J Am Coll Cardiol, 2003; 41:565-571, doi:10.1016/S0735-1097(02)02896-6 © 2003 by the American College of Cardiology Foundation |
,*




* Department of Internal Medicine, Cardiology, University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma, USA
Department of Physiology, University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma, USA
Mid-America Heart Institute, Kansas City, Kansas, USA
Cardiology Department, Karolinska Hospital, Stockholm, Sweden
|| University Clinic, Homburg/Saar, Germany
¶ Medtronic Heart Failure Management, Minneapolis, Minnesota, USA
Manuscript received June 18, 2002; revised manuscript received September 23, 2002, accepted October 10, 2002.
* Reprint requests and correspondence: Dr. Philip B. Adamson, University of Oklahoma Health Sciences Center, Medicine/Cardiology, P.O. Box 26901, WP3120, Oklahoma City, Oklahoma 73190, USA.
philip-adamson{at}ouhsc.edu
| Abstract |
|---|
|
|
|---|
BACKGROUND: Hemodynamic monitoring might improve the day-to-day management of patients with chronic heart failure (CHF). Little is known about the characteristics of long-term hemodynamic information in patients with CHF or how such information relates to meaningful clinical events.
METHODS: Thirty-two patients with CHF received a permanent RV IHM system similar to a single-lead pacemaker. Right ventricular systolic and diastolic pressures, heart rate, and pressure derivatives were continuously measured for nine months without using the data for clinical decision-making or management of patients. Data were then made available to clinical providers, and the patients were followed up for 17 months. Pressure characteristics during optimal volume, clinically determined volume-overload exacerbations, and volume depletion events were examined. The effect of IHM on hospitalizations was examined using the patients historical controls.
RESULTS: Long-term RV pressure measurements had either marked variability or minimal time-related changes. During 36 volume-overload events, RV systolic pressures increased by 25 ± 4% (p < 0.05) and heart rate increased by 11 ± 2% (p < 0.05). Pressure increases occurred in 9 of 12 events 4 ± 2 days before the exacerbations requiring hospitalization. Hospitalizations before using IHM data for clinical management averaged 1.08 per patient year and decreased to 0.47 per patient-year (57% reduction, p < 0.01) after hemodynamic data were used.
CONCLUSIONS: Long-term ambulatory pressure measurements from an IHM may be helpful in guiding day-to-day clinical management, with a potentially favorable impact on CHF hospitalizations.
| ||||||||||||||||||||
Continuous hemodynamic monitoring with an implantable device is technically feasible and can provide information during normal activity (510). Newer device designs have long-term accuracy (5), providing information over a variety of physiologic conditions (5) and during short-term medication changes (6). This study is the first to examine the long-term characteristics of ongoing hemodynamics measured from a permanently implanted device in patients with CHF and to correlate changes to clinical events. These data help develop the hypothesis that incorporation of long-term hemodynamic data in heart failure (HF) management has the potential to affect meaningful outcome measures in patients with CHF.
| Methods |
|---|
|
|
|---|
Patient inclusion criteria. Patients between the ages of 21 and 75 years, with New York Heart Association class II/III CHF, were enrolled in the trial if they were able to provide informed consent. Subjects were excluded if they had an existing indication for a pacemaker or an implantable cardiac defibrillator (ICD). Additionally, patients with known atrial or ventricular septal defects, tricuspid or pulmonic stenosis, or mechanical right heart valves, as well as those with a terminal illness unrelated to HF with a life-expectancy of <12 months, were excluded. Institutional review boards at each site approved the study design, and patients gave informed consent to be involved in this investigational protocol.
Device design and implantation
The Chronicle IHM (Medtronic, Inc., Minneapolis, Minnesota) is a long-term implanted device designed to record ongoing right ventricular (RV) pressures, heart rate (HR), and pressure derivatives. The implantation procedure was similar to a single-lead permanent pacemaker system with a programmable memory device (128-K random-access memory) placed in the pectoral area and a transvenous electrode carrying a pressure sensor in the RV outflow tract. The device continuously measured RV systolic and diastolic pressures (RVSP and RVDP), estimated pulmonary artery diastolic pressure (ePADP), maximum change in pressure over time (dP/dt), HR, activity, and temperature. The ePADP was derived from RV pressures at the time of maximum dP/dt, the moment of pulmonary valve opening. This concept was validated in previous work (11,12).
Absolute pressure values were obtained directly from the pressure sensor, and HR was determined from a unipolar electrogram recorded at the tip of the lead. Long-term performance of the IHM sensor was recently established (5). The sensor provides a continuous pressure trace for analysis by the IHM (Fig. 1) and is known to be stable over time (5,7,8). A brief pre-insertion calibration procedure was performed immediately before implantation. Once the sensor was implanted, no additional or repeated calibration was required. Previous validation of the sensors accuracy and precision was performed by using serial right heart catheterizations and comparing IHM signals with simultaneously acquired balloon-tipped catheter values at 3, 6, and 12 months after implantation (5). These studies revealed a small baseline error at 12 months that was similar to the error at the time of implantation (<1.0 mm Hg). This indicated that no significant drift occurred in the sensor (5). Although the sensors response is linear to >100 Hz, the IHM signal was low-pass filtered at 60 Hz to minimize noise. An external device that patients kept with them recorded ambient barometric pressure once per minute. When stored hemodynamic information was interrogated from the monitor, the external pressure reference information was subtracted from monitor data to provide relative pressures in the RV.
|
Long-term monitoring and clinical end points
The daily maximum, minimum, median, and range of pressures were analyzed, along with nighttime values between midnight and 4 AM, when no activity was sensed. During the blinded phase of the trial, HF physicians determined volume-overload exacerbations on the basis of elevated jugular venous pressure, rhonchi on lung auscultation, and abdominal-jugular reflux. Volume-overload events were considered major if hospitalization was required and minor if outpatient adjustments in therapy treated the exacerbation. Volume depletion was reported on the basis of an orthostatic drop in systolic blood pressure and typical signs and symptoms of dehydration that resulted in discontinuation of diuretic therapy or volume infusion. Individual baseline pressure parameters were determined at times when the patient had optimal volume control. Percent changes in pressures were calculated for each day during the week before the event. A sustained change of >20% above baseline for any pressure parameter before a clinical event was considered meaningful.
The CHF hospitalization data (verified by ICD-9 and DRG codes) collected from 12 months before IHM placement, and during the 9-months blinded phase, were compared with the 17-months of follow-up that began when information was made available for management decisions. Hospitalization incidence was expressed in patient-years to reduce the statistical impact of patients who underwent transplantation or died during follow-up.
Statistics
The two-tailed paired Student t test was used to determine a difference between baseline values and peak pressure values. Hospitalization rates were evaluated using an unpaired t test comparing pre-hemodynamic management rates to post-hemodynamic rates. An alpha level of p < 0.05 was considered statistically significant. Data are reported as the mean value ± SD, unless otherwise noted in the text.
| Results |
|---|
|
|
|---|
|
Volume-related exacerbations
Thirty-six volume-overload events occurred in 14 patients, and 12 events (major) were severe enough to require hospitalization. Sustained increases (>20%) in at least one pressure parameter were observed in nine of the 12 events 4 ± 2 days before exacerbations, leading to hospitalizations (Fig. 2, left panel). In contrast, early pressure increases occurred in only 9 of 24 events with minor exacerbations. In all exacerbations, pressures increased 24 h before clinical intervention (Fig. 2). The RVDP increased by 265 ± 16% (2.3 ± 4 to 8.4 ± 4 mm Hg), whereas RVSP increased by 25 ± 4% (54 ± 17 to 67 ± 10 mm Hg), and ePADP rose 26 ± 4% (28 ± 8 to 35 ± 9 mm Hg), when comparing baseline values with peak events (all p < 0.05). The HR increased by 11 ± 2% during the peak of exacerbations (p < 0.05).
|
Hospitalizations
Hospitalization data were examined in 28 patients, as one patient died, two underwent heart transplantation, and one was lost to follow-up before hemodynamic data were used for clinical management. A total of 52 CHF-related hospitalizations occurred before use of hemodynamic information for management had begun. After hemodynamic data were available to help manage patients, six patients received a transplant and five died. There were 18 CHF-related hospitalizations in 14 of the 28 patients in the 17 months of follow-up. When expressed in patient-years, there were 1.08 admissions per patient-year in the 21 months before hemodynamic information was used and 0.47 hospitalizations per patient-year in the 17 months after data were made available (57% reduction, p < 0.01) (Fig. 3).
|
| Discussion |
|---|
|
|
|---|
Individual characteristics of ongoing pressures. The individual characteristics of ongoing pressures were unique, which required using the patients values as their own control. Many patients had large pressure variability, whereas others had less variable measurements with a similar clinical course (Fig. 1). This observation may be important, as highly variable systemic arterial pressures cause more end-organ damage than higher mean, but less variable pressures (14). Whether RV pressure variability carries the same risk remains to be determined. Furthermore, inherent pressure variability may increase the possibility of inappropriate volume assessment, unless monitoring algorithms account for individual changes. Other markers of physiologic control systems, such as HR variability, which indirectly quantifies autonomic HR control, may be important parameters to include in continuous hemodynamic monitoring systems.
This study established that long-term pressures have individual variability, but it was not designed to examine the exact predictive value of continuous hemodynamic monitoring, as there were no control subjects in the phase of the trial in which hemodynamic information was used for management. However, these observations will provide a means to develop future monitoring algorithms that may facilitate processing of continuously acquired data. This is important because pressures changed beyond their normal variability when volume-overload exacerbations were severe enough to require hospitalization. Often the pressure changes occurred several days before clinical intervention. Pressures further increased in the 24 h before hospitalization. This pattern suggests that the earlier changes possibly reduced the patients "tolerable reserve" but did not produce severe symptoms until pressures increased further. Therapy guided by information from IHM systems could thus be useful by providing early warning of an impending, severe exacerbation or, in the case of minor volume overload, confirming the patients status and guiding therapy. This may result in a substantial reduction in the need for health care utilization, especially with the advent of trans-telephone data transmission. Furthermore, when hospitalization cannot be avoided, the presence of an IHM may reduce the need for invasive hemodynamic monitoring using traditional balloon-tipped catheters. Prospective studies designed to specifically evaluate how an IHM may reduce costs of CHF care are still required.
Implications for future use
Device therapy to treat CHF now includes biventricular pacemakers in patients with significant conduction delays (15,16) and ICDs in those with ischemic heart disease (17). Because biventricular pacemakers reduce hospitalizations (16) and ICDs tend to increase them (17), combination devices, such as biventricular pacemakers with ICDs, may provide therapy that has a favorable impact on morbidity as well as mortality. Similarly, combining IHM capability with ICDs or other devices may have a favorable, meaningful impact clinical outcomes. A device that combines IHM information with ICD capability will potentially result in more cost-effective, device-based, disease management systems. Furthermore, value may be added to other interventions such as ventricular assist devices or artificial hearts and may offer a way to reduce costs while enhancing therapeutic benefits.
The patients in this study already benefited from organized CHF treatment programs, which significantly reduce hospitalizations (1). Although this study was not designed to examine the effect of IHMs on health care utilization, the results provide substantial background to more fully examine the hypothesis that IHMs use may have a broad range of applications and will likely have a significant impact on disease morbidity, even beyond those benefits of an organized system. The device used in this study continuously measures hemodynamic information. This is theoretically superior to one-time "spot measurements," typically used in CHF management and diagnosis, such as body weight and B-type natriuretic protein. Validation of this concept will require prospective trials specifically designed to compare other modalities with IHM information.
Conclusions
Continuous monitoring of ambulatory RV pressures is feasible, accurate, and safe. Furthermore, RV pressure changes reflect the patients volume status, which provides meaningful information to guide day-to-day management of CHF. Long-term management of patients with CHF, guided by ambulatory hemodynamic information, promises to have an effective impact on the severe morbidity associated with this syndrome.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. C. Bourge, W. T. Abraham, P. B. Adamson, M. F. Aaron, J. M. Aranda Jr, A. Magalski, M. R. Zile, A. L. Smith, F. W. Smart, M. A. O'Shaughnessy, et al. Randomized controlled trial of an implantable continuous hemodynamic monitor in patients with advanced heart failure: the COMPASS-HF study. J. Am. Coll. Cardiol., March 18, 2008; 51(11): 1073 - 1079. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. E. Verdejo, P. F. Castro, R. Concepcion, M. A. Ferrada, M. A. Alfaro, M. E. Alcaino, C. C. Deck, and R. C. Bourge Comparison of a Radiofrequency-Based Wireless Pressure Sensor to Swan-Ganz Catheter and Echocardiography for Ambulatory Assessment of Pulmonary Artery Pressure in Heart Failure J. Am. Coll. Cardiol., December 18, 2007; 50(25): 2375 - 2382. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ritzema, I. C. Melton, A. M. Richards, I. G. Crozier, C. Frampton, R. N. Doughty, J. Whiting, S. Kar, N. Eigler, H. Krum, et al. Direct Left Atrial Pressure Monitoring in Ambulatory Heart Failure Patients: Initial Experience With a New Permanent Implantable Device Circulation, December 18, 2007; 116(25): 2952 - 2959. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-P. Lau, H.-F. Tse, A. J. Camm, and S. S. Barold Evolution of pacing for bradycardias: sensors Eur. Heart J. Suppl., December 1, 2007; 9(suppl_I): I11 - I22. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Stellbrink and H.-J. Trappe The follow-up of cardiac devices: what to expect for the future? Eur. Heart J. Suppl., December 1, 2007; 9(suppl_I): I113 - I115. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. E. Wolfel Can We Predict and Prevent the Onset of Acute Decompensated Heart Failure? Circulation, October 2, 2007; 116(14): 1526 - 1529. [Full Text] [PDF] |
||||
![]() |
E. Page, S. Cazeau, P. Ritter, D. Galley, and C. Casset Physiological approach to monitor patients in congestive heart failure: application of a new implantable device-based system to monitor daily life activity and ventilation Europace, August 1, 2007; 9(8): 687 - 693. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Rozenman, R. S. Schwartz, H. Shah, and K. H. Parikh Wireless Acoustic Communication With a Miniature Pressure Sensor in the Pulmonary Artery for Disease Surveillance and Therapy of Patients With Congestive Heart Failure J. Am. Coll. Cardiol., February 20, 2007; 49(7): 784 - 789. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Elkayam, P. Hatamizadeh, and M. Janmohamed The Challenge of Correcting Volume Overload in Hospitalized Patients With Decompensated Heart Failure J. Am. Coll. Cardiol., February 13, 2007; 49(6): 684 - 686. [Full Text] [PDF] |
||||
![]() |
E. Zima, M. Lippert, G. Czygan, and B. Merkely Determination of left ventricular volume changes by intracardiac conductance using a biventricular electrode configuration Europace, July 1, 2006; 8(7): 537 - 544. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Braunschweig, B. Kjellstrom, M. Soderhall, N. Clyne, and C. Linde Dynamic changes in right ventricular pressures during haemodialysis recorded with an implantable haemodynamic monitor Nephrol. Dial. Transplant., January 1, 2006; 21(1): 176 - 183. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Wadas The Implantable Hemodynamic Monitoring System Crit. Care Nurse, October 1, 2005; 25(5): 14 - 26. [Full Text] [PDF] |
||||
![]() |
D. F. Pauly Cardiac Function and Heart Failure J. Am. Coll. Cardiol., June 7, 2005; 45(11_Suppl_B): B24 - B29. [Full Text] [PDF] |
||||
![]() |
M. Gheorghiade, W. A. Gattis, C. M. O'Connor, K. F. Adams Jr, U. Elkayam, A. Barbagelata, J. K. Ghali, R. L. Benza, F. A. McGrew, M. Klapholz, et al. Effects of Tolvaptan, a Vasopressin Antagonist, in Patients Hospitalized With Worsening Heart Failure: A Randomized Controlled Trial JAMA, April 28, 2004; 291(16): 1963 - 1971. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Shabetai Monitoring heart failure hemodynamics with an implanted device:its potential to improve outcome J. Am. Coll. Cardiol., February 19, 2003; 41(4): 572 - 573. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||