J Am Coll Cardiol, 2003; 42:1900-1905, doi:10.1016/j.jacc.2003.06.012 © 2003 by the American College of Cardiology Foundation |
* Cardiovascular Research Foundation, Lenox Hill Heart and Vascular Institute, New York, New York, USA
Manuscript received March 19, 2003; revised manuscript received June 1, 2003, accepted June 3, 2003.
* Reprint requests and correspondence: Dr. George Dangas, Lenox Hill Heart and Vascular Institute, Cardiovascular Research Foundation, 55 East 59th Street, 6th Floor, New York, New York 10022, USA.
gdangas{at}crf.org
This study was presented at the 52nd Annual Scientific Session of the American College of Cardiology, Atlanta, Georgia, March 2002.
| Abstract |
|---|
|
|
|---|
BACKGROUND: Elevation of CK-MB after percutaneous coronary interventions has been associated with late mortality.
METHODS: We identified 989 consecutive patients who underwent intravascular ultrasound-guided stenting of 1,015 coronary lesions. Patients were divided into three groups according to stent expansion, defined as the ratio of final lumen over the reference lumen cross-sectional areas: Group 1 (ratio <70%, n = 117 patients with 126 lesions); Group 2 (ratio 70% to 100%, n = 551 patients with 562 lesions); Group 3 (ratio >100%, n = 321 patients with 327 lesions).
RESULTS: The peak CK-MB values increased significantly with increasing stent expansion: CK-MB = 3 to 5x normal occurred 16%, 18%, and 25% in Groups 1, 2, and 3, respectively, p = 0.02; CK-MB >5 times normal occurred 9%, 13%, and 16% respectively, p = 0.02. Conversely, at one year follow-up there was a stepwise decrease in target lesion revascularization (11% vs. 19% and 17%, respectively, p = 0.04) and major adverse cardiac events with increasing stent expansion. In addition, there was a trend toward lower mortality in Group 3 (9% vs. 4.4% vs. 4.0%, p = 0.07).
CONCLUSIONS: Intravascular ultrasound-guided stent overexpansion (final lumen greater than reference lumen cross-sectional area) is accompanied by a higher periprocedural CK-MB release but a lower target lesion revascularization and a trend toward lower mortality at one year. Increased periprocedural CK-MB release appears as a trade-off for optimal stent implantation and lower clinical restenosis.
| ||||||||||||||||||
Several intravascular ultrasound (IVUS) criteria have been used to optimize stent expansion in an attempt to reduce clinical and angiographic restenosis. The purpose of the current study was to determine the relationship between stent expansion, as assessed by IVUS, and both post-PCI CK-MB elevation and clinical restenosis.
| Methods |
|---|
|
|
|---|
Procedures. All procedures were performed after written informed consent. All patients underwent PCI according to current stent-implantation guidelines and techniques (10). All patients received aspirin 325 mg daily at least 24 h before the procedure and indefinitely afterwards. Patients were treated concomitantly with a thienopyridine: either ticlopidine 250 mg twice daily or clopidogrel 75 mg daily for four weeks. Weight-adjusted heparin was administered during the procedure and was routinely discontinued at the end of the procedure. Platelet glycoprotein IIb/IIIa inhibitors were used in <4% of cases.
Angiographic analysis. Cineangiograms were analyzed using a validated system (ARTEK, Quantitative Cardiac Systems, Ann Arbor, Michigan) by a dedicated core laboratory blinded to the IVUS and clinical data. Using the contrast-filled catheter as the calibration standard, minimal lumen diameter, reference diameter, and percent diameter stenosis before and after PCI were obtained from the single "worst" and least foreshortened view. The reference segment diameter was averaged from user-defined, 5-mm-long angiographically normal segments proximal and distal to the stenosis in the projection with the least amount of foreshortening. Ostial lesions were within 3 mm of the coronary ostia or <3 mm distal to a major proximal side branch. These are standard qualitative and quantitative analyses and definitions, and they have been published previously (11).
IVUS imaging and analysis. Intravascular ultrasound imaging was performed after administration of 0.2 mg of intracoronary nitroglycerin, using a system made by one of the following manufacturers: CVIS/Inter Therapy, Hewlett-Packard and Boston Scientific Corporation, or Cardiovascular Imaging Systems/Boston Scientific Corporation. The IVUS catheter was advanced approximately 10 mm distal to the lesion, the video recorder was turned on, the transducer and pullback device was activated, and the entire artery was imaged and recorded on 0.5-inch high-resolution s-VHS for off-line analysis.
Validations of normal coronary anatomy, plaque composition, and measurements using IVUS have been reported (12). Using computer planimetry (TapeMeasure, Indec Systems), the following lesion and reference measurements were made in diastole (13,14). The external elastic membrane (EEM) CSA was measured by tracing the leading edge of the adventitia. The lesion site was the cross-sectional slice with the smallest lumen; among sections with the same lumen area, the one with the most plaque was selected. If the plaque was "packed" around the catheter, the lumen was assumed to be the physical (not acoustical) size of the catheter. Because IVUS cannot measure media thickness accurately, plaque and media were the measure of plaque mass. Cross-sectional narrowing (plaque and media CSA divided by EEM CSA) has also been called the plaque burden or percent plaque area. The reference segment averaged the most visually normal cross-sections (largest lumen with least plaque) within 10 mm proximal and distal to the lesion but between major branches; a distal reference was used for ostial lesions. Percent stent expansion was measured as final stent CSA divided by the average of the proximal and distal reference lumen CSA. These IVUS methods have been validated and published previously (15,16).
Clinical definitions and follow-up. Pre-specified clinical and laboratory demographic information was obtained from hospital charts that were reviewed by independent research personnel who were unaware of the objectives of the study; data were entered prospectively into a database. Standard definitions included: 1) Q-wave MIthe presence of new pathologic Q-waves associated with an elevation of CK or CK-MB at least two times the upper normal value and 2) nonQ-wave MICK-MB elevation at least five times the upper normal value without new Q-waves.
CK-MB levels were systematically evaluated before intervention and at 6 h and 24 h after the procedure. If CK-MB levels were elevated, serial measurements were performed every 8 h and the peak level was recorded. All CK-MB determinations were performed in the clinical chemistry laboratory by the mass-determination method (normal range 0 to 4 ng/ml). For this study, only the peak CK-MB values post-PCI were used.
One-year clinical follow-up was performed by either telephone contact or office visit. Multiple adverse cardiac events (MACE) were recorded, including death (all-cause), Q-wave MI, and target lesion revascularization (TLR) (whether PCI or surgical). These data collection methods have been published previously (17).
Statistical analysis. Statistical analysis was performed using SAS (Statistical Analysis Systems, SAS Institute Inc, Cary, North Carolina) by a dedicated data analysis center. Continuous variables are expressed as mean ± 1 SD and categorical variables as frequency (%). Continuous variables were compared using analysis of variance or the unpaired Student t test. Categorical variables were compared with chi-squared statistics. A p value <0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Relatively low-level release of cardiac enzymes occurs after 6% to 30% of otherwise successful PCIs (68). It has also been shown that elevation of CK-MB is associated with increased late mortality (6,1821). However, the impact of CK-MB release on clinical outcome is rather small in the absence of new Q-waves or elevation less than eight times the upper limit of normal (20,2224). It has also been suggested that a larger post-PCI lumen dimension, attained after successful new device PCI, primarily determines subsequent cardiac events and is associated with lower TLR (25,26); thus, the potentially adverse influence of CK-MB elevation may be overshadowed by the other favorable outcomes after optimal PCI stent (22).
In the present study, the aggressive stent expansion strategy was associated with higher CK-MB elevation post PCI. It is of note that atheroablation was more frequent, and plaque and media were larger, in Group 1 (least stent expansion group). As previously reported, periprocedural enzyme release varies markedly with different devices, being most common with combined stenting and atheroablation (20). We have also suggested that the increased incidence of CK-MB elevation after stenting and atheroablation may simply reflect the baseline presence of more diffuse atherosclerotic disease and greater plaque burden (5).
Stent overexpansion had been considered elsewhere to result in an increased late loss or even higher restenosis rates (27). Indeed, we documented potentially greater vessel wall trauma, as suggested by the greater balloon-to-artery ratio in Group 3; however, this finding was not associated with an increased rate of late events. To the contrary, increased stent expansion correlated with reduced one-year event rates and, in particular, an arithmetically lower one-year mortality. This finding is in accordance with studies in which stent overexpansion resulted in larger final lumen dimensions that were maintained at follow-up, despite greater lumen loss due to exaggerated intimal hyperplasia in response to the greater vessel wall trauma (28,29).
It is of note that Group 1 lesions were longer than lesions in the other two groups. Conversely, Group 3 vessels were smaller, as suggested by the smaller EEM CSA. Longer lesion length and smaller vessel size are well-recognized risk factors for in-stent restenosis (30). There are two main determinants of the final stent CSA: vessel size and implantation technique (30). Intravascular ultrasound guidance can be used to optimize final stent implantation results, mainly because of bigger balloons (greater balloon-to-artery ratio, as seen in this study) or higher inflation pressures (30,31). After IVUS-guided overdilation, minimal stent lumen CSA and diameter have been shown to increase by as much as 11% to 80% (16,29,32). In the present study, we examined various IVUS-derived stent expansion cutoff values in relation to clinical restenosis and found that percent stent expansion >90% had the lowest one-year TLR rate.
Limitations. First, this was a retrospective analysis; in an attempt to offset this limitation, the data were collected prospectively by independent monitors and entered into a dedicated database, and independent laboratories interpreted all angiographic and IVUS studies. Second, complete follow-up was available in 92% of the whole patient population (89%, 92%, and 93% for Groups 1, 2, and 3, respectively). We did not use more sensitive markers of myocardial damage such as troponin T or I tests. However, measurement of CK-MB release is an accepted, widely used method of assessment of myocardial injury after PCI, used in a large number of previous studies (5,8,9,18,22). In addition, we do not report data on time of ischemia and information about side branch occlusion. Nevertheless, the absence of reporting these data does not alter the basic hypothesis of the current study.
Conclusions. Stent overexpansion (final lumen greater than reference lumen CSA) is accompanied by a higher periprocedural CK-MB release but a lower TLR and a trend toward lower mortality at one year. Increased periprocedural CK-MB release appears as a trade-off for optimal stent implantation and lower clinical restenosis and one-year MACE.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
T. Uetani, T. Amano, H. Ando, K. Yokoi, K. Arai, M. Kato, N. Marui, M. Nanki, T. Matsubara, H. Ishii, et al. The correlation between lipid volume in the target lesion, measured by integrated backscatter intravascular ultrasound, and post-procedural myocardial infarction in patients with elective stent implantation Eur. Heart J., July 2, 2008; 29(14): 1714 - 1720. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. K. Shah Distal Embolization After Percutaneous Coronary Interventions: Prediction, Prevention, and Relevance J. Am. Coll. Cardiol., October 23, 2007; 50(17): 1647 - 1648. [Full Text] [PDF] |
||||
![]() |
T. Kawamoto, H. Okura, Y. Koyama, I. Toda, H. Taguchi, K. Tamita, A. Yamamuro, Y. Yoshimura, Y. Neishi, E. Toyota, et al. The Relationship Between Coronary Plaque Characteristics and Small Embolic Particles During Coronary Stent Implantation J. Am. Coll. Cardiol., October 23, 2007; 50(17): 1635 - 1640. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. K. Steigen, M. Maeng, R. Wiseth, A. Erglis, I. Kumsars, I. Narbute, P. Gunnes, J. Mannsverk, O. Meyerdierks, S. Rotevatn, et al. Randomized Study on Simple Versus Complex Stenting of Coronary Artery Bifurcation Lesions: The Nordic Bifurcation Study Circulation, October 31, 2006; 114(18): 1955 - 1961. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Balian, M. Galli, C. Marcassa, G. Cecchin, M. Child, F. Barlocco, E. Petrucci, G. Filippini, R. Michi, and M. Onofri Intracoronary ST-Segment Shift Soon After Elective Percutaneous Coronary Intervention Accurately Predicts Periprocedural Myocardial Injury Circulation, October 31, 2006; 114(18): 1948 - 1954. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Herrmann Peri-procedural myocardial injury: 2005 update Eur. Heart J., December 1, 2005; 26(23): 2493 - 2519. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Babuin and A. S. Jaffe Troponin: the biomarker of choice for the detection of cardiac injury Can. Med. Assoc. J., November 8, 2005; 173(10): 1191 - 1202. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. L. Bhatt and E. J. Topol Periprocedural Cardiac Enzyme Elevation Predicts Adverse Outcomes Circulation, August 9, 2005; 112(6): 906 - 922. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |