J Am Coll Cardiol, 2006; 47:296-300, doi:10.1016/j.jacc.2005.09.024
(Published online 29 December 2005). © 2006 by the American College of Cardiology Foundation |


,2
* Division of Invasive Cardiology, Rambam Medical Center and the Technion, Haifa, Israel
Cardiovascular Disease Institute "Prof. C. C. Iliescu," Fundeni Hospital, Bucharest, Romania
Department of Cardiology, Shaarei Zedek Medical Center, Jerusalem, Israel
NaviCath Ltd., Haifa, Israel
Manuscript received May 14, 2005; revised manuscript received September 2, 2005, accepted September 12, 2005.
* Reprint requests and correspondence: Dr. Rafael Beyar, Womens Division/Dr. Phillip and Sarah Gotlieb Chair in Medicine and Biomedical Engineering, Head, Division of Invasive Cardiology, Rambam Medical Center, Haifa, Israel. (Email: rafael{at}tx.technion.ac.il).
| Abstract |
|---|
|
|
|---|
BACKGROUND: Percutaneous coronary interventions (PCIs) are manually performed under fluoroscopic guidance, requiring lead protection for the operators. A system in which the operator can remotely, safely, and precisely navigate the procedure during PCI would have clear advantages.
METHODS: The RNS involves a computer-controlled wire and delivery system navigator. Following preclinical validation, the system was assessed in patients undergoing single-vessel PCI.
RESULTS: The study involved 18 patients (age 55.9 years, 16% women). The RNS successfully crossed lesions with the guidewire in 17 patients. The stent was then advanced by the advance/rotate mode and adequately positioned in 15 of 17 cases. Technical malfunction was encountered in three patients in whom the procedure was successfully completed manually. Direct stenting was employed in 10 of 18 patients, pre-dilation in 7 patients, and after-stent balloon dilation in 5 patients. The total fluoroscopy time for 17 RNS patients was compared with the corresponding time of 20 consecutive patients who underwent standard single-lesion PCI. Fluoroscopy time was similar for both procedures, with 8.8 ± 4.8 min with the RNS versus 9.1 ± 3.5 min with the standard techniques (p = NS). Clinical success was 100% and technical success 94% for the guidewire and 83% for the overall procedure.
CONCLUSIONS: The use of the RNS for guidewire, balloon, and stent manipulation during PCI appears safe and feasible for the treatment of patients with coronary stenosis. The system offers operator radiation safety and may enhance precision of stent placement and balloon dilation strategies.
| |||||||||
| Methods |
|---|
|
|
|---|
|
|
|
Preclinical studies. The system was initially tested on a transparent glass coronary model. The model showed that the guidewire can be easily manipulated through branches and that the angioplasty device could be easily positioned at the required location. The system was then tested in an anesthetized coronary sheep model. The procedure was repeated in each animal several times; after stent implantations by the RNS, the results were angiographically assessed.
Pilot clinical study. A total of 18 patients (age 55.9 ± 12.0 years; 15 men, 3 women) underwent RNS-guided PCI at three medical centers under an approved protocol. After femoral cannulation with a 6- or 7-F sheath, the target coronary artery was engaged with the appropriate guiding catheter, which was hooked to the RNS. Contrast injections were done either manually or by Automatic Coronary Injection System (ACIST Medical Systems Inc., Eden Prairie, Minnesota). The RNS was loaded with standard 0.014-inch coronary guide wires that were navigated across the lesion by the operator. Subsequently, the device was driven to the lesion using both continuous- and discrete-mode movements for precise positioning. The lesion was predilated in cases of severe stenosis (n = 8), and high-pressure post-stent dilation was performed in cases of suboptimal results (n = 5). Baseline and final angiograms were recorded. The patients underwent follow-up for immediate, in-hospital, and 30-day major adverse coronary events (MACE) (death, myocardial infarction, urgent revascularization).
The study end points were success in navigating the guidewire across the lesion and success in precise positioning of the device. Technical success was defined as the ability to complete the procedure without reverting to manual mode. Clinical success was defined as the ability to successfully complete the procedure without complications. Offline quantitative coronary angiography was performed at the Rambam Core Laboratory using a CAAS II system (Pie Medical Imaging, Maastricht, the Netherlands). Total catheterization times and total fluoroscopy time (TFT) were recorded for 17 cases (1 unavailable) and were compared with a control group consisting of 20 consecutive patients who underwent single-lesion PCI during the same time period at the Rambam Medical Center.
| Results |
|---|
|
|
|---|
Pilot clinical study. The majority of patients (16 of 18) had stable angina pectoris and typical risk factors for coronary artery disease (diabetes, 22%; hypertension, 39%; smoking history, 32%; hyperlipidemia, 83%; previous myocardial infarction, 39%). The majority (n = 11) of patients underwent treatment of the left anterior descending coronary artery (LAD) 11 patients, left circumflex artery (LCX) (2 patients), and right coronary artery (RCA) (5 patients). Lesion types ranged between A and C with minimal calcification or angulations.
Clinical success was achieved in all patients. The guidewire was successfully navigated across the lesion in 17 of 18 cases. In one case a technical problem with the RNS (the system stalled) was successfully managed by immediate transition to manual operation. The stent was successfully delivered in 15 of 17 cases. In two cases, the procedures were completed manually because of device malfunction. Overall, technical success was achieved in 15 of 18 of the patients. Ten patients were treated with direct stenting. In five patients, high-pressure after-stent dilation was performed. All patients were treated with a single stent, except for two patients who required additional stents because of incomplete lesion coverage and distal dissection. There were no MACE during the procedure or during hospitalization. One patient had a non-target vessel myocardial infarction three weeks after the procedure.
With respect to procedural times, TFT was 8.8 ± 4.8 min with the RNS versus 9.1 ± 3.5 min with the standard techniques (p = NS). Total catheterization time was 44 ± 32.7 min with the RNS versus 61 ± 19 min with the regular cases (p = NS). Although the groups are not randomized, they are comparable with respect to the single-vessel treatment, the mixture of vessels (LAD, 10; RCA, 8; and LCX, 2), and the requirements for additional stents (3 in the control group vs. 2 in the treatment group). Therefore, procedural times for the RNS are comparable to those with standard PCI, despite the early phase of the technology and the limited clinical experience.
By QCA, the reference vessel diameter was 2.73 ± 0.66 mm before PCI and 2.95 ± 0.50 mm after PCI. Minimal luminal diameter increased from 1.06 ± 0.42 mm to 2.43 ± 0.40 mm and the diameter stenosis decreased from 60.6 ± 13.5% to 17.0 ± 7.8%.
| Discussion |
|---|
|
|
|---|
Robotic manipulation of surgical devices has been reported for several applications, including computed tomography-guided biopsy (9) and surgery (1012). The shared feature of robotic surgery and the current RNS is the ability to translate the operator maneuvers to remotely manipulate devices and potential increased accuracy (13,14). However, there are basic differences between the surgical and endovascular systems. The surgical systems enable working through small ports under direct visual control and offer patient benefit related to the minimally invasive feature of the technique. The current RNS works through the standard port and does not alter the procedure for the patient, but may have a precision benefit for the patient and a benefit for the operator.
Remote-control magnetically driven manipulation of catheters was recently applied to electrophysiologic procedures as well as to coronary wire navigation (15). In contrast to the magnetically driven system, which has shown magnetic navigation of specific guide wires, the current RNS performs the entire stent-assisted PCI using standard equipment.
The potential advantages of a remote catheterization system can be summarized as:
Some limitations are inherent to this early phase of device development and evaluation. The system was applied to relatively simple lesions. The ability of the system to tackle complex cases will have to be addressed in the future by upgrading the engineering design and adding additional features that have been suggested for the surgical systems, such as haptic mechanical force feedback (16). In addition, guiding catheter control may be important for online modification of catheter support that may be needed in some cases. Therefore, the results of this study should be considered only as encouraging and facilitating further testing.
In summary, we have shown that remote-controlled, stent-assisted PCI by the RNS is feasible and appears safe. The RNS provides convenient conditions for the operator in an X-rayfree environment and can increase the precision of stent deployment. This report opens the door for ergonomic technologic developments in the catheterization laboratory, aiding physicians in achieving their therapeutic goals.