EDITORIAL COMMENT
Cardiac resynchronization: a cornerstone in the foundation of device therapy for heart failure*
Leslie A. Saxon, MD, FACC*,a and
Teresa De Marco, MD, FACCa
a University of California, San Francisco, California, USA
* Reprint requests and correspondence: Dr. Leslie A. Saxon, University of California, San Francisco, 500 Parnassus Avenue, Box 1354, San Francisco, California 94143 USA
Accepted goals of heart failure treatment include: 1) improvement in symptom status, 2) prevention of disease progression, and 3) reduction in morbidity and mortality (1). Complex pharmacologic therapies achieve these goals for many, but not all, patients with heart failure. For patients who remain symptomatic despite receiving "optimal" medical therapy, novel adjunctive therapies are needed. Technological advancements are permitting the exploration of the so-called device therapies for heart failure. One such device incorporates biventricular stimulation to resynchronize cardiac contraction. The study by Stellbrink et al. (2) in this issue of the Journal is consistent with other emerging data finding that chronic resynchronization therapy may achieve several treatment goals in heart failure including symptom reduction and slowing of heart failure progression (25).
Cardiac resynchronization therapy is only applicable to a subset of heart failure patients, namely those with myocardial conduction system delay. Myocardial conduction system delay, manifest as bundle branch block, impacts 20% to 30% of New York Heart Association (NYHA) functional class IIIIV heart failure patients and consists predominately of left bundle branch block (69). The presence of bundle branch block introduces contractile inefficiency and dysynchrony, causing further diminishment of ventricular function (9,10). When left ventricular (LV) conduction delay is superimposed upon ventricular dysfunction, it appears to be a marker of disease severity, predicting an increased risk of both heart failure progression and susceptibility to ventricular arrhythmias (1115). Cardiac resynchronization attempts to correct the deleterious effects of dysynchrony. The question remains whether this approach satisfies some or all heart failure treatment goals.
The echocardiographic data provided by Stellbrink et al. (2) are derived from a subset of patients enrolled in one of the first clinical studies of resynchronization therapy in Europe, the Pacing Therapies in Congestive Heart Failure (PATH-CHF) trial (2). Because of rapid advances in device and lead technology, this trial differs from studies subsequently performed in Europe and the U.S., particularly in relation to duration of follow-up and implant tools and techniques. The PATH-CHF trial enrolled 42 patients with advanced NYHA functional class III or IV heart failure due predominately to nonischemic etiologies. Primary end points of the study were measures of functional capacity. The PATH-CHF trial had a unique design, common to two clinical trials of resynchronization therapy performed in Europe (16). At the time of device implantation, subsequent programmed parameters were determined on the basis of acute hemodynamic data measuring aortic pulse pressure and +dP/dt. Hemodynamics obtained in sinus rhythm were compared with right, left and simultaneous biventricular stimulation performed at variable atrioventricular intervals. Specifically, the hemodynamic data were acquired, analyzed off-line and used to determine whether resynchronization was chronically provided by right ventricular, LV or simultaneous biventricular stimulation. Patients enrolled in PATH-CHF had LV stimulation provided from an epicardial LV lead placed via a limited thoracotomy. Patients also received two separate pacemakers and two endocardial right atrial leads to allow for independent programming of right and left ventricle. In contrast, all U.S. trials completed to date provided resynchronization by simultaneous biventricular stimulation (5). Patients were then paced for six months, at which time echocardiography was performed and compared to baseline.
The primary end points of the PATH-CHF trial were positive, but they are not reported in the study by Stellbrink et al. (2), and have only been published in abstract form (1719). The investigators do report that NYHA functional class significantly improved over six months of resynchronization therapy. The echocardiographic data demonstrate improvements in LV volumes and dimensions, well-validated measures of LV reverse remodeling. Neither LV fractional shortening, mass, sphericity index nor heart rate changed with therapy. There are no data evaluating the effect of therapy on measures of diastolic function or right heart function. Of interest, the beneficial effects of resynchronization on LV volume and dimension did not correlate with either acute hemodynamic or NYHA functional class benefit. Also, there were patients who had progressive ventricular enlargement on therapy but had improvement in acute hemodynamic parameters and/or improvement in functional class. This suggests that the well-described acute hemodynamic benefit of LV or biventricular stimulation may be uncoupled from the potential for a reverse ventricular remodeling effect (20,21). It further suggests that functional benefit can be seen in patients who do not attain reverse remodeling. Accordingly, hemodynamic parameters may not be useful in determining the full potential of the therapy to improve both symptom status and ventricular remodeling.
Neither the study by Stellbrink et al. (2) nor a similar study by Touiza et al. (22) in this issue of the Journal was designed to answer whether chronic pacing of one ventricle was equivalent to chronic simultaneous stimulation of both ventricles. In the Stellbrink et al. (2) report, the site(s) of stimulation was determined by the location that produced the most favorable acute hemodynamic profile. In the Touiza et al. (22) study, the stimulation site(s) was determined simply by physician preference, which in part was influenced by individual patient anatomic characteristics that dictated the practicalities of pacer lead placement. A prospective, randomized trial will be needed to determine whether single or biventricular stimulation is important in achieving the treatment goals of chronic resynchronization therapy.
Among other limitations of the Stellbrink et al. (2) study is that the majority of patients had heart failure due to idiopathic cardiomyopathy. The study results may not apply to patients with an ischemic etiology for heart failure, characterized by more regional myocardial dysfunction. As there were changes in heart failure medications throughout the study interval, this may have influenced LV volumes and dimensions at six months independent of resynchronization. Further, as only 60% of patients were receiving beta-blockers, it is difficult to say whether chronic resynchronization is positively or adversely affected by chronic beta-blockade. The Stellbrink et al. (2) study also attempted to define a subgroup of "responders" to resynchronization (as defined by a 15% decrease in LV end systolic volume compared to baseline) and to determine whether baseline characteristics of such patients were different from "nonresponders." The "responders" group, however, included patients whose LV end systolic volumes were stable following resynchronization therapy and may themselves have different baseline characteristics than true responders. Little can therefore be said of this subgroup analysis.
Finally, although it is clear that the natural history of heart failure is characterized by progressive myocardial remodeling, studies of beta-blocker therapy that include a control group find that some echocardiographic measures of progression or regression of remodeling may not be detected for up to 18 months. The study by Stellbrink et al. (2) may be too short to understand the full impact (if any) of chronic resynchronization therapy on myocardial remodeling (23).
Ever since the PATH-CHF study was initiated, device technology has rapidly evolved. In the U.S., a resynchronization device has recently gained regulatory approval and a resynchronization ICD is on the cusp of approval. Devices are now available, on clinical protocol, that can be programmed to achieve resynchronization with LV or biventricular stimulation, provide early diagnosis and treatment of atrial arrhythmias, and offer real-time hemodynamic data. All leads can now be implanted with a transvenous approach, thus avoiding thoracotomy. Left ventricular stimulation is now achieved reliably and safely using a coronary sinus branch vein (3,22,2425). This is important as it reduces the inherent operative morbidity and mortality of thoracotomy in order to place the LV epicardial lead. One patient died of postthoracotomy complications in the Stellbrink et al. (2) study. The addition of defibrillation capability to the device adds little to device size or operative risk (25). Thus, technologic advancements have rapidly improved the safety and simplicity of LV lead placement.
Resynchronization devices may be simple and safe, but does chronic resynchronization therapy fulfill its promise to act as an adjunct to pharmacologic agents in the management of heart failure patients with myocardial conduction delay and achieve some or all of the goals of heart failure treatment? In this calendar year, the results of the first controlled clinical trials of resynchronization therapies providing six months of follow-up have been published or presented (2,3,22,2425). The data are consistent across these studies, demonstrating the therapys effectiveness at improving heart failure symptoms and stabilizing or reducing LV volumes and dimensions. Clinical trials initiated in 1999 are evaluating the ability of chronic resynchronization therapy (with or without an ICD), compared to standard medical therapies, to improve all-cause mortality and hospitalization (4,26).
Those of us involved in the clinical studies of cardiac resynchronization already appreciate that these devices promote optimal medical management of the heart failure patient. Improvements in systolic blood pressure resulting from resynchronization therapy allow for upward titration of vasodilating drugs (22,25). The atrial rate support feature is also very useful in this era of increasing use of beta-receptor blocking agents. Although extremely attractive theoretically, the mortality benefit of the defibrillating capability is yet to be determined.
Why have technical devices been developed for cardiac diseases? Simply put, because the need was there. The first implantable pacemakers were devised by Chardack, Wilson, and Greatbach in 1960 to treat heart block because medications were unable to correct this problem. The need was there. The first ICD was proposed in 1966 by Mirowski when a close friend experienced sudden death. Medications were not preventing sudden death. The need was there. Many heart failure patients still experience debilitating symptoms and die prematurely despite "optimal" medical therapy. The need for adjunctive therapy is here. The development of resynchronization therapy is addressing this need. The development of resynchronization therapy is a unique cooperation of electrophysiologists and heart failure pharmacotherapists that is only in its infancy. As Henry Ford observed, "Coming together is a beginning; keeping together is progress; working together is success." The need is here and success is our goal. We commend the early work of Stellbrink et al. (2) for providing a rationale to establish resynchronization as a legitimate adjunctive therapy in the management of heart failure.
 |
Footnotes
|
|---|
* Editorials published in the Journal of the American College of Cardiology reflect the views of the authors and do not necessarily represent the views of JACC or the American College of Cardiology. 
 |
References
|
|---|
1. Cleland JGF, Swedberg K, Poole-Wilson PA. Successes and failures of current treatment of heart failure. Lancet. 1998;352(Suppl I):SI19SI28
2. Stellbrink C, Breithardt O-A, Franke A, et al. Impact of cardiac resynchronization therapy using hemodynamically optimized pacing on left ventricular remodeling in patients with congestive heart failure and ventricular conduction disturbances. J Am Coll Cardiol. 2001;38:19571965[Abstract/Free Full Text]
3. Abraham WT. Report from the MIRACLE (Multicenter Insync Randomized Clinical Evaluation) study. 50th Annual Scientific Session, American College of Cardiology, 2001, Late-Breaking Clinical Trials.
4. Vigor-CHF investigatorsSaxon LA, De Marco T, Dibs SR, et al. Chronic biventricular pacing improves indices of systolic function and reduces left ventricular volume. (abstr)Pacing Clin Electrophysiol. 2000;23-24:II635
5. Saxon LA, Kumar UN, De Marco T. Heart failure and cardiac resynchronization therapies: US experience in the year 2000. Ann Noninvas Electrophysiol. 2000;5:188194[CrossRef]
6. Shamim W, Francis DP, Yousufuddin M, et al. Intraventricular conduction delay: a prognostic marker in chronic heart failure. Int J Cardiol. 1999;70:171178[CrossRef][Medline]
7. Cianfocca C, Pelliccia F, Nigri A, et al. Resting and ambulatory ECG predictors of mode of death in dilated cardiomyopathy. J Electrocardiol. 1992;25:295303[CrossRef][Medline]
8. Aaronson KD, Schwartz JS, Chen T-M, Wong K-L, Goin JE, Mancini DM. Development and prospective validation of a clinical index to predict survival in ambulatory patients referred for cardiac transplant evaluation. Circulation. 1997;95:26602667[Abstract/Free Full Text]
9. Xiao HB, Brecker SJD, Gibson DG. Effects of abnormal activation on the time course of the left ventricular pressure pulse in dilated cardiomyopathy. Br Heart J. 1992;68:403407[Abstract/Free Full Text]
10. Kerwin WF, Botvinick EH, OConnell JW, et al. Ventricular contraction abnormalities in dilated cardiomyopathy: effect of biventricular pacing to correct intraventricular dyssynchrony. J Am Coll Cardiol. 2000;35:12211227[Abstract/Free Full Text]
11. Packer M, Carver JR, Rodeheffer RJ, et al. Effort of oral milrinone on mortality in severe chronic heart failure: the PROMISE study research group. N Engl J Med. 1991;325:14681475[Abstract]
12. Xiao HB, Roy C, Fujimoto S, et al. Natural history of abnormal conduction and its relation to prognosis in patients with dilated cardiomyopathy. Int J Cardiol. 1996;53:163170[CrossRef][Medline]
13. Eriksson P, Hansson P-O, Eriksson H, Dellborg M. Bundle-branch block in a general male population: the study of men born 1913. Circulation. 1998;98:24942500[Abstract/Free Full Text]
14. McAnulty HJ, Rahimtoola HS, Murphy E, et al. Natural history of "high-risk" bundle-branch block. N Engl J Med. 1982;307:137143[Abstract]
15. Englund A, Bergfeldt L, Rehnqvist N, ström H, Rosenqvist M. Diagnostic value of programmed ventricular stimulation in patients with bifascicular block: a prospective study of patients with and without syncope. J Am Coll Cardiol. 1995;26:15081515[Abstract]
16. Auricchio A, Stellbrink C, Sack S, et al. The pacing therapy for congestive heart failure (PATH CHF) study: rationale, design and endpoints of a prospective randomized multicenter study. Am J Cardiol. 1999;83:130D135D[CrossRef][Medline]
17. Auricchio A, Stellbrink C, Sack S, et al. Impact of short-term chronic ventricular stimulation on exercise tolerance and well-being in CHF: results of PATH-CHF trial. (abstr)J Card Fail. 2000;6:25[Medline]
18. Sack S, Auricchio A, Kadhiresan V, et al. Long-term improvement with ventricular resynchronization therapy in heart failure patients: does etiology of heart failure matter? (abstr)Pacing Clin Electrophysiol. 2000;23:II555
19. Auricchio A, Stellbrink C, Sack S, et al. Long-term benefit as a result of pacing resynchronization in congestive heart failure: results of the Path-CHF trial (abstr). Circulation. 2000;102:II693
20. Kass DA, Chen CH, Curry C, et al. Improved left ventricular mechanics from VDD pacing in patients with dilated cardiomyopathy and ventricular conduction delay. Circulation. 1999;99:15671573[Abstract/Free Full Text]
21. Auricchio A, Stellbrink C, Block M, et al. Effects of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure. Circulation. 1999;99:29933001[Abstract/Free Full Text]
22. Touiza A, Etienne Y, Gilard M, Fatemi M, Mansourati J, Blanc J-J. Long-term left ventricular pacing: assessment and comparison with biventricular pacing in patients with severe congestive heart failure. J Am Coll Cardiol. 2001;38:19661970[Abstract/Free Full Text]
23. Hall SA, Cigarroa CG, Marcoux L, Risser RC, Grayburn PA, Eichhorn EJ. Time course of improvement in left ventricular function, mass and geometry in patients with congestive heart failure treated with beta-adrenergic blockade. J Am Coll Cardiol. 1995;25:11541161[Abstract]
24. Cazeau S, Leclercq C, Lavergne T, et al. Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay. N Engl J Med. 2001;344:873880[Abstract/Free Full Text]
25. Cannon D. Report of the CONTAK CD clinical trial results. Presentation at Guidant Symposium at the 22nd Annual Scientific Session of the North American Society of Pacing and Electrophysiology (NASPE), Boston, MA, 2001.
26. Bristow MR, Feldman AM, Saxon LA. Heart failure management using implantable devices for ventricular resynchronization: comparison of medical therapy, pacing, and defibrillation in chronic heart failure (COMPANION) trial. J Card Fail. 2000;6:276285[CrossRef][Medline]
This article has been cited by other articles:

|
 |

|
 |
 
R. G. Turcott and T. J. Pavek
Hemodynamic sensing using subcutaneous photoplethysmography
Am J Physiol Heart Circ Physiol,
December 1, 2008;
295(6):
H2560 - H2572.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. Steendijk, S. A. F. Tulner, J. J. Schreuder, J. J. Bax, L. van Erven, E. E. van der Wall, R. A. E. Dion, M. J. Schalij, and J. Baan
Quantification of left ventricular mechanical dyssynchrony by conductance catheter in heart failure patients
Am J Physiol Heart Circ Physiol,
February 1, 2004;
286(2):
H723 - H730.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Jessup and S. Brozena
Heart Failure
N. Engl. J. Med.,
May 15, 2003;
348(20):
2007 - 2018.
[Full Text]
[PDF]
|
 |
|
|