CLINICAL STUDY: INTERVENTIONAL CARDIOLOGY
Cutting balloon angioplasty for the treatment of in-stent restenosis: a matched comparison with rotational atherectomy, additional stent implantation and balloon angioplasty
Milena Adamian, MD, PhD ,
Antonio Colombo, MD, FACC*,
Carlo Briguori, MD, PhD*,
Takahiro Nishida, MD*,
Federica Marsico, MD*,
Carlo Di Mario, MD, PhD, FACC*,
Remo Albiero, MD*,
Issam Moussa, MD and
Jeffrey W. Moses, MD, FACC
* EMO Centro Cuore Columbus, Milan, Italy
Lenox Hill Heart and Vascular Institute, New York, New York, USA
Manuscript received January 23, 2001;
revised manuscript received April 27, 2001,
accepted June 8, 2001.
Reprint requests and correspondence: Dr. Antonio Colombo, EMO Centro Cuore Columbus, Via M. Buonarroti 48, 20145 Milan, Italy columbus{at}micronet.it
 |
Abstract
|
|---|
OBJECTIVES
The purpose of the study was to determine whether cutting balloon angioplasty (CBA) has advantages over other modalities in treatment of in-stent restenosis (ISR).
BACKGROUND
Controversies exist regarding optimal treatment for ISR. Recently, CBA emerged as a tool in management of ISR.
METHODS
A total of 648 lesions treated for ISR were divided into four groups according to the treatment strategy: CBA, rotational atherectomy (ROTA), additional stenting (STENT), and percutaneous transluminal coronary angioplasty (PTCA). Following the matching process, 258 lesions were entered into the analysis.
RESULTS
Baseline clinical and angiographic characteristics were similar among the groups (p = NS). Acute lumen gain was significantly higher in the STENT group (2.12 ± 0.7 mm), whereas in the CBA group the gain was similar to one achieved following ROTA and following PTCA (1.70 ± 0.6 vs. 1.79 ± 0.5 mm and 1.56 ± 0.7 mm, respectively; p = NS). The lumen loss at follow-up was lower for the CBA versus ROTA and versus STENT (0.63 ± 0.6 vs. 1.30 ± 0.8 mm and 1.36 ± 0.8 mm, respectively; p < 0.0001), yielding a lower recurrent restenosis rate (20% vs. 35.9% and 41.4%, respectively; p < 0.05). By multivariate analysis, CBA (odds ratio [OR] = 0.17; confidence interval [CI], 0.06 to 0.51; p = 0.001) and diffuse restenosis type at baseline (OR = 2.07; CI, 1.15 to 3.71; p = 0.02) were identified as predictors of target lesion revascularization.
CONCLUSIONS
We conclude that CBA is a safe and efficient technique for treatment of ISR, with immediate results similar to atheroablation and better clinical and angiographic outcomes at follow-up. This approach might be implemented as a viable option in management of focal ISR and to prepare diffuse ISR for brachytherapy treatment.
|
Abbreviations and Acronyms
| | CABG | = coronary artery bypass graft surgery | | CBA | = cutting balloon angioplasty | | CI | = confidence interval | | CK | = creatine kinase | | CK-MB | = creatine kinase-myocardial band | | ISR | = in-stent restenosis | | IVUS | = intravascular ultrasound | | MACE | = major adverse cardiac events | | MI | = myocardial infarction | | MLD | = minimal lumen diameter | | NQMI | = nonQ-wave myocardial infarction | | OR | = odds ratio | | PTCA | = percutaneous transluminal coronary angioplasty | | QCA | = quantitative coronary angiography | | QMI | = Q-wave myocardial infarction | | ROTA | = rotational atherectomy | | STENT | = additional stenting | | TLR | = target lesion revascularization |
|
With the expanded application of coronary stenting beyond the criteria for inclusion in the randomized trials, stent restenosis emerged as a new "disease entity." With over 800,000 stents a year in the U.S., and a stent restenosis incidence of 20% to 50% of lesions, it becomes clear that in-stent restenosis (ISR) is a significant problem (1). The most common modality to treat ISR is percutaneous transluminal coronary angioplasty (PTCA), but long-term results are unsatisfactory, with a high recurrence rate, particularly after treatment of diffuse (lesion length >10 mm) ISR (2). Additional stent implantation (STENT) may have merit because it reduces acute recoil and helps tissue extrusion; however, this acute benefit did not translate into improved long-term outcome (35). Atheroablative therapies (i.e., rotational atherectomy [ROTA] and excimer laser angioplasty) have been used in an attempt to decrease the rate of recurrence. However, the beneficial effect of these devices was not proven (6). Recent reports suggested that cutting balloon angioplasty (CBA) might be of benefit in the treatment of ISR, decreasing the need for repeat revascularization procedures (7,8).
The aim of our study was to compare acute and long-term angiographic and clinical outcomes between the various techniques that have been used to treat ISR in matched lesion subsets.
 |
Methods
|
|---|
Patient and lesion characteristics.
Between January 1997 and February 1999, a total of 1,410 patients (1,980 lesions) underwent successful coronary stent implantation at Centro Cuore Columbus, Milan, Italy. Out of this population, 684 lesions (38%) had ISR. These lesions were divided into four groups based on the ISR treatment modality: PTCA, STENT, ROTA and CBA. The groups were matched according to clinical and angiographic characteristics: presence of diabetes mellitus, reference vessel size, minimal lumen diameter (MLD), lesion length and type of ISR. Following the matching process and the random selection of only one lesion per patient, the final patient population totaled 258 (258 lesions). The numbers of patients/lesions in each group were: CBA, 57; ROTA, 48; STENT, 79; and PTCA, 74.
Angiographic follow-up was obtained in 85% of the eligible patients at an average duration of 6.2 ± 3.2 months. Clinical follow-up was obtained through outpatient clinic visits, by direct telephone interview and with the referring physician when additional information was necessary. Clinical follow-up was obtained in 100% of patients at an average of 11 ± 8 months.
 |
Procedure
|
|---|
Before angioplasty, 70 U/kg of intravenous heparin and 0.2 mg nitroglycerin intracoronary were administered. All patients received aspirin 300 mg q.d. and ticlopidine 250 mg b.i.d. starting three days before the procedure when feasible. Each operator selected a particular procedure according to his judgment. In general, we can state that ROTA, PTCA and stenting were used more frequently in the early period while CBA was used more recently. Each patient was encouraged to return for angiographic follow-up between the fifth and sixth month following the procedure.
Cutting balloon angioplasty.
Cutting balloon (InterVentional Technologies, San Diego, California) is a conventional balloon catheter 10 or 15 mm in length with three (on smaller balloon sizes) or four microblades (on balloon sizes >3.5 mm in diameter). These blades, mounted longitudinally on the surface of balloon, are 0.25 mm in height and three to five times sharper than conventional surgical blades (9). During dilation, the device produces three or four endovascular surgical incisions. The CBA was performed with multiple inflations, increasing the balloon pressure, usually up to 12 atm.
ROTA with adjunct balloon angioplasty.
The techniques used for ROTA have been described elsewhere (6,10). Rotablator or Rotalink devices (Boston Scientific, Scimed, Maple Grove, Minnesota) were used with an incremental burr size approach to achieve a burr/artery ratio 0.7. After ROTA, adjunct PTCA was routinely performed with balloon-to-artery ratio of 1:1 utilizing medium to high pressure (12 to 16 atm).
Additional stent implantation.
Restenotic lesions treated by additional stent implantation were predilated using a traditional balloon angioplasty. If necessary, multiple stents were used to cover the entire lesion. Postdilation was performed with an appropriately sized balloon at high pressure ( 14 atm).
PTCA.
The PTCA procedure was done using noncompliant or semicompliant balloons that matched the size of the final balloon used at the time of stent implantation. Single or multiple high-pressure balloon inflations ( 12 atm) were generally performed with the goal of achieving a near 0% residual stenosis. A larger balloon (0.5 mm) than the original postdilation balloon was used at the operators discretion if the stent was considered undersized compared to the reference vessel size.
Intravascular ultrasound evaluation.
Intravascular ultrasound (IVUS) was performed only in situations where a possible stent underexpansion was thought to be present. On average, IVUS was used in 21%, 22%, 24%, and 34% in CBA, ROTA, STENT and PTCA groups, respectively (p = NS).
Definitions.
In-stent restenosis was defined as >50% diameter stenosis by quantitative coronary angiography (QCA) inside the stent or within 5 mm from the edges of a stented lesion presenting at least eight weeks after the stent implantation. In addition, ISR was defined as focal (<10 mm in length) or diffuse (>10 mm in length).
Major adverse cardiac events (MACE) were defined as death, any myocardial infarction (MI) and/or repeat revascularization (coronary artery bypass graft surgery [CABG]/repeat PTCA). A diagnosis of Q-wave myocardial infarction (QMI) was made when there was documentation of new pathologic Q-waves (>0.04 s) on an electrocardiogram in conjunction with an elevation of total creatine kinase (CK) more than twice the normal value, with a concomitant elevation of creatine kinase-myocardial band. A diagnosis of nonQ-wave myocardial infarction (NQMI) was made when an elevation of total CK to greater than twice the upper limit of normal value, with a concomitant elevation of CK-MB, was documented without development of new pathologic Q-waves. Procedural success was defined as achievement of a residual diameter stenosis <30% at the lesion site without the occurrence of death, bypass surgery, or QMI. Recurrent restenosis at follow-up angiography was defined as 50% diameter stenosis by quantitative coronary angiography (QCA) occurring not earlier than eight weeks after successful treatment of ISR.
QCA.
For each lesion, the single view showing the most severe degree of stenosis was used for QCA with a computer-assisted automated edge detection algorithm (QCA-CMS version 4.0, MEDIS, Leiden, the Netherlands). Both absolute reference and MLD in millimeters were determined using the guiding catheter filled with contrast for calibration. The lesion length was measured as the distance from shoulder to shoulder. In complex lesions with the involvement of adjacent segments proximal and distal, a user-defined reference lumen diameter of a proximal and distal angiographically normal-appearing segment was chosen.
Acute lumen gain was defined as the MLD immediately after the procedure minus the MLD before the procedure. Late lumen loss was defined as the MLD after the procedure minus the MLD at follow-up. The late-loss index was defined as the late lumen loss divided by the acute lumen gain.
Matching process.
Matching was performed by a computerized program and was based on principles of the matching process derived from Umans et al. (11). The database was reviewed sequentially; for each restenotic lesion treated with the cutting balloon, the first lesion encountered in each of the three different treatment groups (ROTA, STENT, CBA) that satisfied the matching parameters was chosen. The matching parameters in order of sequential selection were: 1) diabetes; 2) reference diameter ± 0.3 mm; 3) baseline MLD ± 0.1 mm; 4) lesion length ± 1 mm; and 5) type of restenosis. The order we selected to perform the lesion selection for the matching process influenced the parameter, which could not always find a corresponding lesion with the same parameter.
Statistical analysis.
Statistical analysis was performed using the StatView statistical package (StatView 5, SAS Institute, Cary, North Carolina). To remove any confounding factor due to the behavior of different lesions in the same patient, we used the patients as the unit of the analysis. For patients with multiple lesions, only one randomly selected lesion was included in the analysis according to the following randomization scheme; each lesion was assigned a random number between 0 and 1. The lesion corresponding to the smallest number was included in the analysis.
Comparison among the groups was performed by the 2 test (or the Fisher exact test) for categorical data. A value of p < 0.05 was considered statistically significant. The analysis of variance test was performed for continuous data. Multiple comparisons were accounted for by using the Bonferroni-Dunn method to preserve the overall significance level. Results were expressed as mean value ± SD. Independent risk factors were determined using a multivariate logistic regression model. Results were presented as the odds ratio (OR) and 95% confidence interval (CI) for each variable. Values with a significance level of p < 0.05 were accepted as significant.
 |
Results
|
|---|
Patient characteristics and procedural outcome.
There was no difference in the baseline clinical and lesion characteristics among the four groups (Tables 1 and 2). Procedural success was achieved in all patients in the CBA group, whereas there were one NQMI in the ROTA group, two QMIs in the STENT group, and three events in the PTCA group (1 death; 1 emergency CABG; and 1 QMI). However, the difference in incidence of in-hospital MACE among the four groups did not reach statistical significance.
Angiographic analysis.
Quantitative angiographic measurements, both pre- and postintervention and at follow-up, are summarized in Table 2. No significant differences were seen in baseline angiographic measurements among the four groups. After intervention, the greatest MLD was achieved in the STENT group, without statistical significance, compared to the ROTA and CBA groups. Lesions treated with PTCA alone had the smallest final MLD. In the STENT group, the balloon was inflated at a higher pressure (15.7 ± 3.9 atm) than in the CBA group (10.7 ± 2.8 atm; p < 0.0001). A significantly higher late loss at follow-up was observed in the ROTA, STENT, and PTCA groups (1.30 ± 0.8 mm vs. 1.36 ± 0.8 mm vs. 1.07 ± 0.8, respectively) compared to the CBA group (0.63 ± 0.7 mm; p < 0.0001). In addition, the loss index was significantly lower in the group of lesions treated with CBA (0.38 ± 0.3 mm) than in the other groups.
Long-term clinical outcome.
The mean duration of clinical follow-up was 11 ± 8 months. Death, CABG and MI occurred with equal proportions in all groups (Table 3). Recurrent ISR rate was significantly lower in the CBA group compared to other groups (CBA, 20%; ROTA, 35.9%; STENT, 41.4%; and PTCA, 45.2%; p = 0.04) with a significantly lower target lesion revascularization (TLR) rate (CBA, 15.8%; ROTA, 31.9%; STENT, 35.5%; and PTCA, 37.8%; p = 0.03). A diffuse pattern of recurrence was more common in lesions treated with PTCA, whereas focal pattern of recurrence was more common in the CBA group (Table 3).
Predictors of TLR.
By univariate analysis, the presence of final MLD, diffuse ISR at baseline, and CBA were all identified as risk factors of TLR. However, in the multivariate model, CBA (OR 0.33; p = 0.003) and diffuse pattern of ISR at baseline (OR 2.07; p = 0.02) were distinguished as predictors of repeat revascularization procedures at follow-up (Table 4).
Using logistic regression analysis we also looked for factors triggering the occurrence of diffuse recurrent restenosis. Using multivariate analysis, the diffuse character of the initial ISR and treatment with PTCA were identified as independent predictors of recurrence of diffuse ISR.
 |
Discussion
|
|---|
The results of our study indicate that CBA can be safely used to treat ISR with acceptable acute angiographic and clinical outcome. Moreover, CBA significantly decreases the rate of repeat revascularization procedures at follow-up compared to other commonly used techniques for management of ISR.
Current treatment of ISR.
In contrast to angioplasty, where restenosis is predominantly caused by elastic recoil and vascular remodeling, ISR is almost exclusively due to neointimal hyperplasia (12,13). Although the optimal treatment of ISR has not yet been well defined, three treatment approaches are commonly used: 1) PTCA; 2) atheroablation (ROTA, excimer laser angioplasty [ELCA] and directional coronary atherectomy); and 3) STENT. Recently, brachytherapy has been effectively used to prevent angiographic and clinical recurrence following optimal mechanical dilation (14). Most of the studies comparing different treatment strategies for ISR were not randomized or matched comparisons. Of importance is that not all ISR lesions have a similar risk of recurrence (15).
The simplest way to treat ISR is repeat dilation using a conventional PTCA balloon. Mechanism of lumen enlargement after PTCA for ISR is controversial. Although Gordon et al. (16) attributed it to neointimal tissue compression and extrusion of out-of-stent struts, an IVUS study by Mehran et al. (17) showed that tissue extrusion and additional stent expansion contributed equally to lumen improvement. In the Mehran et al. study, there was 32% of residual neointimal tissue of the stent area following PTCA.
In general, recurrence after PTCA treatment of ISR has been reported to range from 37% to 50% (angiographic restenosis) and from 14% to 30% (clinical restenosis) (2,1719).
In our series, PTCA treatment gave the lowest acute gain (1.56 ± 0.7 mm) with a late loss of 0.89 ± 0.6 mm at follow-up associated with the highest rates of angiographic and clinical recurrence (45.2% and 41.4%, respectively). In the past we reported better long-term results following PTCA for ISR with a target vessel revascularization of 11% (19). The discordance between the results of the current study and our former report can be explained by the larger reference vessel size, the shorter lesion length, and the very low angiographic follow-up rate reported in that study.
Atheroablative techniques have been used in an attempt to improve the acute results of PTCA by increasing lumen dimensions through ablation of neointimal tissue (20,21). Clinical recurrence after ROTA has ranged from 28% to 50%, with an angiographic restenosis rate up to 45% (10,2224). In one study (25), 28% of TLRs occurred after ROTA with adjunctive PTCA versus 46% with PTCA alone.
In another study, vom Dahl et al. (6) recently reported that the occurrence of re-restenosis correlated with length of the primarily stented lesion and was as high as 49% (TLR, 35%) for long diffuse lesions treated with ROTA and adjunctive balloon angioplasty. A subsequent randomized study confirmed the lack of additional benefit of ROTA and adjunctive PTCA versus simple angioplasty (26).
Preliminary observations indicate that additional stent implantation achieves the best acute results (QCA diameter stenosis of 10%) by recovering all the lumen area of the original stenting procedure, primarily via neointimal tissue extrusion out of the stent, with some additional stent expansion (5,18). However, angiographic recurrence after additional stent implantation has ranged from 30% to 35%, with clinical recurrence from 17% to 40% (3,4). The highest acute gain and the largest acute MLD were obtained after STENT. However, these acute results were offset by a high late lumen loss (about 47%) at follow-up. Different amounts of net gain obtained with PTCA, STENT and ROTA may explain the different angiographic and clinical restenosis rates. A comparison of PTCA, ELCA, ROTA and STENT in a large series of 821 restenotic lesions was concluded with the statement that "all interventional strategies are disappointing," demonstrating similar late clinical outcome independent of device choice (27).
CBA.
The cutting balloon catheter is a relatively novel device. The concept of CBA rests on the presence of microblades intended to incise the atherosclerotic plaque or neointimal tissue at the beginning of balloon inflation and to develop a controlled fault line along which dilation will occur (9). The main application of CBA was in noncalcified lesions with concentric plaque; however, this technique was subsequently applied effectively to treat several types of lesions including ISR (8,28). The Japanese Multicenter Registry of CBA for ISR reported data on 194 lesions treated. Angiographic restenosis occurred in 29% and target lesion revascularization in 22% of the lesions (29). These findings are consistent with those reported by Chevalier et al. (7) demonstrating better acute and follow-up angiographic results in treatment of ISR when CBA is compared to PTCA. The acute gain obtained was significantly bigger in the CBA group compared to the PTCA group (2.1 ± 0.47 mm vs. 1.74 ± 0.58 mm; p < 0.05). The TLR at nine months was 12% in the CBA group (7). In our series, despite the fact that lumen size immediately after CBA was not as large as after additional stent implantation or ROTA with adjunctive PTCA, follow-up MLD in the CBA group was significantly larger compared to those obtained with other techniques (Fig. 1). Both late lumen loss and loss index were significantly lower in the CBA group compared to the other techniques, resulting in a lower angiographic restenosis rate (20%) and need for TLR (15.7%).

View larger version (32K):
[in this window]
[in a new window]
|
Figure 1 Cumulative frequency distribution of the late lumen loss at follow-up in cutting balloon angioplasty (CBA), percutaneous transluminal coronary angioplasty (PTCA), rotational atherectomy (ROTA), and additional stenting (STENT) groups. Significantly lower lumen loss noticed in CBA group compared to ROTA and STENT groups (0.63 ± 0.6 mm, 1.30 ± 0.8 mm, and 1.36 ± 0.8 mm; p < 0.0001).
|
|
The possible mechanisms for dilation with CBA for ISR are probably related to plaque extrusion through the stents struts (30,31) compared to other techniques, possibly, with less tissue injury. Figure 2 demonstrates a typical IVUS finding following CBA for ISR. Another added advantage of the technique of CBA for ISR is the lack of occurrence of the "watermelon seeding effect." This fact could become a problem when contemplating additional brachytherapy, a setting where there is the need to carefully control the boundaries of the injured segment.

View larger version (82K):
[in this window]
[in a new window]
|
Figure 2 Acute angiographic and intravascular ultrasound (IVUS) results obtained after cutting balloon angioplasty (CBA) for in-stent restenosis (ISR) in left circumflex artery. (a) Preintervention IVUS image shows complete neointimal filling of the lumen in the stent (lumen cross-sectional area 1.3 mm2; (b) Final result following CBA with 2.5-mm balloon (lumen cross-sectional area 6.0 mm2).
|
|
Predictors of recurrent ISR.
Various investigators reported a recurrence rate following treatment of ISR of between 31% and 85% with relation to the focal or diffuse type of original restenosis (2,6,22).In our study, use of CBA predicted a lower recurrence rate of restenosis but did not predict a lower occurrence of diffuse ISR at the time of the second recurrence. For this reason, and, more importantly, for the high frequency of focal ISR at baseline, we caution to state that CBA could be a viable solution to the problem of diffuse ISR. The important and positive features of the CBA approach for diffuse ISR are to provide a reasonable large final lumen even in large vessels, with no need to use a large guiding catheter; with a lower risk of distal embolization compared to ROTA or directional atherectomy; with the possibility to select carefully the length of the injury segment, and with very limited need for additional stent implantation. When dealing with a diffuse or proliferative ISR, the CBA approach seems the most logical way to prepare the lesion for a more definite therapy such as brachytherapy (32).
Study limitations.
There are several limitations to the present study. It is a retrospective study and therefore inherently contains all the disadvantages of such a comparative analysis. The matching process was performed only for a certain set of variables, and this technique may suffer from biases of retrospective evaluations. There were no data regarding IVUS assessment of the treated lesions so as to understand better the mechanism of CBA in ISR.
Conclusions.
Finally, CBA seems to be a reasonable approach for patients with ISR. This technique provides immediate results superior to simple angioplasty and quite similar to more complex and expensive atheroablative techniques. Follow-up results do not seem to penalize the CBA approach by a higher late loss compared to other approaches.
We believe that this technique can be considered a reasonable first line and possible definite approach for most focal ISRs and for the preparation of diffuse ISR to other more definite treatment modalities. To evaluate better the definite value of CBA versus PTCA to treat ISR, a randomized study has recently been launched in Europe.
 |
Footnotes
|
|---|
No financial support was received for this study.
 |
References
|
|---|
1. Baim DS, Levine MJ, Leon MB, Levine S, Ellis SG, Schatz RA. Management of restenosis within the Palmaz-Schatz coronary stent (the U.S. multicenter experience). The U.S. Palmaz-Schatz Stent Investigators. Am J Cardiol. 1993;71:364366[CrossRef][Medline]
2. Eltchaninoff H, Koning R, Tron C, Gupta V, Cribier A. Balloon angioplasty for the treatment of coronary in-stent restenosis: immediate results and 6-month angiographic recurrent restenosis rate. J Am Coll Cardiol. 1998;32:980984[Abstract/Free Full Text]
3. Elezi S, Kastrati A, Hadamitzky M, Dirschinger J, Neumann FJ, Schomig A. Clinical and angiographic follow-up after balloon angioplasty with provisional stenting for coronary in-stent restenosis. Cathet Cardiovasc Interv. 1999;48:151156[CrossRef][Medline]
4. Al-Sergani HS, Ho PC, Nesto RW, et al. Stenting for in-stent restenosis: a long-term clinical follow-up. Cathet Cardiovasc Interv. 1999;48:143148[CrossRef][Medline]
5. Antoniucci D, Valenti R, Moschi G, et al. Stenting for in-stent restenosis. Cathet Cardiovasc Interv. 2000;49:376381[CrossRef][Medline]
6. vom Dahl J, Radke PW, Haager PK, et al. Clinical and angiographic predictors of recurrent restenosis after percutaneous transluminal rotational atherectomy for treatment of diffuse in-stent restenosis. Am J Cardiol. 1999;83:862867[CrossRef][Medline]
7. Chevalier B, Guyon P, Glatt B. Treatment of in-stent restenosis: short- and mid-term results of a pilot study between balloon and cutting balloon. (abstr)J Am Coll Cardiol. 1999;33:62A
8. Kurbaan AS, Foale RA, Sigwart U. Cutting balloon angioplasty for in-stent restenosis. Cathet Cardiovasc Interv. 2000;50:480483[Medline]
9. Barath P, Fishbein MC, Vari S, Forrester JS. Cutting balloon: a novel approach to percutaneous angioplasty. Am J Cardiol. 1991;68:12491252[CrossRef][Medline]
10. Radke PW, Klues HG, Haager PK, et al. Mechanisms of acute lumen gain and recurrent restenosis after rotational atherectomy of diffuse in-stent restenosis: a quantitative angiographic and intravascular ultrasound study. J Am Coll Cardiol. 1999;34:3339[Abstract/Free Full Text]
11. Umans VA, Keane D, Quaedvlieg P, Serruys PW. Matching to guide the design and predict the outcome of randomized atherectomy trials. Am Heart J. 1995;130:11351143[CrossRef][Medline]
12. Hoffmann R, Mintz GS, Dussaillant GR, et al. Patterns and mechanisms of in-stent restenosis. A serial intravascular ultrasound study. Circulation. 1996;94:12471254[Abstract/Free Full Text]
13. Kornowski R, Mintz GS, Kent KM, et al. Increased restenosis in diabetes mellitus after coronary interventions is due to exaggerated intimal hyperplasia. A serial intravascular ultrasound study. Circulation. 1997;95:13661369[Abstract/Free Full Text]
14. Kastrati A, Schomig A, Elezi S, et al. Predictive factors of restenosis after coronary stent placement. J Am Coll Cardiol. 1997;30:14281436[Abstract]
15. Mehran R, Dangas G, Abizaid AS, et al. Angiographic patterns of in-stent restenosis: classification and implications for long-term outcome. Circulation. 1999;100:18721878[Abstract/Free Full Text]
16. Gordon PC, Gibson CM, Cohen DJ, Carrozza JP, Kuntz RE, Baim DS. Mechanisms of restenosis and redilation within coronary stentsquantitative angiographic assessment. J Am Coll Cardiol. 1993;21:11661174[Abstract]
17. Mehran R, Mintz GS, Popma JJ, et al. Mechanisms and results of balloon angioplasty for the treatment of in-stent restenosis. Am J Cardiol. 1996;78:618622[CrossRef][Medline]
18. Bauters C, Banos JL, Van Belle E, McFadden EP, Lablanche JM, Bertrand ME. Six-month angiographic outcome after successful repeat percutaneous intervention for in-stent restenosis. Circulation. 1998;97:318321[Abstract/Free Full Text]
19. Reimers B, Moussa I, Akiyama T, et al. Long-term clinical follow-up after successful repeat percutaneous intervention for stent restenosis. J Am Coll Cardiol. 1997;30:186192[Abstract]
20. Lee SG, Lee CW, Cheong SS, et al. Immediate and long-term outcomes of rotational atherectomy versus balloon angioplasty alone for treatment of diffuse in-stent restenosis. Am J Cardiol. 1998;82:140143[CrossRef][Medline]
21. Mahdi NA, Pathan AZ, Harrell L, et al. Directional coronary atherectomy for the treatment of Palmaz-Schatz in-stent restenosis. Am J Cardiol. 1998;82:13451351[CrossRef][Medline]
22. Mehran R, Dangas G, Mintz GS, et al. Treatment of in-stent restenosis with excimer laser coronary angioplasty versus rotational atherectomy: comparative mechanisms and results. Circulation. 2000;101:24842489[Abstract/Free Full Text]
23. Sharma SK, Duvvuri S, Dangas G, et al. Rotational atherectomy for in-stent restenosis: acute and long-term results of the first 100 cases. J Am Coll Cardiol. 1998;32:13581365[Abstract/Free Full Text]
24. Goldberg SL, Berger P, Cohen DJ, et al. Rotational atherectomy or balloon angioplasty in the treatment of intra-stent restenosis: BARASTER multicenter registry. Cathet Cardiovasc Interv. 2000;51:407413[CrossRef][Medline]
25. Dauerman HL, Baim DS, Cutlip DE, et al. Mechanical debulking versus balloon angioplasty for the treatment of diffuse in-stent restenosis. Am J Cardiol. 1998;82:277284[CrossRef][Medline]
26. Vom Dahl J, Silber S, Buettner H, et al. Rotational atherectomy versus balloon angioplasty for diffuse in-stent restenosis: preliminary results of a randomized multicenter trial (ARTIST trial). (abstr)Am J Cardiol. 1999;84(Suppl 6A):237[Medline]
27. Mehran R. In-stent restenosis: "The great equalizer"disappointing outcomes with all interventional strategies. (abstr)J Am Coll Cardiol. 1999;33:63A
28. Kurbaan AS, Kelly PA, Sigwart U. Cutting balloon angioplasty and stenting for aorto-ostial lesions. Heart. 1997;77:350352[Abstract/Free Full Text]
29. Nakamura M, Matsubara T. Results of cutting baloon angioplasty for stent restenosis. Japanese Multicenter Registry. (abstr)J Am Coll Cardiol. 1998;31:235A
30. Albiero R, Nishida T, Karvouni E, et al. Cutting balloon angioplasty for the treatment of in-stent restenosis. Cathet Cardiovasc Interv. 2000;50:452459[CrossRef][Medline]
31. Kinoshita Y, Matsuno Y, Ohta T, et al. Dilatation mechanism of cutting balloon angioplasty for in-stent restenosis: comparison with plain old balloon angioplasty using volumetric intravascular ultrasound analysis. (abstr)Am J Cardiol. 2000;86(Suppl 8A):120i
32. Teirstein PS, Massullo V, Jani S, et al. Three-year clinical and angiographic follow-up after intracoronary radiation: results of a randomized clinical trial. Circulation. 2000;101:360365[Abstract/Free Full Text]
This article has been cited by other articles:

|
 |

|
 |
 
P. M Patel, J. Eisenberg, M A. Islam, A. O Maree, and K. A Rosenfield
Percutaneous revascularization of persistent renal artery in-stent restenosis
Vascular Medicine,
August 1, 2009;
14(3):
259 - 264.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
V. G. Cin, H. Pekdemir, M. N. Akkus, A. Camsari, O. Doven, and S. Yenihan
Cutting Balloon Angioplasty for the Treatment of In-Stent Restenosis in Diabetics: A Matched Comparison of 6 Months' Outcome With Conventional Balloon Angioplasty
Angiology,
August 1, 2006;
57(4):
445 - 452.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Alfonso, R. Melgares, V. Mainar, R. Lezaun, N. Vazquez, J. Tascon, F. Pomar, A. Cequier, J. Angel, M.-J. Perez-Vizcayno, et al.
Therapeutic implications of in-stent Restenosis located at the stent edge.: Insights from the Restenosis Intra-stent Balloon angioplasty versus elective Stenting (RIBS) randomized trial
Eur. Heart J.,
October 2, 2004;
25(20):
1829 - 1835.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F Saia, P A Lemos, C A Arampatzis, A Hoye, M Degertekin, K Tanabe, G Sianos, P C Smits, W J van der Giessen, P J de Feyter, et al.
Routine sirolimus eluting stent implantation for unselected in-stent restenosis: insights from the rapamycin eluting stent evaluated at Rotterdam cardiology hospital (RESEARCH) registry
Heart,
October 1, 2004;
90(10):
1183 - 1188.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. Albiero, S. Silber, C. Di Mario, C. Cernigliaro, S. Battaglia, B. Reimers, A. Frasheri, V. Klauss, J. M. Auge, P. Rubartelli, et al.
Cutting balloon versus conventional balloon angioplasty for the treatment of in-stent restenosis: Results of the restenosis cutting balloon evaluation trial (RESCUT)
J. Am. Coll. Cardiol.,
March 17, 2004;
43(6):
943 - 949.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Degertekin, P. A. Lemos, C. H. Lee, K. Tanabe, J.E. Sousa, A. Abizaid, E. Regar, G. Sianos, W. J. van der Giessen, P. J. de Feyter, et al.
Intravascular ultrasound evaluation after sirolimus eluting stent implantation for de novo and in-stent restenosis lesions
Eur. Heart J.,
January 1, 2004;
25(1):
32 - 38.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Alfonso, J. Zueco, A. Cequier, R. Mantilla, A. Bethencourt, J. R. Lopez-Minguez, J. Angel, J. M. Auge, M. Gomez-Recio, C. Moris, et al.
A randomized comparison ofrepeat stenting with balloon angioplasty in patients with in-stent restenosis
J. Am. Coll. Cardiol.,
September 3, 2003;
42(5):
796 - 805.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. E. Ajani, R. Waksman, E. Cheneau, D.-H. Cha, S. McGlynn, M. Castagna, R. C. Chan, L. F. Satler, K. M. Kent, A. D. Pichard, et al.
The outcome of percutaneous coronary intervention in patients with In-Stentrestenosis who failed intracoronary radiation therapy
J. Am. Coll. Cardiol.,
February 19, 2003;
41(4):
551 - 556.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. Tanabe, P. W. Serruys, E. Grube, P. C. Smits, G. Selbach, W. J. van der Giessen, M. Staberock, P. de Feyter, R. Muller, E. Regar, et al.
TAXUS III Trial: In-Stent Restenosis Treated With Stent-Based Delivery of Paclitaxel Incorporated in a Slow-Release Polymer Formulation
Circulation,
February 4, 2003;
107(4):
559 - 564.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Degertekin, E. Regar, K. Tanabe, P. C. Smits, W. J. van der Giessen, S. G. Carlier, P. de Feyter, J. Vos, D. P. Foley, J. M. R. Ligthart, et al.
Sirolimus-eluting stent for treatment of complex in-stent restenosis: The first clinical experience
J. Am. Coll. Cardiol.,
January 15, 2003;
41(2):
184 - 189.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Engelke, C. Sandhu, R. A. Morgan, and A.-M. Belli
Using 6-mm Cutting Balloon Angioplasty in Patients with Resistant Peripheral Artery Stenosis: Preliminary Results
Am. J. Roentgenol.,
September 1, 2002;
179(3):
619 - 623.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. C. Lowe, S. N. Oesterle, and L. M. Khachigian
Coronary in-stent restenosis: Current status and future strategies
J. Am. Coll. Cardiol.,
January 16, 2002;
39(2):
183 - 193.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. B. King III
A cutting edge technology, or is it?
J. Am. Coll. Cardiol.,
September 1, 2001;
38(3):
680 - 681.
[Full Text]
[PDF]
|
 |
|
|