J Am Coll Cardiol, 2005; 46:2258-2263, doi:10.1016/j.jacc.2005.04.067 © 2005 by the American College of Cardiology Foundation |




,*
* Cardiac Arrhythmia Center, University of Minnesota Medical School, Minneapolis, Minnesota
Central Minnesota Heart Center, St. Cloud Hospital, St. Cloud, Minnesota.
Manuscript received March 13, 2005; revised manuscript received April 2, 2005, accepted April 13, 2005.
* Reprint requests and correspondence: Dr. David G. Benditt, MMC 508, University of Minnesota Hospital-Fairview, 420 Delaware Street SE, Minneapolis, Minnesota 55455. (Email: bendi001{at}umn.edu).
| Abstract |
|---|
|
|
|---|
BACKGROUND: Cardiac resynchronization therapy is an effective addition to conventional treatment of HF in many patients with left ventricular systolic dysfunction. However, whether CRT-induced improvements in HF status also reduce susceptibility to life-threatening arrhythmias is less certain.
METHODS: Clinical and ICD electrogram data were evaluated in 18 consecutive ICD patients who underwent an upgrade to CRT-ICD. Pharmacologic HF therapy was not altered during follow-up. The definition of ventricular tachycardia (VT) and ventricular fibrillation (VF) for each patient was as determined by device programming. Statistical comparisons used paired t tests.
RESULTS: Findings were recorded during two time periods: 47 ± 21 months (range 24 to 70 months) before and 14 ± 2 months (range 9 to 18 months) after CRT upgrade. At time of upgrade, patient age was 69 ± 11 years and ejection fraction was 21 ± 8%. Before CRT the frequency of VT, VF, and appropriate ICD shocks was 0.31 ± 1.23, 0.047 ± 0.083, and 0.048 ± 0.085 episodes/month/patient, respectively. After CRT-ICD, VT and VF arrhythmia burdens and frequency of shocks were respectively 0.13 ± 0.56, 0.001 ± 0.004, and 0.003 ± 0.016 episodes/month/patient (p = 0.59, 0.03, and 0.05 vs. pre-CRT).
CONCLUSIONS: Arrhythmia frequency and number of appropriate ICD treatments were reduced after upgrade to CRT-ICD for HF treatment. Thus, apart from hemodynamic benefits, CRT may also ameliorate ventricular tachyarrhythmia susceptibility in HF patients.
| |||||||||||||
| Methods |
|---|
|
|
|---|
Demographic features, LVEF, pharmacological treatment, and HF functional class (New York Heart Association [NYHA]) are summarized in Table 1. The primary end point during follow-up was ventricular tachyarrhythmia burden (see definitions in the following text).
|
2-fold voltage safety factor in each chamber and a sufficiently rapid base rate to assure biventricular pacing for as much time as possible. Follow-up procedures. At a minimum, all ICD patients included in this study were seen in the clinic and examined at every three months after implantation. Individual patients were also seen and ICD interrogation undertaken at intervening times as circumstances dictated. At each visit, clinical status was documented, pharmacological therapy was recorded, and the ICD was interrogated. All ICD generators permitted full disclosure of arrhythmia recurrence date, duration, and cycle length, as well as the nature and effectiveness of delivered therapies. As appropriate, intracardiac electrograms were reviewed, and to the extent possible, the appropriateness of ICD shocks and/or antitachycardia pacing (ATP) applications were determined.
None of the patients included in this study were deemed "pacemaker-dependent." Thus, before initiation of CRT therapy, none of the patients were exposed to prolonged periods of right ventricular apex pacing. The percent of cardiac cycles that were ventricular paced based on findings obtained at the most recent device interrogation before "upgrade" ranged from 0% to 15%. Subsequent to initiation of CRT, device follow-up confirmed biventricular pacing >85% (range 85% to 100%) of the time in all cases. However, technological limitations preclude being able to confirm that biventricular capture was consistently achieved at all points in time.
Definitions.
Arrhythmia burden
The total number of episodes and total duration of an arrhythmia in a given time frame, presented on a per month basis.
Ventricular Tachycardia (VT)
Ventricular tachyarrhythmia with cycle length <400 ms but
320 ms leading to ICD therapy.
Ventricular Fibrillation (VF)
Ventricular tachyarrhythmia with cycle length
320 ms leading to ICD therapy.
Statistical analysis. Comparison of arrhythmia frequency and characteristics before and after initiation of CRT therapy utilized the two-tail paired t test. A p value of <0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
Patients had been followed for 47 ± 21 months (range 24 to 70 months) before CRT upgrade, and were thereafter followed prospectively for an additional 14 ± 2 months (range 9 to 18 months). Presenting arrhythmias were VT in 55%, VF in 28%, and nonsustained VT in 17%. During the baseline period, 13 of 18 (72%) patients had at least one VT and/or VF event. After CRT-ICD placement, only 2 of 18 (11%) had tachyarrhythmic events. The total number of events before upgrade was 32 ± 80.5 compared with 0.7 ± 2.2 after the upgrade (p = 0.01). Findings were essentially unchanged when re-examined using comparable durations of pre- and post-CRT-ICD follow-up in each patient. No statistically significant correlation between event rate and change in LVEF was detected.
Before CRT the frequency of VT and VF was 0.31 ± 1.23 and 0.047 ± 0.083 episodes per patient per month, respectively (Table 2). After CRT the frequency of VT and VF was 0.13 ± 0.56 and 0.001 ± 0.004 per patient per month, respectively (p = 0.01) (Table 2). The decrease in event rates after CRT for VT and VF was 0.18 (p = 0.59) and 0.046 (p = 0.03) episodes per patient per month, respectively.
|
There were two deaths recorded during follow-up. One in-hospital death occurred at 18 months after upgrade to CRT-ICD and was due to progressive intractable HF. This patient had VT episodes after CRT-ICD, but at a decreased frequency than previously. The second death was unwitnessed, but occurred in an otherwise apparently stable patient at approximately 16 months after CRT-ICD upgrade. The ICD was not interrogated. This death was deemed to be a sudden death, and occurred in a patient who had not exhibited any arrhythmias during the CRT-ICD phase of follow-up.
| Discussion |
|---|
|
|
|---|
Impact of CRT therapy on ventricular arrhythmias. Recent clinical trials have provided ample evidence supporting the effectiveness of CRT therapy in the treatment of HF patients with poor left ventricular systolic function (111). The most important positive findings have been diminished frequency of hospitalization, enhanced exercise capacity, and improved quality of life. However, the impact of CRT on arrhythmia susceptibility and mortality has been less clear, with multiple studies providing differing outcomes (9,1115,19,20).
The CONTAK-CD (9) and Multicenter InSync Randomized Clinical Evaluation-ICD (MIRACLE-ICD) (10) trials directly addressed the question of whether CRT-ICD therapy offers additional antiarrhythmic benefit not available with conventional ICDs. CONTAK-CD (9) compared CRT-ICD therapy to conventional ICD treatment. A parallel two-arm design (i.e., CRT-ICD vs. conventional ICD) was employed. Ultimately, 490 HF patients (NYHA functional class II to IV) were enrolled. Apart from HF, these individuals exhibited intraventricular conduction delays and sufficiently severe ventricular arrhythmias to warrant ICD therapy. All patients received a CRT-capable device; in 245 patients the CRT feature was disabled, whereas in the remaining patients (n = 245) both features were enabled. After six months, the frequency of observed arrhythmias did not differ in the two treatment groups; 15% of CRT patients received appropriate ICD shocks compared with 16% of no-CRT patients. Further, the distribution of arrhythmias was not substantially different for CRT-ICD versus no-CRT-ICD (VT: 10% vs. 11%, VF: 3% vs. 2%, both VT and VF: 2% vs. 2%). Thus, an antiarrhythmic benefit for CRT was not demonstrable. Similarly, MIRACLE-ICD (10) enrolled 369 patients (182 randomized to CRT off, and 187 to ICD-CRT on) with LVEF
35% and abnormally prolonged QRS duration (
130 ms). All enrollees were considered to be NYHA functional class III (n = 328) or class IV (n = 41) despite best available pharmacological treatment. The primary end points were changes between baseline and six months in quality of life, functional class, and 6-min walk distance. However, survival, incidence of ventricular arrhythmias, and rates of hospitalization were also compared. At six months, there was no evident antiarrhythmic impact of CRT. In terms of "intention-to-treat" numbers, 26% of control patients had appropriate ICD shocks versus 24% in CRT-ICD patients (p = 0.76). Similarly, the frequency of ATP did not differ between the two groups (31% of controls vs. 33% of CRT-ICD patients, p = 0.89).
The absence of CRT-associated antiarrhythmic benefit in CONTAK-CD and MIRACLE-ICD trials is in contrast to our findings. In part, this difference may be due to the fact that the CONTAK-CD and MIRACLE-ICD studies examined treatment effect in parallel patient populations rather than within the same individuals. In this regard, important findings previously reported by Higgins et al. (13) from the Ventak-CD trial tend to support this view. The latter report restricted its observations to two periods of three months each in patients subjected to biventricular ICD therapy versus ICD treatment without pacing. Among 32 patients enrolled in the Ventak CHF study, ventricular tachyarrhythmias were recorded in 16% of patients during the CRT phase versus 34% in the no-CRT portion of the follow-up (p = 0.035).
With regard to mortality, both the Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) trial (11) and the Cardiac Resynchronization Heart Failure (CARE-HF) trial (20) were sufficiently powered to address this end point. These studies differed in that COMPANION combined both CRT and ICD, whereas CARE-HF examined CRT effect alone. Nevertheless, both tended to support a mortality benefit; COMPANION reported a mortality reduction that approached conventionally accepted statistical significance (11) and CARE-HF (20) reported a clear-cut statistically significant reduction of both the primary combined end point (i.e., mortality and major cardiovascular events) as well as mortality alone (a secondary end point). Over an approximate 29 months of average follow-up, CRT was associated with a 20% mortality, compared to 30% mortality in controls (95% confidence interval 0.48 to 0.85, p < 0.002).
Antiarrhythmic and proarrhythmic balance in CRT. The potential for CRT to offer a beneficial antiarrhythmic effect was anticipated by virtue of a theoretically diminished risk of myocardial ischemia due to improved cardiac output with reduced wall stress, and a more advantageous neurohumoral impact on diseased myocardium (1922). On the other hand, the potential for an epicardial pacing-induced proarrhythmic effect to occur during CRT by altering the direction and duration of left ventricular repolarization has raised concern (23). Thus, the possibility exists that the beneficial and adverse effects of CRT on arrhythmia susceptibility may tend to counteract each other making it difficult to discern a true arrhythmia or mortality benefit. Large study populations may overcome this difficulty, as suggested by the demonstration of CRT mortality benefit in a recent meta-analysis combining multiple CRT studies (24).
In terms of a possible CRT-triggered proarrhythmic effect, early concern arose in the Multisite Stimulation in Cardiomyopathy (MUSTIC) trial (4) as a result of deaths that occurred shortly after cross-over from conventional pacing to biventricular stimulation. Ultimately, it was concluded that the deaths were attributable to specific clinical circumstances in each case. More recently, despite improved hemodynamic state in CARE-HF, the investigators did observe a 7% sudden death rate in their CRT patients. A similar trend was noted in COMPANION. However, rather than a proarrhythmic effect, it is more likely that CRT, by reducing HF deaths, may seem to tip the balance toward sudden deaths. In any case, such an outcome further favors the need for CRT-ICD combinations.
Study limitations. The analysis in this study is subject to important limitations. First, by virtue of the fact that our patient cohort included only individuals in whom upgrade to CRT was based on hemodynamic indications, we may have selected a relatively sick patient population. Such a population may have had a higher ambient level of cardiac arrhythmia than would be observed in a more typical group of CRT candidates. If true, one could argue that any CRT-ICD benefit was magnified by the study design, thereby allowing benefit to be detectable in a relatively small cohort. On the other hand, even if the benefit were small and only discernable in this manner, one would not expect the direction of this treatment effect to have been altered by virtue of being studied in a very sick population. A demonstrable benefit or, at worst, a neutral effect, would be a reasonable expectation in less sick patients. Second, we have inferred from the findings that the CRT feature was the principal factor reducing arrhythmia susceptibility. However, CRT therapy is inherently accompanied by "overdrive" pacing in order to maintain control over the ventricular activation sequence. Although unlikely, it is possible that pacing alone could have been primarily responsible for diminishing arrhythmia susceptibility in our cohort. Third, the patient population, although consecutive, was not randomized and the pre-upgrade arrhythmia status was determined retrospectively. While a prospective approach is generally conceded to be superior, this aspect of the study design eliminated inclusion of patients in whom upgrade to CRT was motivated principally by a desire to try to reduce arrhythmia burden. Fourth, the study population was small, and the two deaths during follow-up might be considered to represent a mortality concern. On the other hand, the crude mortality rate in this study was only approximately 7.4% per year, which is reasonable given the nature of the HF population. Fifth, we were unable to demonstrate a correlation between reduction of arrhythmia event rates and positive change in LVEF. Conceivably, more sensitive measures of altered hemodynamic status, such as changes in circulating catecholamines or brain natriuretic peptide levels, might be more effective in evaluating such a relationship. Finally, the small study population leaves open concern that the detected apparent CRT benefit was overestimated by chance. Consequently, given this last concern, as well as the previously stated limitations, the findings reported here must be interpreted with caution, and perhaps are best used as the basis for designing a prospective randomized trial.
Conclusions. This study examined ventricular arrhythmia burden and ICD treatment frequency in patients in whom HF treatment dictated the need for replacing a pre-existing conventional ICD system with a CRT-ICD. The availability in each of these individuals of a full-featured ICD, both before and after introduction of CRT, along with absence of substantial alterations of drug therapy, permitted detailed assessment of the impact of CRT on arrhythmia susceptibility in these patients. The findings suggest that, in the setting of diminished left ventricular systolic function and HF, CRT does diminish both tachyarrhythmia susceptibility and the frequency of either ICD shock or ATP. Thus, while potential mortality benefits cannot be addressed in this study, CRT appears to reduce tachyarrhythmia risk and also the need for ICD treatment intervention in these high-risk patients.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. G. Benditt, D. Yannopoulos, K. G. Lurie, S. Sakaguchi, L. Van Heel, S. Milstein, and C. Ermis Reply. J. Am. Coll. Cardiol., February 12, 2008; 51(6): 677 - 677. [Full Text] [PDF] |
||||
![]() |
G Lin, R F Rea, S C Hammill, D L Hayes, and P A Brady Effect of cardiac resynchronisation therapy on occurrence of ventricular arrhythmia in patients with implantable cardioverter defibrillators undergoing upgrade to cardiac resynchronisation therapy devices Heart, February 1, 2008; 94(2): 186 - 190. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Leclercq Upgrading from right ventricular pacing to biventricular pacing in pacemaker patients with chronic heart failure Heart, January 1, 2008; 94(1): 102 - 107. [Full Text] [PDF] |
||||
![]() |
F. A McAlister Device therapy in heart failure BMJ, November 3, 2007; 335(7626): 895 - 896. [Full Text] [PDF] |
||||
![]() |
D. Yannopoulos, K. G. Lurie, S. Sakaguchi, S. Milstein, C. Ermis, L. VanHeel, and D. G. Benditt Reduced Atrial Tachyarrhythmia Susceptibility After Upgrade of Conventional Implanted Pulse Generator to Cardiac Resynchronization Therapy in Patients With Heart Failure J. Am. Coll. Cardiol., September 25, 2007; 50(13): 1246 - 1251. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. A. McAlister, J. Ezekowitz, N. Hooton, B. Vandermeer, C. Spooner, D. M. Dryden, R. L. Page, M. A. Hlatky, and B. H. Rowe Cardiac Resynchronization Therapy for Patients With Left Ventricular Systolic Dysfunction: A Systematic Review JAMA, June 13, 2007; 297(22): 2502 - 2514. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Burkhardt and B. L. Wilkoff Interventional Electrophysiology and Cardiac Resynchronization Therapy: Delivering Electrical Therapies for Heart Failure Circulation, April 24, 2007; 115(16): 2208 - 2220. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. K. Kantharia, J. A. Patel, B. S. Nagra, and G. S. Ledley Electrical storm of monomorphic ventricular tachycardia after a cardiac-resynchronization-therapy-defibrillator upgrade Europace, August 1, 2006; 8(8): 625 - 628. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Ermis and D. G. Benditt Cardiac resynchronization pacing without defibrillator capability: is this a viable option? Europace, July 1, 2006; 8(7): 499 - 501. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||