| Abstract |
|---|
|
|
|---|
The aim of this study was to evaluate the use of endovascular stents in native and recurrent coarctation of the aorta (CoA).
BACKGROUND
Stents have been used successfully in various locations. Their use in CoA can be an alternative to surgery or balloon angioplasty (BA).
METHODS
Thirty-four patients with CoA (13 native and 21 re-coarctation after surgery or BA) with a mean age of 16 ± 8 years (range 4 to 36 years) underwent attempted stent implantation between 1993 and 1999. Successful outcome was defined as peak systolic pressure gradient after stent implantation <20 mm Hg.
RESULTS
Stents were implanted in 33/34 patients, and successful outcome occurred in 32/33 patients. Peak systolic pressure gradient decreased from 32 ± 12 mm Hg to 4 ± 11 mm Hg (p < 0.001). Coarctation site to descending aorta diameter ratio increased from 0.46 ± 0.16 to 0.92 ± 0.16 (p < 0.001). Two patients underwent successful stent re-dilation 16 and 21 months after initial implantation. Six patients (18%) developed complications, including two patients who underwent surgery. Follow-up for 29 ± 17 months (range: 5 to 81 months) demonstrated no evidence of re-coarctation, aneurysm formation, stent displacement or fracture. Systolic blood pressure (SBP) decreased from 136 ± 21 mm Hg before stent placement to 122 ± 19 mm Hg at follow-up (p = 0.002). The SBP gradient decreased from 39 ± 18 mm Hg to 4 ± 6 mm Hg, and peak Doppler gradient decreased from 51 ± 26 mm Hg to 13 ± 11 mm Hg at follow-up (p < 0.001).
CONCLUSIONS
Intravascular stent placement for native and recurrent CoA has excellent results in the short and intermediate terms. Long-term outcome remains to be evaluated.
| ||||||||||||||||||||
| Methods |
|---|
|
|
|---|
Definition.
Coarctation of the aorta was defined as systolic arm-to-leg blood pressure gradient
20 mm Hg or echocardiographic, magnetic resonance imaging (MRI) or angiographic evidence of CoA. Isthmus or transverse aortic arch (TAR) hypoplasia was defined as a ratio of the diameter of these structures to the descending aorta (DAo) at the level of the diaphragm <0.6 (911). Successful outcome was defined as a peak systolic residual gradient after stent implantation of <20 mm Hg.
Patients. Thirty-four consecutive patients with CoA underwent cardiac catheterization with the intention to treat by stent implantation at our institution between May 1993 and July 1999. Characteristics of the patients and of the CoA are summarized in Tables 1 and 2. The results of six patients were previously reported (3). Eighteen patients (53%) had discrete coarctation, and 16 (47%) patients had tubular coarctation.
|
|
Before discharge, all patients underwent complete clinical evaluation, echocardiography and chest films. The patients were placed on antiplatelet therapy with aspirin for three to six months after the procedure unless they were already on anticoagulation for other reasons. One patient with obstruction in the TAR was maintained on coumadin for four months after the procedure because of the proximity of the stent to the left common carotid artery.
Statistical analysis. Data are presented as the mean ± SD. Comparison for individual parameters before and after stenting were performed using the two-tailed paired t test. Comparison between patients who had native CoA and those with re-coarctation, as well as between those with tubular versus discrete coarctation, was made using the two-tailed unpaired t test for continuous-valued data. For categorical data, chi-square analysis was used. Linear and multiple regression analyses were performed to test the effect of growth and age of the patients adjusting for variable lengths of follow-up using SAS statistical software. A p value <0.05 was considered statistically significant.
Follow-up. Clinical details, blood pressure data and results of echocardiogram, chest X-ray, MRI and catheterization were determined by retrospective chart review.
| Results |
|---|
|
|
|---|
1 because larger balloons were necessary to expand the stent adequately. The ratio of coarctation diameter to DAo diameter increased by 50% from 0.46 ± 0.16 to 0.92 ± 0.16 (p < 0.001). Balloons used had a mean diameter of 17 ± 2 mm (range: 12 to 20 mm), while the arterial sheaths used ranged from 8F to 14F.
|
Complications. Major complications (defined as those requiring surgical treatment) occurred in two patients (6%), and minor complications occurred in four patients (12%). Major complications included a retroperitoneal hematoma secondary to bleeding from the external iliac artery in a 16-year-old child because of high femoral artery puncture and a fragment of ruptured balloon embolized to the left axillary artery in another 19-year-old patient.
Minor complications included: additional stent placement in the DAo distal to the coarctation site in two patients after the partially inflated stent migrated and had to be fully expanded in the DAo. In one patient, the procedure was aborted after the stent migrated and had to be inflated in the right external iliac artery (as mentioned above). Another patient developed a small femoral arteriovenous fistula two weeks after the procedure with spontaneous resolution. Acute dissection or aneurysm formation did not occur in any of the patients.
Deaths. There were no deaths related to the procedure. One 16-year-old patient with Shones complex who had a documented influenza A sepsis and Adult Respiratory Distress syndrome had stent placement for CoA and balloon valvuloplasty of mitral stenosis to try to improve her low output state before she was placed on extracorporeal membrane oxygenation (ECMO). She died two weeks later from complications related to ECMO.
Repeat catheterization and stent dilation. Four patients underwent repeat cardiac catheterization 16 ± 5 months after their initial stent implantation (range 10 to 21 months). Two patients were catheterized for follow-up of their stenting procedure, and the other two for pre-surgical evaluation of additional cardiac defects. None of these patients was found to have aneurysm formation, stent displacement or fracture (Fig. 1). One patient had a recurrent peak gradient of 24 mm Hg across the stent secondary to somatic growth and increased cardiac index without intimal hyperplasia. He underwent successful stent dilation from 13 to 17 mm using an 18-mm HP balloon with reduction of the peak gradient to 8 mm Hg. Another patient had stent implantation at 12 years of age (and weight of 57 kg) for CoA using a 13-mm LP balloon. Re-catheterization 16 months later (at weight of 70 kg) showed a 10-mm Hg gradient across the stent, which was eliminated as the stent was dilated from 11 to 14 mm using an 18-mm HP balloon to accommodate somatic growth. The other two patients had no gradient across the stent.
Follow-up procedure. The 31 patients who were successfully stented and alive were followed for 29 ± 17 months (median: 28 months, range: 5 to 81 months). Follow-up blood pressure data are available for 28/31 patients. Two patients had aberrant right subclavian arteries, and the repair involved the left subclavian artery, while the third had no recorded blood pressure after the procedure. Sphygmomanometric systolic blood pressure (SBP) gradient decreased from 39 ± 18 mm Hg before stent placement to 4 ± 6 mm Hg at follow-up (p < 0.001), while SBP decreased from 136 ± 21mm Hg to 122 ± 19 mm Hg at the last follow-up (p = 0.002). Doppler echocardiographic data are available for 28/31 patients. Peak Doppler pressure gradient decreased from 51 ± 26 mm Hg before stent implantation to 13 ± 11 mm Hg at follow-up (p < 0.001). A total of 22/31 patients (71%) were hypertensive and 13 (42%) were on antihypertensive medications before stent implantation. At their last follow-up visit, only 8/31 (26%) remain hypertensive and on antihypertensive medications (p < 0.05), and all the patients are asymptomatic.
There was no evidence of recurrent coarctation, aneurysm formation or stent displacement or fracture identified by follow-up echocardiography and chest X-ray (in 30 patients), MRI (in 1 patient) or repeat catheterization (in 4 patients). There were no significant differences in the initial or follow-up results between patients with native and those with recurrent CoA or between patients with tubular versus discrete coarctation. Linear and multiple regression analyses, adjusting for the variable duration of follow-up, showed no significant effect of age or body surface area on the change in SBP (p = 0.6 and 0.4, respectively), the change in peak Doppler echocardiography gradient (p = 0.3 and 0.9, respectively) or the change in SBP cuff gradient (p = 0.1 and 0.5, respectively).
Residual arch gradient.
In 7/31 (23%) patients, a mild gradient in the TAR of 5 to 20 mm Hg became unmasked after the stent was successfully placed in the original coarctation site, despite having no gradient in the TAR before stent implantation. At follow-up, 3/7 continue to have a mild gradient (
20 mm Hg) and isthmus/arch hypoplasia.
| Discussion |
|---|
|
|
|---|
Initial results.
Stent implantation was performed in 33/34 patients catheterized with the intention to treat, and excellent initial results occurred in 32 of 33 patients stented, regardless to the type of their coarctation, comparable to what has been previously reported (6,7,2022). This is considerably better than BA alone where suboptimal initial results were reported in 19% to 25% of patients, especially in tight coarctation sites (
3.5 mm) and long segments of narrowing (1012,16,23,24).
Complications. Stent migration is a recognizable complication of stent placement (6,20,22). The use of the balloon-in-balloon catheter (BIB, NuMED Inc.) may provide more control of the inflation by using the better anchoring mechanism of the inner smaller balloon that is inflated first. Balloon rupture occurred in five patients, resulting in stent migration in three, but the ruptured balloons were retrieved successfully through the sheaths in four patients. Balloon rupture can be minimized by using a balloon that has the same length or slightly shorter than the stent and keeping the exchange wire as straight as possible during inflation, a factor that could be more difficult to achieve in the TAR than in the DAo or other locations.
Complications related to the arterial access were reported in up to 19% of patients with CoA after BA or stent placement (6,16,1820). This is higher than the 6% (2/34) in our study. The vascular complications in our study were related to the femoral artery cannulation and could have been prevented.
Stent re-dilation. Despite one report in animals of acute aortic rupture and death upon attempted stent re-dilation (25), several studies in animal models and in humans with CoA have shown that stents can be further expanded to accommodate somatic growth, up to three years after implantation (3,7,8,13,22). Our experience is still limited though in regards to the safety and long-term effects of multiple stent expansions, especially in younger patients. It is also unknown to what diameter a stent can be safely expanded at initial implantation. The diameter of the stented coarctation segment should not generally exceed the diameter of the isthmus and/or DAo at the level of the diaphragm. In patients with hypoplastic isthmus or tight coarctation (<4 mm), the stent should be expanded to a reasonable diameter (12 to 14 mm) and then re-dilated in six months.
Follow-up. We had no cases of aneurysm formation detected by echocardiography, repeat catheterization or MRI. Echocardiography, however, may be less sensitive than angiography, spiral computed tomography or MRI in detecting aneurysms after stent placement. Magnetic resonance imaging has a limited role in CoA after stent placement since the metallic artifact (or noise) prevents detailed evaluation of the aortic segment within the stent, despite adequate visualization of the aorta proximal and distal to the stent. Magee et al. (20) reported one patient with a small aneurysm detected by angiography, which remained stable for two years upon follow-up with serial computed tomograms.
Residual arch obstruction. Stents can be very effective in relieving arch obstruction. Three patients had obstruction in the TAR and had no residual gradient either initially after stent implantation or at follow-up. On the other hand, the relief of distal obstruction in the DAo unmasked TAR stenosis in three patients who also have isthmus or TAR hypoplasia and continue to have a mild gradient at follow-up. Isthmus and arch hypoplasia were shown to be risk factors for early and late failure after BA in CoA (10,11,19,23,24,26).
Study limitations. The use of stents in CoA remains limited to older children and adults because of the large delivery systems required and the need of multiple stent expansions to accommodate growth. It is unclear whether the noncompliant aortic stents would alter the flow pattern in a way that contributes to the occurrence of late re-stenosis and aneurysm formation in the growing aorta. We still lack adequate long-term follow-up of these patients.
Conclusions. Coarctation of the aorta can be successfully and safely managed with endovascular stents. They provide an alternative form of therapy in patients with native or recurrent CoA in the short and intermediate terms. Questions remain to be answered on the impact of stents on blood flow dynamics and the best management in patients with arch hypoplasia. Long-term follow-up is, thus, needed.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. F. Rhodes, Z. M. Hijazi, and R. J. Sommer Pathophysiology of Congenital Heart Disease in the Adult, Part II: Simple Obstructive Lesions Circulation, March 4, 2008; 117(9): 1228 - 1237. [Full Text] [PDF] |
||||
![]() |
P. F. Ford and M. A. Farber Role of Endovascular Therapies in the Management of Diverse Thoracic Aortic Pathology Perspectives in Vascular Surgery and Endovascular Therapy, June 1, 2007; 19(2): 134 - 143. [Abstract] [PDF] |
||||
![]() |
A. Eicken, U. Pensl, W. Sebening, A. Hager, T. Genz, C. Schreiber, D. Lang, H. Kaemmerer, R. Busch, and J. Hess The fate of systemic blood pressure in patients after effectively stented coarctation Eur. Heart J., May 1, 2006; 27(9): 1100 - 1105. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Tzifa, P. Ewert, G. Brzezinska-Rajszys, B. Peters, M. Zubrzycka, E. Rosenthal, F. Berger, and S. A. Qureshi Covered Cheatham-Platinum Stents for Aortic Coarctation: Early and Intermediate-Term Results J. Am. Coll. Cardiol., April 4, 2006; 47(7): 1457 - 1463. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Carr The Results of Catheter-Based Therapy Compared With Surgical Repair of Adult Aortic Coarctation J. Am. Coll. Cardiol., March 21, 2006; 47(6): 1101 - 1107. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Chessa, M. Carrozza, G. Butera, L. Piazza, D. G. Negura, C. Bussadori, E. Bossone, A. Giamberti, and M. Carminati Results and mid-long-term follow-up of stent implantation for native and recurrent coarctation of the aorta Eur. Heart J., December 2, 2005; 26(24): 2728 - 2732. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-L. Tan, S. V. Babu-Narayan, M. Y. Henein, M. Mullen, and W. Li Doppler Echocardiographic Profile and Indexes in the Evaluation of Aortic Coarctation in Patients Before and After Stenting J. Am. Coll. Cardiol., September 20, 2005; 46(6): 1045 - 1053. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. N. Raval, J. D. Telep, M. A. Guttman, C. Ozturk, M. Jones, R. B. Thompson, V. J. Wright, W. H. Schenke, R. DeSilva, R. J. Aviles, et al. Real-Time Magnetic Resonance Imaging-Guided Stenting of Aortic Coarctation With Commercially Available Catheter Devices in Swine Circulation, August 2, 2005; 112(5): 699 - 706. [Abstract] [Full Text] [PDF] |
||||
![]() |
P Ewert, S Schubert, B Peters, H Abdul-Khaliq, N Nagdyman, and P E Lange The CP stent--short, long, covered--for the treatment of aortic coarctation, stenosis of pulmonary arteries and caval veins, and Fontan anastomosis in children and adults: an evaluation of 60 stents in 53 patients Heart, July 1, 2005; 91(7): 948 - 953. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Carr, J. J. Amato, and R. S.D. Higgins Long-Term Results of Surgical Coarctectomy in the Adolescent and Young Adult With 18-Year Follow-Up Ann. Thorac. Surg., June 1, 2005; 79(6): 1950 - 1956. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Zoghbi, A. Serraf, S. Mohammadi, E. Belli, F. Lacour Gayet, B. Aupecle, J. Losay, J. Petit, and C. Planche Is surgical intervention still indicated in recurrent aortic arch obstruction? J. Thorac. Cardiovasc. Surg., January 1, 2004; 127(1): 203 - 211. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Sebastia, S. Quiroga, R. Boye, M. Perez-Lafuente, E. Castella, and A. Alvarez-Castells Aortic Stenosis: Spectrum of Diseases Depicted at Multisection CT RadioGraphics, October 1, 2003; 23(90001): S79 - 91. [Abstract] [Full Text] [PDF] |
||||
![]() |
C Duke, E Rosenthal, and S A Qureshi The efficacy and safety of stent redilatation in congenital heart disease Heart, August 1, 2003; 89(8): 905 - 912. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||