BACKGROUND: A shortage of transplant donors necessitates the testing of alternative treatments. The Jarvik-2000 is an axial flow pump with a percutaneous retro-auricular power connector, designed for permanent use.
METHODS: Patients with severe heart failure (HF), unsuitable for heart transplantation or conventional LVAD support, were offered implantation. The surgical approach included a left lateral thoracotomy. The device was implanted into the left ventricular apex on femoro-femoral bypass. It is set to allow pulsatile flow with an aortic valve opening. Anticoagulation is adjusted the same as for patients with a heart valve.
RESULTS: Between May 2001 and August 2001, we implanted the Jarvik-2000 in two patients with dilated cardiomyopathy and in one with cardiac amyloidosis, all with severe HF (cardiac index 1.8 ± 0.3 l/m2 per min). One patient required preoperative inotropic support. All patients did well, with no repeat operations or infections. Patients received 4.3 ± 3.2 packed red blood cells and were intubated at 14 ± 3 h, and the intensive care unit stay was 7.0 ± 0.5 days. The cardiac index increased from 3.7 ± 1.5 l/min per m2 at 8,000 rpm to 5.9 ± 2.9 l/min per m2 at 12,000 rpm. All patients currently have mild hemolysis not requiring transfusion. The following postoperative events were recorded: a transient ischemic attack with complete recovery, a short re-intubation due to ventricular arrhythmia, loss of consciousness with a battery change while standing, knee-joint effusion after ergometry training, a minor wound problem and a short hospital re-admission due to dehydration. Patients were discharged home after 49 ± 7 days; one has returned to work. All quality-of-life scores have improved.
CONCLUSIONS: The permanent Jarvik-2000 appears safe. It can be used for dilative or restrictive disease. The Jarvik-2000 might prove a valid option for the long-term treatment of patients with severe HF.
| ||||||||||||||||||
|
1.5 kg, allowing patients to lead an independent life-style. We report our early experience with this device. | Methods |
|---|
|
|
|---|
Patient selection.
Patients with end-stage HF, who had no conventional treatment option, were considered for Jarvik-2000 implantation. Table 1 lists the inclusion and exclusion criteria for implantation of the Jarvik-2000 LVAD. Heart failure criteria for transplantation or mechanical LV support had to be present. Patients who did not qualify for heart transplantation, because of advanced age or elevated pulmonary vascular resistance, or patients unsuitable for a conventional LVAD because of a small body frame, with a body surface area
1.5 m2, were eligible for Jarvik-2000 implantation. We did not offer the Jarvik-2000 to patients with fixed pulmonary vascular resistance >7 wood units (WU), because of the high risk of postoperative right ventricular (RV) failure. Only patients with a Columbia University risk score index
5 were considered for implantation (10). This score, used for patient risk stratification, is based on the following clinical variables: central venous pressure, urine output, respiratory status, hepatic synthetic function, previous cardiac surgery and leukocyte count. Most patients with a score >5 will die after LVAD implantation (10).
|
4 mm). A full evaluation by a neurologist was performed. Preoperative quality of life was assessed by use of a standardized questionnaire (12,13). Patient and family information includes detailed discussions and a brochure in layman language explaining the potential risks, benefits and alternatives of the planned procedure. Sufficient time is allowed for questions and the decision-making process before full informed consent is obtained.
Intraoperative management
Intraoperative monitoring included TEE and a Swan-Ganz catheter. Aprotinin and nitric oxide (5 to 80 ppm) were used routinely. The retro-auricular pedestal site was prepared. A left postero-lateral thoracotomy incision was performed. The internal power cable of the device was passed through the posterior apex of the chest and tunneled to the pedestal, which was attached to the skull (Fig. 1B, inset). Then the outflow graft was sewn to the descending thoracic aorta in partial occlusion. Conventional femoro-femoral bypass was instituted. A sewing ring was attached to the left ventricular apex. The heart was fibrillated. The device was implanted into the LV apex and secured to the sewing ring. Complete de-airing was confirmed with TEE before unclamping the outflow graft. Cardiopulmonary bypass was discontinued, and meticulous hemostasis was secured before thoracotomy closure.
Postoperative management
The pump was set to the minimal speed, keeping the LV in a partially loaded state, with aortic valve opening. Infection prophylaxis included 48 h of perioperative antibiotics, antibiotic-coated catheters, catheter removal within two days, early extubation followed by incentive spirometry and aggressive mobilization. In the absence of bleeding, intravenous heparin (500 IU/h) was started 12 h after the operation. On the second postoperative day, heparin was adjusted to full anticoagulation, with an activated partial thromboplastin time of 50 to 70 s. Coumadin (warfarin), as well as the pharmacologic HF treatment, including angiotensin-converting enzyme inhibitors and beta-blockers, was started as soon as the patient could tolerate oral intake. Diuretics could be gradually reduced. Patients visited a formal two-day teaching course for Coumadin management and received a device to check their international normalized ratio (INR) at home. Rehabilitation was started in the hospital and consisted of bicycle ergometry, walking and strengthening exercises with physical therapy. Systematic teaching of the functions and handling of the device components, as well as emergency algorithms, was provided to the patients and their relatives. After hospital discharge, a multidisciplinary team provided outpatient follow-up.
| Results |
|---|
|
|
|---|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Patient selection
Choosing the right patients for LVAD support poses a difficult problem. Oz et al. (10), at Columbia University, have established a score for risk stratification of patients, which we have used for bedside decision-making. Implantation of the LVAD in patients with multi-system organ failure will result in a prohibitively high mortality rate, regardless of the device chosen (10). In the early phase of device testing, we preferred to offer this therapy to patients with a risk score index
5 in order to avoid offering this expensive therapy to patients who are unlikely to benefit and to recognize the potential morbidity inflicted by the device. It is important to emphasize that the patients in our series were in better preoperative condition than many patients requiring mechanical support. Yet, all of the patients were in dire need of LV support and had contraindications to heart transplantation or conventional ventricular support, as illustrated in Table 2.
Bleeding and RV failure
Conventional LVAD implantation is associated with a high rate of bleeding (4,8), often requiring a repeat operation, thus increasing the risk of postoperative RV failure (17). In our patients, we noted no bleeding complications, no RV failure and a low transfusion requirement after Jarvik-2000 implantation. Several factors might have been responsible for this: the surgeons awareness of the high morbidity associated with bleeding, the routine use of aprotinin, a smaller surgical procedure, compared with conventional LVAD implantation, using a lateral thoracotomy, short cardiopulmonary bypass times and the absence of severe preoperative coagulopathy. Aprotinin reduces the transfusion requirements and mortality after LVAD implantation and also potentially reduces the incidence of RV failure (17). In addition, unloading of the right ventricle with routine use of nitric oxide (5 to 80 ppm) decreases the pulmonary artery pressure and improves LVAD flow, although no beneficial effect on mortality has been shown (18,19).
Infection
Infections occur in 30% to 50% of patients in a series of conventional LVAD implantations (5,7,20). Drive-line and peri-implant infections, as well as LVAD endocarditis, cannot generally be cleared without removing the device at the time of transplantation. The majority of infections associated with LVAD do not appear to have a negative effect on the outcomes of heart transplantation (5). We have encountered no infectious complications after Jarvik-2000 implantation. Factors that may potentially reduce the risk of infection include a strict infection control policy and a small amount of surgically implanted foreign material. The different power supply by means of an immobile skin exit site in the left retro-auricular area probably has been the most important factor (Fig. 1B). Clinical experience with early cochlear implants, using similar percutaneous skull-mounted pedestals for power supply, demonstrated a low rate of complications, including infection (21). Other advantages of this mode of power supply include only minimal interference with daily activities and the waterproof power plug, which allows the patients to shower. This skull-mounted power connector will soon be released for clinical testing in the United States.
Thromboembolism
Thromboembolic events are often devastating complications after successful LVAD implantation. To potentially reduce thromboembolic events, the Jarvik-2000 is implanted directly into the LV with no inflow cannula, and the outflow graft is anastomosed into the descending thoracic aorta. The device is set at a minimal pump speed to allow aortic valve opening, thus preventing potential thrombus formation in areas of stasis in the LV outflow tract and aortic root. All of these factors can potentially reduce the incidence of thromboembolism. One of our patients had a transient ischemic attack, likely due to a small embolus. At the time, his INR was subtherapeutic at 1.8. Based on this single event, we did not change the anticoagulation regimen, adjusting Coumadin to an INR of 2.0 to 2.5. To date, we intend to avoid giving platelet inhibitors together with Coumadin because of the increased risk of bleeding complications in other patient cohorts (22).
Chronic nonpulsatile flow
The effects of chronic nonpulsatile flow are unknown; therefore, we avoided exposing our patients to it. The Jarvik-2000 is set to allow aortic valve opening with a palpable pulse. The concept of this device is new, as it is truly assisting and not replacing LV function, keeping the ventricle in a slightly loaded state. It should therefore not be used in patients with virtually absent contractile function. Conventional LVADs are more suitable to support such patients. To be more accurate, they should probably be called "LV replacement systems," as they completely replace native heart function, with the aortic valve permanently closed. Interestingly, the Jarvik-2000 alone appears to maintain, at least temporarily, sufficient cardiac output, as our patient with sustained ventricular tachycardia remained fully awake.
Hemolysis
Hemolysis is one of the main concerns of axial flow pumps, given the devices tremendous rotational speeds (8,000 to 12,000 rpm). As illustrated in Figure 3, we found evidence of mild blood trauma, likely due to mechanical forces affecting the cellular blood components. Hemolysis does not appear to pose a clinical problem thus far, as no patient required a transfusion or had complications due to increased red-cell turnover. We will perform ultrasonography of the gallbladder every six months because of the increased risk of pigment gallstone formation and related complications.
Psychology
Many patients require psychological therapy after LVAD implantation (23). Denial is a common coping mechanism, but patients with an LVAD cannot ignore or deny their condition. The patient has to change the batteries several times a day and is constantly reminded of the device. Professional psychological help, focusing on coping strategies and treating depression, was perceived as beneficial by two of our patients. All patients had an improved quality-of-life score (Table 3).
Prognosis
We can only speculate on the long-term prognoses of these patients. It will depend on the devices durability, the incidence of late complications and the course of the underlying myocardial disease. The wear and tear of the device is very low according to the manufacturer (Jarvik Heart, Inc., New York, New York). We are attempting to keep late complications at a minimum, with close outpatient multidisciplinary follow-up. Long-term unloading could potentially allow partial myocardial recovery or stabilize the disease. Changes in myocardial gene expression have been described with long-term unloading (24). Only a few patients with a conventional LVAD implanted because of chronic HF had sufficient myocardial recovery to allow device explantation. Many of them later developed recurrent HF (25). Native myocardial function is crucial, as the Jarvik-2000 only assists the LV and depends on the hearts contractility. Therefore, continuous HF treatment with administration of angiotensin-converting enzyme inhibitors and beta-blockers appears to be essential in the long-term management of these patients.
Conclusions
Our early experience with the permanent Jarvik-2000 LVAD has been encouraging. It can be used in patients with dilative or restrictive disease. There were no device failures. All patients were discharged home. We noted a low incidence of complications. The Jarvik-2000 might prove to be a valid tool in the physicians armamentarium for the long-term treatment of severe HF.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
U. Geisen, C. Heilmann, F. Beyersdorf, C. Benk, M. Berchtold-Herz, C. Schlensak, U. Budde, and B. Zieger Non-surgical bleeding in patients with ventricular assist devices could be explained by acquired von Willebrand disease Eur. J. Cardiothorac. Surg., April 1, 2008; 33(4): 679 - 684. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Markl, C. Benk, D. Klausmann, A. F. Stalder, A. Frydrychowicz, J. Hennig, and F. Beyersdorf Three-dimensional magnetic resonance flow analysis in a ventricular assist device. J. Thorac. Cardiovasc. Surg., December 1, 2007; 134(6): 1471 - 1476. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Zimpfer, P. Zrunek, S. Sandner, H. Schima, M. Grimm, A. Zuckermann, E. Wolner, and G. Wieselthaler Post-transplant survival after lowering fixed pulmonary hypertension using left ventricular assist devices Eur. J. Cardiothorac. Surg., April 1, 2007; 31(4): 698 - 702. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Zimpfer, P. Zrunek, W. Roethy, M. Czerny, H. Schima, L. Huber, M. Grimm, A. Rajek, E. Wolner, and G. Wieselthaler Left ventricular assist devices decrease fixed pulmonary hypertension in cardiac transplant candidates J. Thorac. Cardiovasc. Surg., March 1, 2007; 133(3): 689 - 695. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Beyersdorf Transapical transcatheter aortic valve implantation Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 7 - 8. [Full Text] [PDF] |
||||
![]() |
A. J. Clegg, D. A. Scott, E. Loveman, J. L. Colquitt, P. Royle, and J. Bryant Clinical and cost-effectiveness of left ventricular assist devices as a bridge to heart transplantation for people with end-stage heart failure: a systematic review and economic evaluation Eur. Heart J., December 2, 2006; 27(24): 2929 - 2938. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. P. Siegenthaler, J. Martin, R. Gutwald, R. Bahr, S. Westaby, R. Schmelzeisen, and F. Beyersdorf Anterior Approach to Implant the Jarvik 2000 With Retroauricular Power Supply Ann. Thorac. Surg., August 1, 2005; 80(2): 745 - 747. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. P. Siegenthaler, S. Westaby, O.H. Frazier, J. Martin, A. Banning, D. Robson, J. Pepper, P. Poole-Wilson, and F. Beyersdorf Advanced heart failure: feasibility study of long-term continuous axial flow pump support Eur. Heart J., May 2, 2005; 26(10): 1031 - 1038. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. H. Frazier, T. J. Myers, S. Westaby, and I. D. Gregoric Clinical experience with an implantable, intracardiac, continuous flow circulatory support device: physiologic implications and their relationship to patient selection Ann. Thorac. Surg., January 1, 2004; 77(1): 133 - 142. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. A. Nussmeier, C. B. Probert, D. Hirsch, J. R. Cooper Jr., I. D. Gregoric, T. J. Myers, and O. H. Frazier Anesthetic Management for Implantation of the Jarvik 2000TM Left Ventricular Assist System Anesth. Analg., October 1, 2003; 97(4): 964 - 971. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||