VIEWPOINT
Choice of prosthetic heart valves:update for the next generation
David S. Bach, MD, FACC*,*
* Department of Medicine, Division of Cardiology, University of Michigan, Ann Arbor, Michigan, USA
Manuscript received April 3, 2003;
revised manuscript received April 30, 2003,
accepted June 3, 2003.
* Reprint requests and correspondence: Dr. David S. Bach, L3119 Women's0273, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109-0273, USA.
dbach{at}umich.edu
 |
Abstract
|
|---|
Intervention for valvular heart disease poses unique clinical challenges in cardiology because the diseases are of relatively low prevalence, the interventions do not lend themselves to randomized comparative trials, and important clinical end points are assessed only after decades of follow-up. In addition, continuing advances in prosthetic heart valve technology make follow-up a moving target because long-term data by definition are available only for older prostheses. Newer tissue and mechanical prostheses afford superior hemodynamics compared with their older counterparts, and data suggest that durability and patient mortality are superior with newer compared with older bioprostheses. Arbitrary cutoffs dictating valve choice based predominantly on patient age may not give appropriate weight to individual patient perspectives. In educating and counseling patients regarding choices in heart valve prostheses, the clinician should help the patient weigh the relative merits for the individual patient of projected mortality, valve durability, and requirement for anticoagulation, with associated freedom from re-operation, hemorrhagic and thromboembolic risk, and impact on lifestyle.
Valvular heart disease and heart valve surgery pose unique clinical challenges compared with other cardiovascular diseases. As examples, the treatment of congestive heart failure and acute coronary syndromes have been revolutionized in the past few years by large, multi-center, randomized, placebo-controlled trials with statistically (and clinically) significant end points available in months if not in a few years. Valvular heart disease and heart valve surgery, however, do not lend themselves to similar investigations. Valvular heart disease is far less prevalent and its clinical end points far more indolent; clinical outcomes associated with prosthetic heart valves (PHVs) are of interest decades later, rather than acutely after intervention; and heart valve surgery does not lend itself to randomization. Other than two trials conducted nearly 30 years ago, there are no (and are unlikely to be any) large-scale randomized comparisons between valve prostheses. In addition, advances in valve technology make the determination of valve durability a moving target: by nature, 15-year durability data exist for valves that now may be obsolete. For these reasons, clinicians are forced to draw conclusions and make recommendations based on incomplete information, extrapolating from limited data, clinical experience, and common sense. In short, for PHV choices, randomized data alone are insufficient to dictate clinical practice.
The recent review in the Journal (1) concerning choices in PHVs raises important issues with respect to how physicians should inform patients and recommend therapy when existing data are incomplete or imperfect. The review draws heavily on the randomized trials performed in the late 1970s (2,3). Based on the assessment that outcomes are the same for newer prostheses, the review perpetuates traditional recommendations that tissue valves are appropriate for old patients and mechanical valves are appropriate for all others. But much in medicine has changed in the ensuing decades since these data were derived, raising questions of how to extrapolate data from these trials to modern practice.
 |
New valves versus old valves
|
|---|
The Edinburgh heart valve trial (2) was conducted between 1975 and 1979; the Veterans Affairs trial (3) was conducted between 1977 and 1982. In these trials, late structural deterioration occurred more often with bioprostheses, which was felt to limit their usefulness among patients <65 years of age. Because there are no randomized trials comparing valve prostheses that enrolled patients after 1982, the clinician must draw conclusions about old versus new prostheses based on extrapolation of data.
In the decades since these trials were performed, the evolution of tissue technology has resulted in newer bioprostheses with superior durability. For example, the first-generation Hancock MO bioprosthesis (Medtronic, Inc., Minneapolis, Minnesota) was fixed at high pressure using no tissue treatment to mitigate calcification, whereas the second-generation Hancock II bioprosthesis used low-pressure fixation and an early form of tissue anticalicification treatment. The 15-year actuarial freedom from structural valve deterioration was 57 ± 4% for the first-generation valve in a population 69 years of age at the time of implantation (4) and 81 ± 5% for the second-generation valve in a population 65 years of age at the time of implantation (5). Although long-term data are lacking for yet newer devices, data from animal studies (6) and early human experience (79) suggest that third-generation bioprostheses will be even more durable compared with earlier devices. Significantly, case-controlled studies of newer versus older bioprostheses demonstrate statistically significant survival advantages associated with some newer bioprostheses (1013).
In addition to advances resulting in greater bioprosthesis durability and lower patient mortality, both tissue and mechanical prostheses have evolved to provide superior hemodynamics compared with older counterparts. For bioprostheses, this includes supra-annular seating and changes in sewing ring shape and strut materials that provide greater orifice area, as well as the development and ongoing refinement of stentless aortic bioprostheses (7,8). Similarly, mechanical prostheses have evolved to provide larger effective orifice area and superior hemodynamics compared with earlier valves (14). Although it remains debatable as to whether hemodynamics impact patient mortality (15,16), there is no question that superior prosthesis hemodynamics minimizes the occurrence of symptomatic prosthesis-patient mismatch. Albeit in the absence of randomized, double-blind trials, there are ample data to suggest that in hemodynamics, bioprosthesis durability, and mortality, good "new" valves are superior to good "old" valves.
 |
Mechanical versus tissue
|
|---|
In the randomized trials between mechanical and tissue valves that were performed in the late 1970s, there was either a trend (2) or statistical significance (3) favoring improved survival in association with a mechanical prosthesis. Freedom from re-operation was superior with a mechanical prosthesis, with higher rates of bioprosthesis failure among patients <65 years of age. However, freedom from all valve-related complications for tissue and mechanical prostheses was indistinguishable at 12 years (Fig. 1) (3). (Morbidity among patients with a bioprosthesis was composed predominantly of late structural failure, whereas morbidity among patients with a mechanical prosthesis occurred largely as the result of a constant, cumulative occurrence of major hemorrhage.) Although physicians concentrate on mortality and freedom from re-operation as dominant end points, patients may be more interested in avoiding all valve-related morbidity after surgery. Using this end point, bioprostheses were superior to mechanical valves prior to 12 years after surgery and were equivalent thereafter.

View larger version (30K):
[in this window]
[in a new window]
|
Figure 1 Freedom from all valve-related complications for tissue and mechanical prostheses (from Veterans Affairs trial [3], reprinted with permission from the American College of Cardiology Foundation). Morbidity was composed predominantly of late structural failure among patients with a bioprosthesis and cumulative occurrence of major hemorrhage among patients with a mechanical prosthesis. Freedom from all valve-related complications for tissue and mechanical prostheses was indistinguishable at 12 years AVR = aortic valve replacement; MVR = mitral valve replacement.
|
|
In the decades since these trials were performed, anticoagulation therapy has become more refined, with closer control of anticoagulation raising the possibility that bleeding rates in a similar trial would now be lower. Concurrently, as previously described, freedom from structural valve deterioration is superior for newer compared with older bioprostheses owing to advances in tissue technology that result in mitigation of calcification. Since the randomized trials were performed, there have been changes that favorably affect morbidity and mortality for both tissue and mechanical prostheses. Extrapolating from available data to modern practice, it is likely that freedom from valve-related complications continues to favor bioprostheses early after surgery, that total complications are similar for mechanical and tissue valves late after surgery, and that the curves would likely now superimpose later than 12 years after surgery.
 |
Differences between tissue prostheses
|
|---|
There are no large prospectively randomized comparisons among tissue prostheses, and published comparisons often involved dissimilar populations. In such circumstances, the devil is in the details. The conclusion that pericardial valves have superior durability to porcine tissue valves (1,17) may be flawed when the pericardial valve was studied in a population substantially older than that for the porcine valve and the porcine valve used in the comparison was a first-generation valve no longer in clinical use. As already described, second-generation porcine valves have greater 15-year freedom from structural deterioration compared with first-generation prostheses, and there is reason to believe that there is incremental improvement associated with third-generation bioprostheses. In short, there are no available data to suggest that modern pericardial bioprostheses have any advantage over modern porcine bioprostheses. To the contrary, data reported in similar populations suggest equivalent actuarial and actual freedom from structural valve deterioration at 15 years for pericardial and porcine bioprostheses (Table 1) (5,18).
 |
Patient quality of life
|
|---|
There are ample data to show that, regardless of prosthesis choice, mortality is higher among patients after valve replacement than among age-matched controls. Further, any differences in mortality found among prostheses are of small magnitude and of arguable clinical relevance. As such, in addition to end points of mortality and valve-related morbidity, prosthesis choices should take into consideration the difficult-to-quantify end point of patient quality of life after valve replacement. Individual patients may place different emphasis on mortality, freedom from re-operation, risk of thromboembolism (stroke), risk of anticoagulation-related hemorrhage, and the lifestyle modification required with chronic anticoagulation. When given the chance, some patients choose to accept a near certainty of requiring re-operation 10 to 20 years in the future in exchange for the benefit of avoiding anticoagulation, whereas others opt to minimize the likelihood of re-operation.
As physicians who counsel and help educate our patients before valve surgery and who continue to care for them after, we should allow greater latitude in the options that we present. Simple algorithms that assign patients under an arbitrary age to receive a mechanical prosthesis minimize our role to help optimize patient quality of life. If there is a new "state-of-the-art" for internists and cardiologists in valve surgery and prosthesis choices, it lies in understanding the developments in the field and how these changes can favorably impact our patients' lives.
 |
References
|
|---|
- Rahimtoola SH. Choice of prosthetic heart valve for adult patients. J Am Coll Cardiol. 2003;41:893904[Abstract/Free Full Text]
- Bloomfield P, Wheatley DJ, Prescott RJ, Miller HC. Twelve-year comparison of a Bjork-Shiley mechanical heart-valve with porcine bioprosthesis. N Engl J Med. 1991;324:573579[Abstract]
- Hammermeister KE, Sethi GK, Henderson WC, Grover FL, Oprian C, Rahimtoola SH. Outcomes 15 years after valve replacement with a mechanical versus bioprosthetic valve: final report of the Veterans Affairs randomized trial. J Am Coll Cardiol. 2000;36:11521158[Abstract/Free Full Text]
- Cohn LH, Collins JJ Jr., Rizzo RJ, Adams DH, Cooper GS, Avanki SF. Twenty-year follow-up of the Hancock modified orifice porcine aortic valve. Ann Thorac Surg. 1998;66:S3034[CrossRef][Medline]
- David TE, Ivanor J, Armstrong S, Feindel CM, Cohen G. Late results of heart valve replacement with the Hancock II bioprosthesis. J Thorac Cardiovasc Surg. 2001;121:268278[CrossRef][Medline]
- Gott JP, Girardot JM, Girardot NM, et al. Refinement of the alpha aminooleic acid bioprosthetic valve anticalcification technique. Ann Thorac Surg. 1997;64:5058[Abstract/Free Full Text]
- Doty DB, Cafferty A, Cartier P, et al. Aortic valve replacement with Medtronic Freestyle bioprosthesis: five year results. Semin Thorac Cardiovasc Surg. 1999;11(Suppl 1):3541[Medline]
- Bach DS, Goldman B, Verrier E, et al. Eight year hemodynamic follow-up after aortic valve replacement with the Toronto SPV stentless aortic valve. Semin Thor Cardiovasc Surg. 2001;13(Suppl 1):173179
- Melina G, Rubens MB, Amrani M, Khaghani A, Yacoub MH. Electron bean tomography for cusp calcification in homograft versus Freestyle xenografts. Ann Thorac Surg. 2001;71(Suppl):S368370
- David TE, Puschmann R, Ivanov J, et al. Aortic valve replacement with stentless and stented porcine valves: a case-match study. J Thorac Cardiovasc Surg. 1998;116:236241[Abstract/Free Full Text]
- Del Rizzo DF, Abdoh A, Cartier P, Doty D, Westaby S. The effect of prosthetic valve type on survival after aortic valve surgery. Semin Thorac Cardiovasc Surg. 1999;11(Suppl 1):18[Medline]
- Westaby S, Horton M, Jin XY, et al. Survival advantage of stentless aortic bioprostheses. Ann Thorac Surg. 2000;70:785790[Abstract/Free Full Text]
- Luciani GB, Auriemma S, Santini F, Casali G, Barozzi L, Mazzucco A. Comparison of late outcome after stentless versus stented xenograft aortic valve replacement. Semin Thorac Cardiovasc Surg. 2001;13(Suppl 1):136142[Medline]
- Bach DS, Sakwa MP, Goldbach M, Petracek M, Mohr F. Hemodynamic and early clinical performance of the St. Jude Medical Regent mechanical aortic valve. Ann Thorac Surg. 2002;74:20032009[Abstract/Free Full Text]
- Medalion B, Blackstone EH, Lytle BW, White J, Arnold JH, Cosgrove DM. Aortic valve replacement: is valve size important? J Thorac Cardiovasc Surg. 2000;119:963974[Abstract/Free Full Text]
- Rao V, Jamieson WRE, Ivanor J, Armstrong S, David T. Prosthesispatient mismatch affects survival after aortic valve replacement. Circulation. 2000;102(Suppl III):III59[Medline]
- Grunkemeier GL, Li H-H, Naftel DC, Starr A, Rahimtoola SH. Long-term performance of heart valve prosthesis. Curr Probl Cardiol. 2000;25:73156[Medline]
- Banbury MK, Cosgrove DM III, White JA, Blackstone EH, Frater RWM, Okies JE. Age and valve size effect on the long-term durability of the Carpentier-Edwards aortic pericardial prosthesis. Ann Thorac Surg. 2001;72:753757[Abstract/Free Full Text]
This article has been cited by other articles:

|
 |

|
 |
 
E B Schelbert, M S Vaughan-Sarrazin, K F Welke, and G E Rosenthal
Valve type and long-term outcomes after aortic valve replacement in older patients
Heart,
September 1, 2008;
94(9):
1181 - 1188.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. El Oakley, P. Kleine, and D. S. Bach
Choice of Prosthetic Heart Valve in Today's Practice
Circulation,
January 15, 2008;
117(2):
253 - 256.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. L. Grunkemeier, R. Jin, and A. Starr
Prosthetic heart valves: objective performance criteria versus randomized clinical trial.
Ann. Thorac. Surg.,
September 1, 2006;
82(3):
776 - 780.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. S. Bach, N. D. Kon, J. G. Dumesnil, C. F. Sintek, and D. B. Doty
Ten-Year Outcome After Aortic Valve Replacement with the Freestyle Stentless Bioprosthesis
Ann. Thorac. Surg.,
August 1, 2005;
80(2):
480 - 487.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. Bloomfield
Choice of prosthetic heart valves: 20-year results of the Edinburgh Heart Valve Trial
J. Am. Coll. Cardiol.,
August 4, 2004;
44(3):
667 - 667.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Rahimtoola
Reply
J. Am. Coll. Cardiol.,
August 4, 2004;
44(3):
667 - 668.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. S. Bach
Reply
J. Am. Coll. Cardiol.,
August 4, 2004;
44(3):
668 - 668.
[Full Text]
[PDF]
|
 |
|