| The relationship between implant testing and training requirements for implanters |
|---|
|
|
|---|
First, no study has established that a shortage of trained implanters creates a barrier to ICD therapy, even in remote regions. High cost, lack of expert consensus on indications, lack of perceived benefit by patients or physicians for primary prevention, disparities in insurance coverage, or other factors may be more important.
Second, the authors assume that more implant capacity will result in better outcomes on a national level by servicing unmet need, but they provide no estimate of either need or present capacity. A work-time study sponsored by the North American Society of Pacing and Electrophysiology-Heart Rhythm Society determined that electrophysiologists have substantial excess implant capacity. They perform an average of three implants per month, spending only 12% of their time in implant-related activities (G. Naccarelli, personal communication, January 2003), but a recent study (2) estimated that ICDs were underused in urban areas despite a concentration of electrophysiologists who have unused implant capacity. If we make the conservative assumption that an electrophysiologist can implant two ICDs per week with testing, present implant capacity probably is in the range of two to three times the present volume. Once this capacity is saturated, it is not axiomatic that "more is better." After a threshold fraction of patients undergo coronary interventions, more interventions do not consistently improve outcomes (3).
Third, although the authors focus on implant testing, their implicit thesis is that a putative shortage of electrophysiologists is the principal barrier to ICD therapy. Having an insufficient supply of electrophysiologists perform all ICD care except implantation will not remove this barrier. Thus, the authors' fundamental assumption is that physicians who are not trained to perform defibrillation testing can nevertheless provide comprehensive, quality care to patients at risk for sudden death, but proficiency at implant testing is only one of a number of differences between training requirements for ICD and pacemaker implantation (4). Both VF induced by inappropriate therapy for supraventricular tachycardia and untreated ventricular tachycardia due to programming errors have caused sudden deaths in ICD recipients (5). Multiple unnecessary shocks delivered because of suboptimal programming may have a major adverse effect on quality of life. To be meaningful, the authors' study must test their fundamental assumption, that physicians without electrophysiologic training can provide comprehensive care for ICD patients, including selection, preoperative evaluation, implantation, programming, troubleshooting, follow-up, and management of drug-device interactions and recurrent ventricular tachyarrhythmias. Their proposed trial does not address this comprehensive issue.
Fourth, the authors assume that the outcome of their study will not be influenced strongly by the type of implant testing used. However, different testing methods result in different recommendations about programming or system revision, and experts disagree about which method is best. Before considering a multicenter study of testing versus no testing, it would be prudent either to identify optimal testing or demonstrate that a clinically relevant range of testing strategies results in similar outcomes.
The authors demonstrate that a randomized controlled trial will be large, expensive, and time consuming. We need better data in at least four key planning areas before considering one: 1) What is the optimal method for implant testing? 2) Does implant testing limit access to indicated ICDs? If so, to what degree? 3) How safe is a trial of "no testing" in a large prospectively selected subset of ICD recipients? Detection of VF may be unreliable if the R-wave is small. The high rate at which left-sided pectoral ICDs with right ventricular, apical defibrillation electrodes pass the 10-J safety margin is in part due to implant testing. Success rates
95% are achieved only after implanters change shock waveform, lead position, or generator position in
10% of patients (6). Defibrillation thresholds (DFTs) are higher with right pectoral implants. Even with left pectoral implants, experts disagree about whether lead positions other than the apex have consistently low DFTs. Some of these situations cannot be anticipated preoperatively. 4) How will the study test the fundamental assumption that physicians without electrophysiologic training can provide comprehensive care for ICD patients?
| Alternative perspective |
|---|
|
|
|---|
However, limited testing provides limited assurance of reliable defibrillation. The commonly used, 10-J safety margin identifies only
50% of patients who have unreliable defibrillation (8). Increases in DFTs that compromise defibrillation efficacy occur in up to 15% of patients who are programmed to a 10-J safety margin (9). Implantable cardioverter defibrillators reduce the risk of sudden death only by
50% (10). In one study of ICD recipients, more than half of sudden deaths were caused by shock failure or post-shock electromechanical dissociation (5). Some of the former may be prevented by a more specific implant criterion and some of the latter by programming lower, patient-specific shock strengths that cause less post-shock mechanical dysfunction. Even if it were practical to implant ICDs that deliver uniformly effective shocks, unnecessarily strong shocks would cause serious mechanical dysfunction in many patients (7).
Strict, conventional defibrillation implant criteria subject all patients to multiple VF episodes and a significant fraction to unnecessary system revisions and retesting. There is an inverse relationship between the fraction of patients who pass any defibrillation implant criterion and the subsequent success rate for conversion of spontaneous device-detected VF (8). Given the high, a priori probability of successful defibrillation at implants performed by trained implanters, the rules of conditional probability dictate that many patients with reliable defibrillation will not pass a highly specific implant criterion.
Thus, the implanter's dilemma is that programming based on more specific testing may increase long-term benefits of ICD therapybut only at the price of greater implant risk and complexity. Experts disagree on optimal implant criteria because they disagree about the trade-off between accuracy and risk. In this context, "no testing" is one end of a continuum that progresses with increasing rigor through limited safety-margin testing and DFT testing to research DFT testing used only in animals.
New testing methods improve the predictive accuracy of implant testing with fewer or no episodes of VF. Bayesian defibrillation testing is the most efficient method (11). Testing based on the upper limit of vulnerability permits the assessment of defibrillation efficacy without inducing VF in 80% to 90% of implants (1214). An interleaved Bayesian combination of defibrillation and vulnerability testing is one method to maximize the information derived from implant testing. Clinical studies will be required to determine whether new implant testing methods can improve outcomes.
| Financial considerations |
|---|
|
|
|---|
| The big picture |
|---|
|
|
|---|
Both the reliable sensing of VF and safe and effective defibrillation are crucial for effective ICD therapy, but neither can be assured at implant without some risk to the patient. Ideal implant testing would identify both the optimal ICD system for each patient and its optimal programming, without the need for either VF or shocks. Until such a method is developed, implanters will benefit from new, practical, evidence-based guidelines on optimal use of available methods. In the short term, guidelines may recommend less testing, more testing, or different (e.g., vulnerability) testing, depending on the clinical situation. In the long term, they may recommend no testing if supported by the level of evidence proposed by the authors. The goal of reassessing implant testing should be to determine the optimal balance between implant safety and long-term benefits of ICD therapy. To abandon implant testing is to squander a valuable opportunity to improve outcomes for ICD patients.
Available evidence indicates that nationwide ICD implant capacity is not presently restricted by either the requirement for implant testing or training requirements for implanters. There are no waiting lists for ICDs. Although implant capacity should be sufficient for unmet need, excess implant capacity should not drive need. The analysis of barriers to therapy for prevention of arrhythmic death should include all steps in the complex chain required for safe, effective, and comprehensive antiarrhythmic therapy.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. A. Blatt, J. E. Poole, G. W. Johnson, D. J. Callans, M. H. Raitt, R. K. Reddy, F. E. Marchlinski, R. Yee, T. Guarnieri, M. Talajic, et al. No Benefit From Defibrillation Threshold Testing in the SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial) J. Am. Coll. Cardiol., August 12, 2008; 52(7): 551 - 556. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Day, B. Olshansky, S. Moore, S. Brown, K. Q. Stolen, D. R. Lerew, and for the INTRINSIC RV Study Investigators High defibrillation energy requirements are encountered rarely with modern dual-chamber implantable cardioverter-defibrillator systems Europace, March 1, 2008; 10(3): 347 - 350. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. K. Kantharia Defibrillation thresholds testing of implantable cardioverter-defibrillators at implantation J. Am. Coll. Cardiol., February 1, 2005; 45(3): 468 - 468. [Full Text] [PDF] |
||||
![]() |
C. Swerdlow Reply J. Am. Coll. Cardiol., February 1, 2005; 45(3): 468 - 468. [Full Text] [PDF] |
||||
![]() |
D. A.M.J. Theuns, T. Szili-Torok, and L. J. Jordaens Defibrillation efficacy testing: Long-term follow-up and mortality Europace, January 1, 2005; 7(6): 509 - 515. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||