J Am Coll Cardiol, 2006; 47:1214-1223, doi:10.1016/j.jacc.2006.01.049
(Published online 8 February 2006). © 2006 by the American College of Cardiology Foundation |


* Department of Cardiology, Charles Nicolle Hospital, Rouen, France
Department of Cardiac Surgery, Charles Nicolle Hospital, Rouen, France
Manuscript received December 20, 2005; revised manuscript received January 19, 2006, accepted January 24, 2006.
* Reprint requests and correspondence: Dr. Alain Cribier, Service de Cardiologie, Hopital Charles Nicolle, 1 rue de Germont, 76 031 Rouen Cedex, France (Email: Alain.Cribier{at}chu-rouen.fr).
| Abstract |
|---|
|
|
|---|
BACKGROUND: We developed a PHV (equine pericardium valve in a balloon-expandable, stainless-steel stent) to treat patients with inoperable aortic stenosis (AS).
METHODS: Thirty-six patients (aortic valve area
0.7 cm2, New York Heart Association [NYHA] functional class IV, and severe comorbidities), formally declined for surgery, were recruited on a compassionate basis. The PHV was implanted by retrograde or antegrade trans-septal approach. Clinical and echocardiographic outcomes were assessed serially.
RESULTS: Twenty-seven patients were implanted successfully (23 antegrade, 4 retrograde) in the subcoronary position with improvement in valve area (0.60 ± 0.11 cm2 to 1.70 ± 0.10 cm2, p < 0.0001) and transvalvular gradient (37 ± 13 mm Hg to 9 ± 2 mm Hg, p < 0.0001). Paravalvular aortic regurgitation was grade 0 to 1 (n = 10), grade 2 (n = 12), and grade 3 (n = 5). One week post-procedure, improvement in left ventricular function (45 ± 18% to 53 ± 14%, p = 0.02) was most pronounced in patients with ejection fraction <50% (35 ± 10% to 50 ± 16%, p < 0.0001). Thirty-day major adverse events after successful implantation were 26% (pericardial tamponade, stroke, arrhythmia, urosepsis, and one death unexplained at autopsy). Eleven patients are currently alive with follow-up of 9 months (n = 2), 10 months (n = 3), 11 months (n = 1), 12 months (n = 2), 23 months (n = 1), and 26 months (n = 2). All patients experienced amelioration of symptoms (>90% NYHA functional class I to II). Percutaneous heart valve function remained unchanged during follow-up, and no deaths were device-related.
CONCLUSIONS: Percutaneous heart valve implantation is feasible in inoperable patients with end-stage AS leading to hemodynamic and clinical improvement. Continued advances and improved patient selection should decrease adverse events in the near future.
| |||||||||||
The initial attempts to treat non-surgical patients with advanced aortic stenosis began with balloon aortic valvuloplasty (BAV) in 1985 (3). This technique, initially met with enthusiasm, was largely abandoned by clinicians as the benefits of valvuloplasty rarely lasted more than one year (4). The problem of valve restenosis after balloon dilation was finally addressed in 1999 with the development of a bioprosthetic heart valve, which was sutured onto a balloon expandable stent (Percutaneous Valve Technologies, Edwards Lifesciences, Irvine, California) (Fig. 1). Though previous attempts at percutaneous valve replacement in the aortic position had been limited by the applicability to humans (5,6), this percutaneous heart valve (PHV) was successfully implanted on April 16, 2002, in a patient with inoperable aortic stenosis and life-threatening comorbidities (7).
|
| Methods |
|---|
|
|
|---|
0.7 cm2) with associated symptoms (by New York Heart Association [NYHA] functional class IV dyspnea) that were expected to benefit from isolated valve replacement. Patients had to be refused for standard aortic valve replacement by two independent cardiac surgeons on the basis of their high risk for surgery (Parsonnets score
30) (9). Exclusion criteria included the following: vascular disease that precluded access, severe deformation of the chest, intracardiac thrombus, unprotected stenosis of the left main coronary artery not amenable to percutaneous intervention, myocardial infarction (MI) within seven days, prosthetic heart valves, active infection, leukopenia (<3,000 white blood cells/mm3), coagulopathy, active bleeding, or acute anemia (hemoglobin <9 mg/dl). Patients that could not be fully dilated with a 23-mm aortic valvuloplasty balloon (notable waist) and patients with a native aortic valve annulus size >24 mm or <19 mm were also excluded. The study was approved and performed in accordance with the regulations of our institutional ethics committee (CCPPRB, Comité Consultatif de Protection des Personnes dans la Recherche Biomédicale) and the French government (AFSSAPS, Agence Francaise de Sécurité Sanitaire des Produits de Santé).
Definitions and baseline measurements.
The primary objectives of this study were to evaluate the feasibility and safety of PHV implantation. Feasibility was evaluated by procedural success defined as accurate placement of the PHV in the subcoronary position with associated improvement of hemodynamic parameters (
30% reduction in mean transvalvular aortic gradient) and absence of severe aortic regurgitation (grade 4) (10). Safety was evaluated by the occurrence of major adverse cardiac and cerebrovascular events (MACCE) or other valve-related adverse events. Major adverse cardiac and cerebrovascular events were defined as death, MI, emergent cardiac surgery, and cerebrovascular accident. Any MACCE that occurred either during or within 30 days of PHV implantation was considered a procedure-related complication. Any adverse event that was secondary to PHV malfunction after implantation was considered a device-related complication. Safety was assessed at 1, 3, 6, and 12 months; after 1 year, patients were followed every 6 months.
The secondary objectives of the study were to obtain data regarding the efficacy and durability of the PHV. Efficacy and durability were assessed by NYHA functional classification and hemodynamic improvement (measured by echocardiography). Hemodynamic variables, such as native and prosthetic aortic valve gradient, area, and regurgitation (valvular or paravalvular, graded from 0 to 4) (10), were evaluated by standard methods of echocardiographic measurement. Clinical evaluation and echocardiographic examination were performed immediately post-procedure, on day 1, day 7, 1 month, 3 months, 6 months, and 12 months; afterwards, patients were followed every 6 months.
Valve anatomy and function were determined in all patients by transesophageal echocardiography and transthoracic echocardiography (TTE) pre-procedure; coronary anatomy was also evaluated pre-procedure. In addition to echocardiographic measurements, right-sided heart pressures, cardiac output, oxygen saturations, and aortic transvalvular pressure gradient were measured on the day of implantation by invasive methods. Valve calcification was assessed qualitatively by fluoroscopic evaluation.
Procedure. The materials and technique for implantation have been previously described in detail (7,8,11). Briefly, each procedure was performed under mild sedation and local anesthesia. All patients received aspirin (160 mg) and clopidogrel (300 mg) 24 h before valve placement; antibiotics for procedural prophylaxis (first generation cephalosporin) were given 1 h before. After measurement of baseline hemodynamics, supra-aortic angiography and placement of a right ventricular pacing lead were performed. Heparin 5,000 IU was given intravenously before retrograde catheterization of the aortic valve. Retrograde pre-dilation of the aortic valve was done in all patients (except two who were dilated antegrade) with a 23-mm Z-MED balloon (NuMED Inc., Hopkinton, New York) during rapid ventricular pacing (200 to 220 stimulations/min) (Fig. 2). Delivery of the valve was performed by an antegrade trans-septal or retrograde approach.
|
|
Regardless of the approach used, the final steps of PHV implantation were similar for both methods. The PHV was mounted onto a 22-mm Z-MED II balloon (NuMed Canada Inc., Cornwall, Ontario, Canada) using a specially designed crimper. The supra-aortic angiogram and native valve calcifications were used as anatomical landmarks for valve placement in the anteroposterior projection (mid-line of the stent frame was placed at the level of the calcifications). All valves were deployed (Figs. 3C and 3D) during rapid pacing except two patients that were implanted during cardiac arrest. Hemodynamic improvement was measured immediately afterwards, and a supra-aortic angiogram was performed in patients without renal insufficiency to verify placement as well as the presence of aortic regurgitation (Fig. 3E). A cranial view of the stent-valve was used to evaluate uniform expansion of the PHV (Fig. 3F). Arterial access was managed using closure devices and/or surgical repair before device use or in cases of device failure. Venous access was managed by manual compression. Antibiotics were given up to 48 h after the procedure. Subcutaneous enoxaparin (40 mg/day) was administered until the day of discharge. Clopidogrel (75 mg/day) was continued for one month, and aspirin (160 mg/day) was continued indefinitely.
Statistical analysis. Comparison of clinical and echocardiographic variables with baseline values was performed using the non-parametric Wilcoxon rank sum test. Comparison of procedural times was done by the Mann-Whitney U test. Differences were considered statistically significant for p < 0.05. All values were expressed as mean ± SD.
| Results |
|---|
|
|
|---|
3 per patient). Other significant comorbid conditions not detailed in the table included severe kyphoscoliosis, severe osteoporosis, hepatic cirrhosis, systemic scleroderma, angiodysplasia of the colon with chronic anemia (hematocrit <30%), myelodysplastic syndrome with chronic anemia, chronic steroid use, antiphospholipid syndrome with previous thrombotic history, peripheral neuropathy, severe obesity, severe right heart failure, and Parkinsons disease. One patient was a Jehovahs Witness. All native aortic valves were tricuspid and calcified (12 heavy, 17 moderate, 7 mild).
|
|
Echocardiographic data. A statistically significant increase in aortic valve area (AVA) was seen in the group of patients with successful implantation. Twenty-four hours after the procedure, the AVA, measured by TTE, increased from 0.60 ± 0.09 cm2 to 1.70 ± 0.11 cm2 (p < 0.0001, n = 25 surviving patients). At 1, 3, 6, 12, and 24 months, the AVA remained stable at 1.68 ± 0.11 cm2 (n = 16), 1.66 ± 0.09 cm2 (n = 12), 1.63 ± 0.07 cm2 (n = 7), 1.75 ± 0.18 cm2 (n = 3), and 1.64 ± 0.04 cm2 (n = 2), respectively (p = NS) (Fig. 5A). Similarly, the mean aortic gradient decreased from 37 ± 13 mm Hg to 9 ± 2 mm Hg post-procedure (p < 0.0001, n = 25), and remained 10 ± 2 mm Hg, 11 ± 2 mm Hg, 11 ± 2 mm Hg, 10 ± 1 mm Hg, and 12 ± 1 mm Hg (1, 3, 6, 12, and 24 months, respectively) during follow-up (p = NS) (Fig. 5B). Peak velocity across the aortic valve decreased from 4.01 ± 0.60 m/s to 2.25 ± 0.40 m/s post-implantation (p < 0.0001), and was 2.36 ± 0.25 m/s, 2.45 ± 0.25 m/s, 2.42 ± 0.31 m/s, 2.33 ± 0.11 m/s, and 2.36 ± 0.17 m/s upon follow-up (p = NS).
|
|
Clinical follow-up. Six of the 27 patients with PHV implantation had a complication during the procedure. Two patients died as a consequence of cardiac tamponade. One of these patients, with severe dextrorotation of the heart, had a complicated trans-septal puncture. The other patient, on chronic steroid therapy for pulmonary fibrosis, had a slow bleeding perforation of the right ventricle from the pacing lead; infection and subsequent sepsis developed after surgical repair. Another patient on chronic steroids for treatment of rheumatologic disease developed urosepsis three days after the procedure and died two days later. Complete heart block with temporary loss of pacing lead contact occurred in one case; despite successful PHV implantation, prolonged resuscitation lead to irreversible brain damage. One patient developed a stroke during retrograde catheterization of the aortic valve. She expired at 33 days secondary to multi-organ failure. Intractable hypotension in the procedure room after removal of the 24-F sheath from the femoral artery was the cause of another death. No etiology was found at autopsy; the PHV appeared normal and appropriately positioned (Table 2).
|
To date, 11 patients are alive (3 from the I-REVIVE trial, 8 from the RECAST trial) and have returned to a normal life, limited only by their previous medical conditions (NYHA functional class I in 4 patients, class II in 6 patients, class III in 1 patient). Their follow-up is 9 months (n = 2), 10 months (n = 3), 11 months (n = 1), 12 months (n = 2), 23 months (n = 1), and 26 months (n = 2). The valve function by the most recent echocardiographic exam (valve area 1.69 ± 0.10 cm2, mean transvalvular gradient 11 ± 3 mm Hg, measured at 3 to 24 months post-procedure) is unchanged compared to the day of implantation (valve area 1.72 ± 0.13 cm2, mean transvalvular gradient 9 ± 3 mm Hg, p = NS). The amount of paravalvular leak in these patients is grade 0 to 1 (n = 7) and grade 2 (n = 4). Ejection fraction in this group is 53 ± 12% (baseline 51 ± 15%, p = NS). One patient has not returned for repeat TTE, but remains clinically improved.
Seven patients who underwent BAV without a successful valve implantation or attempted valve implantation, survived several months after their procedure. One patient survived 19 months but suffered from recurrent heart failure, multiple strokes, and, eventually, a neurovegetative state. One patient died 10 months after dilation, spending most of her remaining life hospitalized for heart failure. Another patient died of heart failure (three months), and one succumbed to pancreatic cancer (two months). Three patients are alive at eight, six, and four months post-procedure. One (month 8) had return of NYHA functional class IV dyspnea.
| Discussion |
|---|
|
|
|---|
Feasibility and technique. Twenty-seven patients (82% of the 33 patients attempted) with debilitating and life-threatening aortic stenosis were successfully implanted with a PHV in the subcoronary position without coronary occlusion or interference of the mitral valve apparatus. Both the antegrade and retrograde approach were used in the I-REVIVE part of the trial; only the antegrade approach was used in the RECAST part of the trial (per protocol amendment). With the antegrade approach, the success rate was very high (85%), and the problems associated with introducing a 24-F sheath into the femoral artery were avoided (two cases of closure device failure of four attempted). It was easier to cross the native valve with the PHV in the antegrade rather than the retrograde direction, especially in patients with excessive calcification. This approach, however, was technically more demanding and required more case experience. Hemodynamic collapse used to occur during implantation until we learned that the stiff guidewire could exert traction on the anterior leaflet of the mitral valve. Straightening of this guidewire loop in the left ventricle caused massive mitral regurgitation and hemodynamic collapse that recovered only when the loop shape was resumed. The problems of the guidewire loop were markedly improved by pushing the Sones catheter retrogradely into the left ventricle, maintaining the guidewire shape during passage of the PHV in the left heart. Guidewire-induced hemodynamic collapse, consequently, did not occur during PHV implantation in the RECAST trial. In the I-REVIVE trial, however, five patients experienced this problem during implantation.
The success of retrograde implantation (57%, four of seven patients) was lower than that of the antegrade approach. The procedure, however, was quicker and less technically demanding without the problem of guidewire loops. Many patients were not candidates for this method because of diseased or tortuous femoral arteries that could not accommodate such a large sheath size. In some patients with heavily calcified valves, we were unable to retrogradely cross the native valve with the PHV because of wire bias into a commissure (the three failures). In this situation, we continued the procedure by the antegrade route in one patient. When the valve could not be implanted in the heart, the PHV was surgically removed from the femoral artery (one patient) or safely deployed above the bifurcation of the iliac arteries to avoid the complications of vascular surgery, mechanical ventilation, and general anesthesia (two patients). In one of the patients with a valve in the descending aorta, angiographic and clinical follow-up (10 months) confirmed the valve in the open position without impedance to flow. In the other patient, autopsy evaluation revealed the valve also in the open position without evidence of migration (12 h). Likewise, in two patients with a failed procedure secondary to migration, implantation of the PHV in the thoracic aorta was without sequelae (up to six months follow-up). In our previous animal experience (A. Cribier and H. Eltchaninoff, unpublished data, 2001), the PHV implanted in the aorta remains in the open position and is of no clinical consequence.
In both the antegrade and the retrograde approach, rapid ventricular pacing was fundamental for the success of the procedure. At a rate of 200 to 220 beats/min, rapid pacing causes an immediate and predictable decrease in stroke volume, preventing the ejection of the PHV-mounted balloon catheter by the left ventricle. Rapidly pacing the heart for <10 s during PHV deployment or BAV is well tolerated and has not provoked malignant ventricular arrhythmias in our series. The pacemaker lead is also helpful to prevent episodes of significant bradycardia that can occur during the procedure.
Safety. Thirty-day MACCE was 26% in those with a successful implantation. Though none of these patients expired because of prosthetic valve dysfunction, these deaths were related to complications from the procedure. Obviously, this group was also at an extremely high risk for surgical valve replacement. A Parsonnets surgical risk score of 47 and a EuroSCORE of 12 has been respectively associated with a >25% and >20% mortality within 30 days of cardiac surgery (9,13). The high mortality reflects the nature of intervening on this precarious group. The rate of stroke in our series (including patients who underwent PHV implantation or just BAV with a failed implantation) was 2.8% (1 of 35 patients), which is similar to the incidence reported with surgical valve replacement (14).
The six-month MACCE was also very high (37% excluding MACCE at one month). Even after successful implantation with improvement in functional status, many of our patients (n = 10) remained fragile, unable to tolerate non-cardiac surgery or succumbing to conditions such as stroke, pneumonia, progressive renal failure, cancer, and pulmonary embolism. None of the mortality that occurred after 30 days was related to the PHV or to the procedure.
Efficacy and durability. The AVA increased to 1.7 cm2 consistently after PHV implantation and did not significantly change even up to two years. Similarly, the mean transvalvular gradient decreased to 10 mm Hg post-procedure and remained at this value throughout the follow-up. Left ventricular function improved as early as one week with the most significant increase seen in patients with depressed myocardial function (15). This phenomenon is well described in the surgical literature after appropriate selection of patients with depressed ejection fraction for valve replacement (16). And though our patients were not pre-selected by dobutamine echocardiography, recovery of left ventricular dysfunction was seen in several cases, consistent with afterload mismatch pathophysiology.
Improvements in echocardiographic variables correlated with clinical improvement as well. All patients had symptomatic relief after the procedure, with the majority improving from NYHA functional class IV to class II. Obviously, the largest amelioration of symptoms was seen in patients without comorbidities that limited functional status such as severe lung disease. In patients without PHV implantation, improvement in symptoms after only BAV were short lived. Patients either died of heart failure or had the return of class IV dyspnea by six months, which is consistent with previously published data on BAV (4).
Future directions. Future improvements in technique and hardware will be directed at two different issues: improved deliverability of the PHV via the retrograde approach and resolution of adverse events that are related to PHV size. Currently, a new, lower-profile system with a steerable delivery catheter has been developed to revive the retrograde approach. The results of this new technology and the concomitant implantation of a larger PHV (26 mm) in a series of less severely ill patients are promising and should soon be reported. In our series, we have been implanting patients that have different size aortic annuli with one size PHV (23 mm). Keeping this limitation in mind, grade 3 paravalvular leak occurred in five patients (four in I-REVIVE, one in RECAST). The regurgitation was well tolerated, and all patients maintained an improved functional class post-procedure (one patient has 10 months follow-up and paravalvular leak is now grade 2). In these cases, however, a larger size prosthesis could have been better apposed to the native leaflets, resulting in less regurgitation. Similarly, a larger PHV may have been important in decreasing the risk of migration. Although this problem is less common than paravalvular leak, patients with a larger annulus and mild aortic valve calcification seem to be at increased risk. Currently, we believe that patients with an aortic annulus >21 mm would benefit from a larger size PHV.
With increasing experience, complications related to the procedure should decrease, and innovations in hardware should make PHV implantation technically less demanding, with less risk to the patient. Soon, multicenter trials will begin in Europe to study the feasibility, safety, and efficacy of PHV implantation in patients at high surgical risk with less severe comorbidity. In this cohort, we hope to study the long-term performance of the PHV and the continuing clinical benefits of this technique. With continued advances in this field, percutaneous aortic valve implantation should become a very realistic way of treating a select population of patients with degenerative aortic stenosis in the near future.
| Appendix |
|---|
|
|
|---|
Echocardiography Core Laboratory: Michael Picard, MD, Massachusetts General Hospital, Boston, MA.
Pathology Core Laboratory: Renu Vermani, MD, Armed Forces Institute of Pathology, Department of Cardiovascular Pathology, Washington, DC.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. Ramaraj and V. L Sorrell Degenerative aortic stenosis BMJ, March 8, 2008; 336(7643): 550 - 555. [Full Text] [PDF] |
||||
![]() |
J. F. Rhodes, Z. M. Hijazi, and R. J. Sommer Pathophysiology of Congenital Heart Disease in the Adult, Part II: Simple Obstructive Lesions Circulation, March 4, 2008; 117(9): 1228 - 1237. [Full Text] [PDF] |
||||
![]() |
B. Murtuza, J. R. Pepper, R. DeL Stanbridge, C. Jones, C. Rao, A. Darzi, and T. Athanasiou Minimal Access Aortic Valve Replacement: Is It Worth It? Ann. Thorac. Surg., March 1, 2008; 85(3): 1121 - 1131. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Zegdi, G. Sleilaty, A. Lafont, and J.-N. Fabiani Percutaneous Aortic Valve Replacement With the CoreValve Prosthesis J. Am. Coll. Cardiol., January 15, 2008; 51(2): 170 - 170. [Full Text] [PDF] |
||||
![]() |
E. A. Grossi, C. F. Schwartz, P.-J. Yu, U. P. Jorde, G. A. Crooke, J. B. Grau, G. H. Ribakove, F. G. Baumann, P. Ursumanno, A. T. Culliford, et al. High-Risk Aortic Valve Replacement: Are the Outcomes as Bad as Predicted? Ann. Thorac. Surg., January 1, 2008; 85(1): 102 - 107. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. F. Padera Jr. and F. J. Schoen Pathology of Cardiac Surgery Card. Surg. Adult, January 1, 2008; 3(2008): 111 - 178. [Full Text] |
||||
![]() |
N. D. Desai and G. T. Christakis Bioprosthetic Aortic Valve Replacement: Stented Pericardial and Porcine Valves Card. Surg. Adult, January 1, 2008; 3(2008): 857 - 894. [Full Text] |
||||
![]() |
M. J. Davidson and D. S. Baim Percutaneous Aortic Valve Interventions Card. Surg. Adult, January 1, 2008; 3(2008): 963 - 971. [Full Text] |
||||
![]() |
T. Feldman and M. B. Leon Prospects for Percutaneous Valve Therapies Circulation, December 11, 2007; 116(24): 2866 - 2877. [Full Text] [PDF] |
||||
![]() |
S. R. Dixon, C. L. Grines, and W. W. O'Neill The Year in Interventional Cardiology J. Am. Coll. Cardiol., July 17, 2007; 50(3): 270 - 285. [Full Text] [PDF] |
||||
![]() |
E. Grube, G. Schuler, L. Buellesfeld, U. Gerckens, A. Linke, P. Wenaweser, B. Sauren, F.-W. Mohr, T. Walther, B. Zickmann, et al. Percutaneous Aortic Valve Replacement for Severe Aortic Stenosis in High-Risk Patients Using the Second- and Current Third-Generation Self-Expanding CoreValve Prosthesis: Device Success and 30-Day Clinical Outcome J. Am. Coll. Cardiol., July 3, 2007; 50(1): 69 - 76. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Tabata, Z. Khalpey, S. F. Aranki, G. S. Couper, L. H. Cohn, and P. S. Shekar Minimal Access Surgery of Ascending and Proximal Arch of the Aorta: A 9-Year Experience Ann. Thorac. Surg., July 1, 2007; 84(1): 67 - 72. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Bergler-Klein, G. Mundigler, P. Pibarot, I. G. Burwash, J. G. Dumesnil, C. Blais, C. Fuchs, D. Mohty, R. S. Beanlands, Z. Hachicha, et al. B-Type Natriuretic Peptide in Low-Flow, Low-Gradient Aortic Stenosis: Relationship to Hemodynamics and Clinical Outcome: Results From the Multicenter Truly or Pseudo-Severe Aortic Stenosis (TOPAS) Study Circulation, June 5, 2007; 115(22): 2848 - 2855. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Coats and P. Bonhoeffer NEW PERCUTANEOUS TREATMENTS FOR VALVE DISEASE Heart, May 1, 2007; 93(5): 639 - 644. [Full Text] [PDF] |
||||
![]() |
A. N. DeMaria, O. Ben-Yehuda, G. K. Feld, G. S. Ginsburg, B. H. Greenberg, W. Y.W. Lew, J. A.C. Lima, A. S. Maisel, J. Narula, D. J. Sahn, et al. Highlights of the Year in JACC 2006 J. Am. Coll. Cardiol., January 30, 2007; 49(4): 509 - 527. [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, A. Vahanian, H. Baumgartner, J. Bax, E. Butchart, R. Dion, G. Filippatos, F. Flachskampf, R. Hall, B. Iung, et al. Guidelines on the management of valvular heart disease: The Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology Eur. Heart J., January 26, 2007; (2007) ehl428v1. [Full Text] [PDF] |
||||
![]() |
S. H. Rahimtoola The Year in Valvular Heart Disease J. Am. Coll. Cardiol., January 23, 2007; 49(3): 361 - 374. [Full Text] [PDF] |
||||
![]() |
J. Ye, A. Cheung, S. V. Lichtenstein, S. Pasupati, R. G. Carere, C. R. Thompson, A. Sinhal, and J. G. Webb Six-month outcome of transapical transcatheter aortic valve implantation in the initial seven patients Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 16 - 21. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Grube, J. C. Laborde, U. Gerckens, T. Felderhoff, B. Sauren, L. Buellesfeld, R. Mueller, M. Menichelli, T. Schmidt, B. Zickmann, et al. Percutaneous Implantation of the CoreValve Self-Expanding Valve Prosthesis in High-Risk Patients With Aortic Valve Disease: The Siegburg First-in-Man Study Circulation, October 10, 2006; 114(15): 1616 - 1624. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. V. Lichtenstein, A. Cheung, J. Ye, C. R. Thompson, R. G. Carere, S. Pasupati, and J. G. Webb Transapical Transcatheter Aortic Valve Implantation in Humans: Initial Clinical Experience Circulation, August 8, 2006; 114(6): 591 - 596. [Abstract] [Full Text] [PDF] |
||||
![]() |
Percutaneous Aortic Valve Implantation: Mid-Term Follow-Up Journal Watch Cardiology, May 11, 2006; 2006(511): 2 - 2. [Full Text] |
||||
| ||||||||||||||||||||||||