|
|
||||||||||
|
J Am Coll Cardiol, 2001; 38:1622-1627 © 2001 by the American College of Cardiology Foundation |
a University of California, Los Angeles Center for Health Sciences, Los Angeles, California, USA
Manuscript received April 10, 2001; revised manuscript received July 20, 2001, accepted August 15, 2001.
* Reprint requests and correspondence: Dr. Jonathan M. Tobis, 10833 Le Conte Avenue, Room BL-394 CHS, Los Angeles, California 90095-1717 USA
jtobis{at}mednet.ucla.edu
| Abstract |
|---|
|
|
|---|
The purpose of this study was to assess whether the newer stent delivery systems provide a stented lumen cross-sectional area (CSA) that is equal to the delivery balloon nominal dimensions.
BACKGROUND
First generation stents were often not adequately expanded with their delivery system and frequently required higher pressure or a larger balloon after deployment. Newer stents were designed to optimize expansion with noncompliant, high-pressure balloons provided as the delivery systems.
METHODS
Intravascular ultrasound (IVUS) was used to evaluate 38 stents in 32 patients after deployment at 14 to 16 atm with their delivery balloon system. Minimum stent lumen CSA and stent minimum lumen diameter (MLD) were measured by IVUS imaging. The manufacturers expected stent diameter was defined as the balloon diameter measured by the company at the maximum pressure used. The manufacturers expected stent area was calculated based on the manufacturers expected stent diameter.
RESULTS
The MLD (2.5 ± 0.5 mm) and minimum stent CSA (6.0 ± 1.7 mm2) by IVUS were significantly smaller than the manufacturers expected stent diameter (3.5 ± 0.4 mm) and area (9.5 ± 1.9 mm2) (p < 0.0001, respectively). The mean MLD by IVUS was 72 ± 8% of the expected stent diameter, and the mean minimum stent CSA by IVUS was 62 ± 10% of the expected stent area.
CONCLUSIONS
Despite moderately high-pressure inflations, the mean minimum stent CSA actually achieved was, on average, only 62% of the manufacturers expected stent area. To optimize stent deployment, these IVUS observations should be considered during coronary artery stenting.
| ||||||||||
| Methods |
|---|
|
|
|---|
Coronary angiography and stent deployment. Coronary angiography was performed in a routine manner from the femoral artery with a 6F or 8F catheter. Direct stenting was used in eight lesions. Preparation for the stent was performed on 30 lesions with conventional balloon angioplasty alone in 12 cases, one with directional coronary atherectomy, six with rotational coronary atherectomy and eight with cutting balloon angioplasty. In addition, four lesions were treated with an experimental thrombectomy device before stenting (X-SIZER, Endicor Inc., San Clemente, California). Guidant Tristar and AVE/Medtronic S670 stents were deployed at 14 or 16 atm with their delivery balloon system. No patient had additional dilation after stent insertion with another balloon catheter before the IVUS examination.
IVUS. After stent deployment, the IVUS study was performed with a 3.2F monorail system using a 30 MHz transducer-tipped catheter (Ultracross, CVIS/SCIMED Inc, Sunnyvale, California). The IVUS imaging catheter was passed over the guide wire at least 10 mm distal to the stent. The IVUS imaging was performed during motorized pullback (0.5 mm/s) of the catheter. The IVUS images were recorded on S-VHS videotapes for subsequent review and quantitative analysis.
Quantitative coronary angiography (QCA) measurements. Quantitative coronary angiography measurements were performed with an automated computer-based system (CRS-PC+, General Electric Company, Fairfield, Connecticut) by an experienced angiographer. The external diameter of the contrast-filled catheter was used as the calibration standard. Before and after stenting, lesion minimum lumen diameter (MLD) or stent MLD, lesion length and reference diameter were measured. The percent diameter stenosis was calculated automatically as the reference lumen diameter the MLD divided by the reference diameter. Lesions were characterized according to the modified American College of Cardiology-American Heart Association classification (12). The maximum balloon diameter during stent deployment was measured and compared with the manufacturers expected values and with the stent lumen diameter measured by QCA.
IVUS measurements. The IVUS images were digitized with a Macintosh G4 computer (Apple Corp., Cupertino, California). Measurements were made with the use of computerized planimetry with public-domain software (NIH image). The digitized IVUS images were reviewed, and the image slice with the minimum stent lumen CSA was chosen for analysis. The minimum stent lumen CSA and stent MLD were measured.
In addition, in 14 lesions from 11 patients, measurements of stent lumen CSA were made every 1-mm along the length of the stent from the distal to the proximal edge of the stent. The minimum stent lumen CSA was selected from each of these 1-mm measurements and was then compared with the minimum stent lumen CSA chosen by the visual scanning method to determine the best method for identifying the minimum IVUS image slice.
A proximal reference site was chosen from the IVUS images at the lumen cross section 3-mm proximal from the stent edge. The MLD at this reference site was measured and was compared with the reference diameter measured by QCA. Reference measurements were not performed if there was a major side branch within 3 mm of the proximal stent edge or if there were two stents placed continuously within 3 mm of each other.
Calculation of expected stent area and diameter.
The manufacturers stated balloon diameter at the inflation pressure written on the package was defined as the expected stent diameter. Based on this expected diameter, the expected stent area was calculated as the square of half of the expected diameter multiplied by
. The expected stent diameter and area were compared with the actual stent diameter and stent area achieved as assessed by IVUS.
Statistical analysis. All values are expressed as mean ± SD. The paired Student t test was used to compare any two measurements made on the same patients. Since different stents within the same patient may not respond independently, the patient was chosen as the unit of analysis. The stent with the largest percent of expected stent area was chosen for data analysis from each patient who had two or more stents. Analysis of variance for repeated measures with a post hoc factorial Bonferroni/Dunn test was used when more than two measurements were compared. A linear regression model was used to assess the relationship of stent lumen CSA between the manufacturers expected values and the in vivo IVUS measurements and between IVUS measurements and QCA measurements for lumen diameter. The level of significance was set at p <0.05.
| Results |
|---|
|
|
|---|
Systemic arterial hypertension was present in 19 patients (59%), diabetes mellitus in 7 (22%), hypercholesterolemia in 26 (81%), and three patients were currently smoking (9%).
Angiographic results. Angiographic characteristics are summarized in Table 1. A total of 38 stents were deployed in 32 patients. There were 28 de novo lesions. Stent deployment was successfully completed in all cases, and no major complication occurred during these procedures. Balloon inflation was performed using 16 atm during stent deployment in 31 lesions, and 14 atm pressure was used in seven cases. No lesion received a postdeployment second dilation before the IVUS examination.
|
IVUS results. The IVUS imaging was successfully completed in all stents. As shown in Table 2, the mean stent MLD was 2.5 ± 0.5 mm, and the mean minimum stent CSA was 6.0 ± 1.7 mm2.
|
Comparison of expected stent diameter and area. The MLD measured either by IVUS (2.5 ± 0.5 mm) or QCA (3.0 ± 0.4 mm) was significantly smaller than the manufacturers stated mean balloon diameter (3.5 ± 0.4 mm) at 32 stent sites from 32 patients (Fig. 1). The minimum lumen CSA measured by IVUS (6.0 ± 1.7 mm2) was also significantly less compared with the manufacturers expected stent balloon area (calculated based on the balloon diameter at 9.5 ± 1.9 mm2, p < 0.0001) (Fig. 2). There was no difference in these results between the two types of stent. The stent lumen CSA measured by IVUS correlated with the manufacturers expected stent area (r = 0.85, p < 0.0001) (Fig. 3). The mean MLD of the stents measured by IVUS was 72 ± 8% of the expected stent diameter, and the mean minimum CSA of the stents was 62 ± 10% of the expected stent area (Table 2, Fig. 4).
|
|
|
|
| Discussion |
|---|
|
|
|---|
The manufacturers expected stent diameter and area. The results of this study demonstrate that there is a significant discrepancy between the manufacturers estimation of the stented lumen dimensions and those that are actually obtained. As measured by IVUS imaging, the minimum stent lumen CSA was only 62% of the manufacturers expected stent area. The observed stent MLD was only 72% of the expected stent diameter. These results were obtained despite the use of moderately high pressures to deploy the stents (14 or 16 atm during balloon inflation). If a lower pressure had been used, such as the recommended nominal balloon inflation pressure, then a smaller diameter and CSA would have been achieved (9,13,14). Since a strong predictor of restenosis is minimum stent lumen CSA, a smaller than expected lumen would predictably increase the chance for developing restenosis (15). Based on the IVUS results, 14 lesions (37%) were redilated with larger balloons or higher pressures.
Manufacturers provide stent dimensions that are obtained by balloon inflation in air. The difference between expanding a stent in air versus an artery with resistance might explain the mismatch between the measured and predicted stent dimensions. In the seven lesions with the lowest stent expansion, extensive calcium was present with an arc of 270 to 360°. Four of these seven lesions had rotational atherectomy performed before stenting. The maximum balloon diameter during stent deployment was not significantly different from the stented lumen diameter measured by QCA. This suggests that stent recoil does not play a role in inadequate stent expansion.
Comparison with previous studies of first generation stents. The initial observations of coronary artery stenting with IVUS imaging demonstrated that the recommended pressures for deployment using the balloons available at that time resulted in smaller lumen diameters than expected by the balloon size or the measurements obtained by angiographic QCA (8,10,16). Bermejo et al. (17) reported that the final acute luminal gain achieved with the first generation stents (Palmaz-Schatz and Wiktor) at an average of 14 atm inflation pressure was only 55% of the theoretical value. Because these observations have been generally accepted by the interventional cardiology community, it has been assumed that the current state of technology has advanced such that the balloon size predicted by the manufacturers in vitro testing will be achieved at the time of actual stent deployment in vivo. This study indicates that this assumption is incorrect and that a "look up" table is necessary to interpret the expected stent deployment results provided by the manufacturers.
The differences in measurements of MLD between IVUS and QCA. In this study, the MLD measured by IVUS was significantly smaller than the MLD measured by QCA at both the in-stent and reference sites. The correlation between these measurements was weak (r = 0.5) but consistent with previous studies (1820). These results may have been caused by the two measurements not being performed at exactly the same cross section. In addition, the lumen may not be circular; thus, the minimal diameter chosen by IVUS may be less than QCA, even at the same site.
Study limitations. This study is limited because preintervention IVUS imaging was not performed routinely. Thus, tissue characterization of the underlying atherosclerotic plaque was not assessed in all lesions. Hard calcified plaque would be expected to yield less expansion than softer plaque. However, if plaque characterization were taken into account, it is likely that the manufacturers expected stent area and diameter would be even less when calcified plaque is present. In addition, in patients who received directional atherectomy, the resistance to stent expansion might be reduced due to removal of plaque bulk. However, this series of patients was nonselected by lesion type and is representative of our patient population.
Conclusions. Despite moderately high inflation pressure, the mean minimum stent CSA was only 62% of the manufacturers expected stent area. To optimize stent deployment, even with the newer generation of flexible, low-profile stents, these IVUS observations should be considered during coronary artery stenting, especially if lower pressures are used during deployment.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E. Romagnoli, G. M. Sangiorgi, J. Cosgrave, E. Guillet, and A. Colombo Drug-Eluting Stenting: The Case for Post-Dilation J. Am. Coll. Cardiol. Intv., February 1, 2008; 1(1): 22 - 31. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Aziz, J. L Morris, R. A Perry, and R. H Stables Stent expansion: a combination of delivery balloon underexpansion and acute stent recoil reduces predicted stent diameter irrespective of reference vessel size Heart, December 1, 2007; 93(12): 1562 - 1566. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Spanos, G. Stankovic, J. Tobis, and A. Colombo The challenge of in-stent restenosis: insights from intravascular ultrasound Eur. Heart J., January 2, 2003; 24(2): 138 - 150. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |