Please click here to obtain permission to reproduce this image.
Click on image to view larger version.

Figure 9 Combined receiver operator curve for clot strength (Thrombelastograph [TEG] MA), reaction time (TEG R), and 20 µM adenosine-diphosphate-induced posttreatment platelet aggregation (light transmittance aggregometry [LTA]). High TEG MA has 74% sensitivity and 89% specificity; low TEG R has 42% sensitivity and 79% specificity; high LTA has 37% sensitivity and 79% specificity.
|