in Patients Undergoing Pulmonary Vein Isolation for Atrial Fibrillation J. Am. Coll. Cardiol., July 17, 2007; 50(3): 234 - 242. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. L. Hollingsworth, T. T. Yoshizumi, D. P. Frush, F. P. Chan, G. Toncheva, G. Nguyen, C. R. Lowry, and L. M. Hurwitz Pediatric Cardiac-Gated CT Angiography: Assessment of Radiation Dose Am. J. Roentgenol., July 1, 2007; 189(1): 12 - 18. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Inoue, T. Kobayashi, A. Morikawa, A. Taddio, C. D. Rose, G. F. Greil, W. J. Manning, J. W. Newburger, L. A. Sleeper, and J. C. Burns Treatment of Kawasaki Disease N. Engl. J. Med., June 28, 2007; 356(26): 2746 - 2748. [Full Text] [PDF] |
||||
![]() |
R. F. Redberg Computed Tomographic Angiography: More Than Just a Pretty Picture? J. Am. Coll. Cardiol., May 8, 2007; 49(18): 1827 - 1829. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. R. Karsch, D. R. Coles, M. A. Smail, I. S. Negus, P. Wilde, M. Oberhoff, and A. Baumbach Reply J. Am. Coll. Cardiol., February 20, 2007; 49(7): 829 - 829. [Full Text] [PDF] |
||||
![]() |
C. R. Conti Comparing Radiation Doses From 2 Diagnostic Tests J. Am. Coll. Cardiol., February 20, 2007; 49(7): 828 - 828. [Full Text] [PDF] |
||||
![]() |
S. Achenbach Computed Tomography Coronary Angiography J. Am. Coll. Cardiol., November 21, 2006; 48(10): 1919 - 1928. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. B. Meijboom, N. R. Mollet, C. A.G. Van Mieghem, J. Kluin, A. C. Weustink, F. Pugliese, E. Vourvouri, F. Cademartiri, A. J.J.C. Bogers, G. P. Krestin, et al. Pre-Operative Computed Tomography Coronary Angiography to Detect Significant Coronary Artery Disease in Patients Referred for Cardiac Valve Surgery J. Am. Coll. Cardiol., October 17, 2006; 48(8): 1658 - 1665. [Abstract] [Full Text] [PDF] |
||||
![]() |