J Am Coll Cardiol, 2004; 44:1036-1046, doi:10.1016/j.jacc.2004.05.056
© 2004 by the American College of Cardiology Foundation
This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Tsutsui, J. M.
Right arrow Articles by Porter, T. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tsutsui, J. M.
Right arrow Articles by Porter, T. R.

ECHOCARDIOGRAPHY

Detection of retained microbubbles in carotid arteries with real-time low mechanical index imaging in the setting of endothelial dysfunction

Jeane M. Tsutsui, MD*, Feng Xie, MD*, Martin Cano, BS{dagger}, James Chomas, PhD{ddagger}, Patrick Phillips, PhD{ddagger}, Stanley J. Radio, MD{dagger}, John Lof, MS* and Thomas R. Porter, MD, FACC*,*

* Section of Cardiology, University of Nebraska Medical Center, Omaha, Nebraska, USA
{dagger} Department of Pathology, University of Nebraska Medical Center, Omaha, Nebraska, USA
{ddagger} Siemens Medical Solutions, Mountain View, California, USA

Manuscript received March 30, 2004; revised manuscript received May 13, 2004, accepted May 18, 2004.

* Reprint requests and correspondence: Dr. Thomas R. Porter, University of Nebraska Medical Center, 981165 Nebraska Medical Center, Omaha, Nebraska 68198-1165 (Email: trporter{at}unmc.edu).


    Abstract
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
OBJECTIVES: We sought to determine if intravenously injected microbubbles would be retained by the carotid arteries (CAs) in the setting of endothelial dysfunction (ED) using a linear transducer equipped with a low mechanical index pulse sequence scheme (PSS).

BACKGROUND: Microbubbles normally pass freely through large and small vessels but are retained in regions with ED. New high-frequency low mechanical index PSS can potentially be utilized to image these retained microbubbles.

METHODS: Intravenous albumin- and lipid-encapsulated microbubbles were administered in seven pigs while imaging the CAs before and after a 20% intralipid infusion to induce hypertriglyceridemia. The degree of microbubble retention was quantified by measuring endothelial acoustic intensity (AI) after clearance of free-flowing microbubbles. Microbubble adherence was also evaluated after selective balloon injury of the CAs. The CA diameter responses to acetylcholine were quantified.

RESULTS: After induction of hypertriglyceridemia, adherence of albumin-encapsulated microbubbles was visually evident in all CAs, and endothelial AI increased significantly (p < 0.001 compared with baseline). The CA responses to acetylcholine went from vasodilation at baseline to vasoconstriction during hypertriglyceridemia. Endothelial AI also increased in the balloon-stretched vessels (p < 0.01 compared with uninjured vessels) after albumin-encapsulated microbubble injection, with a ring of microbubbles selectively adhering to the injured segment. This retention was not observed with lipid-encapsulated microbubbles. Scanning electron microscopy confirmed that albumin-coated microbubbles adhered to endothelial cells.

CONCLUSIONS: Retention of intravenously injected albumin microbubbles occurs in the setting of both global and regional ED in large vessels and can be noninvasively imaged with high-frequency low mechanical index PSS.

Abbreviations and Acronyms
  AI = acoustic intensity
  CA = carotid artery
  ED = endothelial dysfunction
  PESDA = perfluorocarbon-exposed sonicated dextrose albumin
  PSS = pulse sequence scheme
  ROI = region of interest
  SEM = scanning electron microscopy


Endothelial dysfunction (ED) is considered one of the earliest events in the process of atherosclerosis (1) and is associated with most cardiovascular risk factors, including diabetes mellitus, hyperlipidemia, smoking, and hypertension (2–4). Endothelial dysfunction also appears to be the first identifiable functional abnormality observed with a high-fat diet, balloon- or stent-induced arterial injury, reperfusion injury, or transplant rejection (5–8).

Because ED is an early reversible risk factor in the development of atherosclerosis, its identification is an important clinical end point. Current methods for detecting ED are the evaluation of endothelium-dependent, flow-mediated dilation, or response to an endothelium-dependent vasodilator infusion, usually assessed by ultrasound or plethysmography of the brachial artery (5,6). It is known that in patients with coronary artery disease there is correlation between impaired endothelium-dependent vasodilation in the brachial artery and the presence of ED in the coronary arteries (7). However, a noninvasive direct method to visualize ED has not been developed.

Intravenously injected perfluorocarbon-filled microbubbles are currently indicated for the improvement of endocardial border resolution. These microbubbles exhibit rheologic properties similar to those of red blood cells and normally pass freely through the large and small vessels (8). However, several investigators have demonstrated that microbubbles are retained in the microvasculature when ED exists (9–12). It is unknown whether regional or global arterial ED would also result in retention of these microbubbles.

One of the major limitations for imaging a small number of microbubbles retained along the endothelial border is that they are easily destroyed by the interrogating beam. However, a new low mechanical index pulse sequence scheme (PSS) has been developed which is sensitive enough to detect nonlinear behavior from small numbers of microbubbles with minimal destruction while simultaneously canceling background linear signals from tissue (13). In this study, we utilized a high-frequency version of this PSS to determine whether endothelial retention of microbubbles occurs with regional or global arterial ED.


    Methods
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
Animal preparation.   The study was approved by the Institutional Animal Care and Use Committee of the University of Nebraska Medical Center and conformed to the Position of the American Heart Association on Research Animal Use. Nine pigs weighing 36 ± 2 kg were sedated, endotracheally intubated, and placed under general anesthesia using isoflurane. In each animal, 7-F catheters were placed in both femoral veins for administration of microbubbles and intralipid emulsion and for withdrawal of blood samples. Both femoral arteries were also cannulated for hemodynamic monitoring and for placing an 8-F guide catheter into the proximal left carotid artery (CA). All pigs were treated with 40 mg of Ketorolac intravenously to prevent pulmonary hypertensive responses to microbubbles, which occur exclusively in this species (14), and ventilated with supplemental oxygen during the microbubble injections.

Evaluation of endothelial function.   Endothelial function was evaluated by measuring the responses of the diameter of the CAs to intra-arterial infusion of the endothelium-dependent vasodilator acetylcholine. Local infusions of acetylcholine normally result in either no change in vessel diameter or mild vasodilation, whereas abnormal responses produce endothelin-mediated vasoconstriction (15). Longitudinal sectional images of the vessel were obtained with a standard 7-4 MHz linear transducer (HDI 5000, Philips Medical Systems, Bothell, Washington) at a mechanical index of 0.8. After adequate transcutaneous imaging of the CA with this transducer, the skin overlying the vessel was marked to permit monitoring of the vessel at one constant location. Also, a radiopaque marker was positioned at this site to permit its adequate identification by angiography. The CAs were scanned longitudinally, and their diameters were measured from two-dimensional images, from the anterior to the posterior inner borders of the vessels at end-diastole, coincident with the R wave on a continuously recorded electrocardiogram. Mean flow velocity was obtained with pulsed Doppler by placing the sample volume in the center of the artery. The average of three cardiac cycles was used for analysis.

After baseline CA measurements, acetylcholine was infused to achieve estimated intra-arterial concentrations of 10–7 and 10–6 mol/l, at 10-min intervals through the 8-F catheter. The flow in the CA obtained by ultrasound was used to determine the molar concentrations of acetylcholine to be infused, as follows:

{4011325.1036.SI1}
where 1.0 ml/min was the infusion rate for acetylcholine; Q = CA flow calculated by the product of CA area and mean velocity; and MAch = molar concentration of acetylcholine desired (mol/l) (16).

The response of CA diameter to acetylcholine was considered as the highest variation from baseline during the 10–7 or 10–6 mol/l infusion.

Carotid imaging of retained microbubbles.   The evaluation of both albumin- and lipid-encapsulated microbubble transit in the CAs was performed with a novel 15L8 linear array transducer equipped with a low mechanical index PSS (Contrast Pulse Sequencing; Siemens Acuson Sequoia, Mountain View, California) operating at 7 MHz. This technique transmits pulses of alternating polarity and amplitude to detect nonlinear responses from microbubbles while simultaneously canceling the linear responses from tissue (13).

Perfluorocarbon-exposed sonicated dextrose albumin (PESDA) microbubbles were injected as a bolus of 0.5 ml followed by a 10-ml saline flush into the femoral vein. The PESDA was prepared according to the method previously described (17). Briefly, a mixture of three parts 5% dextrose and one part 5% human serum albumin (total, 16 ml) was added to 8 ml decafluorobutane gas, hand-agitated, and then sonicated with an electromechanical sonicator (Heat Systems Inc., Farmingdale, New York) for 70 ± 5 s. The resulting microbubble size by hemocytometry was 4.6 ± 0.4 µm, and mean concentration measured by a Coulter counter was 1.4 x 109 bubbles/ml.

Lipid-encapsulated microbubbles (Definity; Bristol-Myers Squibb Medical Imaging Inc., North Billerica, Massachusetts) were injected as a bolus of 0.05 ml followed by a 10-ml saline flush into the femoral vein. Definity microbubbles are composed of octafluoropropane gas encapsulated in a lipid shell with a polyethylene glycol spacer on its surface that inhibits any complement bound to the microbubbles from binding to the vascular endothelium (18,19). The reported diameter of Definity is 1.1 to 3.3 µm, and the mean concentration is 1.2 x 1010 bubbles/ml. Therefore, we reduced the dose of Definity microbubbles 10-fold to match the concentration of PESDA microbubbles injected.

Carotid arteries were imaged in both the transverse and longitudinal sectional views, in the same position used for the evaluation of diameter responses to acetylcholine. The depth, power output, gain, and compression settings were initially adjusted to obtain adequate detection of the microbubbles by real-time imaging and kept constant throughout the examination. The mechanical index used was 0.3, and the frame rate was 24 Hz. Retained microbubbles were defined qualitatively as microbubbles adherent to the arterial wall which persisted in one location at the vessel luminal border late after injection. Examinations were recorded on videotape, and the images were digitally stored for subsequent analysis.

Quantification of the contrast acoustic intensity (AI) was performed by analyzing digitally acquired cine loops with quantitative contrast software (CUSQ 1.3, Siemens Acuson, Mountain View, California). Transverse sectional views of the CAs were used for this analysis. To determine the AI of microbubbles retained to the arterial endothelium, we performed a simultaneous quantification of the entire lumen AI and the central lumen AI. The endothelial AI was defined as the difference between the entire lumen AI minus the central lumen AI (Fig. 1A).



View larger version (80K):
[in this window]
[in a new window]
 
Figure 1 (A) Example of the region of interest (ROI) used for the measurements of contrast acoustic intensity (AI) of the entire lumen (white line) and central lumen (yellow line). (B) Two-dimensional (2D) imaging of a normal carotid artery in the transverse view and with pulse sequence scheme (PSS) before intravenous microbubbles (MB) injection showing no background imaging, early after MB injection showing the complete luminal vessel opacification, and late after MB injection showing few MB in the central lumen. The time-intensity curve demonstrates the typical quantification of the mean AI in both the entire lumen ROI and the central lumen ROI. The late 50 frames after clearance of free-flowing MB were chosen for the calculation of endothelial AI, defined as the difference between the mean entire lumen AI minus the central lumen AI (see text for details).

 
Quantification of the entire lumen AI (encompassing the endothelial border and central lumen) was obtained by placing a region of interest (ROI) just inside the lumen border on the transverse carotid image, thus avoiding the analysis of contrast signals coming from the vasa vasorum. A second ROI, with a diameter 20% smaller than the larger ROI, was then placed in the central lumen of the vessel (Fig. 1A). The mean AI of both ROI was measured simultaneously. For each sequence of images at baseline, post intralipid infusion and post balloon injury, 50 consecutive frames in the late period following microbubble injections after clearance of free-flowing microbubbles were analyzed (Fig. 1B). A typical bolus injection of microbubbles resulted in a 20- to 25-s period of complete vessel opacification, followed by a washout period. Therefore, the sequence of frames chosen for quantification of endothelial AI was initiated 30 s after the first detection of microbubbles in the CA. Endothelial AI was expressed as the mean difference between the entire lumen AI and the central lumen AI in the 50 analyzed frames. The results of AI were expressed in units (range 0 to 100).

Study protocol.   Phase I: induction of hypertriglyceridemia
This portion of the study involved seven pigs. After baseline microbubble injections and measurements of endothelial function, transient hypertriglyceridemia was induced by a continuous intravenous infusion of 20% intralipid (Fresenius Kabi Clayton, L.P., Clayton, North Carolina), at a rate of 0.3 ml/kg/min, for 20 min (5,20). Measurements of serum levels of triglycerides and cholesterol were performed before and after intralipid infusion. In a similar way, the serum levels of complement (C3) were measured at baseline, before and after microbubble injection, and again during hypertriglyceridemia. The CA responses to intra-arterial acetylcholine infusion, and imaging of CAs after albumin- and lipid-encapsulated microbubble injections, were repeated 30 min after intralipid infusion. Three pigs were euthanized at this phase of the study for scanning electron microscopy (SEM) verification of microbubble binding.

Serum was also obtained before and after induction of hypertriglyceridemia for flow cytometry, to determine whether hypertriglyceridemia influenced serum complement attachment to the microbubble surface. For this procedure, 0.7 x 109 PESDA microbubbles were combined with 0.5 ml of pig serum, incubated at 37°C for 5 min, and washed twice. The serum-exposed microbubbles were then combined with anti-human C3b murine monoclonal antibody (Gamma Biologicals Inc., Houston, Texas) for 20 min, washed twice, and subsequently combined with a R-phycoerythrin conjugated secondary antibody (Becton Dickinson Pharmigen, San Jose, California) for 20 min. Aliquots of 0.7 x 109 PESDA microbubbles were also incubated with immunoglobulin G control antibody (Alpha Diagnostic, San Antonio, Texas) and R-phycoerythrin conjugated secondary antibody, or only with R-phycoerythrin conjugated secondary antibody. Microbubbles were then analyzed on a FACSCalibur (Becton Dickinson, San Jose, California) to generate graphs of red fluorescent intensity.

Phase II: balloon-stretching of the CAs
This portion of the study involved six pigs. After a single intravenous bolus of heparin (100 USP U/kg), baseline carotid angiography was performed using injections of iohexol 300 (Omnipaque; Nycomed, Princeton, New Jersey), and the vessel diameter was determined with a manual caliper. A 40- to 60-mm segment of the left CA was stretched using an oversized balloon dilation catheter (Guidant, Advanced Cardiovascular Systems, Temecula, California) that measured 140% of baseline angiographic vessel diameter, as previously described (14). In one pig there was dissection and occlusion of the left CA during the procedure. Dilation of the right CA was subsequently performed and used for analysis. In the remaining five pigs, the uninjured right CA served as control.

Intravenous PESDA and Definity injections were repeated for imaging of microbubble transit in both the balloon-injured and the control CAs within 1 h of arterial injury. The CA responses to local acetylcholine infusions were also re-evaluated. At the end of the protocol, the pigs were euthanized, and sections of both CAs were analyzed by SEM. In five of these pigs, a final 0.5-ml, albumin-encapsulated microbubble injection was performed without any ultrasound imaging 15 min before the pigs were euthanized.

Pathologic analysis of the CA
The balloon-injured, hypertriglyceridemic, and control CAs were excised, washed in 0.9% saline solution, and fixed in 10% formalin for a minimum of 24 h. Histologic analysis was performed in sections of these vessels to assess the extent of injury. For SEM, the CAs were longitudinally sectioned and dehydrated through a series of ethanol concentrations increasing from 50% to 100%. The specimens were then immersed in Freon 113, critical point dried, mounted on aluminum stubs, and sputter-coated with gold (Polaron E5100, Polaron Inc., Watford, United Kingdom).

To determine the appearance of microbubbles by SEM, we also evaluated an in vitro preparation of PESDA. The microbubbles were fixed in 10% formalin, placed over a 0.22-µm Millipore filter (Millipore Corporation, Bedford, Massachusetts), and submitted to the same processing as described for the CA specimens. Scanning electron microscopy was performed using a Philips 515 system (Philips Inc., Eindhoven, the Netherlands).

Statistical analysis.   Continuous data are expressed as mean ± SD. Comparisons of endothelial AI before and after procedures were performed by a paired t test. The CA diameter responses to acetylcholine infusion were compared by a Wilcoxon signed rank test. An unpaired t test was used for the comparison of endothelial AI in the balloon-injured and control vessels. Triglyceride and cholesterol levels at baseline, immediately after intralipid infusion, and 2 h after intralipid infusion were compared by repeated measures analysis of variance. A p value <0.05 was considered statistically significant.


    Results
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
Complement binding to microbubbles.   Flow cytometry revealed complement binding to PESDA microbubbles in the presence of pig serum, which did not occur with the control antibody or with the anti-mouse secondary antibody alone (Fig. 2, curves A and B). The binding of anti-human C3b to PESDA microbubbles was increased during hypertriglyceridemia, expressed as a shift of the mean fluorescent intensity curve to the right (Fig. 2, curves C and D).



View larger version (27K):
[in this window]
[in a new window]
 
Figure 2 Example of flow cytometry showing histograms of red fluorescent intensity for perfluorocarbon-exposed sonicated dextrose albumin microbubbles incubated with R-phycoerythrin–labeled control monoclonal antibody (A), with R-phycoerythrin anti-mouse alone (B), and with serum and R-phycoerythrin–labeled monoclonal antibody to complement C3b, both at baseline (C) and after intralipid infusion (D). Note that incubation of microbubbles and anti-human C3b in the presence of serum drawn in the setting of hypertriglyceridemia resulted in a shift of the curve to the right, reflecting an increase in the mean fluorescent intensity.

 
Serum triglyceride and complement levels.   Serum triglyceride and cholesterol levels were low in all pigs at baseline. Intralipid infusion resulted in a significant increase in the serum triglyceride levels, returning to near baseline values at 2 h after infusion. No significant change was observed in the cholesterol measurements.

Complement levels were normal in all pigs at baseline and remained unchanged after microbubble injection. However, during hypertriglyceridemia there was a significant increase in the serum levels of complement C3 (Table 1).


View this table:
[in this window]
[in a new window]
 
Table 1. Lipid and Complement Levels at Baseline and Post-Intralipid Infusion
 
CA response to acetylcholine.   The CA response to the acetylcholine at baseline was vasodilation, reflecting normal endothelial function, in all but one pig in which a vasoconstrictive response was obtained. The presence of ED in the remaining pigs after intralipid infusion was confirmed by the CA response to acetylcholine, which went from vasodilation at baseline to vasoconstriction during hypertriglyceridemia (p = 0.024) (Fig. 3A). In a similar way, a vasoconstrictive response to intra-arterial infusion of acetylcholine was also observed in the balloon-injured CAs (Fig. 3B).



View larger version (15K):
[in this window]
[in a new window]
 
Figure 3 Carotid artery (CA) diameter response to intra-arterial infusion of acetylcholine, at baseline and during hypertriglyceridemia (A) and after CA balloon injury (B). Carotid response went from vasodilation at baseline to vasoconstriction after intralipid infusion and also after balloon injury, confirming the presence of endothelial dysfunction in both conditions.

 
There were no significant differences in the arterial blood pressures or heart rates between baseline, post-intralipid, and post-dilation conditions.

Carotid imaging of retained microbubbles.   A clear visualization of the PESDA and Definity microbubble transit in real time was obtained by imaging the CAs with the 7-MHz contrast pulse sequencing PSS. No visually evident retention of microbubbles to the arterial endothelium was detected late (>30 s) after PESDA injections at baseline, except in the pig that exhibited a vasoconstrictive response to acetylcholine. In this pig, retention of PESDA microbubbles to the arterial endothelium could be observed (pig 4) (Fig. 4).



View larger version (109K):
[in this window]
[in a new window]
 
Figure 4 Transverse sectional images of carotid arteries obtained with low mechanical index real-time pulse sequence scheme at baseline (panels at left) and during hypertriglyceridemia (panels at right). Note that at baseline there was some adherence of microbubbles to the endothelium only in pig 4 (arrow) and no adherence in the other pigs. After intralipid infusion, we can see microbubbles adherence to the endothelium constituting an echogenic ring around the vessel lumen in all pigs.

 
After induction of hypertriglyceridemia, imaging of both right and left CAs before microbubble injections was performed and confirmed no background signals emanating from the endothelium. After intravenous PESDA injection, microbubble retention to the arterial endothelium was evident in both CAs of all pigs. This microbubble retention appeared as numerous bright, circular densities creating a ring of enhancement around the vessel (Fig. 4).

Visual adherence of PESDA microbubbles to arterial endothelium was also observed in all balloon-injured CAs and in none of the control vessels (Fig. 5, Video 1). This retention of microbubbles to the arterial endothelium, however, was not observed with Definity microbubbles, either during hypertriglyceridemia or after balloon injury of the CAs.



View larger version (111K):
[in this window]
[in a new window]
 
Figure 5 Pulse sequence scheme imaging showing microbubble retention to the endothelium in the carotid arteries (CAs) submitted to balloon-stretching (at left) and absence of retention in the contralateral control vessels (at right). In pig 5, the left CA was initially dissected, and the right CA was then submitted to the balloon dilation and considered for analysis. In this pig, the control demonstrated in the figure is the baseline imaging before dilation of the vessel. For an accompanying video corresponding to Figure 5 (Video 1), please see the September 1 issue of JACC at www.cardiosource.com/jacc.html.

 
Endothelial AI quantification.   After intralipid infusion, a significant increase in the endothelial AI was measured in all CAs after PESDA injection, except in the right CA of the pig in which microbubble adherence to the CA wall was detected even at baseline. Baseline endothelial acoustic intensities were less than zero owing to residual microbubbles in the central portion of the lumen. The mean endothelial AI measured late after PESDA injections increased by 1.1 ± 1.2 U during hypertriglyceridemia (p < 0.001) (Table 2). However, there was no difference in the endothelial AI after intravenous Definity when comparing baseline (–1.09 ± 1.03 U) and hypertriglyceridemia (–1.05 ± 1.05 U; p = 0.56). Figure 6 is an illustrative example of time-intensity curves obtained after intravenous PESDA and Definity microbubble injections during hypertriglyceridemia. The analysis of the curves in the late period after PESDA injection demonstrates that the entire lumen AI was higher than central lumen AI, resulting in a positive value of mean endothelial AI. After Definity injection, the central lumen AI was slightly higher than the entire lumen AI at the same time interval, with a resulting negative value of endothelial AI.


View this table:
[in this window]
[in a new window]
 
Table 2. Values of Endothelial Acoustic Intensity, Expressed in Units, at Baseline, During Hypertriglyceridemia, and After Balloon-Stretching of the Carotid Arteries Following Albumin- and Lipid-Encapsulated Microbubble Injections
 


View larger version (42K):
[in this window]
[in a new window]
 
Figure 6 Example of time-intensity curve showing the difference between the mean acoustic intensity (AI) of the entire lumen (dark line) and the central lumen (gray line) regions (A), and transverse images of carotid artery after PESDA and Definity microbubbles (MB) injection in the setting of hypertriglyceridemia (B). Note the retention of PESDA MB around the vessel, not seen with Definity. Amplification of the curves in the late period after PESDA MB injection (arrow) demonstrates a resulting positive value of mean endothelial AI. On the other hand, after Definity MB injection the central lumen AI was slightly higher than the entire lumen AI with a resulting negative value of endothelial AI.

 
The endothelial AI after PESDA microbubble injections in the balloon-injured vessels (+0.68 ± 1.28 U) was significantly higher than the endothelial AI in the control CAs (–0.53 ± 0.98 U; p < 0.01). No difference in the endothelial AI between these two CAs (balloon-injured vs. contralateral control) was observed with intravenous Definity (Table 2).

SEM.   Histologic analysis demonstrated preserved endothelium in the balloon-injured CAs, with minimal endothelium denudation, and no other damage in the media or adventitia layers.

The results of in vitro studies confirmed that microbubbles fixed with 10% formalin were stable and survived the SEM processing. When examined under SEM, microbubbles were characterized as a less electron-dense structure, with sizes ranging from 1 to 5 µm, spherically shaped (differentiating them from the biconcave-shaped erythrocytes), and not exhibiting any of the surface characteristics of hematopoietic cells, such as microvilli or microridges (Fig. 7). These criteria were then used to differentiate the microbubbles from red blood cells or endothelial protusions in the CA specimens.



View larger version (111K):
[in this window]
[in a new window]
 
Figure 7 The PESDA microbubbles fixed with formalin and examined under scanning electron microscopy (SEM) using a 0.22-µm Millipore filter, at (A) low magnification (bar = 10 µm; magnification 1,420x) and (B) high magnification (bar = 10 µm; magnification 7,400x). Under SEM the microbubbles were characterized as less electron-dense structures of different sizes, spherically shaped and not exhibiting any other surface features of hematopoietic cells.

 
Scanning electron microscopy of CAs demonstrated evidence of endothelial stretching with minimal denudation in all balloon-dilated vessels, and normal-appearing endothelium in the control vessels. Denudation of the endothelial layer was observed in <5% of the scanned regions of the injured CAs. However, microbubbles were attached to both the denuded areas and the endothelial cells within the stretched segments (Fig. 8A) and were not found in the normal control vessels (Fig. 8B). Scanning electron microscopy also confirmed the presence of microbubbles attached to the otherwise normal-appearing arterial endothelium of all three pigs euthanized during hypertriglyceridemia (Fig. 9).



View larger version (163K):
[in this window]
[in a new window]
 
Figure 8 High magnification (bar = 10 µm; magnification 1,420x) scanning electron microscopy pictures (left panels) and their respective low mechanical index pulse sequence scheme (PSS) images (right panels). Panel A demonstrates the presence of retained microbubbles in the endothelium detected by PSS in the left injured carotid artery, and Panel B reveals the absence of microbubbles in the endothelium in the control right side. Scanning electron microscopy revealed sites of injury with endothelial denudation and attachment of microbubbles (black arrows) to the denuded endothelium only in the injured vessel (A) and normal-appearing endothelium in the control vessel (B).

 


View larger version (86K):
[in this window]
[in a new window]
 
Figure 9 Scanning electron microscopy of carotid artery of a pig euthanized during hypertriglyceridemia at (A) low magnification (bar = 0.1 mm; magnification 462x) and (B) high magnification (bar = 10 µm; magnification 1,420x) showing the presence of microbubbles attached to the structurally normal endothelium (arrows).

 

    Discussion
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
In this study, we were able to detect and quantify microbubble adherence to dysfunctional endothelium using a high-resolution low mechanical index PSS. Using this low mechanical index prevented microbubble destruction and thus minimized any possible damage to the endothelium caused by the ultrasound. Dysfunctional endothelium of the CAs was induced globally by transient hypertriglyceridemia, and then regionally with balloon injury. In both settings, ED in the CAs was confirmed by a vasoconstrictive response to intra-arterial acetylcholine. The retained microbubbles were seen only in CAs that exhibited abnormal endothelial vasomotion.

The term "endothelial dysfunction" refers to a broad alteration in the endothelial phenotype that may lead to the development of atherosclerosis (21). Both traditional and newly recognized cardiovascular risk factors initiate a chronic inflammatory process that is accompanied by a loss of endothelium-dependent vasodilatory and anti-thrombotic properties. Current diagnostic tests designed to detect ED have centered around the detection of abnormalities in nitric oxide production. However, more recent observations have emphasized the anti-inflammatory properties of the endothelium, and that inflammatory markers such as C-reactive protein may be a mediator in the vasoconstrictive response to agents such as acetylcholine (22). Microbubble binding to endothelium has been consistently associated with increased endothelial expression of inflammatory markers (9–11). This inflammation process appears to play a role in the downregulation of nitric oxide synthase and augmentation of endothelium-derived vasoconstrictor production (22). Thus, microbubble adherence could indicate the presence of inflammatory markers that initiate the atherosclerotic process and mediate the loss of appropriate vasodilator responses.

This study is the first to demonstrate direct visualization of microbubble binding to dysfunctional endothelium induced by both a local arterial injury and systemic hypertriglyceridemia. Others have demonstrated in animal models that microbubbles specifically targeted to adhesion molecules may be necessary to concentrate retained microbubbles in sufficient quantity that they could be visualized with ultrasound (11,12). Our success with nontargeted microbubbles most likely results from the sensitivity of the contrast pulse sequence imaging scheme in detecting even low concentrations of microbubbles and from the use of a higher frequency transducer, which permitted a higher axial resolution image of the endothelial surface in the transverse plane.

Detection of both global and regional ed using intravenous microbubbles.   Endothelial dysfunction has been observed with transient hypertriglyceridemia induced by an intravenous or oral fat load (23,24). In our study, the presence of hypertriglyceridemia resulted in albumin-encapsulated microbubble retention in anatomically normal-appearing CAs. These data indicate that exposure of interstitial matrix beneath the endothelium is not necessary for binding. We also observed direct binding of albumin-encapsulated microbubbles to intact endothelial cells within the balloon-injured segment of the CAs. Others have observed albumin- and lipid-encapsulated microbubble binding to activated neutrophils and monocytes along the inflamed or injured venular endothelial surface (25,26). The absence of such an observation in our study by SEM indicates that a direct interaction between an endogenous receptor on the endothelial cell surface and the microbubble resulted in binding. The only known molecule within serum which binds to untargeted microbubbles is complement (18). An increase in complement receptor density has been observed with acute endothelial injury induced by hypoxia, resulting in increased C3b deposition (27). The C3 levels are augmented with both acute and chronic elevations in serum triglycerides (28,29). With hypertriglyceridemia, the increased complement on the surface of the microbubbles could have increased their binding to complement receptors already on the surface of endothelial cells (30). This would also explain why the lipid-encapsulated microbubbles did not appear to bind in this same setting. Although they also take up complement within serum, they contain a polyethylene glycol spacer that prevents complement-mediated binding to the endothelium (18).

Study limitations.   The ring of contrast enhancement around the vessel lumen during hypertriglyceridemia and after balloon injury was considered to reflect retained microbubbles to the arterial endothelium. One could argue that this signal could be coming from microbubbles flowing through the vasa vasorum in the arterial wall. However, we did not observe any increase in endothelial AI after lipid-encapsulated microbubbles in any of these situations. Although the Definity microbubbles were smaller, they still produced excellent vessel lumen opacification and would be expected to produce an increase in signal from the vasa vasorum. Secondly, we utilized the outline from the complete opacification of the border between lumen and endothelium early after injection of microbubbles to differentiate the bubbles present in the vessel lumen from those in the immediate peri-luminal area. This also allowed us to put the ROI exactly inside the vessel lumen for quantification of AI.

It has already been demonstrated that the acoustically driven diffusion of gas from the microbubbles is a slower mechanism of bubble destruction, and the rate of this diffusion increases during insonation even when using a low mechanical index. This produces changes in the microbubble diameter on the time scale of the acoustic pulse length (31) and may have caused some microbubble destruction even with the low mechanical index PSS used in this study. This may have prevented the accurate quantification of the number of adherent microbubbles in our study. As the mean size of Definity microbubbles is smaller than PESDA, acoustically driven diffusion from these microbubbles may have resulted in a more rapid loss of signal, and played a role in our observation that retention of microbubbles to the arterial endothelium was not detected after Definity injections.

The low endothelial AI values that we observed reflect the small retention fraction of microbubbles that actually adhere to the dysfunctional endothelium. However, the values of endothelial acoustic intensities under resting conditions were all less than zero, indicating that even late after injection there is still a slightly greater AI from a few residual bubbles in the central lumen. Furthermore, the mean value and range of values for endothelial AI remained less than zero at all stages of the study for intravenous Definity. However, the values for intravenous PESDA in the setting of balloon-induced ED all ranged greater than zero (range 0.09 to 2.31 U). Thus, we could both visualize retained microbubbles and quantify a significant change in AI despite the low retention fraction of microbubbles to the endothelium.

Although we observed microbubble retention with acute hypertriglyceridemia in this animal model, we do not know if a similar phenomenon would be observed in humans with chronic or acute hyperlipidemia. Further investigation in this area is necessary.


    Conclusions
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
A high-frequency linear transducer with a contrast-sensitive PSS was able to image retention of albumin-encapsulated microbubbles in dysfunctional carotid arterial endothelium. Although ED during transient hypertriglyceridemia had already been demonstrated by indirect measurement of vasoreactivity of the brachial artery by ultrasound, this is the first study to visualize retained arterial microbubbles in the setting of ED. Given the critical role of ED and inflammation in the pathogenesis of atherosclerosis and its close relationship with acute coronary syndromes, the detection of endothelial microbubble retention in vivo after intravenous injection of microbubbles could be a powerful clinical tool to identify the presence of vascular inflammation and serially follow it during treatment.


    Footnotes
 
Limited grant support was provided by Acuson-Siemens, Mountain View, California.


    References
 Top
 Abstract
 Methods
 Results
 Discussion
 Conclusions
 References
 
1. Celermajer DS. Endothelial dysfunction: does it matter? Is it reversible? J Am Coll Cardiol 1997;30:325-333.[Abstract]

2. Panza JA, Quyyumi AA, Brush JE, Epstein SEAbnormal endothelium-dependent vascular relaxation in patients with essential hypertension. N Engl J Med 1990;323:22-27.[Abstract]

3. Johnstone MT, Creager SJ, Scales KM, Cusco JA, Lee BK, Creager MA. Impaired endothelium-dependent vasodilation in patients with insulin-dependent diabetes mellitus Circulation 1993;88:2510-2516.[Abstract/Free Full Text]

4. Celermajer DS, Adams MR, Clarkson P, et al. Passive smoking and impaired endothelium-dependent arterial dilatation in healthy young adults N Engl J Med 1996;334:150-154.[Abstract/Free Full Text]

5. Lundman P, Eriksson M, Schenck-Gustafsson K, Karpe F, Tornvall P. Transient triglyceridemia decreases vascular reactivity in young, healthy men without risk factors for coronary heart disease Circulation 1997;96:3266-3268.[Abstract/Free Full Text]

6. Lupattelli G, Lombardini R, Schillaci G, et al. Flow-mediated vasoactivity and circulating adhesion molecules in hypertriglyceridemia: association with small, dense LDL cholesterol particles Am Heart J 2000;140:521-526.[CrossRef][Medline]

7. Anderson TJ, Uehata A, Gerhard MD, et al. Close relation of endothelial function in the human coronary and peripheral circulations J Am Coll Cardiol 1995;26:1235-1241.[Abstract]

8. Jayaweera AR, Edwards N, Glasheen WP, Villanueva FS, Abbott RD, Kaul S. In vivo myocardial kinetics of air-filled albumin microbubbles during myocardial contrast echocardiography: comparison with radiolabeled red blood cells Circ Res 1994;74:1157-1165.[Abstract/Free Full Text]

9. Villanueva FS, Jankowski RJ, Manaugh C, Wagner WR. Albumin microbubble adherence to human coronary endothelium: implications for assessment of endothelial function using myocardial contrast echocardiography J Am Coll Cardiol 1997;30:689-693.[Abstract]

10. Christiansen JP, Leong-Poi H, Klibanov AL, Kaul S, Lindner JR. Noninvasive imaging of myocardial reperfusion injury using leukocyte-targeted contrast echocardiography Circulation 2002;105:1764-1767.[Abstract/Free Full Text]

11. Lindner JR, Song J, Christiansen J, Klibanov AL, Xu F, Ley K. Ultrasound assessment of inflammation and renal tissue injury with microbubbles targeted to P-selectin Circulation 2001;104:2107-2112.[Abstract/Free Full Text]

12. Lindner JR, Song J, Xu F, et al. Noninvasive ultrasound imaging of inflammation using microbubbles targeted to activated leukocytes Circulation 2000;102:2745-2750.[Abstract/Free Full Text]

13. Philips P. Contrast pulse sequence (CPS): imaging nonlinear microbubbles. IEEE Trans Ultrason Ferroelectr Freq Control Symposium 2001;2:1739–45..

14. Porter TR, Hiser WL, Kricsfeld D, et al. Inhibition of carotid artery neointimal formation with intravenous microbubbles Ultrasound Med Biol 2001;27:259-265.[CrossRef][Medline]

15. Traupe T, D'Uscio LV, Muenter K, Morawietz H, Vetter W, Barton M. Effects of obesity on endothelium-dependent reactivity during acute nitric oxide synthase inhibition: modulatory role of endothelin Clin Sci (Lond) 2002;103(Suppl 48):13S-15S.

16. Porter TR, Taylor DO, Cycan A, et al. Endothelium-dependent pulmonary artery responses in chronic heart failure: influence of pulmonary hypertension J Am Coll Cardiol 1993;22:1418-1424.[Abstract]

17. Porter TR, Xie F, Kricsfeld D, Armbruster RW. Improved myocardial contrast with second harmonic transient ultrasound response imaging in humans using intravenous perfluorocarbon-exposed sonicated dextrose albumin J Am Coll Cardiol 1996;27:1497-1501.[Abstract]

18. Fisher NG, Christiansen JP, Klibanov A, Taylor RP, Kaul S, Lindner JR. Influence of microbubble surface charge on capillary transit and myocardial contrast enhancement J Am Coll Cardiol 2002;40:811-819.[Abstract/Free Full Text]

19. Du H, Chandaroy P, Hui SW. Grafted poly-(ethylene glycol) on lipid surfaces inhibits protein adsorption and cell adhesion Biochim Biophys Acta 1997;1326:236-248.[Medline]

20. Rim SJ, Leong-Poi H, Lindner JR, Wei K, Fisher NG, Kaul S. Decrease in coronary blood flow reserve during hyperlipidemia is secondary to an increase in blood viscosity Circulation 2001;104:2704-2709.[Abstract/Free Full Text]

21. Widlansky ME, Gokce N, Keaney JF, Vita JA. The clinical implications of endothelial dysfunction J Am Coll Cardiol 2003;42:1149-1160.[Abstract/Free Full Text]

22. Verma S, Buchanan MR, Anderson TJ. Endothelial function testing as a biomarker of vascular disease Circulation 2003;108:2054-2059.[Free Full Text]

23. Bae JH, Schwemmer M, Lee IK, et al. Postprandial hypertriglyceridemia-induced endothelial dysfunction in healthy subjects is independent of lipid oxidation Int J Cardiol 2003;87:259-267.[CrossRef][Medline]

24. Bae JH, Bassenge E, Kim KB, et al. Postprandial hypertriglyceridemia impairs endothelial function by enhanced oxidant stress Atherosclerosis 2001;155:517-523.[CrossRef][Medline]

25. Lindner JR, Dayton PA, Coggins MP, et al. Noninvasive imaging of inflammation by ultrasound detection of phagocytosed microbubbles Circulation 2000;102:531-538.[Abstract/Free Full Text]

26. Lindner JR, Coggins MP, Kaul S, Klibanov AL, Brandenburger GH, Ley K. Microbubble persistence in the microcirculation during ischemia/reperfusion and inflammation is caused by integrin- and complement-mediated adherence to activated leukocytes Circulation 2000;101:668-675.[Abstract/Free Full Text]

27. Collard CD, Bukusoglu C, Agah A, et al. Hypoxia-induced expression of complement receptor type 1 (CR1, CD35) in human vascular endothelial cells Am J Physiol 1999;276:C450-8.

28. Halkes CJ, van Dijk H, de Jaegere PP, et al. Postprandial increase of complement component 3 in normolipidemic patients with coronary artery disease: effects of expanded-dose simvastatin Arterioscler Thromb Vasc Biol 2001;21:1526-1530.[Abstract/Free Full Text]

29. Ylitalo K, Pajukanta P, Meri S, et al. Serum C3 but not plasma acylation-stimulating protein is elevated in Finnish patients with familial combined hyperlipidemia Arterioscler Thromb Vasc Biol 2001;21:838-843.[Abstract/Free Full Text]

30. Langeggen H, Berge KE, Johnson E, Hetland G. Human umbilical vein endothelial cells express complement receptor 1 (CD35) and complement receptor 4 (CD11c/CD18) in vitro Inflammation 2002;26:103-110.[CrossRef][Medline]

31. Chomas JE, Dayton P, Allen J, Morgan K, Ferrara KW. Mechanisms of contrast agent destruction IEEE Trans Ultrason Ferroelectr Freq Control 2001;48:232-248.[CrossRef][Medline]




This article has been cited by other articles:


Home page
Arterioscler. Thromb. Vasc. Bio.Home page
R. P. Choudhury and E. A. Fisher
Molecular Imaging in Atherosclerosis, Thrombosis, and Vascular Inflammation
Arterioscler. Thromb. Vasc. Biol., July 1, 2009; 29(7): 983 - 991.
[Abstract] [Full Text] [PDF]


Home page
Cardiovasc ResHome page
B. A. Kaufmann
Ultrasound molecular imaging of atherosclerosis
Cardiovasc Res, July 1, 2009; (2009) cvp179v2.
[Abstract] [Full Text] [PDF]


Home page
J Ultrasound MedHome page
J. A. Kopechek, T. M. Abruzzo, B. Wang, S. M. Chrzanowski, D. A. B. Smith, P. H. Kee, S. Huang, J. H. Collier, D. D. McPherson, and C. K. Holland
Ultrasound-Mediated Release of Hydrophilic and Lipophilic Agents From Echogenic Liposomes
J. Ultrasound Med., November 1, 2008; 27(11): 1597 - 1606.
[Abstract] [Full Text] [PDF]


Home page
StrokeHome page
M. J. Martin, E. M.L. Chung, A. H. Goodall, A. D. Martina, K. V. Ramnarine, L. Fan, S. V. Hainsworth, A. R. Naylor, and D. H. Evans
Enhanced Detection of Thromboemboli With the Use of Targeted Microbubbles
Stroke, October 1, 2007; 38(10): 2726 - 2732.
[Abstract] [Full Text] [PDF]


Home page
Cardiovasc ResHome page
D. R. Anderson, J. M. Tsutsui, F. Xie, S. J. Radio, and T. R. Porter
The role of complement in the adherence of microbubbles to dysfunctional arterial endothelium and atherosclerotic plaque
Cardiovasc Res, February 1, 2007; 73(3): 597 - 606.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Tsutsui, J. M.
Right arrow Articles by Porter, T. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tsutsui, J. M.
Right arrow Articles by Porter, T. R.

 