| Abstract |
|---|
|
|
|---|
BACKGROUND: We have previously shown that, for two stents with similar design, the risk for restenosis is dependent on the strut thickness. It is unknown whether strut thickness preserves its relevance as a determinant of restenosis even in the presence of different stent designs.
METHODS: A total of 611 patients with symptomatic coronary artery disease were randomly assigned to receive either the thin-strut ACS RX Multilink stent (Guidant, Advanced Cardiovascular Systems, Santa Clara, California) (strut thickness 50 µm, interconnected ring design; n = 309) or the thick-strut BX Velocity stent (Cordis Corp., Miami, Florida) (strut thickness 140 µm, closed cell design; n = 302). The primary end point was angiographic restenosis (
50% diameter stenosis at follow-up angiography). Secondary end points were the incidence of target-vessel revascularization (TVR) and the combined rate of death and myocardial infarction (MI) at one year.
RESULTS: The incidence of angiographic restenosis was 17.9% in the thin-strut group and 31.4% in the thick-strut group, relative risk, 0.57 (95% confidence interval, 0.39 to 0.84), p < 0.001. A TVR due to restenosis was required in 12.3% of the thin-strut group and 21.9% of the thick-strut group, relative risk, 0.56 (95% confidence interval, 0.38 to 0.84), p = 0.002. No significant difference was observed in the combined incidence of death and MI at one year.
CONCLUSIONS: When two stents with different design are compared, the stent with thinner struts elicits less angiographic and clinical restenosis than the thicker-strut stent.
| ||||||||||||||
| Methods |
|---|
|
|
|---|
After successful passage of the guidewire through the target vessel, the patients were randomly assigned to receive one of the following two premounted stent types: the thin-strut stent, ACS RX Multi-Link (Guidant, Advanced Cardiovascular Systems, Santa Clara, California), has a strut thickness of 50 µm, a strut width of 100 µm, and an interconnected ring design; it was available in lengths of 15, 25, and 35 mm. The thin-strut stent model is no longer commercially available. The thick-strut stent, BX Velocity (Cordis Corporation, Miami, Florida), has a strut thickness of 140 µm, a strut width of 130 µm, and a closed cell design; it was available in lengths of 8, 13, 18, 23, 28, and 33 mm. The operator decided which balloon pressure to use for final stent expansion on the basis of the evidence provided by previous randomized trials (10). The procedure was considered successful when stent placement was associated with a residual stenosis of <30% and Thrombolysis in Myocardial Infarction flow grade
2. During the intervention, patients received heparin and aspirin, intravenously. The patients considered at a higher risk for thrombotic events (e.g., patients with acute myocardial infarction (MI), thrombus-containing lesions, and suboptimal procedural results) received abciximab as a bolus, followed by a 12-h intravenous infusion and heparin dosage reduced by 50%. Post-procedural therapy comprised aspirin 100 mg twice daily indefinitely and clopidogrel 75 mg/day for at least four weeks. A follow-up angiography at six months was scheduled for all patients.
Quantitative coronary angiographic evaluation.
Operators of the core angiographic laboratory who performed the quantitative assessment were blinded to the randomly assigned treatment. Angiograms recorded before and immediately after the procedure as well as at six-month follow-up were assessed with the aid of the automated edge-detection system CMS (Medis Medical Imaging System, Nuenen, the Netherlands). All measurements were performed on cineangiograms recorded after intracoronary nitroglycerin administration. The same projections were used at all time points. The contrast-filled non-tapered catheter tip was used for calibration. Early lumen gain was calculated as the difference between minimal lumen diameter at the end of the intervention and before the balloon dilation. Late lumen loss was the difference in the minimal lumen diameter between that immediately after the procedure and that at follow-up, and loss index was the ratio between late lumen loss and early lumen gain. Angiographic restenosis was defined as a diameter stenosis
50% at follow-up angiography measured at any point within the stented segment or in the 5-mm proximal or distal segments adjacent to the stent.
Clinical evaluation. Adverse events were monitored throughout the follow-up period: by telephone interview at 30 days, a clinical visit at six months, and an additional telephone interview at one year after the intervention. If patients reported cardiac symptoms during the telephone interview, at least a clinical and electrocardiographic follow-up visit was performed at the outpatient clinic or by the referring physician. All information available from hospital re-admission records, the referring physician, or the outpatient clinic was entered into a computer database. Death, MI, and target-vessel revascularization (TVR) (percutaneous transluminal coronary angioplasty or bypass surgery) were considered as adverse events. The diagnosis of acute MI required the presence of at least two of the following criteria: prolonged (>20-min duration) and typical chest pain, new pathological Q waves on the electrocardiogram, and an increase of creatine kinase (CK) or its MB isoenzyme >2 times the upper limit; CK was determined before and immediately after the procedure, every 8 h for the first 24 h after stenting, and daily afterward until discharge. The criteria for TVR included the presence of restenosis accompanied by symptoms and/or positive exercise test.
End points of the study and sample size calculation.
The primary end point of the study was the incidence of angiographic restenosis (see the definition in the preceding text). The sample size of the trial was calculated on the basis of the findings of a previous randomized trial (7): a 15.0% restenosis rate with the thin-strut stent and 25.8% with the thick-strut stent. We intended to be able to show a similar difference for the two stents in this study as well, with an 80% power and a two-sided
-level of 0.05. To achieve this objective, 440 patients with follow-up angiography were required. We included 611 patients to accommodate for expected missing angiographic examinations at six months.
Secondary end points focused on clinical events occurring during one year after the procedure: the incidence of TVR and the combined incidence of death and MI.
Statistical analysis. The main analyses were performed on an intention-to-treat basis. Data are presented as mean ± SD or as proportions (%). The differences between groups were assessed by chi-square test or Fisher exact test for categorical data and t test for continuous data. Survival parameters were compared using the log-rank test. The relative risk (RR) and its 95% confidence interval (CI) were calculated for each study end point. A p value < 0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
Clinical data at one-year follow-up. During one year after the procedure, a TVR due to restenosis was required in 38 patients of the thin-strut group (12.3%) and 66 patients of the thick-strut group (21.9%) (Fig. 2, right panel) (p = 0.002). This indicates a 44% risk reduction with the use of the thin-strut stent (RR, 0.56 [95% CI, 0.38 to 0.84]). During the same follow-up period, 15 patients of the thin-strut group (4.9%) and 19 patients of the thick-strut group (6.3%) died or incurred an MI (RR, 0.77 [95 CI, 0.39 to 1.52], p = 0.46). No significant differences were observed in one-year mortality rates, with 3.9% in the thin-strut group and 4.6% in the thick-strut group (p = 0.66).
| Discussion |
|---|
|
|
|---|
Comparison with previous trials. The incidence of restenosis at follow-up angiography with the thin-strut stent was 17.9%. This appears to be slightly higher than the incidence of restenosis of 15.0% reported from in the previous ISAR-STEREO trial with the same stent type (7). The reason may be found in the difference in vessel size between the populations included in the two ISAR-STEREO trials, with smaller coronary vessels treated in the present trial. Altogether, the findings of the ISAR-STEREO trials as well as those of another randomized trial (12) demonstrate that consistently low restenosis rates can be achieved with thin-strut stents.
In the present trial, the incidence of restenosis at follow-up angiography with the thick-strut stent was 31.4%. The same type of stent has been used in the uncoated stent arm of the Randomized Study With the Sirolimus Coated Velocity Balloon Expandable Stent in the Treatment of Patients With de novo Native Coronary Lesions (RAVEL) (13). In the RAVEL trial, which included a control group of 118 patients treated with the uncoated BX-Velocity stent, the angiographic restenosis rate in this arm was 26% (13). Considering that the study lesions in the ISAR-STEREO 2 trial were more complex, there is essentially no difference between our trial and the RAVEL trial with respect to the incidence of angiographic restenosis observed in the uncoated, thick-strut arm. Consistent with these results are also the findings of clinical restenosis, with a revascularization rate of 21.9% in the present trial and 22% in the RAVEL trial (13).
Clinical implications. Although there was no difference in mortality and the incidence of MI, the group assigned to the thin-strut stent had a 43% decrease in the risk for restenosis at follow-up angiography and a 44% reduction in the risk for re-intervention, compared with the group assigned to the thick-strut stent. Similar reduction in the risk for restenosis was also observed in the group of the thin-strut stent in the first ISAR-STEREO trial (7). The consistent reduction of restenosis with the thin-strut stent compared with two thick-strut stents with different architecture in both ISAR-STEREO trials strongly supports a role for the strut thickness that is largely independent of other stent characteristics. We do not know the exact mechanisms underlying the reduction of restenosis after the use of thin-strut stents. A few possibilities have already been discussed in the previous publication on the same subject (7). Despite the paucity of the mechanistic data that may help to explain the influence of strut thickness on restenosis (9), the findings of the present trial have important implications because most of the stents currently in use have a strut thickness similar to that of the thick-strut stent used in this and in the previous ISAR-STEREO trial. Drug-eluting stent technology has surely opened new promising prospects for our efforts against restenosis (13), and hopefully, large studies in the near future will be able to prove the abolishment of restenosis through this technology.
Study limitations. The design of the present study aimed at complementing the previous ISAR-STEREO trial (7). Although both trials suggest a relevant role for strut thickness in in-stent restenosis, other stent-related factors might also have contributed to the differences in restenosis observed between the study groups. The impact of stent geometry on restenosis has not been evaluated in controlled clinical trials, and it is not known which geometrical characteristics enable reduction of restenosis. The differences in stent geometry between the thick- and the thin-strut models, which were subtle in the previous ISAR-STEREO trial (7) and much more conspicuous in the present trial, might also have contributed to the overall better result achieved with the thin-strut stent. Therefore, further work is needed to quantify the degree in which strut thickness per se influences in-stent restenosis and to identify the optimal strut thickness to use during stent manufacturing.
Two additional limitations should also be acknowledged. First, due to the failure of thin-strut device, the procedural success in 12% of the patients in the thin-strut group was achieved after crossover to the thick-strut model. We tried to adjust for this inconvenience by performing an additional analysis confined to patients without crossover, but both the nature and magnitude of the possible bias introduced by the crossover phenomenon remain still unknown. Second, the difference in radiolucency between the thin- and the thick-strut models might have interfered with the quantitative angiographic assessment. A specifically designed study concluded that quantitative angiography can be used as an accurate method of lumen assessment after implantation of steel stents with different thickness and that only the highly opaque tantalum stent may reduce the accuracy to some degree (14). However, our angiographic data are also supported by the incidence of reinterventions, which favors the thin-strut model.
| APPENDIX |
|---|
|
|
|---|
Steering Committee: A. Schömig (chairman), A. Kastrati, J. Dirschinger.
Data Coordinating Center: A. Kastrati, M. Hadamitzky, Deutsches Herzzentrum, Munich.
Angiographic Core Laboratory: J. Mehilli, A. Redl, D. Kiemoser, S. Pinieck, Deutsches Herzzentrum, Munich.
Clinical Follow-up Center: H. Holle, K. Hösl, F. Rodrigues, S. Koch, M. Müller; Deutsches Herzzentrum, Munich.
Patient-Enrolling Centers: Deutsches Herzzentrum, Munich: J. Dirschinger (principal investigator), C. Schmitt, R. Blasini, M. Gawaz, J. Pache; 1. Medizinische Klinik rechts der Isar, Munich: F.-J. Neumann (principal investigator), H. Schühlen, M. Seyfarth; Medizinische Klinik I, Garmisch-Partenkirchen: F. Dotzer (principal investigator), M. Fleckenstein. Medizinische Klinik I, Ingolstadt: C. Pfafferott (principal investigator), U. Sattelberger.
| Footnotes |
|---|
Please see the Appendix for the complete list of the centers and investigators participating in the ISAR-STEREO 2 trial.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
B. Chevalier, C. Di Mario, F.-J. Neumann, F. Ribichini, P. Urban, J. J. Popma, P. J. Fitzgerald, D. E. Cutlip, D. O. Williams, J. Ormiston, et al. A Randomized, Controlled, Multicenter Trial to Evaluate the Safety and Efficacy of Zotarolimus- Versus Paclitaxel-Eluting Stents in De Novo Occlusive Lesions in Coronary Arteries: The ZoMaxx I Trial J. Am. Coll. Cardiol. Intv., October 1, 2008; 1(5): 524 - 532. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. C. John, R. Wessely, A. Kastrati, A. Schomig, M. Joner, M. Uchihashi, J. Crimins, S. Lajoie, F. D. Kolodgie, H. K. Gold, et al. Differential Healing Responses in Polymer- and Nonpolymer-Based Sirolimus-Eluting Stents J. Am. Coll. Cardiol. Intv., October 1, 2008; 1(5): 535 - 544. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Joner, G. Nakazawa, A. V. Finn, S. C. Quee, L. Coleman, E. Acampado, P. S. Wilson, K. Skorija, Q. Cheng, X. Xu, et al. Endothelial Cell Recovery Between Comparator Polymer-Based Drug-Eluting Stents J. Am. Coll. Cardiol., July 29, 2008; 52(5): 333 - 342. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Vlaar, B. J. G. L. de Smet, and F. Zijlstra DES or BMS in acute myocardial infarction? Eur. Heart J., November 2, 2007; 28(22): 2693 - 2694. [Full Text] [PDF] |
||||
![]() |
C.-C. Fang, Yeun Tarl Fresner Ng Jao, Yi-Chen, C.-L. Yu, C.-L. Chen, and S.-P. Wang Angiographic and Clinical Outcomes of Rosiglitazone in Patients With Type 2 Diabetes Mellitus After Percutaneous Coronary Interventions: A Single Center Experience Angiology, November 1, 2007; 58(5): 523 - 534. [Abstract] [PDF] |
||||
![]() |
M. Menichelli, A. Parma, E. Pucci, R. Fiorilli, F. De Felice, M. Nazzaro, A. Giulivi, D. Alborino, A. Azzellino, and R. Violini Randomized Trial of Sirolimus-Eluting Stent Versus Bare-Metal Stent in Acute Myocardial Infarction (SESAMI) J. Am. Coll. Cardiol., May 15, 2007; 49(19): 1924 - 1930. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Tung, S. Kaul, G. A. Diamond, and P. K. Shah Narrative review: drug-eluting stents for the management of restenosis: a critical appraisal of the evidence. Ann Intern Med, June 20, 2006; 144(12): 913 - 919. [Abstract] [Full Text] [PDF] |
||||
![]() |
C Stettler, S Allemann, M Egger, S Windecker, B Meier, and P Diem Efficacy of drug eluting stents in patients with and without diabetes mellitus: indirect comparison of controlled trials Heart, May 1, 2006; 92(5): 650 - 657. [Abstract] [Full Text] [PDF] |
||||
![]() |
Additional Information JAMA, March 15, 2006; 295(11): E1 - E6. [Full Text] [PDF] |
||||
![]() |
U Dietz, N Holz, C Dauer, and H Lambertz Shortening the stent length reduces restenosis with bare metal stents: matched pair comparison of short stenting and conventional stenting Heart, January 1, 2006; 92(1): 80 - 84. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. A. Holzapfel, G. Sommer, C. T. Gasser, and P. Regitnig Determination of layer-specific mechanical properties of human coronary arteries with nonatherosclerotic intimal thickening and related constitutive modeling Am J Physiol Heart Circ Physiol, November 1, 2005; 289(5): H2048 - H2058. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. S. Monraats, N. M.M. Pires, W. R.P. Agema, A. H. Zwinderman, A. Schepers, M. P.M. de Maat, P. A. Doevendans, R. J. de Winter, R. A. Tio, J. Waltenberger, et al. Genetic Inflammatory Factors Predict Restenosis After Percutaneous Coronary Interventions Circulation, October 18, 2005; 112(16): 2417 - 2425. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Butany, K Carmichael, S W Leong, and M J Collins Coronary artery stents: identification and evaluation J. Clin. Pathol., August 1, 2005; 58(8): 795 - 804. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. W. Serruys, G. Sianos, A. Abizaid, J. Aoki, P. den Heijer, H. Bonnier, P. Smits, D. McClean, S. Verheye, J. Belardi, et al. The Effect of Variable Dose and Release Kinetics on Neointimal Hyperplasia Using a Novel Paclitaxel-Eluting Stent Platform: The Paclitaxel In-Stent Controlled Elution Study (PISCES) J. Am. Coll. Cardiol., July 19, 2005; 46(2): 253 - 260. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Pache, A. Dibra, J. Mehilli, J. Dirschinger, A. Schomig, and A. Kastrati Drug-eluting stents compared with thin-strut bare stents for the reduction of restenosis: a prospective, randomized trial Eur. Heart J., July 1, 2005; 26(13): 1262 - 1268. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Windecker, R. Simon, M. Lins, V. Klauss, F. R. Eberli, M. Roffi, G. Pedrazzini, T. Moccetti, P. Wenaweser, M. Togni, et al. Randomized Comparison of a Titanium-Nitride-Oxide-Coated Stent With a Stainless Steel Stent for Coronary Revascularization: The TiNOX Trial Circulation, May 24, 2005; 111(20): 2617 - 2622. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. S. Rana, P. S. Monraats, A. H. Zwinderman, M. P.M. de Maat, J. J.P. Kastelein, P. A.F. Doevendans, R. J. de Winter, R. A. Tio, R. R. Frants, A. van der Laarse, et al. Metabolic Syndrome and Risk of Restenosis in Patients Undergoing Percutaneous Coronary Intervention Diabetes Care, April 1, 2005; 28(4): 873 - 877. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. V. Finn, I. F. Palacios, A. Kastrati, and H. K. Gold Drug-eluting stents for diabetes mellitus: A rush to judgment? J. Am. Coll. Cardiol., February 15, 2005; 45(4): 479 - 483. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. R.P Agema, P. S Monraats, A. H Zwinderman, R. J de Winter, R. A Tio, P. A.F.M Doevendans, J. Waltenberger, M. P.M de Maat, R. R Frants, D. E Atsma, et al. Current PTCA practice and clinical outcomes in The Netherlands: the real world in the pre-drug-eluting stent era Eur. Heart J., July 1, 2004; 25(13): 1163 - 1170. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Moreno, C. Fernandez, F. Alfonso, R. Hernandez, M. J. Perez-Vizcayno, J. Escaned, M. Sabate, C. Banuelos, D. J. Angiolillo, L. Azcona, et al. Coronary stenting versus balloon angioplasty in small vessels: A meta-analysis from 11 randomized studies J. Am. Coll. Cardiol., June 2, 2004; 43(11): 1964 - 1972. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Kereiakes, A. M. Szyniszewski, D. Wahr, H. C. Herrmann, D. I. Simon, C. Rogers, P. Kramer, W. Shear, A. C. Yeung, K. A. Shunk, et al. Phase I Drug and Light Dose-Escalation Trial of Motexafin Lutetium and Far Red Light Activation (Phototherapy) in Subjects With Coronary Artery Disease Undergoing Percutaneous Coronary Intervention and Stent Deployment: Procedural and Long-Term Results Circulation, September 16, 2003; 108(11): 1310 - 1315. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. B. Berger Can euphoria be bad? J. Am. Coll. Cardiol., April 16, 2003; 41(8): 1289 - 1292. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |