Early and late complications associated with transcatheter occlusion of secundum atrial septal defect
Massimo Chessa, MD, PhDa,*,
Mario Carminati, MDa,
Gianfranco Butera, MD, PhDa,
Roberta Margherita Bini, MDa,
Manuela Drago, MDa,
Luca Rosti, MDa,
Alessandro Giamberti, MDa,
Giuseppe Pomè, MDa,
Eduardo Bossone, MD, PhDa and
Alessandro Frigiola, MDa
a Pediatric Cardiology, Istituto Policlinico San Donato, San Donato Milanese, Italy
Manuscript received June 28, 2001;
revised manuscript received December 4, 2001,
accepted December 21, 2001.
* Reprint requests and correspondence: Dr. Massimo Chessa, Pediatric Cardiology, Istituto Policlinico San Donato, Via Morandi, 30, 20097 San Donato Milanese, Italy. massimo.chessa{at}lycos.com
 |
Abstract
|
|---|
OBJECTIVES: The goal of this study was to report the early and late complications experienced in atrial septal defect (ASD) transcatheter closure.
BACKGROUND: Atrial septal defect transcatheter occlusion techniques have become an alternative to surgical procedures. A number of different devices are available for transcatheter ASD closure. The type and rate of complications are different for different devices.
METHODS: Between December 1996 and January 2001, 417 patients (mean age: 26.6 ± 19 years) underwent transcatheter occlusion of secundum type ASD. Complications were categorized into major and minor. Two different devices were used: the CardioSEAL/STARFlex in 159 patients and the Amplatzer septal occluder in 258 patients.
RESULTS: Thirty-four patients experienced 36 complications during the hospitalization (8.6%, 95% confidence interval: 6.1% to 11.1%). Ten patients underwent elective surgical repair because of device malposition (three patients) or device embolization (seven patients). Twenty-four patients experienced 25 minor complications: unsatisfactory device position or embolization. Devices were retrieved using a gooseneck snare and/or a basket; 11 patients experienced arrhythmic problems. Other complications were: pericardial effusion, thrombus formation on the left atrial disc, right iliac vein dissection, groin hematoma, hemorrhage in the retropharynx and sizing balloon rupture. Two patients had late complications: peripheral embolization in the left leg one year after implantation of an Amplatzer device and sudden death 1.5 year later.
CONCLUSIONS: Our series of patients with ASD by transcatheter occlusion shows that the procedure is safe and effective in the vast majority of cases. To further reduce the complications rate, the criteria of device selection according to ASD morphology and some technical tips during implantation are discussed.
|
Abbreviations and Acronyms
| | CS/SF | | ASD | | atrial septal defect | | ASO | | Amplatzer septal occluder | | CS/SF | | CardioSEAL/STARFlex | | TEE | | transesophageal echocardiography |
|
The secundum type atrial septal defect (ASD) is the fourth most common congenital heart defect with an incidence of 3.78 per 10,000 live births (1), corresponding to 5.9% of diagnosed congenital heart disease in children (2). The ASD can now be treated by transcatheter occlusion technique. We report the early and late complications experienced in ASD transcatheter closure.
 |
Methods
|
|---|
Patients.
Between December 1996 and January 2001, 417 patients (mean age: 26.6 ± 19 years) underwent transcatheter occlusion of secundum type ASD (Table 1). Of 417 patients, 34 experienced complications due to the procedure. Complications were categorized into major and minor (3). Major complications included all events leading to one of the following: 1) death; 2) life-threatening hemodynamic decompensation requiring immediate therapy; 3) need for surgical intervention; and 4) significant permanent anatomic or functional lesion resulting from catheterization. Minor complications were defined as events that were transient and resolved with specific treatment.
Devices and implantation technique.
Two different devices were used: the CardioSEAL/STARFlex (CS/SF) (from 1996) in 159 patients and the Amplatzer septal occluder (ASO) (from 1999) in 258 patients.
The CS/SF (Nitinol Medical Technical Inc., Boston, Massachusetts) is a second-generation double umbrella device that was developed from the Clamshell occluder. It consists of a metallic framework covered, in umbrella-like fashion, by knitted polyester fabric. The device is available in five different sizes, 17, 23, 28, 33 and 40 mm, which can be delivered through an 11F long sheath. The CardioSEAL has been recently modified (StarFLEX) by the addition of a flexible self-centering mechanism, and it can be inserted on a front loading system of a 10F.
The ASO is constructed from 0.004-in. to 0.0075-in. nitinol wires that are tightly woven into two flat buttons (discs) with a 4-mm connecting waist that dictates the device diameter. The device is available from 4-mm to 38-mm. There are a total of three Dacron polyester patches sewn securely with polyester thread into each disc and the connecting waist to increase the thrombogenicity of the device. The device is delivered through a 6F up to 12F long sheath.
The criteria for using each device in our center changed during time; from 1996 to 1999 only the CS/SF has been used. After 1999, when the ASO became available, we preferred to close an ASD bigger then 18 mm only with the new device.
In all patients, implantation was carried out under general anesthesia. Before starting the catheterization, transesophageal echocardiographic examination was undertaken using a multiplane probe. Standard catheterization of the right heart was then performed through the right femoral vein, taking recordings of pressures and blood samples to calculate Qp/Qs ratio. Angiographic visualization of the defects was achieved by injection of a contrast material in the left atrium or the right upper pulmonary vein in the left anterior oblique view with cranial angulation. Heparin (100 IU/kg) and antibiotic prophylaxis were given routinely. Meditech or Numed balloons were used to establish the stretched diameter of the defect. Implantation was performed under fluoroscopic and echocardiographic control. After positioning a long sheath in the left atrium, the device was attached to its delivery wire and advanced within the sheath until the distal disc was deployed in the left atrial cavity. Both the sheath and delivery system were then slowly withdrawn towards the atrial septum. The proper position of the distal disc is confirmed on transesophageal echocardiography (TEE), and then the proximal disc was opened on the right side of the atrial septum by withdrawing the sheath. Interference of the device with the caval or pulmonary veins or with the atrioventricular valves was checked echocardiographically once deployment was complete. If all findings were satisfactory, the device was released.
Protocol for follow-up.
All patients underwent clinical examination, electrocardiography, chest radiography in two projections and transthoracic echocardiography before discharge. The same procedures were performed at 1, 6 and 12 months after the implantation. Aspirin, at a dose of 5 mg kg daily, was recommended from six months after implantation.
 |
Results
|
|---|
The overall incidence of complications was 8.6% (95% confidence interval: 6.1% to 11.1%) (36/417). Thirty-four patients experienced 36 complications during the hospitalization: 11 major and 25 minor. This incidence includes the learning curve for each device. The embolization/malposition was the most common complication (3.5%), accounting for almost half of the events (16 complications in 15 patients; 47% of all complications).
Seven patients had device embolization that needed surgical retrieval. In four patients the device was a CS/SF (Patients 1, 3, 4, 10); the other three patients had an ASO implanted (Patients 19, 23, 29). Three more patients underwent elective surgical repair because of malposition of the device (Patients 6, 13, 16). In four patients the device was delivered but, because of an unsatisfactory position (Patients 2, 7, 24) or embolization (Patient 8), it was retrieved using a gooseneck snare and/or a basket; a second device was successfully implanted. The embolized/malpositioned CS/SF were 40-mm in five cases, 33-mm in four cases and 28-mm in two cases. Embolization of the ASO occurred with devices approximately or bigger than 24-mm. Malposition occurred with 19-mm and 22-mm devices, and, in one of those, it was related to accidental detachment of the device inside the delivery sheath. Several maneuvers had been performed in order to find the proper position for the deployment of the distal disc unscrewing the device from its delivery cable. One patient (Patient 11) experienced both malposition (CS/SF 33-mm) and embolization in the right ventricle at the second attempt (CS/SF 40-mm) 6 h after implantation. The ASD was closed with an ASO 26-mm. One patient (Patient 24) had a malposition of his device (CS/SF 28-mm); using a gooseneck snare, the device was pulled back into a 14F introducer up to the common femoral vein where it was surgically retrieved. The ASD was then closed using a CS/SF 33-mm. Arrhythmias are the next most common complication (2.6%). Eleven patients experienced arrhythmic problems: atrial fibrillation requiring electrical cardioversion (Patients 13, 14, 15, 18, 23, 31), atrial fibrillation with spontaneous resolution (Patients 11, 22) and supraventricular tachycardia with spontaneous resolution (Patients 12, 26). In Patient 23 during electrical cardioversion the device embolized a few minutes after implantation; at the TEE control, no residual shunt was detected before cardioversion. Patient 9 had a complete atrioventricular block immediately after device implantation. The device (ASO 18-mm) was then removed with a complete sinus rhythm recovery 3 h later. He underwent successful closure with a 14-mm ASO one year later. Electrical cardioversion was effective in five cases. In the remaining patients, sinus rhythm returned spontaneously. None of these patients experienced palpitations before the procedure, and the electrocardiogram was normal for all of them.
Pericardial effusion was experienced by two patients (Patients 17 and 27). Patient 17 was treated medically and showed complete resolution of the effusion 20 days later; the preprocedure echocardiogram on this patient did not show pericardial effusion. The mechanism of this complication in that patient is not clear.
Patient 27 had a perforation due to the guidewire and needed surgical drainage of the effusion in the cath lab. This patient underwent complete ASD closure.
Patient 20 had a thrombus formation on the left atrial disc immediately after delivery; she was fully heparinized. She was treated with anticoagulant therapy for six months with complete resolution at the TEE control (Fig. 1). There was right iliac vein dissection (Patient 5) treated by implantation of two stents, groin hematoma (Patient 25) requiring surgical revision, hemorrhage in the retropharynx (Patient 28).

View larger version (33K):
[in this window]
[in a new window]
|
Figure 1 Transesophageal echocardiography follow-up of a patient with thrombus in the left Amplatzer septal occluder (ASO) disc. (A) Thrombus (arrow) in the left ASO disc immediately after device implantation. (B) Thrombus reduction (arrow) after three months anticoagulant therapy. (C) Left ASO disc thrombus free (arrow) after six months anticoagulant therapy. LA = left atrium; RA = right atrium.
|
|
In Patient 33 the sizing balloon ruptured (MediTech 33-mm) and was retained along the long wire. The ruptured balloon was retrieved with a basket catheter from the right femoral vein. Two patients had late complications. Patient 30 was reported to have experienced peripheral embolization in the left leg one year after implantation of an Amplatzer device. She required surgical retrieval of the distal peripheral embolus followed by anticoagulation treatment for one month. Patient 34 died suddenly 1.5 years later. No post mortem data are available, and we do not know whether the event is related to the device implanted.
 |
Discussion
|
|---|
In the last 50 years cardiac catheterization has changed its primary role from a diagnostic investigation to that of a therapeutic procedure. Most recently, ASD transcatheter occlusion techniques have become an alternative to surgical procedure using cardiopulmonary bypass. A number of different devices are available for transcatheter ASD closure, all of them with demonstrated advantages and disadvantages. The type and rate of complications are different among different devices (410) (Table 2). In our experience the embolization/malposition was the most common complication. Devices usually embolize in the main pulmonary artery (89% in our group). Once the device embolizes, two different options are possible: 1) retrieve the device by a gooseneck snare or a basket catheter, 2) refer the patient to the surgeon. The last option is indicated when the size of the device is among the largest; the surgeon will retrieve the device and close the ASD at the same time. If we decide to attempt the retrieval, an introducer of at least 12F to 14F size is inserted to accommodate the device captured by a gooseneck snare or a basket catheter. Such an attempt can be carried out with CS/SF. An embolized/malpositioned ASO, in our experience, is irretrievable. Among the five cases with the CS/SF 40-mm device, at least four had a satisfactory position after release; nevertheless, they embolized within 24 h. Therefore, we became convinced that arms are too weak to grab the septal margins and hold the device in place with sufficient stability. The 40-mm device was then abandoned in our practice, and the ASO device became the first choice for closing defects larger than 18 mm. Our personal guidelines are: - for ASD larger than 18 mm, use ASO device;
- where several maneuvers are needed for positioning ASO, remember to screw the cable on the device again to avoid accidental detachment; and
- once the ASO has detached from its cable, it becomes difficult to retrieve.
Arrhythmias are the next most common complication. For patients who experienced atrial fibrillation or supraventricular tachycardia after ASO device implantation, a possible explanation could be the stretching of the interatrial septum by the central waist of the device.
Thrombus formation in the left atrial disk immediately after the procedure was experienced in a 67-year-old patient suffering from systemic hypertension and with evidence of spontaneous contrast on echocardiogram. To avoid this complication, it is now our policy to start oral antiaggregation therapy one day before the procedure in all patients. We do not start anticoagulation therapy because we do not have criteria to define the high-risk patients that need it. The rupture of the sizing balloon is an unpredictable event because it seems to be related to the balloon itself. This complication was managed by locating the balloon outlined by some contrast material along the long wire and by retrieving it using a basket catheter.
The other complications (right iliac vein dissection, groin hematoma, retropharynx hemorrhage) were related to human mismanagement during the procedure. The major complication reported in Patient 34, that is, sudden death, is quite worrisome and needs further investigations in other large series. We cannot understand how a device implanted 1.5 year before could cause sudden death. It is, however, our responsibility to report it because there are no other known reasons to explain this event in a young man of good health of 34 years of age. Peripheral embolization in Patient 19 was clearly related to the presence of thrombi in the device as detected at the TTE. A nonhypercoagulable state was found. The patient had experienced severe gastrointestinal tract infectious disease. This complication is very important because it tells us that endothelization of the device is not completed after six months, and some predisposing factors, such as an infection, can favor thrombus formation on the device beyond the period of antiaggregation treatment.
Conclusions.
Our series of patients with ASD treated with transcatheter occlusion shows that the procedure is safe, with a low rate of early and late complications. However, some aspects need further and longer investigations, especially the possibility of late thrombus formation on the device, systemic embolization and, the most worrisome event, sudden death.
 |
Acknowledgments
|
|---|
The authors thank Miss Amanda Green, medical secretary, for her help reviewing the English style of this article.
 |
References
|
|---|
1. Emmanoulides GC, Allen HD, Reimenschneider FA, et al. Heart disease in infants, children and adolescents including fetus and young adults. Clark FB, Gutgesell HP. Baltimore, MD: Williams and Wilkins; 1995. p. 6069
2. Dickinson DE, Arnold R, Wilkinson JL. Congenital heart disease among 160,480 liveborn children in Liverpool 1960 to 1969. Implications for surgical treatment. Br Heart J. 1981;46:5562[Abstract/Free Full Text]
3. Vitiello R, McCrindle BW, Nykanem D, Freedom MR, Benson LN. Complications associated with pediatric cardiac catheterization. J Am Coll Cardiol. 1998;32:14331440[Abstract/Free Full Text]
4. Walsh KP, Tofeig M, Kitchiner DJ, Peart I, Arnold R. Comparison of the Sideris and Amplatzer septal occlusion devices. Am J Cardiol. 1999;83:933936[CrossRef][Medline]
5. Sievert H, Babic UU, Hausdorf G, et al. Transcatheter closure of atrial septal defect and patent foramen ovale with the ASDOS device (a multi-institutional European trial). Am J Cardiol. 1998;82:14051413[CrossRef][Medline]
6. Carminati M, Chessa M, Butera G, et al. Transcatheter closure of atrial septal defects with the STARFlex device: early results and follow-up. J Intervent Cardiol. 2001;14:319324[Medline]
7. Berger VM, Dahert I, Ewert P, Lange PE. Treatment of atrial septal defects in symptomatic children aged less than 2 years of age using the Amplatzer septal occluder. Cardiol Young. 2000;10:534537[Medline]
8. Chan KC, Godman MJ, Walsh K, Wilson N, Redington A, Gibbs JL. Transcatheter closure of atrial septal defect and interatrial communications with a new self expanding Nitinol double disc device (Amplatzer septal occluder): multicentre UK experience. Heart. 1999;82:300306[Abstract/Free Full Text]
9. Waight DJ, Koenig PR, Cao Q, Hijazi ZM. Transcatheter closure of secundum atrial septal defects using the Amplatzer septal occluder: clinical experience and technical considerations. Curr Intervent Cardiol Rep. 2000;2:7077[Medline]
10. Hijazi ZM, Cao Q, Patel H, Rhodes J. Transcatheter closure of atrial communications using the Amplatzer septal occluder. J Intervent Cardiol. 1999;12:5157
This article has been cited by other articles:

|
 |

|
 |
 
F. Chen, X. Zhao, X. Zheng, S. Chen, R. Xu, and Y. Qin
Incomplete Endothelialization and Late Dislocation After Implantation of an Amplatzer Septal Occluder Device
Circulation,
August 9, 2011;
124(6):
e188 - e189.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
X. Wei, W. Yi, X. Xu, J. Zhang, J. Li, S. Yu, and D. Yi
Transthoracic occlusion for secundum atrial septal defects unsuitable for transcatheter occlusion approach
J. Thorac. Cardiovasc. Surg.,
July 1, 2011;
142(1):
113 - 119.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
U. Gupta, S. J. Al-anani, and A. C. Polimenakos
Timing for successful surgical management of heart block after placement of an Amplatzer occlusion device for secundum atrial septal defect repair
J. Thorac. Cardiovasc. Surg.,
May 1, 2011;
141(5):
1319 - 1321.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Humenberger, R. Rosenhek, H. Gabriel, F. Rader, M. Heger, U. Klaar, T. Binder, P. Probst, G. Heinze, G. Maurer, et al.
Benefit of atrial septal defect closure in adults: impact of age
Eur. Heart J.,
March 1, 2011;
32(5):
553 - 560.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Jerath, A. Jaura, J. Karski, C. Feindel, J. Butany, and A. Vegas
Left-sided migration of Sideris button atrial septal occluder device
Eur Heart J Cardiovasc Imaging,
June 1, 2010;
11(5):
E21 - E21.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. B. Clark, D. Chowdhury, L. B. Pauliks, and H. S. Weber
Resolution of Heart Block After Surgical Removal of an Amplatzer Device
Ann. Thorac. Surg.,
May 1, 2010;
89(5):
1631 - 1633.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Balzer, S. van Hall, T. Rassaf, Y.-C. Boring, A. Franke, R. M. Lang, M. Kelm, and H. P. Kuhl
Feasibility, safety, and efficacy of real-time three-dimensional transoesophageal echocardiography for guiding device closure of interatrial communications: initial clinical experience and impact on radiation exposure
Eur Heart J Cardiovasc Imaging,
January 1, 2010;
11(1):
1 - 8.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Schoof, K. Norozi, T. Breymann, A. Wessel, and H. Bertram
Cough: A Potentially Life-Threatening Condition After Interventional Closure of Atrial Septal Defect
Circ Cardiovasc Imaging,
September 1, 2009;
2(5):
e30 - e31.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. J. Dibardino, D. B. McElhinney, A. K. Kaza, and J. E. Mayer Jr.
Analysis of the US Food and Drug Administration Manufacturer and User Facility Device Experience database for adverse events involving Amplatzer septal occluder devices and comparison with the Society of Thoracic Surgery congenital cardiac surgery database.
J. Thorac. Cardiovasc. Surg.,
June 1, 2009;
137(6):
1334 - 1341.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Balzer, M. Kelm, and H. P. Kuhl
Real-time three-dimensional transoesophageal echocardiography for guidance of non-coronary interventions in the catheter laboratory
Eur Heart J Cardiovasc Imaging,
May 1, 2009;
10(3):
341 - 349.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Sharp and I. Malik
Secundum atrial septal defects: time to close them all?
Heart,
September 1, 2008;
94(9):
1120 - 1122.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. M. Gaca, J. J. Jaggers, L. T. Dudley, and G. S. Bisset III
Repair of Congenital Heart Disease: A Primer--Part 2
Radiology,
July 1, 2008;
248(1):
44 - 60.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. Hongxin, S. Lijun, Z.-j. Wang, J. Zi, W.-l. Zhang, H.-z. Zhang, G. Wenbin, and C.-w. Zou
Intraoperative device closure of large secundum atrial septal defects; a safe alternative to transcatheter closure
Eur J Cardiothorac Surg,
June 1, 2008;
33(6):
1055 - 1060.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
W. Scholtz, S. Jategaonkar, L. Faber, and D. Horstkotte
Unusual Complication With Transcatheter Closure of an Atrial Septal Defect Prevented by Adequate Imaging
Circulation,
March 11, 2008;
117(10):
e181 - e183.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Z. Luqman, J. A. Ansari, H. Ullah, and M. M. Amanullah
Retrograde aortic perfusion dislodges a dislodged Amplatzer device
J. Thorac. Cardiovasc. Surg.,
March 1, 2008;
135(3):
698 - 699.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. De Decker, R. Hamerlijnck, D. Goosens, and P. Pletinckx
Surgical Removal of an Atrial Septal Defect Closure Device Embolized Into the Pulmonary Trunk by Port Access Technique
Ann. Thorac. Surg.,
February 1, 2008;
85(2):
672 - 674.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. R. Raghuram, R. Krishnan, S. Kumar, and K. Balamurugan
Complications in atrial septal defect device closure
Interact CardioVasc Thorac Surg,
February 1, 2008;
7(1):
167 - 169.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Y. Sheikh, S. Schrepfer, W. Stein, J. West, J. Lombard, T. Burdon, B. Pinsker, and M. P. Pelletier
Right Atrial Mass After Primary Repair of an Atrial Septal Defect: Thrombus Masquerading as a Myxoma
Ann. Thorac. Surg.,
November 1, 2007;
84(5):
1742 - 1744.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. Hongxin, G. Wenbin, S. Lijun, W. Zhengjun, L. Hao, Z. Chengwei, D. Liang, and Y. Guidao
Intraoperative device closure of secundum atrial septal defect with a right anterior minithoracotomy in 100 patients.
J. Thorac. Cardiovasc. Surg.,
October 1, 2007;
134(4):
946 - 951.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. A. Diab, Q.-L. Cao, E. A. Bacha, and Z. M. Hijazi
Device closure of atrial septal defects with the Amplatzer septal occluder: Safety and outcome in infants.
J. Thorac. Cardiovasc. Surg.,
October 1, 2007;
134(4):
960 - 966.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. Butera, M. Carminati, M. Chessa, L. Piazza, A. Micheletti, D. G. Negura, R. Abella, A. Giamberti, and A. Frigiola
Transcatheter Closure of Perimembranous Ventricular Septal Defects: Early and Long-Term Results
J. Am. Coll. Cardiol.,
September 18, 2007;
50(12):
1189 - 1195.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Misra, A. Sadiq, N. Namboodiri, and J. Karunakaran
The 'aortic rim' recount: embolization of interatrial septal occluder into the main pulmonary artery bifurcation after atrial septal defect closure
Interact CardioVasc Thorac Surg,
June 1, 2007;
6(3):
384 - 386.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
I. E. Konstantinov, P. Saxena, L. Friederich, and M. A.J. Newman
Emergency surgery after failed device closure of the atrial septal defect
J. Thorac. Cardiovasc. Surg.,
May 1, 2007;
133(5):
1370 - 1371.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Sigler and C. Jux
Biocompatibility of septal defect closure devices
Heart,
April 1, 2007;
93(4):
444 - 449.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. A. Alsaileek, A. Omran, M. Godman, and H. K. Najm
Echocardiographic visualization of laceration of atrial septum during balloon sizing of atrial septal defect
Eur Heart J Cardiovasc Imaging,
March 1, 2007;
8(2):
155 - 157.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. Butera, M. Carminati, M. Chessa, L. Piazza, R. Abella, D. G. Negura, A. Giamberti, B. Claudio, A. Micheletti, Y. Tammam, et al.
Percutaneous closure of ventricular septal defects in children aged <12: early and mid-term results
Eur. Heart J.,
December 1, 2006;
27(23):
2889 - 2895.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. A. Contrafouris, A. C. Chatzis, N. M. Giannopoulos, M. Milonakis, T. Kousi, G. Kirvassilis, and G. E. Sarris
Emergency surgical intervention for runaway atrial septal defect closure devices: A word of caution.
J. Thorac. Cardiovasc. Surg.,
November 1, 2006;
132(5):
1234 - 1235.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. S. Maimon, S. Ratnapalan, A. Do, J. A. Kirsh, G. J. Wilson, and L. N. Benson
Cardiac Perforation 6 Weeks After Percutaneous Atrial Septal Defect Repair Using an Amplatzer Septal Occluder
Pediatrics,
November 1, 2006;
118(5):
e1572 - e1575.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Jux, H. Bertram, P. Wohlsein, M. Bruegmann, and T. Paul
Interventional Atrial Septal Defect Closure Using a Totally Bioresorbable Occluder Matrix: Development and Preclinical Evaluation of the BioSTAR Device
J. Am. Coll. Cardiol.,
July 4, 2006;
48(1):
161 - 169.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. Teoh, E. Wilton, S. Brecker, and M. Jahangiri
Simultaneous removal of an Amplatzer device from an atrial septal defect and the descending aorta
J. Thorac. Cardiovasc. Surg.,
April 1, 2006;
131(4):
909 - 910.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M S Spence and S A Qureshi
Complications of transcatheter closure of atrial septal defects
Heart,
December 1, 2005;
91(12):
1512 - 1514.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Ruge, S.M. Wildhirt, P. Libera, M. Vogt, K. Holper, and R. Lange
Left Atrial Thrombus on Atrial Septal Defect Closure Device as a Source of Cerebral Emboli 3 Years After Implantation
Circulation,
September 6, 2005;
112(10):
e130 - e131.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
V. Costache, O. Chavanon, F. Thony, and D. Blin
Aortic arch embolization of an Amplatzer(R) occluder after an atrial septal defect closure: hybrid operative approach without circulatory arrest
Eur J Cardiothorac Surg,
August 1, 2005;
28(2):
340 - 342.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. Pastuszko, J. J. Rome, M. M. Gleason, R. D. Bellah, L. M. Montenegro, and J. W. Gaynor
Hemoptysis after CardioSEAL device embolization of a venous collateral after the Fontan operation
J. Thorac. Cardiovasc. Surg.,
June 1, 2005;
129(6):
1447 - 1448.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F P Casselman, H Dom, B De Bruyne, Y Vermeulen, and H Vanermen
Thoracoscopic ASD closure is a reliable supplement for percutaneous treatment
Heart,
June 1, 2005;
91(6):
791 - 794.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Vojacek, M. Mates, J. Popelova, and P. Pavel
Perforation of the right atrium and the ascending aorta following percutaneous transcatheter atrial septal defect closure
Interact CardioVasc Thorac Surg,
June 1, 2005;
4(3):
157 - 159.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Bartel, T. Konorza, U. Neudorf, T. Ebralize, H. Eggebrecht, A. Gutersohn, and R. Erbel
Intracardiac echocardiography: an ideal guiding tool for device closure of interatrial communications
Eur Heart J Cardiovasc Imaging,
March 1, 2005;
6(2):
92 - 96.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Masura, P. Gavora, and T. Podnar
Long-term outcome of transcatheter secundum-type atrial septal defect closure using Amplatzer septal occluders
J. Am. Coll. Cardiol.,
February 15, 2005;
45(4):
505 - 507.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J.-F. Piechaud
Closing down: transcatheter closure of intracardiac defects and vessel embolisations
Heart,
December 1, 2004;
90(12):
1505 - 1510.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E. Y. Lee, M. J. Siegel, C. M. Chu, F. R. Gutierrez, and H. W. Kort
Amplatzer Atrial Septal Defect Occluder for Pediatric Patients: Radiographic Appearance
Radiology,
November 1, 2004;
233(2):
471 - 476.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Massimo, B. Gianfranco, and C. Mario
Risk of thrombus formation on devices used to close transcatheter atrial septal defect and patent foramen ovale
J. Am. Coll. Cardiol.,
October 19, 2004;
44(8):
1712 - 1712.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Jux and H. Bertram
Thrombus formation on intracardiac devices: A complex issue
J. Am. Coll. Cardiol.,
October 19, 2004;
44(8):
1712 - 1713.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
U. Krumsdorf, S. Ostermayer, K. Billinger, T. Trepels, E. Zadan, K. Horvath, and H. Sievert
Reply
J. Am. Coll. Cardiol.,
October 19, 2004;
44(8):
1714 - 1716.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C K Silversides, S C Siu, P R McLaughlin, K L Haberer, G D Webb, L Benson, and L Harris
Symptomatic atrial arrhythmias and transcatheter closure of atrial septal defects in adult patients
Heart,
October 1, 2004;
90(10):
1194 - 1198.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. D. Webb and E. M. Horlick
Lessons from cardiopulmonary testing after device closure of secundum atrial septal defects: A tale of two ventricles
J. Am. Coll. Cardiol.,
May 19, 2004;
43(10):
1892 - 1893.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. Suda, M.-J. Raboisson, E. Piette, N. S. Dahdah, and J. Miro
Reversible atrioventricular block associated with closure of atrial septal defects using the amplatzer device
J. Am. Coll. Cardiol.,
May 5, 2004;
43(9):
1677 - 1682.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F E Willcoxson, J D R Thomson, and J L Gibbs
Successful treatment of left atrial disk thrombus on an Amplatzer atrial septal defect occluder with abciximab and heparin
Heart,
May 1, 2004;
90(5):
e30 - e30.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
O. Preventza, S. Sampath-Kumar, J. Wasnick, and J. P. Gold
Late cardiac perforation following transcatheter atrial septal defect closure
Ann. Thorac. Surg.,
April 1, 2004;
77(4):
1435 - 1437.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D J R Hildick-Smith, M O'Sullivan, C R Wisbey, J H Mackay, E M Lee, and L M Shapiro
Amplatzer device closure of atrial septal defects in mature adults: analysis of 76 cases
Heart,
March 1, 2004;
90(3):
334 - 335.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
U. Krumsdorf, S. Ostermayer, K. Billinger, T. Trepels, E. Zadan, K. Horvath, and H. Sievert
Incidence and clinical course of thrombus formation on atrial septal defect and patient foramen ovale closure devices in 1,000 consecutive patients
J. Am. Coll. Cardiol.,
January 21, 2004;
43(2):
302 - 309.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. W. Moore and D. S. Levi
Transcatheter closure of atrial shunts: Focus on a lingering issue
J. Am. Coll. Cardiol.,
January 21, 2004;
43(2):
310 - 312.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. Butera, G. De Rosa, M. Chessa, L. Rosti, D. G. Negura, P. Luciane, A. Giamberti, E. Bossone, and M. Carminati
Transcatheter closure of atrial septal defect in young children: Results and follow-up
J. Am. Coll. Cardiol.,
July 16, 2003;
42(2):
241 - 245.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Heggie, N. Poirer, W. G. Williams, and J. Karski
Anesthetic Considerations for Adult Cardiac Surgery Patients with Congenital Heart Disease
Seminars in Cardiothoracic and Vascular Anesthesia,
June 1, 2003;
7(2):
141 - 152.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Bartel, T. Konorza, J. Arjumand, T. Ebradlidze, H. Eggebrecht, G. Caspari, U. Neudorf, and R. Erbel
Intracardiac Echocardiography Is Superior to Conventional Monitoring for Guiding Device Closure of Interatrial Communications
Circulation,
February 18, 2003;
107(6):
795 - 797.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Transcatheter Atrial Septal Defect Closure
Journal Watch Cardiology,
May 24, 2002;
2002(524):
3 - 3.
[Full Text]
|
 |
|
|