Objectives: We sought to test how catheter ablation using an open irrigation catheter (OIC) compares with standard catheters for pulmonary vein antrum isolation.
Background: Open irrigation catheters have the advantage of delivering greater power without increasing the temperature of the catheter tip, which enables deeper and wider lesions without the formation of coagulum on catheters.
Methods: Catheter ablation was performed using an 8-mm catheter (8MC) or an OIC. Patients were randomized to 3 groups: 8MC; OIC-1, OIC with a higher peak power (50 W); and OIC-2, OIC with lower peak power (35 W).
Results: A total of 180 patients were randomized to the 3 treatment strategies. Isolation of pulmonary vein antra was achieved in all patients. The freedom from atrial fibrillation was significantly greater in the 8MC and OIC-1 groups compared with the OIC-2 group (78%, 82%, and 68%, respectively, p = 0.043). Fluoroscopy time was lower in OIC-1 compared with OIC-2 and 8MC (28 ± 1 min, 53 ± 2 min, and 46 ± 2 min, respectively, p = 0.001). The mean left atrium instrumentation time was lower in the OIC-1 compared with the OIC-2 and 8MC groups (59 ± 3 min, 90 ± 5 min, and 88 ± 4 min, respectively, p = 0.001). However, there was a greater incidence of "pops" in the OIC-1 (100%, 0%, 0%, p < 0.001) along with higher incidences of pericardial effusion (20%, 0%, 0%, p < 0.001) and gastrointestinal complaints (17% in OIC-1, 3% in 8MC, and 5% in OIC-2, p = 0.031).
Conclusions: Although there was a decrease in fluoroscopy and left atrium instrumentation time with the use of OIC at higher power, this setting was associated with increased cardiovascular and gastrointestinal complications.
| |||||||||||