|
|
||||||||||
|
J Am Coll Cardiol, 1999; 33:1735-1742 © 1999 by the American College of Cardiology Foundation |
a Department of Physiology, Cardiovascular Research Institute Maastricht, University of Maastricht, Maastricht, The Netherlands
* Department of Biomedical Engineering, Johns Hopkins University Medical School, Baltimore, Maryland, USA
Manuscript received April 24, 1998; revised manuscript received December 15, 1998, accepted January 20, 1999.
Reprint requests and correspondence: Dr. Frits W. Prinzen, Department of Physiology, Cardiovascular Research Institute Maastricht, University of Maastricht, P.O. Box 616, 6200 MD Maastricht, The Netherlands
frits.prinzen{at}fys.unimaas.nl
| Abstract |
|---|
|
|
|---|
The purpose of this study was to determine the spatial distribution of myocardial function (myofiber shortening and work) within the left ventricular (LV) wall during ventricular pacing.
BACKGROUND
Asynchronous electrical activation, as induced by ventricular pacing, causes various abnormalities in LV function, perfusion and structure. These derangements may be caused by abnormalities in regional contraction patterns. However, insight into these patterns during pacing is as yet limited.
METHODS
In seven anesthetized dogs, high spatial and temporal resolution magnetic resonancetagged images were acquired in three orthogonal planes. Three-dimensional deformation data and LV cavity pressure and volume were used to determine midwall circumferential strain and external and total mechanical work at 192 sites around the left ventricle.
RESULTS
During ventricular pacing, systolic fiber strain and external work were approximately zero in regions near the pacing site, and gradually increased to more than twice the normal value in the most remote regions. Total mechanical work, normalized to the value during right atrial pacing, was 38 ± 13% (right ventricular apex [RVapex] pacing) and 61 ± 23% (left ventricular base [LVbase] pacing) close to the pacing site, and 125 ± 48% and 171 ± 60% in remote regions, respectively (p < 0.05 between RVapex and LVbase pacing). The number of regions with reduced work was significantly larger during RVapex than during LVbase pacing. This was associated with a reduction of global LV pump function during RVapex pacing.
CONCLUSIONS
Ventricular pacing causes a threefold difference in myofiber work within the LV wall. This difference appears large enough to regard local myocardial function as an important determinant for abnormalities in perfusion, metabolism, structure and pump function during asynchronous electrical activation. Pacing at sites that cause more synchronous activation may limit the occurrence of such derangements.
| ||||||||||||||||||||||||||
Better knowledge of the effect of asynchronous activation on regional ventricular function is not only relevant to understanding the acute derangements mentioned above, but also because asynchronous activation appears to have potentially deleterious effects over time. In patients, it has been shown that perfusion deficits and wall motion abnormalities, as induced by pacing, become more pronounced with increasing duration of pacing, and this is accompanied by impaired LV pump function (12). Ventricular pacing and LBBB also lead to structural changes, such as asymmetric hypertrophy (13,14) and, in immature dogs, fiber disarray (15).
The present study was designed to obtain detailed insight into the effects of ventricular pacing on local myocardial function. This was obtained by determining deformation of the entire LV wall at high spatial resolution with use of the magnetic resonance imaging (MRI) tagging technique. This technique uses markers (tags), magnetically induced into the myocardium at end-diastole, that persist during systole. Myocardial motion patterns were translated to myofiber work using data on LV midwall deformation, LV cavity pressure and LV cavity to wall volume ratio using a previously validated approach (8). Asynchronous activation was induced by pacing at two ventricular sites (RVapex and left ventricular base [LVbase]).
| Methods |
|---|
|
|
|---|
In the MRI, scanner-specialized filters were used to eliminate the voltages induced in the pacing leads by the radiofrequency (RF) and magnetic field gradient pulses (16). Through these filters the pacing leads were connected to the pacing and monitoring system outside the scanning room. Electrograms and LV pressure signals were recorded on a four-channel recorder (Gould, Cleveland, Ohio) and A/D converted at 200 Hz with 10 bits. Left ventricular cavity volume was calculated from the endocardial contours in the end-diastolic short-axis MRI and from the deformation of the endocardial layers (see the following section).
MRI. The imaging procedure has been described in more detail previously (16). During stopped ventilation, MRI tagging images were acquired in a 1.5 T scanner (Signa, General Electric Medical Systems, Milwaukee, Wisconsin) with a segmented k-space acquisition. A 6-ms RF gradient tagging pulse sequence was used to produce parallel planes of saturation bands (Fig. 1). The tag pattern was induced at end-diastole, and a total of 12 to 16 images were acquired at 19 to 21-ms intervals, covering the entire systolic phase. Each measurement consisted of six to eight short axis sections with tags at 0°, the same set of sections with tags at 90°, and a set of nine radially oriented long-axis sections (Fig. 1) at 20° intervals around the LV long axis (16).
|
Finally, the dogs were sacrificed by IV injection of an overdose of pentobarbital. The heart was removed from the chest, and the atria and ventricles were weighed. The LV wall volume was calculated as LV weight/1.05 to correct for the specific mass of the myocardium.
Strain analysis. The analysis of strains from the images has been described in detail elsewhere (16). Briefly, the tag lines and heart contours in the images were identified off-line using a semiautomated algorithm. In each set of sections, displacement in the direction normal to the undeformed tag planes was obtained along the taglines. Three-dimensional (3D) deformation was calculated from the three independent sets of one-dimensional displacement data (a total of 2,000 to 3,000 points) (17). Myocardial deformation was calculated in a 24 (circumferential) x 8 (longitudinal) x 3 (radial) mesh of material points. A material point is an infinitesimal volume element of the myocardium, which moves and deforms with the surrounding tissue. The mesh was defined in the first image, and the material points were subsequently tracked in 3D space using the information on the displacement of the tags (17).
In each region Lagrangian strain (fractional change in length as compared with the length at end-diastole, when taglines were exactly parallel) was calculated. Although MRI tagging enables analysis of complete 3D deformation and pacing leads to changes in 3D deformation (18), we focused on circumferential strain in the midwall as derived from the complete 3D motion analysis. Muscle fibers in the midwall run within 20 degrees of the circumference (19). Therefore, in the midwall, circumferential strain closely approximates strain along the fibers. We were interested in myofiber strain, rather than strain in other directions, because in the myocardium work is generated in the direction of the muscle fibers. Fiber strain in other layers than the midwall is harder to determine because fiber angle is more variable and deformation data are not always complete in these layers.
Calculation of local fiber stress and work.
According to the LaPlace equation and more recently developed mathematical models, wall stress in the LV wall is a function of cavity pressure and the ratio of cavity diameter and wall thickness (20). A robust approximation for fiber stress (
) in a thick-walled, anisotropic structure, which yields more physiologic values than most other models, is obtained using the ratio of LV cavity and wall volume (Vc/Vw) (20):
![]() | (1) |
is inversely proportional to the cross-sectional area of the muscle fiber, and thus proportional to fiber length. Using this property local
can be calculated using (8):
![]() | (2) |
![]() | (3) |
Local mechanical work was determined using the local fiber stressfiber length relations (8), in analogy to the pressure-volume area (PVA) concept for the ventricle as a whole (22). The PVA represents total mechanical energy per heart beat and is equal to the sum of external work and potential energy. The PVA is closely related to myocardial oxygen consumption, even during heart beats with negative external work (22). Analogously, regional external work, potential energy and total work were calculated from regional stresslength diagrams (Fig. 2). Potential energy was calculated assuming that no stress is developed at estimated zero cavity volume (8). In epicardial regions of the LV free wall, thus calculated total work was highly correlated with local oxygen consumption when pacing from various sites (8).
|
| Results |
|---|
|
|
|---|
Figure 3 presents midwall circumferential strain as a function of time in one experiment. During RA pacing (upper row), differences in time course and extent of shortening between the various regions were relatively small. During ventricular pacing close to the pacing site (asterisk in Fig. 3), a rapid onset of shortening (negative strain) during the early systolic phase was followed by a rebound stretch and a second phase of shortening. In regions remote from the pacing site (left and right panels of lower row, and middle panels of middle row of Fig. 3), considerable early systolic stretch was followed by pronounced shortening during the ejection phase. Strain patterns gradually changed from the pacing site to the most remote regions.
|
|
|
Figure 6 displays polar plot maps of external work during the three pacing modes in one experiment. A fairly homogeneous distribution was found during RA pacing, but large variations were present during ventricular pacing. External work ranged from negative at the pacing site to supranormal values in remote regions. Note that the area with reduced external work was larger during RVapex pacing than during LVbase pacing.
|
|
| Discussion |
|---|
|
|
|---|
Abnormal contraction patterns. The effect of abnormal electrical activation on LV deformation appears to depend on the spatial resolution used. At the level of the whole ventricle, the principal directions and magnitudes of deformation (23) and systolic torsion (24) were hardly affected by ventricular pacing. In contrast, the present and previous studies show dramatic changes in local deformation, characterized by reciprocal contraction patterns in opposing sites of the LV wall (6). The contraction patterns change gradually from earlier to later activated regions (7,8,25). With use of the MRI tagging technique and additional hemodynamic measurements, the present study demonstrates that ventricular pacing reduces mechanical work in the septum during RVapex pacing by 50% and increases it by 50% in the LV free wall. Previous studies showed that during ventricular pacing blood flow varied in parallel with regional function, presumably due to local metabolic regulation of myocardial blood flow (7,8). Accordingly, a good correlation was found between local mechanical work and oxygen consumption during ventricular pacing (8). Together with the present findings, this indicates that during RVapex pacing septal oxygen consumption is reduced by half.
The regionally different contraction patterns are presumably caused by regional differences in fiber length, induced by the asynchronous activation. Early activated regions start to shorten early and vigorously (by up to 10%) because their afterload is still very low. This forces later activated regions on the opposite LV wall to be prestretched (by up to 15%, Figs. 3 and 4). Consequently, when all regions have been activated and LV pressure rises, late activated fibers are longer than early activated fibers. By virtue of a local "Frank-Starling" relation, the later activated regions are stronger and shorten more during the ejection phase. Therefore, the regional differences in contraction pattern during ventricular pacing can be regarded as differences in effective local preload.
Possible implications of the abnormal contraction patterns. Because regional work is a good predictor of regional myocardial oxygen demand (8), the abnormal contraction patterns occurring during RVapex pacing and LBBB most likely reduce metabolic demands in the septum. This may explain the reduced septal perfusion and 18-fluoro-deoxyglucose uptake frequently observed in LBBB (4,5). At higher heart rates, regional perfusion abnormalities are amplified, which has been attributed to impaired perfusion due to the abnormal contraction patterns (11,26).
Although adaptation of the heart as a whole to increased workload is well known, adaptations to locally different workload, induced by asynchronous activation, also become evident. Asymmetrical hypertrophy occurs in LBBB patients (13) and in dogs with chronic ventricular pacing at physiologic heart rate (14). In the LBBB patients and in the paced dogs, the late activated regions, which are the regions shown to have higher workload in the present study, were hypertrophied most pronouncedly. In immature dog hearts, pacing induced fiber disarray (15). In patients, septal perfusion and wall motion deteriorates with increasing duration of pacing. Such changes are associated with deterioration of ventricular pump function (12). Strategies to decrease the asynchrony of activation, like pacing high in the ventricular septum, ameliorated the reduction in pump function (27) and prevented the development of fiber disarray (28). The present study indicates that MRI tagging may be a powerful tool to study the effects of new, potentially more physiologic pacing regimes on regional cardiac mechanics.
Relation between local contraction patterns and ventricular function. In the present study systolic LV pressure and stroke volume significantly decreased during RVapex pacing, but not during LVbase pacing. Other studies have also shown a larger depression of ventricular function during pacing at the RVapex than at other sites (23,29,30). The exact cause of the dependency of ventricular function on the site of pacing is not clear. Grover and Glantz (23) reported a smaller end-diastolic volume, indicating disturbed filling during RVapex pacing. However, in the present study, end-diastolic pressure and volume were similar during the three modes of pacing. Burkhoff et al. (30) found evidence that contractile failure is due to ineffective contractions in the early activated regions, thus leading to a "loss of effective muscle mass." The present data support the loss of contractile function in early activated regions but also demonstrate hyperfunctioning late activated regions. Because ventricular pump function was not impaired during LVbase pacing, hyperfunction in the late activated regions has presumably compensated the loss of function in the early activated regions. However, during RVapex pacing the size of the region with reduced function appears to be too large to be compensated. Because a direct relation exists between the moment of local electrical activation and systolic fiber shortening (25), the large septal hypofunctional area during RVapex pacing suggests that large clusters of regions in the septum are activated almost simultaneously. This may imply that RVapex pacing couples into the right Purkinje system but is only slowly conducted transseptally. Instead, pacing from the epicardium of the LVbase does not couple into the Purkinje system, or much later.
Evaluation of experimental approach. The present study highlights the ability of MRI tagging to reveal regional differences in myofiber shortening and work during abnormal electrical activation. Moreover, it is completely noninvasive and, therefore, directly applicable to patients. Pacing and measurement of local electrograms in the MRI scanner are also possible when using specialized RF filters. Development of such a system for humans is required before human studies can be performed.
In the present study, atrio-ventricular (A-V) synchronous pacing was used with an A-V delay of 0 ms to ensure activation of the entire ventricle by the ectopic stimulus and to prevent intermittent forward conduction through the A-V node. The latter could have induced arrhythmias, which spoil MRI quality. Although a short A-V delay may itself have deteriorated ventricular function (29), in the present study reduced ventricular pump function was only found during RVapex pacing and not during LVbase pacing.
The calculations of regional myocardial work used in this study are based on measured fiber strains and estimations of fiber stress, enabling noninvasive estimation of local work at multiple sites. The approach to calculate work has been validated previously (8). Although myofiber orientation in the midwall may not be exactly circumferential, values of midwall circumferential strain were virtually equal to values of epicardial fiber strain, as measured in previous studies using optical markers. Because, at least during a normal heart beat, fiber strain is uniform throughout the wall (31), this indicates that in the midwall circumferential strain is a fair estimate of fiber strain. Moreover, qualitatively the effects of pacing are the same for midwall strain in the circumferential and longitudinal direction, so calculating work with use of strain in another direction would only cause quantitative, but not qualitative, differences. Another validation of this approach is that global external work was estimated to be 280 mJ per beat by multiplying stroke volume and LV pressure (see Results). This value correlates favorably with a regionally based estimate of myocardial work, obtained by integrating average local work (2.8 mJ/g, middle panel of Fig. 7) over the entire LV (average weight 94 g): 263 mJ per beat. Moreover, the decrement in global systolic work seen from hemodynamic measurements during RVapex pacing versus LVbase pacing also correlated with the decrement of regional work integrated over the entire ventricle (Fig. 7). The mechanical analysis in this study is limited to the midwall. However, comparable patterns of midwall mechanics in the present study and epicardial mechanics in a previous study (8) indicate that transmural differences are small compared with the regional differences induced by ventricular pacing.
Conclusions. Pacing at the LVbase and at the RVapex leads to local functional abnormalities throughout the LV wall, ranging from virtually absent systolic fiber shortening and external work at the site of pacing, to values twice above normal in regions most remote from the pacing site. The reduction in systolic shortening and external work as well as the size of the hypofunctioning zone around the pacing site were significantly larger during RVapex pacing than during LVbase pacing. These local changes may be responsible for the depression in ventricular function during RVapex pacing.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
Z. B. Popovic and J. D. Thomas In Search of a Holy Grail: Predicting Cardiac Resynchronization Therapy Outcomes by Echocardiography Circ Cardiovasc Imaging, July 1, 2008; 1(1): 3 - 5. [Full Text] [PDF] |
||||
![]() |
C. Miyazaki, G. Lin, B. D. Powell, R. E. Espinosa, C. J. Bruce, F. A. Miller Jr, B. L. Karon, R. F. Rea, D. L. Hayes, and J. K. Oh Strain Dyssynchrony Index Correlates With Improvement in Left Ventricular Volume After Cardiac Resynchronization Therapy Better Than Tissue Velocity Dyssynchrony Indexes Circ Cardiovasc Imaging, July 1, 2008; 1(1): 14 - 22. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. W. Prinzen and A. Auricchio Is echocardiographic assessment of dyssynchrony useful to select candidates for cardiac resynchronization therapy?: Echocardiography Is Not Useful Before Cardiac Resynchronization Therapy if QRS Duration Is Available Circ Cardiovasc Imaging, July 1, 2008; 1(1): 70 - 78. [Full Text] [PDF] |
||||
![]() |
A. Auricchio and F. W. Prinzen Update on the pathophysiological basics of cardiac resynchronization therapy Europace, July 1, 2008; 10(7): 797 - 800. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. O. Sweeney and F. W. Prinzen Ventricular Pump Function and Pacing: Physiological and Clinical Integration Circ Arrhythmia Electrophysiol, June 1, 2008; 1(2): 127 - 139. [Full Text] [PDF] |
||||
![]() |
E. S. Chung, A. R. Leon, L. Tavazzi, J.-P. Sun, P. Nihoyannopoulos, J. Merlino, W. T. Abraham, S. Ghio, C. Leclercq, J. J. Bax, et al. Results of the Predictors of Response to CRT (PROSPECT) Trial Circulation, May 20, 2008; 117(20): 2608 - 2616. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Miyazaki, B. D. Powell, C. J. Bruce, R. E. Espinosa, M. M. Redfield, F. A. Miller, D. L. Hayes, Y.-M. Cha, and J. K. Oh Comparison of Echocardiographic Dyssynchrony Assessment by Tissue Velocity and Strain Imaging in Subjects With or Without Systolic Dysfunction and With or Without Left Bundle-Branch Block Circulation, May 20, 2008; 117(20): 2617 - 2625. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Kypta, C. Steinwender, J. Kammler, F. Leisch, and R. Hofmann Long-term outcomes in patients with atrioventricular block undergoing septal ventricular lead implantation compared with standard apical pacing Europace, May 1, 2008; 10(5): 574 - 579. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Zanon, E. Bacchiega, L. Rampin, S. Aggio, E. Baracca, G. Pastore, T. Marotta, G. Corbucci, L. Roncon, D. Rubello, et al. Direct His bundle pacing preserves coronary perfusion compared with right ventricular apical pacing: a prospective, cross-over mid-term study Europace, May 1, 2008; 10(5): 580 - 587. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Chakir, S. K. Daya, R. S. Tunin, R. H. Helm, M. J. Byrne, V. L. Dimaano, A. C. Lardo, T. P. Abraham, G. F. Tomaselli, and D. A. Kass Reversal of Global Apoptosis and Regional Stress Kinase Activation by Cardiac Resynchronization Circulation, March 18, 2008; 117(11): 1369 - 1377. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z B Popovic, R A Grimm, A Ahmad, D Agler, M Favia, G Dan, P Lim, F Casas, N L Greenberg, and J D Thomas Longitudinal rotation: an unrecognised motion pattern in patients with dilated cardiomyopathy Heart, March 1, 2008; 94(3): e11 - e11. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Pirat, D. S. Khoury, C. J. Hartley, L. Tiller, L. Rao, D. G. Schulz, S. F. Nagueh, and W. A. Zoghbi A novel feature-tracking echocardiographic method for the quantitation of regional myocardial function: validation in an animal model of ischemia-reperfusion. J. Am. Coll. Cardiol., February 12, 2008; 51(6): 651 - 659. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Bordachar, L. Labrousse, J.-B. Thambo, P. Reant, S. Lafitte, M. D. O'Neill, P. Jais, M. Haissaguerre, J. Clementy, and P. Dos Santos Haemodynamic impact of the left ventricular pacing site during graded ischaemia in an open-chest pig model Europace, February 1, 2008; 10(2): 242 - 248. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E. Albertsen, J. C. Nielsen, S. H. Poulsen, P. T. Mortensen, A. K. Pedersen, P. S. Hansen, H. K. Jensen, and H. Egeblad DDD(R)-pacing, but not AAI(R)-pacing induces left ventricular desynchronization in patients with sick sinus syndrome: tissue-Doppler and 3D echocardiographic evaluation in a randomized controlled comparison Europace, February 1, 2008; 10(2): 127 - 133. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Pu and W. T Abraham Resynchronisation therapy in heart failure: searching for predictors Heart, January 1, 2008; 94(1): 4 - 5. [Full Text] [PDF] |
||||
![]() |
K. M. Kavanagh, I. Belenkie, and H. J. Duff Transmural temporospatial left ventricular activation during pacing from different sites: potential implications for optimal pacing Cardiovasc Res, January 1, 2008; 77(1): 81 - 88. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. W. Kimmel, N. D. Skadsberg, C. L. Byrd, D. J. Wright, T. G. Laske, and P. A. Iaizzo Single-site ventricular and biventricular pacing: investigation of latest depolarization strategy Europace, December 1, 2007; 9(12): 1163 - 1170. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Lionetti, L. Guiducci, A. Simioniuc, G. D. Aquaro, C. Simi, D. De Marchi, S. Burchielli, L. Pratali, M. Piacenti, M. Lombardi, et al. Mismatch between uniform increase in cardiac glucose uptake and regional contractile dysfunction in pacing-induced heart failure Am J Physiol Heart Circ Physiol, November 1, 2007; 293(5): H2747 - H2756. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Byrne, R. H. Helm, S. Daya, N. F. Osman, H. R. Halperin, R. D. Berger, D. A. Kass, and A. C. Lardo Diminished Left Ventricular Dyssynchrony and Impact of Resynchronization in Failing Hearts With Right Versus Left Bundle Branch Block J. Am. Coll. Cardiol., October 9, 2007; 50(15): 1484 - 1490. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. F. Tops, M. S. Suffoletto, G. B. Bleeker, E. Boersma, E. E. van der Wall, J. Gorcsan III, M. J. Schalij, and J. J. Bax Speckle-Tracking Radial Strain Reveals Left Ventricular Dyssynchrony in Patients With Permanent Right Ventricular Pacing J. Am. Coll. Cardiol., September 18, 2007; 50(12): 1180 - 1188. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. C. Shuros, R. W. Salo, V. G. Florea, J. Pastore, M. A. Kuskowski, Y. Chandrashekhar, and I. S. Anand Ventricular Preexcitation Modulates Strain and Attenuates Cardiac Remodeling in a Swine Model of Myocardial Infarction Circulation, September 4, 2007; 116(10): 1162 - 1169. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Arita, G. P. Sorescu, B. T. Schuler, L. S. Schmarkey, J. D. Merlino, J. Vinten-Johansen, A. R. Leon, R. P. Martin, and D. Sorescu Speckle-tracking strain echocardiography for detecting cardiac dyssynchrony in a canine model of dyssynchrony and heart failure Am J Physiol Heart Circ Physiol, July 1, 2007; 293(1): H735 - H742. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. A. Coppola, J. W. Covell, A. D. McCulloch, and J. H. Omens Asynchrony of ventricular activation affects magnitude and timing of fiber stretch in late-activated regions of the canine heart Am J Physiol Heart Circ Physiol, July 1, 2007; 293(1): H754 - H761. [Abstract] [Full Text] [PDF] |
||||