Advertisement







Click here for more guidelines.
CME Topic Collections Past Issues Search Current Issue Home
     

J Am Coll Cardiol, 2005; 45:770-779, doi:10.1016/j.jacc.2004.11.036
© 2005 by the American College of Cardiology Foundation
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (8)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Buck, T.
Right arrow Articles by Levine, R. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Buck, T.
Right arrow Articles by Levine, R. A.

CLINICAL RESEARCH: ECHOCARDIOGRAPHY

Broad-beam spectral Doppler sonification of the vena contracta using matrix-array technology

A new solution for semi-automated quantification of mitral regurgitant flow volume and orifice area

Thomas Buck, MD, FACC*,*, Björn Plicht, MD*, Peter Hunold, MD{dagger}, Ronald A. Mucci, PhD{ddagger}, Raimund Erbel, MD, FACC* and Robert A. Levine, MD, FACC{ddagger}

* West German Heart Center Essen, University Duisburg-Essen, Essen, Germany
{dagger} Department for Diagnostic and Interventional Radiology, University Duisburg-Essen, Essen, Germany
{ddagger} Cardiac Ultrasound Laboratory, Massachusetts General Hospital, Boston, Massachusetts

Manuscript received August 9, 2004; accepted October 6, 2004.

* Reprint requests and correspondence: Dr. Thomas Buck, West German Heart Center Essen, Department of Cardiology, University Duisburg-Essen, Hufelandstrasse 55, 45122 Essen, Germany (Email: thomas.buck{at}uni-essen.de).

OBJECTIVES: The objective of this study was to evaluate broad-beam spectral Doppler sonification of the vena contracta using a matrix-array transducer for quantification of mitral regurgitation (MR).

BACKGROUND: Noninvasive assessment of the severity of valvular regurgitation remains challenging. A recent technique measures regurgitant flow directly at the vena contracta based on the product of velocity times backscattered Doppler power (proportional to orifice area). That approach, however, has been limited by relatively narrow conventional beamwidths. Matrix-array transducers, recently developed for three-dimensional imaging, can potentially provide broader beams. Therefore, we addressed the hypothesis that deliberate broadening of the Doppler beam can encompass larger regurgitant cross-sectional areas to capture a broader range of regurgitant flows.

METHODS: A matrix-array transducer system was modified to provide a three-dimensionally expanded spectral Doppler sample volume. Calculations of orifice area, flow rate, and regurgitant stroke volume (RSV) from Doppler power were automated on board a routinely used echocardiographic scanner and tested in vitro. In 24 patients with isolated MR, RSV was compared with magnetic resonance imaging (MRI) mitral inflow minus aortic outflow from phase-velocity maps.

RESULTS: The calculated flow rate and RSV correlated and agreed well with reference values in vitro (r = 0.98 to 0.99) and in patients (r = 0.93, mean difference 0.4 ± 3.2 ml, p = NS vs. 0), with sufficient sonification to measure flow orifices up to 0.85 cm in diameter. Agreement with MRI was comparable in 17 patients with central and seven with eccentric jets (p = NS vs. 0).

CONCLUSIONS: The broad-beam spectral Doppler technique provides accurate, largely automated quantification of regurgitant flow rate and integrated RSV directly at the lesion. The accuracy related to broader sonification is made possible by the new matrix-array transducer design.

Abbreviations and Acronyms
  CSA = cross-sectional area
  LV = left ventricle/ventricular
  MR = mitral regurgitation
  MRI = magnetic resonance imaging
  PVI = power-velocity integral
  ROA = regurgitant orifice area
  RSV = regurgitant stroke volume
  VC = vena contracta




This article has been cited by other articles:


Home page
Eur J EchocardiogrHome page
C. Li, J. Zhang, X. Li, C. Zhou, H. Li, H. Tang, and L. Rao
Quantification of chronic aortic regurgitation by vector flow mapping: a novel echocardiographic method
Eur J Echocardiogr, November 21, 2009; (2009) jep175v1.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
T. Buck, B. Plicht, P. Kahlert, I. M. Schenk, P. Hunold, and R. Erbel
Effect of Dynamic Flow Rate and Orifice Area on Mitral Regurgitant Stroke Volume Quantification Using the Proximal Isovelocity Surface Area Method
J. Am. Coll. Cardiol., August 26, 2008; 52(9): 767 - 778.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
A. N. DeMaria, O. Ben-Yehuda, D. Berman, G. K. Feld, G. S. Ginsburg, B. H. Greenberg, W. Y.W. Lew, D. Sahn, and S. Tsimikas
Highlights of the Year in JACC 2005
J. Am. Coll. Cardiol., January 3, 2006; 47(1): 184 - 202.
[Full Text] [PDF]



 
  CME Topic Collections Past Issues Search Current Issue Home

Advertisement