J Am Coll Cardiol, 2005; 46:253-260, doi:10.1016/j.jacc.2005.03.069
(Published online 5 July 2005). © 2005 by the American College of Cardiology Foundation |












* Erasmus Medical Center, Rotterdam, the Netherlands
Instituto Dante Pazzanese de Cardiologia, Sao Paulo, Brazil
Amphia Ziekenhuis, Breda, the Netherlands
Catharina Ziekenhuis, Eindhoven, the Netherlands
|| Medisch Centrum Rijnmond-Zuid, Rotterdam, the Netherlands
¶ Christchurch Hospital, Christchurch, New Zealand
# Academisch Ziekenhuis Middelheim, Antwerp, Belgium
** Instituto Cardiovascular de Buenos Aires, Buenos Aires, Argentina

Hospital Miguel Perez Carreno, Caracas, Venezuela

Herzzentrum Bodensee, Kreuzlingen, Switzerland

Conor Medsystems, Menlo Park, California
|||| Cardialysis, Rotterdam, the Netherlands.
Manuscript received November 30, 2004; revised manuscript received March 17, 2005, accepted March 29, 2005.
* Reprint requests and correspondence: Dr. Patrick W. Serruys, Erasmus Medical Center, Thoraxcenter, Bd-406, Dr. Molewaterplein 40, 3015 GD Rotterdam, The Netherlands. (Email: p.w.j.c.serruys{at}erasmusmc.nl).
| Abstract |
|---|
|
|
|---|
BACKGROUND: Conventional paclitaxel-eluting stents use a durable polymer coating as a vehicle for drug delivery. The Conor stent (Conor Medsystems, Menlo Park, California) with intra-strut wells and erodable polymer is specifically designed for drug delivery with programmable pharmacokinetics.
METHODS: Two hundred and forty-four patients with single vessel disease received either a bare metal Conor stent (n = 53) or one of six different release formulations that varied in dose (10 or 30 µg) and elution release kinetics (first order, zero order), direction (abluminal, luminal), and duration (5, 10, and 30 days). End points at six months (bare stent group) and at four months (eluting stent groups) were angiographic late loss and neointimal tissue volume by intravascular ultrasound and the rate of major adverse cardiac events (MACE).
RESULTS: The lowest in-stent late loss (0.38 mm, p <0.01, and 0.30 mm, p <0.01) and volume obstruction (8%, p <0.01, and 5%, p <0.01) were observed with the 10-µg and 30-µg doses in the 30-day release groups respectively, whereas the highest in-stent late loss (0.88 mm), volume obstruction (26%), and restenosis rate (11.6%) were observed in the bare stent group. The overall MACE rate of the eluting stent group was 8.6%: death 0.5%, myocardial infarction 2.7%, and target lesion revascularization (TLR) 5.3%. Sub-acute thrombosis was 0.5%. The TLR rates in the two 30-day release groups were 0% and 3.4%.
CONCLUSIONS: This novel eluting stent platform, using an erodable polymer with complete elution of low doses of paclitaxel, is safe. The inhibition of the in-stent neointimal hyperplasia was best in the long release groups.
| |||||||||||||
Each DES comprises three components: the stent platform, the active pharmacologic compound, and a drug carrier vehicle, usually a polymer, that controls drug elution. Conventional research for development of new and more effective DES has focused on the use of new drugs and advanced biocompatible polymers coated on stent struts. This approach has limitations, including limited control of kinetic profiles and drug-loading capacity.
A novel metallic stent has been engineered specifically as a coronary drug delivery system. It is designed to permit precise and programmable control over spatial and temporal release profiles and to enhance the drug-loading capacity (5).
The present study evaluates this new stent platform without drug and with six different pharmacokinetic release formulations of paclitaxel. These formulations have previously been investigated for their safety and efficacy in animal models (6). The objective of the Paclitaxel In-Stent Controlled Elution Study (PISCES) trial is to compare the safety and performance of different doses and release rates from the Conor paclitaxel-eluting stent for reducing clinical events and angiographic late loss at four months when used in native coronary vessels.
| Methods |
|---|
|
|
|---|
The PISCES trial was a prospective, multi-center, sequentially enrolled, non-randomized, open-label trial in which the patient data collected from six release formulations were compared with one another and to the historical bare stent cohort.
One hundred and ninety-one subjects from the ten participating sites were enrolled in the PISCES trial and received a paclitaxel-eluting stent with one of the six release formulations. At least one study stent was implanted in each subject. When additional stents were required for treatment of edge dissection, stents from the same formulation type were used.
Clinical follow-up was conducted at one and four months after the index procedure. Quantitative angiography and IVUS were performed at four months. Provisional angiographic and IVUS follow-up is planned at 12 months in the dose treatment groups showing efficacy.
End points. The safety end point of the study is a composite of major adverse cardiac events (MACE), defined as death, Q-wave or non-Q-wave myocardial infarction (MI), and target lesion revascularization (TLR; coronary artery bypass grafting or percutaneous coronary intervention) at four months. All deaths were considered cardiac unless they were unequivocally documented to be non-cardiac. Myocardial infarction was diagnosed by a rise in the creatine kinase level to more than twice the upper normal limit with an increased creatine kinase-MB accompanied by new abnormal Q-waves in the surface electrocardiogram (Q-wave MI) or not (non-Q-wave MI). Target lesion revascularization was defined as revascularization of the stented and the peri-stent segments (5 mm proximal and distal). Target vessel revascularization (TVR) was defined as revascularization due to narrowing (>50% diameter stenosis) of any portion of the target vessel outside the peri-stent segment, but was not included as an event in the MACE rate. Both TLR and TVR are reported as actual rates without adjudication for clinical indication.
The efficacy end points included the in-stent and peri-stent angiographic late loss and binary restenosis rate as well as percent volume obstruction of the stent and neointimal hyperplasia of the stented and peri-stent segments, as determined by quantitative IVUS.
Patient selection. Subjects were eligible for the study if they were 18 to 80 years of age, had single de novo lesions with a reference diameter of 2.5 to 3.5 mm and a lesion length that could be covered by a single 17-mm stent, and if they had stable or unstable angina pectoris or documented silent ischemia.
Subjects were excluded from the study if they had an acute MI 72 h before the procedure, an ejection fraction <30%, stroke/gastro-intestinal bleeding within six months, severe hepatic disease or renal insufficiency with a serum creatinine level >2.5 mg/dl, or known intolerance or contraindication to aspirin and/or clopidogrel. Further angioplasty exclusion criteria included total occlusions (Thrombolysis In Myocardial Infarction flow grade 0), bifurcational (adjacent branch >2 mm) and ostial lesions, left main disease, and tortuous target vessel. Written informed consent was obtained from all patients.
The inclusion and exclusion criteria for both the bare metal registry phase and the PISCES trial were identical.
Study device. Figure 1 shows the balloon-expandable, 316L stainless steel Conor stent. The unique design features include struts with holes along their length, linked to flexible sinusoidal bridges by specially contoured features called ductile hinges. Unlike conventional stents consisting of repeating units wherein the entire structure is deformed by expansion forces placed on it, the Conor stent differs in that deformation is confined to the 10% of the stent comprising the ductile hinges, rendering the struts as passive elements. This allows the struts to be cored with holes or reservoirs for drug delivery with no effect on the strength or crush resistance of the struts. The holes can be inlaid with polymer/drug that will not deform or separate from the stent during stent expansion, and bench testing shows no extrusion of polymer (5). The strut thickness is 127 µm (range 122 to 132 µm).
|
Dose and kinetic release profiles. The drug under evaluation in the PISCES study was paclitaxel. The Conor stent is designed to deliver a similar dose density (measured in µg/mm2 of vessel surface area) to all vessels treated; stents comprise varying numbers of stent cells. As the stents increase in diameter, more stent cells are used to complete the structure. In this way, the total contained dose on the stent will increase in proportion to the vessel diameter, thereby maintaining approximately a consistent dose per unit of vessel surface area (6). The PISCES trial evaluated one of six different release formulations that varied in dose (10 or 30 µg) and elution release kinetics (first order, zero order), direction (abluminal, luminal), and duration (5, 10, and 30 days). The release profiles for the formulations are shown in Figure 2 and Table 1.
|
|
80 mg daily and clopidogrel 75 mg daily) was prescribed for at least six months after procedure. A loading dose of 300 mg clopidogrel was given before the procedure. Quantitative coronary angiography evaluation. Quantitative coronary angiography (QCA) was performed by means of the CAAS II analysis system (Pie Medical BV, Maastricht, the Netherlands). In each patient, the following segments were analyzed:
The following QCA parameters were computed: computer-defined minimal luminal diameter (MLD), reference diameter obtained by an interpolated method, and percentage diameter stenosis. Binary restenosis was defined in every segment as diameter stenosis >50% at follow-up. Late loss was defined as the difference between MLD after procedure and MLD at follow-up.
Quantitative IVUS. Post-procedure and follow-up stented vessel segments were examined with mechanical IVUS (Cardio Vascular Imaging System, CVIS, Sunnyvale, California) with automated pullback at 0.5 mm/s. A coronary segment beginning 5 mm distal to and extending 5 mm proximal to the stented segment was also examined. A computer-based contour detection program was used for automated three-dimensional reconstruction of the stented and adjacent segments. The lumen, stent boundaries, and external elastic membrane (vessel boundaries) were detected with a minimum cost algorithm. The stent volume (SV) and lumen volume (LV) were calculated according to Simpsons rule. The intra-stent neointimal volume was calculated as "SV-LV." The percentage obstruction of the SV was calculated as intra-stent neointimal volume/SV x 100. Feasibility, reproducibility, and inter- and intra-observer variability of this system have been validated in vitro and in vivo (7,8). The IVUS data of the proximal and distal peri-stent segments are expressed in mean area (vessel and lumen area) instead of volume, because the analyzable lengths of these segments were variable, owing to anatomical (side branch) and/or technical limitations. The "plaque" behind the stent struts was also expressed in mean area.
Statistical analysis. Continuous parameters were presented as mean values and standard deviations, and discontinuous parameters as percentages. For lesion characteristics and procedural outcomes, the following tests were applied to calculate the differences among the seven groups (one bare stent and six paclitaxel-eluting stent groups): F test from an analysis of variance (ANOVA), two-sample t test, likelihood ratio chi-square test, Fisher exact test, and Cochran-Mantel-Haenszel test. Overall QCA and IVUS parameters were compared using general linear models (i.e., one-way ANOVA) for unbalanced data. As per protocol, the Student t test was performed between each release formulation and the bare stent group when ANOVA was positive, and, hence, no correction for multiple comparisons was performed. The statistical significance of all tests was defined at the p <0.05 level.
| Results |
|---|
|
|
|---|
|
|
Serial QCA analyses. Table 5 shows the serial QCA analyses. Each QCA parameter is the average of multiple angiographic views. Angiographic follow-up was available in 93% of the patients. Groups D5 and D6 show the lowest in-stent late loss, which is reduced approximately by 57% and 66%, respectively, when compared with the loss in the bare stent group. In D5 and D6, there were no cases of edge restenosis. The overall restenosis rates in the stented and peri-stent segments in D5 and D6 are 0% and 3.8%, respectively, compared with 14% in the bare stent.
|
|
| Discussion |
|---|
|
|
|---|
The precise reasons for these observations are unclear but may be related to several factors. First, molecular biology studies have demonstrated activation of genes potentially responsible for proliferation for periods up to 21 days (11). As such, the inhibitory compound may need to be present for some minimum period of time. Second, animal studies with the Conor drug delivery system have demonstrated that, at 30 days, all the doses used in the PISCES trial were effective, but slightly higher indexes of injury (e.g., fibrin deposition, eosinophilic deposits) were observed with the shorter release formulations (6). There were no significant differences in the balloon-to-artery ratios between the different groups, suggesting that the histologic variability was pharmacologic rather than mechanical. In these same studies, at 90 days, there was durability of the 30-day results in the longer release formulations that was not seen in the shorter release formulations. One may speculate that the anti-proliferative effects of the drug may be blunted by a secondary injury, induced either by the drug itself or by the bio-absorbable polymer, an effect which is not seen in the longer releases. Our data may help explain some of the discordant data found in other paclitaxel clinical trials (3,4,1214).
The TAXUS II study, a blinded, randomized trial with two paclitaxel treatment arms and two control arms, enrolled over 500 patients. At drug concentrations of 1 µg/mm2 (equivalent to a total drug loading of 108 µg/16-mm stent), delivered in either moderate or slow release formulations, restenosis rates of <5% and reductions in late loss of 60% versus bare-metal control stent were reported. Although there were no reported toxic side-effects related to the use of paclitaxel after 6 or 12 months, concern has been raised about the long-term biological effect of the non-erodable polymer used as well as persistence of significant quantities of drug still present in the polymer at 30 days; 92.5% for the slow release formulation or 78.1% for the moderate release (personal communication from Mary Russell, June 2004).
The DELIVER trial also used paclitaxelwithout a polymer carrierand failed, at a dose density of 3.04 µg/mm2, to demonstrate significant improvement versus the bare stent. The reasons for such discordant results with the same drug are unclear, but suggest that release rates may impact efficacy (12). The present observation may therefore have relevance for optimizing paclitaxel efficacy and, potentially, even have implications for other therapeutic compounds.
A further finding of our study was that in dose 3, which had abluminal release only, and dose 2, which had bi-directional release, there was no differential efficacy either on the in-stent segment or edges of the stent.
The IVUS observations made with the TAXUS polymer-coated DES have been duplicated in these limited populations. First, a significant tissue growth behind the struts of the stent has been observed, accompanied by expansive remodeling of the external elastic membrane (15). Second, lumen reductions at the edge of the stentpredominantly at the proximal edgeare usually seen after bare metal stent implantation, owing to constrictive remodeling combined with plaque growth. This phenomenon is prevented by the paclitaxel-eluting stent, because the plaque growth is partially accommodated by expansive remodeling (16).
Study limitations. This study is subject to several limitations. It was not designed as a pivotal efficacy study but rather as a first-in-man study and, therefore, was not blinded or randomized. As a pilot study, the primary follow-up was conducted at four months for safety reasons similar to other first-in-man studies. Comparison of the neointimal hyperplasia at four months in the paclitaxel-eluting stent with the neointimal hyperplasia at six months in the bare stent was not aimed to show superiority, but to demonstrate that the novel platform itself was not providing unusual results compared with conventional stainless steel stents. Consequently, the study was not statistically modeled to demonstrate efficacy but rather to develop insight into how to optimize the pharmacokinetics and to understand the relative importance of different doses versus duration of elution for paclitaxel.
Conclusions. The PISCES trial demonstrates for the first time that kinetic variations play a key role in the efficacy of DES systems. These findings may have significant implications for future research and development. Ongoing research activities with this reservoir-based technology involve the ultra-thin cobalt-chromium stent in place of stainless steel and long-release paclitaxel formulations. Evaluation of other compounds for indications other than restenosis is also underway. Future studies with respect to restenosis will be larger and include longer-term follow-up, randomized cohorts, and non-inferiority trials with other DES.
|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. W. Krucoff, D. J. Kereiakes, J. L. Petersen, R. Mehran, V. Hasselblad, A. J. Lansky, P. J. Fitzgerald, J. Garg, M. A. Turco, C. A. Simonton III, et al. A novel bioresorbable polymer paclitaxel-eluting stent for the treatment of single and multivessel coronary disease: primary results of the COSTAR (Cobalt Chromium Stent With Antiproliferative for Restenosis) II study. J. Am. Coll. Cardiol., April 22, 2008; 51(16): 1543 - 1552. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Wessely, A. Kastrati, J. Mehilli, A. Dibra, J. Pache, and A. Schomig Randomized trial of rapamycin- and paclitaxel-eluting stents with identical biodegradable polymeric coating and design Eur. Heart J., November 2, 2007; 28(22): 2720 - 2725. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Windecker and B. Meier Late Coronary Stent Thrombosis Circulation, October 23, 2007; 116(17): 1952 - 1965. [Full Text] [PDF] |
||||
![]() |
J. Daemen and P. W. Serruys Drug-Eluting Stent Update 2007: Part I: A Survey of Current and Future Generation Drug-Eluting Stents: Meaningful Advances or More of the Same? Circulation, July 17, 2007; 116(3): 316 - 328. [Full Text] [PDF] |
||||
![]() |
A H Gershlick and G Richardson Drug eluting stents BMJ, December 16, 2006; 333(7581): 1233 - 1234. [Full Text] [PDF] |
||||
![]() |
J. S. Forrester, M. S. Lee, N. Kapoor, and R. R. Makkar The Janus Face of Drug-Eluting Stents J. Am. Coll. Cardiol., July 18, 2006; 48(2): 375 - 376. [Full Text] [PDF] |
||||
![]() |
A. Colombo and S. J. Corbett Drug-Eluting Stent Thrombosis: Increasingly Recognized But Too Frequently Overemphasized J. Am. Coll. Cardiol., July 4, 2006; 48(1): 203 - 205. [Full Text] [PDF] |
||||
![]() |
P. W. Serruys Fourth Annual American College of Cardiology International Lecture: A Journey in the Interventional Field J. Am. Coll. Cardiol., May 2, 2006; 47(9): 1754 - 1768. [Full Text] [PDF] |
||||
![]() |
R. Wessely, A. Schomig, and A. Kastrati Sirolimus and Paclitaxel on Polymer-Based Drug-Eluting Stents: Similar But Different J. Am. Coll. Cardiol., February 21, 2006; 47(4): 708 - 714. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |