Advertisement






Click here for more guidelines.
CME Topic Collections Past Issues Search Current Issue Home
     

Right arrow Help viewing high resolution images
Right arrow Return to article

Please click here to obtain permission to reproduce this image.

Click on image to view larger version.



Figure 4 The intervals separating the two components of double potentials (DP1-2) recorded along the ablation line during coronary sinus pacing before and after complete block in the cavo-tricuspid isthmus are shown. Also depicted are the mean values ± SD. Note that a DP1-2 interval <90 ms was always associated with incomplete block and a DP1-2 interval ≥110 ms was always associated with complete block. Although there were 30 patients in this study, the sample size during incomplete block was 25 patients, because a DP was not always present during incomplete isthmus block.