EXPEDITED REVIEW
Incidence and clinical course of thrombus formation on atrial septal defect and patient foramen ovale closure devices in 1,000 consecutive patients
Ulrike Krumsdorf, MD*,
Stefan Ostermayer, MD*,
Kai Billinger, MD*,
Thomas Trepels, MD*,
Elisabeth Zadan, MD*,
Kathrin Horvath, MD* and
Horst Sievert, MD, PhD*,*
* Cardiovascular Center Frankfurt, Sankt Katharinen, Frankfurt, Germany
Manuscript received May 2, 2003;
revised manuscript received October 13, 2003,
accepted October 27, 2003.
* Reprint requests and correspondence: Prof. Dr. med. Horst Sievert, Cardiovascular Center Frankfurt, Sankt Katharinen, Seckbacher Landstrasse 65, 60389 Frankfurt, Germany. horst.sievert{at}dgn.de
 |
Abstract
|
|---|
OBJECTIVES: The purpose of this study was to investigate the incidence, morphology, and clinical course of thrombus formation after catheter closure of intra-atrial shunts.
BACKGROUND: Post-procedure detailed information about thrombotic material on different devices for transcatheter closure is missing.
METHODS: A total of 1,000 consecutive patients were investigated after patent foramen ovale (PFO) (n = 593) or atrial septal defect (ASD) (n = 407) closure. Transesophageal echocardiography (TEE) was scheduled after four weeks and six months. Additional TEEs were performed as clinically indicated.
RESULTS: Thrombus formation in the left atrium (n = 11), right atrium (n = 6), or both (n = 3) was found in 5 of the 407 (1.2%) ASD patients and in 15 of the 593 (2.5%) PFO patients (p = NS). The thrombus was diagnosed in 14 of 20 patients after four weeks and in 6 of 20 patients later on. The incidence was: 7.1% in the CardioSEAL device (NMT Medical, Boston, Massachusetts); 5.7% in the StarFLEX device (NMT Medical); 6.6% in the PFO-Star device (Applied Biometrics Inc., Burnsville, Minnesota); 3.6% in the ASDOS device (Dr. Ing, Osypka Corp., Grenzach-Wyhlen, Germany); 0.8% in the Helex device (W.L. Gore and Associates, Flagstaff, Arizona); and 0% in the Amplatzer device (AGA Medical Corp., Golden Valley, Minnesota). The difference between the Amplatzer device on one hand and the CardioSEAL device, the StarFLEX device, and the PFO-Star device on the other hand was significant (p < 0.05). A pre-thrombotic disorder as a possible cause of the thrombus was found in two PFO patients. Post-procedure atrial fibrillation (n = 4) and persistent atrial septal aneurysm (n = 4) had been found as significant predictors for thrombus formation (p < 0.05). In 17 of the 20 patients, the thrombus resolved under anticoagulation therapy with heparin or warfarin. In three patients, the thrombus was removed surgically.
CONCLUSIONS: The incidence of thrombus formation on closure devices is low. The thrombus usually resolves under anticoagulation therapy.
|
Abbreviations and Acronyms
| | AF | = atrial fibrillation | | ASA | = atrial septal aneurysm | | ASD | = atrial septal defect | | PFO | = patent foramen ovale | | TEE | = transesophageal echocardiographic/echocardiography | | TIA | = transient ischemic attack | | TTE | = transthoracic echocardiographic/echocardiography |
|
Transcatheter closure of atrial septal defect (ASD) and patent foramen ovale (PFO) has become a less invasive alternative to open heart surgery. The first attempt at transcatheter closure of an ASD was described by King and Mills in 1974 (1,2). Since then, there has been a remarkable evolution of devices with different materials, mechanism of closure, and methods of implantation (319). Although excellent results have been reported for transcatheter occlusion with different devices (319), concerns have arisen about peri- and/or post-interventional complications (5,20,21). Information about thrombus formation on devices and its clinical outcome is very limited (2123). The purpose of our study was to determine and analyze the incidence, morphology, possible predictors, and clinical management of thrombus formation on ASD and PFO closure devices in 1,000 consecutive patients.
 |
Methods
|
|---|
Patient population.
Between August 1992 and January 2003, transcatheter closure of a PFO (n = 593) or an ASD (n = 407) was performed in 1,000 consecutive patients (579 females and 421 males) between the age 13 and 90 years (mean age 48 ± 15 years). Nine different devices were used: 1) Rashkind (USCI Bard, Billerica, Massachusetts); 2) Buttoned (Custom Medical Devices, Amarillo, Texas); 3) ASDOS (Dr. Ing, Osypka Corp., Grenzach-Wyhlen, Germany); 4) Angel Wings (Das Angel Wings, Microvena Corp., Vadnais, Minnesota); 5 and 6) CardioSEAL and StarFLEX (NMT Medical, Boston, Massachusetts); 7) Amplatzer (AGA Medical Corp., Golden Valley, Minnesota); 8) Helex (W. L. Gore and Associates, Flagstaff, Arizona); and 9) PFO-Star (Applied Biometrics Inc., Burnsville, Minnesota).
The stretched ASD diameter, measured with a balloon catheter, varied between 6 and 37 mm (mean 20 ± 6 mm) and pulmonary/systemic output ratio from 0.4 to 4.7 (mean 2.0 ± 0.7). Of the ASD patients, 50 (12%) had multiple defects. In 15 (30%) of these 50 patients, two devices were implanted. An atrial septal aneurysm (ASA) defined as an excursion of the septum of more than 10 mm was present in 52 (13%) of the ASD patients. The following concomitant diseases were present in the ASD patients: pulmonary hypertension in 133 (33%) patients, atrial fibrillation (AF) in 56 (14%), coronary artery disease in 34 (8%) patients, systemic hypertension in 88 (22%) patients, diabetes in 17 (4%) patients, phlebothrombosis in 18 (4%) patients, and embolic events (stroke, transient ischemic attack [TIA], and peripheral embolism) in 65 (16%) patients.
In the PFO patients, an ASA was present in 157 (27%) cases. The stretched diameter of the PFO was also defined by balloon sizing and ranged from 4 to 21 mm (mean 9 ± 3 mm). In these patients, 741 embolic events occurred before PFO closure: 353 strokes, 315 TIAs, 38 prolonged reversible ischemic neurologic deficits, and 35 peripheral embolisms with no other known source. On average, the patients had 1.2 embolic events before the procedure. The annual recurrence rate after the first event was 25%.
Some ASD and PFO patients had coagulation disorders, such as protein C deficiency (ASD: n = 2; PFO: n = 7; p = NS), protein S deficiency (ASD: n = 3, PFO: n = 5; p = NS), or activated protein C resistance (ASD: n = 9; PFO: n = 16; p = NS). Overall, 14 of 221 (6.3%) ASD patients and 28 of 235 (12%) PFO patients had one of these disorders. This difference was not significant. Of the patients with an ASD, 65 had an embolic event before defect closure. In total, 300 patients had an embolic event (235 PFOs and 65 ASDs). In these 300 patients a coagulation disorder was diagnosed in 39 (13%) cases compared with only 3 (19%) cases in those 156 patients without an embolic event (p < 0.001).
Closure devices.
Table 1 provides an overview of the devices used in this series. These have been described in detail elsewhere (319).
Occlusion procedure.
The procedures were performed under local anesthesia unless the patient did not tolerate the transesophageal echocardiographic (TEE) probe. Detailed echocardiographic assessment of the size and morphology of the ASD or PFO and the correct positioning of the device was performed using a multiplane 5-MHz probe. Patients were fully heparinized (10,000 to 20,000 U heparin), and antibiotic prophylaxis was given. Measurements of the defect and occlusion techniques with the different devices have been reported previously (319). Until October 2001 (in 817 patients) heparin was antagonized with protamine at the end of the procedure. Therefore, the patients were usually discharged on the same or following day.
Follow-up.
A clinical examination, an electrocardiograph, a chest radiograph in two projections, and a transthoracic echocardiograph (TTE) were performed before discharge and repeated after 1, 6, and 12 months. In addition, TEE follow-up has been performed according to different study protocols approved by the local institutional review board. The TEE was scheduled four weeks and six months after closure with the newer devices: StarFLEX, Amplatzer, Helex, and PFO-Star. In older device protocols (Cardio-SEAL, Angel Wings, ASDOS, Buttoned Device, and Rashkind), control TEE after four weeks was optional, and at six months it was recommended. Cardiac catheterization was performed in patients with moderate and large residual shunts. Fluoroscopy to evaluate the metallic structures of the device was also performed after six months. Thereafter, follow-up was performed by a questionnaire every 6 to 12 months. In the event of a recurrent embolic situation, additional clinical investigations were performed as necessary. In event of a thrombus, repeat TEEs were scheduled until resolution of the thrombus.
Anticoagulation therapy.
According to different device protocols, different regimens of anticoagulation and antiplatelet therapy were begun after implantation. Anticoagulation with warfarin (target international normalized ratio 2 to 3 for patients with the Rashkind and ASDOS devices) or anti-aggregation with aspirin (100 mg daily for patients with the Buttoned, Angel Wings, CardioSEAL, StarFLEX, Amplatzer, and Helex devices) was conducted for six months. As for the PFO-Star study protocol, clopidogrel (75 mg daily) was additionally recommended for three months. Since January 2001 (in 264 patients), protocols were changed for all PFO patients to a combination of aspirin and clopidogrel for six months.
Statistical analysis.
Data were described as mean ± SD, and percentages were expressed where appropriate. Comparisons of results in patients with and without thrombus formation and between different devices were assessed by chi-square test; this was replaced by the Fisher exact test in cases in which the cell frequency was <5. A multiple regression analysis was performed to determine how variables (predisposing factors) were related to the risk of thrombus formation on ASD and PFO closure devices after implantation. A value of p < 0.05 was considered to be statistically significant.
 |
Results
|
|---|
Implantation of the device was technically successful in all patients. However, two attempts were necessary in 13 patients. During a follow-up period of 1 to 108 months (mean 36 ± 17 months; 1,327 patient-years), a thrombus formation was detected in 20 of 1,000 (2%) patients.
Thrombus formation in the left atrium (n = 11), right atrium (n = 6), or both (n = 3) was found in 5 of the 407 (1.2%) ASD patients and in 15 of the 593 (2.5%) PFO patients (p = NS). The TEE had been performed four weeks after closure in 709 of 1,000 (71%) patients and after six months in 714 of 1,000 (71%) patients. Thrombotic structures were observed by TEE after four weeks in 14 patients (Table 2), after six months in 3 patients (Table 2), after one year in 2 patients, and after five years in 1 patient. None of these thrombi could be detected by TTE, which had also been performed in all patients at the same time. The diameter of the thrombi varied from 5 to 30 mm (mean 11.7 ± 7 mm). The thrombus appeared as pedunculated and mobile in 9 (45%) cases (Fig. 1A). Immobile thrombi were detected in 11 (55%) patients (Fig. 1B). In two of these patients, a right atrial thrombus was attached to the lateral atrial wall (Fig. 1C) without direct contact to the occlusion device.

View larger version (75K):
[in this window]
[in a new window]
|
Figure 1 (A) Transesophageal echocardiography four-chamber view: left-sided mobile thrombus attached to a StarFLEX occluder detected four weeks after catheter closure. (B) Transesophageal echocardiography short-axis: right- and left-sided immobile thrombus surrounding an ASDOS occluder detected four weeks after catheter closure. (C) Transesophageal echocardiography short-axis: large mobile thrombus (30 x 18 mm) attached to the right atrial wall (without direct contact to the ASDOS device) detected one year after catheter closure. LA = left atrium; RA = right atrium.
|
|
In devices with star-shaped metallic arms (CardioSEAL, StarFLEX, and PFO-Star occluder; n = 14) thrombus formations seemed to be attached either at the center of the right or left atrial disc (n = 5), at the transitional epithelium between device and septum (n = 1), or attached to the expanded nitinol or metal arms (n = 8).
Four weeks after implantation, the TEE revealed a thrombus formation in 14 of 709 (2%) patients, with an incidence of 0% to 7.1% for different devices (Table 2). The difference between the Amplatzer (0%) versus the CardioSEAL (7.1%), the StarFLEX (5.7%), and the PFO-Star (6.6%) was significant (p < 0.05, Fisher exact test) (Table 2). The Amplatzer had the lowest incidence of thrombi against each of those devices. Post-procedure AF (n = 4; p < 0.05) and persistent ASA (n = 4; p < 0.01) were significant predictors for thrombus formation determined by the Fisher exact test. These and other possible risk factors for thrombus formation are summarized in Table 3. Pre-thrombotic disorders were found in two PFO patients (thrombocytosis in one patient and hyperactivity of factor VIII in one patient). Two minor strokes caused by a left-sided mobile thrombus on a StarFLEX device (one of these patients had a hyperactivity of factor VIII). One minor stroke (patient with a thrombocytosis) and one TIA occurred in two PFO-Star patients. No other pre-thrombotic disorders were found in these 20 thrombi patients. Of the 20 patients, 17 had a good outcome with disappearance of the thrombus following anticoagulation therapy with heparin (n = 1), warfarin (n = 12), or both (n = 4). Repetitive TEE examinations showed that the thrombotic material on the device was completely resolved in 11 of 20 cases after one month and in 6 of 20 cases after a time period of between two and six months. In three patients the thrombus was removed surgically: - Patient 1: A left-sided mobile thrombus on a StarFLEX device was removed together with the device after cerebral embolism and unsuccessful fibrinolysis.
- Patient 2: A large right-sided thrombus (30 x 18 mm) attached to the lateral atrial wall did not disappear under long-term warfarin treatment and was removed surgically. The ASDOS occluder was left in place. Only a tip of wire sticking out of the covered device had been cut (Fig. 2).
- Patient 3: In this asymptomatic patient diagnosed with AF and four small mobile left-sided thrombi seen on TEE on the StarFLEX occluder, the referring physician decided to send the patient to surgery. During surgery, the left-sided thrombus could not be found anymore, but an 8-mm right-sided thrombus not diagnosed before was detected and removed together with the device (Fig. 3).

View larger version (73K):
[in this window]
[in a new window]
|
Figure 2 Unsuccessful resolution of a large right-sided thrombus (30 x 18 mm) attached to the lateral atrial wall under long-term warfarin treatment. During surgery the ASDOS occluder was left in place. Only a tip of wire sticking out of the covered device had been cut.
|
|

View larger version (77K):
[in this window]
[in a new window]
|
Figure 3 A short transesophageal view of small mobile left-sided thrombi on a StarFLEX occluder. During surgery, the absence of the left-sided thrombi but detection of a right-sided thrombus (8 mm) not diagnosed before was removed together with the device. AO = aorta; LA = left atrium; RA = right atrium.
|
|
 |
Discussion
|
|---|
Atrial thrombi after transcatheter ASD and PFO closure are rare complications (20,24,25). We investigated the incidence of thrombus formation on nine different devices in 1,000 consecutive patients to determine possible predictors. A thrombus formation was found in 5 of the 407 (1.2%) ASD patients and in 15 of the 593 (2.5%) PFO patients (p = NS). La Rosee et al. (23) described thrombus formation in 3 of the 38 (10.5%) ASD patients and in 8 of the 60 (13.3%) PFO patients (p = NS). These were diagnosed early, two to three days after implantation. The investigators found a thrombotic structure in 7 of the 40 (17.5%) patients on the ASDOS occluder, in 3 of the 33 (9%) patients on the PFO-Star occluder, and in 1 of the 25 (4%) patients on the Amplatzer occluder (p = NS). In contrast to La Rosee et al. (23), we found atrial thrombi after four weeks only in 1 of the 28 (3.6%) ASDOS devices, 5 of the 76 (6.6%) PFO-Star devices, and 0 of the 328 (0%) Amplatzer devices. Other previous reports revealed a thrombus incidence of 6.5% (9 of 139) in the ASDOS device (17) and 11.1% (3 of 27) in the Buttoned device (26). According to the data of nine different devices implanted in our center, we verified that thrombus formation is not only a clinical event of older devices; newer devices also demonstrate the potential to develop early thrombus formation after four weeks: 1 of 14 (7.1%) CardioSEAL, 6 of 105 (5.7%) StarFLEX, 5 of 76 (6.6%) PFO-Star (1st and 2nd generation), and 1 of 123 (0.8%) Helex.
Furthermore, we found that the difference between the Amplatzer device (0%) on one hand and the CardioSEAL device (7.1%), the StarFLEX device (5.7%), and the PFO-Star device (6.6%) on the other hand was significant (p < 0.05). The Amplatzer nitinol wire frame filled with polyester fabric and the Helex nitinol wire covered by an ultra thin membrane of expanded polytetrafluoroethylene seem to be less thrombogenic than the metallic framework and Dacron fabric in the CardioSEAL and StarFLEX devices and the nitinol framework and Ivalon foam in the PFO-Star occluder. Uncoated metal arms in the CardioSEAL and StarFLEX devices and outside nitinol arms in the PFO-Star devices (1st and 2nd generation) may increase the risk for excessive thrombus formation. This has also been affirmed by Braun et al. (19). In their largest PFO-Star device study, adherent thrombus formations could only be detected in generations 1 and 2 when expanded nitinol arms were located on the outer side of the foam. In the 3rd generation, the PFO-Star device nitinol arms were placed from the outer to the inner side of the left atrial foam. After these changes, no further thrombi were noted by TEE.
Possible predictors of thrombi.
After the procedure we found paroxysmal AF and persistent ASA to be significant predictors for thrombus formation (Table 3). The prevalence of AF and persistent ASA was higher among patients with thrombus formation than in those without (AF: 20% vs. 6.2%, p < 0.05; ASA: 20% vs. 1.3% p < 0.001). However, this has to be addressed carefully because of the small number of thrombus formations (n = 20) detected in our study. Moreover, there is some evidence that patients with ASA benefit from occlusion in reducing the mobility of the septum. In a previous study (27), the septal excursion had been significantly reduced in 48 of 51 patients six months after closure. A persistent ASA could be found in only 3 of 51 patients. None of them developed an atrial thrombus post-procedure. In our thrombus patients with persistent ASA, no recurrent thromboembolic events occurred. Windecker et al. (28) affirmed that after transcatheter closure the presence of ASA and PFO does not result in a higher recurrence rate of cerebrovascular events; thus, this group of patients would particularly benefit from closure.
No other significant predictors for thrombus formation on several devices could be found. A post-procedural shunt was detected in 3 of the 20 (15%) thrombus patients. A wire device frame fracture was also observed in 3 of 20 (15%) patients: 1 ASDOS, 1 CardioSEAL, and 1 PFO-Star occluder. In all three patients the side of the wire fracture was consistent with the side of the atrial thrombus. These patients all had right-sided atrial thrombi attached to right-sided broken arms or frames. In a study with the Buttoned device, Lambert et al. (26) noted that tilting the occluder created a large residual defect and was associated with left atrial thrombosis. Of the 27 Buttoned devices displaced in that way, 3 (11%) developed thrombus formation. Coagulation disorders such as protein C deficiency, protein S deficiency, and activated protein C resistance did not exist in our thrombus cases; only a hyperactivity of factor VIII and a thrombocytosis could be found in two patients. Gastmann et al. (24) reported a case of a left-sided thrombus formation on an ASDOS device resulting in cerebral thromboembolism. This patient had a deficiency of coagulation factor XII.
Other risk factors for cardiovascular disease, such as diabetes, coronary artery disease, and hypertension may not conduct thrombus formation on implanted devices. Among patients with (n = 20) and without (n = 980) thrombus formation, no significant differences could be determined in terms of age or gender. According to different study protocols, ASD and PFO patients received different medical prophylaxes after catheter closure either with antiplatelet agents or with anticoagulants (warfarin: n = 98, aspirin: n = 511, aspirin + clopidogrel: n = 391). Our results do not favor any medical regimen. Patients without thrombi (n = 980) were treated with warfarin in 10% of cases (n = 95), with aspirin in 52% of cases (n = 505), and with aspirin and clopidogrel in 39% of cases (n = 380), compared to patients with thrombi (n = 20), who received warfarin in 15% of cases (n = 3), aspirin in 30% of cases (n = 6), and aspirin + clopidogrel in 55% of cases (n = 11). However, these differences were not significant.
There are no additional studies comparing these three therapeutic regimes relating to the incidence of thrombi formations on devices. In a PFO-Star device study of 276 patients, Braun et al. (19) found that ever since change in the antiplatelet therapy from aspirin alone to a combination of aspirin and clopidogrel, no further thrombotic material occurred. Unfortunately, no statistical evaluation comparing both therapeutic regimes concerning the incidence of device-adherent thrombus was made.
In a smaller Amplatzer device study by Brandt et al. (29), 37 patients (ASD: n = 21; PFO: n = 16) received clopidogrel and aspirin after transcatheter closure for six months. In an absence of any thrombus at four weeks' and six months' TEE, the investigators concluded that this antiplatelet therapy is safe and effective in preventing thrombus formation on the septal occluder surface. However, they did not have a control group who received aspirin alone. In our study, uncertainty remains regarding the combination of aspirin + clopidogrel because there seemed to be no significant benefit in this particular thrombotic prophylaxis. Of 391 patients with this treatment, 11 (2.8%) developed thrombi after transcatheter closure compared to the patient group, who received only aspirin and detected thrombi in 6 of 511 (1.2%) patients (p = NS). It has to be mentioned that heparin was antagonized with protamine immediately after the procedure in 267 of 391 (68%) patients and 10 of 11 (91%) thrombus cases who received aspirin and clopidogrel. Anti-heparin therapy with protamine in the first hours after implantation might be a disadvantage and thus enhance thrombogenesis. Therefore, we stopped antagonizing heparin in November 2001. Since then only one thrombus occurred in 183 patients. In all 183 patients a four-week TEE had been performed. However, this has not yet been statistically significant.
In general, resolution of thrombi had been achieved under medical treatment in 17 of 20 patients within four weeks to six months; 12 patients received warfarin, 1 patient received heparin, and 4 patients received a combination of heparin and warfarin. In 11 of 17 patients, the thrombus disappeared after four weeks. Therefore, we currently recommend initial anticoagulation therapy for three months, similar to other studies (19,20) with repetitive TEE examinations depending upon diameter and mobility of device-attached thrombi.
Complications related to thrombus formations.
Cardiac surgery was performed in only three patients. One left-sided and one right-sided mobile thrombus attached to a StarFLEX occluder in two patients was removed together with the device, after cerebral embolism and unsuccessful fibrinolysis in one and as decided by the referring physician in the other. In the third patient a large right-sided thrombusnot device-relatedattached to the lateral atrial wall did not disappear under long-term warfarin treatment. This thrombus was removed surgically, leaving the ASDOS occluder in place. No further thrombi or other complications occurred in these three patients.
During follow-up, 4 of 20 (20%) patients became symptomatic and suffered from thromboembolic events. Two minor strokes occurred in patients with a StarFLEX occluder, and one minor stroke and one TIA occurred in two PFO-Star device patients. There have been other reports about left-sided thrombus formation on ASDOS (24,25) and Buttoned (26) devices complicated by systemic embolism. In the ASDOS device case, reported by Gastmann et al. (24), the patient had a deficiency of coagulation factor XII that is known to be associated with thrombotic complications. The large left-sided thrombus completely resolved after one month. In two of our four thrombus cases resulting in acute strokes, coagulation disorders were also found (hyperactivity of factor VIII in a patient with an ASD occluded with a StarFLEX and thrombocytosis in a patient with a PFO-Star occluder). All patients fully recovered from these neurological deficits.
If some patients endangered by thromboembolic complications nevertheless undergo transcatheter closure, anticoagulation monitoring requires exceptional attention. Similar to other investigators (20,24), we state that TTE is usually unable to detect thrombus formation on the device. According to our results we recommend that all adult patients, including those with good echocardiographic windows, should have a routine TEE examination after 4 weeks, 6 months, and 12 months. For younger children, follow-up by TTE as an imaging perspective might be sufficient. To shed light on the performed follow-up studies of these 1,000 consecutive patients, a group of them did not return for one of the TEE examinations at four weeks (n = 291) and at six months (n = 286), factors relating either to different device protocols or to personal declining. These patients were checked by interviews, clinical examination, electrocardiography, chest X-ray, and TTE at 1, 6, and 12 months following implantation. None of them suffered from a TIA or a stroke.
Thrombus formation and cerebrovascular events after surgical defect closure.
There is still an ongoing discussion (30,31) about the gold standard for the best therapeutic approach (catheter closure, medical therapy, or even surgery) in ASD and PFO patients. Surgical closure has also been attempted with mixed results in terms of recurrent embolic events derived by thrombus formation. Most recently, Rodriguez et al. (32) noted an intra-atrial thrombus with a small residual PFO six months after surgical PFO closure. Homma et al. (33) reported in a series of 28 patients, four recurrent neurological events (one stroke and three TIAs) after surgical closure at a mean follow-up of 19 months. In these particular patients (two were treated with aspirin and two received no medical treatment), no intracardiac mass or thrombus was seen by TEE, and only a small residual shunt was found in two cases. The recurrence only occurred in the older population (>45 years). Thus, the investigators concluded there must be some undefined causes for their patients' strokes.
In a larger study, Dearani et al. (34) documented eight TIAs during a mean follow-up of two years following surgical closure, one caused by temporal arteritis. Using TEE they also demonstrated that all closures had been intact without thrombotic structures. Having multiple embolic events before PFO closure was the only significant risk factor for recurrence after closure.
In a different study, Konstantinides et al. (35) found AF in six of nine patients who suffered from a recurrent stroke or TIA after surgical ASD closure. Only two of the six patients with AF received anticoagulants at the time of the event, that occurred in the late postoperative phase (>30 days after surgery). Reviewing additional surgical closure data it has to be mentioned that there are surgical studies (36,37) with excellent results and no report of neurological recurrences following patients who were embolic free up to an average of 24 and 41 months. However, prospective and randomized clinical trials are warranted to compare percutaneous closure directly with other treatment strategies such as medical therapy and surgical repair of intra-atrial shunts.
Conclusions.
The incidence of thrombus on closure devices is low. We found significant differences between different devices. Atrial fibrillation and persistent ASA after transcatheter closure are significant risk factors for thrombus formation. In most of the patients the thrombus resolved under medical therapy without clinical consequences.
 |
References
|
|---|
1. King TD, Mills NL. Nonoperative closure of atrial septal defects. Surgery. 1974;75:383388[Medline]
2. King TD, Thompson SL, Steiner C, et al. Secundum atrial septal defect nonoperative closure during cardiac catheterization. JAMA. 1976;235:25062509[Abstract/Free Full Text]
3. Sievert H, Ensslen R, Fach A, et al. Transcatheter closure of patent ductus arteriosus with the Rashkind occluder: acute results and angiographic follow-up in adults. Eur Heart J. 1997;18:10141018[Abstract/Free Full Text]
4. Berger F, Ewert P, Björnstad PG, et al. Transcatheter closure as standard treatment for most interatrial defects: experience in 200 patients treated with the Amplatzer septal occluder. Cardiol Young. 1999;9:468473[Medline]
5. Chan KC, Godman MJ, Walsh K, et al. Transcatheter closure of atrial septal defect and interatrial communications with a new self expanding nitinol double disc device (Amplatzer septal occluder): multicentre UK experience. Heart. 1999;82:300306[Abstract/Free Full Text]
6. Kort HW, Balzer DT, Johnson MC, et al. Resolution of right heart enlargement after closure of secundum atrial septal defect with transcatheter technique. J Am Coll Cardiol. 2001;38:15281532[Abstract/Free Full Text]
7. Han Y, Gu X, Titus JL, Rickers C, et al. New self-expanding patent foramen ovale occlusion device. Catheter Cardiovasc Interv. 1999;47:370376[CrossRef][Medline]
8. Hausdorf G, Schneider M, Franzbach B, et al. Transcatheter closure of secundum atrial septal defects with the atrial septal occlusion system (ASDOS): initial experience in children. Heart. 1996;75:8388[Abstract/Free Full Text]
9. Hausdorf G, Kaulitz R, Paul T, et al. Transcatheter closure of atrial septal defect with a new flexible, self-centering device (the STARFlex occluder). Am J Cardiol. 1999;84:11131116[CrossRef][Medline]
10. Mullen MJ, Dias BF, Walker F, et al. Intracardiac echocardiography guided device closure of atrial septal defects. J Am Coll Cardiol. 2003;41:285292[Abstract/Free Full Text]
11. Kreutzer J, Ryan CA, Wright JA, et al. Acute animal studies of the STARFlex system: a new self-centering CardioSeal septal occluder. Catheter Cardiovasc Interv. 2000;49:225233[CrossRef][Medline]
12. Lock JE, Rome JJ, Davis R, et al. Transcatheter closure of atrial septal defectsexperimental studies. Circulation. 1989;79:10911099[Abstract/Free Full Text]
13. Magni G, Hijazi Z, Pandian NG, et al. Two- and three-dimensional transesophageal echocardiography in patient selection and assessment of atrial septal defect closure by the new DAS-Angel Wings device: initial experience. Circulation. 1997;96:17221728[Abstract/Free Full Text]
14. Veldtman GR, Razack V, Siu S, et al. Right ventricular form and function after percutaneous atrial septal defect device closure. J Am Coll Cardiol. 2001;37:21082113[Abstract/Free Full Text]
15. Sharafuddin MJ, Gu X, Titus JL, et al. Transvenous closure of secundum atrial septal defects: preliminary results with a new self-expanding nitinol prosthesis in a swine model. Circulation. 1997;95:21622168[Abstract/Free Full Text]
16. Sideris EB, Sideris SE, Fowlkes P, et al. Transvenous atrial septal defect occlusion in piglets with the "buttoned" double-disc device. Circulation. 1990;81:312318[Abstract/Free Full Text]
17. Sievert H, Babic UU, Hausdorf G, et al. Transcatheter closure of atrial septal defect and patent foramen ovale with the ASDOS device (a multi-institutional European trial). Am J Cardiol. 1998;82:14051413[CrossRef][Medline]
18. Zahn EM, Wilson N, Cutright W, Latson LA, et al. Development and testing of the Helex septal occluder, a new expanded polytetrafluoroethylene atrial septal defect occlusion system. Circulation. 2001;104:711716[Abstract/Free Full Text]
19. Braun M, Fassbender D, Schoen S, et al. Transcatheter closure of patent foramen ovale in patients with cerebral ischemia. J Am Coll Cardiol. 2002;39:20192025[Abstract/Free Full Text]
20. Cooke J, Gelman J, Menahem S, et al. Thrombus on an ASD closure device: a call for caution. Heart, Lung and Circulation. 2000;9:3031[CrossRef]
21. Pinto F, Sousa L, Fragata J, et al. Late cardiac tamponade after transcatheter closure of atrial septal defect with CardioSeal device. Cardiol Young. 2001;11:233235[CrossRef][Medline]
22. Chessa M, Carminati M, Butera G, et al. Early and late complications associated with transcatheter occlusion of secundum atrial septal defect. J Am Coll Cardiol. 2002;39:10611065[Abstract/Free Full Text]
23. La Rosee K, Krause D, Becker M, et al. Transcatheter closure of atrial septal defect in adults: practicability and safety of four different occluder systems used in 102 patients. Dtsch Med Wochenschr. 2001;126:10301036[CrossRef][Medline]
24. Gastmann O, Werner GS, Babic UU, et al. Thrombus formation on transcatheter ASD occluder device in a patient with coagulation factor XII deficiency. Cathet Cardiovasc Diagn. 1998;43:8183[CrossRef][Medline]
25. Khatchatourov G, Kalangos A, Anwar A, et al. Massive thromboembolism due to transcatheter ASD closure with ASDOS device. J Invasive Cardiol. 1999;11:743745[Medline]
26. Lambert V, Losay J, Piot JD, et al. Late complications of percutaneous closure of atrial septal defects with the Sideris occluder. Arch Mal Coeur Vaiss. 1997;90:245251[Medline]
27. Krumsdorf U, Keppeler P, Horvath K, et al. Catheter closure of atrial septal defects and patent foramen ovale in patients with an atrial septal aneurysm using different devices. J Interv Cardiol. 2001;14:4956[Medline]
28. Windecker S, Wahl A, Chatterjee T, et al. Percutaneous closure of patent foramen ovale in patients with paradoxical embolism: long-term risk of recurrent thromboembolic events. Circulation. 2000;101:893898[Abstract/Free Full Text]
29. Brandt RR, Neumann T, Neuzner J, et al. Transcatheter closure of atrial septal defect and patent foramen ovale in adult patients using the Amplatzer occlusion device: no evidence or thrombus deposition with antiplatelet agents. J Am Soc Echocardiogr. 2002;15:10941098[CrossRef][Medline]
30. Du ZD, Hijazi ZM, Kleinman CS, et al. Comparison between transcatheter and surgical closure of secundum atrial septal defect in children and adults: results of a multicenter nonrandomized trial. J Am Coll Cardiol. 2002;39:18361844[Abstract/Free Full Text]
31. Formigari R, Di Donato RM, Mazzera E, et al. Minimally invasive or interventional repair of atrial septal defects in children: experience in 171 cases and comparison with conventional strategies. J Am Coll Cardiol. 2001;37:17071712[Abstract/Free Full Text]
32. Rodriguez CJ, Di Tullio MR, Sacco RL, et al. Inter-atrial thrombus after surgical closure of patent foramen ovale. J Am Soc Echocardiogr. 2001;14:6366[CrossRef][Medline]
33. Homma S, Di Tullio M, Sacco RL, et al. Surgical closure of patent foramen ovale in cryptogenic stroke patients. Stroke. 1997;28:23762381[Abstract/Free Full Text]
34. Dearani JA, Ugurlu BS, Danielsen GK, et al. Surgical patent foramen ovale closure for prevention of paradoxical embolism-related cerebrovascular ischemic events. Circulation. 1999;100(Suppl):II171, 175
35. Konstantinides S, Geibel A, Olschewwski M, et al. A comparison of surgical and medical therapy for atrial septal defect in adults. N Engl J Med. 1995;333:469473[Abstract/Free Full Text]
36. Hanna JP, Sun JP, Furlan AJ, et al. Patent foramen ovale and brain infarct: echocardiographic predictors, recurrence and prevention. Stroke. 1994;25:782786[Abstract]
37. Devuyst G, Bogousslavsky J, Ruchat P, et al. Prognosis after stroke followed by surgical closure of patent foramen ovale: a prospective follow-up study with brain MRI and simultaneous transesophageal and transcranial Doppler ultrasound. Neurology. 1996;47:11621166[Abstract/Free Full Text]
This article has been cited by other articles:

|
 |

|
 |
 
A. A. Alsheikh-Ali, D. E. Thaler, and D. M. Kent
Patent Foramen Ovale in Cryptogenic Stroke: Incidental or Pathogenic?
Stroke,
July 1, 2009;
40(7):
2349 - 2355.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Wohrle, M. Kochs, J. Spiess, T. Nusser, V. Hombach, and N. Merkle
Impact of Percutaneous Device Implantation for Closure of Patent Foramen Ovale on Valve Insufficiencies
Circulation,
June 16, 2009;
119(23):
3002 - 3008.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. J. Dibardino, D. B. McElhinney, A. K. Kaza, and J. E. Mayer Jr.
Analysis of the US Food and Drug Administration Manufacturer and User Facility Device Experience database for adverse events involving Amplatzer septal occluder devices and comparison with the Society of Thoracic Surgery congenital cardiac surgery database.
J. Thorac. Cardiovasc. Surg.,
June 1, 2009;
137(6):
1334 - 1341.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A.-R. Fathi, P. Eshtehardi, and B. Meier
Patent foramen ovale and neurosurgery in sitting position: a systematic review
Br. J. Anaesth.,
May 1, 2009;
102(5):
588 - 596.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Pelliccia, C. Cianfrocca, V. Pasceri, A. Granatelli, G. Speciale, A. Roncella, M. Turturo, G. Richichi, and C. Pristipino
Transcathether occlusion of interatrial communications: postprocedural transoesophageal echocardiography allows timely detection and treatment of intracardiac thrombus formation
Eur J Echocardiogr,
May 1, 2009;
10(3):
439 - 441.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. Meier
Stroke and migraine: a cardiologist's headache
Heart,
April 1, 2009;
95(7):
595 - 602.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. Yared, A. L. Baggish, J. Solis, R. Durst, J. J. Passeri, I. F. Palacios, and M. H. Picard
Echocardiographic Assessment of Percutaneous Patent Foramen Ovale and Atrial Septal Defect Closure Complications
Circ Cardiovasc Imaging,
March 1, 2009;
2(2):
141 - 149.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Wahl, T. Tai, F. Praz, M. Schwerzmann, C. Seiler, K. Nedeltchev, S. Windecker, H. P. Mattle, and B. Meier
Late Results After Percutaneous Closure of Patent Foramen Ovale for Secondary Prevention of Paradoxical Embolism Using the Amplatzer PFO Occluder Without Intraprocedural Echocardiography: Effect of Device Size
J. Am. Coll. Cardiol. Intv.,
February 1, 2009;
2(2):
116 - 123.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A Wahl and B Meier
Patent foramen ovale and ventricular septal defect closure
Heart,
January 1, 2009;
95(1):
70 - 82.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Windecker and B. Meier
Patent Foramen Ovale and Cryptogenic Stroke: To Close or Not to Close? Closure: What Else!
Circulation,
November 4, 2008;
118(19):
1989 - 1997.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. Hongxin, S. Lijun, Z.-j. Wang, J. Zi, W.-l. Zhang, H.-z. Zhang, G. Wenbin, and C.-w. Zou
Intraoperative device closure of large secundum atrial septal defects; a safe alternative to transcatheter closure
Eur. J. Cardiothorac. Surg.,
June 1, 2008;
33(6):
1055 - 1060.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. R. Raghuram, R. Krishnan, S. Kumar, and K. Balamurugan
Complications in atrial septal defect device closure
Interactive CardioVascular and Thoracic Surgery,
February 1, 2008;
7(1):
167 - 169.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Ritzema, I. C. Melton, A. M. Richards, I. G. Crozier, C. Frampton, R. N. Doughty, J. Whiting, S. Kar, N. Eigler, H. Krum, et al.
Direct Left Atrial Pressure Monitoring in Ambulatory Heart Failure Patients: Initial Experience With a New Permanent Implantable Device
Circulation,
December 18, 2007;
116(25):
2952 - 2959.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. K. Mareedu, M. S. Shah, J. E. Mesa, and C. S. McCauley
Percutaneous Closure of Patent Foramen Ovale: A Case Series and Literature Review
Clin. Med. Res.,
December 1, 2007;
5(4):
218 - 226.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. Ak, T. Aybek, G. Wimmer-Greinecker, F. Ozaslan, F. Bakhtiary, A. Moritz, and S. Dogan
Evolution of surgical techniques for atrial septal defect repair in adults: A 10-year single-institution experience
J. Thorac. Cardiovasc. Surg.,
September 1, 2007;
134(3):
757 - 764.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. R. Sukernik and E. Bennett-Guerrero
The Incidental Finding of a Patent Foramen Ovale During Cardiac Surgery: Should It Always Be Repaired? A Core Review
Anesth. Analg.,
September 1, 2007;
105(3):
602 - 610.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Hara, T. K. Jones, E. R. Ladich, R. Virmani, D. C. Auth, J. E. Eichinger, R. J. Sommer, R. A. Van Tassel, and R. S. Schwartz
Patent Foramen Ovale Closure by Radiofrequency Thermal Coaptation: First Experience in the Porcine Model and Healing Mechanisms Over Time
Circulation,
August 7, 2007;
116(6):
648 - 653.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. Divchev, E. K. Podewski, M. Mengel, G. P. Meyer, H. Drexler, and A. Schaefer
Inflammatory, abscess-forming foreign body reaction mimics a thrombus formation on an atrial septal defect closure device: A commented case report
Eur J Echocardiogr,
August 1, 2007;
8(4):
298 - 302.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Giardini and A. Donti
Enhanced Thrombogenesis is Physiologic After Transcatheter Closure of Patent Foramen Ovale
Stroke,
July 1, 2007;
38(7):
e55 - e55.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Rodes-Cabau, E. Bedard, O. F. Bertrand, and C. Houde
Response to Letter by Giardini and Donti
Stroke,
July 1, 2007;
38(7):
e56 - e56.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. Krieg, H. Lapp, K. Pethig, J. F. Gummert, and T. Bossert
Removal of a Left Atrial Thrombus Adherent to a Patent Foramen Ovale Occluder
Ann. Thorac. Surg.,
April 1, 2007;
83(4):
1539 - 1541.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Sigler and C. Jux
Biocompatibility of septal defect closure devices
Heart,
April 1, 2007;
93(4):
444 - 449.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. Divchev, A. Schaefer, M. Fuchs, T. Breymann, H. Drexler, and G. P. Meyer
Thrombus formation on an atrial septal defect closure device: A case report and review of the literature
Eur J Echocardiogr,
January 1, 2007;
8(1):
53 - 56.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E. Bedard, J. Rodes-Cabau, C. Houde, A. Mackey, D. Rivest, S. Cloutier, M. Noel, A. Marrero, J.-M. Cote, P. Chetaille, et al.
Enhanced Thrombogenesis but Not Platelet Activation Is Associated With Transcatheter Closure of Patent Foramen Ovale in Patients With Cryptogenic Stroke
Stroke,
January 1, 2007;
38(1):
100 - 104.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. C. Post and W. Budts
The Relationship Between Migraine and Right-to-Left Shunt: Fact or Fiction?
Chest,
September 1, 2006;
130(3):
896 - 901.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Jux, H. Bertram, P. Wohlsein, M. Bruegmann, and T. Paul
Interventional Atrial Septal Defect Closure Using a Totally Bioresorbable Occluder Matrix: Development and Preclinical Evaluation of the BioSTAR Device
J. Am. Coll. Cardiol.,
July 4, 2006;
48(1):
161 - 169.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Spies, R. Strasheim, I. Timmermanns, and R. Schraeder
Patent foramen ovale closure in patients with cryptogenic thrombo-embolic events using the Cardia PFO occluder
Eur. Heart J.,
February 1, 2006;
27(3):
365 - 371.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. D.R. Thomson
Percutaneous PFO closure, further data but many unanswered questions
Eur. Heart J.,
February 1, 2006;
27(3):
258 - 259.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
V. L. Vida, J. Barnoya, M. O'Connell, J. Leon-Wyss, L. A. Larrazabal, and A. R. Castaneda
Surgical Versus Percutaneous Occlusion of Ostium Secundum Atrial Septal Defects: Results and Cost-Effective Considerations in a Low-Income Country
J. Am. Coll. Cardiol.,
January 17, 2006;
47(2):
326 - 331.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Schwerzmann and O. Salehian
Hazards of percutaneous PFO closure
Eur J Echocardiogr,
December 1, 2005;
6(6):
393 - 395.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M S Spence and S A Qureshi
Complications of transcatheter closure of atrial septal defects
Heart,
December 1, 2005;
91(12):
1512 - 1514.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Hara, R. Virmani, E. Ladich, S. Mackey-Bojack, J. Titus, M. Reisman, W. Gray, M. Nakamura, M. Mooney, A. Poulose, et al.
Patent Foramen Ovale: Current Pathology, Pathophysiology, and Clinical Status
J. Am. Coll. Cardiol.,
November 1, 2005;
46(9):
1768 - 1776.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Ruge, S.M. Wildhirt, P. Libera, M. Vogt, K. Holper, and R. Lange
Left Atrial Thrombus on Atrial Septal Defect Closure Device as a Source of Cerebral Emboli 3 Years After Implantation
Circulation,
September 6, 2005;
112(10):
e130 - e131.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. W. Schuchlenz, S. Mannweiler, and D. Martin
Marked foreign body reaction and thrombus formation after transcatheter closure of a patent foramen ovale
J. Thorac. Cardiovasc. Surg.,
August 1, 2005;
130(2):
591 - 592.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
W. H. Maisel and W. K. Laskey
Patent Foramen Ovale Closure Devices: Moving Beyond Equipoise
JAMA,
July 20, 2005;
294(3):
366 - 369.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
W. Budts
ASD closure for migraine: is there a scientific basis?: reply
Eur. Heart J.,
July 2, 2005;
26(14):
1446 - 1447.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B Meier
Closure of patent foramen ovale: technique, pitfalls, complications, and follow up
Heart,
April 1, 2005;
91(4):
444 - 448.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F A Flachskampf and W G Daniel
Closure of patent foramen ovale: is the case really closed as well?
Heart,
April 1, 2005;
91(4):
449 - 450.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Masura, P. Gavora, and T. Podnar
Long-term outcome of transcatheter secundum-type atrial septal defect closure using Amplatzer septal occluders
J. Am. Coll. Cardiol.,
February 15, 2005;
45(4):
505 - 507.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. N. DeMaria, O. Ben-Yehuda, D. Berman, G. K. Feld, B. H. Greenberg, J. D. Knoke, K. U. Knowlton, W. Y.W. Lew, J. Narula, D. Sahn, et al.
Highlights of the year in JACC 2004
J. Am. Coll. Cardiol.,
January 4, 2005;
45(1):
137 - 153.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. W. Schuchlenz, G.P. Anzola, E. Morandi, F. Casilli, and E. Onorato
Contrast Ultrasound Techniques in the Detection and Quantification of Patent Foramen Ovale: Myth Versus Reality * Response:
Stroke,
December 1, 2004;
35(12):
2755 - 2756.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H.P. Kuhl and P. Hanrath
The impact of transesophageal echocardiography on daily clinical practice
Eur J Echocardiogr,
December 1, 2004;
5(6):
455 - 468.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
W. T. Mahle, K. Lundine, K. R. Kanter, J. M. Forbess, P. Kirshbom, S. R. Tosone, and R. N. Vincent
The short term effects of cardiopulmonary bypass on neurologic function in children and young adults
Eur. J. Cardiothorac. Surg.,
November 1, 2004;
26(5):
920 - 925.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Massimo, B. Gianfranco, and C. Mario
Risk of thrombus formation on devices used to close transcatheter atrial septal defect and patent foramen ovale
J. Am. Coll. Cardiol.,
October 19, 2004;
44(8):
1712 - 1712.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. J. Landzberg, L. Brass, A. Furlan, W. Hellenbrand, H. Hermann, P. Kramer, I. F. Palacios, M. Reisman, R. Sommer, C. Thomas, et al.
Transcatheter intracardiac device implantation for atrial level defects and thrombosis: A call for randomized, controlled data
J. Am. Coll. Cardiol.,
October 19, 2004;
44(8):
1713 - 1714.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. Schrader
Incidence and clinical course of thrombus formation on atrial septal defect and patent foramen ovale closure devices
J. Am. Coll. Cardiol.,
October 19, 2004;
44(8):
1714 - 1714.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Jux and H. Bertram
Thrombus formation on intracardiac devices: A complex issue
J. Am. Coll. Cardiol.,
October 19, 2004;
44(8):
1712 - 1713.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
U. Krumsdorf, S. Ostermayer, K. Billinger, T. Trepels, E. Zadan, K. Horvath, and H. Sievert
Reply
J. Am. Coll. Cardiol.,
October 19, 2004;
44(8):
1714 - 1716.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. L. Blackshear
Closure of patent foramen ovale in cryptogenic stroke: Ready or not, here come the trials
J. Am. Coll. Cardiol.,
August 18, 2004;
44(4):
759 - 761.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. P. Graham Jr
The year in congenital heart disease
J. Am. Coll. Cardiol.,
June 2, 2004;
43(11):
2132 - 2141.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Thrombus Formation on Interatrial Septal Closure Devices
Journal Watch Cardiology,
March 26, 2004;
2004(326):
3 - 3.
[Full Text]
|
 |
|
|