| Abstract |
|---|
|
|
|---|
The goal of the present study was to develop a strategy for three-dimensional (3D) volume acquisition along the major axes of the coronary arteries.
BACKGROUND
For high-resolution 3D free-breathing coronary magnetic resonance angiography (MRA), coverage of the coronary artery tree may be limited due to excessive measurement times associated with large volume acquisitions. Planning the 3D volume along the major axis of the coronary vessels may help to overcome such limitations.
METHODS
Fifteen healthy adult volunteers and seven patients with X-ray angiographically confirmed coronary artery disease underwent free-breathing navigator-gated and corrected 3D coronary MRA. For an accurate volume targeting of the high resolution scans, a three-point planscan software tool was applied.
RESULTS
The average length of contiguously visualized left main and left anterior descending coronary artery was 81.8 ± 13.9 mm in the healthy volunteers and 76.2 ± 16.5 mm in the patients (p = NS). For the right coronary artery, a total length of 111.7 ± 27.7 mm was found in the healthy volunteers and 79.3 ± 4.6 mm in the patients (p = NS). Comparing coronary MRA and X-ray angiography, a good agreement of anatomy and pathology was found in the patients.
CONCLUSIONS
Double-oblique submillimeter free-breathing coronary MRA allows depiction of extensive parts of the native coronary arteries. The results obtained in patients suggest that the method has the potential to be applied in broader prospective multicenter studies where coronary MRA is compared with X-ray angiography.
| ||||||||||||||||||||||||||||||||||
By expeditiously planning the 3D volume along the major axis of the coronary vessels (9,10), coronary coverage may be more optimal. Such a targeted approach allows acquisition of smaller 3D volumes which nevertheless covers the major extent of the coronary system and which depicts the arteries themselves with high in-plane spatial resolution.
In the present study, we present a novel approach for defining the plane of the relevant coronary artery anatomy using a fast, 3D free-breathing navigator sequence for scout scanning. In conjunction with a three-point planscan software tool, the double-oblique imaging plane oriented in parallel to the major coronary arteries is then defined. The technique has been applied to 15 volunteers and seven patients with X-ray angiographically defined coronary artery disease.
| Methods |
|---|
|
|
|---|
Protocol. All participants were examined supine with electrocardiographic (ECG) leads on the anterior left hemithorax. All scans were performed during free breathing. For coronary artery localization and for navigator positioning at the right hemidiaphragm (RHD), two scout scans were utilized. The first scout was an ECG triggered, free-breathing, multislice 2D segmented gradient echo (repetition time [TR] = 11 ms, echo time [TE] = 2.4 ms, 256 x 128 matrix, 450 mm field of view) scan with nine transverse, nine coronal and nine sagittal interleaved acquisitions of the thorax. The total duration of this acquisition was 1 min. On these images, the navigator position at the RHD dome as well as the localized transverse 3D volume for the subsequent scout scan were planned.
Three-dimensional turbo field echo-echo planar imaging (TFE-EPI) scout scan
For the second scout, the MR data were acquired in diastole and at end-expiration using navigator gating and real-time motion correction (11,12) about a volume which included the coronary arteries as defined in the prior scout (Fig. 1). One TFE-EPI shot (13) included five radiofrequency excitations, each followed by 11 EPI readouts. One shot was applied per cardiac cycle, resulting in an acquisition window of <70 ms (TR = 8.8 ms, TE = 5.3 ms). A transverse 3D slab covering a volume of 30 mm thickness was scanned with an in-plane resolution of 1.3 x 1.8 mm. A total of 30 overlapping slices with a reconstructed slice thickness of 3 mm were acquired. For contrast enhancement between blood and myocardial muscle, a T2 preparation (T2-Prepulse) (14,15) preceded the navigator and the imaging portion of the sequence (Fig. 2).
|
|
|
The 3D segmented k-space TFE sequence (TE = 2.4 ms, TR = 8.8 ms) used a modular structure as displayed in Figure 2. The flow-insensitive T2-Prepulse (14,15) for contrast enhancement is followed by a localized anterior saturation prepulse (Fig. 2, REST), the navigator, the spectrally selective fat saturation pulse and finally the TFE imaging sequence (Fig. 2, HR-Scan). The utilized receiver bandwidth was 135 Hz/pixel. A double-oblique 3D volume was imaged as prescribed by the three-point planscan tool. To compensate for suboptimal slice profiles of the slice selective excitations, a 3.9-cm thick volume was imaged with subsequent oversampling (in slice selection direction) by a factor of 1.3. The 20 center slices with a slice thickness of 1.5 mm each were subsequently reconstructed rejecting the six most peripheral slices. A field of view of 360 mm and a 512 x 360 matrix yielded an in-plane resolution of 0.7 x 1.0 mm. Eight phase encoding steps were sampled during each cardiac cycle (acquisition window = 70 ms). K-space was sampled using centric ordering with priority for the low ky-profiles. One signal average was performed, and no flow-compensating gradients were used. Data acquisition was performed in middiastole.
Navigator technique
A vertical 2D selective real-time navigator for gating and motion tracking (11,12,16) was applied at the dome of the RHD (Fig. 1). A navigator excitation angle of 60° and a diameter of 20 mm were used. The gating window for the diaphragm was 5 mm, and a correction factor of 0.6 was used to relate the superiorinferior position of the diaphragm to the superiorinferior position of the proximal coronary arteries (17). For real-time motion correction, the frequency of the slice-selective radiofrequency excitation as well as mixing phase and frequency were adapted based on the detected position (within the gating window) of the lungdiaphragm interface.
Evaluation.
Signal-to-noise and contrast-to-noise ratios
At the level of the LM and the proximal RCA, the SNR was quantified in the ascending aorta (16). Hereby, the average signal intensity (IBlood) in a user-defined region of interest (ROI) was divided by the standard deviation (SDEVBlood) of the signal in the same ROI. The SNR of muscle,
![]() | (1) |
![]() | (2) |
Length measurements
For 3D visualization and length measurements of the coronary arteries, the image data sets were transferred to a commercially available EasyVision workstation (Philips Medical Systems, Best, The Netherlands). The three orthogonal sections through the 3D data set are displayed, and the user navigates interactively through the entire data set. The LM, LAD and RCA were manually identified in all the three displayed planes. The 3D pathway of the coronary artery was then multiplanar reformatted, and the lengths of the individual segments of the native coronary arteries was determined manually.
Scan efficiency
During the high resolution coronary MRA scans, the scan efficiency was recorded in all individuals. Scan efficiency (%) was defined as the number of shots accepted divided by the total number of heart beats used for completing a scan. A scan efficiency of 50%, for example, refers to a scan in which on average, data acquired from every other heartbeat are accepted for reconstruction.
Statistics
All values are presented as mean ± SD. Comparisons between healthy volunteers and patients were made using an unpaired Student t test. In all cases, a two-tailed test was performed with p values
0.05 considered significant.
| Results |
|---|
|
|
|---|
Plane orientations. For the RCA (Fig. 4), a double-oblique coronal orientation in parallel to the atrioventricular groove with a leftright angulation of 25.1 ± 9.6° (range 4 to 40°) and a foothead angulation of 32.6 ± 10.2° (range 7 to 54°) was found in the healthy volunteers using the three-point planscan tool. For patients, these two angles were 4.3 ± 11.2° (p = 0.002, range 8° to +27°) and 34.3 ± 10.3° (p = NS, range 21 to 56°), respectively. For the left coronary system, a transverse orientation with an anterioposterior angulation of 12.5 ± 10.0° (range 0 to 38°) and a leftright angulation of 2.5 ± 5.7° (range 13° to +17°) was found in the healthy volunteers. The corresponding angles for the left coronary system in patients were 5.2 ± 3.5° (p = NS, range 0 to 10°) and 5.0 ± 4.0° (p = NS, range 0 to 17°).
|
|
|
|
|
|
|
Signal-to-noise and contrast-to-noise ratios. In the healthy volunteers, an SNR of 12.3 ± 2.6 was measured on the images showing the RCA, whereas 11.9 ± 2.7 was found in the patients at the same location (p = NS). For the double-oblique images showing the left coronary system, a SNR of 10.2 ± 1.0 was found in the volunteers and 9.0 ± 1.2 in the patients (p = NS). The signal of myocardium is markedly reduced when compared with blood. This could be observed in the patients (CNR left: 5.9 ± 1.4, CNR right: 8.0 ± 1.2) as well as in the volunteers (CNR left: 6.2 ± 1.6, CNR right: 9.7 ± 2.0) (p = NS).
Scan efficiency. Combined (left and right coronary system), scan efficiency was 60% in healthy volunteers and 43% in patients (p < 0.01). No significant difference between the left and right coronary system scan efficiency was found between patients or healthy volunteers (p = NS). Average scanning time per double-oblique high resolution 3D scan was less than 16 min.
| Discussion |
|---|
|
|
|---|
Due to the application of the T2-Prepulse, the bloodmuscle contrast does not primarily depend on inflow, therefore allowing for acquisition of imaging planes that are nonorthogonal to the blood flow. Major portions of the coronary arteries including branches are parallel to the imaging plane and thus displayed with the high in-plane resolution. Lower through-plane resolution minimally affects the reconstruction and thus, locally reduced resolution in the reconstructed images due to anisotropic voxel size may be avoided. No increase of the imaged volume and thus no increase in scanning time were needed for enhanced coverage of the coronary artery tree. With a purely transverse 3D volume orientation, an increase of more than 100% in scanning time would be required to cover the same extent of the right coronary system as found in the present study. For the left coronary system, a 20% increase in scanning time would be required utilizing a transverse 3D slab.
Coordinate system of the native coronary system. The advantages of the targeted double-oblique acquisition strategy result in part from the fact that the left and right coronary arteries lie predominantly in the approximate planes of the intraventricular septum and of the atrioventricular groove, respectively. Thus, these anatomically defined planes constitute a natural coordinate system for coronary MRA. Our data suggest, however, that the particular orientations of the optimal imaging planes are patient specific. As a consequence, scout-scanning techniques such as the three-point plan are required for coronary MRA to maximize the patient-specific length of the vessels oriented in-plane.
Scout scanning. For a successful application of the three-point planscan software tool, a sufficient image quality is required for the localization of the double-oblique 3D volume. By the application of fast 3D imaging techniques such as TFE-EPI for scout scanning, the relevant portions of the coronary arteries can be identified very efficiently (2 min scan duration during free breathing) and reliably.
Three-dimensional acquisition. Three-dimensional techniques allow for the acquisition of thin and contiguous sections. Motion artifacts during the rather long scanning time may degrade the quality of the slice profiles of such thin sections. Such effects may be minimized by the application of prospective adaptive volume tracking (11). Although the slice profile of the 3D slab may be compromised with respect to 2D techniques, the excitation of a thicker slab with subsequent oversampling in slice selection direction minimizes the adverse effect on image quality if the most peripheral slices are rejected during reconstruction. This approach, however, is associated with a prolongation of measurement time. For the present study, where no dynamic information is assessed, the duty cycle is also relatively low, because the 3D data are acquired in only a small fraction of the RR interval. However, this also allows taking maximum advantage of the potential SNR improvement associated with 3D acquisition techniques.
Patient data. By a visual comparison of the X-ray angiograms and the coronary MR angiograms, anatomy and localization of the stenoses could be successfully confirmed on the MR angiograms. Anatomic as well as pathologic details are confirmed in the MR images with an objective, quantitative description of the pathology remaining to be defined. The accuracy of this methodology for identification of coronary stenoses needs to be investigated in large multicenter clinical studies.
In the patients, the average visible length of the RCA was lower than in healthy volunteers. This might be explained by the total occlusion of the proximal RCA (three cases) and a 90% stenosis of the mid-RCA (two cases) which was reported from the X-ray angiograms.
Conclusions. Double-oblique submillimeter free-breathing 3D coronary MRA is a robust technique that has been successfully applied in healthy volunteers and in patients with X-ray angiographically documented coronary disease. The native left and right coronary systems can be displayed with a good contrast, and the 3D reconstructed vessels are seen with high in-plane resolution. Disadvantageous effects of anisotropic voxel sizes in the reconstructed images are suppressed using the double-oblique approach, and the coverage of the coronary artery tree may be enhanced without an increase in scanning time.
The orientation of the left and right coronary system is patient specific. Therefore, accurate localization strategies such as the three-point planscan technique in conjunction with the navigator-gated and motion-corrected TFE-EPI scout scanning are crucial to take full advantage of double-oblique imaging.
The good agreement between the information obtained on the X-ray angiograms and coronary MRA suggest that the currently proposed free-breathing double-oblique technique has the potential to be applied in broader multicenter prospective patient studies, in which coronary MRA is compared with X-ray angiography.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. A. Bluemke, S. Achenbach, M. Budoff, T. C. Gerber, B. Gersh, L. D. Hillis, W. G. Hundley, W. J. Manning, B. F. Printz, M. Stuber, et al. Noninvasive Coronary Artery Imaging: Magnetic Resonance Angiography and Multidetector Computed Tomography Angiography: A Scientific Statement From the American Heart Association Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention, and the Councils on Clinical Cardiology and Cardiovascular Disease in the Young Circulation, July 29, 2008; 118(5): 586 - 606. [Full Text] [PDF] |
||||
![]() |
A. M. Gharib, V. B. Ho, D. R. Rosing, D. A. Herzka, M. Stuber, A. E. Arai, and R. I. Pettigrew Coronary Artery Anomalies and Variants: Technical Feasibility of Assessment with Coronary MR Angiography at 3 T Radiology, April 1, 2008; 247(1): 220 - 227. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Saam, T. S. Hatsukami, N. Takaya, B. Chu, H. Underhill, W. S. Kerwin, J. Cai, M. S. Ferguson, and C. Yuan The Vulnerable, or High-Risk, Atherosclerotic Plaque: Noninvasive MR Imaging for Characterization and Assessment Radiology, July 1, 2007; 244(1): 64 - 77. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Takemura, A. Suzuki, R. Inaba, T. Sonobe, K. Tsuchiya, M. Omuro, and T. Korenaga Utility of Coronary MR Angiography in Children with Kawasaki Disease Am. J. Roentgenol., June 1, 2007; 188(6): W534 - W539. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Ustun, M. Desai, K. Z. Abd-Elmoniem, M. Schar, and M. Stuber Automated Identification of Minimal Myocardial Motion for Improved Image Quality on MR Angiography at 3 T Am. J. Roentgenol., March 1, 2007; 188(3): W283 - W290. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. J. Botman, H. Post, O. Penn, and N. Pijls Value of Magnetic Resonance Imaging, Angiography, and Fractional Flow Reserve to Evaluate the Left Main Coronary Artery After Direct Surgical Angioplasty Ann. Thorac. Surg., February 1, 2007; 83(2): 490 - 494. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Y. Kim, A. S. Astrup, M. Stuber, L. Tarnow, E. Falk, R. M. Botnar, C. Simonsen, L. Pietraszek, P. R. Hansen, W. J. Manning, et al. Subclinical Coronary and Aortic Atherosclerosis Detected by Magnetic Resonance Imaging in Type 1 Diabetes With and Without Diabetic Nephropathy Circulation, January 16, 2007; 115(2): 228 - 235. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. M. Bansmann, A. N. Priest, K. Muellerleile, A. Stork, G. K. Lund, M. G. Kaul, and G. Adam MRI of the Coronary Vessel Wall at 3 T: Comparison of Radial and Cartesian k-Space Sampling Am. J. Roentgenol., January 1, 2007; 188(1): 70 - 74. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. J. McMahon, J. T. Su, M. D. Taylor, R. Krishnamurthy, R. Muthupillai, J. P. Kovalchin, T. Chung, and G. W. Vick III Detection of Active Coronary Arterial Vasculitis Using Magnetic Resonance Imaging in Kawasaki Disease Circulation, November 8, 2005; 112(19): e315 - e316. [Full Text] [PDF] |
||||
![]() |
M. Y. Desai, S. Lai, C. Barmet, R. G. Weiss, and M. Stuber Reproducibility of 3D free-breathing magnetic resonance coronary vessel wall imaging Eur. Heart J., November 1, 2005; 26(21): 2320 - 2324. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Sakuma, Y. Ichikawa, N. Suzawa, T. Hirano, K. Makino, N. Koyama, M. Van Cauteren, and K. Takeda Assessment of Coronary Arteries with Total Study Time of Less than 30 Minutes by Using Whole-Heart Coronary MR Angiography Radiology, October 1, 2005; 237(1): 316 - 321. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Katoh, M. Stuber, A. Buecker, R. W. Gunther, and E. Spuentrup Spin-labeling Coronary MR Angiography with Steady-State Free Precession and Radial k-Space Sampling: Initial Results in Healthy Volunteers Radiology, September 1, 2005; 236(3): 1047 - 1052. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ozgun, A. Hoffmeier, M. Kouwenhoven, R. M. Botnar, M. Stuber, H. H. Scheld, W. J. Manning, W. Heindel, and D. Maintz Comparison of 3D Segmented Gradient-Echo and Steady-State Free Precession Coronary MRI Sequences in Patients with Coronary Artery Disease Am. J. Roentgenol., July 1, 2005; 185(1): 103 - 109. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Sommer, M. Hackenbroch, U. Hofer, A. Schmiedel, W. A. Willinek, S. Flacke, J. Gieseke, F. Traber, R. Fimmers, H. Litt, et al. Coronary MR Angiography at 3.0 T versus That at 1.5 T: Initial Results in Patients Suspected of Having Coronary Artery Disease Radiology, March 1, 2005; 234(3): 718 - 725. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. L. Gerber, E. Coche, A. Pasquet, E. Ketelslegers, D. Vancraeynest, C. Grandin, B. E. Van Beers, and J.-L. J. Vanoverschelde Coronary Artery Stenosis: Direct Comparison of Four-Section Multi-Detector Row CT and 3D Navigator MR Imaging for Detection--Initial Results Radiology, January 1, 2005; 234(1): 98 - 108. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Plein, J. P. Greenwood, J. P. Ridgway, G. Cranny, S. G. Ball, and M. U. Sivananthan Assessment of non-ST-segment elevation acute coronary syndromes with cardiac magnetic resonance imaging J. Am. Coll. Cardiol., December 7, 2004; 44(11): 2173 - 2181. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. U. Herborn, M. Schmidt, O. Bruder, E. Nagel, K. Shamsi, and J. Barkhausen MR Coronary Angiography with SH L 643 A: Initial Experience in Patients with Coronary Artery Disease Radiology, November 1, 2004; 233(2): 567 - 573. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Botnar, A. Buecker, A. J. Wiethoff, E. C. Parsons Jr, M. Katoh, G. Katsimaglis, R. M. Weisskoff, R. B. Lauffer, P. B. Graham, R. W. Gunther, et al. In Vivo Magnetic Resonance Imaging of Coronary Thrombosis Using a Fibrin-Binding Molecular Magnetic Resonance Contrast Agent Circulation, September 14, 2004; 110(11): 1463 - 1466. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. R. Edelman Contrast-enhanced MR Imaging of the Heart: Overview of the Literature Radiology, September 1, 2004; 232(3): 653 - 668. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Jahnke, I. Paetsch, B. Schnackenburg, A. Bornstedt, R. Gebker, E. Fleck, and E. Nagel Coronary MR Angiography with Steady-State Free Precession: Individually Adapted Breath-hold Technique versus Free-breathing Technique Radiology, September 1, 2004; 232(3): 669 - 676. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Spuentrup, M. Katoh, A. Buecker, W. J. Manning, T. Schaeffter, T.-H. Nguyen, H. P. Kuhl, M. Stuber, R. M. Botnar, and R. W. Gunther Free-breathing 3D Steady-State Free Precession Coronary MR Angiography with Radial k-Space Sampling: Comparison with Cartesian k-Space Sampling and Cartesian Gradient-Echo Coronary MR Angiography--Pilot Study Radiology, May 1, 2004; 231(2): 581 - 586. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Mavrogeni, G. Papadopoulos, M. Douskou, S. Kaklis, I. Seimenis, P. Baras, P. Nikolaidou, C. Bakoula, E. Karanasios, A. Manginas, et al. Magnetic resonance angiography isequivalent to X-Ray coronary angiography for the evaluation of coronary arteries in kawasaki disease J. Am. Coll. Cardiol., February 18, 2004; 43(4): 649 - 652. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Maintz, F. C. Aepfelbacher, K. V. Kissinger, R. M. Botnar, P. G. Danias, W. Heindel, W. J. Manning, and M. Stuber Coronary MR Angiography: Comparison of Quantitative and Qualitative Data from Four Techniques Am. J. Roentgenol., February 1, 2004; 182(2): 515 - 521. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Budoff, S. Achenbach, and A. Duerinckx Clinical utility of computed tomography and magnetic resonance techniques for noninvasive coronary angiography J. Am. Coll. Cardiol., December 3, 2003; 42(11): 1867 - 1878. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. P. Salm, J. J. Bax, H. J. Lamb, M. G. Hazekamp, A. de Roos, E. E. van der Wall, and H. W. Vliegen Evaluation of rerouting surgery of a coronary artery anomaly by magnetic resonance angiography Ann. Thorac. Surg., November 1, 2003; 76(5): 1748 - 1748. [Full Text] [PDF] |
||||
![]() |
C. U. Herborn, J. Barkhausen, I. Paetsch, P. Hunold, M. Mahler, K. Shamsi, and E. Nagel Coronary Arteries: Contrast-enhanced MR Imaging with SH L 643A--Experience in 12 Volunteers Radiology, October 1, 2003; 229(1): 217 - 223. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Plein, S. Bulugahapitiya, T. R. Jones, G. J. Bainbridge, J. P. Ridgway, and M. U. Sivananthan Cardiac MR Imaging with External Respirator: Synchronizing Cardiac and Respiratory Motion—Feasibility Study Radiology, June 1, 2003; 227(3): 877 - 882. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Bogaert, R. Kuzo, S. Dymarkowski, R. Beckers, J. Piessens, and F. E. Rademakers Coronary Artery Imaging with Real-time Navigator Three-dimensional Turbo-Field-Echo MR Coronary Angiography: Initial Experience Radiology, March 1, 2003; 226(3): 707 - 716. [Abstract] [Full Text] [PDF] |
||||
![]() |
|