J Am Coll Cardiol, 2006; 47:933-938, doi:10.1016/j.jacc.2005.10.054
(Published online 8 February 2006). © 2006 by the American College of Cardiology Foundation |

* Heart Center of Chonnam National University Hospital, Chonnam National University Research Institute of Medical Sciences, Gwangju, Korea.
Chemical Engineering of Chonnam National University, Gwangju, Korea.
Manuscript received August 31, 2005; revised manuscript received October 10, 2005, accepted October 18, 2005.
* Reprint requests and correspondence: Dr. Myung Ho Jeong, Cardiac Catheterization Laboratory, Heart Center of Chonnam National University Hospital, 8 Hak-dong, Dong-gu, Gwangju, 501-757, Korea. (Email: myungho{at}chollian.net).
| Abstract |
|---|
|
|
|---|
BACKGROUND: Recently we have demonstrated that abciximab-coated stents have inhibitory effects in the prevention of coronary restenosis.
METHODS: Ninety-six patients with AMI were randomly allocated into two groups; group I received abciximab-coated stents (n = 48, 57.1 ± 12.0 years), and group II received bare metal control stents (n = 48, 58.4 ± 11.6 years).
RESULTS: At baseline, clinical characteristics, percent diameter stenosis, and minimal luminal diameter were no different between the two groups. One patient in group II had reinfarction and target lesion reintervention during hospital stay. Follow-up coronary angiography was obtained in 77.1% (37 of 48) in group I and 75.0% (36 of 48) in group II. Percent diameter stenosis and late loss were significantly lower in group I than group II (18.9 ± 5.54% vs. 37.9 ± 6.25%, p = 0.008; and 0.39 ± 0.29 mm vs. 0.88 ± 0.45 mm; p = 0.008, respectively). At follow-up intravascular ultrasound, intrastent lumen area and intrastent neointimal hyperplasia (NIH) area were 5.4 ± 1.8 mm2 and 2.2 ± 1.5 mm2, respectively, in group I and 4.3 ± 1.6 mm2 and 3.4 ± 1.8 mm2, respectively, in group II (p = 0.045). And, in-stent restenosis rate was lower in group I than group II (p = 0.011 and p = 0.008, respectively). During 1-year follow-up, two patients in group II (4.1%) had AMI, whereas no patient in group I suffered AMI. Target lesion revascularization and total major adverse cardiac events rates were relatively lower in group I compared with those in group II (10.4% [5 of 48] vs. 20.8% [10 of 48], p = 0.261, and 10.4% vs. 25.0%, p = 0.107, respectively).
CONCLUSIONS: Abciximab-coated stent implantation was safe and effective without stent thrombosis in AMI patients.
| ||||||||
Abciximab, a potent antiplatelet agent that blocks the final pathways to platelet aggregation, improves outcomes of high-risk PCI and decreases the incidence of major adverse cardiac events (MACE) (811). Differently from other kinds of platelet glycoprotein IIb/IIIa receptor blockers, it also binds to Mac-1 (CD11b/18) on vascular endothelial cells and macrophages, thereby inhibiting inflammatory responses and smooth muscle cell proliferation after vascular injury (1217). Recently, we demonstrated that abciximab-coated stents were safe and effective in the prevention of coronary restenosis in humans as well as in a porcine model (18,19).
The aim of the present study is to evaluate clinical and angiographic outcomes of abciximab-coated stent implantation in patients with acute myocardial infarction (AMI).
| Methods |
|---|
|
|
|---|
Manufacturing process of the abciximab-coated stent. Abciximab-coated stents were used according to the protocol previously described (18,19). Briefly, a plasma polymerization reaction was performed to attach amine radicals to the stent surface. For the attachment of amine radicals to the stent surface, diaminocyclohexane monomer was drifted to the tubular reactor in a constant dose, and plasma was generated with a radiofrequency power generator. The abciximab used was a human-murine chimeric antibody Fab fragment, ReoPro (Eli Lilly and Company, Indianapolis, Indiana). The carboxy radical of abciximab was introduced to the amine radicals attached to the stent to achieve covalent bonding and improved attachment power between the stent and abciximab. The abciximab coating on the surface of the stent was confirmed by scanning electron microscopy. For the release, kinetics of abciximab from the stent was done.
Study procedure, stent implantation, angiography, and IVUS. All other procedures were performed by standard techniques: randomly selected stents were deployed at 10 to 16 atmospheric pressure after predilation with a balloon catheter. In cases with residual stenosis, additional balloon dilatation was performed after stenting. None of the study patients received glycoprotein IIb/IIIa receptor blockers. All patients received aspirin (300 mg loading and 100 to 200 mg/day indefinitely) and clopidogrel (300 mg loading and 75 mg/day for 2 months). Heparin was administered as 5,000-U bolus, followed by 1,000 U/h and additional 5,000 U immediately before PCI to keep the activated clotting time (ACT) at 250 to 300 s. Successful PCI was defined as a patent vessel at the treatment site with anterograde Thrombolysis In Myocardial Infarction (TIMI) flow grade 3 and angiographic residual stenosis <20% without occurrence of any cardiac events.
Acute myocardial infarction was defined as the presence of typical chest pain, ischemic change on the electrocardiogram in two or more contiguous leads, and peak elevation of plasma creatine kinase (CK) and CK-MB to at least twice normal. Stent thrombosis was defined as an acute coronary syndrome with angiographic documentation of either vessel occlusion (TIMI flow grade 0 or 1) or thrombus within or adjacent to a previously successfully stented vessel (TIMI flow grade 1 or 2).
Bleeding events were classified as major, minor, or insignificant according to the criteria of the TIMI Study Group. Major bleeding events were defined as intracranial bleeding or a bleeding event that caused fall in hemoglobin of >4 g/dl or that required transfusion of >3 U of blood. Thrombocytopenia was defined as a platelet count <100 x 103/mm3.
Coronary angiography was performed at baseline, immediately after the procedure, and 6 months later. Quantitative diameter measurements of the coronary arteries were obtained by a blinded reviewer using a workstation with dedicated software (Philips H5000 or Allura DCI program, Philips Medical Systems, Eindhoven, the Netherlands) by a standard technique. In-stent restenosis was defined as an in-stent luminal diameter stenosis
40%. Late lumen loss was defined as the difference between the minimal luminal diameter immediately after stenting and the minimal luminal diameter at follow-up. The intravascular ultrasound (IVUS) imaging was performed in all patients before and after procedure and at follow-up. The IVUS images were acquired with motorized pullback at a constant speed of 1 mm/s (Galaxy, Boston Scientific, Natick, Massachusetts, or Endosonics, EndoSonics Corporation, Rancho Cordova, California).
Study end points. Clinical evaluation was done at 30 days, 6 months, and 1 year to assess patient symptom frequency and cardiac event rates. The primary end point was in-stent late lumen loss and the secondary end point was intrastent luminal volume by IVUS and composite of major cardiac events, including cardiac death, any myocardial infarction, and percutaneous or surgical revascularization of the target lesion 12 months after procedure.
Statistical analysis. We calculated that with a sample of 96 patients, the study would have 80% power to detect a difference in the mean late luminal loss of 0.30 mm between the two groups, assuming a standard deviation of 0.5 mm in each group, with the use of a two-group t test and a two-sided significance level of 0.5. All analyses were based on the intention-to-treat principle. For continuous variables, differences between the treatment groups were evaluated by analysis of t test. For categorical variables, differences were expressed as counts and percentages and were analyzed with the chi-square test. Statistical analysis was performed with the aid of commercially available software (SPSS Version 10.0, SPSS Inc., Chicago, Illinois).
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
This study is the first randomized, prospective clinical trial evaluating effects of abciximab-coated stents in patients with AMI. Patients who received abciximab-coated stents were free of procedure-related complications, and the incidence of in-hospital and 30-day MACE in this patient population was not significantly different when compared with patients who received uncoated control stents. Because all cases recruited for this study had not received PCI within 6 h, it is difficult to interpret our results in the same context as that of other primary PCI studies. In terms of restenosis and target lesion revascularization rates, however, our results (13.9% and 10.4%, respectively) were superior to those of the Stent PAMI (22) trial (20.3% and 7.7%) or the CADILLAC (23) study (22.2% and 8.9%), which compared outcomes of stenting in AMI patients. Moreover, these results are less favorable compared with those of sirolimus-eluting stent implantation by Saia et al. (24) (late loss: 0.04 ± 0.25 mm) and Lemos et al. (25) (target lesion revascularization rate: 1.1%; MACE rate: 9.4%). The incidence of overall MACE was lower in the abciximab-coated stent group, which was primarily attributed to a decrease in the incidence of AMI. In fact, as was suggested by Jeremia et al. (26), who reported that the incidence of stent thrombosis after sirolimus-eluting stent implantation was only 1.1%, which is similar to that after conventional stenting, acute or subacute stent thrombosis associated with drug-eluting stent implantation is not thought to be clinically significant.
Although the patient number was relatively small, the absence of episodes of AMI by acute or subacute thrombotic occlusion during one-year clinical follow-up in the patients who received abciximab-coated stents suggests that platelet aggregation was effectively inhibited with use of abciximab, and this effect could be maintained for a long-term period. Postprocedural clopidogrel was given for two months according to the same protocol for conventional stenting, whereas it was prescribed for a more prolonged period of time in the trials using DES. Abciximab-coated stent implantation was not associated with significant bleeding complications. No patient in either group received platelet glycoprotein IIb/IIIa blockers during PCI, and there was no difference in the postprocedural antiplatelet regimen between the two groups.
One of the main limitations of this study is that PCI was not performed at least within 6 h of onset of chest pain in all cases, whereby the true effects of abciximab-coated stents in AMI might have been rendered less pronounced. And then, this clinical study included only 98 patients with de novo, single vessel stenting, and follow-up coronary angiogram was performed in only 75% to 77% patients. Furthermore, it is only a single center study.
In conclusion, abciximab-coated stents in AMI patients were feasible and safe. Low rates of restenosis and revascularization were observed. At one-year clinical follow-up, these results appear to be sustained. These observations warrant further investigation with a large, randomized multicenter study.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. S. Rihal Drug-Eluting Stents for ST-Segment Elevation Myocardial Infarction: Treatment of Choice or Is Discretion the Better Part of Valor? J. Am. Coll. Cardiol. Intv., June 1, 2008; 1(3): 233 - 235. [Full Text] [PDF] |
||||
![]() |
K. E. Kip, K. Hollabaugh, O. C. Marroquin, and D. O. Williams The Problem With Composite End Points in Cardiovascular Studies The Story of Major Adverse Cardiac Events and Percutaneous Coronary Intervention. J. Am. Coll. Cardiol., February 19, 2008; 51(7): 701 - 707. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. R. Dixon, C. L. Grines, and W. W. O'Neill The Year in Interventional Cardiology J. Am. Coll. Cardiol., July 17, 2007; 50(3): 270 - 285. [Full Text] [PDF] |
||||
![]() |
Abciximab-Coated Stents in AMI: Saving Muscle . . . or Not Journal Watch Emergency Medicine, April 25, 2006; 2006(425): 5 - 5. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |