JACC
HOME SUBSCRIPTIONS CURRENT ISSUE PAST ISSUES CARDIOSOURCE SEARCH HELP FEEDBACK
 QUICK SEARCH:   [advanced]


     


J Am Coll Cardiol, 1986; 8:971-974
© 1986 by the American College of Cardiology Foundation
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mann, D.
Right arrow Articles by Weyman, A.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Mann, D.
Right arrow Articles by Weyman, A.

Doppler and two-dimensional echocardiographic diagnosis of Bjork-Shiley prosthetic valve malfunction: importance of interventricular septal motion and the timing of onset of valve flow

DL Mann, LD Gillam, JE Marshall, ME King, and AE Weyman

In a 69 year old woman with a "sticking" Bjork-Shiley mitral prosthesis, the diagnosis was suggested by both the two-dimensional and the Doppler ultrasound examinations. In particular, the findings of early diastolic paradoxic septal motion, intermittent delayed opening of the prosthetic disc and variable timing of the onset of mitral valve inflow were believed to be diagnostic of a sticking tilting disc prosthesis.





HOME SUBSCRIPTIONS CURRENT ISSUE PAST ISSUES CARDIOSOURCE SEARCH HELP FEEDBACK
Copyright © 1986 by the American College of Cardiology Foundation.