Doppler echocardiographic evaluation of Hancock and Bjork-Shiley prosthetic values
KB Sagar,
LS Wann,
WH Paulsen,
and
DW Romhilt
Doppler echocardiographic characteristics of normally functioning Hancock and Bjork-Shiley prostheses in the mitral and aortic positions were studied in 50 patients whose valvular function was considered normal by clinical evaluation. Doppler studies were also performed in 46 patients with suspected malfunction of Hancock and Bjork-Shiley valves and who subsequently underwent cardiac catheterization. Mean gradients were estimated for both mitral and aortic valve prostheses and valve area was calculated for the mitral prostheses. Doppler prosthetic mitral valve gradient and valve area showed good correlation with values obtained with cardiac catheterization (r = 0.93 and 0.97, respectively) for both types of prosthetic valves. The correlation coefficient (r = 0.93) for mean prosthetic aortic valve gradient was also good, although Doppler echocardiography overestimated the mean gradient at lower degrees of obstruction. Regurgitation of Hancock and Bjork-Shiley prostheses in the mitral and aortic positions was correctly diagnosed. These results suggest that Doppler echocardiography is a reliable method for the characterization of normal and abnormal prosthetic valve function.
This article has been cited by other articles:

|
 |

|
 |
 
M. Josa, M. Castella, C. Pare, J. L. Bedini, R. Cartana, C. A. Mestres, J. L. Pomar, and J. Mulet
Hemolysis in Mechanical Bileaflet Prostheses: Experience With the Bicarbon Valve
Ann. Thorac. Surg.,
April 1, 2006;
81(4):
1291 - 1296.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. Becassis, M. Hayot, J.-M. Frapier, F. Leclercq, L. Beck, J.e. Brunet, E. Arnaud, C. Prefaut, P.-A. Chaptal, J. M. Davy, et al.
Postoperative exercise tolerance after aortic valve replacement by small-size prosthesis: Functional consequence of small-size aortic prosthesis
J. Am. Coll. Cardiol.,
September 1, 2000;
36(3):
871 - 877.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J CHAMBERS
Should we ever treat the echo rather than the patient?
Heart,
December 1, 1999;
82(6):
652 - 653.
[Full Text]
|
 |
|

|
 |

|
 |
 
N. Cam, H. Gercekoglu, S. Celik, M. Gursurer, G. Tayyareci, H. Karabulut, A. Narin, T. Tezel, and B. Yigiter
Dobutamine Stress Test to Evaluate Different Sizes of Prosthetic Aortic Valves
Asian Cardiovasc Thorac Ann,
September 1, 1998;
6(3):
166 - 173.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. De Paulis, L. Sommariva, G. M. De Matteis, E. Caprara, F. Tomai, A. P. de Peppo, P. Polisca, C. Bassano, and L. Chiariello
EXTENT AND PATTERN OF REGRESSION OF LEFT VENTRICULAR HYPERTROPHY IN PATIENTS WITH SMALL SIZE CARBOMEDICS AORTIC VALVES
J. Thorac. Cardiovasc. Surg.,
May 1, 1997;
113(5):
901 - 909.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
T. Carrel, U. Zingg, R. Jenni, B. Aeschbacher, and M. I. Turina
Early In Vivo Experience With the Hemodynamic Plus St. Jude Medical Heart Valves in Patients With Narrowed Aortic Annulus
Ann. Thorac. Surg.,
May 1, 1996;
61(5):
1418 - 1422.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
M. B. Izzat, I. Birdi, P. Wilde, A. J. Bryan, and G. D. Angelini
COMPARISON OF HEMODYNAMIC PERFORMANCES OF ST. JUDE MEDICAL AND CARBOMEDICS 21 MM AORTIC PROSTHESES BY MEANS OF DOBUTAMINE STRESS ECHOCARDIOGRAPHY
J. Thorac. Cardiovasc. Surg.,
February 1, 1996;
111(2):
408 - 415.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
R. De Paulis, L. Sommariva, F. Russo, F. Tomai, A. Tondo, C. Pagliaricci, C. Bassano, and L. Chiariello
Doppler echocardiography evaluation of the CarboMedics valve in patients with small aortic anulus and valve prosthesis-body surface area mismatch
J. Thorac. Cardiovasc. Surg.,
July 1, 1994;
108(1):
57 - 62.
[Abstract]
[Full Text]
|
 |
|
|