J Am Coll Cardiol, 2000; 36:871-877 © 2000 by the American College of Cardiology Foundation |




* Services de Cardiologie, Hopital Arnaud de Villeneuve, Montpellier, France
Service dExploration Fonctionnelle, Hopital Arnaud de Villeneuve, Montpellier, France
Service de Chirurgie Cardiaque et Vasculaire, Hopital Arnaud de Villeneuve, Montpellier, France
Manuscript received August 25, 1998; revised manuscript received March 1, 2000, accepted April 12, 2000.
Reprint requests and correspondence: Dr. Pierre Becassis, Service de Cardiologie B, Hopital A. de Villeneuve, 371 Avenue Doyen Gaston Giraud, 34000 Montpellier, France
| Abstract |
|---|
|
|
|---|
The objective of this study was to determine whether a small-size valve prosthesis contributes to exercise intolerance, as assessed by VO2 measurement during an exhaustive cycle ergometer exercise.
BACKGROUND
The determinants of exercise capacity after mechanical aortic replacement are not well known. The selection of small valve sizes has, however, been described as an independent predictor of exercise intolerance as assessed by exercise duration. Maximal oxygen uptake (VO2 max) is a good index of exercise tolerance.
METHODS
Fourteen patients were eligible, with a mean age of 62 ± 6 years. Before surgery, the mean left ventricular ejection fraction (LVEF) was 73 ± 8%. Two valve types with small diameter (19 to 21 mm) were used: Medtronic Hall and St Jude Medical. A healthy sedentary control group (n = 14) paired for age, weight and size was constituted. After one year of follow-up, cardiorespiratory tests were performed. In addition, the gradients through the prostheses were determined by continuous pulse Doppler at rest and immediately after the cardiorespiratory test.
RESULTS
The exercise tolerance was not significantly different between the control group and patient group: VO2 peak (21.7 vs. 20.4 ml/kg/min; p = 0.42), workloads (115 vs. 93 W; p = 0.13) and ventilatory parameters were similar. The mean and peak gradients at rest and during exercise were not correlated with VO2max.
CONCLUSIONS
Valve replacement by small aortic prosthesis does not seem to be a factor of exercise intolerance as assessed by VO2 max in patients without LVEF dysfunction before surgery.
| ||||||||||||||||||||||||||||
A negative influence of small prosthesis size on exercise capacity has been suggested. This is thought to be due mainly to high transaortic gradients during exercise that are not observed at rest and also to other factors such as a smaller left ventricular cavity (4). After aortic valve replacement, the patients functional level usually improves, but the effective prosthetic valve area is smaller than that of healthy native valves. Few studies have described the exercise tolerance after valve implantation. Indeed, the determinants of exercise capacity after mechanical aortic replacement are not well known. Tatineni et al. (5) showed that, as expressed by exercise duration, prosthesis size is an independent predictor of exercise tolerance.
Maximal oxygen uptake (VO2max) is a good index of exercise capacity and is used for prognosis in cardiac disease (6). To our knowledge, the evaluation of exercise capacity as assessed by VO2max, compared with that of control subjects, has not been studied in patients with small aortic prosthesis (
21 mm). Hirooka et al. (7) compared the percentage of predicted peak oxygen consumption in relationship to four valve sizes (19 to 25 mm) without statistically significant results. In their study, predicted peak oxygen consumption tended to be lower, however, in patients with a prosthesis size of 19 mm than in those with a size of 25 mm. This suggests that patients with small aortic prostheses present an exercise intolerance. The mechanisms may involve a residual outflow obstruction at rest that is increased during exercise (2); however, we do not know whether hemodynamic characteristics of these small prostheses were a determinant of exercise tolerance. Hemodynamic evaluation of prosthesis function by Doppler echocardiography is frequently performed at rest (8) and less often during exercise (9). Several studies have found strong correlations between the pressure gradient across the prosthetic valve at rest and during exercise (4,5,9).
Consequently, we do not know whether patients with a small aortic valve prosthesis have a real exercise limitation, or if hemodynamic characteristics have a role in this limitation. The aims of this study were 1) to determine exercise tolerance, by VO2max measurement during exhaustive exercise, in patients after aortic valve replacement with small-size prostheses (
21 mm) compared to a paired control group without prosthesis, and 2) to establish whether this capacity is correlated with hemodynamic data collected at rest and during exercise. In addition, we compared the hemodynamic and exercise tolerance capacity characteristics of patients with two types of small mechanical prosthesis in aortic location: St. Jude Medical (SJM) and Medtronic Hall (MH).
| Methods |
|---|
|
|
|---|
21 mm) were implanted. Two types of valve were used: Medtronic Hall (no. 21, n = 6) and St. Jude Medical (no. 19, n = 7; no. 21, n = 1). After surgery, the patients were prospectively studied with a mean follow-up of 28 ± 6 months. The mean age of patients was 62.4 ± 6 years, with nine women and five men. Before surgery, 13 patients had aortic stenosis; one had a predominant insufficiency. All were free of significant coronary artery disease: mean left ventricular ejection fraction (LVEF) was 73 ± 8% (using digital substraction angiography), and the mean stenotic aortic gradient was 71 ± 25 mm Hg. All patients were in sinus rhythm before and after surgery and without exercise contraindication.
Control group. The healthy control group, matched for gender, height and weight, was constituted at the same period. It was composed of 14 normal sedentary subjects without history of cardiovascular or pulmonary disease based on the results of clinical, echocardiographic, electrocardiographic (ECG), and spirographic examination. All subjects gave signed consent to participate in the study after the design and risks of protocol were explained to them. The study was also approved by the hospital ethics committee.
Spirographic measurements. Conventional spirographic measurements were performed on a digital spirometer (Pulmonet III, SensorMedics, Yorba Linda, California). The lung function study included force expiratory volume in 1 s (FEV1) and vital capacity (VC). The FEV1/VC ratio was then calculated. The predicted values were those of Quanjer (10).
Doppler echocardiography. A complete echocardiographic and Doppler examination was performed at rest in the sitting position with SONOS 1500 Hewlett Packard using a 2.5 duplex mechanical transducer. The usual data were collected using M-mode, two-dimensional and continuous-wave Doppler.
Maximal velocities across the prosthesis were measured by continuous-wave Doppler from the apical view. Indeed, the high-velocity jets were best recorded from the cardiac apex, correlated with cardiac catheterization findings (11). From the highest velocities obtained, the transprosthetic pressure drop was calculated according to the Bernoulli equation (12). Mean gradients were obtained by integrating the Doppler velocity signals (13). The effective prosthesis area was calculated with the continuity equation, using the pulsed Doppler (13,14). Doppler measurements were performed at rest and early after peak exercise, during the first minute of recovery, using the same location of the Doppler sampler, in the sitting position.
Cardiorespiratory exercise testing. Exercise tests were performed on an electrically braked cycle ergometer with heart rate, blood pressure and ECG monitoring under the supervision of a physician. The subjects breathed through a low-resistance valve (2700 Hans Rudolf, Kansas City, Kansas) with the nose clamped. Expired flow and oxygen and carbon dioxide partial pressures were continuously monitored on a breath-by-breath basis using a cardiopulmonary exercise test (CPX Medical Graphics System, Medical Graphics, St. Paul, Minnesota). Ventilation (VE, liter/min BTPS), O2 uptake (VO2 ml/min and ml/kg/min STPD), and CO2 production (VCO2, liter/min STPD) were calculated by averaging the breath-by-breath data over 30 s. We also determined the respiratory gas exchange ratio (R = VCO2/VO2) and the ventilatory equivalents for O2 (VE/VO2) and CO2 (VE/VCO2). The ventilatory threshold (Vt) was determined as the level of VO2 at which an increase in VE/VO2 was observed without a simultaneous change in VE/VCO2 (15). The maximal minute ventilation at peak exercise (VEmax) values were predicted by VEmax = measured FEV1 x 35 (16). The ventilatory reserve (VR) at maximal exercise was calculated as follows: (100 [100 x measured VEmax/predicted VEmax]) (17).
The exercise test consisted of a 3-min rest period, followed by the maximal exercise test, which started with a 3-min warmup at 20 W with a minimum constant pedalling rate of 60 rpm. For the patients, the workload was then increased by 10 W every minute. For the healthy control subjects, the workload increment (20 or 30 W) was chosen to obtain an exercise duration between 8 and 12 min, which is the optimal range for obtaining the highest values of peak VO2 (18). Subjects were asked to continue exercise until exhaustion (VO2 peak). The criteria for prematurely stopping the test were defined as follows: serious cardiac arrhythmia, fall in blood pressure, ECG changes compared to resting value, and chest pain.
For every subject, the observation of at least three of the four following criteria was necessary to consider that VO2 peak was obtained: stability of heart rate (HR) at a value close to the predicted maximal HR; stability of oxygen uptake despite the increase in workload; respiratory ratio >1.10; and the inability of the subject to maintain a pedalling rate of 50 rpm (17).
Protocol. The study protocol included spirometric values at rest and maximal cardiopulmonary exercise testing for both groups. The echocardiography Doppler measurements at rest and early after peak exercise, during the first minute of recovery, were performed for the aortic prosthesis group using the same location of Doppler sample, in the sitting position and by the same physician.
Statistical analysis. All data are expressed as mean ± standard deviation (SD). The Student t test for unpaired observations was used for between-group comparison when the normality distribution (Kolmogorov-Smirnov test) and the equality of variance (Levene median test) were verified. When these conditions of normality were not obtained, a Mann-Whitney rank sum test was used. The relationships between the gas exchange and hemodynamic data were assessed using linear regression analysis. A p value <0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Methods. All patients in our study had a body surface area of less than 1.7 m2, for which the usual recommendation is implantation of a small aortic prosthesis (3). The evaluation of exercise tolerance by VO2max measurement in patients with a small aortic prosthesis was easily obtained without adverse complication. In addition, a noninvasive hemodynamic study by Doppler echocardiography was also possible at rest and early after peak exercise in all patients. However, we did not compare the Doppler echocardiographic data with reference standard catheterization data. Indeed, in the smaller 19- and 21-mm sizes, the difference between Doppler and catheter gradients may be more than 20 mm Hg, particularly at high flow rates (19). In previous studies where close agreement between catheter and Doppler data was shown, the subgroup of small valve sizes was not well represented (19,20). Nevertheless, from an ethical point of view, performance of this invasive evaluation did not appear to be sufficiently justified in our study.
Exercise tolerance. In terms of VO2 peak, we found a preserved exercise tolerance; no statistically significant difference existed in the level of maximal exercise between the two groups when this parameter was expressed as %predicted VO2max or maximal workload. Only maximal HR was slightly higher in the control group, probably linked to a nonsignificantly higher maximal workload (115 ± 39 vs. 93 ± 36; p = 0.13; NS). Despite a mild restrictive respiratory syndrome, the patients had a normal breathing pattern, with no differences between the two groups for VEmax or ventilatory equivalents. The ventilatory threshold occurred slightly sooner in the patients but was not significantly different, which suggested there was no muscular impairment. The use of a small aortic prosthesis did not seem to induce an exercise intolerance.
The hemodynamic performance and exercise capacity did not differ between the patients with SJM and those with MH. In the subgroups, with two different diameters (19 and 21 mm) and two types of prostheses, the exercise tolerance did not significantly differ.
It would of course be helpful to compare these exercise tolerance measurements with preoperative data, but in this study the main etiology for which aortic valve replacement was indicated was a predominant aortic stenosis (13/14 patients). Therefore, we did not conduct cardiorespiratory testing before surgery because it is classically contraindicated in this pathology.
All patients presented a preserved left ventricular systolic function before surgery. Respiratory functional exploration performed two years after surgery found impaired respiratory function with a mild restrictive syndrome, probably associated with the thoracic surgery. The spirometric abnormalities we observed have already been described at the two-month follow-up after cardiac surgery (21).
Doppler hemodynamic data and VO2max. The VO2max was not correlated with the hemodynamic parameters (i.e., the transprosthetic gradients at rest and early after peak exercise) and the functional surface of the valve. Many previous studies have used cardiac catheterization for hemodynamic valuation, but this procedure introduced additional risks to patients, especially during exercise. The use of noninvasive Doppler echocardiography during exercise was an excellent means of studying the hemodynamic modifications.
Similar to the data reported by Wiseth et al. (4), the gradients during exercise were correlated with the resting gradients. In addition, rest and exercise prosthetic gradients did not differ between the two valve types or between the two sizes. This was true despite the mechanical differences, with a bi-leaflet tilting-disk design in SJM prostheses and a single tilting-disk design, with a less homogeneous velocity distribution across the valve orifice, in MH prostheses (19).
The transaortic Doppler exercise measurement was performed early after peak exercise, in the first minute of active recovery. This was not really the "peak exercise," but hemodynamic conditions are very similar between peak and early active recovery of a maximal exercise test (22). Indeed, although HR and cardiac output rapidly decrease, the stroke volume is maintained or increased due to an increment of venous return (23).
Comparison with previous studies. Very few studies have specifically investigated the parameters influencing exercise capacity and exercise hemodynamics in patients after aortic valve replacement, especially with a small-size prosthesis. Tatineni et al. (5) showed in patients with an aortic prosthesis that the ejection fraction and other parameters of left ventricular function were not predictive of exercise capacity. Only age and valvular size were independent predictors of exercise tolerance, which might be explained by the higher gradients in small prostheses (2). But in their study, the exercise tolerance was evaluated by exercise duration, and the correlation with prosthesis size was weak (r = 0.41; p < 0.05). Moreover, these data were not compared with those of a control group, and during these invasive studies the ability to complete exercise may have been limited in the patients, which would explain the moderate increase in gradient. Nitter-Hauge et al. (24) showed, in 10 patients with MH of usual size (21 to 27 mm), a mean gradient of 2.9 mm Hg at rest and increasing to 6.8 mm Hg during exercise.
Using Doppler echocardiography for a subgroup of 14 patients with MH 21 mm, Wiseth et al. (4) reported mean and peak gradients, at rest in supine position, of, respectively, 15 ± 4 and 30 ± 8, increased to 24 ± 6 and 47 ± 11 mm Hg during exercise. Aris et al. (25) reported, for 12 patients with SJM 19 mm, a mean resting gradient of 22 ± 7 mm Hg, increased during supine exercise to 32 ± 10 mm Hg, and for 16 patients with MH 20 mm, a mean resting gradient of 17 ± 5 mm Hg, increased to 24 ± 8 mm Hg.
In another study by Tatieni et al. (5), 12 patients had prosthesis sizes
21 mm (data extracted from figure), and their peak exercise gradients ranged from 22 to 52 mm Hg. For Wiseth et al. (4), the range was from 25 to 67 mm Hg, and in our study it was from 25 to 103 mm Hg. Thus, our data show mean gradients at rest and during exercise that are comparable to those of Aris et al. (25) (21 ± 8 vs. 22 ± 7 and 33 ± 14 vs. 32 ± 10 mm Hg, respectively), but the peak exercise gradients were higher in our study probably because one patient had values near 100 mm Hg at exercise.
More recently, Kadir et al. (26) studied the hemodynamic performance in patients with small-size SJM, using dobutamine echocardiography. They observed an increase in mean transvalvular gradient from 22 ± 4.9 mm Hg at rest to 41.9 ± 9 mm Hg at maximum stress (respectively, 21 ± 8 mm Hg and 33 ± 14 in our study).
Concerning the comparison between the exercise capacity in patients with SJM and MH, our finding was similar to that observed in the prospective randomized trial of Fiore et al. (27): their study could not detect a difference in clinical performance or dobutamine stress transesophageal Doppler echocardiography between SJM and MH.
If we summarized these data, we may conclude that our method of measuring the transvalvular gradient in sitting position, at rest, and immediately after exercise was reliable and comparable to other methods (dobutamine stress).
Conclusions. In patients without preliminary ventricular dysfunction who presented a small body surface (<1.7 m2), the implantation of small aortic prostheses did not provoke an alteration in exercise tolerance compared with controls. This demonstrates that the high values of gradient observed in small-sized prostheses at rest and during exercise, in comparison with those of medium sizes, are not correlated with exercise tolerance. Thus, the residual outflow obstruction seems to have no functional consequence.
Finally, in cases of small aortic root, the use of SJM or MH valve prostheses with diameters
21 mm did not appear to be a factor of execise intolerance. This was so despite higher rest and exercise gradients than those observed in "medium"-sized prostheses.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
21 mm) aortic valve prostheses assessed by Doppler echocardiography. Am Heart J. 1993;125:138146[CrossRef][Medline]
This article has been cited by other articles:
![]() |
J.-M. Frapier, C. Sportouch, V. Rauzy, P. Rouviere, S. Cade, R. G. Demaria, J.-M. Davy, and B. Albat Mitral valve repair by Alfieri's technique does not limit exercise tolerance more than Carpentier's correction. Eur. J. Cardiothorac. Surg., June 1, 2006; 29(6): 1020 - 1025. [Abstract] [Full Text] [PDF] |
||||
![]() |