J Am Coll Cardiol, 2003; 41:1926-1932, doi:10.1016/S0735-1097(03)00426-1 © 2003 by the American College of Cardiology Foundation |



* John Muir/Mount Diablo Medical Center, Walnut Creek, California, USA
Ferrell-Duncan Clinic and Cox Medical Center, Springfield, Missouri, USA
Florida Heart Institute and Florida Hospital, Orlando, Florida, USA
MidCarolina Cardiology and Presbyterian Hospital, Charlotte, North Carolina, USA
|| Cedar-Sinai Medical Center and UCLA School of Medicine, Los Angeles, California, USA
¶ St. Josephs Hospital Health Center, Syracuse, New York, USA
** Medtronic Inc., Minneapolis, Minnesota, USA
Manuscript received September 27, 2002; revised manuscript received February 20, 2003, accepted February 25, 2003.
* Reprint requests and correspondence: Dr. Michael A. Lee, 365 Hawthorne Avenue, #201, Oakland, California 94609, USA.
drmichaelalee{at}yahoo.com
| Abstract |
|---|
|
|
|---|
BACKGROUND: The effect of the combination of atrial prevention and termination algorithms on AT/AF burden and frequency in pacemaker patients is unknown.
METHODS: A DDDRP pacemaker (AT500, Medtronic Inc., Minneapolis, Minnesota) with three atrial preventive pacing algorithms and two ATP algorithms was implanted in 368 patients. Patients were randomized one-month post-implant to all prevention and ATP therapies ON or OFF and followed for three months. The OFF group had DDDR pacing at a lower programmed rate of 60 ppm. The AT/AF burden and frequency were determined from daily device counters in 324 patients treated according to protocol.
RESULTS: In 17,018 episodes with stored electrograms, appropriate detection was confirmed in 17,004 (99.9%). The median percentage of atrial pacing was 98% in the ON group versus 75% in the OFF group (p < 0.001). Using device-defined criteria for successful termination, ATP terminated 8,590 (54%) of 15,789 treated episodes. The median AT/AF burden during the three-month study period was 4.2 h/month ON versus 1.1 h/month OFF (p = 0.20). The median AT/AF frequency was 1.3 episodes/month ON versus 1.2 episodes/month OFF (p = 0.65). System-related, complication-free survival at four months was 90.2% (Kaplan-Meier estimate).
CONCLUSIONS: This DDDRP pacemaker is safe, has accurate AT/AF detection, and provides ATP with 54% efficacy as defined by the device. The atrial prevention and termination therapies combined did not reduce AT/AF burden or frequency in this patient population.
| ||||||||||||
| Methods |
|---|
|
|
|---|
|
The device also has three pacing algorithms designed to prevent AT/AF recurrence. These include an atrial overdrive algorithm designed to maintain the programmed atrial escape interval just ahead of the intrinsic rate (atrial preference pacing), a temporary atrial overdrive mode designed to prevent short-long intervals after a premature atrial contraction (atrial rate stabilization), and a post-mode-switch overdrive pacing mode designed to inhibit the early recurrence of AT/AF after episode termination (6,7).
Study design and patient population. Eligible patients were those having American College of Cardiology/American Heart Association class I or II indications for a standard dual-chamber pacemaker (8) and a history of AT/AF. Patients were required to have at least one episode of paroxysmal or persistent AT/AF documented by electrocardiogram within the past year and one symptomatic episode in the past three months. Patients with permanent AF, an indication for an implantable defibrillator, or New York Heart Association functional class IV heart failure were excluded. All patients gave written informed consent according to a protocol approved by the local human subjects committee at the implanting center.
Bipolar right atrial and ventricular leads were required at implant, with
10 mm interelectrode spacing recommended. Atrial detection was programmed ON at implant in all patients, and DDDR pacing was programmed at a lower rate of 60 ppm and an upper rate of 120 ppm. A one-month run-in period was allowed for lead fixation and drug stabilization. At the one-month visit, patients were randomized to all prevention and termination therapies ON or OFF (Table 1) and followed for three months.
|
28 days of follow-up with strict programming compliance (Table 1). In addition, patients were excluded if they developed chronic AF before randomization that persisted throughout the study period. Secondary efficacy end points included the accuracy of the detection algorithm, ATP efficacy, and symptomatic AT/AF frequency. The accuracy of the detection algorithm was determined by investigator review of stored AT/AF episodes. The ATP efficacy was expressed as the incidence of successful terminations based on the device counters. Symptomatic frequency was determined similar to total frequency except that only episodes were counted that were associated with one or more manual symptom markers (Fig. 1).
The primary safety end point was system-related, complication-free survival in all patients enrolled. System-related complications were defined as clinical events with potential adverse effects requiring invasive interventions specifically related to the device, the implanted leads, or the implant procedure. Adverse clinical events were analyzed and classified by an independent physician advisory committee.
Statistical analyses.
Statistical analyses were performed using SAS statistical software (SAS Institute, Cary, North Carolina). Continuous data were reported as the mean ± SD or as the median along with the 25th to 75th percentile. A Kaplan-Meier estimate was obtained for complication-free survival. Comparisons between groups were performed using a Wilcoxon rank-sum test for unpaired data, a Wilcoxon signed rank test for paired data, or a chi-square test. For the evaluation of ATP efficacy and the efficacy of the detection algorithm, the generalized estimating equation (GEE) method was used with an exchangeable correlation structure (9). This technique adjusts for multiple episodes in a single patient and yields an average estimate that is based on the number of patients as well as the number of episodes per patient. For all statistical comparisons, the null hypothesis was rejected when p
0.05.
| Results |
|---|
|
|
|---|
The demographics of the patients in the efficacy cohort are shown in Table 2. There were no significant differences in baseline characteristics between the ON and OFF groups. The antiarrhythmic drugs used are shown in Table 3. There were no significant differences in the use of drugs between the ON and OFF groups at either follow-up time, and there were no significant changes in drug use over time in the two groups.
|
|
Atrial and ventricular pacing. The number of atrial and ventricular paced and sensed beats were determined from the daily trend counters (atrial beats were not counted during an AT/AF episode) and summed over the study period in each patient. The median amount of atrial pacing was 98% (25th to 75th percentile, 97% to 99%) in the ON group and 75% (25th to 75th percentile, 38% to 95%) in the OFF group (p < 0.001). The median amount of ventricular pacing was 99% (25th to 75th percentile, 95% to 100%) in the ON group and 98% (25th to 75th percentile, 81% to 100%) in the OFF group (p = 0.005).
Atrial ATP. A total of 94 patients in the ON group experienced 54,976 AT/AF episodes during the three-month study period, and 15,789 of these episodes were treated with ATP in 85 patients. There were 8,590 treated episodes that were classified by the device as successfully terminated, yielding an efficacy of 54.4% (GEE adjusted estimate, 41.4%). An example of an episode treated successfully by ATP is shown in Figure 2.
|
0.70).
|
0.40). It is also conceivable that there may have been a subgroup of patients receiving a particular type of antiarrhythmic drug or with a certain demographic profile that may have derived benefit from the therapies. Therefore, multiple subgroup analyses were performed after stratifying patients on the basis of their baseline characteristics (Table 2) or the use of antiarrhythmic drugs at one month (Table 3). The results of these analyses did not identify a responder subgroup.
Episode duration.
Figure 3 shows a histogram of AT/AF episodes based on duration. The device counters automatically grouped episodes into eight duration bands. Only seven bands are displayed because episodes lasting <1 min were not eligible to receive ATP. There were 28,735 episodes <1 min in the ON group and 20,395 episodes in the OFF group. There were 26,241 episodes
1 min in the ON group and 28,229 episodes in the OFF group. The majority of all episodes (
80%) terminated within 10 min in both groups. Only 60% of the episodes
1 min in the ON group were treated with ATP therapy. For each patient in the ON and OFF groups, the number of episodes in each duration band was divided by the follow-up time. This yielded eight band-specific AT/AF frequencies. The median frequency in each band was compared between the ON and OFF groups (Wilcoxon rank-sum test), and no significant differences were observed (p
0.17).
|
There were nine deaths over the course of the four-month follow-up period; six occurred before randomization and one was classified as procedure related. Causes of death included cardiac arrest (n = 3), cardiogenic shock (n = 2), congestive heart failure (n = 1), ventricular arrhythmia (n = 1), and renal failure (n = 1). One death was classified as unknown owing to inability of the site to acquire patient records from an outlying facility.
There were 65 external cardioversions to treat AT/AF in 59/368 patients during the four-month follow-up period. The largest number of cardioversions occurred at the time of implant (48%). In the efficacy cohort, there were 7 (4.6%) cardioversions in the ON group and 2 (1.2%) in the OFF group during the study period (p = 0.13).
| Discussion |
|---|
|
|
|---|
The atrial prevention and termination therapies combined did not significantly alter burden, total frequency, or symptomatic frequency. The median burden was higher in the ON group during the study period, but this was also observed during the run-in period. The atrial therapies failed to reduce burden and frequency despite an increase in the relative amount of atrial pacing from a median of 75% to 98% and an ATP efficacy of 54%. It is possible that the high proportion of ventricular pacing (median = 99%) may have exerted a deleterious effect and limited any benefit from the atrial therapies. The ATP efficacy of 54% confirms the results of previous findings with this device in patients with bradycardia and AT/AF (6). Although a neutral effect of the atrial therapies on burden and frequency was reported in the previous study, patients were not randomized and the analysis was retrospective (6). In the present study, the effect of the atrial therapies on burden and frequency was evaluated prospectively with strict regulation of device programming. Because the outcome of the prospective analysis was neutral, several retrospective analyses were performed that stratified patients on the basis of demographics, antiarrhythmic drug use, or the presence of episodes during the run-in period. These retrospective analyses failed to identify a responder subgroup.
The failure of the atrial therapies to reduce burden stands in contrast to a previous study that evaluated burden in patients with a history of both atrial and ventricular tachyarrhythmias implanted with a device that had atrial defibrillation capabilities as well as pacing therapies (10). In that study, the median AT/AF burden was significantly reduced from 2.7 to 0.1 h/month in a highly selected subgroup of patients (41/269) who had atrial pacing (prevention and termination) therapies programmed ON, but did not receive atrial shocks. Although the explanation for the difference in the outcome of the two studies is not clear, several factors may have been operative. There were differences in the study designs, patient populations, antiarrhythmic drug use during follow-up, and the pacing algorithms employed for prevention and termination. In addition, episodes with cycle lengths <220 ms were not treated with ATP in the present study but were treated with ATP and/or 50-Hz burst pacing in the previous study (10).
Given the inclusion criteria of at least one symptomatic episode in the three months before implant in the present study, it was interesting that about 40% of the patients in both groups had no episodes during the study period. Because bradycardia may play a role in the initiation of AT/AF, DDDR pacing may have suppressed AT/AF in some patients. Moreover, changes in type or dosage of antiarrhythmic drugs before randomization may have played a role in AT/AF suppression during follow-up.
In addition to providing information on burden and frequency, the device also generated a histogram that grouped episodes into eight different duration bands. Because the precise duration of each episode was not stored, it was not possible to directly link the histogram to burden in a given patient. The histogram showed that the majority of episodes (
80%) in both groups terminated within 10 min. The atrial therapies did not have a significant effect on the frequency of episodes in any of the eight duration bands. The lack of a significant effect of the atrial therapies on the duration histogram is consistent with the lack of an effect of the therapies on burden.
Study limitations. There are several important limitations of this study. The device stores 35 episodes between interrogations and, thus, only about 10% of all episodes logged by the daily counters were available for review. The daily counters did not store the episode type (AT or AF); thus, it was not possible to distinguish AT burden from AF burden. The daily counters also did not store the time of each episode; therefore, the time interval between episodes could not be evaluated. There was no mechanism in place to ensure that patients used their patient activators to log all symptomatic episodes during the study period; therefore, the number of true symptomatic episodes may have been underestimated. Furthermore, a three-month follow-up period after randomization may have been insufficient to adequately characterize burden and frequency.
| Conclusions |
|---|
|
|
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons J. Am. Coll. Cardiol., May 27, 2008; 51(21): e1 - e62. [Full Text] [PDF] |
||||
![]() |
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons Circulation, May 27, 2008; 117(21): e350 - e408. [Full Text] [PDF] |
||||
![]() |
M. E.W. Hemels, J. H. Ruiter, G. P. Molhoek, N. J.G.M. Veeger, A. C.P. Wiesfeld, A. V. Ranchor, M. van Trigt, A. Pilmeyer, I. C. Van Gelder, and for The Features in AT500TM study; Chances for pat Right atrial preventive and antitachycardia pacing for prevention of paroxysmal atrial fibrillation in patients without bradycardia: a randomized study Europace, March 1, 2008; 10(3): 306 - 313. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Mont, R. Ruiz-Granell, J. G. Martinez, J. R. Carmona, M. Fidalgo, E. Cobo, M. Riera, X. Navarro, and on behalf of the Prevention or Termination (POT) S Impact of anti-tachycardia pacing on atrial fibrillation burden when added on top of preventive pacing algorithms: results of the prevention or termination (POT) trial Europace, January 1, 2008; 10(1): 28 - 34. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. W. Israel Studying atrial fibrillation: what can we learn from the AFTherapy study? Europace, December 1, 2007; 9(12): 1107 - 1109. [Full Text] [PDF] |
||||
![]() |
A.J. Camm, N. Sulke, N. Edvardsson, P. Ritter, B.A. Albers, J.H. Ruiter, T. Lewalter, P.A. Capucci, E. Hoffmann, and on behalf of the AFTherapy investigators Conventional and dedicated atrial overdrive pacing for the prevention of paroxysmal atrial fibrillation: the AFTherapy study Europace, December 1, 2007; 9(12): 1110 - 1118. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Padeletti, P. Pieragnoli, V. Jentzen, and A. Schuchert The comorbidity of atrial fibrillation and heart failure: a challenge for electrical therapies Eur. Heart J. Suppl., December 1, 2007; 9(suppl_I): I81 - I86. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. W. Israel What is the role of the pacing rate in the prevention of atrial tachyarrhythmias? Europace, November 1, 2007; 9(11): 999 - 1001. [Full Text] [PDF] |
||||
![]() |
B. K. Kantharia, R. A. Freedman, D. Hoekenga, G. Tomassoni, S. Worley, R. Sorrentino, D. Steinhaus, J. M. Wolkowicz, Z. A. Syed, and and the AOP Study Investigators Increased base rate of atrial pacing for prevention of atrial fibrillation after implantation of a dual-chamber pacemaker: insights from the Atrial Overdrive Pacing Study Europace, November 1, 2007; 9(11): 1024 - 1030. [Abstract] [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, P. E. Vardas, A. Auricchio, J.-J. Blanc, J.-C. Daubert, H. Drexler, H. Ector, M. Gasparini, C. Linde, F. B. Morgado, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm Association Europace, October 1, 2007; 9(10): 959 - 998. [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, P. E. Vardas, A. Auricchio, J.-J. Blanc, J.-C. Daubert, H. Drexler, H. Ector, M. Gasparini, C. Linde, F. B. Morgado, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm Association Eur. Heart J., September 2, 2007; 28(18): 2256 - 2295. [Full Text] [PDF] |
||||
![]() |
S. Petrutiu, A. V. Sahakian, and S. Swiryn Abrupt changes in fibrillatory wave characteristics at the termination of paroxysmal atrial fibrillation in humans Europace, July 1, 2007; 9(7): 466 - 470. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Sulke, J. Silberbauer, L. Boodhoo, N. Freemantle, K. Kamalvand, S. O'Nunain, D. Hildick-Smith, G. W. Lloyd, N. R. Patel, and V. Paul The use of atrial overdrive and ventricular rate stabilization pacing algorithms for the prevention and treatment of paroxysmal atrial fibrillation: the Pacemaker Atrial Fibrillation Suppression (PAFS) study Europace, June 11, 2007; (2007) eum111v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. H. Schoenfeld Contemporary Pacemaker and Defibrillator Device Therapy: Challenges Confronting the General Cardiologist Circulation, February 6, 2007; 115(5): 638 - 653. [Full Text] [PDF] |
||||
![]() |
Writing Committee Members, V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, et al. ACC/AHA/ESC 2006 guidelines for the management of patients with atrial fibrillation: full text: A report of the American College of Cardiology/American Heart Association Task Force on practice guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation) Developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society Europace, September 1, 2006; 8(9): 651 - 745. [Full Text] [PDF] |
||||
![]() |
V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, J. E. Lowe, et al. ACC/AHA/ESC 2006 Guidelines for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation) Developed in Collaboration With the European Heart Rhythm Association and the Heart Rhythm Society J. Am. Coll. Cardiol., August 15, 2006; 48(4): e149 - e246. [Full Text] [PDF] |
||||
![]() |
V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, J. E. Lowe, et al. ACC/AHA/ESC 2006 Guidelines for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation): Developed in Collaboration With the European Heart Rhythm Association and the Heart Rhythm Society Circulation, August 15, 2006; 114(7): e257 - e354. [Full Text] [PDF] |
||||
![]() |
M. Gulizia, S. Mangiameli, S. Orazi, G. Chiaranda, G. Boriani, G. Piccione, N. DiGiovanni, A. Colletti, C. Puntrello, G. Butera, et al. Randomized comparison between Ramp and Burst+ atrial antitachycardia pacing therapies in patients suffering from sinus node disease and atrial fibrillation and implanted with a DDDRP device Europace, July 1, 2006; 8(7): 465 - 473. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Azzolini, G. Critelli, V. De Giorgi, G. B. Del Giudice, G. Ibba, D. Melissano, A. Scaccia, and A. Puglisi Atrial tachyarrhythmia burden modelling by some electrophysiological parameters in pacemaker-recipient patients with Brady-Tachy syndrome Europace, July 1, 2006; 8(7): 474 - 481. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Drago, M. S. Silvetti, G. Grutter, and A. De Santis Long term management of atrial arrhythmias in young patients with sick sinus syndrome undergoing early operation to correct congenital heart disease Europace, July 1, 2006; 8(7): 488 - 494. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Hoffmann, N. Sulke, N. Edvardsson, J. Ruiter, T. Lewalter, A. Capucci, A. Schuchert, S. Janko, J. Camm, and on behalf of the Atrial Fibrillation Therapy (AFT) New Insights Into the Initiation of Atrial Fibrillation: A Detailed Intraindividual and Interindividual Analysis of the Spontaneous Onset of Atrial Fibrillation Using New Diagnostic Pacemaker Features Circulation, April 25, 2006; 113(16): 1933 - 1941. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Gulizia, S. Mangiameli, G. Chiaranda, V. Spadola, N. Di Giovanni, A. Colletti, V. Bulla, A. Circo, O. Pensabene, L. Vasquez, et al. Design and rationale of a randomized study to compare amiodarone and Class IC anti-arrhythmic drugs in terms of atrial fibrillation treatment efficacy in patients paced for sinus node disease: the PITAGORA trial. Europace, April 1, 2006; 8(4): 302 - 305. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E.W. Hemels, A. C.P. Wiesfeld, B. Inberg, P. F.H.M. Van Dessel, W. Nieuwland, E. S. Tan, H. Mulder, D. J. Van Veldhuisen, and I. C. Van Gelder Right atrial overdrive pacing for prevention of symptomatic refractory atrial fibrillation. Europace, February 1, 2006; 8(2): 107 - 112. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Musilli and L. Padeletti Pacemaker selection: time for a rethinking of complex pacing systems? Eur. Heart J., January 2, 2006; 27(2): 132 - 135. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Capucci, M. Santini, L. Padeletti, M. Gulizia, G. Botto, G. Boriani, R. Ricci, S. Favale, F. Zolezzi, N. Di Belardino, et al. Monitored Atrial Fibrillation Duration Predicts Arterial Embolic Events in Patients Suffering From Bradycardia and Atrial Fibrillation Implanted With Antitachycardia Pacemakers J. Am. Coll. Cardiol., November 15, 2005; 46(10): 1913 - 1920. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. P. Knight, B. J. Gersh, M. D. Carlson, P. A. Friedman, R. L. McNamara, S. A. Strickberger, H. F. Tse, A. L. Waldo, and for the AHA Writing Group Role of Permanent Pacing to Prevent Atrial Fibrillation: Science Advisory From the American Heart Association Council on Clinical Cardiology (Subcommittee on Electrocardiography and Arrhythmias) and the Quality of Care and Outcomes Research Interdisciplinary Working Group, in Collaboration With the Heart Rhythm Society Circulation, January 18, 2005; 111(2): 240 - 243. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Padeletti, M. Santini, G. Boriani, G. Botto, A. Capucci, M. Gulizia, R. Ricci, A. Spampinato, P. Pieragnoli, E. Warman, et al. Temporal variability of atrial tachyarrhythmia burden in bradycardia-tachycardia syndrome patients Eur. Heart J., January 2, 2005; 26(2): 165 - 172. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Vollmann, E. Kollet, L. Lüthje, S. Busse, and C. Unterberg Comparison of immediate and delayed automatic antitachycardia pacing for the termination of atrial tachyarrhythmias Europace, January 1, 2005; 7(3): 248 - 254. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Flammang, V. Loteanu, D. Hamani, M. Lambiez, and G. Flammang-Dorie Brady-tachy syndrome: Rapid atrial pacing efficacy in preventing atrial fibrillation recurrence assessed by reliable electrograms: the prefib pilot study Europace, January 1, 2005; 7(5): 482 - 489. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Zimetbaum and M. E. Josephson Is There a Role for Maintaining Sinus Rhythm in Patients with Atrial Fibrillation? Ann Intern Med, November 2, 2004; 141(9): 720 - 726. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. R.J. Mitchell and N. Sulke How do atrial pacing algorithms prevent atrial arrhythmias? Europace, January 1, 2004; 6(4): 351 - 362. [Abstract] [Full Text] [PDF] |
||||
|
|