J Am Coll Cardiol, 2002; 40:2195-2200
© 2002 by the American College of Cardiology Foundation
This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Chouraqui, P.
Right arrow Articles by Sheffy, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chouraqui, P.
Right arrow Articles by Sheffy, J.

CLINICAL STUDY

Assessment of peripheral artery tonometry in the detection of treadmill exercise-induced myocardial ischemia

Pierre Chouraqui, MD, FACC*,*, Robert P. Schnall, DSc{dagger}, Itsik Dvir, DSc{dagger}, Alan Rozanski, MD, FACC{ddagger}, Ehtasham Qureshi, MD{ddagger}, Alexander Arditti, MD§, Jerold Saef, MD, FACC||, Paul D. Feigin, PhD and Jacob Sheffy, PhD{dagger}

* Nuclear Cardiology Unit, The Heart Institute, Chaim Sheba Medical Center, Tel Hashomer, and the Sackler School of Medicine, Tel Aviv University, Tel Aviv, Israel
{dagger} Itamar Medical Ltd., Caesarea, Israel
{ddagger} Department of Medicine, St. Luke’s-Roosevelt Hospital, New York, New York, USA
§ The Heart Institute, Kaplan Medical Center, Rehovot, Israel
|| Batey Cardiovascular Center, Bradenton, Florida, USA
Faculty of Industrial Engineering and Management, Technion, Israel Institute of Technology, Haifa, Israel

Manuscript received December 1, 2001; revised manuscript received June 25, 2002, accepted August 30, 2002.

* Reprint requests and correspondence: Dr. Pierre Chouraqui, Nuclear Cardiology Unit, Heart Institute, Chaim Sheba Medical Center, Tel-Hashomer, 52621, Israel.
chouraq{at}post.tau.ac.il


    Abstract
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
OBJECTIVES: We sought to assess the added diagnostic value of peripheral artery tonometric (PAT) measurements, based on finger pulsatile arterial volume changes, to standard 12-lead stress electrocardiography (ECG), for detecting exercise-induced myocardial ischemia, using single-photon emission computed tomography (SPECT) as the standard of comparison in a double-blinded, multicenter protocol.

METHODS: An automated algorithm for identifying myocardial ischemia from PAT was derived from 345 training cases. The PAT outcome was combined with the ECG result (ischemic, nonischemic, or equivocal), giving a PAT-enhanced value. A threshold of normality was determined to optimize agreement with the SPECT results in the training sample. The PAT-enhanced analysis was then validated in 616 subjects, only two of whom had technically unacceptable PAT studies.

RESULTS: In the validation cohort, receiver operating characteristic curve analysis of the PAT-enhanced diagnosis yielded an area under the curve of 0.72, a sensitivity of 63.5%, compared with 44.7% for ECG alone (p < 0.0001), and a specificity of 67.8% common to both ECG and PAT-enhanced diagnoses. Similar results were found in the training sample. Although over 10% of validation subjects had equivocal ECG results, with the aid of PAT, it was possible to provide diagnostic information for all but one subject.

CONCLUSIONS: Peripheral artery tonometry may be useful for improving the diagnosis of exercise-induced myocardial ischemia by both enhancing the sensitivity without impairing the specificity and increasing the percentage of definitive test results.

Abbreviations and Acronyms
  AUC
  area under the curve
  BP
  blood pressure
  ECG
  electrocardiogram/electrocardiography
  PAT
  peripheral artery tonometer/tonometry
  PWA
  pulse wave amplitude
  ROC
  receiver operating characteristics
  SPECT
  single-photon emission computed tomography


Of all currently available noninvasive diagnostic techniques for detecting myocardial ischemia during exercise stress testing, electrocardiography (ECG) is the simplest and most widely used (1). The main limitation of ECG exercise testing is its relatively low accuracy (2). We postulated that an additional channel of physiologic information related to peripheral vascular responsiveness during exercise could supplement standard stress ECG and provide additional diagnostic value for the detection of exercise-induced myocardial ischemia. The basis for this speculation lies in earlier reports that myocardial ischemia induced by balloon inflation during percutaneous transluminal coronary angioplasty of major coronary arteries was found to be associated with increased systemic vascular resistance tightly linked to the induced ischemia (3). Further studies reported that increases in systemic vascular resistance were associated with myocardial ischemia (4,5).

Recent studies utilizing a specially designed device for measuring the peripheral artery tone of the finger—the peripheral artery tonometric (PAT) device, which was originally developed by one of the authors (Dr. Schnall) (6)—indicate that pulsatile blood volume through peripheral regions is subject to intense transient vasoconstriction during clinical events that involve activation of the sympathetic nervous system, such as the arousal phase from sleep apnea (6,7) and during rapid-eye-movement sleep (8).

In light of these findings, it was thought that finger vasomotor changes might provide appropriate hemodynamic information that could be associated with the occurrence of myocardial ischemia in clinically relevant situations.

In a recent study, a high concordance between exercise-induced, PAT-detected peripheral vasoconstriction and single-photon emission computed tomographic (SPECT) evidence of exercise-induced reversible myocardial hypoperfusion has been shown (9). Another study showed a significant difference between the PAT signal patterns of healthy volunteers and patients with documented coronary artery disease, where the latter group showed a higher incidence of vasoconstriction (10).

The goal of this study was to assess the possible additional diagnostic value of finger peripheral artery tone changes using a fully automated algorithmic system in combination with ECG, as compared with ECG alone, for detecting exercise-induced myocardial ischemia.


    Methods
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
Study population and test conditions.   This was a prospective, multicenter study involving a total of 1,053 randomly selected subjects (820 males; mean age 56.2 ± 11 years) who were referred for diagnostic exercise myocardial perfusion SPECT imaging. These tests were independently performed at four medical centers: Sheba Medical Center, Tel Hashomer; Kaplan Medical Center, Rehovot, Israel; St. Luke’s-Roosevelt Medical Center, New York, New York; and Batey Medical Center, Bradenton, Florida. The study was approved by each hospital’s ethics committee. Patients gave informed, written consent after the purpose and nature of the study were explained. Patients with known nonischemic cardiomyopathy, chronic severe pulmonary disease (asthma or chronic obstructive pulmonary disease), or previous coronary artery bypass graft surgery were excluded. Patients on vasoactive drugs were instructed to stop taking short-acting nitrates for 6 h, long-acting nitrates and calcium channel blockers for 24 h, and beta-blockers for at least 48 h. Before the test, patients fasted for 3 h. Tests were conducted in a thermally controlled environment (21°C to 23°C).

The efficacy analysis involved two stages; in both, the SPECT results served as a gold standard. In the first stage, a random sample of 314 subjects served as the training sample for developing the analysis algorithm. The remaining 708 subjects served as the validation set, 46 of whom were omitted from the analysis because of a nondiagnostic ECG (e.g., complete left bundle branch block, left ventricular hypertrophy with a strain pattern, Wolf-Parkinson-White syndrome, nonspecific ST-T–segment changes) and 2 because of technical failure of PAT; another 44 subjects had a nondefinitive (equivocal) SPECT study, leaving a total of 616 subjects (497 males; overall mean age 55.3 ± 10.5 years; body mass index 27.3 ± 4.0 kg/m2) who completed the study according to the protocol. We chose not to include the equivocal SPECT results when making the comparison between stress ECG and PAT results, as the SPECT findings constituted our referent standard for comparison of ECG versus PAT in our study, and the statistical analysis required that the independent variable consist of only two categories

Description of the PAT device and PAT signal
A detailed description of the device has been previously described in reports related to other clinical conditions (6,8,10).

The essential features of the PAT device are as follows: 1) a specialized probe and a computer-based system that pneumatically controls the probe and filters (0.3 to 30 Hz) and digitizes and records the PAT signal; 2) a uniform pressure field surrounding the whole of the distal part of the finger, including the finger tip; this facilitates a desirable arterial wall unloading effect, while inhibiting distal venous distention, which can cause veno-arterial–mediated vasoconstriction (11); 3) a proximal extension of the pressure field, which buffers the sensing region at the finger’s tip from extraneous and artifact signals such as perturbations in the venous system; and 4) an active clamp to the finger, due to the split thimble design. These features are illustrated in Figure 1.



View larger version (31K):
[in this window]
[in a new window]
 
Figure 1 Schematic diagram of the sensor’s structure. The sensor is partitioned into two separate sections that are independently pressurized. The sensor cap has a split thimble design that imparts a two-point clamping effect to lock the sensor to the finger tip while measuring pulsatile volume changes. The adjacent open-ended annular cuff provides a buffering effect against retrograde shock waves, to reduce noise, but is not used for sensing. Both compartments are designed to keep the venous transmural pressure negative to prevent venous pooling and distention and to ensure that only arterial volume changes are recorded.

 
Representative time courses of PAT signals are given in Figure 2. A normal PAT response (upper tracing) typically shows a progressive increase in the signal amplitude, whereas a progressive decrease (lower tracing) is the hallmark of a vasoconstrictive response, typically representing an abnormal PAT response.



View larger version (26K):
[in this window]
[in a new window]
 
Figure 2 Representative examples of the peripheral arterial tonometer (PAT) signal’s time course in PAT-normal and PAT-abnormal cases. (Upper tracing) A negative PAT responder showing the typical progressive increase in the signal throughout exercise. (Lower tracing) A progressive decrease in PAT signal amplitude, which was the hallmark of the vasoconstrictive response, representing a typical abnormal PAT response. Inserts show higher resolution samples at the start and end of exercise periods.

 
PAT signal processing
Processing of the PAT signal involved the removal of noisy signals and the subsequent derivation of trends of pulse wave amplitude (PWA), heart rate, and a reflection coefficient parameter (k3), describing wave propagation properties of the signal (12) and, thus, is related to the contractile properties of the arterial bed. These trends were derived before, during, and after exercise. A function of the percent predicted maximal heart rate and PWA product at peak exercise, the k3 value at baseline, and the ECG outcome was then generated to predict the SPECT outcome.

The rationale for combining the percent maximal heart rate and PWA features was to take into account the fact that subjects reached relatively different levels of stress and predicted maximal heart rates.

Stress testing procedures
The PAT and standard 12-lead ECG measurements were obtained continuously before, during, and after the exercise stress test. A multistage symptom-limited treadmill exercise test was performed according to the Bruce protocol, where patients were encouraged to reach at least 85% of their predicted maximal heart rate (220 – age).

Exercise was terminated before reaching 85% of the predicted maximal heart rate if clinical symptoms (e.g., severe chest pain, hypotension, severe hypertension) or significant ECG changes (>2.5-mm ST-segment depression, ST-segment elevation, three or more consecutive ventricular premature beats) occurred or if fatigue or dyspnea limited further exercise performance. Blood pressure (BP) was measured by standard sphygmomanometry before and after exercise and at the conclusion of each stage of exercise.

Myocardial perfusion imaging
The SPECT imaging was performed using conventional methodology with either thallium-201 (3 to 4 mCI) or technetium-99m sestamibi (9 to 10 mCi at rest; 30 to 31 mCi at peak exercise). Standard SPECT myocardial scintigrams were independently assessed using standard techniques (13). Myocardial SPECT results were classified as normal, equivocal, or abnormal (reversible hypoperfusion).

Assessment of exercise stress ECG
The ECG response during stress was considered ischemic if horizontal or downsloping ST-segment depression of ≥1 mm or upsloping ST-segment depression of ≥1.5 mm occurred at 0.08 s after the J point, relative to the baseline ECG. The response was considered equivocal if the ECG demonstrated 1.0 to 1.4 mm of upsloping or 0.5 to 0.9 of horizontal or downsloping ST-segment depression, relative to the baseline ECG.

PAT algorithm and statistical analysis
The objective of the PAT algorithm was to optimize agreement between the combined PAT responses/ECG results and the SPECT results. This was performed using a logistic regression model to generate an overall test value based on the PAT features of the training set signals, and then the ECG outcomes were classified as ischemic, nonischemic, or equivocal. When considered as independent predictors of the SPECT outcome, multivariate analysis showed that the PAT amplitude ratio was the most significant predictive factor, followed by ECG, percent maximal heart rate, and finally the k3 value. A threshold of normality for the PAT-enhanced value was then determined to optimize agreement with the SPECT results in the training sample.

After applying the PAT algorithm to the PAT data of the validation set and calculating the overall score when the ECG result was added, a receiver operating characteristic (ROC) curve was derived. The area under the receiver operating characteristic curve (AUC) provides a measure of the overall efficacy of the score. McNemar’s test, which provides an evaluation of the significance of difference in the number of accurate determinations for a given test, was used to evaluate the contribution of PAT to the efficacy of diagnosis of exercise-induced myocardial ischemia. This was used to compare the combined PAT/ECG analysis with the ECG diagnosis when considered in terms of the SPECT result.

McNemar’s test was also used to test the significance of differences when the study population was segmented according to a variety of clinically and demographically relevant criteria. Statistical significance was set at p < 0.05.


    Results
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
Sensitivity and specificity values and ROC curves at the specific ECG working points.   Figure 3 shows the ROC curve for the PAT-enhanced determinations in the 616 validation subjects. The AUC for this sample was 0.72 (p < 0.0001). The same AUC value was also obtained for the PAT-enhanced value in the training set, demonstrating the robust nature of the analysis. Also shown in Figure 3 is the ROC curve based on the PAT alone response data only, without ECG, for the same 616 subjects. The AUC in this case was 0.675.



View larger version (17K):
[in this window]
[in a new window]
 
Figure 3 (Thickest curve) The receiver-operating characteristic (ROC) curve for the validation sample of 616 subjects in whom the peripheral arterial tonometer (PAT) outcome was used in combination with the electrocardiograph (ECG) result (PAT enhanced). The area under the curve (AUC) was 0.72. At a specificity value of 67.8%, the sensitivity of the combined PAT and ECG model was 62.4%, compared with 44.7% for ECG alone (p < 0.0001). The specificity value selected represents the specificity value of ECG for this study population. (Thinner curve) The ROC curve for the validation sample of 616 subjects in whom only the PAT outcome (PAT alone) was used. The AUC was 0.675. At a specificity value of 67.8%, the sensitivity of PAT was 58.4%, compared with 44.7% for ECG alone (p < 0.01).

 
The reference point marked on the ROC curve represents ECG specificity (67.8%) and allows the different sensitivity values for ECG, the PAT alone value, and the PAT-enhanced value to be compared at matching levels of specificity. At this specificity, ECG gave a sensitivity of 44.7%, whereas the sensitivity for the PAT-enhanced assessment was 63.5% (chi-squared statistic = 35.10, p < 0.0001) and that for PAT alone was 58.4% (chi-squared statistic = 7.67, p < 0.01). The sensitivity of the PAT diagnosis without ECG was not significantly different from the PAT-enhanced diagnosis. The PAT-enhanced ECG assessment increased the sensitivity by over 18%, while maintaining the specificity at a fixed value. When PAT was used alone, it increased the sensitivity by 14%, compared with exercise ECG alone. Very similar values were observed in the training sample, with a specificity of 70.1% and ECG sensitivity was 46.1%, whereas the sensitivity for PAT-enhanced assessment was 65.8% (chi-squared statistic = 13.24, p < 0.0003) and that for PAT alone was 52.6% (p = NS). The sensitivity of the PAT diagnosis without ECG was not significantly different from that of the PAT-enhanced diagnosis. These results are summarized inTable 1.


View this table:
[in this window]
[in a new window]
 
Table 1 ECG, PAT Alone, and PAT-Enhanced Sensitivities and Specificities

 
Sensitivity and specificity of PAT and ECG by demographic data and clinical features
Table 2 summarizes the specificity and sensitivity results when comparing dichotomous groupings of patients based on certain demographic and clinical features of the patient population. None of the comparisons were statistically significantly different, suggesting that the PAT-enhanced diagnosis provided similar results for the opposing sides of each category. The ECG sensitivity and specificity results were also not significantly different in any of these comparisons.


View this table:
[in this window]
[in a new window]
 
Table 2 Sensitivities and Specificities of Stress ECG and PAT Based on Demographic Data and Clinical Features

 
Comparative ECG sensitivity and specificity values show that the diagnostic advantage of PAT-enhanced analysis over ECG alone was consistent throughout the range of clinical and demographic subgroups studied.

Diagnostic value of PAT versus ECG
Nondiagnostic or equivocal outcomes of ECG were frequent occurrences in this study, as were SPECT equivocal findings. Overall, 17.5% of the ECG tests failed to provide a definitive diagnosis (10.9% equivocal, 6.5% nondiagnostic), whereas SPECT was equivocal in 6.4% of the subjects, and PAT was nondiagnostic in only 0.3% of the subjects and equivocal in none.


    Discussion
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
We assessed the diagnostic value of adding a novel technique, based on the measurement of vasomotor responses of the finger, to standard 12-lead ECG scoring for detecting exercise-induced myocardial ischemia.

Possible significance of finger vasomotor responses during circulatory adjustments to exercise.   The major peripheral vascular change associated with exercise-induced myocardial ischemia was the degree of vasoconstriction. The mechanism for vasoconstriction during exercise is poorly understood. Previous studies of the interaction between exercise and thermal stress have shown that skin blood flow regulation is affected by both baroregulatory and thermoregulatory homeostatic requirements (14). As the core temperature increases, skin vessels dilate (15). In the face of impaired myocardial contractility due to ischemia, a tendency of peripheral vasoconstriction to increase peripheral resistance is important for maintaining BP, but, however, would result in reduced heat dissipation (14).

The palmar surfaces of the hands and fingers are important sites of thermoregulation (16–18) and may be better suited for measuring exercise-related vasomotor changes versus generalized systemic vascular changes. Goldberg et al. (19) showed that exercise-induced myocardial ischemia occurs amid an overall reduction in systemic vascular resistance, so it is probable that measuring total peripheral resistance could fail to identify the transition of peripheral vascular tone in the manner demonstrated by finger PAT (9,10). Therefore, it appears that the choice of the peripheral site for vascular tone measurement is crucial to the appropriate determination of myocardial ischemia during exercise.

Other factors affecting finger PWA during exercise
Several other factors can have immediate short-term effects on finger blood flow. These include local finger temperature changes, heating or cooling of the torso, and other large body masses. Because the studies were performed in a temperature-controlled environment, it is unlikely that any of these factors could have played a significant role or have selectively biased the ischemic patients.

Reductions in systemic BP or local finger BP are also improbable causes of finger PWA during exercise, because it was found that during exercise, diastolic BP was generally maintained and systolic BP generally showed large work load-related increases. Thus, the pulse pressure at the more distal finger vasculature was at least maintained if not increased. Because the amplitude of the pulsatile volume change recorded by PAT is a function of both the effective pulse pressure and the local vascular compliance, it is most likely that a reduction in compliance (i.e., vasoconstriction) was the cause of the reduction in the pulsatile volume amplitude.

Clinical implications
Our validated results showed that the PAT device contributed to an approximately 40% increase in diagnostic sensitivity over ECG, with no loss of specificity. This benefit was quite consistent in referred populations with substantially different ECG sensitivity levels (42.6% in Israel vs. 61.9% in the U.S.). It was also consistent among populations with substantially different positive SPECT prevalences (40.7% in Israel vs. 11.0% in the U.S.) and between patients taking cardiac medications (49.8% vs. 21.4%) and those not taking cardiac medications. The PAT-enhanced diagnosis also proved to be very robust, as it was able to provide a diagnosis in virtually all subjects and facilitated the conversion of equivocal ECG results to definitive diagnoses, without impairing the specificity.

In contrast to imaging methods and, to some extent, ECG, the PAT method has two limitations: 1) it is not configured to provide information on the site of coronary pathology, as it does not actually derive information from the myocardium itself; and 2) it currently does not provide an index of the severity of disease. Further studies are required to address these issues. Application of the PAT method would be possible, however, in virtually any clinical exercise stress testing setting, because the analysis is automatic and entirely user-independent, and the device is extremely simple to use.

The very close agreement between ECG sensitivity and specificity values for the training and validation populations shows that the independent findings were consistent, thus lending support to the accuracy of the determined ECG performance relative to thallium SPECT in this study. The close agreement between PAT-enhanced sensitivity and specificity values for the two independent patient populations in this study demonstrates the robust nature of the measurement.

The relative sensitivity of ECG for the detection myocardial ischemia by SPECT imaging was relatively low in this study; moreover, it varied by nearly 20% between the Israeli and U.S. subpopulations. We have previously noted that "apparent" test accuracy can vary markedly according to the presence and magnitude of various pre-test and post-test referral biases (20,21). The effect of referral bias is also present when evaluating the apparent accuracy of exercise ECG testing in a population referred for stress radionuclide testing, as there is a preferential tendency to refer either "nondefinitive" or discordant exercise ECG test responders for radionuclide stress testing. Such referral could tend to result in an underestimation of the true sensitivity of exercise ECG relative to SPECT evidence of ischemia. Furthermore, differences in clinical practice in Israel versus the U.S. could serve to accentuate the potential effects of referral bias, although this would require more prospective studies to evaluate it formally. The differences in test sensitivity among male and female subjects that was observed in this study were probably also secondary to the impact of referral bias, as the female population in our study constituted a subgroup in which the prevalence of SPECT ischemia was much lower than that in our male population. The limited number of females with SPECT-positive results in this study further precludes the drawing of firmer conclusions.

Conclusions
In light of: 1) the improved performance of the combined ECG and PAT analysis over ECG alone; 2) the increased yield of definitive outcomes; 3) the robustness of the combined ECG and PAT analysis; and 4) the benign and simple nature of the PAT technique, we believe that this approach may have a useful contributory role in the diagnosis of exercise-induced myocardial ischemia.


    Acknowledgments
 
We thank Barry L. Zaret, MD, from Yale University, for reviewing this manuscript and offering very helpful advice.


    Footnotes
 
This study was supported in part by a grant from Itamar Medical Ltd.


    References
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
1. Berman DS, Kiat H, Friedman JD, Diamond G. Clinical applications of exercise nuclear cardiology studies in the era of healthcare reform. Am J Cardiol. 1995;75:3–13D

2. Philbrick JT, Horwitz RI, Feinstein AR. Methodologic problems of exercise testing for coronary artery disease: groups, analysis and bias. Am J Cardiol. 1980;46:807–812[CrossRef][Medline]

3. Mohr R, Rath S, Meir O, et al. Changes in systemic vascular resistance detected by the arterial Resistometer: preliminary report of a new method tested during percutaneous transluminal coronary angioplasty. Circulation. 1986;74:780–785[Abstract/Free Full Text]

4. Mohr R, Meir O, Smolinsky A, Goor DA. A method for continuous on-line monitoring of systemic vascular resistance (COMS) after open-heart procedures. J Cardiovasc Surg. 1987;28:558–565[Medline]

5. Mohr R, Dinbar I, Bar-El Y, Goldbourt U, Abel M, Goor DA. Correlation between myocardial ischemia and changes in arterial resistance during coronary artery bypass surgery. J Cardiothorac Vasc Anesth. 1992;6:33–41[Medline]

6. Schnall RP, Shlitner A, Sheffy J, et al. Periodic, profound peripheral vasoconstriction—a new marker of obstructive sleep apnea. Sleep. 1999;22:939–946[Medline]

7. Pillar G, Malholtra A, Fogel R, et al. Detection of obstructive sleep disordered breathing events by utilizing peripheral arterial tonometry and oximetry. (abstr)Sleep. 2000;23:A258

8. Lavie P, Schnall RP, Sheffy J, et al. Peripheral vasoconstriction during REM sleep detected by a novel plethysmographic device. Nat Med. 2000;6:606[CrossRef][Medline]

9. Qureshi EA, Carnendran L, Bauman M, et al. Cold hands, cold heart? Paradoxical finger blood flow responses during treadmill exercise. (abstr)Circulation. 2000;102(Suppl 2):538

10. Rozanski A, Qureshi E, Bauman M, Reed G, Pillar G, Diamond GA. Peripheral arterial responses to treadmill exercise among healthy subjects and atherosclerotic patients. Circulation. 2001;103:2084–2089[Abstract/Free Full Text]

11. Henriksen O, Sejrsen P. Local reflex in microcirculation in human cutaneous tissue. Acta Physiol Scand. 1976;98:227–231[Medline]

12. Porat B. Digital Processing of Random Signals. Englewood Cliffs, NJ: Prentice Hall, 1994

13. Wackers F, Soufer R, Zaret BL. Nuclear cardiology. Braunwald E, Zipes DP, Libby P. Heart Disease: A Textbook of Cardiovascular Medicine. 6th ed. Philadelphia: W. B. Saunders; 2000. p. 273–323

14. Rowell LB, O’Leary DS, Kellog DL Jr. Integration of cardiovascular control systems in dynamic exercise (chapter 17). In: American Physiological Society, Handbook of Physiology. Section 12: Exercise Regulation and Integration of Multiple Systems. Oxford, England: Oxford University Press, 1996;770–838

15. Nielsen B, Rowell LB, Bonde-Petersen F. Cardiovascular responses to heat stress and blood volume displacements during exercise in man. Eur J Appl Physiol. 1984;52:370–374

16. Burton AC. The range and variability of the blood flow in the human fingers and the vasomotor regulation of body temperature. Am J Physiol. 1939;127:437–453[Free Full Text]

17. Russell A, Kappagoda T. Reactive hyperemia in patients with Reynaud’s phenomenon. J Rheumatol. 1988;15:1653–1657[Medline]

18. Hassan AAK, Rayman G, Tooke JE. Effects of indirect heating on the postural control of skin blood flow in the human foot. Clin Sci. 1986;70:577–582[Medline]

19. Goldberg AD, Becker LC, Bonsall R, et al. Ischemic, hemodynamic and neurohumoral responses to mental and exercise stress: experience from the Physiological Investigation of Myocardial Ischemia (PIMI) study. Circulation. 1996;94:2402–2409[Abstract/Free Full Text]

20. Rozanski A, Diamond GA, Berman D, Forrester JS, Morris D, Swan HJC. Specificity of exercise radionuclide ventriculography. N Engl J Med. 1984;310:594–595[Medline]

21. Rozanski A, Diamond GA, Forrester JS, Berman D, Morris D, Swan HJC. Comparison of alternative referent standards for cardiac normality: implications for diagnostic testing. Ann Intern Med. 1984;101:164–171[Abstract/Free Full Text]




This article has been cited by other articles:


Home page
Psychosom. Med.Home page
M. M. Burg, B. Graeber, A. Vashist, D. Collins, C. Earley, J. Liu, R. Lampert, and R. Soufer
Noninvasive Detection of Risk for Emotion Provoked Myocardial Ischemia
Psychosom Med, January 1, 2009; 71(1): 14 - 20.
[Abstract] [Full Text] [PDF]


Home page
Emerg. Med. J.Home page
M Gorenberg, A Marmor, and H Rotstein
Detection of chest pain of non-cardiac origin at the emergency room by a new non-invasive device avoiding unnecessary admission to hospital
Emerg. Med. J., July 1, 2005; 22(7): 486 - 489.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Chouraqui, P.
Right arrow Articles by Sheffy, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chouraqui, P.
Right arrow Articles by Sheffy, J.

 
  CME Topic Collections Past Issues Search Current Issue Home

Advertisement