| Abstract |
|---|
|
|
|---|
BACKGROUND: The SPWMD predicted clinical benefit with CRT in two previous studies from the same center.
METHODS: In this retrospective analysis of the CONTAK-CD trial, SPWMD was measured from the baseline echocardiogram of 79 heart failure patients (ejection fraction 22 ± 7%, QRS duration 159 ± 27 ms, 72% ischemic, 84% male) randomized to CRT and compared with six-month changes in echocardiographic and clinical parameters. Patients with a left ventricular end-systolic volume index (LVESVI) reduction of at least 15% were considered responders.
RESULTS: The feasibility and reproducibility of performing the SPWMD measurements were poor. Larger values for SPWMD did not correlate with six-month changes in left ventricular end-diastolic volume index (p = 0.26), LVESVI (p = 0.41), or left ventricular ejection fraction (p = 0.36). Responders did not have a significantly different SPWMD than non-responders (p = 0.26). The SPWMD did not correlate with measures of clinical improvement. At a threshold of SPWMD >130 ms, the test characteristics to predict reverse remodeling or a clinical response were inadequate.
CONCLUSIONS: The previous findings that SPWMD predicts reverse remodeling or clinical improvement with CRT were not reproducible in patients randomized in the CONTAK-CD trial.
| |||||||||
By simultaneously stimulating the right ventricle and the free wall of the LV, cardiac resynchronization therapy (CRT) has been shown to decrease LV volumes and dimensions, increase ejection fraction, and improve functional status in some, but not all patients with severe LV systolic dysfunction, interventricular conduction delay, and symptomatic heart failure (5,6). In previous studies, a greater degree of electrical and mechanical dyssynchrony at baseline have been associated with greater likelihood of improvement in LV structure and function with CRT (7,8).
Current inclusion criteria for CRT devices use only the QRS duration to assess dyssynchrony. Given the costs and risks associated with CRT, identifying reliable non-invasive measures of ventricular dyssynchrony that predict therapeutic benefit are needed to optimize patient selection.
In a non-randomized study of 20 CRT patients with predominantly non-ischemic cardiomyopathy, it was demonstrated that septal-to-posterior wall motion delay (SPWMD) measured by M-mode echocardiography was a useful predictor of reverse remodeling (9). More recently, a study of 60 CRT patients from the same center demonstrated that the SPWMD was useful to predict heart failure progression. In both studies, a longer SPWMD was associated with greater improvement with CRT, and a SPWMD
130 ms was found to be a useful cutoff point to predict success. Based on these results, SPWMD has been suggested as a screening method for patient selection for CRT. However, the feasibility and utility of this parameter have not been tested in a largely ischemic population in whom endocardial excursion of either the septal or posterior wall may be markedly diminished and therefore not clearly identifiable.
This study was designed to test the hypothesis that SPWMD would predict greater reverse remodeling and an improved clinical response in heart failure patients randomized to CRT in the CONTAK-CD trial.
| Methods |
|---|
|
|
|---|
35%, a QRS duration
120 ms; age
18 years, and normal sinus node function. The exclusion criteria and study design have been described in a previous publication (5). Originally, all patients received a CRT device and were randomized to CRT turned on or to CRT turned off (phase 1, n = 248). Because of regulatory concern, the study was modified from a crossover to a six-month parallel design in phase 2 (n = 333). Of the 581 patients enrolled between the two phases, 501 were successfully implanted with the investigational system (5). Only phase 2 patients randomized to CRT turned on with paired baseline and six-month echo data are included in this analysis. Echocardiographic analysis. Two-dimensional Doppler echocardiograms were performed at multiple investigational sites according to a standardized protocol at baseline, three months, and six months after randomization. For the purposes of this study, the baseline and six-month echocardiograms were analyzed. A two-dimensionally guided (M-mode) echocardiogram at a level just below the tips of the mitral leaflets, basal to the tips of the papillary muscles, was used for measurement of the SPWMD as described subsequently. Left ventricular end-diastolic volumes and LV end-systolic volumes were obtained by tracing the endocardium at the end-diastole (maximum cavity dimension) and end-systole (frame before mitral valve opening or the minimum cavity area) in the two-chamber and four-chamber views and derived according to Simpsons rule (biplane method of discs) (10). The left ventricular end-systolic volume index (LVESVI) and left ventricular end-diastolic volume index (LVEDVI) are normalized for body surface area. The LVEF was calculated as the total stroke volume divided by the end-diastolic volume. Echocardiograms were excluded when less than 80% of the endocardium could be visualized. Responders were defined as those who had greater than a 15% decrease in LVESVI (9,11). Clinical improvement measures included six-month change in peak exercise O2 consumption, 6-min walk distance, quality of life score as assessed by the Minnesota Living with Heart Failure Questionnaire, and New York Heart Association (NYHA) functional class.
The SPWMD represents the delay between the motion of the septum and the posterior wall of the LV. We adopted the method described by Pitzalis et al. (9) and measured the SPWMD on the M-mode echocardiogram obtained at baseline before randomization: the maximum displacement of the septal wall and the maximum displacement of the posterior wall were both measured from the onset of the QRS; the QRS to septal wall deflection value was then subtracted from the QRS to posterior wall deflection value to calculate the SPWMD (Fig. 1). Three measurements were made for each parameter and then averaged. These measurements were performed for each subject by two independent observers for 60% of the subjects for the primary and first observer and 47% of the subjects for the primary and second observer. In adopting this method, we recognize that the level of M-mode echocardiogram differs from the original description (9): Pitzalis obtained the M-mode at the level of the papillary muscles. In contrast, the measurements for the present study were made at a more basal level, representing the standard location for measuring LV dimensions.
|
Operators blinded to the success of resynchronization therapy for the individual patients performed the echocardiographic analysis of intraventricular dyssynchrony. Inter-observer reliability was measured with two independent observers compared with the primary observer.
Statistical analysis.
Linear comparisons between SPWMD and LVESVI, LVEDVI, and LVEF were made by calculating the Pearson product-moment correlation coefficient. Differences in categorical and continuous variables between responders and non-responders were assessed by the chi-square test or Student t test, as appropriate. In addition, six-month change in peak exercise O2 consumption, 6-min walk distance, quality of life score, and NYHA functional class were correlated with all dyssynchrony parameters. To reveal whether patients with longer baseline SPWMD measurements experienced a significantly greater improvement in LVESVI, LVEDVI, and LVEF with CRT than other patients, linear regressions with respect to these variables and their effect on treatment group were performed. To be consistent with previous studies, a SPWMD value
130 ms was used to determine the sensitivities, specificities, and positive and negative predictive values for six-month improvement in echocardiographic and clinical parameters (8,9). In determining associations between patients with and without a SPWMD
130 ms and clinical outcomes, continuous measures were calculated from a Student t test and values for categorical measures were calculated by a chi-square test.
Within-subjects standard deviation was calculated using the analysis of variance between the two independent observers and compared with the primary observer, and repeatability of measurements was calculated by multiplying the within-subjects standard deviation by 2.77 to account for 2 SD from the mean and two independent measurements of the same value (1214).
| Results |
|---|
|
|
|---|
|
Reproducibility of SPWMD measurements. There was poor inter-observer precision of SPWMD (observer 1 with the primary observer = 141 ms, n = 37; observer 2 with the primary observer = 194 ms, n = 47). For example, the within-subjects standard deviation for SPWMD between observer 1 and the primary observer was 51 ms, yielding a repeatability of 141 ms. These suggest that 95% of pairs of measurements of the same echocardiographic parameter by different observers should fall between 141 ms.
Correlations between baseline SPWMD and six-month evidence of reverse remodeling. The reduction in LVESVI and LVEDVI and the improvement in ejection fraction were determined by comparing the baseline and six-month echocardiograms. Baseline SPWMD did not correlate with six-month changes in LVESVI (0.10, p = 0.41), LVEDVI (0.14, p = 0.26), or ejection fraction (0.11, p = 0.36). Correlations remained non-significant when examining the same echocardiographic parameters in patients with and without ischemic etiologies of heart failure.
The baseline SPWMD did not significantly differ between responders (mean 77 ± 141 ms, range 340 ms to 312 ms) and non-responders (mean 59 ± 160, range 326 ms to 525 ms, p = 0.63). The lack of any relationship between baseline SPWMD and responders persisted when examining patients with ischemic etiologies of heart failure alone or non-ischemic etiologies of heart failure alone. Patients with a SPWMD
130 ms did not have significantly improved six-month changes in LVESVI, LVEDVI, or ejection fraction compared with those with SPWMD <130 ms; again, this finding was independent of heart failure etiology. In addition, the test characteristics for a SPWMD
130 ms to predict a responder were poor (Table 2).
|
130 ms were no more likely to have six-month improvement in any of the clinical parameters (Table 3). The test characteristics for a SPWMD
130 ms to predict an improvement in at least one NYHA functional class were poor (Table 2).
|
| Discussion |
|---|
|
|
|---|
Primarily, the limited feasibility of obtaining the necessary M-mode measurements in order to calculate the SPWMD suggests that this method of screening for CRT selection would likely not be applicable in clinical practice. In this cohort, nearly one-half (45%) of patients had neither a definitive septal nor posterior wall deflection visualized on M-mode echocardiography. Twenty-seven (34%) had only one of the two walls exhibiting a definite systolic excursion. The remaining 13 patients (16%) of the cohort with M-mode echocardiograms showed both definitive septal and posterior wall deflections. However, as demonstrated in Figure 2, the SPWMD measurement remained difficult even when both septal and posterior wall deflections are well visualized. The difficulty was further confounded by the thinning and akinesis of infarcted segments frequently present in many of these predominantly ischemic patients. Given the practical limitations of obtaining these measurements, the poor inter-observer reproducibility in our study was not surprising.
|
As described in the Methods section, the M-mode images in the previous studies were taken in the parasternal short-axis view at the level of the papillary muscle (8,9), whereas our M-mode images were taken in the parasternal long-axis view at the tips of the mitral valve leaflets. Although it is theoretically possible that a measurement taken a few millimeters towards the base would make such a significant difference, we doubt that it would rectify the significant difficulties in identifying a definite systolic excursion in one or both walls. If such a difference were found, this would mean that the measurement at the papillary muscle would have to be quite precise in location, potentially limiting applicability in clinical practice. In our opinion, it is more likely that defining the complex activation sequence of the LV using a single dimensional view is inherently unreliable independent of the precise location of the ultrasound beam. Moreover, because septal infarcts tend to spare the proximal septum, basal beam angulation may theoretically produce better results in ischemic patients. Nevertheless, in their most recent publication, Pitzalis et al. considered the SPWMD to have a value of 0 ms when septal akinesis was observed, potentially biasing the results of the study (8).
Our data, collected in a variety of echocardiography laboratories across the U.S. and involving a majority of ischemic cardiomyopathy patients, are more representative of clinical practice than those acquired in a single center with a majority of non-ischemic cardiomyopathy patients.
Tissue Doppler imaging (11,1517), strain rate imaging (11,17), and methods based on phase analysis (18) all show great promise in the measurement of intraventricular dyssynchrony. One promising method for detecting and measuring dyssynchrony uses an index based on the standard deviation of the time to peak systolic shortening in 12 myocardial segments: Yu et al. (17) have shown that a standard deviation of 33 ms or greater predicts response to CRT with a high degree of sensitivity and specificity . A limitation of this method is that nearly all of the published series have been performed with echocardiographic systems available from a single vendor and the generalizability of these measurements to other systems has not been proven. In contrast, most of the current echocardiographic systems are capable of recording pulsed tissue Doppler velocities, a potentially useful measurement that can be used to obtain timing differences between peak systolic velocities among different segments. Using this technique, one study demonstrated that a septal to lateral wall delay of greater than 60 ms predicted a favorable response to CRT (15). A recognized limitation of the pulsed Doppler technique is that only one segment can be recorded in a given cardiac cycle and may be influenced by timing shifts from beat to beat (16). These methods remain investigational, are time-intensive, and are not currently available in the great majority of clinical non-invasive laboratories. On the other hand, although M-mode echocardiography is clinically widely available and technically easy to perform, our findings suggest that it should not be used for patient selection for CRT.
Our study demonstrates that this parameter based on M-mode echocardiography, while attractive in its apparent simplicity, is not adequately robust to be applied in clinical practice. A major concern is that patients, especially those with ischemic cardiomyopathy, will not be offered this therapy on the basis of the previously reported studies.
Study limitations. The major limitation of this study is that the M-mode images were not targeted to prospectively collect these data. However, all M-mode imaging was performed in accordance with the American Society of Echocardiography recommendations. All centers followed a standardized protocol, and all echocardiograms were analyzed in an experienced core laboratory.
| Conclusions |
|---|
|
|
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
L. J. Anderson, C. Miyazaki, G. R. Sutherland, and J. K. Oh Patient Selection and Echocardiographic Assessment of Dyssynchrony in Cardiac Resynchronization Therapy Circulation, April 15, 2008; 117(15): 2009 - 2023. [Full Text] [PDF] |
||||
![]() |
C. Leclercq, G. B. Bleeker, C. Linde, E. Donal, J. J. Bax, M. J. Schalij, and C. Daubert Cardiac resynchronization therapy: clinical results and evolution of candidate selection Eur. Heart J. Suppl., December 1, 2007; 9(suppl_I): I94 - I106. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. B Bleeker, C.-M. Yu, P. Nihoyannopoulos, J. de Sutter, N. Van de Veire, E. R Holman, M. J Schalij, E. E van der Wall, and J. J Bax Optimal use of echocardiography in cardiac resynchronisation therapy Heart, November 1, 2007; 93(11): 1339 - 1350. [Abstract] [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, P. E. Vardas, A. Auricchio, J.-J. Blanc, J.-C. Daubert, H. Drexler, H. Ector, M. Gasparini, C. Linde, F. B. Morgado, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm Association Europace, October 1, 2007; 9(10): 959 - 998. [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, P. E. Vardas, A. Auricchio, J.-J. Blanc, J.-C. Daubert, H. Drexler, H. Ector, M. Gasparini, C. Linde, F. B. Morgado, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm Association Eur. Heart J., September 2, 2007; 28(18): 2256 - 2295. [Full Text] [PDF] |
||||
![]() |
T. Arita, G. P. Sorescu, B. T. Schuler, L. S. Schmarkey, J. D. Merlino, J. Vinten-Johansen, A. R. Leon, R. P. Martin, and D. Sorescu Speckle-tracking strain echocardiography for detecting cardiac dyssynchrony in a canine model of dyssynchrony and heart failure Am J Physiol Heart Circ Physiol, July 1, 2007; 293(1): H735 - H742. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Schmidt, J. Bromsen, C. Herholz, K. Adler, F. Neff, C. Kopf, and M. Block Evidence of left ventricular dyssynchrony resulting from right ventricular pacing in patients with severely depressed left ventricular ejection fraction Europace, January 1, 2007; 9(1): 34 - 40. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. F. Tops, M. J. Schalij, E. R. Holman, L. van Erven, E. E. van der Wall, and J. J. Bax Right Ventricular Pacing Can Induce Ventricular Dyssynchrony in Patients With Atrial Fibrillation After Atrioventricular Node Ablation J. Am. Coll. Cardiol., October 17, 2006; 48(8): 1642 - 1648. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. V. Pitzalis, M. Iacoviello, R. Romito, G. Luzzi, M. Anaclerio, and C. Forleo Role of Septal to Posterior Wall Motion Delay in Cardiac Resynchronization Therapy J. Am. Coll. Cardiol., August 1, 2006; 48(3): 596 - 597. [Full Text] [PDF] |
||||
![]() |
N. M. Hawkins, M. C. Petrie, M. R. MacDonald, K. J. Hogg, and J. J.V. McMurray Selecting patients for cardiac resynchronization therapy: electrical or mechanical dyssynchrony? Eur. Heart J., June 1, 2006; 27(11): 1270 - 1281. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. J. Bax, T. Abraham, S. S. Barold, O. A. Breithardt, J. W.H. Fung, S. Garrigue, J. Gorcsan III, D. L. Hayes, D. A. Kass, J. Knuuti, et al. Cardiac Resynchronization Therapy: Part 1--Issues Before Device Implantation J. Am. Coll. Cardiol., December 20, 2005; 46(12): 2153 - 2167. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||