J Am Coll Cardiol, 2005; 45:1201-1205, doi:10.1016/j.jacc.2004.10.078 © 2005 by the American College of Cardiology Foundation |












,


* Washington Hospital Center, Washington, DC
Boston Scientific Corp., Natick, Massachusetts
Mid Carolina Cardiology, Charlotte, North Carolina
St. Vincents Hospital, Indianapolis, Indiana
|| Elyria Memorial Hospital, Elyria, Ohio
¶ WakeMed, Raleigh, North Carolina
# Washington Adventist Hospital, Tacoma Park, Maryland
** St. Josephs Hospital, Syracuse, New York

Sacred Heart Medical Center, Eugene, Oregon

Florida Hospital, Orlando, Florida

Wake Forest University Baptist Medical Center, Winston-Salem, North Carolina
|||| New York Presbyterian Hospital, New York, New York
¶¶ Huntsville Hospital, Huntsville, Alabama
## Abbott Northwestern Hospital, Minneapolis, Minnesota
*** Cleveland Clinic Foundation, Cleveland Ohio


College of Physicians and Surgeons, Columbia University, New York, New York


Cardiovascular Research Foundation, New York, New York.
Manuscript received August 5, 2004; revised manuscript received September 26, 2004, accepted October 4, 2004.
* Reprint requests and correspondence: Dr. Neil J. Weissman, 110 Irving Street, NW, Suite EB-5123, Washington, DC 20010. (Email: Neil.J.Weissman{at}medstar.net).
| Abstract |
|---|
|
|
|---|
BACKGROUND: The TAXUS-IV trial demonstrated that the slow-release, polymer-based, paclitaxel-eluting stent reduces angiographic restenosis and the need for repeat revascularization procedures. Serial IVUS studies reveal details of the pattern of vascular responses provoked by stent implantation that provide insight into device safety and efficacy.
METHODS: In the TAXUS-IV trial, patients were randomized to the slow-release, polymer-based, paclitaxel-eluting TAXUS stent or a bare-metal EXPRESS stent (Boston Scientific Corp., Natick, Massachusetts). As part of a formal substudy, complete volumetric IVUS data were available in 170 patients, including 88 TAXUS patients and 82 controls, at implantation and at nine-month follow-up.
RESULTS: No baseline differences were present in the clinical characteristics or IVUS parameters between the control and TAXUS groups. At nine-month follow-up, IVUS lumen volumes were larger in the TAXUS group (123 ± 43 mm3 vs. 104 ± 44 mm3, p = 0.005), due to a reduction in neointimal volume (18 ± 18 mm3 vs. 41 ± 23 mm3, p < 0.001). Millimeter-by-millimeter analysis within the stent demonstrated uniform suppression of neointimal growth along the entire stent length. Late lumen loss was similar at the proximal edge of the stent between the two groups, and reduced with the TAXUS stent at the distal edge (p = 0.004). Incomplete stent apposition at nine months was observed in only 3.0% of control and 4.0% of TAXUS stents (p = 0.12).
CONCLUSIONS: Polymer-based, paclitaxel-eluting TAXUS stents are effective in inhibiting neointimal tissue proliferation, and do not result in late incomplete stent apposition.
| ||||
| Methods |
|---|
|
|
|---|
The current IVUS analysis represents the results of a formal prespecified substudy of 268 patients at 28 centers, designed to recruit a representative cohort. Clinical sites were selected based on their IVUS experience, volume, and willingness to enroll all study patients. Intravascular ultrasound was performed immediately after stent implantation and at nine-month follow-up in all patients at the IVUS substudy sites until the prespecified enrollment was obtained.
IVUS imaging and analysis. Intravascular ultrasound imaging was performed after intracoronary administration of 0.1 to 0.2 mg nitroglycerin using motorized pullback (0.5 mm/s) and contemporary, commercial scanners. Images were continuously recorded throughout the stent and at least 5 mm distal and proximal to the stent. Images were recorded onto SVHS videotape or digitally onto CD or MO disc and analyzed at an independent core laboratory at the Washington Hospital Center (Washington, DC). All IVUS core lab analyses were performed blinded to treatment arm (TAXUS vs. bare-metal stent), and adequacy of the image for analysis was determined before unblinding the data.
Using computerized planimetry (Tapemeasure, Indec Inc., Mountain View, California), the reference segment external elastic membrane (EEM), stent, and lumen were measured every millimeter within the stent and 5 mm proximal and distal to each stent edge. Neointima was calculated as stent minus lumen measures. To correct for different stent lengths, because all stents were at least 16 mm in length, the proximal and distal 7 mm of each stent (14 mm total) were aligned and reported. Volumes were calculated using Simpsons rule, and baseline and follow-up images were aligned using the stent edge. Volumes were calculated only if the vascular interface was visualized every millimeter throughout the stent. Incomplete stent apposition was defined as a separation of at least one stent strut from the intimal surface of the arterial wall; incomplete apposition at nine months was considered "late-acquired" if it was not present after implantation (5). The primary prespecified end point of this analysis was the comparison of intrastent neointimal volume at follow-up between control and TAXUS stents.
Statistical analysis. Anticipating neointimal hyperplasia of 21.3 mm3 (±10.7 mm3) in the control group, 200 patients enrolled provided 80% power to detect a 4.3-mm3 absolute decrease in neointimal volume in the TAXUS group, a relative 20% reduction. Thus, 268 patients were prespecified for enrollment to allow for 25% attrition due to noncompliance with invasive follow-up procedures, IVUS technical failures, and suboptimal IVUS quality.
Data are presented as frequencies or mean ± 1 SD. Comparisons between control and TAXUS stents were performed with two-tailed, unpaired t tests for continuous parameters, paired t tests for change from post-procedure to follow-up, and Fisher exact test for categorical variables. Significance was set at alpha of 0.05.
| Results |
|---|
|
|
|---|
|
0.01) (Fig. 1). Overall vessel volumes were similar between TAXUS and control both at baseline and follow-up. However, positive remodeling assessed as the absolute increase in the EEM volume over time tended to be slightly more prominent with the TAXUS stent (7.66 ± 48.64 mm3 vs. 12.29 ± 36.05 mm3, respectively, p = 0.064).
|
|
Incomplete apposition. As seen in Figure 2, incomplete stent apposition was infrequent and comparable for TAXUS and control groups. Notably, there was only one case of late-acquired incomplete apposition with the TAXUS stent, which was focal (<1 quadrant in circumference and <1 mm in length), compared to two cases with control stents (p = 0.62). No episode of incomplete apposition documented by IVUS was associated with a major adverse cardiac event.
|
| Discussion |
|---|
|
|
|---|
The degree of neointimal hyperplasia seen in the control stent (29.4%) is similar to other bare-metal stents, which has ranged from 25% to 35% in prior reports (1,3,68). The observed reduction in neointimal net volume obstruction with the TAXUS stent to 12.2% in this study is also consistent with the findings from the TAXUS-I (14.8%) and TAXUS-II (7.8%) trials, which were completed with different investigator pools and core labs, indicating comparable efficacy in different patient and lesion types (1,7).
In contrast, sirolimus-eluting stents have been characterized in prior IVUS substudies by the near absence of measurable in-stent tissue (3,4). Yet from currently available data, the clinical efficacy (target lesion and vessel revascularization) of paclitaxel- and sirolimus-eluting stents appears roughly comparable, suggesting that accumulation of the small amount of in-stent neointima produced with the TAXUS stent is below the threshold of physiologic significance in most patients.
In the present study, the frequency of late incomplete stent apposition with the TAXUS stent was comparable to the control bare-metal stent, also consistent with prior studies with paclitaxel-eluting stents (1,7). In contrast, the incidence of late incomplete apposition with the sirolimus-eluting stent has ranged from 19% to 21% compared to 4% to 9% for bare-metal stents (3,4,6). The slight excess amount of neointimal hyperplasia with paclitaxel-eluting compared to sirolimus-eluting stents may fill in gaps present either immediately after the procedure, or emerge from positive remodeling during the follow-up period. While the occurrence of late incomplete stent apposition has not been definitively linked to adverse cardiac events (9), patients who develop major adverse events during follow-up often do not undergo routine IVUS follow-up, and optimal patient management when stent malapposition is seen is unsettled.
This study confirms and expands the results from previous studies showing no aggravation of edge stenosis ("candy wrapper" phenomenon) with the TAXUS stent. An attenuated luminal loss was present at the first millimeters of the distal edge with the TAXUS stent compared to controls, similar to that reported from the TAXUS-II trial (7). The absence of a beneficial effect at the proximal edge is at variance with the larger TAXUS-IV angiographic substudy, in which greater angiographic dimensions were present at both stent margins with the TAXUS stent. Possible explanations for the differences between angiographic and IVUS findings may be the smaller sample size and more select nature of patients in the IVUS compared to the angiographic substudy.
Finally, although nine-month vessel volumes were similar when comparing the TAXUS and control stents, a trend was present for slightly more positive remodeling from baseline to follow-up with the TAXUS stent. This pattern was also seen in the TAXUS-II trial, with even more pronounced positive remodeling with the moderate-release formulation. It is possible that paclitaxel may inhibit the inflammatory and or fibrotic responses that affect vessel remodeling. To date, no adverse consequences have been ascribed to this effect, and serial longitudinal IVUS follow-up is being performed in the TAXUS-II trial at two years to evaluate the stability and implications of this observation.
Study limitations. Serial IVUS interrogations were performed in only a small subset of the patients enrolled in the TAXUS-IV trial, which may have introduced selection bias, though IVUS substudy patients had similar baseline characteristics compared to the entire study population. It should also be recognized that IVUS follow-up tends not to be performed in severely obstructive vessels, and in patients with major adverse cardiac events; as such, follow-up IVUS represents a "best-case scenario," providing complementary data to follow-up angiography.
Conclusions. The polymer-based, slow-release, paclitaxel-eluting TAXUS stent attenuates vascular responses that lead to restenosis without adverse edge effects or late incomplete stent apposition.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. Aoki, G. S. Mintz, N. J. Weissman, J. T. Mann, L. Cannon, J. Greenberg, E. Grube, A.R. Z. Masud, J. Koglin, L. Mandinov, et al. Chronic Arterial Responses to Overlapping Paclitaxel-Eluting Stents: Insights From Serial Intravascular Ultrasound Analyses in the TAXUS-V and -VI Trials J. Am. Coll. Cardiol. Intv., April 1, 2008; 1(2): 161 - 167. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. J. Weissman, S. G. Ellis, E. Grube, K. D. Dawkins, J. D. Greenberg, T. Mann, L. A. Cannon, P. A. Cambier, S. Fernandez, G. S. Mintz, et al. Effect of the polymer-based, paclitaxel-eluting TAXUS Express stent on vascular tissue responses: a volumetric intravascular ultrasound integrated analysis from the TAXUS IV, V, and VI trials Eur. Heart J., July 1, 2007; 28(13): 1574 - 1582. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. S. Petronio, M. De Carlo, G. Branchitta, B. Papini, N. Ciabatti, R. Gistri, B. Cortese, G. Gherarducci, and A. Barsotti Randomized Comparison of Sirolimus and Paclitaxel Drug-Eluting Stents for Long Lesions in the Left Anterior Descending Artery: An Intravascular Ultrasound Study J. Am. Coll. Cardiol., February 6, 2007; 49(5): 539 - 546. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kimmelstiel Multislice Computed Tomography After Left Main Drug-Eluting Stenting: Are We Putting the CarT Before the Horse? Circulation, August 15, 2006; 114(7): 616 - 619. [Full Text] [PDF] |
||||
![]() |
G. S. Mintz and N. J. Weissman Intravascular Ultrasound in the Drug-Eluting Stent Era J. Am. Coll. Cardiol., August 1, 2006; 48(3): 421 - 429. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Jimenez-Quevedo, M. Sabate, D. J. Angiolillo, M. A. Costa, F. Alfonso, J. A. Gomez-Hospital, R. Hernandez-Antolin, C. Banuelos, J. Goicolea, F. Fernandez-Aviles, et al. Vascular Effects of Sirolimus-Eluting Versus Bare-Metal Stents in Diabetic Patients: Three-Dimensional Ultrasound Results of the Diabetes and Sirolimus-Eluting Stent (DIABETES) Trial J. Am. Coll. Cardiol., June 6, 2006; 47(11): 2172 - 2179. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.-K. Hong, G. S. Mintz, C. W. Lee, D.-W. Park, K.-M. Park, B.-K. Lee, Y.-H. Kim, J.-M. Song, K.-H. Han, D.-H. Kang, et al. Late Stent Malapposition After Drug-Eluting Stent Implantation: An Intravascular Ultrasound Analysis With Long-Term Follow-Up Circulation, January 24, 2006; 113(3): 414 - 419. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Kaneda Intravascular Ultrasound Analysis of Polymer-Based Paclitaxel-Eluting Stents J. Am. Coll. Cardiol., January 17, 2006; 47(2): 470 - 470. [Full Text] [PDF] |
||||
![]() |
N. J. Weissman and G. W. Stone Reply J. Am. Coll. Cardiol., January 17, 2006; 47(2): 470 - 470. [Full Text] [PDF] |
||||
![]() |
D. J. Kereiakes, R. E. Kuntz, L. Mauri, and M. W. Krucoff Surrogates, substudies, and real clinical end points in trials of drug-eluting stents J. Am. Coll. Cardiol., April 19, 2005; 45(8): 1206 - 1212. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |