| Abstract |
|---|
|
|
|---|
Background. The ICD is very effective in terminating ventricular tachycardia (VT) or ventricular fibrillation (VF), but protection against SD is not absolute. Little is known about the nature and potential causes of SD in patients with ICDs.
Methods. We analyzed 25 cases of out-of-hospital SD among patients enrolled in the clinical investigation of the Cadence Tiered-Therapy Defibrillator System.
Results. All patients (24 men and 1 woman, mean age 62 ± 10 years) received epicardial lead systems. The majority (92%) had coronary artery disease and a previous myocardial infarction (MI), with a mean left ventricular ejection fraction 0.25 ± 0.07. At device implantation, the mean defibrillation threshold was 13 ± 5 J. Sudden death occurred 13 ± 11 months later. Twenty patients (80%) had received appropriate ICD therapies before death, and 18 (72%) were receiving
1 antiarrhythmic drugs at the time of death. Sudden death was tachyarrhythmia-associated in 16 patients (64%), non-tachyarrhythmia-associated in 7 (28%) and indeterminate in 2 (8%). In the 16 patients with tachyarrhythmia-associated SD, the overall first therapy success rate in tachycardia and fibrillation zones was 60% and 67%, respectively. However, despite protracted therapies (
2 shocks) in 7 (66%) of 12 patients who received fibrillation therapies, the final tachyarrhythmic episode was ultimately terminated by the ICD in 15 (94%) of the 16 patients, whereas 1 patient died after multiple (initially successful) internal and external shocks for intractable VT/VF during exercise. In 10 patients (40%) one or more, primarily clinical, factors potentially contributory to SD were identified: heart failure (n = 8), angina (n = 2), hypokalemia (n = 1), adverse antiarrhythmic drug treatment (n = 1) and acute MI (n = 1). An additional 10 patients (40%) had experienced an increase in frequency of ICD shocks within 3 months of SD. Appropriate battery voltages and normal circuitry function were found in all devices interrogated and analyzed after death.
Conclusions. In this select group of patients receiving a third-generation ICD, SD was associated with VT or VF events in nearly two-thirds of patients, and death occurred despite ultimately successful, although often protracted, device therapies. These observations, along with evidence of recent worsening clinical status, suggest acute cardiac mechanical dysfunction as a frequent terminal factor. In recipients with ICDs, SD directly attributable to device failure seems to be rare.
| ||||||||||||||||||||||
The event storage capability of the more recent generation of ICDs, including intracardiac electrograms (6), offers the possibility of better characterization of terminal arrhythmias associated with SD (7). At the same time, various technologic features of the newer devices, such as biphasic shock (which provides for lower defibrillation energy [8]) and antibradycardia pacing, might have a positive impact on patient outcome, whereas enhanced programmability, with the potential for suboptimal device responses (9), might have adverse consequences.
The objectives of this study were twofold: first, to determine the arrhythmic events associated with SD in recipients of a third-generation ICD and, second, to identify clinical and possible device-related factors that might have contributed to the patients death, and whether or not such factors, if any, might be potentially preventable.
| Methods |
|---|
|
|
|---|
Device system and implantation.
The Cadence ICD is a multiprogrammable device capable of backup bradycardia pacing and "noncommitted" antitachycardia pacing, cardioversion and defibrillation with biphasic shocks (12). Detailed information on the detected tachyarrhythmic events, including each delivered therapy and its duration and outcome, is available for each of the most recent 11 episodes. The device stores intracardiac electrograms corresponding to the most recent (
3) tachyarrhythmic events, along with the time and date of the events. The devices extended high rate (EHR) detection feature incorporates a "timer" that allows antitachycardia therapies to be attempted for only a programmable length of time (10 to 120 s); if the timer expires, the device proceeds to the more aggressive fibrillation zone therapies (12).
Written informed consent for implantation of this device (investigational at the time) was obtained from all patients after approval by the Institutional Review Boards of each of the participating centers (see Appendix). All 25 patients received epicardial lead systems consisting of two defibrillating patches and epicardial (n = 14) or endocardial (n = 11) rate-sensing leads. Four patients had concomitant myocardial revascularization.
Patient follow-up and data collection. A complete evaluation of device function, including VT and VF therapy, was performed before and 6 weeks after hospital discharge, and per the discretion of the individual electrophysiologist depending on clinical circumstances (e.g., addition or change in antiarrhythmic drug therapy). Thereafter, patients were seen at a minimum of once every 3 months, or sooner in cases of device discharges. At each visit, the device was interrogated and all diagnostic-therapy data, including intracardiac electrograms, stored since the previous visit were recorded. Retrieved data obtained from postmortem device evaluation were analyzed with emphasis on the number of detected and treated tachyarrhythmic episodes and the outcome of each therapy. These and any other pertinent data were reviewed after being submitted to Ventritex from each of the participating centers.
Death classification. Sudden death was defined as death occurring within 1 h of symptoms or during sleep or unwitnessed in a previously medically stable patient (13), and was further classified on the basis of the presence or absence of tachyarrhythmias at the time of death: 1) Tachyarrhythmia-associated: In witnessed patients, documented ICD shocks or VT/VF during resuscitation, plus new tachyarrhythmic events (from ICD data log) at the time of death; or, in unwitnessed patients, evidence of delivered ICD therapy on the day of death. Stored intracardiac electrograms of specific tachyarrhythmias were classified according to published criteria (6). Tachyarrhythmia cycle length was determined on the basis of an average of 10 beats immediately preceding a delivered or aborted therapy. 2) Non-tachyarrhythmia-associated: Death in the absence of stored, new tachyarrhythmic events in unwitnessed patients, or documented asystole or EMD not immediately preceded by antitachyarrhythmic ICD therapy in witnessed, monitored patients. 3) Indeterminate: Insufficient information regarding either postmortem device data or witnessed shocks or documented tachyarrhythmia at the time of death.
Statistical analysis. Numeric values are expressed as the mean value ± SD or percentage. Comparisons were made using the Student t test, with p values <0.05 considered significant.
| Results |
|---|
|
|
|---|
|
1 postimplantation assessment of device function showing appropriate detection and termination of induced VF and, when applicable, VT. Twenty-one patients (84%) had a history of
1 appropriate device therapy for either VT or VF, or both. Eighteen patients (72%) were receiving
1 antiarrhythmic drug at the time of death, including amiodarone alone or combined with another agent in five patients. Characterization of terminal events. Death was witnessed in 18 (72%) of the 25 patients. In 13 of the 18 witnessed patients with available relevant data, 8 (62%) had no premonitory symptoms before collapse, and the remaining 5 (38%) complained of chest pain or respiratory distress within minutes of, or immediately preceding, collapse. Eight (44%) of the 18 witnessed patients were reported to have received ICD shocks at the time of collapse; 4 (22%) were observed specifically not to have received shocks; and in the remaining 6 (34%), information on shock delivery was available. In 20 patients for whom EMS was summoned soon after collapse, the following rhythms were reported (Table 1): VF in 5, VT in 3 and paced/EMD in the remaining 12. All 20 patients received advanced resuscitative measures with an unsuccessful outcome. Six of the seven unwitnessed patients had been seen in stable condition within 1 to 24 h before death.
Postmortem device analysis.
Device interrogation was completed after death in 22 (88%) of 25 patients: 15 (68%) had new tachyarrhythmic events stored at the time (witnessed patients) or on the day of death (unwitnessed patients), and the remaining 7 (32%) had no new stored tachyarrhythmic events. Retrieved electrograms corresponding to the most recent
3 stored tachyarrhythmic events showed VT or VF in 14 patients and "make-break" potentials resulting from cut rate-sensing leads during autopsy in one patient (no. 25). The cycle lengths of stored VT and VF events were 260 to 370 ms and
250 ms, respectively. In each of the 14 patients these final
3 VT or VF events were stored within seconds to minutes apart, consistent with closely timed recurrent VT/VF. Appropriate battery voltages (>5.1 V) were noted in all 22 patients whose devices were interrogated after death, and shock impedance was in the normal range (28 to 60 ohms) in the 8 patients with available data (Table 1). Analysis performed at Ventritex on the 21 returned generators showed no evidence of circuitry malfunction.
Tachyarrhythmia-versus nontachyarrhythmia-associated SD. Based on documentation of VT/VF during attempted resuscitation, witnessed ICD shocks (presumptive VT/VF) or stored ICD data, or a combination, SD was considered tachyarrhythmia-associated in 16 patients, nontachyarrhythmia-associated in 7 and indeterminate in 2 (Fig. 1). Of the two indeterminate cases (both witnessed), information on ICD shocks, EMS documented rhythm or device interrogation was not available in one patient (no. 24); the other patient (no. 25) most likely represented a nontachyarrhythmia-associated death because the patient had received no shocks at the time of collapse.
|
|
40 beats/min), including one patient (no. 19) who had a separate VVI pacemaker with a programmed lower rate of 70 beats/min; and a similar proportion of witnessed patients in each group had no premonitory symptoms before collapse. Results of antitachyarrhythmia therapy in cases of tachyarrhythmia-associated SD. Table 3 summarizes tachyarrhythmia detections and ICD therapies in patients with tachyarrhythmia-associated SD, excluding those patients (nos. 7 and 8) who did not have postmortem ICD interrogation.
|
330 ms) and, in addition, only a single therapy was specified for the tachycardia zone (i.e., the second through fourth tachycardia zone therapies were programmed "OFF" without a high-rate "timer" limit).
Tachyarrhythmia detections and therapy data
Between the time of last ICD interrogation and death (42 ± 29 days [range 5 to 90]), an average of 18 (range 1 to 108) and 11 (range 0 to 80) VT and VF detections, respectively, were recorded, for which an average of 23 (range 0 to 94) shocks were delivered and 3.7 (range 0 to 11) were aborted (for presumed nonsustained tachyarrhythmic episodes). Three patients (nos. 5, 12 and 14) received
2 shocks and one patient (no. 13) received only antitachycardia pacing for seven episodes of tachyarrhythmias.
Success rate of first therapies
Although most tachyarrhythmia therapies were ultimately successful (see later discussion), the first and, in some cases, subsequent therapies did not terminate the tachyarrhythmias. Specifically, of 11 patients who received shocks for tachyarrhythmias in the fibrillation zone, the first therapy was unsuccessful on
1 occasion in 7 (64%), yielding an overall first-therapy success rate of only 67% based on the total number of VF episodes (i.e., 62 successes in 92 episodes). Among the seven patients who received at least two consecutive fibrillation zone shocks on one or more occasion, four (nos. 2, 6, 9 and 11) (Table 3) required all six shocks (the maximal number delivered by the device for any given tachyarrhythmia episode). For tachyarrhythmias in the tachycardia zone, in 6 (75%) of 8 patients initial antitachycardia pacing therapy failed on
1 occasion, representing an overall success rate of 60% (i.e., 30 of 50 episodes).
Outcome of final therapy
Despite delivery of consecutive
2 fibrillation zone therapies in seven patients, all VT/VF episodes were ultimately terminated by the device in 11 (85%) of 13 patients; one patient (no. 10) had only nonsustained (aborted) events. One patient (no. 4) with an ethmozine-induced rise in DFT died during a treadmill test after multiple, initially effective, ICD and, after device inactivation, external shocks for intractable VT/VF. In an additional witnessed patient (no. 5) who received two appropriate ICD shocks (Fig. 2, A and B), the final intracardiac electrogram (Fig. 2C) registered
38 min after the patients collapse, showed inappropriate redetection of "sinus rhythm" in the presence of ongoing polymorphic VT. The two patients (nos. 7 and 8) without postmortem ICD data were both witnessed at the time of collapse and reported by emergency medical staff and emergency room personnel to have received successful ICD shocks for recurrent VF, but eventually developed EMD.
|
Battery depletion or generator component failures were not identified in any of the 22 devices that were analyzed after death. Postshock tachyarrhythmia detection failure (patient no. 5) and inappropriate device programming (patient no. 11) theoretically might have played a role, because both can result in a delay or complete withholding of needed antitachyarrhythmia therapy. However, these potential device-related factors could not be definitively implicated in the causation of SD in any of the cases in our series (Fig. 2).
| Discussion |
|---|
|
|
|---|
Tachyarrhythmic versus nontachyarrhythmic SD. The finding in the present third-generation ICD experience that some two-thirds of SDs were associated with tachyarrhythmia is consistent with observations from earlier devices in which at least one-half of the cases of SDs occurred in association with VT or VF events (24,7).
The proportion of out-of-hospital SDs deemed nontachyarrhythmic (28%) in the present series is quite similar to that reported in patients without ICDs (1416) and in patients with first-generation (nonpacing) ICDs (5), even though the incorporation of antibradycardia pacing in current devices might be expected to prevent, or at least reduce (17), the occurrence of bradyarrhythmic SD reported to occur in patients with nonpacing devices (18,19). This suggests that the mechanism(s) involved in bradyarrhythmic SD is far more complex than in cases of bradyarrhythmias due to conduction system disease, which, in general, are readily amenable to pacing therapy.
Sudden death and device-related factors.
In contrast to earlier reports of SD in patients with first-generation ICDs (35), device-related factors, such as battery-depleted or deactivated ICDs and malfunctioning generators or defibrillation leads, did not appear to play a role in the present third-generation experience. However, relevant findings in two of our study patients bear mention: potentially deleterious, inappropriate device programming and device malsensing. In one patient (no. 11) unusual device programming (described earlier) might have resulted in prolonged, untreated recurrent VT that eventually led to VF, which, although ultimately terminated after multiple shocks (Table 3), may have caused irrevocable hemodynamic insult and death. The second patient (no. 5) had evidence on the final stored intracardiac electrogram of inappropriate redetection of sinus rhythm in the setting of ongoing polymorphic VT (Fig. 2C); however, this event occurred
38 min after conversion of VT (without restoration of pulse) by the ICD (Fig. 2A). Previously, failure to redetect VF has been reported only in connection with laboratory-induced arrhythmias (2023).
Reduced effectiveness of antitachyarrhythmia therapy.
Although the ICD ultimately terminated most of the tachyarrhythmic episodes, the observed 64% first-shock (biphasic waveform) success rate for the final (
11) fibrillation zone tachyarrhythmic episodes was lower than the
93% value reported for similar patients with comparable DFTs (12,24,25), and even below that described with monophasic shocks (26). In addition, we observed an overall antitachycardia pacing success rate of only 60% compared with a reported success rate of 92% to 97% in patients similar to ours (12,2426).
Despite the statistical nature of defibrillation success (27), a purely random basis for all the first-shock therapy failures in this series is unlikely given the low implant DFTs, the high proportion of patients known to have received appropriate and successful ICD therapies (before death) and the one to three normal premortem VF tests in our study group. Conceivably, some first-shock failures might have been due to a subthreshold programmed energy, because in 43% of the patients the programmed first defibrillation shock voltage was only marginally greater, and in some cases equal to or lower than the implant DFT, not in keeping with the recommended "safety margin" (28). Under ordinary conditions, the DFT for an epicardial system is expected to remain stable over time (29). The increased frequency of first-shock failures in our series, therefore, implicates dynamic factors that subacutely or acutely altered adversely the energy requirement for terminating VT or VF, as suggested by premortem worsening CHF in 40% of our patients and preterminal symptoms suggestive of ischemia in one-third of witnessed SDs.
Potential mechanisms of SD patients with ICDs.
Although the ultimate rhythm in patients dying suddenly with (or without) an ICD is expected to be asystole or EMD, such an outcome can arise directly as an initial, pacing-unresponsive bradyarrhythmia, or indirectly as the sequel to one or more VT or VF events. Acute EMD (without antecedent VT or VF) can occur on a primary cardiac (e.g., acute MI) or noncardiac basis (e.g., massive pulmonary embolism). Interpretation of VT or VF in the SD setting is more complex. Clearly, an acute pathophysiologic derangement, such as ischemia, can trigger VT or VF; in some cases, however, even if VT or VF so initiated is promptly detected and treated appropriately by the ICD ("effective" therapy), asystole or EMD can follow (25,18,19). Moreover, rapidly recurrent VT or VF events can, despite successful treatment, lead to eventual mechanical failure either from repeated bouts of myocardial hypoperfusion or from potentially harmful effects of defibrillating shocks on ventricular function in chronically failing myocardium (30,31), as suggested by the markedly reduced baseline LVEF (mean 0.25) in the present patient group compared with the
0.30 to 0.35 range typical of patients in large ICD series (12,13,26).
Submaximal first-shock energies for VF and, more importantly, a lower than 100 V (
10 J) first-shock safety margin, as programmed in 86% and 43%, respectively, of our cases of tachyarrhythmia-associated SD (Table 3) may be sufficient under relatively stable laboratory testing conditions but may be inadequate during acute pathophysiologic changes.
Study limitations. The complexities of clarifying the nature of "terminal" events associated with SD are well known (32,33) and may not be clarified in the case of the ICD recipient dying suddenly (7,33). Despite availability of event markers and time-stamped stored electrograms, there often remain uncertainties about the precise timing of the fatal event. Moreover, autopsies were performed in only three of our patients. A higher autopsy rate might have provided additional insight into the mechanisms of SD in our patient group; for instance, evidence for some acute events would help to explain both the immediate cause of death (in some cases noncardiac [7]) and the reduced effectiveness of initial ICD therapies. In addition, the present series involved epicardial lead systems only and devices from a single manufacturer; however, there are no reasons to suspect device-specific differences in the final event(s), although with the current widespread use of transvenous lead systems, ICD lead failures might play a larger role in the cause of SD in the future (2023,34).
Conclusions. The present study suggests that, by and large, SD in patients with ICD reflects acute myocardial mechanical deterioration, with "incidental" tachyarrhythmias or bradyarrhythmias that, though effectively treated by the device, may still lead to death. However, whether or not the already remarkably low SD rate in device-treated patients can be further reduced ultimately would depend on the mechanism(s) of SD in ICD recipients. The possibility that some cases of SD in patients with ICDs might be preventable merits continued investigation.
| Appendix |
|---|
|
|
|---|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
T. W. Baron, T. S. Faber, A. Grom, T. Schwab, M. Brunner, A. Geibel, H. Just, C. Bode, and M. Zehender Real-time assessment of acute myocardial ischaemia by an intra-thoracic 6-lead ECG: evaluation of a new diagnostic option in the implantable defibrillator Europace, November 1, 2006; 8(11): 994 - 1001. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. V. Huikuri, T. H. Makikallio, M. J. P. Raatikainen, J. Perkiomaki, A. Castellanos, and R. J. Myerburg Prediction of Sudden Cardiac Death: Appraisal of the Studies and Methods Assessing the Risk of Sudden Arrhythmic Death Circulation, July 8, 2003; 108(1): 110 - 115. [Full Text] [PDF] |
||||
![]() |
W. G. Stevenson and L. M. Epstein Predicting Sudden Death Risk for Heart Failure Patients in the Implantable Cardioverter-Defibrillator Age Circulation, February 4, 2003; 107(4): 514 - 516. [Full Text] [PDF] |
||||
![]() |
A. E. Buxton, K. L. Lee, G. E. Hafley, D. G. Wyse, J. D. Fisher, M. H. Lehmann, L. A. Pires, M. R. Gold, D. L. Packer, M. E. Josephson, et al. Relation of Ejection Fraction and Inducible Ventricular Tachycardia to Mode of Death in Patients With Coronary Artery Disease: An Analysis of Patients Enrolled in the Multicenter Unsustained Tachycardia Trial Circulation, November 5, 2002; 106(19): 2466 - 2472. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. B. Mitchell, E. A. Pineda, J. L. Titus, P. M. Bartosch, and D. G. Benditt Sudden death in patients with implantable cardioverter defibrillators: The importance of post-shock electromechanical dissociation J. Am. Coll. Cardiol., April 17, 2002; 39(8): 1323 - 1328. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Gronefeld and S. H Hohnloser What do implantable cardioverter/defibrillators teach us about the mechanisms of sudden cardiac death? Cardiovasc Res, May 1, 2001; 50(2): 232 - 241. [Abstract] [Full Text] [PDF] |
||||
![]() |
R J P LEWIN, D J FRIZELLE, and G C KAYE A rehabilitative approach to patients with internal cardioverter-defibrillators Heart, April 1, 2001; 85(4): 371 - 372. [Full Text] |
||||
![]() |
B. J. Maron, W.-K. Shen, M. S. Link, A. E. Epstein, A. K. Almquist, J. P. Daubert, G. H. Bardy, S. Favale, R. F. Rea, G. Boriani, et al. Efficacy of Implantable Cardioverter-Defibrillators for the Prevention of Sudden Death in Patients with Hypertrophic Cardiomyopathy N. Engl. J. Med., February 10, 2000; 342(6): 365 - 373. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |