J Am Coll Cardiol, 2005; 45:2054-2060, doi:10.1016/j.jacc.2005.03.039 © 2005 by the American College of Cardiology Foundation |
German Heart Center Munich, Department of Cardiovascular Surgery, Munich, Germany
Manuscript received November 20, 2004; revised manuscript received December 24, 2004, accepted March 10, 2005.
* Reprint requests and correspondence: Dr. Florian Botzenhardt, Department of Cardiology, Krankenhaus Muenchen Bogenhausen, Englschalkinger Str. 77, 81925 Munich, Germany (Email: f.botzenhardt{at}gmx.de).
| Abstract |
|---|
|
|
|---|
BACKGROUND: Aortic valve replacement in patients with small aortic annulus using stented bioprostheses is often associated with unsatisfactory hemodynamic results and high incidence of patient-prosthesis mismatch.
METHODS: Between February 2000 and January 2004, 156 patients with aortic valve disease and an aortic annulus of 18 to 23 mm in diameter received the stented bovine Soprano (Sorin Biomedica Cardio, Saluggia, Italy) (n = 18), Perimount (Edwards Lifesciences, Irvine, California) (n = 52), Perimount Magna (Edwards Lifesciences) (n = 42), or the stented porcine Mosaic (Medtronic Inc., Minneapolis, Minnesota) (n = 44) bioprostheses. Intraoperatively, the surgeon measured the aortic annulus diameter by inserting a hegar dilator. Thus, postoperative hemodynamic results could be referred to the patients aortic annulus diameter instead of referring the results to the labeled valve size. This allows for objective comparisons between different valve types.
RESULTS: There was no significant difference in hemodynamic results between the different valve types in patients with an aortic annulus 18 to 20 mm. In patients with an annulus 21 to 23 mm, the Magna was significantly superior to the other investigated devices in mean pressure gradient, effective orifice area, and incidence of patient-prosthesis mismatch. There was no significant difference between the complete supra-annular bioprostheses Mosaic and Soprano and the intra-supra-annular Perimount valve.
CONCLUSIONS: In patients with an aortic annulus of 18 to 20 mm in diameter, hemodynamic performance is independent of the implanted stented valve type and the annular position. Root enlargement or stentless valves may be beneficial alternatives. Patients with annulus diameter 21 to 23 mm benefit from the Magna in complete supra-annular position leading to superior hemodynamic results.
| |||||||
Aortic root enlargement and, consecutively, the implantation of a larger valve or the implantation of stentless bioprostheses are strategies to achieve an optimal flow orifice and to minimize the transvalvular gradient in small aortic roots. Another concept is the construction of stented prostheses, which are suitable for a complete supra-annular position, so that the stent material is totally placed on top of the annulus and no stent material impairs the blood stream. This technique was realized in the Mosaic (Medtronic Inc., Minneapolis, Minnesota) bioprosthesis, followed by the Perimount Magna (Edwards Lifesciences, Irvine, California), and the Sorin Soprano (Sorin Biomedica Cardio, Saluggia, Italy) bioprostheses. The aim of the present study was to evaluate the hemodynamic performance of these bioprostheses in patients with small aortic annulus and to analyze the existence of hemodynamic benefits resulting from complete supra-annular placement.
| Methods |
|---|
|
|
|---|
The Carpentier-Edwards Perimount Magna (Edwards Lifesciences) valve is a stented bovine pericardial bioprosthesis that corresponds to the standard Perimount valve in terms of tissue preparation and prosthetic material. However, the stent design is modified to allow for complete supra-annular placement. Thus, the implantation of a Magna of larger size, compared to the standard Perimount, can be considered. This prosthesis has been in clinical use since September 2002. Up to now there are no hemodynamic and clinical data available.
The Medtronic Mosaic (Medtronic Inc.) valve is a stented porcine bioprosthesis that has been in clinical use since 1994 (Europe) and 2000 (U.S.), respectively, and has proven low-pressure gradients and high freedom rates from adverse events in mid-term follow-up studies (79).
The Sorin Soprano (Sorin Biomedica Cardio) valve is a stented bovine bioprosthesis. It corresponds to the Sorin Pericarbon More, which has been in clinical use for over 15 years (1012), in terms of tissue preparation and prosthetic material, but differs in stent design. It is constructed to allow for complete supra-annular implantation, intended to improve the hemodynamic performance. It has been in clinical use since August 2003.
The geometric dimensions of these prostheses are depicted in Table 1.
|
|
|
|
x (velocity time integral [VTI]LVOT/VTIValve). Effective orifice area index (EOAI) (cm2/m2) was calculated by dividing EOA by body surface area. Peak pressure gradient (PG) was calculated by using the Bernoulli equation: [PG (mm Hg) = 4 x (VmaxValve)2]. Mean PG was provided by a software application integrated in the ultrasound machine based on Bernoulli. Aortic annulus diameter was measured intraoperatively by inserting a hegar dilator into the patients aortic annulus. Aortic annulus area was calculated by assuming that the aortic annulus approaches the circle shape [aortic annulus area (cm2) = (annulus diameter x 0.5)2 x
]. Moreover, we calculated the ratio of the prosthetic orifice area and the patients anatomic orifice area, represented by the ratio of EOA and aortic annulus area. It is called effective orifice fraction (EOF) (in %) and has been described earlier by our group (13). Statistical analysis. Data are expressed as mean values ± 1 SD. Comparisons between groups were made using the t test for independent samples in case of normal data distribution, and the Mann-Whitney U test in case of non-normal data distribution. Comparisons within one group were performed using the t test for dependent samples or the Wilcoxon matched pairs test, respectively. Normal data distribution was assessed by using Q-Q plots. Differences were considered significant with p < 0.05. Correlations were tested with the Pearson product-moment method.
| Results |
|---|
|
|
|---|
|
Patient-prosthesis mismatch.
Patient-prosthesis mismatch is rated in three grades according to the results in EOAI: no mismatch (EOAI >0.85 cm2/m2), moderate mismatch (EOAI
0.85 cm2/m2), and severe mismatch (EOAI
0.65 cm2/m2) (14). The incidence of patient-prosthesis mismatch after aortic valve replacement grouped by aortic annulus diameter is shown in Table 5.
|
In Table 6, the hemodynamic results are shown grouped by aortic annulus diameter and by the presence or absence of severe patient-prosthesis mismatch.
|
| Discussion |
|---|
|
|
|---|
The aortic root can be enlarged by a patch after the techniques described by Nicks et al. (15) or Manougian and Seybold-Epting (16) to realize the implantation of a larger prosthesis. Castro et al. (17) reported that aortic root enlargement can be performed with minimal added risk relative to standard aortic valve replacement as a preventive strategy to minimize patient-prosthesis mismatch.
Stentless prostheses implanted in the subcoronary position or as full root replacement are associated with superior hemodynamic results. However, the implantation of a stentless valve requires surgical experience to achieve morbidity and mortality rates as low as with stented prostheses.
The concept of the so-called complete supra-annular bioprostheses is to maintain the advantage of the easy and safe implantation of stented valves, but to place them on top of the aortic annulus. The aortic root orifice is naturally narrowed by the aortic annulus ring. The seating of the complete supra-annular prosthesis is upon this annulus ring. Thus, there should be no narrowing of the flow area by the prosthetic stent in addition to the narrowing by the annulus ring.
The Mosaic, Magna, and Soprano bioprostheses belong to the latter category of prostheses. The aim of this study was to evaluate their hemodynamic performance in patients with small aortic annulus. The intra-supra-annular Perimount prosthesis is considered as a reference valve, which has proven low-pressure gradients even in small valve sizes (4,18).
Hemodynamic results. To allow representative comparisons between different bioprostheses of various manufacturers, we introduced a new concept of grouping the prostheses and the obtained results. As shown in Table 1, all prostheses differ in their geometric dimensions. These differences do not allow the comparison of their hemodynamic performance based on the same valve size label. Moreover, our interest is not focused on the performance of a certain valve type in a certain size. Emphasizing our patient-centered point of view, we are more interested in the following question: what hemodynamic performance can a patient with an aortic annulus of 20 mm expect after aortic valve replacement with certain prosthesis? This point of view is also the content of the proposals of the International Organization for Standardization (ISO) concerning the valve size labeling of heart valve prostheses (ISO/CD 5840): the ISO defines "valve size" as "manufacturers designation of a valve that indicates the tissue annulus diameter of the patient into which the valve is intended to be implanted." The tissue annulus diameter is, in aortic valve replacement, the patients aortic annulus diameter.
To investigate the prosthetic performance in certain aortic annulus diameters, we started to refer our results to the aortic annulus diameter: patients with an aortic annulus of 18 to 20 mm in diameter and patients with an aortic annulus of 21 to 23 mm in diameter.
In patients with an aortic annulus of 18 to 20 mm in diameter, the implanted valve type and the implantation position (intra-supra-annular, complete supra-annular) did not influence the hemodynamic outcome after aortic valve replacement. This result may be explained by the adversarial ratio of the prosthetic stent to the small blood flow area in this patient group, causing an artificial flow obstruction, independently of the implanted valve type. These patients may benefit form aortic root enlargement and implantation of a larger valve.
In contrast, in patients with an aortic annulus of 21 to 23 mm in diameter, the Magna prosthesis is significantly superior regarding mean pressure gradient, EOA, EOAI, and EOF compared with the other devices. The improvement of the Magna can be attributed to the renewed stent design with smaller sewing ring and lower profile in the Magna compared with the standard Perimount, whereas there is no difference in the internal diameter between both prostheses. Practically, a larger Magna than a standard Perimount can be implanted in the same patient.
Patient-prosthesis mismatch.
Effective orifice area index is the decisive parameter to describe the incidence of patient-prosthesis mismatch. In case of mismatch, there can be a high residual transvalvular gradient, although the valve is absolutely normal. It has been demonstrated that, to avoid any significant gradient at rest and during exercise, EOAI of an aortic valve prosthesis should not be <0.85 cm2/m2 (1921). We rated the extent of patient-prosthesis mismatch as not present in EOAI >0.85 cm2/m2, moderate in EOAI
0.85 cm2/m2, and severe in EOAI
0.65 cm2/m2 (14). This graduation corresponds to the general concept that moderate aortic stenosis of a native valve is present with EOAI <0.90 cm2/m2 (22).
Obviously, the stent is positioned within the blood flow and causes a relative flow obstruction. The smaller the aortic annulus, the smaller the implanted prostheses, and, thus, the ratio of stent to effectively used blood flow area worsens.
However, the incidence of patient-prosthesis mismatch of various valve types should not be compared based on the labeled valve size. This method would not consider the benefit of the complete supra-annular implantation technique: the implantation of a larger valve compared with conventional prostheses in the same patient. Therefore, we referred the incidence of patient-prosthesis mismatch to the aortic annulus diameter as an invariant variable.
Our data show that there is no significant difference in the incidence of patient-prosthesis mismatch among the various prostheses in patients with an aortic annulus of 18 to 20 mm in diameter. Regarding the patients with an aortic annulus of 21 to 23 mm, there is a clear superiority of the Magna prosthesis concerning the incidence of moderate and severe mismatch, when compared with the Perimount, Mosaic, and Soprano. This result shows that the complete supra-annular position alone is not decisive for a low incidence of patient-prosthesis mismatch. It is the combination of the complete supra-annular position and a stent design that is characterized by a narrow sewing ring and a low profile, realized in the Magna prosthesis.
When analyzing the effect of patient-prosthesis mismatch on the hemodynamic results, the transprosthetic pressure gradient is expected to increase with decreasing EOAI. In our series, mean pressure gradient in patients with EOAI
0.65 cm2/m2 was 16.7 ± 4.6 mm Hg vs. 10.8 ± 5.0 mm Hg (p < 0.001) in patients with EOAI >0.65 cm2/m2. The consequence of elevated pressure gradients is a reduced or missing left ventricular mass regression after aortic valve replacement (23,24), which may negatively influence late survival, systolic and diastolic cardiac function, and physical capacity (25,26). The Magna bioprosthesis significantly minimizes the risk of patient-prosthesis mismatch in patients with an aortic annulus of 21 to 23 mm. The impact of this reduced incidence of mismatch on the transvalvular pressure gradients could have been shown in this study. The effect of this benefit on left ventricular mass regression and late clinical outcome will have to be studied in long-term investigational series (27).
Patients mean age at operation date was 75.6 years (range 55 to 89 years), representing the increasing number of elderly patients undergoing aortic valve replacement. In our institution, patients from the age of 65 receive a bioprosthesis, so that anticoagulation can be limited to three postoperative months to prevent embolic events caused by prosthetic thrombogenity and episodic atrial fibrillation. Then the risk of major bleeding significantly exceeds the risk of embolic events. Especially older women often present with narrow LVOT, small aortic annulus, and heavy calcification. According to the recommendations by the American College of Cardiology, the surgical procedure in this patient group, according to our results properly requiring root enlargement, must consider the balance between the potential for improved symptoms and survival and the morbidity and mortality of surgery (28). In our institution, we do not always perform aortic root enlargement in case of patient-prosthesis mismatch in this patient group, setting priority to achieve low rates of perioperative adverse events. However, we sometimes have to accept suboptimal hemodynamic performance.
Conclusions. After aortic valve replacement in patients with an aortic annulus of 18 to 20 mm in diameter, there is no statistical difference in the hemodynamic outcome between intra-supra-annular and complete supra-annular stented bioprostheses, and there is no statistical difference among the various complete supra-annular valves. All stented prostheses in this patient group are associated with high incidence of patient-prosthesis mismatch. Hence, there is no benefit of the complete supra-annular prostheses in these patients. Therefore, we recommend for patients with an aortic annulus up to 20 mm the use of aortic root enlargement and, consecutively, the implantation of a larger stented bioprostheses.
Patients with an aortic annulus of 21 to 23 mm in diameter benefit from the implantation of the complete supra-annular Magna bioprosthesis, resulting in a significantly superior hemodynamic performance and a minimized risk of patient-prosthesis mismatch. This benefit could not be demonstrated for the other complete supra-annular valves, Mosaic and Soprano. There was no significant difference between those valves and the intra-supra-annular Perimount prosthesis.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. Nozohoor, J. Nilsson, C. Luhrs, A. Roijer, and J. Sjogren Influence of Prosthesis-Patient Mismatch on Diastolic Heart Failure After Aortic Valve Replacement Ann. Thorac. Surg., April 1, 2008; 85(4): 1310 - 1317. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Eichinger, S. Dabritz, and R. Lange Update of the European standards for inactive surgical implants in the area of heart valve prostheses Eur. J. Cardiothorac. Surg., November 1, 2007; 32(5): 690 - 695. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Bakhtiary, O. Dzemali, U. Steinseiffer, C. Schmitz, B. Glasmacher, A. Moritz, and P. Kleine Opening and closing kinematics of fresh and calcified aortic valve prostheses: An in vitro study J. Thorac. Cardiovasc. Surg., September 1, 2007; 134(3): 657 - 662. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Bottio and G. Gerosa Conflicting Echocardiographic Evaluations of Supra-Annular Tissue Valves: A Mystery Not Yet Resolved Ann. Thorac. Surg., August 1, 2007; 84(2): 713 - 714. [Full Text] [PDF] |
||||
![]() |
R. Garcia Fuster, V. Estevez, I. Rodriguez, S. Canovas, O. Gil, F. Hornero, and J. Martinez-Leon Prosthesis patient mismatch with latest generation supra-annular prostheses. The beginning of the end? Interactive CardioVascular and Thoracic Surgery, August 1, 2007; 6(4): 462 - 469. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Borger, A. F. Nette, M. Maganti, and C. M. Feindel Carpentier-Edwards Perimount Magna Valve Versus Medtronic Hancock II: A Matched Hemodynamic Comparison Ann. Thorac. Surg., June 1, 2007; 83(6): 2054 - 2058. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. M. Wagner, W. B. Eichinger, S. Bleiziffer, F. Botzenhardt, I. Gebauer, R. Guenzinger, R. Bauernschmitt, and R. Lange Influence of completely supra-annular placement of bioprostheses on exercise hemodynamics in patients with a small aortic annulus J. Thorac. Cardiovasc. Surg., May 1, 2007; 133(5): 1234 - 1241. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Bleiziffer, W. B Eichinger, I. Hettich, R. Guenzinger, D. Ruzicka, R. Bauernschmitt, and R. Lange Prediction of valve prosthesis-patient mismatch prior to aortic valve replacement: which is the best method? Heart, May 1, 2007; 93(5): 615 - 620. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Totaro and V. Argano Improved prosthesis-annulus interaction and postoperative hemodynamic performance of new-generation aortic bioprosthesis J. Thorac. Cardiovasc. Surg., March 1, 2007; 133(3): 846 - 847. [Full Text] [PDF] |
||||
![]() |
G. Gerosa, V. Tarzia, G. Rizzoli, and T. Bottio Reply to the Editor J. Thorac. Cardiovasc. Surg., March 1, 2007; 133(3): 847 - 849. [Full Text] [PDF] |
||||
![]() |
D. Pavoni, L. P. Badano, S. F. Musumeci, R. Frassani, P. Gianfagna, E. Mazzaro, and U. Livi Results of Aortic Valve Replacement With a New Supra-Annular Pericardial Stented Bioprosthesis Ann. Thorac. Surg., December 1, 2006; 82(6): 2133 - 2138. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. B. Eichinger, I. M. Hettich, and R. Lange Small aortic annulus: The hydrodynamic performances of 5 commercially available tissue valves J. Thorac. Cardiovasc. Surg., December 1, 2006; 132(6): 1499 - 1500. [Full Text] [PDF] |
||||
![]() |
G. Gerosa, V. Tarzia, G. Rizzoli, and T. Bottio Reply to the Editor J. Thorac. Cardiovasc. Surg., December 1, 2006; 132(6): 1500 - 1501. [Full Text] [PDF] |
||||
![]() |
P Pibarot and J G Dumesnil Prosthesis-patient mismatch: definition, clinical impact, and prevention Heart, August 1, 2006; 92(8): 1022 - 1029. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Dumesnil and P. Pibarot Prosthesis-patient mismatch and clinical outcomes: The evidence continues to accumulate J. Thorac. Cardiovasc. Surg., May 1, 2006; 131(5): 952 - 955. [Full Text] [PDF] |
||||
![]() |
G. Gerosa, V. Tarzia, G. Rizzoli, and T. Bottio Small aortic annulus: The hydrodynamic performances of 5 commercially available tissue valves J. Thorac. Cardiovasc. Surg., May 1, 2006; 131(5): 1058 - 1064. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. H. Cohn Comparison of Bioprostheses in Patients With Small Aortic Annulus J. Am. Coll. Cardiol., March 7, 2006; 47(5): 1088 - 1088. [Full Text] [PDF] |
||||
![]() |
F. Botzenhardt, W. B. Eichinger, S. Bleiziffer, R. Guenzinger, I. M. Wagner, R. Bauernschmitt, and R. Lange Reply J. Am. Coll. Cardiol., March 7, 2006; 47(5): 1088 - 1088. [Full Text] [PDF] |
||||
![]() |
S. H. Rahimtoola The Year in Valvular Heart Disease J. Am. Coll. Cardiol., January 17, 2006; 47(2): 427 - 439. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |