J Am Coll Cardiol, 2006; 47:319-325, doi:10.1016/j.jacc.2005.09.028 © 2006 by the American College of Cardiology Foundation |



* Section of Pediatric Cardiology, Department of Pediatrics, University of Chicago Comer Childrens Hospital, Pritzker School of Medicine, Chicago, Illinois
Division of Pediatric Cardiology, University of Minnesota, Minneapolis, Minnesota
Joint Division of Pediatric Cardiology, University of Nebraska/Creighton University, Childrens Hospital of Omaha, Omaha, Nebraska
Section of Pediatric Cardiology, MUSC Childrens Hospital, Charleston, South Carolina, and Childrens Hospital of Pittsburgh, Pittsburgh, Pennsylvania
|| Section of Pediatric Cardiology, Department of Pediatrics, Columbus Childrens Hospital, Columbus, Ohio
¶ Section of Pediatric Cardiology, Department of Pediatrics, Columbia University, New York, New York
# Section of Pediatric Cardiology, Department of Pediatrics, St. Louis Childrens Hospital, St. Louis, Missouri
Manuscript received March 22, 2005; revised manuscript received August 20, 2005, accepted September 8, 2005.
* Reprint requests and correspondence: Dr. Ziyad M. Hijazi, Section of Pediatric Cardiology, University of Chicago Comer Childrens Hospital, 5841 South Maryland Avenue, MC4051, Chicago, Illinois 60637. (Email: zhijazi{at}peds.bsd.uchicago.edu).
| Abstract |
|---|
|
|
|---|
BACKGROUND: The most common congenital heart disease is PmVSD. Surgical repair is widely accepted, but still carries a small but definite risk of morbidity and mortality.
METHODS: Between October 2003 and August 2004, a total of 35 patients with PmVSD underwent an attempt of transcatheter closure under transesophageal and/or intracardiac echocardiographic guidance. The median age was 7.7 years (range, 1.2 to 54.4 years) and median weight was 25 kg (range, 8.3 to 110 kg). The median Qp/Qs ratio was 1.8 (range, 1 to 4), and the median VSD size as assessed by echocardiography was 7 mm (range, 4 to 15 mm).
RESULTS: The attempt to place a device was successful in 32 patients (91%). The median device size used was 10 mm (range, 6 to 16 mm). The complete closure rates by echocardiography at 10 min (transesophageal/intracardiac), 24 h, 1 month, and 6 months (transthoracic) were 47% (15/32), 63% (20/32), 78% (25/32), and 96% (27/28), respectively. The median fluoroscopy time was 36 min (range, 14 to 191 min), and the median total procedure time was 121 min (range, 67 to 276 min). Three patients (8.6%) had serious adverse events of complete heart block, peri-hepatic bleeding, and rupture of tricuspid valve chordae tendineae. No other patient encountered serious adverse events during the follow-up.
CONCLUSIONS: Transcatheter closure of a PmVSD is technically feasible and seems safe enough in children over 8 kg in weight to warrant continuation of clinical trials to assess the long-term safety and efficacy.
| |||||||
Since 1987, the Rashkind and buttoned devices have been used to close PmVSDs, which were originally designed for patent ductus arteriosus and atrial septal defect, respectively (69). The major drawbacks of these devices were the large delivery sheaths (11-F) required, complex implantation techniques, inability to reposition and redeploy the device, interference with the aortic and tricuspid valves and significant residual shunts (25% to 60%) (610). The Amplatzer Membranous VSD Occluder (AGA Medical Corp., Golden Valley, Minnesota) is the only device that is specifically designed for PmVSDs (11). Its initial use in six patients in 2002 was associated with complete closure in all and absence of significant complications (12). After that, several centers reported similar encouraging results of initial complete closure rates of 90% to 92% (1316). However, they were all reports with limited case numbers. The purpose of this phase I study was to report the initial safety and efficacy results of transcatheter closure of PmVSD using the new Amplatzer Membranous VSD Occluder (AGA Medical Corp.) in the U.S.
| Methods |
|---|
|
|
|---|
The inclusion criteria of patients included: 1) PmVSD as shown by echocardiography, and 2) symptoms of heart failure or evidence of left atrial and or left ventricular enlargement for body surface area. Exclusion criteria included: 1) body weight <8 kg, 2) subaortic rim as shown by echocardiography in the long-axis view <2 mm, 3) left ventricle to right atrial shunting, 4) right to left shunting through the defect, 5) PmVSD with an aneurysm and multiple shunts that could not be successfully closed with one device, 6) sepsis, 7) complex heart lesions such as tetralogy of Fallot, 8) contraindication to antiplatelet therapy, 9) patients unable to be followed up for the duration of the trial, and 10) inability to obtain informed consent.
Between October 2003 and August 2004, a total of 35 patients (23 male, 12 female) underwent an attempt at transcatheter device closure (Table 1). Among them were 3 (9%) patients with postoperative residual VSDs, 16 (46%) patients had aneurysm formations around the VSDs, and 11 patients had associated other congenital heart diseases, including patent foramen ovale in 4, pulmonary valvular stenosis/dysplasia in 3, bicuspid aortic valve in 2, small muscular VSDs in 1, patent ductus arteriosus in 1, mild Ebstein anomaly of the tricuspid valve in 1, and dextrocardia with situs inversus in 1. Trivial and mild aortic regurgitation with no aortic valve prolapse before the procedure was noted in nine (25.7%) patients. Six patients had congestive heart failure, three had recurrent respiratory infections, three had failure to thrive, two had pulmonary hypertension, one had cyanosis attributable to her underlying cardiac anatomy of corrected transposition and right-to-left shunt at the atrial level, and one had palpitations.
|
All other adverse events were considered observations. The accuracy of the submitted data was verified for each participating institution using source documents (catheter reports, case notes, medical records) that were analyzed by monitors from the sponsor during control visits to these institutions.
Measured outcome parameters.
Procedure success was defined by device release in the appropriate position without embolization. The residual shunt was classified as trivial (width of the color jet as assessed by color Doppler echocardiography as it exited the septum of <1 mm), small (width of 1 to 2 mm), moderate (width of 3 to 4 mm) or large (width
4 mm), in a way similar to the protocol reported by Boutin et al. (17) for shunt assessment after device closure of atrial septal defects. Experienced echocardiographers at the echocardiography core laboratory (based at the sponsors headquarters) verified the pre-closure, closure, and post-closure echocardiograms.
The device. The Amplatzer Membranous VSD Occluder (AGA Medical Corp.) is a self-expandable double-disk device made of a nitinol wire mesh. The details of the device description were previously reported (12).
Closure protocol.
The protocol used has been described in detail in previous reports (11,12). Briefly, access was obtained in the femoral artery (4- to 5-F sheath) and the femoral vein (7- to 9-F sheath). Heparin was given to keep an activated clotting time of >250 s. Routine right and left heart catheterizations were performed to assess the degree of shunting and to evaluate pulmonary vascular resistance. Left ventriculography in the long axial oblique view (60° left anterior oblique/20° cranial) defined the location and size of the VSD. The appropriate device size was chosen to be at least 1 to 2 mm larger than the VSD size as measured by color Doppler echocardiography or ventriculography. Transesophageal echocardiography (TEE) and/or intracardiac echocardiography (ICE) were used to guide and monitor the procedure. Figures 1 and 2
show the various steps of closure by cine fluoroscopy and echocardiography, respectively. The follow-up protocol included physical examination and electrocardiogram and echocardiography at 10 min (TEE/ICE), pre-discharge, one month, and six months (transthoracic echocardiography [TTE]) after the procedure. Chest X-ray was required before discharge and 24 h after the procedure. Patients were routinely maintained on aspirin 3 to 5 mg/kg daily or equivalent antiplatelet therapy for six months. Patients were instructed to receive infective endocarditis prophylaxis when needed until complete closure was documented at the six-month follow-up visit.
|
|
| Results |
|---|
|
|
|---|
Follow-up data. Table 2 summarizes the data before and after device insertion, including presence of heart murmur, electrocardiogram, chest radiograph, and echocardiographic assessment of shunt status, size of the left atrium and ventricle, and presence or absence of valvar regurgitation. The complete closure rate increased from 20 of 32 at 24 h after the procedure to 27 of 28 at six-month follow-up. Furthermore, as noted in Table 2, aortic regurgitation (AR) was present in 9 of 35 (25.7%) patients before closure (6 with trivial and 3 with mild AR). At discharge, 17 of 32 (53.1%) patients had AR (11 with trivial and 6 with mild AR). At one-month follow-up, 15 of 32 (46.9%) patients had AR (10 with trivial and 5 with mild AR), and at six-month follow-up, 11 of 28 (39.3%) patients had AR (7 with trivial and 4 with mild AR). Because of the small sample size in each category of AR, we were unable to predict who would develop or have an increase in the severity of AR. Data for the tricuspid and mitral valves are shown in Table 2.
|
| Discussion |
|---|
|
|
|---|
Because the membranous septum is very thin, the operator may not feel any resistance during device pullback from the left ventricle to the septum. As a result, the device might be misplaced in the right ventricle. Therefore, echocardiography (TEE or ICE) and angiography are crucial to identify correct device deployment. In addition, echocardiography can recognize immediate possible adverse events of AR caused by impingement of the device on the valve leaflets and mitral or tricuspid regurgitation caused by the interference with the chordae tendineae.
Our results showed that presence of an aneurysm did not prevent device closure. If the aneurysm was large and the opening was solitary, the device could be accommodated within the aneurysmal sac, away from the aortic valve, thereby preventing AR (9). In most situations, the device was placed in the usual septal position (Figs. 1 and 2).
This series reports the largest cohort of patients with PmVSDs who underwent transcatheter closure using the new Amplatzer Membranous VSD Occluder (AGA Medical Corp.). The successful procedure rate was 91%. The complete closure rate was excellent, 96% at six months after the procedure, with the remaining 4% (one patient) having only trivial residual shunt.
The left atrial and ventricular sizes decreased after the procedure. Both left atrial and ventricular dimensions achieved a significant decrease in the Z score at 30 days and six months (Table 2). This indicates that the device was effective in abolishing the left-to-right shunt that resulted in a decrease in the left heart chambers. Mitral and tricuspid regurgitation improved in some patients after the procedure. The proportion of mitral regurgitation decreased from 34.3% before device implantation to 12.5% at one-month follow-up. This again indicated that patients benefited from elimination of the shunt.
No mortality, stroke, or neurologic deficit was noted in our series. This compares favorably to the current surgical results for isolated PmVSDs (4,5).
Hobbins et al. (18) reported that complete right bundle branch block occurred in 79% and 33% of patients after surgical repair of VSD through the right atrium or right ventricle approach, respectively. Yeager et al. (3) reported that 75.3% of patients had persistent postoperative conduction abnormality including complete heart block (2.3%), complete right bundle branch block (64%), and bifascicular block (9%). However, patients who were in that study (3) were younger than one year of age; therefore, direct comparison with our series of patients, who were older, should be viewed with caution. The current study found that bundle branch block occurred in only 6% of patients (2 of 32) who underwent device closure. Permanent complete heart block requiring a pacemaker developed in one patient.
The development of AR is an important concern for device closure of PmVSDs. In the 32 patients undergoing device closure, new trivial AR without clinical or hemodynamic significance within 24 h developed in 5 patients (15.6%), and new mild AR within 24 h developed in 3 patients (9%), for a total of 17 of 32 (53%) patients with AR (Table 2). This might be related to the configuration change of the subaortic septum or repeated crossing of the aortic valve by catheters and guidewires. At six-month follow-up, 11 of 28 (39%) patients had AR. This indicates that some of the aortic regurgitation was transient.
Study limitations. The patients in our study were larger and older than those in whom surgical closure of a VSD is considered. The application of the Amplatzer Membranous VSD Occluder (AGA Medical Corp.) in small infants may carry a higher risk and remains to be determined. Our follow-up period was only six months. The preliminary results and possible delayed adverse effects, including heart block, need to be confirmed by a larger trial and by long-term follow-up.
Conclusions. Transcatheter closure of PmVSDs using the Amplatzer Membranous VSD Occluder (AGA Medical Corp.) is technically feasible and seems safe enough in larger children to warrant continuation of clinical trials to assess the long-term safety and efficacy of this device.
| Addendum |
|---|
|
|
|---|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E. Bedard, D. F. Shore, and M. A. Gatzoulis Adult congenital heart disease: a 2008 overview Br. Med. Bull., March 1, 2008; 85(1): 151 - 180. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Klcovansky, L. Sondergaard, M. Helvind, and H. O. Andersen Cardiac surgery in grown-up congenital heart patients. Will the surgical workload increase? Interactive CardioVascular and Thoracic Surgery, February 1, 2008; 7(1): 84 - 89. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. J. Zeng, S. Q. Sun, X. F. Chen, X. J. Ma, Y. H. Luo, Y. P. Lim, and L. Tao Device Closure of Perimembranous Ventricular Septal Defects With a Minimally Invasive Technique in 12 Patients Ann. Thorac. Surg., January 1, 2008; 85(1): 192 - 194. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Carminati, G. Butera, M. Chessa, J. De Giovanni, G. Fisher, M. Gewillig, M. Peuster, J. F. Piechaud, G. Santoro, H. Sievert, et al. Transcatheter closure of congenital ventricular septal defects: results of the European Registry Eur. Heart J., October 1, 2007; 28(19): 2361 - 2368. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. M. Tucker, L. A. Pyles, J. L. Bass, and J. H. Moller Permanent Pacemaker for Atrioventricular Conduction Block After Operative Repair of Perimembranous Ventricular Septal Defect J. Am. Coll. Cardiol., September 18, 2007; 50(12): 1196 - 1200. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Butera, M. Carminati, M. Chessa, L. Piazza, A. Micheletti, D. G. Negura, R. Abella, A. Giamberti, and A. Frigiola Transcatheter Closure of Perimembranous Ventricular Septal Defects: Early and Long-Term Results J. Am. Coll. Cardiol., September 18, 2007; 50(12): 1189 - 1195. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. R. Dixon, C. L. Grines, and W. W. O'Neill The Year in Interventional Cardiology J. Am. Coll. Cardiol., July 17, 2007; 50(3): 270 - 285. [Full Text] [PDF] |
||||
![]() |
I. Inglessis and M. J. Landzberg Interventional Catheterization in Adult Congenital Heart Disease Circulation, March 27, 2007; 115(12): 1622 - 1633. [Full Text] [PDF] |
||||
![]() |
P. Acar, S. Abadir, and Y. Aggoun Transcatheter closure of perimembranous ventricular septal defects with Amplatzer occluder assessed by real-time three-dimensional echocardiography Eur J Echocardiogr, March 1, 2007; 8(2): 110 - 115. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. D Sullivan Transcatheter closure of perimembranous ventricular septal defect: is the risk of heart block too high a price? Heart, March 1, 2007; 93(3): 284 - 286. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Rajakaruna, J. Hill, E. J. H. Turner, A. Sirker, B. S. Rana, and O. Wendler A novel surgical approach to close an acute ventricular septal defect using an occluder device J. Thorac. Cardiovasc. Surg., February 1, 2007; 133(2): 579 - 580. [Full Text] [PDF] |
||||
![]() |
M. S. Minette and D. J. Sahn Ventricular Septal Defects Circulation, November 14, 2006; 114(20): 2190 - 2197. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. P. Graham Jr The Year in Congenital Heart Disease J. Am. Coll. Cardiol., June 20, 2006; 47(12): 2545 - 2553. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |