cardiology careers collections past issues search home
     

J Am Coll Cardiol, 2004; 44:88-91, doi:10.1016/j.jacc.2003.11.068
© 2004 by the American College of Cardiology Foundation
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Strickberger, S. A.
Right arrow Articles by Klein, G. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Strickberger, S. A.
Right arrow Articles by Klein, G. J.

CLINICAL RESEARCH: ELECTROPHYSIOLOGY: VIEWPOINT

Is defibrillation testing required for defibrillator implantation?

S. Adam Strickberger, MD*,* and George J. Klein, MD{dagger}

* Division of Cardiology, Washington Hospital Center, Washington, DC, USA
{dagger} Division of Cardiology, University of Ontario, London, Ontario, Canada

Manuscript received September 11, 2003; revised manuscript received November 17, 2003, accepted November 24, 2003.

* Reprint requests and correspondence: Dr. S. Adam Strickberger, 110 Irving Street, NW #5A-12, Washington, DC 20010-2975, USA.
strickberger{at}medstar.net


    Abstract
 Top
 Abstract
 History of DFT testing
 The DFT systems
 Techniques for determining DFT...
 DFT testing or not
 Conclusions
 References
 
The assessment of defibrillation (DFT) efficacy has long been the standard of care during defibrillator implantation. To ensure an acceptable DFT safety margin, early defibrillator systems frequently required that the shock polarity and the location, type, or number of electrodes had to be altered. Advances in defibrillator and lead technology have resulted in lower and more consistent DFT thresholds in the range of 10 J, with an infrequent requirement to modify the DFT system. Yet, one can make an argument for and against continuation of DFT testing at the time of defibrillator implantation. The goal of this paper is to address both the data that do support and the data that do not support continuation of DFT testing at the time of device implantation. Scientifically, DFT testing should be abandoned only when prospective evidence demonstrates that defibrillator implantation without testing is as safe and has the same mortality benefits as implantation with testing. The most attractive aspect of eliminating DFT efficacy testing is that more patients may have the opportunity to be treated with this life-saving therapy. Perhaps there are alternative strategies to improve accessibility to defibrillator therapy without possibly eroding its effectiveness. In the end, will lives be saved or lost if we discontinue DFT efficacy testing and lower the barriers to implantable defibrillator therapy?

Abbreviations and Acronyms
  DFT = defibrillation
  VF = ventricular fibrillation


The assessment of defibrillation (DFT) efficacy has long been the standard of care during defibrillator implantation (1,2). To ensure an acceptable DFT safety margin, early defibrillator systems frequently required that the shock polarity and the location, type, or number of electrodes had to be altered. Advances in defibrillator and lead technology have resulted in lower and more consistent DFT thresholds in the range of 10 J, with an infrequent requirement to modify the DFT system (3,4). Because DFT efficacy testing adds time, possibly adds cost, and involves some morbidity, it is fair to review the evidence and ask whether the reliability of DFT with standard implantation techniques is sufficient to preclude DFT testing at the time of defibrillator implantation.


    History of DFT testing
 Top
 Abstract
 History of DFT testing
 The DFT systems
 Techniques for determining DFT...
 DFT testing or not
 Conclusions
 References
 
Defibrillation efficacy traditionally has been performed with successive DFT attempts using approximately 20 J in a device that delivered 25 to 30 J (1,2). A safety margin of 10 J was required for the implantation of the earliest defibrillators. The inventors and early researchers of the implantable defibrillator selected 10 J for the safety margin based on the maximum available energy of 25 J and on a few animal and human studies that suggested early DFT systems would generally be effective with <15 J (Mower MM, personal communication, 2003) (5,6). The initial implants required successful DFT with the maximum device output of 25 J, which therefore assumed a 10-J safety margin (M. M. Mower, personal communication, 2003). The 10-J safety margin was subsequently confirmed in patients with an implantable defibrillator (7) and validated as defibrillators were demonstrated to reduce mortality (8–11) and because patients with inadequate DFT safety margins were thought to have a higher mortality rate (12,13). As technology improved, DFT thresholds were substantially reduced. Most studies of new DFT systems required documentation of a 10-J safety margin, with only the rare study requiring testing beyond documentation of an adequate safety margin (14). Today, our general impression is that only a small percentage of physicians do more than confirm a 10-J safety margin at implant.


    The DFT systems
 Top
 Abstract
 History of DFT testing
 The DFT systems
 Techniques for determining DFT...
 DFT testing or not
 Conclusions
 References
 
Current implantable defibrillator systems provide an average DFT threshold of 8 to 10 J. These systems use one or two intracavitary DFT electrodes while the defibrillator pulse generator functions as a subcutaneous electrode (active can). Only in unusual cases does the DFT system need to be modified to achieve adequate DFT efficacy, for example, with the addition of another subcutaneous DFT electrode, reversal of shock polarity, and/or with the use of a pulse generator with greater delivered energy (15). The frequency with which these approaches are used to achieve adequate DFT efficacy is probably around 5%.


    Techniques for determining DFT efficacy
 Top
 Abstract
 History of DFT testing
 The DFT systems
 Techniques for determining DFT...
 DFT testing or not
 Conclusions
 References
 
Defibrillation efficacy may be assessed in various ways. The most time-honored approach confirms a 10-J safety margin. Considerably less frequently, a step-down DFT protocol that uses decrementing energy levels for subsequent DFT attempts until DFT fails or an abbreviated step-down protocol that is terminated after two or three DFT attempts have been used. Determining the upper limit of vulnerability as an alternative to direct DFT testing has been proposed (16,17). The upper limit of vulnerability demonstrates DFT efficacy by estimating the DFT threshold from the lowest energy level that will not induce ventricular fibrillation (VF) with a "shock on T" (16). Each of the aforementioned techniques that assess DFT efficacy will identify the infrequent patient in whom modification of the DFT system is required to achieve an adequate DFT safety margin.

Documentation of DFT efficacy provides an additional safety evaluation of the lead position and the system's ability to sense, detect, and defibrillate VF. The induction of VF at the time of defibrillator implantation poses little risk. Although precise data are not available, the intraoperative mortality rate is probably in the range of 0.1%.


    DFT testing or not
 Top
 Abstract
 History of DFT testing
 The DFT systems
 Techniques for determining DFT...
 DFT testing or not
 Conclusions
 References
 
What is the rationale for abandoning DFT testing?.   First, the probability of a high DFT threshold and a failed implant is quite small. Second, the majority of ventricular arrhythmias treated by an implantable defibrillator are ventricular tachycardia. The cardioversion energy for ventricular tachycardia is probably significantly less than the DFT energy required for VF (18). In addition, the probabilistic nature of DFT suggests that shocks less than the determined threshold, when repeated several times, have some probability of success even in the much less likely event of VF (4,19). Thus, even the infrequent patient with a high DFT threshold who does receive a DFT system that is incapable of defibrillating VF at the time of device implantation may not to succumb to an episode of VF. Indeed, defibrillator testing is withheld in some patients with profound left ventricular dysfunction, a group where DFT testing is arguably the most necessary. Next, the determination of DFT efficacy requires additional anesthetics, and the shocks themselves may cause hemodynamic compromise (20). Finally, DFT testing may be a barrier to implantable defibrillator therapy in regions with few or no electrophysiologists. It may be argued that other physicians, such as those currently implanting pacemakers, would more readily implant defibrillators if not for the requirement of DFT testing.

What are the arguments for continued DFT testing?.   First, all the data supporting increased survival with implantable defibrillators were attained with some assessment of DFT efficacy at implant (8–11). Arguably, the requirement for testing was essential in demonstrating device efficacy. Second, there is no doubt that some patients receive a more effective DFT system because DFT testing is performed. For instance, patients with a frank inability to defibrillate are identified, and corrective measures such as the use of a higher energy device, addition of electrodes, or reversal of shock polarity can be used (15). Ensuring that the system provides effective DFT and appropriate sensing of VF is intuitively very compelling, although not proven to improve survival. There are no data that specifically demonstrate increased mortality among patients with high DFT thresholds. Epstein et al. (12) observed that 42% of patients with an elevated DFT threshold who were treated with an implantable defibrillator died suddenly. However, the two-year mortality rate was 20%. An annual mortality rate in an implantable defibrillator population from the early 1990s of approximately 10% may not be considered excessive, and it may be possible that patients who cannot be defibrillated easily have intrinsically higher mortality rates independent of defibrillator usage. Third, an accurate estimate of the DFT threshold should facilitate programming of shocks to a multiple of the DFT threshold, allowing lower energy shocks (3,4). This may result in less syncope and near-syncope and improve post-shock hemodynamics (20). Finally, the medico-legal implications of an unexplained death or documented failure to defibrillate are uncertain. Arguably, failure by the physician to verify DFT efficacy may be a compelling argument that the implant procedure fell below the standard of care, a position that would be supported by many electrophysiologists.

What evidence would be required to establish equivalence or superiority of a strategy of no testing versus current DFT testing?.   Data from a randomized trial would probably be required. Patients undergoing defibrillator implantation would be randomized to DFT testing or no DFT testing. It seems obvious that mortality should be the primary end point, or at least a major component of the primary end point, if a composite primary end point was developed. Factors such as cost and procedural complications can be readily assessed but are unlikely to be acceptable as sole primary end points. To estimate the number of patients required for a randomized mortality study, the first assumption is that 5% of patients will not be successfully defibrillated with the maximum output of a standard device that uses the usual lead position and shock polarity. Second, <75% of this 5% of patients will receive appropriate therapies after implant (i.e., 3.75% of the total number of patients who receive an implantable defibrillator). If 25% of appropriate shocks are for VF and represent excess mortality, then the rate of excess mortality as a result of not assessing DFT efficacy would be 25% of 3.75% of patients, or about 1% of initial implants. If the usual death rate for implantable defibrillator patients is 10% per year, a reasonable estimate, the study population would need to differentiate between mortality rates of 10% and 11%. Using these assumptions, which probably overestimate the frequency of high DFT thresholds and the frequency of VF, approximately 29,000 patients would need to be randomized to achieve a statistical power of 80%. Although different assumptions can be made and a creative study design can reduce the sample size, it is clear that such a study would involve a large number of patients.

A looming question is who will do the implantation procedure with the study? If the goal is to demonstrate that experienced defibrillator implanters can safely and effectively implant a defibrillator without DFT efficacy testing, then these physicians should implant the devices and position the leads. If the goal is to demonstrate that experienced pacemaker implanting physicians without defibrillator implantation experience can safely and appropriately implant defibrillators, then these physicians should participate in the trial and at least perform the implants in the patients randomized to implantation without DFT efficacy testing. This would allow the study to test the hypothesis that physicians with pacemaker implantation experience, but not defibrillator implantation experience, can implant defibrillators with an appropriate mortality rate.


    Conclusions
 Top
 Abstract
 History of DFT testing
 The DFT systems
 Techniques for determining DFT...
 DFT testing or not
 Conclusions
 References
 
The assessment of DFT efficacy at the time of implant has long been the standard of care. It is intuitively reasonable to ensure that the system has an acceptable DFT safety margin at the time of implant. One might ask simply whether the downsides of ensuring device sensing and DFT capability are sufficient to discontinue this practice. DFT efficacy testing may be minimally assessed with as few as one or two inductions of VF with only rare complications. The expertise involved in performing this task should not be outside the capabilities for an implanting team involved in the care of patients requiring these devices. With the evolution of devices that defibrillate more efficiently and predictably, it is fair to ask whether initial DFT testing is still necessary. Scientifically, this question should be answered only with prospective evidence that demonstrates at least equivalent patient safety, equivalent mortality reduction, and cost savings. Proponents of eliminating DFT testing need to demonstrate the safety and benefits of such a strategy with objective evidence before this element of defibrillator implantation is abandoned. The most attractive aspect of eliminating DFT efficacy testing is that more patients may have the opportunity to be treated with this life-saving therapy. Perhaps there are alternate strategies to improve accessibility to defibrillator therapy without possibly eroding its effectiveness. In the end, will lives be saved or lost if we discontinue DFT efficacy testing and lower the barriers to implantable defibrillator therapy?


    References
 Top
 Abstract
 History of DFT testing
 The DFT systems
 Techniques for determining DFT...
 DFT testing or not
 Conclusions
 References
 

  1. Echt DS, Armstrong K, Schmidt P, Oyer PE, Stinson EB, Winkle RA. Clinical experience, complications, and survival in 70 patients with the automatic implantable cardioverter/defibrillator. Circulation. 1985;71:289–296[Abstract/Free Full Text]
  2. Marchlinski FE, Flores BT, Buxton AE, et al. The automatic implantable cardioverter-defibrillator: Efficacy, complications, and device failures. Ann Intern Med. 1986;104:481–488[Medline]
  3. Gold MR, Higgins S, Klein R, et al. Efficacy and temporal stability of reduced safety margins for ventricular defibrillation: Primary results from the low energy safety study (LESS). Circulation. 2002;105:2043–2048[Abstract/Free Full Text]
  4. Strickberger SA, Daoud EG, Davidson T, et al. Probability of successful defibrillation at multiples of the defibrillation energy requirement in patients with an implantable defibrillator. Circulation. 1997;96:1217–1223[Abstract/Free Full Text]
  5. Mirowski M, Mower MM, Gott VL, Brawley RK, Denniston RH. Transvenous automatic defibrillator-preliminary clinical tests of the defibrillating subsystem. Trans Am Soc Art Intern Organs. 1972;18:520–524
  6. Mirowski M, Mower MM, Gott VL, Brawley RK. The feasibility and effectiveness of low-energy catheter defibrillation in man. Circulation. 1973;47:79–85[Abstract/Free Full Text]
  7. Marchlinski FE, Flores B, Miller JM, Gottlieb CD, Hargrove WC. Relation of the intraoperative defibrillation threshold to successful postoperative defibrillation with an automatic implantable cardioverter defibrillator. Am J Cardiol. 1988;62:393–398[CrossRef][Medline]
  8. Moss AJ, Hall WJ, Cannom DS, et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators. N Engl J Med. 1996;335:1933–1940[Abstract/Free Full Text]
  9. Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G. A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators. N Engl J Med. 1999;341:1882–1890[Abstract/Free Full Text]
  10. The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med. 1997;337:1576–1583[Abstract/Free Full Text]
  11. Multicenter Automatic Defibrillator Implantation Trial II InvestigatorsMoss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877–883[Abstract/Free Full Text]
  12. Epstein AE, Ellenbogen KA, Kirk KA, Kay GN, Dailey SM, Plumb VJ. Clinical characteristics and outcome of patients with high defibrillation thresholds. A multicenter study. Circulation. 1992;86:1206–1216[Abstract/Free Full Text]
  13. Participating InvestigatorsLehmann MH, Thomas A, Nabih M, et al. Sudden death in recipients of first-generation implantable cardioverter-defibrillators: Analysis of terminal events. J Interv Cardiol. 1994;7:487–503[Medline]
  14. Strickberger SA, Brownstein SL, Wilkoff BL, Zinner AJ. Clinical predictors of defibrillation energy requirements in patients treated with a non-thoracotomy defibrillator system. Am Heart J. 1996;131:257–260[CrossRef][Medline]
  15. Gold MR, Shorofsky SR. Transvenous defibrillation lead systems. J Cardiovasc Electrophysiol. 1996;7:570–580[Medline]
  16. Swerdlow CD, Ahern T, Kass RM, Davie S, Mandel WJ, Chen PS. Upper limit of vulnerability is a good estimator of shock strength associated with 90% probability of successful defibrillation in humans with transvenous implantable cardioverter-defibrillators. J Am Coll Cardiol. 1996;27:1112–1118[Abstract]
  17. Green UB, Garg A, Al-Kandari F, Ungab G, Tone L, Feld GK. Successful implantation of cardiac defibrillators without induction of ventricular fibrillation using upper limit of vulnerability testing. J Interv Card Electrophysiol. 2003;8:71–75[CrossRef][Medline]
  18. Zipes DP, Jackman WM, Heger JJ, et al. Clinical transvenous cardioversion of recurrent life-threatening ventricular tachyarrhythmias: Low energy synchronized cardioversion of ventricular tachycardia and termination of ventricular fibrillation in patients using a catheter electrode. Am Heart J. 1982;103:789–794[CrossRef][Medline]
  19. Davy JM, Fain ES, Dorian P, Winkle RA. The relationship between successful defibrillation and delivered energy in open-chest dogs: Reappraisal of the "defibrillation threshold" concept. Am Heart J. 1987;113:77–84[CrossRef][Medline]
  20. Tokano T, Chang J, Davis J, et al. The effect of ventricular shock strength on cardiac hemodynamics. J Cardiovasc Electrophysiol. 1998;9:791–797[Medline]



This article has been cited by other articles:


Home page
J Am Coll CardiolHome page
A. B. Curtis
Defibrillation Threshold Testing in Implantable Cardioverter-Defibrillators: Might Less Be More Than Enough?
J. Am. Coll. Cardiol., August 12, 2008; 52(7): 557 - 558.
[Full Text] [PDF]


Home page
EuropaceHome page
J. D. Day, B. Olshansky, S. Moore, S. Brown, K. Q. Stolen, D. R. Lerew, and for the INTRINSIC RV Study Investigators
High defibrillation energy requirements are encountered rarely with modern dual-chamber implantable cardioverter-defibrillator systems
Europace, March 1, 2008; 10(3): 347 - 350.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
M. Brignole, G. Raciti, M. G. Bongiorni, G. De Martino, S. Favale, M. Gasparini, R. Luise, E. Occhetta, and A. Proclemer
Defibrillation testing at the time of implantation of cardioverter defibrillator in the clinical practice: a nation-wide survey
Europace, July 1, 2007; 9(7): 540 - 543.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
J. D. Day, R. N. Doshi, P. Belott, U. Birgersdotter-Green, M. Behboodikhah, P. Ott, K. A. Glatter, S. Tobias, H. Frumin, B. K. Lee, et al.
Inductionless or Limited Shock Testing Is Possible in Most Patients With Implantable Cardioverter- Defibrillators/Cardiac Resynchronization Therapy Defibrillators: Results of the Multicenter ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations)
Circulation, May 8, 2007; 115(18): 2382 - 2389.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
B. K. Kantharia
Defibrillation thresholds testing of implantable cardioverter-defibrillators at implantation
J. Am. Coll. Cardiol., February 1, 2005; 45(3): 468 - 468.
[Full Text] [PDF]


Home page
EuropaceHome page
D. A.M.J. Theuns, T. Szili-Torok, and L. J. Jordaens
Defibrillation efficacy testing: Long-term follow-up and mortality
Europace, January 1, 2005; 7(6): 509 - 515.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Strickberger, S. A.
Right arrow Articles by Klein, G. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Strickberger, S. A.
Right arrow Articles by Klein, G. J.

 
  cardiology careers collections past issues search home