J Am Coll Cardiol, 1998; 32:1861-1865 © 1998 by the American College of Cardiology Foundation |
a Department of Internal Medicine (Cardiovascular Division), William Beaumont Hospital, Royal Oak, Michigan, USA
Manuscript received April 10, 1998; revised manuscript received July 31, 1998, accepted August 20, 1998.
Address for correspondence: Dr. Cindy L. Grines, Director of Cardiac Catheterization Laboratories, William Beaumont Hospital, 3601 West Thirteen Mile Road, Royal Oak, Michigan 48073
cgrines{at}beaumont.edu
| Abstract |
|---|
|
|
|---|
Background. Studies have shown that atherosclerotic aortic debris predisposes patients to spontaneous or procedurally related ischemic events.
Methods. In 1,000 consecutive percutaneous interventions, the amount of visible atheromatous material from large-lumen-guiding catheters was recorded. Clinical characteristics and in-hospital complications were prospectively collected and associated with debris production.
Results. Visible aortic debris (1+ to 3+) occurred more frequently with the Judkins left (JL) catheter, followed by the multipurpose (Multi) catheter compared to any other type of guiding catheter (65%, p = 0.001 and 60%, p = 0.01, respectively). Large debris (2+ and 3+) was observed most frequently with the Multi (odds ratio 3.79, C.I. = 2.32 to 6.21, p = 0.001), JL (odds ratio 2.83, C.I. = 1.98 to 4.05, p = 0.001) and voda left (VL) (odds ratio 2.73, C.I. = 1.51 to 4.95, p = 0.001) catheters. The Judkins right (JR) catheter type was least likely to produce any debris (24%, p = 0.001). A history of unstable angina (p = 0.05) or myocardial infarction (p = 0.003) was associated with a decreased incidence of debris production. The presence of debris was not found to be associated with in-hospital ischemic complications.
Conclusions. Studies have shown that atherosclerosis of the aorta is a potential source of systemic embolism in patients undergoing cardiac catheterization. Our study shows that in more than 50% of percutaneous revascularization procedures, guiding catheter placement is associated with scraping debris from the aorta. Design characteristics of the JL, Multi and VL guiding catheters make them most likely to produce such debris. Meticulous attention to allow the debris to exit the back of the catheter is essential to prevent injecting atheromatous debris into the vascular bed.
| ||||||||||||||||||
Atheromatous material within the aorta may be dislodged during invasive aortic procedures (16). Transesophageal echocardiographic recognition of atherosclerotic aortic debris identifies patients at high risk for stroke and peripheral embolism during cardiac catheterization, intraaortic balloon pump placement and cross-clamping of the aorta during coronary bypass surgery (1728). The incidence of clinically apparent embolization and stroke associated with diagnostic cardiac catheterization is less than 1% (29). However, the incidence of catheter-related embolism in necropsy studies has been reported to be as high as 30% (30,31).
The overall complication rates of percutaneous revascularization procedures are higher than diagnostic catheterization (29). In patients undergoing percutaneous interventions, stiff, large-bore-guiding catheters are used. These design characteristics can be more traumatic to the aorta than diagnostic catheters, which are more flexible, have smaller lumens and tapered tips. We noticed that during advancement of guiding catheters, atheromatous debris scraped from the aorta exited the back of the catheter. This was almost never seen during diagnostic catheterization.
This study was designed to determine the incidence and to quantitate aortic debris (atheromatous material that exited the guiding catheter during its advancement into the descending aorta and around the aortic arch) retrieved in 1,000 consecutive patients undergoing percutaneous interventions. We sought to determine whether there were patient or catheter characteristics that predispose to the presence of aortic debris, which, if inadvertently flushed forward, may embolize. Finally, we determined whether the presence of debris was associated with in-hospital recurrent ischemia, myocardial infarction, acute renal failure or ischemic neurologic complications.
| Methods |
|---|
|
|
|---|
|
|
|
|
|
|
|
|
| Results |
|---|
|
|
|---|
A history of unstable angina or previous myocardial infarction was associated with a decreased incidence of any debris production (odds ratio .869 and .810, respectively) (Table 1). When analyzed according to size of debris particles (1+ to 3+), no additional significant association was seen between the presence or size of debris particles and increasing age of the patient. In-hospital recurrent ischemia, myocardial infarction, acute renal failure or ischemic neurologic complications occurred infrequently, and these were not associated with the presence or size of debris particles (1+ to 3+) (Table 2).
| Discussion |
|---|
|
|
|---|
Embolic complications, although infrequent, appear to account for a significant portion of the morbidity and mortality of coronary interventions. They are thought to be the result of clot formation due to prolonged procedures, but they may also be due to dislodgment of atheromatous material from the aorta during catheter and wire manipulation. During percutaneous interventions, large-lumen, stiff-guiding catheters are used. Atheromatous material frequently exits the back of the guiding catheters during their advancement in the descending aorta and around the aortic arch. This study was designed to describe the incidence of aortic debris, to semi-quantitate the amount of aortic debris, to determine whether patient or catheter characteristics have any bearing on debris production, and to determine if the presence of aortic debris is associated with in-hospital complications.
Our data show a high incidence of aortic debris scraped by the guiding catheter (51% of cases overall) but a wide range depending on the shape of the catheter used (24% to 65%). The JL and Multi guiding catheters were associated with a significant increase in overall aortic debris production, particularly large debris, visible at the time of catheter advancement compared to the other guiding catheters evaluated. Voda left catheters also tended to be associated with debris but did not reach statistical significance, likely due to a small sample size. However, when large debris was observed, the voda left catheter (similar to the Multi and JL catheters) was more likely to be responsible. We postulate that the shapes of the JL, Multi and VL catheters are more traumatic to the aorta owing to their long secondary curves, acting as "atherectomy devices" as they pass across friable plaques. The JR catheters were the least likely to lead to debris production.
Interestingly, the patient population at decreased risk for debris production were patients with unstable angina or a history of myocardial infarction. This may be due to a more thrombotic and less atherosclerotic process leading to the acute ischemic syndrome, or lifestyle and medication modification with atherosclerotic regression in the postmyocardial infarction patients.
The presence of aortic debris was not found to be associated with the in-hospital complications of recurrent ischemia, myocardial infarction, acute renal insufficiency or neurologic ischemic complications regardless of the relative size and amount. This may be due to the technique used in this study where all guiding catheters were allowed to bleed back onto a sterile towel after wire removal and prior to attachment of the Y-connector. This technique theoretically could decrease the incidence of embolizing atheromatous material forward into the coronary or cerebral or peripheral circulation. Moreover, the small event rate and innumerable other factors contributing to complications will make the relationship between debris and clinical events difficult to ascertain.
Study limitations. This study has several limitations: 1) all patients included in the study had significant coronary artery disease; therefore, this selection bias may explain why the study failed to demonstrate an association of aortic debris with traditional risk factors for atherosclerosis; 2) the semiquantitative, subjective nature of the "debris score"; 3) none of the patients underwent transesophageal echocardiography at the time of the percutaneous procedure; therefore, it is unknown whether the debris produced by advancement of the catheter within the aorta is representative of the extent of aortic atherosclerosis; and 4) the overall risk of complications during percutaneous interventions is low; consequently, an extremely large sample size would be required to determine whether visible debris correlates with ischemic complications.
Conclusions. Previous studies have shown that atherosclerosis of the aorta is a potential source of systemic embolism, particularly in patients undergoing manipulation of the aorta such as coronary artery bypass surgery and cardiac catheterization. Our study shows that scraping atherosclerotic debris into large lumen-guiding catheters is common and seems to be related to catheter design. Using Judkins left, multipurpose, or voda left guiding catheters during percutaneous revascularization procedures increases the likelihood of dislodging debris from the aorta. Recognition of this phenomenon, and allowing sufficient "back bleeding" from the guiding catheter after wire removal and before attaching the Y-connector, is essential to avoid embolizing atheromatous debris during contrast injections.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Hamon, J.-C. Baron, F. Viader, and M. Hamon Periprocedural Stroke and Cardiac Catheterization Circulation, August 5, 2008; 118(6): 678 - 683. [Full Text] [PDF] |
||||
![]() |
E. J. Heyer, M. I. Gold, E. W. Kirby, J. Zurica, E. Mitchell, H. J. Halazun, L. Teverbaugh, R. R. Sciacca, R. A. Solomon, D. O. Quest, et al. A Study of Cognitive Dysfunction in Patients Having Carotid Endarterectomy Performed with Regional Anesthesia Anesth. Analg., August 1, 2008; 107(2): 636 - 642. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. T. Gutsche, A. T. Cheung, M. L. McGarvey, W. G. Moser, W. Szeto, J. P. Carpenter, R. M. Fairman, A. Pochettino, and J. E. Bavaria Risk Factors for Perioperative Stroke After Thoracic Endovascular Aortic Repair Ann. Thorac. Surg., October 1, 2007; 84(4): 1195 - 1200. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Hamon, S. Gomes, M.-R. Clergeau, S. Fradin, R. Morello, and M. Hamon Risk of Acute Brain Injury Related to Cerebral Microembolism During Cardiac Catheterization Performed by Right Upper Limb Arterial Access Stroke, July 1, 2007; 38(7): 2176 - 2179. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. W. Katzberg and B. J. Barrett Risk of Iodinated Contrast Material-induced Nephropathy with Intravenous Administration Radiology, June 1, 2007; 243(3): 622 - 628. [Full Text] [PDF] |
||||
![]() |
I. Kronzon and P. A. Tunick Aortic Atherosclerotic Disease and Stroke Circulation, July 4, 2006; 114(1): 63 - 75. [Full Text] [PDF] |
||||
![]() |
P. Khatri and S. E. Kasner Ischemic strokes after cardiac catheterization: opportune thrombolysis candidates? Arch Neurol, June 1, 2006; 63(6): 817 - 821. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Presbitero, G. L. Gasparini, and P. Pagnotta Intra-Arterial Thrombolysis for Left Middle Cerebral Artery Embolic Stroke During Coronary Angiography Circulation, February 7, 2006; 113(5): e64 - e66. [Full Text] [PDF] |
||||
![]() |
S. Dukkipati, W. W. O'Neill, K. J. Harjai, W. P. Sanders, D. Deo, J. A. Boura, B. A. Bartholomew, M. W. Yerkey, H. M. Sadeghi, and J. K. Kahn Characteristics of cerebrovascular accidents after percutaneous coronary interventions J. Am. Coll. Cardiol., April 7, 2004; 43(7): 1161 - 1167. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Bittl and L. R. Caplan Stroke after percutaneous coronary interventions J. Am. Coll. Cardiol., April 7, 2004; 43(7): 1168 - 1169. [Full Text] [PDF] |
||||
![]() |
P. Ehlermann, W. Mirau, J. Jahn, A. Remppis, and A. Sheikhzadeh Predictive Value of Inflammatory and Hemostatic Parameters, Atherosclerotic Risk Factors, and Chest X-Ray for Aortic Arch Atheromatosis Stroke, January 1, 2004; 35(1): 34 - 39. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Schwammenthal, Y. Schwammenthal, D. Tanne, A. Tenenbaum, A. Garniek, M. Motro, B. Rabinowitz, M. Eldar, and M. S. Feinberg Transcutaneous detection of aortic arch atheromas by suprasternal harmonic imaging J. Am. Coll. Cardiol., April 3, 2002; 39(7): 1127 - 1132. [Abstract] [Full Text] [PDF] |
||||
![]() |
O S Mortensen, J K Madsen, T Haghfelt, P Grande, K Saunamäki, S Haunsø, E Hjelms, H Arendrup, and t. D. study group Health related quality of life after conservative or invasive treatment of inducible postinfarction ischaemia Heart, November 1, 2000; 84(5): 535 - 540. [Abstract] [Full Text] |
||||
![]() |
P. A. Tunick and I. Kronzon Atheromas of the thoracic aorta: clinical and therapeutic update J. Am. Coll. Cardiol., March 1, 2000; 35(3): 545 - 554. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |