CLINICAL RESEARCH: SAFETY ISSUE
Fluoroscopically Guided Implantation of Modern Cardiac Resynchronization Devices
Radiation Burden to the Patient and Associated Risks
Kostas Perisinakis, PhD*,*,
Nicholas Theocharopoulos, MSc*, ,
John Damilakis, PhD*,
Emmanouel Manios, MD ,
Panayiotis Vardas, MD, PhD, FESC, FACC and
Nicholas Gourtsoyiannis, MD, PhD
* Department of Medical Physics, Faculty of Medicine, University of Crete, Iraklion, Greece
Department of Cardiology, Faculty of Medicine, University of Crete, Iraklion, Greece
Department of Radiology, Faculty of Medicine, University of Crete, Iraklion, Greece
Department of Natural Sciences, Technical Education Institute of Crete, Iraklion, Greece
Manuscript received August 5, 2004;
revised manuscript received January 13, 2005,
accepted January 17, 2005.
* Reprint requests and correspondence: Dr. Kostas Perisinakis, Medical Physics Department, Faculty of Medicine, University of Crete, P.O. Box 2208, 71003 Iraklion, Crete, Greece (Email: perisina{at}med.uoc.gr).
OBJECTIVES: To establish radiation risks for patients undergoing fluoroscopically guided cardiac resynchronization device implantation.
BACKGROUND: Cardiac resynchonization therapy (CRT) may be associated with extended fluoroscopic exposure.
METHODS: The fluoroscopy time, dose-area product (DAP), exposure parameters, and percentage contribution of the fluoroscopic projections commonly used were recorded in a series of 14 consecutive patients referred for cardiac resynchronization device implantation and compared to corresponding data obtained from a control group of 20 patients who underwent a conventional rhythm device implantation operation. The DAP to peak skin dose, DAP to effective dose, and DAP to gonadal dose conversion factors were determined for biventricular pacing and conventional rhythm device implantation using a humanoid phantom and thermoluminescence dosimetry.
RESULTS: The mean total fluoroscopy time and DAP values were 35.2 min and 4,765 cGy cm2, respectively, for biventricular pacing and 8.2 min and 1,106 cGy cm2, respectively, for conventional rhythm device implantation. Patient skin dose from biventricular pacing procedures requiring extended fluoroscopic exposure may exceed threshold dose for the induction of skin effects only if X-ray source-to-skin distance is kept low. The risk values for fatal cancer and severe hereditary disorders, respectively, associated with a typical CRT procedure were 273 per million and 0.2 per million treated patients.
CONCLUSIONS: Radiation risks associated with fluoroscopically guided CRT procedures may be considerable. Present data may be used for the estimation of patient radiation risks from CRT procedures performed in other institutions.
|
Abbreviations and Acronyms
| | cGy cm2 = centigray x centimeter2 | | CRT = cardiac resynchronization therapy | | DAP = dose-area product | | ICD = implantable cardioverter-defibrillator | | ICRP = International Commission on Radiological Protection | | LAO = left-anterior oblique | | PA = posteroanterior | | RAO = right-anterior oblique | | SSD = source-to-skin distance | | TLD = thermoluminescence dosimeter | | Sv1 = sievert1 | | Gy1 = gray1 |
|
This article has been cited by other articles:

|
 |

|
 |
 
G BEDETTI, N BOTTO, M G ANDREASSI, C TRAINO, E VANO, and E PICANO
Cumulative patient effective dose in cardiology
Br. J. Radiol.,
September 1, 2008;
81(969):
699 - 705.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. Greenberg and M. R. Mehra
All Patients With Heart Failure and Intraventricular Conduction Defect or Dyssynchrony Should Not Receive Cardiac Resynchronization Therapy
Circulation,
December 12, 2006;
114(24):
2685 - 2691.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N Theocharopoulos, J Damilakis, K Perisinakis, E Manios, P Vardas, and N Gourtsoyiannis
Occupational exposure in the electrophysiology laboratory: quantifying and minimizing radiation burden
Br. J. Radiol.,
August 1, 2006;
79(944):
644 - 651.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|